K Number
K101533

Validate with FDA (Live)

Date Cleared
2011-08-10

(433 days)

Product Code
Regulation Number
864.5425
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The EXTEM assay is a semi-quantitative in vitro diagnostic assay on the ROTEM® delta Thromboelastometry System to monitor the coagulation process via the extrinsic pathway in citrated whole blood specimens. Clotting characteristics are described by the functional parameters Clotting Time (CT), Speed of Clot Formation (CFT and alpha angle), Clot Firmness (A20/MCF) and Clot Lysis (LOT, ML, LI(x)). Speed of clot formation time (CFT and alpha) is complementary parameter and should be used only in conjunction with the main parameters Clotting time (CT) and Clot Firmness (A20/MCF). The indication for ROTEM® delta use is in adult patients where an evaluation of their blood coagulation properties is desired. Coagulation evaluations with the ROTEM® delta system are commonly used to assess clinical conditions in organ transplantation, cardiovascular surgery, cardiology procedures and trauma to assess post-operative hemorrhage and / or thrombosis.

The FIBTEM assay is a semi-quantitative in vitro diagnostic assay on the ROTEM® delta Thromboelastometry System to monitor the clot firmness of a citrated whole blood specimens after blocking platelet contribution to the clot firmness The fib-TEM® reagent is always used in conjunction with ex-TEM® reagent. Clotting characteristics are described by the functional parameter Clot Firmness (A20/MCF). The indication for ROTEM® delta use is in adult patients where an evaluation of their blood coagulation properties is desired. Coagulation evaluations with the ROTEM® delta system are commonly used to assess clinical conditions in organ transplantation, cardiovascular surgery, cardiology procedures and trauma to assess post-operative hemorrhage and / or thrombosis.

The APTEM assay is a semi-quantitative in vitro diagnostic assay on the ROTEM® delta Thromboelastometry System to monitor the clot firmness of a citrated whole blood specimens after blocking hyperfibrinolysis by aprotinin. The ap-TEM® reagent is always used in conjunction with ex-TEM® reagent. Clotting characteristics are described by the functional parameters Clotting Time (CT), Speed of Clot Formation (CFT and alpha angle), Clot Firmness (A20/MCF) and Clot Lysis (LOT, ML, LI(x)). Speed of clot formation time (CFT and alpha) is complementary parameter and should be used only in conjunction with the main parameters Clotting time (CT) and Clot Firmness (A20/MCF). The indication for ROTEM® delta use is in adult patients where an evaluation of their blood coagulation properties is desired. Coagulation evaluations with the ROTEM® delta system are commonly used to assess clinical conditions in organ transplantation, cardiovascular surgery, cardiology procedures and trauma to assess post-operative hemorrhage and / or thrombosis.

Device Description

The ROTEM® delta Thromboelastometry System consists of a fourcolumn instrument (with integrated computer module, computer controlled electronic pipette, software), system reagents (in-TEM®, hep-TEM®, star-TEM®, ex-TEM®, fib-TEM® and ap-TEM®, quality controls (ROTROL N, ROTROL P) and measurement cells (Cup and Pin pro). The blood sample is filled into a cylindrical cup. A pin oscillates permanently while it is immersed in the blood holding cup. The motion of the pin is detected by an optical detection system. Data are processed and analyzed by a computer with special software. If no clotting takes place, the movement of the pin is not obstructed. When a clot forms and attaches itself to the pin and cup surfaces, the movement is obstructed. As the clot becomes firmer, the rotational movement of the pin is reduced. The rotational movement of the pin is converted into amplitude with the following definitions applying to the thromboelastogram (TEM): Amplitude of 0 mm means unobstructed oscillation, while amplitude of 100 mm can be regarded as infinite firmness and blocking of the pin by the clot. The TEM amplitude is a measure of the clot firmness.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the ROTEM® delta Thromboelastometry System, based on the provided text:

Acceptance Criteria and Device Performance

The acceptance criteria are primarily focused on the precision of the ROTEM® delta system's assays (EXTEM, FIBTEM, APTEM) and their comparability to the predicate device (TEG® 5000).

1. Table of Acceptance Criteria and Reported Device Performance

Test ParameterAcceptance Criteria (Within-run Precision)Reported Performance (Within-run CV%)Acceptance Criteria (Between Operator Precision)Reported Performance (Between Operator CV%)
EXTEM
CT< 10 %4.4 %< 10 %7.9 %
CFT< 20 %5.5 %< 30 %13.4 %
Alpha< 5 %1.4 %< 5 %0.3 %
A20< 5 %1.9 %< 6 %5.5 %
APTEM
CT< 10 %4.4 %< 10 %7.9 %
CFT< 20 %5.5 %< 30 %13.4 %
Alpha< 5 %1.4 %< 5 %0.3 %
A20< 5 %1.9 %< 6 %5.5 %
FIBTEM
A20< 5 %2.9 %< 6 %N/A (Only A20 applies for FIBTEM)

Method Comparison Acceptance Criteria (Implicit from regression analysis):
The study aimed to show equality of clot firmness (MCF vs. MA) and a linear correlation (r>0.8) between ROTEM® and TEG® for kinetic parameters (CT vs. R, CFT vs. K, Alpha Angle vs. Angle). The reported R_OLS values for all comparisons generally exceed 0.9, indicating strong linear correlation.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size:
    • EXTEM Method Comparison: CT vs. R (n=100), CFT vs. K (n=91), α vs. Angle (n=100), MCF vs. MA (n=93).
    • APTEM Method Comparison: CT vs. R (n=84), CFT vs. K (n=73), α vs. Angle (n=82), MCF vs. MA (n=79).
    • FIBTEM Method Comparison: MCF vs. MA (n=88).
    • Precision Studies: Conducted with healthy donor blood and ROTROL N control. Specific sample numbers are not provided beyond "5 runs on each of the 4 channels of one instrument" and "5 operators run ROTROL N in duplicates."
    • Reference Ranges: Estimated using the CSLI C28-A2 guideline on three clinical US reference sample groups, with results consistent with earlier European studies.
  • Data Provenance:
    • Method Comparison: Patient samples from 3 US centers. These patients were during surgery and post-surgery in the intensive care unit (ICU). Contrived samples were added to broaden the range of comparison.
    • Reference Ranges: US and European reference sample groups.
    • Retrospective/Prospective: Not explicitly stated, but the mention of "patient samples during surgery and post surgery at the intensive care unit (ICU)" and "contrived samples were added" for the method comparison, along with "healthy donor blood" for precision, suggests a mix of prospective collections and prepared samples.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

The document does not explicitly state the number of experts or their qualifications for establishing ground truth. The "ground truth" in this context is the measurement provided by the predicate device, TEG® 5000, which is a legally marketed device. The study seeks to establish substantial equivalence to this predicate.

4. Adjudication Method for the Test Set

Not applicable. The study is a direct comparison of measurements between the investigational device (ROTEM® delta) and the predicate device (TEG® 5000), not an interpretation that would require an adjudication method.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

No, an MRMC comparative effectiveness study was not done. This study focuses on the analytical performance (precision and method comparison) of the device itself against a predicate, not on how human readers' performance changes with or without AI assistance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Yes, the studies presented are standalone performance studies of the ROTEM® delta system and its reagents. The device generates quantitative functional parameters (CT, CFT, Alpha, A20/MCF), and the study evaluates the accuracy and precision of these measurements, as well as their correlation to the predicate device. There is no human-in-the-loop component being evaluated for this 510(k) submission.

7. The Type of Ground Truth Used

The primary "ground truth" used for method comparison is the measurements obtained from the predicate device, the TEG® 5000 Thrombelastograph.

For precision studies, the ground truth involves the expected consistency of measurements on controlled samples (healthy donor blood, internal controls like ROTROL N).

For reference ranges, the ground truth is established through analysis of measurements from "clinical US reference sample groups" and consistency with "earlier studies on European reference sample groups."

8. The Sample Size for the Training Set

The document does not mention a "training set" in the context of device development or any machine learning algorithms. The study focuses on the validation of the device's analytical performance.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as no training set for a machine learning model is mentioned or implied in the provided text. The device's measurement principle is based on physical oscillation detection and established thromboelastometry principles, rather than a learned algorithm needing a separate training phase.

{0}------------------------------------------------

4.0 510(k) Summary

4.1 Submitter's name / Contact Person

AUG 1 0 2011

Dr. Volker-Joachim Friemert Head of Quality Management and Regulatory Affairs Tem Innovations GmbH Martin-Kollar-Strasse 13-15 81829 Munich Germany

Contact: Office: +49-89-454295-0 Dr. Volker-Joachim Friemert: +49-89-454295-11 Fax: +49-89-454295-22 E-mail: volker-joachim.friemert@tem-innovations.de

4.2 Identification of the Product

Trade Name:ROTEM® delta Thromboelastometry Systemex-TEM® reagentfib-TEM® reagentap-TEM® reagent
Common Name:Whole Blood Haemostasis System
Classification Name:Multipurpose System for In Vitro CoagulationStudies

4.3 Identification of the predicate

Thrombelastograph® Coagulation Analyzer (TEG®) - 5000 Series K002177, Product Code JPA, Haemoscope Corp.

4.4 Description of the Device

The three assays are system reagents for the previously cleared (K083842) ROTEM® delta Thromboelastometry system. For clarity, the description of ROTEM® delta Thromboelastometry analyzer is provided below.

The ROTEM® delta Thromboelastometry System consists of a fourcolumn instrument (with integrated computer module, computer controlled electronic pipette, software), system reagents (in-TEM®,

{1}------------------------------------------------

hep-TEM®, star-TEM®, ex-TEM®, fib-TEM® and ap-TEM®, quality controls (ROTROL N, ROTROL P) and measurement cells (Cup and Pin pro). The blood sample is filled into a cylindrical cup. A pin oscillates permanently while it is immersed in the blood holding cup. The motion of the pin is detected by an optical detection system. Data are processed and analyzed by a computer with special software. If no clotting takes place, the movement of the pin is not obstructed. When a clot forms and attaches itself to the pin and cup surfaces, the movement is obstructed. As the clot becomes firmer, the rotational movement of the pin is reduced. The rotational movement of the pin is converted into amplitude with the following definitions applying to the thromboelastogram (TEM): Amplitude of 0 mm means unobstructed oscillation, while amplitude of 100 mm can be regarded as infinite firmness and blocking of the pin by the clot. The TEM amplitude is a measure of the clot firmness.

4.5 Intended Use

The EXTEM assay is a semi-quantitative in vitro diagnostic assay used to monitor the coagulation process via the extrinsic pathway in citrated whole blood specimens on the ROTEM® delta. Clotting characteristics are described by the functional parameters Clotting Time (CT), Speed of Clot Formation (CFT and alpha angle), Clot Firmness (A20/MCF) and Clot Lysis (LOT, ML, Ll(x)). CFT and alpha (Speed of clot formation time) are complementary parameters and should be used in conjunction with the main parameters Clotting Time (CT) and Clot Firmness (A20/MCF).

The FIBTEM assay is a semi-quantitative in vitro diagnostic assay on the ROTEM® delta Thromboelastometry System to monitor the clot firmness of a citrated whole blood specimens after blocking platelet contribution to the clot firmness. fib-TEM® is always used in conjunction with ex-TEM®. Clotting characteristics are described by the functional parameter Clot Firmness (A20/MCF).

The APTEM assay is a semi-quantitative in vitro diagnostic assay on the ROTEM® delta Thromboelastometry System to monitor the clot firmness of a citrated whole blood specimens after blocking hyperfibrinolysis by aprotinin. ap-TEM® is always used in conjunction with ex-TEM®. Clotting characteristics are described by the functional parameters Clotting Time (CT), Speed of Clot Formation (CFT and alpha angle), Clot Firmness (A20/MCF) and Clot Lysis (LOT, ML, LI(x)). CFT and alpha (Speed of clot formation time) are

  • 18 -

{2}------------------------------------------------

complementary parameters and should be used in conjunction with the main parameters Clotting Time (CT) and Clot Firmness (A20/MCF).

CONFIDENTIAL

4.6 Summary of Technological Characteristics of the Product, Compared with the Predicate Device

ROTEM® deltaTEG® 5000
Reagents /Accessories:
Extrinsic ContactActivation Reagentex-TEM® (Rabbitbrain thromboplastin)Commercialthromboplastin (e.g.ex-TEM®)
Platelet BlockerReagentfib-TEM®(Cytochalasin D,CaCl2)Reopro®,CaCl2
Antifibrinolytic drugap-TEM®(Aprotinin, CaCl2)Trasylol®,CaCl2

*ex-TEM®, fib-TEM®, and ap-TEM® are the commercial names of the reagents used with the respective EXTEM, FIBTEM, and APTEM assays.

4.7 Executive Summary of the Study Report

The ROTEM® delta (ROTEM) Whole Blood Haemostasis System has recently been FDA cleared (K083842) using the TEG® 5000 (TEG) as a predicate device. In this study, the performance characteristics of the ex-TEM®, fib-TEM® and ap-TEM® reagents for ROTEM® delta were evaluated. For these three tests, no commercially available TEG® 5000 tests are available, but the TEG® 5000 User Manual lists on pp 11-13 recommendations on how to perform these tests using other manufacturers' reagents'.

Based on the recommendations in the TEG® manual and the study protocol the three assays (EXTEM, FIBTEM and APTEM) were run on the ROTEM® and their equivalents (as described in the TEG® manual) were run on the predicate device, the TEG® in support of a 510(k) submission.

Tissue Factor (TF) is the common activator in all three tests on both ROTEM® and on TEG®. Detailed reagent compositions are shown in Table 2.

  1. TEG® 5000 User's Manual. Version 4.2 Software. Remote and TEG-enabled versions. Haemoscope Corporation. Rev 02, 2006-04.

{3}------------------------------------------------

ROTEM® / TEG® testActivation PrincipleComponents (reagents)ROTEM®Components (reagents)TEG®
EXTEM / TF activatedTissue Factor (TF)TF reagent ( ex-TEM®rabbit brainthromboplastin), CaCl2TF reagent (ex-TEM®rabbit brainthromboplastin), CaCl2
FIBTEM / Platelet BlockedTissue Factor (TF)Cytochalasin D, CaCl2ex-TEM®Reopro®, CaCl2TF reagent (ex-TEM®),
APTEM / AntifibrinolyticDrugTissue Factor (TF)Aprotinin, CaCl2ex-TEM®Trasylol® (aprotinin),CaCl2TF reagent (ex-TEM®),

Table 2. Extrinsically activated tests as nerformed in ROTEM® and TEG

ex-TEM® is the name of the TEM Innovations commercial brand of rabbit thromboplastin

The EXTEM assay run on the ROTEM® delta was compared to a "generic" tissue factor reagent run on the TEG® 5000 as recommended in the TEG® User Manual (page 12), The generic tissue factor chosen, was the ex-TEM reagent.

The FIBTEM assay (ex-TEM® reagent plus Cyochalasin D) run on the ROTEM® was compared to the TF reagent plus ReoPro run on the TEG" 5000 and used as source of TF. Cyclochalasin D and ReoPro® are both potent platelet inhibitors that inhibit platelet almost completely 3. ReoPro® is not incorporated into the FIBTEM assay because Cytochalasin D is not readily available for commercial use and is prohibitively expensive as it is an intravenous drug available by prescription only.

The APTEM assay ( ex-TEM® reagent plus aprotinin) run on the ROTEM®, was compared the TF reagent with aprotinin run on the TEG® as recommended in the TEG® 5000 User Manual (page 13) for a hyperfibrinolysis confirmation test. The ex-TEM® reagent was used as source of TF. Trasylol® (Bayer) was used on the TEG® 5000 as recommended in the manual. A generic aprotinin, rather than Trasylol® is incorporated into the APTEM assay because Trasylol® is not readily available for commercial use as it is an intravenous drug available by prescription only.

  1. Khurana S, Mattson J.C, Westley S, O'Neil W.W, Timmis G.C, Safian R.D. Monitoring platelet glycoprotein IIb/Ita-fibrin interaction with tissue factor-activated thrombelastography. J. Lab Clin Med 1997; 4,:401-411

  2. Lang T., Toller W. , Gütl M. , Mahla E. , Metzler H. , Rehak P. , März W. , Hallwachs-Baumann G. Different effects of abciximab and cytochalasin D on clot strength in thromboelastography. J Thromb Haemost 2004;2:147-53

{4}------------------------------------------------

For all three tests, the ROTEM® was shown to have good precision in its primary parameter clot firmness and adequate precision in its secondary coaqulation kinetics parameters (see 4.8.2 - 4.8.4). The main parameters and their definitions are summarized in Table 2.

The method comparison with TEG® showed equality of the clot firmness (MCF vs. MA). The kinetic parameters (CT vs. R, CFT vs. K, Alpha Angle vs. Angle) showed a linear correlation between ROTEM® and TEG® (r>0.8).

Reference ranges for the ROTEM® test were estimated using the CSLI C28-A2 guideline on three clinical US reference sample groups. The reference ranges determined showed no significant center-to-center deviations and were in accordance with the reference ranges determined in earlier studies on European reference sample groups (Table 3).

MCF (mm)A20 (mm)CT (sec)CFT (sec)Alpha (°) ROTEM®
EXTEMEU50-7250-7138-7934-15963-83
US51.7-70.350.2-69.843.2-81.647.6-126.865.0-80.0
FIBTEMEU9-258-24
US7.0-24.07.0-23.8

Table 3: ROTEM® reference ranges from studies in Europe (EU) and US

Three interfering substances widely used in coagulation management, the antifibrinolytic drugs aprotinin, tranexamic acid and epsilon-amino caproic acid (ACA) were investigated. Dose-response curves were investigated for heparin. for dilution and for urokinase on the EXTEM model in order to verify the diagnostic principles of thromboelastometric methods on ROTEM®.

In summary, ROTEM® is a precise Whole Blood Haemostasis System with the typical performance characteristics of a thrombelastographic method (no aprotinin interference on the extrinsically activated tests, high heparin insensitivity of the tests with extrinsic activation and sensitivity to dilution and lysis induced by urokinase in-vitro). Its reference ranges are reproducible from center to center.

The method comparison with TEG® shows equality of the primary parameter clot firmness and a linear regression and good correlation in the secondary kinetic parameters.

In aggregate the data presented in this report demonstrate that the ROTEM® system and the three assays described are substantially equivalent to the predicate TEG® System and corresponding assays.

{5}------------------------------------------------

4.8 Performance Data

4.8.1 Acceptance Criteria for Precision

The following tables show the Acceptance Criteria for ROTEM® reagents:

EXTEM/ FIBTEM/ APTEM:

2

TestCTCFTAlphaA20
Within-run Precision1< 10 %< 20 %<5 %<5 %
Between OperatorPrecision2< 10 %< 30%<5 %<6 %

5 runs on each of the 4 channels of one instrument testing healthy donor blood; CVs in highly pathological samples may vary

5 operators run ROTROL N in duplicates (only EXTEM)

For the FIBTEM test, only A20 applies.

4.8.2 EXTEM

Precision:

CTCV (%)CFTCV (%)α-angleCV (%)A20CV (%)
Within-run14.45.51.41.9
Between-operator27.913.40.35.5

5 runs on each of the 4 channels of one instrument testing healthy donor blood; CVs in highly pathological samples may vary

2 5 operators run ROTROL N in duplicates

Interference:

1

Interference with the antifibrinolytic drugs aprotinin, tranexamic acid (TXA) and epsilon aminocaproic acid (EACA) was investigated. No interference was seen on the EXTEM test with aprotinin (up to 400 kIU/ml spiked in whole blood).

No interference was found with tranexamic acid (up to 60 µg/ml) and EACA (up to 600 µg/ml spiked in whole blood, resp.).

Heparin responsiveness:

When using ex-TEM® as an activator the parameters are unaffected up to a heparin concentration of 4 U/ml UFH in whole blood (corresponding to app. 8 U/ml in the plasma).

{6}------------------------------------------------

Method comparison with TEG® 5000 Thrombelastograph (Haemoscope Corp, Niles, IL):

Method comparison studies were conducted in 3 US centers on patient samples. The patients enrolled comprised patients during surgery and post surgery at the intensive care unit (ICU). In order to broaden the range of comparison, contrived samples were added.

The extrinsically activated EXTEM test on ROTEM® delta was compared to an extrinsically activated tissue factor assay on TEG® 5000.

Scatter Plots CT vs. R

CT: Clotting Time (ROTEM® delta parameter, usually in sec, for regression analysis plotted in min)

R: Reaction time (corresponding TEG® parameter, in min)

Image /page/6/Figure/9 description: The image is a scatter plot titled "EXTEM mean CT [min] vs. mean R [min] Deming Regression". The x-axis is labeled "TEG" and ranges from 0 to 10. The y-axis is labeled "ROTEM" and ranges from 0 to 10. The data points are clustered in the lower left corner and then trend upwards in a linear fashion.

MCF vs. MA

MCF: Maximum Clot Firmness (ROTEM® delta parameter, in mm)) MA: Maximum Amplitude (corresponding TEG® parameter, in mm)

{7}------------------------------------------------

Image /page/7/Figure/3 description: The image is a scatter plot titled "EXTEM mean MCF [mm] vs. mean MA [mm] Deming Regression". The x-axis is labeled "TEG" and ranges from 0 to 80. The y-axis is labeled "ROTEM" and ranges from 0 to 80. The scatter plot shows a positive correlation between TEG and ROTEM, with data points clustered along a regression line.

Regression Statistics:

nMinROTEM®/TEG®MaxROTEM®/TEG®SlopeDemingInterceptDemingSlope OLSInterceptOLSROLS
CT vs. R1001/06/60.990.290.940.350.9497
CFT vs. K910/119/230.790.580.790.610.9855
α vs.Angle10010/3488/871.47-37.051.41-32.910.9736
MCF vs.MA9315/380/780.876.760.857.890.9720

4.8.3 АРТЕМ

Precision:

1

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------a-anale:
41011011
VAlith

5 runs on each of the 4 channels of one instrument testing healthy donor blood; CVs in highly pathological samples may vary

Heparin responsiveness:

When using ex-TEM® as an activator the parameters are unaffected up to a heparin concentration of 4 U/ml UFH in whole blood (corresponding to app. 8 U/ml in the plasma)

{8}------------------------------------------------

Method comparison with TEG® 5000 Thrombelastograph (Haemoscope Corp, Niles, IL):

Method comparison studies were conducted in 3 US centers on patient samples. The patients enrolled comprised patients during surgery and post surgery at the intensive care unit (ICU). In order to broaden the range of comparison, contrived samples were added.The extrinsically (ex-TEM®) activated APTEM test on ROTEM® delta was compared to an extrinsically activated tissue factor plus Trasysol® assay on TEG®5000.

Scatter Plots:

CT vs. R

CT: Clotting Time (ROTEM® delta parameter, usually in sec, for regression analysis plotted in min) R: Reaction time (corresponding TEG® parameter, in min)

Image /page/8/Figure/8 description: The image is a scatter plot titled "APTEM mean CT [min] vs. mean R [min] Deming Regression". The x-axis is labeled "TEG" and ranges from 0 to 10. The y-axis is labeled "ROTEM" and ranges from 0 to 10. The scatter plot shows a positive correlation between TEG and ROTEM, with a regression line drawn through the data points.

MCF vs. MA

MCF: Maximum Clot Firmness (ROTEM® delta parameter, in mm)) MA: Maximum Amplitude (corresponding TEG® parameter, in mm)

{9}------------------------------------------------

Image /page/9/Figure/3 description: The image is a scatter plot titled "APTEM mean MCF [mm] vs. mean MA [mm] Deming Regression". The x-axis is labeled "TEG" and ranges from 0 to 90. The y-axis is labeled "ROTEM" and ranges from 0 to 90. The plot shows a positive correlation between TEG and ROTEM, with data points scattered around a regression line.

Regression Statistics:

nMinROTEM®/TEG®MaxROTEM®/TEG®SlopeDemingInterceptDemingSlope OLSInterceptOLSROLS
CT vs. R841/05/61.020.380.950.480.934
CFT vs. K730/119/220.850.570.790.620.9757
α vs.Angle8210/3688/871.51-39.961.46-37.390.9777
MCF vs.MA7917/1080/810.885.840.867.050.9544

4.8.4 FIBTEM

Precision:

1

A20CV (%)
Within-run12.9

5 runs on each of the 4 channels of one instrument testing healthy donor blood; CVs in highly pathological samples may vary

Heparin responsiveness:

When using ex-TEM® as an activator the parameters are unaffected up to a heparin concentration of 4 U/ml UFH in whole blood (corresponding to app. 8 U/ml in the plasma)

{10}------------------------------------------------

Method comparison with TEG® 5000 Thrombelastograph (Haemoscope Corp, Niles, IL):

Method comparison studies were conducted in 3 US centers on patient samples. The patients enrolled comprised patients during surgery and post surgery at the intensive care unit (ICU).

In order to broaden the range of comparison, contrived samples were added.The extrinsically (ex-TEM®) activated FIBTEM test on ROTEM® delta was compared to an extrinsically activated tissue factor plus Reopro® assay on TEG®5000.

MCF vs. MA

MCF: Maximum Clot Firmness (ROTEM® delta parameter, in mm)) MA: Maximum Amplitude (corresponding TEG® parameter, in mm)

Image /page/10/Figure/8 description: This image is a scatter plot titled "FIBTEM mean MCF [mm] vs. mean MA [mm] Deming Regression". The x-axis is labeled "TEG" and ranges from -10 to 60. The y-axis is labeled "ROTEM" and ranges from -10 to 60. There is a regression line plotted through the data.

Regression Statistics:

nMinROTEM®/TEG®MaxROTEM®/TEG®SlopeDemingInterceptDemingSlope OLSInterceptOLSROLS
MCF vs.MA882/358/571.05-4.530.98-3.090.9384
  • 27 -

{11}------------------------------------------------

4.9 Conclusion (Statement of Equivalence)

The data and information provided in this submission support a substantial equivalence determination, and, therefore, clearance of the 510(k) premarket notification for the EXTEM-, FIBTEM- and APTEM- assays on the ROTEM® delta Thromboelastometry System.

  • 28 -

{12}------------------------------------------------

5.0 Device Description and Specifications

System Description

The previously cleared (K083842) ROTEM delta Thromboelastometry system is based on thromboelastometry, an improved form of the classical thromboelastography / thrombelastography developed by Hartert in 1948. The technique was previously called rotation thrombelastography (ROTEG®), but renamed in 2003 .

Thromboelastometry and thromboelastography are based on the measurement of elasticity of blood by continuous graphic logging of the firmness of a blood clot during clot formation (coagulation factors and inhibitors, platelets and fibrin) and subsequent fibrinolysis.

The ROTEM® Measurement Principle:

The patented ROTEM® technology is based on a disposable measurement cell with a fixed cup in which a pin oscillates permanently. The motion of the pin is detected by an optical detection system. Data are processed and analysed by a computer with special software.

Image /page/12/Figure/9 description: This image shows a diagram of a device with several labeled components. The device includes an oscillating axis, a counterforce spring, a light beam, and a CCD array detector. It also features a cup with a blood sample, a sensor pin, and a temperature-controlled cup holder, along with fibrin strands and platelet aggregates. The device also includes a ball bearing and a data processing unit.

ROTEM® measurement principle Fig. 1)

&quot;TEG" and Thrombelastograph" are trademarks of Haemoscope Corp./ TEM and ROTEM® are registered trademarks of Tem Innovations GmbH

{13}------------------------------------------------

For measurement, the blood sample is filled into the 8 mm diameter cylindrical cup. This cup is placed on a 6 mm diameter pin which is attached to the lower end of a vertical axis. The centre section of the axis is guided by a ball bearing. The axis oscillates to the left and to the right by rotating through an angle of 4.75° via a s pring connector.

The rotation is detected optically via a mirror plate at the upper end of the axis, a diode as light source and a light sensitive sensor (CCD (charge-coupled device) array). If no clotting takes place, the movement is not obstructed. When a clot is formed and attaches itself to the pin and cup surfaces, the movement is obstructed.

The result is a balance between the torque introduced by the spring connector into the pin and the torque generated by the clot on the pin. As the clot becomes firmer, the rotational movement of the pin is reduced. The rotational movement of the pin is converted into amplitude with the following definitions applying to the ROTEM® delta analyzer:

ROTEM®ParameterDefinitionInterpretation
MCFMaximum ClotFirmness [mm]Maximum clot firmness(maximum amplitude) of thedeveloped clot during the test.Normal/reduced/increased clotfirmness.
A10, A20Amplitude 10, 20[mm]Clot firmness (amplitude) at thetime points 10 and 20 minutesafter CT.Normal/reduced/increased clotfirmness.
CTClotting Time [s]The time from test start until firstsignificant level of clot firmness(2 mm) is reached.Normal/reduced/increased speedof coagulation initiation.
CFTClot FormationTime [s]The time from the CT/R until aclot firmness of 20 mm isreached.Normal/reduced/increased speedof coagulation amplification andpropagation(complementary parameter)
αAlpha Angle [°]Angle between the baseline anda tangent to the clotting curvethrough the 2 mm (CT/R) point.Normal/reduced/increased speedof coagulation amplification andpropagation(complementary parameter)

{14}------------------------------------------------

Image /page/14/Figure/3 description: This image shows a graph with the title "Alpha-angle" and "MCF = Maximum Clot Firmness". The graph has an x-axis labeled with numbers 15, 30, 45, and 60. The y-axis is labeled with numbers from 20 to 60. The image also contains the text "CT = Clotting Time" and "CFT = Clot Formation Time".

The ROTEM® TEMogram and its Parameters Fig. 2)

Amplitude of 0 mm means unobstructed rotation, while amplitude of 100 mm can be regarded as infinite firmness and blocking of the pin by the clot.

The result displayed on screen or printed out is called a "TEMogram" or "TEM".

{15}------------------------------------------------

Image /page/15/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features the department's name arranged in a circular pattern around a stylized emblem. The emblem consists of a stylized caduceus, a symbol often associated with healthcare, with a bird-like figure incorporated into its design.

Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993

Tem Innovations GmbH c/o Dr. Volker-Joacham Friemert Manager of Quality Managment and Regulatory Affairs Martin-Kollar-Strasse 13-15 81829 Munich, Germany

AUG 1 0 2011

Re: K101533

Trade/Device Name: EXTEM® Assay, FIBTEM® Assay, APTEM® Assay for the ROTEM® delta Thromboelastometry System Regulation Number: 21 CFR §864.5425 Regulation Name: Multipurpose system for in vitro coagulation studies Regulatory Class: Class II Product Code: JPA Dated: September 15, 2010

Dear Dr. Friemert:

adulteration.

Received: September 21, 2010

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice

{16}------------------------------------------------

Page 2 - Dr. Volker-Joacham Friemert

requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

Travis M. Chan

Maria M. Chan, Ph.D. Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{17}------------------------------------------------

Indications for Use

510(k) Number: K101533

Device Name: EXTEM assay

Indications for Use

The EXTEM assay is a semi-quantitative in vitro diagnostic assay on the ROTEM® delta Thromboelastometry System to monitor the coagulation process via the extrinsic pathway in citrated whole blood specimens.

Clotting characteristics are described by the functional parameters Clotting Time (CT), Speed of Clot Formation (CFT and alpha angle), Clot Firmness (A20/MCF) and Clot Lysis (LOT, ML, LI(x)).

Speed of clot formation time (CFT and alpha) is complementary parameter and should be used only in conjunction with the main parameters Clotting time (CT) and Clot Firmness (A20/MCF).

The indication for ROTEM® delta use is in adult patients where an evaluation of their blood coagulation properties is desired. Coaqulation evaluations with the ROTEM® delta system are commonly used to assess clinical conditions in organ transplantation, cardiovascular surgery, cardiology procedures and trauma to assess post-operative hemorrhage and / or thrombosis.

Prescription Use) × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Division Sign-Off
-------------------

Office of In Vitro Diagnostic
Device Evaluation and Safety

510(k)K101533
-----------------

510kl

{18}------------------------------------------------

Indications for Use

510(k) Number: KIOI533

Device Name: FIBTEM assay

Indications for Use

The FIBTEM assay is a semi-quantitative in vitro diagnostic assay on the ROTEM® delta Thromboelastometry System to monitor the clot firmness of a citrated whole blood specimens after blocking platelet contribution to the clot firmness The fib-TEM® reagent is always used in conjunction with ex-TEM® reagent.

Clotting characteristics are described by the functional parameter Clot Firmness (A20/MCF).

The indication for ROTEM® delta use is in adult patients where an evaluation of their blood coagulation properties is desired. Coagulation evaluations with the ROTEM® delta system are commonly used to assess clinical conditions in organ transplantation, cardiovascular surgery, cardiology procedures and trauma to assess post-operative hemorrhage and / or thrombosis.

Prescription Use) × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Division Sign-Off
---------------------

Office of In Vitro Diagnostic Device Evaluation and Safety

510(k)K101533
-----------------

{19}------------------------------------------------

Indications for Use

510(k) Number: K101533

Device Name: APTEM assay

Indications for Use

The APTEM assay is a semi-quantitative in vitro diagnostic assay on the ROTEM® delta Thromboelastometry System to monitor the clot firmness of a citrated whole blood specimens after blocking hyperfibrinolysis by aprotinin. The ap-TEM® reagent is always used in conjunction with ex-TEM® reagent.

Clotting characteristics are described by the functional parameters Clotting Time (CT), Speed of Clot Formation (CFT and alpha angle), Clot Firmness (A20/MCF) and Clot Lysis (LOT, ML, LI(x)).

Speed of clot formation time (CFT and alpha) is complementary parameter and should be used only in conjunction with the main parameters Clotting time (CT) and Clot Firmness (A20/MCF).

The indication for ROTEM® delta use is in adult patients where an evaluation of their blood coagulation properties is desired. Coagulation evaluations with the ROTEM® delta system are commonly used to assess clinical conditions in organ transplantation, cardiovascular surgery, cardiology procedures and trauma to assess post-operative hemorrhage and / or thrombosis.

Prescription Use) × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)
--------------------------------------------------------
Division Sign-Off
-------------------
Office of In Vitro Diagnostic
Device Evaluation and Safety
510(k)K101533
-----------------

510(k)

§ 864.5425 Multipurpose system for in vitro coagulation studies.

(a)
Identification. A multipurpose system for in vitro coagulation studies is a device consisting of one automated or semiautomated instrument and its associated reagents and controls. The system is used to perform a series of coagulation studies and coagulation factor assays.(b)
Classification. Class II (special controls). A control intended for use with a multipurpose system for in vitro coagulation studies is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.