K Number
K102249

Validate with FDA (Live)

Date Cleared
2010-12-16

(129 days)

Product Code
Regulation Number
888.3070
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The EXPEDIUM Spine System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine.

The EXPEDIUM Spine System is intended for noncervical pedicle fixation and nonpedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudarthrosis; and failed previous fusion in skeletally mature patients.

Device Description

The EXPEDIUM Spine System is a 5.5mm rod-based and platebased system offered in both titanium and stainless steel. The system consists of monoaxial screws, polyaxial screws, uni-planar screws, reduction screws, reduction hooks, hooks, extended tab implants, sacral extenders, lateral connectors, washers, fixed bolts, polyaxial bolts, closed screws, slotted connectors, plates, nuts, washers, drop-entry connectors, modular cross connectors, transverse rod connectors, and wires.

The proposed addition to the EXPEDIUM 5.5mm Spine System is an increased favored angle polyaxial screw. The polyaxial screw head opens posteriorly for placement of a rod which is secured to a screw by either a dual inner or unitized set screw. This additional component is available in various geometries and sizes to accommodate patient anatomy. It will be provided non-sterile.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the K102249 submission, based on the provided text:

Acceptance Criteria and Device Performance

The provided document describes a Special 510(k) submission for an addition to an existing spinal system, the EXPEDIUM® Spine System. The acceptance criteria and "device performance" (in this context, demonstrating that the new component meets the same performance standards as the predicate device) are based on mechanical non-clinical testing.

1. Table of Acceptance Criteria and Reported Device Performance

TestAcceptance CriteriaReported Device Performance
Static cantilever beam in accordance with ASTM F1798-97Not explicitly stated (implied to be meeting the standard's requirements for strength/stability)"The acceptance criteria was/were met."
Static axial slip in accordance with ASTM F1798-97Not explicitly stated (implied to be meeting the standard's requirements for resistance to slip)"The acceptance criteria was/were met."
Dynamic cantilever beam in accordance with ASTM F1798-97Not explicitly stated (implied to be meeting the standard's requirements for fatigue life/durability)"The acceptance criteria was/were met."

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: The document does not specify the exact sample size (number of devices/components tested) for each mechanical test. It only states that the tests were conducted in accordance with ASTM F1798-97.
  • Data Provenance: The data is from non-clinical mechanical testing performed by the manufacturer, DePuy Spine, Inc., presumably in their labs. This is not patient-derived data, so terms like "country of origin" or "retrospective/prospective" are not applicable.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This question is not applicable to this submission. The "ground truth" for mechanical testing is established by the specified ASTM standard (F1798-97), which outlines the methodology and performance metrics. There are no "experts establishing ground truth" in the sense of clinical interpretation or diagnosis for this type of non-clinical mechanical evaluation.

4. Adjudication Method for the Test Set

This question is not applicable. Adjudication typically refers to the process of resolving discrepancies in expert opinions, often in clinical studies. For mechanical testing against an ASTM standard, the results are objectively measured and compared to the predefined criteria of the standard.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The submission explicitly states: "No clinical tests were performed."

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

This question is not applicable. This submission is for a physical medical device (spinal implant components), not a software algorithm or AI-driven system. Therefore, standalone algorithm performance is not relevant.

7. The Type of Ground Truth Used

The "ground truth" for demonstrating the device meets acceptance criteria is adherence to established mechanical testing standards (ASTM F1798-97). The performance of the new components was compared to these standards to ensure they are equivalent to the predicate devices.

8. The Sample Size for the Training Set

This question is not applicable. This submission is for a physical medical device, not a machine learning model. Therefore, there is no "training set."

9. How the Ground Truth for the Training Set Was Established

This question is not applicable. As there is no training set for a machine learning model, there is no ground truth for a training set to be established.

{0}------------------------------------------------

K102249

Special 510(k) Submission - Additions to the EXPEDIUM 5.5mm Spine Systems

5. 510(K) SUMMARY

Submitter:DePuy Spine, Inc.325 Paramount DriveRaynham, MA 02767DEC 16 2010
Contact Person:Daphney GermainRegulatory Affairs AssociateDePuy Spine, Inc.Voice:(508) 880-8395Fax:(508) 828-3797E-Mail:dgermain@its.jnj.com
Date Prepared:August 6, 2010
Device Class:Class III
Tradename:EXPEDIUM® Spine System
Common Name:Appliance, Fixation, Spinal Interlaminal;Orthosis, Spondyloisthesis Spinal Fixation;Orthosis, Spinal Pedicle Fixation;Orthosis, Spinal Pedicle Fixation, For Degenerative Disc Disease
Classification Name:Spinal interlaminar fixation orthosisper 21 CFR §888.3050
Spinal intervertebral body fixation orthosisper 21 CFR §888.3060
Pedicle screw spinal fixationper 21 CFR §888.3070
Classification Panel:Orthopedics
FDA Panel Number:87
Product Code(s):NKB, KWO, KWP, MNH, MNI

Proprietary Name: EXPEDIUM® Spine System

Device Description: The EXPEDIUM Spine System is a 5.5mm rod-based and platebased system offered in both titanium and stainless steel. The system consists of monoaxial screws, polyaxial screws, uni-planar screws, reduction screws, reduction hooks, hooks, extended tab implants, sacral extenders, lateral connectors, washers, fixed bolts, polyaxial bolts, closed screws, slotted connectors, plates, nuts,

{1}------------------------------------------------

washers, drop-entry connectors, modular cross connectors, transverse rod connectors, and wires.

The proposed addition to the EXPEDIUM 5.5mm Spine System is an increased favored angle polyaxial screw. The polyaxial screw head opens posteriorly for placement of a rod which is secured to a screw by either a dual inner or unitized set screw. This additional component is available in various geometries and sizes to accommodate patient anatomy. It will be provided non-sterile.

Intended Use: The EXPEDIUM Spine System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine.

The EXPEDIUM Spine System is intended for noncervical pedicle fixation and nonpedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion in skeletally mature patients.

Manufactured from ASTM F 136 implant grade titanium alloy.

Predicate Devices:

Materials:

EXPEDIUM Spine System (K033901, K051024, K062174, K070387) ISOLA Spine System (K980485)

Summary of Technological Differences:

The purpose of this submission is to obtain market clearance for the proposed additional component to the EXPEDIUM 5.5mm Spine Systems which consists of a favored angle polyaxial screw. The proposed component has the same intended use, design characteristics, performance, and packaging as the predicate devices. The key differences between the subject and predicate devices are:

  • The Medial/Lateral (M/L) degree increase of the . favored angle head
  • . The increase in polyaxial head height.

{2}------------------------------------------------

Nonclinical Test The following mechanical tests were conducted: Summary: . Static cantilever beam in accordance with ASTM F1798-97 test standard Evaluating the Static and Fatigue Properties of Interconnection Mechanisms and Subassemblies Used in Spine Arthrodesis Implants. The acceptance criteria was/were met. : Static axial slip in accordance with ASTM F1798-97 test . standard Evaluating the Static and Fatigue Properties of Interconnection Mechanisms and Subassemblies Used in Spine Arthrodesis Implants. The acceptance criteria was/were met. Dynamic cantilever beam in accordance with ASTM � F1798-97 test standard Evaluating the Static and Fatigue Properties of Interconnection Mechanisms and Subassemblies Used in Spine Arthrodesis Implants. The acceptance criteria was/were met. Clinical Test No clinical tests were performed. Summary: Based on the predicate comparison and testing, the subject addition Conclusion: to the EXPEDIUM 5.5mm Spine Systems is substantially equivalent to the predicate device. :

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three stylized lines.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

DePuy Spine, Inc. % Ms. Daphney Germain Regulatory Affairs Associate 325 Paramount Drive Raynham, Massachusetts 02767

DEC 16 2010

Re: K102249

R102249
Trade/Device Name: EXPEDIUM® Spine System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class III Product Code: NKB, KWQ, KWP, MNH, MNI Dated: December 03, 2010 Received: December 06, 2010

Dear Ms. Germain:

We have reviewed your Section 510(k) premarket notification of intent to market the device indicati we have reviewed your Scctorn 5 ro(t) premained is substantially equivalent (for the indications referenced and nave determined the actrose are arrooms of the devices marketed in interstate for use stated in the encrosule) to regary manotes profithe Medical Device Amendments, or to commerce prior to May 28, 1970, the chacinent with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA).
and Cosmetic Act (Act) that do not require approval of a provisions of the Act and Cosmetic Act (Act) that do not require approval controls provisions of the Act. The You may, therefore, market the device, subject to the general its .
general controls provisions of the Act include requirements for answer misbranding and general controls provisions of the Act mensibitions against misters and devices, good manufacturing placece, not evaluate information related to contract liability.
adulteration. Please note: CDRH does not evaluate information related to traile adulteration. Please note: CDRFI does not evaluate information of the truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it is If your device is classified (see au0v) mio enase regulations affecting your device can be
may be subject to additional controls. Title Regulations affecting your addition. F may be subject to additional controls: "Existing and to 898. In addition, FDA may
found in the Code of Federal Regulations, Title 21, Facts 800 to Begister found in the Code of I cocharters, concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that FDA's issualle of a substantial of a varies with other requirements of the Act
that FDA has made a determination that your device complies with other r

{4}------------------------------------------------

Page 2 - Ms. Daphney Germain

or any Federal statutes and regulations administered by other Federal agencies. You must or any Federal statules and regulations administered of registration and listing (21)
comply with all the Act's requirements, including, but not series of medical comply with all the Act STEquirements, increasing of ever reporting of medical CFR Part 807); labeling (21 CFR 807); intention activery mag (tipes mag (stjer requirements as set
device-related adverse events (21 CFR 803); good manufacturing practice req device-related adverse evenis) (2) CFR Pat 820); and if applicable, the electronic
forth in the quality systems (QS) regulation (21 CFR Patt 820); and if applicable, the elec forth in the quality systems (QS) regulation (DF-SF-S-C-S-F-S-C-S-2-5-2-2-2-2-2-2-2-2-2-2-2-2-2-2-2-2-2-2-2-2-2-2-2

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please
ly of the specific advice and and and and and in a column LSS00 htm for If you desire specific advice for your acvice of var sicontribution of Commission of Compliance. go to http://www.rda.gov/Abour12/0Cchiccs Office.of Compliance. Also, please
the Center for Devices and Radiological Health's (CDRH) starter resifiestion" (21CFR Part the Center for Devices and Radiological Treater of premarket notification" (21CFR Part
note the regulation entitled, "Misbranding by reference to athe MDP regulation (2 note the regulation ennited, "Misoranung of reference to promises
807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

CFR Parl 803), picesse go to http://www.ida.gov/Nicaletrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the You may other general information on your response Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

(800) 038-2041 01 (301) 770 1100 0100 000 000 000 000 000 000 000 000 000 000 000 000 000 000 000 000 000 000 000 000 000 000 000 000 000 000 000 000 000 000 000 000 000 000

Sincerely yours,

Mark A. Millers

Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

K102249

Special 510(k) Submission - Additions to the EXPEDIUM 5.5mm Spine System

INDICATIONS FOR USE STATEMENT 4.

510(k) Number (if known):

DEC 16 2010

Device Name: EXPEDIUM® Spine System

Indications For Use:

The EXPEDIUM Spine System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine.

The EXPEDIUM Spine System is intended for noncervical pedicle fixation and nonpedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); · tumor; pseudarthrosis; and failed previous fusion in skeletally mature patients.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Signature

(Division Sign-off)
Division of Surgical, Orthopedic,
and Restorative Devices

510(k) NumberK 102249
-------------------------

N/A