Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K100313
    Manufacturer
    Date Cleared
    2011-06-14

    (495 days)

    Product Code
    Regulation Number
    878.3300
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K072190,K090778

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MiMedx HydroFix™ Surgical Sheet is indicated for the management and protection of tendon injuries in which there has been no substantial loss of tendon tissue. The surgical sheet minimizes tissue attachment to the device in case of direct contact with the tissues.

    Device Description

    The HydroFix™ Surgical Sheet is a flexible sheet of 30 Wt.% polyvinyl alcohol (PVA) material with dimensions 60 ±6 mm X 50 ±5 mm and 60 ±6 mm X 100 ±10 mm with a thickness of 1.0 ±0.2 mm. The comers of the sheet are rounded. There are no holes or perforations. There are no markings on either side of the sheet, raised (embossed) or printed. The sheet is provided sterile and hydrated in saline solution.

    The HydroFix™ Surgical Sheet will be provided in other shapes and sizes as needed for particular surgical procedures.

    AI/ML Overview

    This document describes the 510(k) submission for the HydroFix™ Surgical Sheet, which aims to demonstrate its substantial equivalence to previously cleared devices. It primarily focuses on comparing the new device's performance and characteristics to those of predicate devices rather than establishing specific acceptance criteria for standalone performance.

    Here's an analysis of the provided information based on your requested points:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state numerical acceptance criteria with pass/fail thresholds for the HydroFix™ Surgical Sheet. Instead, it states that the device's performance was judged as "comparable to the predicate devices."

    Performance CriterionReported Device Performance (vs. Predicate)
    In Vitro Testing
    Suture Pull outComparable
    Tensile StrengthComparable
    Burst StrengthComparable
    Tear resistanceComparable
    In Vivo Testing
    Tissue attachmentComparable
    Ability to suture to tissueComparable
    Ability to cut the sheetComparable
    Ability to secure to tissueComparable
    Ability to manage and protect tendon injuriesComparable

    2. Sample size used for the test set and the data provenance

    • Sample size for test set: Not specified for in vitro tests. For in vivo tests, it mentions studies conducted in "both sheep and rabbit models," implying a sample size of at least two different animal types, but the number of individual animals within each model is not provided.
    • Data provenance: Not explicitly stated (e.g., country of origin, retrospective/prospective). However, in vivo studies typically imply prospective animal studies conducted in a controlled environment.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not provided. The document refers to "in vitro" and "in vivo" testing, implying laboratory and animal model evaluations, not expert-based ground truth for interpretation as might be found in diagnostic imaging studies.

    4. Adjudication method for the test set

    This information is not provided, as the evaluation type doesn't involve human interpretation that would require an adjudication method.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, an MRMC comparative effectiveness study was not done. This device is a surgical sheet, not an AI-assisted diagnostic tool, so such a study would not be applicable.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    This concept is not applicable. The device is a surgical sheet, not an algorithm. The testing performed was to evaluate the physical and biological characteristics of the sheet.

    7. The type of ground truth used

    • For In Vitro Testing: The ground truth is inherent in the physical properties and mechanical measurements themselves (e.g., actual tensile strength, burst strength, etc., measured according to established material testing standards).
    • For In Vivo Testing: The ground truth is the observed biological response and mechanical performance within the animal models as assessed by researchers (e.g., visual assessment of tissue attachment, successful suturing, etc.). This would likely involve histological examination and gross observations post-sacrifice.

    8. The sample size for the training set

    This concept is not applicable. The device is a physical product (surgical sheet), not a machine learning algorithm that requires a training set.

    9. How the ground truth for the training set was established

    This concept is not applicable, as there is no training set for a physical surgical sheet.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1