K Number
K042206

Validate with FDA (Live)

Manufacturer
Date Cleared
2004-10-18

(63 days)

Product Code
Regulation Number
864.5425
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

To verify the performance of the HMS instrument and HMS Heparin Assay cartridges

Device Description

The Purple/Black Heparin Assay control is an in vitro diagnostic device. This control is intended to verify the performance of the HMS instrument and the Purple/Black HMS Heparin Assay cartridge. The cartridge is for use in the HMS Plus (Hemostasis Management System) instrument.

AI/ML Overview

The provided text describes a 510(k) premarket notification for a medical device called the "Purple/Black Heparin Assay Control." However, the document focuses on establishing substantial equivalence to a predicate device based on design, test results, and indications for use. It does not contain the detailed information required to fill out a table of acceptance criteria and reported device performance, nor does it describe a study that explicitly proves the device meets specific acceptance criteria in the manner requested by the prompt.

The document mentions "Validation testing was used to establish the performance characteristic of the modifications of this device from the previously marketed device," but it does not provide any details about this testing.

Therefore, most of the requested information cannot be extracted from the provided text.

Here is what can be inferred or stated as "not provided":

Acceptance Criteria and Study Details

Information CategoryDetails from Provided Text
1. Acceptance Criteria and Reported Device PerformanceNot provided in the document. The summary states, "Validation testing was used to establish the performance characteristic of the modifications of this device from the previously marketed device," but no specific criteria or performance metrics are listed.
2. Sample size and Data Provenance (Test Set)Not provided. The document refers to "Validation testing," but no details on sample size, specific test sets, or data provenance (country, retrospective/prospective) are given.
3. Number and Qualifications of Experts (Test Set GT)Not applicable/Not provided. The document describes an in vitro diagnostic control for an assay, not an AI or imaging device requiring expert ground truth for interpretation. There is no mention of experts establishing ground truth for any test set.
4. Adjudication Method (Test Set)Not applicable/Not provided. No adjudication method is mentioned as there is no expert-based ground truth establishment described.
5. MRMC Comparative Effectiveness Study (AI vs. without AI)Not applicable. This device is an in vitro diagnostic control, not an AI-assisted diagnostic tool that would involve human readers. Therefore, an MRMC study is not relevant here.
6. Standalone Algorithm-Only Performance StudyNot applicable. This is an in vitro diagnostic control, not a standalone algorithm. The "validation testing" likely assessed the control's stability, reproducibility, and expected values.
7. Type of Ground Truth UsedNot applicable/Not explicitly stated for the "validation testing." For a control solution, the "ground truth" would typically be the established lot values (target range with known uncertainty) for the analytes it controls, determined through rigorous laboratory methods (e.g., reference methods, inter-laboratory studies). The document states the control verifies performance against the HMS instrument and cartridges, implying the "truth" is the expected performance within the system.
8. Sample size for the training setNot applicable. As an in vitro diagnostic control for an assay, this device does not involve a "training set" in the context of machine learning.
9. How ground truth for the training set was establishedNot applicable. As this device does not involve a "training set" in the context of machine learning, this question does not apply.

Summary of the Study (as described in the 510(k) Summary):

The document mentions "Validation testing" was performed to establish the performance characteristics of the modifications made to the Purple/Black Heparin Assay Control compared to the previously marketed device. The core of the submission is to demonstrate "substantial equivalence" to the predicate devices (Purple/Black HMS Control, K922031, and Hepcon® HMS Coagulation System and its Associated Cartridges and Controls, K894317). The primary difference noted is the "source of heparin."

Conclusion drawn by the submitter: Medtronic demonstrated substantial equivalence based on "design, test results, and indications for use."

However, no specific details on the "test results" or the criteria against which they were evaluated are provided in this summary. The provided text is a summary seeking FDA clearance based on substantial equivalence, not a detailed study report with specific performance metrics and acceptance criteria.

{0}------------------------------------------------

OCT 1 8 2004

K042206

510(k) Summary

Date Prepared:August 13, 2004
Submitter:Medtronic Perfusion Systems7611 Northland BoulevardBrooklyn Park, MN 55428
Contact Person:Dawn M. StenstromPrincipal Regulatory Affairs SpecialistPhone: (763) 391-9604Fax: (763) 391-9603

Device Name and Classification:

Trade Name:Purple/Black Heparin Assay Control
Common Name:Analyzer, Heparin, Automated
Classification:Class II
Predicate Devices:Purple/Black HMS Control, K922031

Hepcon® HMS Coagulation System and its Associated Cartridges and Controls, K894317

Device Description

The Purple/Black Heparin Assay control is an in vitro diagnostic device. This control is intended to verify the performance of the HMS instrument and the Purple/Black HMS Heparin Assay cartridge. The cartridge is for use in the HMS Plus (Hemostasis Management System) instrument.

Indication for Use

This product is intended to verify the performance of the HMS instrument and HMS Heparin Assay cartridges.

Comparison to Predicate Device

The predicate devices are the currently marketed Purple/Black HMS control and the other colored HMS controls. The currently marketed Purple/Black control has the same indications for use and is the same in all aspects to the modified control with exception of the source of heparin.

Summary of Performance Data

Validation testing was used to establish the performance characteristic of the modifications of this device from the previously marketed device.

Conclusion

Medtronic has demonstrated that the Purple/Black Heparin Assay control is substantially equivalent to the predicate devices based upon design, test results, and indications for use.

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus, which is a symbol often associated with healthcare, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the symbol. The caduceus is depicted with three lines forming the wings and a base that resembles a staff with entwined snakes. The overall design is simple and monochromatic.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Ms. Dawn M. Stenstrom Principal Regulatory Affairs Specialist Medtronic Perfusion Systems 7611 Northland Drive N Minneapolis, MN 55428-1088

OCT 1 8 2004

K042206 Re:

Trade/Device Name: Purple/Black Heparin Assay Control Regulation Number: 21 CFR § 864.5425 Regulation Name: Multipurpose system for in vitro coagulation studies Regulatory Class: II Product Code: GGN, JOX, GFT Dated: September 28, 2004 Received: September 29, 2004

Dear Ms. Stenstrom:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for associated in the encreatment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can he found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed nredicate device results in a classification for your device and thus, permits your device to proceed to the market.

{2}------------------------------------------------

Page 2

If you desire specific information about the application of labeling requirements to your device, of to If you desire specific information as advertising of your device, please contact the Office of In of questions on the promotion and Safety at (301) 594-3084. Also, please note the VITO Diagnostic Dovice Draiaaten and manager and premarket notification" (21CFR Part 807.97). regulation entitled, "Misolanding by reiseous on your responsibilities under the Act from the You may outlif only Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Robert L. Becker Jr.

Robert L. Becker, Jr., M.D., Ph. Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

4042206 510(k) Number (if known):

Device Name:

Purple/Black Heparin Assay Control

Indications for Use:

To verify the performance of the HMS instrument and HMS Heparin Assay cartridges

Prescription Use _____________________________________________________________________________________________________________________________________________________________

(Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Arzeline Bautista

Division Sign-Off

Division Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safety

Page1of1

510(k) KD42206

§ 864.5425 Multipurpose system for in vitro coagulation studies.

(a)
Identification. A multipurpose system for in vitro coagulation studies is a device consisting of one automated or semiautomated instrument and its associated reagents and controls. The system is used to perform a series of coagulation studies and coagulation factor assays.(b)
Classification. Class II (special controls). A control intended for use with a multipurpose system for in vitro coagulation studies is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.