K Number
K083823

Validate with FDA (Live)

Manufacturer
Date Cleared
2009-03-18

(86 days)

Product Code
Regulation Number
878.3300
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Securian™ Tissue Reinforcement Matrix is indicated for the management and protection of soft tissue and tendon injuries in which there has been no substantial loss of tendon tissue. Securian™ Tissue Reinforcement Matrix minimizes tissue attachment to the device in case of direct contact with overlying tissues. Securian™ Tissue Reinforcement Matrix is also indicated for reinforcement of soft tissues repaired by suture anchors during tendon repair surgery including reinforcement of the rotator cuff, patellar, Achilles, biceps, quadriceps, or other tendons. Securian™ Tissue Reinforcement Matrix is not intended to replace normal body structure or provide the full mechanical strength to support tendon repair of the rotator cuff, patellar, Achilles, biceps, quadriceps, or other tendons. Sutures used to repair the tear and sutures or bone anchors used to attach the tissue to the bone provide biomechanical strength for the tendon repair. Securian™ Tissue Reinforcement Matrix reinforces and protects healing soft tissues. The device is indicated for open and endoscopic procedures. Securian™ Tissue Reinforcement Matrix is intended for one-time use.

Device Description

Securian™ Tissue Reinforcement Matrix is a sterile product composed of microbial-derived cellulose. The implantable device is presented in a sterile double-pouched package for appropriate removal in preparation for surgery. The non-resorbable surgical mesh is used for trauma and reconstructive surgical procedures involving tendons or other soft tissues. Securian™ Tissue Reinforcement Matrix maintains the relative position and stability of soft tissues during the healing period.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and study for the Securian™ Tissue Reinforcement Matrix:

It's important to note that the provided documents are a 510(k) summary and an FDA clearance letter for a medical device. This type of regulatory submission typically focuses on demonstrating substantial equivalence to a previously cleared predicate device, rather than presenting a novel clinical study with new acceptance criteria for a first-of-its-kind device. Therefore, many of the requested categories for a new AI/software device study (like sample size for test set, number of experts, adjudication methods, MRMC studies, standalone performance, training set details) are not applicable to this type of submission.

The "study" in this context refers to the demonstration of substantial equivalence through a comparison to predicate devices and performance testing.


Acceptance Criteria and Device Performance

Acceptance CriteriaReported Device Performance
Substantial Equivalence to Predicate Devices (K081882 XYLOS Surgical Mesh, K023237 Ortho-Wrap® Bioresorbable Sheet, K072190 Ortho-Wrap® Bioresorbable Sheet, K021056 Permacol™ Biological Implant): The device must demonstrate identical or similar technological characteristics, principles of operation, and achieve comparable performance and safety as the legally marketed predicate devices."Securian™ Tissue Reinforcement Matrix is identical to previously cleared K081882 XYLOS Surgical Mesh. Therefore Securian™ Tissue Reinforcement Matrix has the same biocompatibility and performance characteristics to those cited in K081882." "The Securian™ Tissue Reinforcement Matrix is substantially equivalent to legally marketed predicate devices and has the same intended uses and the same or similar indications, technological characteristics, and principles of operation as the predicate devices." "Performance data demonstrate that Securian™ Tissue Reinforcement Matrix functions equivalently to the predicate devices. Thus, Securian™ Tissue Reinforcement Matrix is substantially equivalent to the predicate devices." "Securian™ Tissue Reinforcement Matrix met the test requirements, providing reasonable assurance of device performance for its intended use and supporting substantial equivalence."
Biocompatibility: Must be biocompatible for its intended use."Securian™ Tissue Reinforcement Matrix has the same biocompatibility... to those cited in K081882 [XYLOS Surgical Mesh]." (Implied acceptance based on predicate)
Performance for Modified Indications for Use: The device must meet test requirements supporting its modified 'Indications for Use' statement without posing new questions of safety and effectiveness."Additional performance testing was performed to support the modified 'Indications for Use' statement." "Securian™ Tissue Reinforcement Matrix met the test requirements, providing reasonable assurance of device performance for its intended use and supporting substantial equivalence." "The modified 'Indications for Use' in this submission do not pose new questions of safety and effectiveness."

Study Details (Based on the provided 510(k) Summary)

  1. Sample size used for the test set and the data provenance:

    • Sample Size: Not explicitly stated as a separate clinical test set of patients. The "performance data" refers to comparative testing against predicate devices. The study doesn't involve human clinical trials with a patient test set in the traditional sense for demonstrating primary safety and efficacy, but rather bench testing and comparison to existing data from predicate devices.
    • Data Provenance: Not applicable in the context of a de novo clinical data collection from a test set. The data provenance would be from internal performance testing and data cited in the predicate device clearances.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. This submission relies on demonstrating similarity to already approved devices. There is no mention of "ground truth" established by experts for a novel test set in the provided document.
  3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not applicable. There was no independent clinical test set requiring expert adjudication.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This device is a physical surgical mesh, not an AI/software device used by human readers.
  5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is a physical device, not an algorithm.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Not applicable in the AI/software context. For a physical device, "ground truth" would relate to established scientific and engineering principles, biocompatibility standards, and the performance characteristics (e.g., tensile strength, pore size, degradation rate) of the predicate devices. The "truth" is established by the successful prior clearance of the predicate devices.
  7. The sample size for the training set:

    • Not applicable. This device does not use machine learning or a training set.
  8. How the ground truth for the training set was established:

    • Not applicable. This device does not use machine learning or a training set.

{0}------------------------------------------------

MAR 1 8 2009

Section 5: 510(k) Summary

paye K083823

The following information is provided as required by 21 CFR § 807.87 for Securian™ Tissue Reinforcement Matrix 510(k) premarket notification. In response to the Safe Medical Devices Act of 1990, the following is a summary of the safety and effectiveness information upon which the substantial equivalence determination is based.

The safety and effectiveness of Securian™ Tissue Reinforcement Matrix is based upon a determination of the substantial equivalence as well as the safety and effectiveness of its predicate devices.

Sponsor:

Contact:

Xylos Corporation 838 Town Center Drive Langhorne, PA 19047

Joyce Elkins 838 Town Center Drive Langhorne, PA 19047 Phone: 215-867-0220 ext. 642 Facsimile: 215-741-4804 joyce.elkins@xyloscorp.com

Date of Submission: December 19, 2008 Proprietary Name: Securian™ Tissue Reinforcement Matrix Common Name: Surgical Mesh Regulatory Class: Class II Product Codes: FTM XYLOS® Surgical Mesh, K081882 / XYLOS Surgical Mesh, Predicate Device(s): K023237 Ortho-Wrap® Bioresorbable Sheet, K072190 (Ortho-Wrap® Trademark of MAST Biosurgery, Inc.) Permacol™, K021056 (Permacol™ Biological Implant Trademark of Tissue

Science Laboratories, PLC)

Device Description:

Securian™ Tissue Reinforcement Matrix is a sterile product composed of microbial-derived The implantable device is presented in a sterile double-pouched package for cellulose.

CONFIDENTIAL

PAGE 11

{1}------------------------------------------------

appropriate removal in preparation for surgery. The non-resorbable surgical mesh is used for trauma and reconstructive surgical procedures involving tendons or other soft tissues. Securian™ Tissue Reinforcement Matrix maintains the relative position and stability of soft tissues during the healing period.

Indications for Use:

Securian™ Tissue Reinforcement Matrix is indicated for the management and protection of soft tissue and tendon injuries in which there has been no substantial loss of tendon tissue. SecurianTM Tissue Reinforcement Matrix minimizes tissue attachment to the device in case of direct contact with overlying tissues. Securian™ Tissue Reinforcement Matrix is also indicated for reinforcement of soft tissues repaired by sutures or suture anchors during tendon repair surgery including reinforcement of the rotator cuff, patellar, Achilles, biceps, quadriceps, or other tendons. Securian™ Tissue Reinforcement Matrix is NOT intended to replace normal body structure or provide the full mechanical strength to support tendon repair of the rotator cuff, patellar, Achilles, biceps, quadriceps, or other tendons. Sutures used to repair the tear and sutures or bone anchors used to attach the tissue to the bone provide biomechanical strength for the tendon repair. Securian™ Tissue Reinforcement Matrix reinforces and protects healing soft tissues. The device is indicated for open and endoscopic procedures. Securian™ Tissue Reinforcement Matrix is intended for one-time use.

Technological Characteristics and Substantial Equivalence:

Securian™ Tissue Reinforcement Matrix is identical to previously cleared K081882 XYLOS Surgical Mesh. Therefore Securian™ Tissue Reinforcement Matrix has the same biocompatibility and performance characteristics to those cited in K081882. For this 510(k) submission, the device name is changed from "XYLOS® Surgical Mesh" K081882 to "Securian™ Tissue Reinforcement Matrix" and the 'Indications for Use' are modified from those cleared in K081882.

The Securian™ Tissue Reinforcement Matrix is substantially equivalent to legally marketed predicate devices and has the same intended uses and the same or similar indications, technological characteristics, and principles of operation as the predicate devices. Performance data demonstrate that Securian™ Tissue Reinforcement Matrix functions

CONFIDENTIAL

PAGE 12

{2}------------------------------------------------

08383 3

equivalently to the predicate devices. Thus, Securian™ Tissue Reinforcement Matrix is substantially equivalent to the predicate devices.

Discussion of performance testing:

Securian™ Tissue Reinforcement Matrix was subjected to a panel of tests to demonstrate its equivalence to predicate devices. Additional performance testing was performed to support the modified 'Indications for Use' statement.

Securian™ Tissue Reinforcement Matrix met the test requirements, providing reasonable assurance of device performance for its intended use and supporting substantial equivalence.

Conclusion:

Securian™ Tissue Reinforcement Matrix is substantially equivalent to all the predicate devices. The modified 'Indications for Use' in this submission do not pose new questions of safety and effectiveness.

CONFIDENTIAL

PAGE 13

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the eagle. The eagle is black, and the text is also black. The logo is simple and recognizable.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAR 1 8 2009

Xylos Corporation % Ms. Joyce Elkins Director Regulatory Affairs/Quality Assurance 838 Town Center Drive Langhorne, Pennsylvania 19047

Re: K083823

Trade/Device Name: Securian™ Tissue Reinforcement Matrix Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical Mesh Regulatory Class: II Product Code: FTM Dated: December 30, 2008 Received: January 5, 2009

Dear Ms. Elkins:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set

{4}------------------------------------------------

Page 2 - Ms. Joyce Elkins

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

L.Aymed for.

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

K083823 510(k) Number (if known):

Device Name: Securian™ Tissue Reinforcement Matrix

Indications For Use:

The Securian™ Tissue Reinforcement Matrix is indicated for the management and protection of soft tissue and tendon injuries in which there has been no substantial loss of tendon tissue. Securian™ Tissue Reinforcement Matrix minimizes tissue attachment to the device in case of direct contact with overlying tissues. Securian™ Tissue Reinforcement Matrix is also indicated for reinforcement of soft tissues repaired by suture anchors during tendon repair surgery including reinforcement of the rotator cuff, patellar, Achilles, biceps, quadriceps, or other tendons. Securian™ Tissue Reinforcement Matrix is not intended to replace normal body structure or provide the full mechanical strength to support tendon repair of the rotator cuff, patellar, Achilles, biceps, quadriceps, or other tendons. Sutures used to repair the tear and sutures or bone anchors used to attach the tissue to the bone provide biomechanical strength for the tendon repair. Securian™ Tissue Reinforcement Matrix reinforces and protects healing soft tissues. The device is indicated for open and endoscopic procedures. Securian™ Tissue Reinforcement Matrix is intended for one-time use.

Prescription Use(Part 21 CFR 801 Subpart D)XAND/OROver-The-Counter Use(21 CFR 807 Subpart C)X
---------------------------------------------------------------------------------------------------------------

Daniel Kra for MXM March 13, 2009

(Division Sign-Off)

Division of General, Restorative,
and Neurological Devices

510(k) Number K083823

§ 878.3300 Surgical mesh.

(a)
Identification. Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.(b)
Classification. Class II.