K Number
K083842

Validate with FDA (Live)

Manufacturer
Date Cleared
2010-03-23

(454 days)

Product Code
Regulation Number
864.5425
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ROTEM® delta Thromboelastometry System is designed for invitro diagnostic use by professionals in a laboratory environment. The ROTEM® system is intended to be used to provide a qualitative and quantitative indication of the coagulation state of a blood sample. For this purpose the ROTEM® system records the clot firmness changes in a sample of citrated whole blood as the sample clots, retracts and Ivses in real time. The analyzer output consists of a qualitative graphical representation (mirrored coagulation curve - clot firmness over time) and several defined numerical parameters describing the curve quantitatively.

The ROTEM® system provides specific blood modifiers (so-called reagents) intended to be used with the system, as additive to the blood sample.

The results of the ROTEM® analysis should not be the sole basis for a patient diagnosis, but should be evaluated together with the patient's medical history (anamnesis), the clinical picture and, if necessary, further coagulations tests.

The ROTEM® delta is a non-invasive diagnostic instrument designed to monitor and analyze the coagulation state of a blood sample in order to assist in the assessment of patient clinical hemostasis conditions. The indication for ROTEM® delta use is with adult patients where an evaluation of their blood coagulation properties is desired.

Coagulation evaluations with the ROTEM® delta are commonly used to assess clinical conditions in organ transplantation, cardiovascular surgery, cardiology procedures and trauma to assess post-operative hemorrhage and / or thrombosis.

ROTROL N is a quality control material for monitoring accuracy and precision of tests carried out on the ROTEM® delta Thromboelastometry System.

ROTROL P is a quality control material for monitoring accuracy and precision of tests carried out on the ROTEM® delta Thromboelastometry System.

The in-TEM® assay is a semi-quantitative in vitro diagnostic assay used on the ROTEM® delta Thromboelastometry System to monitor the coagulation process via the intrinsic pathway in citrated whole blood specimens. Clotting characteristics are described by the functional parameters Clotting Time (CT), Speed of Clot Formation (CFT and alpha angle), Clot Firmness (A20/MCF) and Clot Lysis (LOT, ML, LI(x)).

The hep-TEM® assay is a semi-quantitative in vitro diagnostic assay used on the ROTEM® delta Thromboelastometry System to monitor the coagulation process, via the intrinsic pathway in the presence of unfractionated heparin, in citrated whole blood specimens. The hep-TEM® reagent is used to inactivate heparin in patients receiving unfractionated heparin. Clotting characteristics are described by the functional parameters Clotting Time (CT), Speed of Clot Formation (CFT and alpha angle), Clot Firmness (A20/MCF) and Clot Lysis (LOT, ML, LI(x)).

The NATEM assay is a semi-quantitative in vitro diagnostic assay used on the ROTEM® delta Thromboelastometry System to monitor the coagulation process, contact-activated by the surface of the measurement cell, in citrated whole blood specimens. Clotting characteristics are described by the functional parameters Clotting Time (CT), Speed of Clot Formation (CFT and alpha angle), Clot Firmness (A20/MCF) and Clot Lysis (LOT, ML, LI(x)).

The star-TEM® Reagent is intended for use as a recalcification reagent in the NATEM and INTEM assays on the ROTEM® delta Thromboelastometry System.

Device Description

The ROTEM® delta Thromboelastometry System consists of a fourcolumn instrument (with integrated computer module, computer controlled electronic pipette, software), system reagents (in-TEM® hep-TEM®, star-TEM®), quality controls (ROTROL N, ROTROL P) and measurement cells (Cup and Pin pro). The blood sample is filled into a cylindrical cup. A pin oscillates permanently while it is immersed in the blood holding cup. The motion of the pin is detected by an optical detection system. Data are processed and analyzed by a computer with special software. If no clotting takes place, the movement of the pin is not obstructed. When a clot forms and attaches itself to the pin and cup surfaces, the movement is obstructed. As the clot becomes firmer, the rotational movement of the pin is reduced. The rotational movement of the pin is converted into an amplitude with the following definitions applying to the thromboelastogram (TEM): An amplitude of 0 mm means unobstructed rotation, while an amplitude of 100mm can be regarded as infinite firmness and blocking of the pin by the clot. The TEM amplitude is a measure of the clot firmness.

AI/ML Overview

The document is a 510(k) summary for the ROTEM® delta Thromboelastometry System, seeking substantial equivalence to the predicate device, the Thrombelastograph® Coagulation Analyzer (TEG®) - 5000 Series. The study primarily focuses on comparing the performance of ROTEM® delta with the predicate device across several coagulation parameters.

The acceptance criteria for the ROTEM® delta were based on precision and a method comparison with the predicate TEG® 5000.

1. Table of Acceptance Criteria and Reported Device Performance

Parameter TypeAcceptance Criteria (Implicit from Study Description)Reported ROTEM® Delta Performance
PrecisionHigh precision expected in primary parameter; adequate precision in secondary parametersPrimary Parameter: Clot Firmness (MCF) - Shown to have high precision. Secondary Parameters: Coagulation Kinetics (CT, CFT, Alpha Angle) - Shown to have adequate precision.
Method Comparison with TEG®- Equality of clot firmness - Linear correlation for kinetic parameters (r > 0.8)Clot Firmness (MCF vs. MA): Demonstrated equality. Kinetic Parameters (CT vs. R, CFT vs. K, Alpha Angle vs. Angle): Showed a linear correlation (r > 0.8). It was noted that the slope of correlation was not equal to one and intercepts were not equal to zero, which was expected due to different activation reagents and distinct reference ranges between the two systems. Heparin Sensitivity: ROTEM® showed different heparin sensitivities compared to TEG®. ROTEM® was more sensitive for higher heparin concentrations, allowing analysis in cases where TEG® was unable to demonstrate coagulation.
Reference RangesReproducible and consistent across centersReference ranges for NATEM and INTEM were estimated using CSL1 protocols on three clinical US reference sample groups. These showed no significant center-to-center deviations and were in accordance with earlier studies on European reference groups.
Interfering SubstancesExpected dose-response curves and interferencesInvestigated for aprotinin, tranexamic acid, EACA. Dose-response curves investigated for heparin, dilution, and urokinase on the INTEM model, verifying diagnostic principles.

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the numerical sample size for the test set used in the method comparison study with the TEG® 5000. It mentions that "three clinical US reference sample groups" were used to estimate reference ranges for NATEM and INTEM. The provenance of this data is stated as US and the study appears to be prospective for the estimation of reference ranges. For the method comparison, it implies testing on various samples to establish correlations and differences in heparin sensitivity, but specific numbers are not provided.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This type of study does not rely on expert consensus or human interpretation to establish ground truth. It is a comparison of quantitative measurements from two medical devices. The "ground truth" implicitly refers to the accuracy and precision of the measurements themselves, and the correlation between the device and its predicate.

4. Adjudication Method for the Test Set

Not applicable. As a device comparison study involving quantitative measurements, there would be no adjudication method in the sense of expert review for image interpretation or diagnosis. The "adjudication" is based on statistical analysis of the measured parameters.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This study is for a Thromboelastometry System, which is an in-vitro diagnostic device for analyzing blood coagulation. It is not an AI-assisted diagnostic tool that involves human readers interpreting images or data. Therefore, an MRMC study or assessment of human reader improvement with/without AI assistance is not relevant.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

The ROTEM® delta is inherently a standalone device in the sense that the instrument's software processes the measurements and produces the quantitative results (e.g., CT, CFT, MCF, etc.) and graphical outputs. The performance described (precision, method comparison, sensitivity to interfering substances) represents the standalone performance of the device's measurement and processing capabilities. Clinicians then interpret these results, but the generation of the results themselves is automated.

7. The Type of Ground Truth Used

The ground truth for this device comparison is established by:

  • Predicate Device Measurements: The TEG® 5000 served as the predicate, and its measurements were essentially the "truth" against which the ROTEM® delta's measurements were compared for substantial equivalence.
  • Known Physiological Responses: The sensitivity to heparin, dilution, and lysis induced by urokinase are based on established physiological and pharmacological effects on coagulation, serving as an expected "truth" for how the device should respond.
  • Reference Ranges: Established through established clinical laboratory protocols (CSL1) on healthy reference populations.

8. The Sample Size for the Training Set

Not applicable. This type of device (Thromboelastometry System) does not typically involve "training sets" in the machine learning sense. The device's algorithms are based on established biophysical principles of coagulation detection and signal processing, not on iterative machine learning from a large dataset. Calibration and control materials are used for device setup and quality control, but these are distinct from training sets for AI algorithms.

9. How the Ground Truth for the Training Set was Established

Not applicable. As stated above, a training set for machine learning is not used for this type of device.

{0}------------------------------------------------

MAR 22 2010

510{k} Submission 2008/12/19

510(k) Summarv

K083842

1. Submitter's Name / Contact Person

Dr. Volker-Joachim Friemert Head of Quality Management and Regulatory Affairs Pentapharm GmbH Stahlgruberring 12 81829 Munich Germany

Contact:

Office:+49-89-454295-0
Dr. Volker-Joachim Friemert:+49-89-454295-11
Fax:+49-89-454295-22
E-mail:volker-joachim.friemert@rotem.de

2. Identification of the Product

Trade Name:ROTEM® delta Thromboelastometry System
Common Name:Whole Blood Haemostasis System
Classification Name:Multipurpose System for In Vitro CoagulationStudies

3. Identification of the Predicate

Thrombelastograph® Coagulation Analyzer (TEG®) - 5000 Series K002177, Product Code JPA, Haemoscope Corp.

4. Description of the Device

The ROTEM® delta Thromboelastometry System consists of a fourcolumn instrument (with integrated computer module, computer controlled electronic pipette, software), system reagents (in-TEM® hep-TEM®, star-TEM®), quality controls (ROTROL N, ROTROL P) and measurement cells (Cup and Pin pro). The blood sample is filled into a cylindrical cup. A pin oscillates permanently while it is immersed in the blood holding cup. The motion of the pin is detected by an optical detection system. Data are processed and analyzed by a computer with special software. If no clotting takes place, the movement of the pin is not obstructed. When a clot forms and attaches itself to the pin and cup surfaces, the movement is obstructed. As the clot becomes firmer, the rotational movement of the pin is reduced. The rotational

{1}------------------------------------------------

movement of the pin is converted into an amplitude with the following definitions applying to the thromboelastogram (TEM): An amplitude of 0 mm means unobstructed rotation, while an amplitude of 100mm can be regarded as infinite firmness and blocking of the pin by the clot. The TEM amplitude is a measure of the clot firmness.

5. Intended Use

The ROTEM® delta Thromboelastometry System is designed for invitro diagnostic use by professionals in a laboratory environment. The ROTEM® system is intended to be used to provide a qualitative and quantitative indication of the coagulation state of a blood sample. For this purpose the ROTEM® system records the clot firmness changes in a sample of citrated whole blood as the sample clots, retracts and Ivses in real time. The analyzer output consists of a qualitative graphical representation (mirrored coagulation curve - clot firmness over time) and several defined numerical parameters describing the curve quantitatively.

The ROTEM® system provides specific blood modifiers (so-called reagents) intended to be used with the system, as additive to the blood sample.

The results of the ROTEM® analysis should not be the sole basis for a patient diagnosis, but should be evaluated together with the patient's medical history (anamnesis), the clinical picture and, if necessary, further coagulations tests.

ROTEM® deltaTEG® 5000
InstrumentFully integratedThromboelastometryinstrumentThrombelastographyinstrument
Measuring TechniqueShear elasticity of acoagulating sampleby motion of pinShear elasticity of acoagulating sampleby motion of cup
Measuring Channels42
Signal GenerationOscillating pin instationary cupOscillation cuparound stationary pin
Signal TransducerOptical 4 CCD chipsElectrical-mechanicaltransducer
Temperature Control30-40 ℃20-40 ℃
Sample Volume300 μl360 μl
Total Reaction Volume320-340 μl360-380 μl
Voltage110-230 V, 50-60 Hz120 V, 60 Hz and 220 V, 50 Hz model available
EnvironmentStable and level surface.Operating temperature: 15-30 ℃.Storage temperature 0-50 °C,rel. humidity 20-85%.Vibration free position, no solar radiation.Operating temperature: 10-35 °C.Storage temperature -30-50 °C,rel. humidity 20-80%
Operating PositionLevel to 5°inclinationLevel ad justed by leveling feet and level
Operable to Height above Sea Level2000 m (6560 ft)3048 m (10000 ft)
Initial Warm Up Time5-15 min (depending on room temperature)5 min
PipettingElectronic pipette (20-320 μl)Manual pipettes (10, 20, 360, 1000 μl)
Reagents / Accessories:in-TEM® (ellagic acid)Kaolin
Intrinsic Contact Activation Reagent
Heparinase I Reagent (for heparin neutralization)hep-TEM® (heparinase I, CaCl2)Cup (single use) coated with heparinase IorHeparinase I reagent CaCl2 reagent
Calcium Chloride Reagent (for recalcification)star-TEM® (CaCl2)
Quality Control Material (level I control)ROTROL NTEG® coagulation control - level I
Quality Control Material (level II control)ROTROL PTEG® coagulation control - level II
Cups & PinsCup & Pin pro (Acrylic plastic)Cups & Pins (Acrylic plastic)

6. Summary of Technological Characteristics of the Product, Compared with the Predicate Device

{2}------------------------------------------------

{3}------------------------------------------------

7. Executive Summary of the Study Report

In this study, the performance characteristics of the ROTEM® delta Whole Blood Haemostasis System (ROTEM®) were investigated and three assays were compared to the predicate device TEG® 5000 (TEG®) in support of a 510(k) submission for ROTEM®. The three ROTEM® tests NATEM, INTEM and HEPTEM were compared to their predicate tests on TEG®. The NATEM test represents the classical non-activated thrombelastographic method corresponding to the non-activated TEG®. The INTEM is activated by the intrinsic activator reagent ellagic acid corresponding to the intrinsically activated Kaolin test on TEG®. The HEPTEM is activated as the INTEM by ellagic acid and additionally contains heparinase for heparin neutralization corresponding to the Kaolin + Heparinase test of TEG®.

For all three tests, the ROTEM® was shown to have high precision in its primary parameter clot firmness and adequate precision in its secondary coagulation kinetics parameters.

The method comparison with TEG® showed equality of the clot firmness (MCF vs. MA). The kinetic parameters (CT vs. R, CFT vs. K, Alpha Angle vs. Angle) showed a linear correlation between ROTEM' and TEG® (r>0.8). As expected, slope of the correlation between the two systems was not equal to one and intercepts were not equal to zero in the kinetic parameters as the respective tests use different activation reagents and each test has its own distinct reference ranges. Heparin sensitivities of the respective intrinsic ROTEM® and TEG® tests also differed. ROTEM® was more sensitive for higher heparin concentrations than TEG® and allowed for an analysis of heparin concentrations normally seen during cardiovascular surgery; while TEG® was unable to demonstrate any coagulation in these cases.

Reference ranges for the ROTEM® tests NATEM and INTEM were estimated using CSL1 protocols on three clinical US reference sample aroups. The reference ranges determined showed no significant center-to-center deviations and were in accordance with the reference ranges determined in earlier studies on European reference sample groups.

Three interfering substances widely used in coagulation management, the antifibrinolytic drugs aprotinin, tranexamic acid and epsilon-amino caproic acid (EACA) were investigated. Dose-response curves were investigated for heparin, for dilution and for urokinase on the INTEM model in order to verify the diagnostic principles of thrombelastographic methods on ROTEM®.

{4}------------------------------------------------

In summary, ROTEM® is a precise Whole Blood Haemostasis System with the typical performance characteristics of a thrombelastographic method (aprotinin interference on the contact activated and intrinsically activated tests, heparin sensitivity of the intrinsic test and sensitivity to dilution and lysis induced by urokinase in-vitro). Its reference ranges are reproducible from center to center. The method comparison with TEG® shows equality of the primary parameter clot firmness and a linear regression and good correlation in the secondary kinetic parameters. In comparison to the TEG® the ROTEM® is optimized for a rapid diagnosis (employing stronger activation for shorter activation and test times) accompanied by a shift of the detection range of heparin in the intrinsically activated test.

In aggregate the data presented in this report demonstrate that the ROTEM® system and the three assays described are substantially equivalent to the predicate TEG® System and corresponding assays.

8. Conclusion (Statement of Equivalence)

The data and information provided in this submission support a substantial equivalence determination, and, therefore, clearance of the 510(k) premarket notification for the ROTEM® delta Thromboelastometry System.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.

Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002

Pentapharm GinbH c/o Dr. Volker-Joachim Friemert Head of Quality Management and Regulatory Affairs Stahlgruberring 12 Munich, Germany 81829

MAR 2 3 2010

Re: K083842 Trade/Device Name: ROTEM® delta Thromboelastometry System: in-TEM® Assay hep-TEM® Assay NATEM® Assay Star-TEM® ROTEM® delta instrument ROTROL N ROTROL P Regulation Number: 21 CFR 864.5425 Regulation Name: Multipurpose system for in vitro coagulation studies Regulatory Class: Class II Product Code: JPA. GGN Dated: July 27, 2009 Received: July 29, 2009

Dear Dr. Friemert:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21,

{6}------------------------------------------------

Page 2 - Dr. Friemert

Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Maria M Chan

Maria M. Chan, Ph.D. Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

Indications for Use

510(k) Number (if known): KO83842

Device Name:

ROTEM® delta Thromboelastometry System

Indications for Use

The ROTEM® delta is a non-invasive diagnostic instrument designed to monitor and analyze the coagulation state of a blood sample in order to assist in the assessment of patient clinical hemostasis conditions. The indication for ROTEM® delta use is with adult patients where an evaluation of their blood coagulation properties is desired.

Coagulation evaluations with the ROTEM® delta are commonly used to assess clinical conditions in organ transplantation, cardiovascular surgery, cardiology procedures and trauma to assess post-operative hemorrhage and / or thrombosis.

Prescription Use) (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)

Division Sign-Off

Office of In Vitro Diagnostic
Device Evaluation and Safety

510(k) K083842

{8}------------------------------------------------

510(k) Submission 2008/12/19

Indications for Use

CONFIDENTIAL

510(k) Number (if known): K083842

Device Name: ROTROL N

Indications for Use

ROTROL N is a quality control material for monitoring accuracy and precision of tests carried out on the ROTEM® delta Thromboelastometry System.

Prescription Use) × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Division Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safety

510/k

{9}------------------------------------------------

510(k) Submission 2008/12/19

Indications for Use

CONFIDENTIAL

510(k) Number (if known): K083842

Device Name: ROTROL P

Indications for Use

ROTROL P is a quality control material for monitoring accuracy and precision of tests carried out on the ROTEM® delta Thromboelastometry System.

' Prescription Use) × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)

Division Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K083842

{10}------------------------------------------------

CONFIDENTIAL

Indications for Use

510(k) Number: K083842 ·

Device Name: in-TEM® Assay

Indications for Use

The in-TEM® assay is a semi-quantitative in vitro diagnostic assay used on the ROTEM® delta Thromboelastometry System to monitor the coagulation process via the intrinsic pathway in citrated whole blood specimens. Clotting characteristics are described by the functional parameters Clotting Time (CT), Speed of Clot Formation (CFT and alpha angle), Clot Firmness (A20/MCF) and Clot Lysis (LOT, ML, LI(x)).

The indication for ROTEM® delta use is with adult patients where an evaluation of their blood coagulation properties is desired. Coagulations with the ROTEM® delta are commonly used to assess clinical conditions in organ transplantation, cardiovascular surgery, cardiology procedures and trauma to assess post-operative hemorrhage and / or thrombosis.

Prescription Use) × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Division Sign-Off

Office of In Vitro Diagnostic

510(k) K083842

{11}------------------------------------------------

CONFIDENTIAL

Indications for Use

510(k) Number: K083842

Device Name: hep-TEM® Assay

Indications for Use

The hep-TEM® assay is a semi-quantitative in vitro diagnostic assay used on the ROTEM® delta Thromboelastometry System to monitor the coagulation process, via the intrinsic pathway in the presence of unfractionated heparin, in citrated whole blood specimens. The hep-TEM® reagent is used to inactivate heparin in patients receiving unfractionated heparin. Clotting characteristics are described by the functional parameters Clotting Time (CT), Speed of Clot Formation (CFT and alpha angle), Clot Firmness (A20/MCF) and Clot Lysis (LOT, ML, LI(x)).

The indication for ROTEM® delta use is with adult patients where an evaluation of their blood coagulation properties is desired. Coagulations with the ROTEM® delta are commonly used to assess clinical conditions in organ transplantation, cardiovascular surgery, cardiology procedures and trauma to assess post-operative hemorrhage and / or thrombosis.

Prescription Use) × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Division Sign-Off

510(k) K083842

{12}------------------------------------------------

510{k} Submission 2008/12/19

Indications for Use

CONFIDENTIAL

510(k) Number: K083842

Device Name: NATEM Assay

Indications for Use

The NATEM assay is a semi-quantitative in vitro diagnostic assay used on the ROTEM® delta Thromboelastometry System to monitor the coagulation process, contact-activated by the surface of the measurement cell, in citrated whole blood specimens. Clotting characteristics are described by the functional parameters Clotting Time (CT), Speed of Clot Formation (CFT and alpha angle), Clot Firmness (A20/MCF) and Clot Lysis (LOT, ML, LI(x)).

The indication for ROTEM® delta use is with adult patients where an evaluation of their blood coagulation properties is desired. Coagulations with the ROTEM® delta are commonly used to assess clinical conditions in organ transplantation, cardiovascular surgery, cardiology procedures and trauma to assess post-operative hemorrhage and / or thrombosis.

Prescription Use) X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Division Sign-Off

Office of In Vitro Diggnostic ice Evaluation and Safety

510(k) K083842

{13}------------------------------------------------

CONFIDENTIAL

Indications for Use

510(k) Number: K083842

Device Name: Star-TEM® Reagent

Indications for Use

The star-TEM® Reagent is intended for use as a recalcification reagent in the NATEM and INTEM assays on the ROTEM® delta Thromboelastometry System.

The indication for ROTEM® delta use is with adult patients where an evaluation of their blood coagulation properties is desired. Coagulations with the ROTEM® delta are commonly used to assess clinical conditions in organ transplantation, cardiovascular surgery, cardiology procedures and trauma to assess post-operative hemorrhage and / or thrombosis.

Prescription Use) × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Division Sign-Off

K083842
510(k)

§ 864.5425 Multipurpose system for in vitro coagulation studies.

(a)
Identification. A multipurpose system for in vitro coagulation studies is a device consisting of one automated or semiautomated instrument and its associated reagents and controls. The system is used to perform a series of coagulation studies and coagulation factor assays.(b)
Classification. Class II (special controls). A control intended for use with a multipurpose system for in vitro coagulation studies is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.