Intended Use

To verify the performance of the HMS Plus and Heparin Assay Cartridges. For In Vitro Diagnostic use.

Device Description

The modification to the current control is to replace old USP heparin with revised USP heparin from the new heparin monograph that became effective October 1, 2009. The source of the heparin remains porcine.

AI/ML Overview

The provided 510(k) summary (K103313) for the Medtronic Heparin Assay Controls describes a device modification and claims substantial equivalence to its predicate devices. However, it does not contain the kind of detailed information typically found in studies that establish acceptance criteria and prove device performance in the context of medical AI or complex diagnostic systems.

This document describes a change in the source of heparin used in the control substance, from "old USP heparin" to "revised USP heparin from the new heparin monograph." The purpose of the submission is to demonstrate that this change does not alter the fundamental scientific technology or intended use, and that the modified device remains substantially equivalent to the previously cleared predicate.

Here's an analysis based only on the provided text, highlighting what is missing or not applicable within the context of your request:

Analysis based on the provided K103313 document:

  1. A table of acceptance criteria and the reported device performance:

    • Not provided. The document states "verification testing indicated that the fundamental scientific technology and the intended use are unchanged," but it does not detail specific acceptance criteria (e.g., accuracy, sensitivity, specificity, or specific range of values) nor does it report quantitative performance results from this verification testing.
  2. Sample size used for the test set and the data provenance:

    • Not provided. The document mentions "verification testing" but does not specify the sample size of the test set or the provenance of any data used. Given the nature of a heparin assay control, the "test set" would likely refer to batches of the control material tested against an assay, not patient data.
  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable/Not provided. For a heparin assay control, the "ground truth" would be established by the known concentration and characteristics of the heparin within the control, likely determined by chemical and analytical methods, not by expert human interpretation of medical images or other diagnostic data. Experts might be involved in formulating the control or validating the analytical methods, but the document does not specify this.
  4. Adjudication method for the test set:

    • Not applicable/Not provided. Adjudication methods like "2+1" or "3+1" are typically used in clinical studies where multiple human readers interpret ambiguous cases. This concept is not relevant to the verification testing of a chemical control product.
  5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This document describes a modification to a chemical control product, not an AI-powered diagnostic device or an imaging system involving human reader interpretation. No MRMC study was performed or is relevant here.
  6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is not an algorithm or an AI device.
  7. The type of ground truth used:

    • Not explicitly stated, but inferred to be analytical/chemical validation. For a heparin assay control, the ground truth would be the known concentration and properties of the heparin within the control, established through rigorous analytical chemistry and adherence to pharmacopeial standards (like the USP heparin monograph). The document only states the source of heparin changed to "revised USP heparin."
  8. The sample size for the training set:

    • Not applicable. This is not an AI or machine learning device that requires a training set.
  9. How the ground truth for the training set was established:

    • Not applicable. (See point 8)

In summary, the provided K103313 document is a 510(k) submission for a minor modification to an in vitro diagnostic control product (Heparin Assay Controls). It focuses on demonstrating substantial equivalence due to a change in a chemical component, not on proving the performance of a new diagnostic algorithm or system using clinical data against defined acceptance criteria in the manner your request describes.

The document's "Conclusion" states: "The modifications to the Medtronic Heparin Assay Controls described in this submission result in a substantially equivalent device because verification testing indicated that the fundamental scientific technology and the intended use are unchanged." This implies that the 'acceptance criteria' were met by demonstrating that the performance of the modified control was equivalent to the predicate, likely through internal analytical testing consistent with existing standards for such controls. However, the specifics of this testing are not detailed in this public summary.

{0}------------------------------------------------

K103313

Image /page/0/Picture/1 description: The image shows the Medtronic logo. The logo consists of a stylized human figure inside a circle with arrows indicating movement, followed by the word "Medtronic" in a bold, sans-serif font. The logo is black and white.

DEC 1 2010

510(k) Summary

Date PreparedOctober 1, 2010
SubmitterMedtronic, Inc.Medtronic Perfusion Systems7611 Northland DriveMinneapolis, MN 55428Establish Registration Number: 2184009
Contact PersonJeffrey L. KollSenior Regulatory Affairs SpecialistPhone: (763) 514-9842Fax: (763) 367-8360Email: jeffrey.l.koll@medtronic.com

Device Name and Classification

Trade Name:Heparin Assay Controls
Common Name:Plasma, Coagulation Control
Regulation Number:21 CFR 864.5425
Product Code:GGN
Classification:Class II

Predicate Device

Medtronic Heparin Assay Controls (K894317 and K042206) cleared on September 5, 1989 and October 1, 2004, respectively.

Comparison to Predicate Device

A comparison of the modified device and the currently marketed Heparin Assay Controls shows the following similarities:

  • Same intended use. .
  • Same operating principle. .
  • Same technological characteristics. .
  • Same performance claims. .

Alleviating Pain · Restoring Health · Extending Life

{1}------------------------------------------------

Description of Device Modification

Heparin

  • . The modification to the current control is to replace old USP heparin with revised USP heparin from the new heparin monograph that became effective October 1, 2009. The source of the heparin remains porcine.

Intended Use

The intended use is unchanged.

Labeling

The current labels are unchanged by this device modification. The current Instructions for Use (IFU) has been updated for style, grammar and readability. Applicable tables have also been updated with information on the construction of the modified controls using revised USP heparin.

  • . Appendix A contains the current IFU.
  • Appendix B contains the draft IFU. .
  • Appendix C contains the current labels .

Conclusion

The modifications to the Medtronic Heparin Assay Controls described in this submission result in a substantially equivalent device because verification testing indicated that the fundamental scientific technology and the intended use are unchanged.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (USA). The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three curved lines.

Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-0609 Silver Spring, MD 20993-0002

Medtronic, Inc. c/o Mr. Jeffrey L. Koll Senior Regulatory Affairs Specialist 8200 Coral Sea Street NE Mounds View, MN 55112

DEC 01 2010

Re: K103313

Trade/Device Name: Heparin Assay Controls Regulation Number: 21 CFR 864.5425 Regulation Name: Multipurpose system for in vitro coagulation studies Regulatory Class: Class II Product Code: GGN Dated: October 1, 2010 Received: November 10, 2010

Dear Mr. Koll:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter

{3}------------------------------------------------

Page 2 - Mr. Jeffrey L. Koll

will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Reena Philip

Maria M. Chan. Ph.D Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known): K 10 331 3

Device Name: Heparin Assay Controls

DEC 1 2010

Indications for Use:

To verify the performance of the HMS Plus and Heparin Assay Cartridges.

For In Vitro Diagnostic use.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Division Sign-Off

Division Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safety

KID3313 510(k)

§ 864.5425 Multipurpose system for in vitro coagulation studies.

(a)
Identification. A multipurpose system for in vitro coagulation studies is a device consisting of one automated or semiautomated instrument and its associated reagents and controls. The system is used to perform a series of coagulation studies and coagulation factor assays.(b)
Classification. Class II (special controls). A control intended for use with a multipurpose system for in vitro coagulation studies is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.