Intended Use

The Exsudex™ Wound Drainage Device is indicated for patients who would benefit from the application of negative pressure to the area of a wound to promote wound healing by drainage of fluids and infected materials from the wound and wound bed.

The Exsudex™ Wound Drainage Device is a compact, portable device indicated for patients who would benefit from the application of negative pressure to the area of a wound, for the aspiration and removal of surgical fluids, irrigation fluids, tissue (including bone), gases, bodily fluids or infectious materials either during surgery or at the patient's bedside particularly as the device may promote wound healing.

Device Description

The Exsudex™ unit is a compact portable suction device that can be powered by either an internal battery pack or wall current and is compatible with U.S. electrical standards. The pump will be used in combination with accessory kits to create localized topical negative pressure and promote wound healing by drainage of fluids and infected materials from the wound and wound bed into a disposable canister.

The Exsudex™ consists of a medium sized housing that contains a vacuum pump and control system with a location for a fluid canister system. There is also a bacterial filter which is plugged directly into a housing in the pump and replaced as necessary. Accessories include a bed clamp, infusion pole clamp and carry bag. For the expanded indication of use for wound healing the unit is used in conjunction with accessory kits conforming to the guidelines for dressings developed by Chariker et al already available in the U.S. market. This kit consists of individually reviewed medical components that have added instructions for use. The kits are considered "Convenience Kits", falling under the General Surgical classification (see 21 CFR 878.4800.)

To use the pump, the suction tube must be connected to the canister and to the drain.

After checking the set up and confirming connection between the drain, the Exsudex™ is activated by pressing the start/stop button.

To ensure the maintenance of the desired level pressure, Exsudex™ monitors and controls:

  • . The buildup of pressure
  • . pressure loss
  • . pressure tolerance

The Exsudex™ is intended for acute, institutional, and long-term care use and only when prescribed by a physician. Product may be indicated for patient home-care, when carefully monitored by licensed home health-care providers, and only when prescribed by a physician.

AI/ML Overview

The provided text is a 510(k) Summary for the Exsudex™ Wound Drainage Pump. It primarily focuses on demonstrating substantial equivalence to predicate devices rather than detailing a specific clinical study with structured acceptance criteria and device performance metrics in the way a clinical trial report would.

Therefore, many of the requested sections about specific acceptance criteria and detailed study results cannot be fully populated from this document. The document describes technological characteristics, functional performance and electrical leakage testing, and states "extensive clinical use in Europe." It doesn't present a randomized controlled trial or a standalone performance study with quantifiable acceptance criteria.

However, I can extract the information that is present and highlight what is not available.

1. A table of acceptance criteria and the reported device performance

The document does not explicitly state quantifiable acceptance criteria with corresponding reported device performance in a table format. The primary "performance" is stated as having "the same intended uses as the predicate devices, with similar technological characteristics" and that "The system performs as intended and raises no new safety or effectiveness issues."

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document mentions "Verification and validation testing of the Exsudex™ device, including functional performance testing and electrical leakage testing." It does not provide:

  • A specific sample size for a test set (e.g., number of patients, number of cases).
  • Data provenance beyond "extensive clinical use in Europe." It does not specify whether this was retrospective or prospective data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not provided in the document. Ground truth establishment, expert review, or qualifications are not detailed for any testing related to clinical performance. The document refers to "established design control procedures" for functional and electrical testing, which typically do not involve expert ground truth in a clinical sense.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not provided in the document.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No MRMC study is mentioned. This device is a powered suction pump, not an AI-assisted diagnostic or therapeutic device for which "human readers improve with AI" would be applicable.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

The document describes functional performance testing and electrical leakage testing. While these are standalone tests of the device itself, they are not a "standalone algorithm" performance study in the context of AI/ML devices. The device is a physical pump, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For the "Verification and validation testing of the Exsudex™ device, including functional performance testing and electrical leakage testing," the ground truth would likely be established engineering specifications and regulatory standards (e.g., electrical safety standards, vacuum pressure targets). The document does not specify these explicitly but implies adherence to them. For the "extensive clinical use in Europe," no specific ground truth methodology for clinical outcomes is described.

8. The sample size for the training set

This information is not provided. The device is not an AI/ML algorithm that would typically have a "training set" in the conventional sense.

9. How the ground truth for the training set was established

This information is not provided, as it's not relevant to this type of device.

{0}------------------------------------------------

OCT 0 8 2008

:2062311 page 1/5

510(k) Summary:

Applicant Information:

Name:The Medical Company
Address:P.O. Box 21163800 CC AmersfoortThe Netherlands
Phone:+31 33 465 0113
Fax:+31 33 465 2191

Email: info@themedicalcompany.nl

Contact: Mr. Harm Jaap Smit

Establishment registration number:

not available yet

Manufacturer:

Name:MediTop BV
Address:Vlasakker 223417 XT MontfoortThe Netherlands
Phone:+31 348 566 100
Fax:+31 348 566 119
Email:info@meditop.nl

Contact: Mr. Gé van Breukelen

Trade Name:

Wound Drainage Pump Exsudex™

Device Classification:

Powered Suction Pump Regulation per FDA 21 CR 878.4780 Class II Product code: BTA Review Panel: General & Plastic Surgery

{1}------------------------------------------------

K082311 page 2/5

  • i. Device name: Exsudex™ Negative Pressure Wound Drainage pump
  • ii. Classification name: Powered suction pump (per 21 CFR 878.4780)
  • iii. Substantial equivalence:

V.A.C.'" Plus 510(k) No.K992448 Ambulatory V.A.C. 510(k) No.K971548 Medela Vario 5 10(k) No. K983552 Versatile 1 510(k) No. K052456

iv. Device description:

The Exsudex™ unit is a compact portable suction device that can be powered by either an internal battery pack or wall current and is compatible with U.S. electrical standards. The pump will be used in combination with accessory kits to create localized topical negative pressure and promote wound healing by drainage of fluids and infected materials from the wound and wound bed into a disposable canister.

The Exsudex™ consists of a medium sized housing that contains a vacuum pump and control system with a location for a fluid canister system. There is also a bacterial filter which is plugged directly into a housing in the pump and replaced as necessary. Accessories include a bed clamp, infusion pole clamp and carry bag. For the expanded indication of use for wound healing the unit is used in conjunction with accessory kits conforming to the guidelines for dressings developed by Chariker et al already available in the U.S. market. This kit consists of individually reviewed medical components that have added instructions for use. The kits are considered "Convenience Kits", falling under the General Surgical classification (see 21 CFR 878.4800.)

To use the pump, the suction tube must be connected to the canister and to the drain.

After checking the set up and confirming connection between the drain, the Exsudex™ is activated by pressing the start/stop button.

To ensure the maintenance of the desired level pressure, Exsudex™ monitors and controls:

  • . The buildup of pressure
  • . pressure loss
  • . pressure tolerance

The Exsudex™ is intended for acute, institutional, and long-term care use and only when prescribed by a physician. Product may be indicated for patient home-care, when carefully monitored by licensed home health-care providers, and only when prescribed by a physician.

CAUTION

Federal law restricts this device to sale by or on the order of a physician.

Canisters

The machine may be used with a disposable canister which is intended for medical use and is registered appropriately. The canister is disposable and can be supplied in a range of sizes up to a maximum fluid capacity of 1300cc and contains an overflow shut off valve. It is secured to the Exsudex™ unit by means of a slide-fit housing on the back of the pump.

The canister intended for primary use with the device is a 1200cc disposable canister, with a secure press-on lid, designed to tightly affix during transport and disposal, eliminating the risk of cross-contamination. The canister lid has a built-in pour spout. The canister is equipped with a filter with positive shut-off value that effectively traps aerosolized microorganisms and particulate matter and that prevents fluid overlow. The canister is easy to read with etched and printed graduations which ensure total volume accuracy and easy-to-read measurements. The canister has a smudge-free surface which provides adequate space for marking patient information. The canister is non-sterile and is designed to be discarded when full, partially full, or on completion of patient care.

{2}------------------------------------------------

K082311 page 3/5

Single use disposables

For the indication of use for wound healing, the device is used in conjunction with accessory kits conforming to the guidelines developed by Katherine Jeter and Dr. Mark Chariker. Similar kits are already available in the U.S. market. The kits consist of individually reviewed medical components from existing manufacturers that have met the regulatory requirements of the FDA (see 21 CFR 878.4800) and include separate, additional instructions for use.

The Kits include:

    1. Silicone or Wound drain
    1. Connecting tubing
    1. Impregnated gauze
    1. Gauze fill material
    1. Polyurethane Cover dressing
    1. Skin prep

All products, including canister and disposable kit of individual components are not designed for cleaning or reuse and should be properly discarded after patient use. The sale or use of this device must be by or on the order of a physician.

Indications for use

The Exsudex™ Wound Drainage Device is indicated for patients who would benefit from the application of negative pressure to the area of a wound to promote wound healing by drainage of fluids and infected materials from the wound and wound bed.

Effective Management of Incisional and Cutaneous Fistuiae With Closed Suction Wound Drainage, Mark E. Chariker MD, Katherine F. Jeter, Ed.D., et al, Contemporary Surgery, Vol. 34 June 1989, pp. 59-63.

{3}------------------------------------------------

Contraindications:

The Exsudex™ is contraindicated for the following reasons for Wound Treatment:

  • Presence of necrotic tissue .
  • Malignancy (except for quality of life reason for terminal patients) .
  • . Untreated Osteomyelitis
  • . Untreated malnutrition
  • Use on exposed arteries, veins or organs.

Precautions:

  • Patients on anticoagulants or with difficult hemostasis should be treated with caution and have to be . controlled regularly for bleeding.
  • Patients with infections in the wound and or other parts of the body have to receive proper systemic treatment.
  • . Non-compliant patients
  • The Exsudex™ has not been studied on pediatric patients. .

General Precautions for all indications for use:

  • Health care provider must evaluate patient to ensure that use of the Exsudex™ is an appropriate therapy. ●

Warnings

  • Physicians should consider patient's size and weight when prescribing device. ◆
  • The device is MR unsafe and must be disconnected from the patient prior to MRI. ♥
  • . The device may be used in the event that defibrillation is needed, provided there is no electrical contact between patient and device. In such case the device must be disconnected. Be sure to remove the dressing if defibrillation is required in the area of dressing placement. Failure to remove the dressing may inhibit transmission of electrical energy and/or patient resuscitation.
  • . The device may not be used in a hyperbaric oxygen chamber.
  • To prevent unintentional gauze retention, all dressings are removed from the wound bed. . Upon removal of dressing(s), the wound bed should be cleaned in accordance with standard wound care practices, prior to the application of new sterile dressing.
  • . If necessary all wounds should be debrided prior to application of the therapy and/or dressings.
  • . Ensure there are no pockets left in the wound bed after application of the dressing.
  • Infected wounds may need more frequent dressing changes, up to twice a day, and the patient and the . wound must be inspected regularly for signs of increased infection or sepsis.
  • . Patients who do not have adequate haemostasis, and on whom anticoagulation or platelet aggregation inhibitors are being used have an increased risk of bleeding with or without the Exsudex™.
  • . The device has not been studied in pediatric patients.
  • All arteries, veins, tendons, ligaments nerves and other organs have to be covered completely prior to . application of the Exsudex™.
  • . Infected tissue such as blood vessels may have a weakened structure and have to be treated with care. Infected blood vessels may bleed more readily than normal blood vessels
  • The device is intended for acute, institutional, and long-term care use and only when prescribed by a . physician. Product may be indicated for patient home-care, when carefully monitored by licensed home health-care providers, and only when prescribed by a physician.

{4}------------------------------------------------

K08 2311 page 5/5

v. Summary of the technological characteristics of the device compared to the predicate device.

Each of the devices providing NPWT consists of the same basic technology, and do not differ significantly in purpose, design, materials, energy source, function, or any other feature related to safety and effectiveness. Each device consists of a vacuum control unit with an integrated collection canister and power supply (battery or AC). The primary differences in these models relate to size and weight.

All models of the NPWT family of devices are designed to help promote wound healing, through the application of controlled negative pressure to the surface and margins of the wound. This negative pressure therapy is applied through the dressing positioned in the wound cavity or over a flap or graft. This pressure distributing dressing helps remove fluids from the wound. The devices are designed to treat wounds such as chronic, acute, traumatic, subacute and dehisced wounds; partial-thickness burns; ulcers (such as diabetic or pressure); flaps; and grafts.

vi. Testing

Verification and validation testing of the Exsudex™ device, including functional performance testing and electrical leakage testing, was conducted in accordance with established design control procedures.

vii. Conclusions

Based upon testing as well as extensive dinical use in Europe, the Exsudex™ Wound Drainage Device has the same intended uses as the predicate devices, with similar technological characteristics. The system performs as intended and raises no new safety or effectiveness issues.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a stylized abstract design resembling a person embracing another, or possibly a bird in flight. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the emblem.

Public Health Service

'APR - 7 2009

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

The Medical Company % Kema Quality B.V. Mr. J. A. Van Vugt 4377 County Line Road Chalfont, Pennsylvania 18914

Re: K082311

Trade/Device Name: Exsudex™ Wound Drainage Pump Regulation Number: 21 CFR 878.4780 Regulation Name: Powered Suction Pump Regulatory Class: II Product Code: OMP Dated: September 25, 2008 Received: September 30, 2008

Dear Mr. Van Vugt:

This letter corrects our substantially equivalent letter of October 8, 2008.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other

{6}------------------------------------------------

Page 2 - Mr. J. A. Van Vugt

limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to continue marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

Indications for Use Statement

510(k) Number (if known)

KD82311

Device Name:

Exsudex™ Wound Drainage Pump

Statement:

The Exsudex™ Wound Drainage Device is a compact, portable device indicated for patients who would benefit from the application of negative pressure to the area of a wound, for the aspiration and removal of surgical fluids, irrigation fluids, tissue (including bone), gases, bodily fluids or infectious materials either during surgery or at the patient's bedside particularly as the device may promote wound healing .

Prescription Use X (Part 21 CFR 801 Subpart D)

. . . . .

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Mark M. Milken

Division Sign-Off) Division of General, Restorative, and Neurological Devices

510(k) Number K082311

§ 878.4780 Powered suction pump.

(a)
Identification. A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.(b)
Classification. Class II.