Intended Use

The Ambulatory V.A.C. device is indicated for patients who whould benefit from a suction device, particularly as the device use may promote wound healing, including for patients who would benefit from vacuum-assisted drainage and removal of infectious material or other fluids from wounds under the influence of continuous and/or alternating suction pressures.

Device Description

The Ambulatory V.A.C. is a portable suction device that can be powered by a separate battery pack or a low voltage power adapter. When activated, it applies continuous or intermittent negative pressure to a wound, promoting wound healing and drainage of fluids and/or infectious materials from the wound into a disposable collection canister.

The Ambulatory V.A.C. consists of a small housing that contains a vacuum pump and control system with a containment chamber for a disposable wound fluid collection canister, and sterile, single-use disposables. Accessories included in the Ambulatory V.A.C. system are separate battery packs, a battery charger, an optional modem and a carrying belt and pouch. The disposables consists of the wound fluid collection canister with an integral hydrophobic filter and an attached length of multi-lumen tubing, a wound dressing pad, and an overlying adhesive film that adheres to the skin surrounding the wound.

The Ambulatory V.A.C. pump housing comprises of a compact plastic molding that weighs approximately 500 grams and measures 15 cm x 11 cm x 6 cm. When in use, the pump can be placed inside the pouch and carried on the belt or, when the patient is stationary, the pump can be placed on a table or flat surface.

The pump is turned ON by applying power either by the battery pack or the low voltage adapter. An LCD and membrane keypad on the top of the unit becomes active when power is applied. Operating parameters are selected using the LCD and keypad. The LCD is also used during operation to display the current negative pressure being applied and any alam condition messages.

The Ambulatory V.A.C. automatically stores the operating parameters from the previous therapy session. When the Ambulatory V.A.C. unit is initialized, the LCD displays the phrase "NEW PATIENT" at the top the display with "NO" and "YES" beneath it. "NO" is displayed over the left-arrow key and "YES" is displayed over the right-arrow key. Both arrow keys are located directly below the LCD. If the left-arrow key is pressed, for NO, the phrase "Will operate at previous settings" is displayed on the LCD. The operating parameters for the current therapy session remain as they were for the previous session. The THERAPY ON/OFF key, located to the left of the LCD, is then pressed to begin therapy.

If the right-arrow key is pressed for YES, indicating a new patient, the phrase "USING DEFAULTS" is displayed on the LCD. The user must specify whether default or user-selected parameters are to be used for the current therapy session. To use default parameters and begin operation, the THERAPY ON/OFF key is pressed and the unit will operate at default settings of 125mmHg with continuous therapy. To specify new parameters, the SELECT OPTIONS key, located to the right of the LCD, is pressed. The LCD displays the words "VAC TARGET =" followed by a number representing the current vacuum pressure setting. Pressing the right-arrow key increases the current setting in 25mmHg increments, while pressing the left-arrow key When the desired vacuum pressure is displayed, the decreases it in 25mmHg increments. SELECT OPTIONS key is again pressed to save the setting.

The LCD next displays a choice of two operation modes: 1) continuous and 2) intermittent. The word "CONTIN." appears over the left-arrow key and the word "INTERMIT." appears over the right-arrow key. The currently selected mode flashes on and off. To select continuous mode (if it is not already the currently selected mode), the user presses the left-arrow key and then the THERAPY ON/OFF key, to begin operation. If the continuous mode is already selected, the user simply presses the THERAPY ON/OFF key. To select the intermittent mode (if it is not already the currently selected mode), the user presses the right-arrow key and then the THERAPY ON/OFF key to being operation of therapy ON for 5 minutes and OFF for 2 minutes. The Ambulatory V.A.C. runs continuously until the THERAPY ON/OFF key is again pressed.

The Ambulatory V.A.C. pump has the following Alarm functions:

  • · Canister Full
  • · Dressing Leak
  • · Battery Low

Each alarm condition is accompanied by an appropriate LCD message and an audible alarm that can be silenced by correcting the condition or pressing the therapy button.

SINGLE USE DISPOSABLES

  • · wound fluid collection canister (50 ml) with integral filter and attached multilumen tube ( flexible medical tubing, meets USP Class VI criteria)
  • polyurethane foam pad .
  • plastic adhesive overlay .
  • · Flexible tube to dressing connector

The wound fluid collection canister is a disposable chamber made of translucent plastic used for receiving fluids emitted from a wound. The canister includes a plastic cap that firmly locates the canister inside the pump housing. The canister is also graduated to allow the user to easily view and measure the volume of wound fluids within the canister.

The collection canister has two parts which locate on spigots inside the pump housing. One spigot is used for applying a vacuum down the central bore of the multilumen tube. The other spigot locates in the second port which in turn is connected to the outer lumen of the tube and is used for sensing negative pressure at the wound site. A 0.2 micron hydrophobic membrane filter is mounted over the canister outlet. Fluid is drawn up through the central bore of the tube and into the canister. When the canister becomes full, the face of the filter is occluded and the negative pressure is reduced at the wound site which triggers the canister full alarm.

The disposable wound pad is fabricated from either polyurethane or PVA foam. In order for wound fluids to be communicated into the canister, the pad must be secured over the wound. This is accomplished by applying a plastic adhesive drape over the wound pad that gently adheres to the skin surrounding the wound. With the wound area sealed, vacuum pressure is transmitted to the wound when the suction pump is activated.

AI/ML Overview

The provided text describes a medical device, the Ambulatory V.A.C., which is a portable suction device used for wound healing. However, the document is a 510(k) summary for premarket notification and does not contain details about a clinical study with acceptance criteria and reported device performance in the way typically found for AI/ML-based devices.

This document focuses on demonstrating substantial equivalence to a predicate device (V.A.C. Plus, 510(k) No. 945062) rather than presenting results from a performance study against specific acceptance criteria for a new device. The content is primarily a device description, indications for use, contraindications, precautions, and a comparison table of specifications between the Ambulatory V.A.C. and its predicate/similar devices.

Therefore, I cannot extract the requested information regarding acceptance criteria, reported device performance, sample size for test set, data provenance, expert details, adjudication methods, MRMC study, standalone performance, ground truth types, training set size, or how training ground truth was established, as this information is not present in the provided 510(k) summary.

The closest information available related to "performance" is the specifications comparison chart which lists physical and functional characteristics of the device and comparable devices. These are not "acceptance criteria" in the sense of a clinical or performance study outcome but rather design specifications.

However, I can extract information related to the device's specifications, which act as de-facto performance metrics for a hardware device in the context of substantial equivalence.

Here's an attempt to derive some of the requested information based on what is available, acknowledging the limitations mentioned above:


1. A table of acceptance criteria and the reported device performance

Since this is a 510(k) summary demonstrating substantial equivalence for a physical medical device, not a performance study of an AI/ML algorithm, the "acceptance criteria" are implied by the similarity to the predicate device and relevant industry standards for powered suction pumps. The "reported device performance" are the physical and functional specifications of the Ambulatory V.A.C.

Acceptance Criteria (Implied by Predicate/Device Type)Reported Device Performance (Ambulatory V.A.C. Specifications)
Weight (Comparable to portable devices)2 lbs
Power Supply (Battery or low voltage adapter capable)Battery powered or low voltage adapter, 115V/60Hz
Therapy Settings Range (Appropriate for wound care)Range 0-125mmHg
Continuous and Intermittent Therapy Settings (Customizable)Custom settings range from 30 secs. to 10 min. for ON/OFF times
Exhaust Filter (Present)YES
Settings Lockout (Present)YES
Pump Type (Oil-less Diaphragm)Oil-less Diaphragm
Canister Capacity (Appropriate for portable use)50 ml
Microprocessor Control (Present)YES
Negative Pressure Feedback (Present)YES
Maximum Pressure (Within therapeutic range)Not explicitly listed, but operates up to 125mmHg
Indicator (Functioning gauge)Gauge
Free Air Displacement (Within acceptable range for suction pump)2 liters/min
Cycle ModesIntermittent and Continuous Option
Vacuum Variability (Manually adjustable)Yes, Manual
Time Meter (Present)Yes
Dimensions6"Wx4.3"Hx3"D
In-take Hose0.25" ID
Illuminated Power SwitchYes

The following information cannot be extracted from the provided text as it pertains to clinical/performance studies, particularly for AI/ML devices, which is not the nature of this 510(k) submission for a physical medical device.

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Not Applicable. The document is a 510(k) summary for a physical medical device, not a data-driven AI/ML device. It relies on design specifications and performance characteristics, comparing them to a predicate device, rather than a clinical "test set" with patient data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Not Applicable. No "ground truth" establishment by experts in the context of a test set is described.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Not Applicable. No test set or adjudication method is described.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not Applicable. This is a hardware device, not an AI-assisted diagnostic tool.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not Applicable. This is a hardware device.

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

  • Not Applicable. No "ground truth" of this nature is used for this type of device submission. The "ground truth" for a physical device is its functional specifications and adherence to design parameters compared to established medical device standards and predicate devices.

8. The sample size for the training set

  • Not Applicable. No training set is described for a hardware device.

9. How the ground truth for the training set was established

  • Not Applicable. No training set or ground truth establishment is described.

{0}------------------------------------------------

510 (k) Summary K971548

I. Name of Device:Ambulatory V.A.C
II. Classification Name:Powered Suction Pump (per 21 CFR 878.4780)
MAY 22 1997
III. Substantial Equivalence:V.A.C. Plus, 510(k) No. 945062

IV. Device Description:

The Ambulatory V.A.C. is a portable suction device that can be powered by a separate battery pack or a low voltage power adapter. When activated, it applies continuous or intermittent negative pressure to a wound, promoting wound healing and drainage of fluids and/or infectious materials from the wound into a disposable collection canister.

The Ambulatory V.A.C. consists of a small housing that contains a vacuum pump and control system with a containment chamber for a disposable wound fluid collection canister, and sterile, single-use disposables. Accessories included in the Ambulatory V.A.C. system are separate battery packs, a battery charger, an optional modem and a carrying belt and pouch. The disposables consists of the wound fluid collection canister with an integral hydrophobic filter and an attached length of multi-lumen tubing, a wound dressing pad, and an overlying adhesive film that adheres to the skin surrounding the wound.

The Ambulatory V.A.C. pump housing comprises of a compact plastic molding that weighs approximately 500 grams and measures 15 cm x 11 cm x 6 cm. When in use, the pump can be placed inside the pouch and carried on the belt or, when the patient is stationary, the pump can be placed on a table or flat surface.

The pump is turned ON by applying power either by the battery pack or the low voltage adapter. An LCD and membrane keypad on the top of the unit becomes active when power is applied. Operating parameters are selected using the LCD and keypad. The LCD is also used during operation to display the current negative pressure being applied and any alam condition messages.

The Ambulatory V.A.C. automatically stores the operating parameters from the previous therapy session. When the Ambulatory V.A.C. unit is initialized, the LCD displays the phrase "NEW PATIENT" at the top the display with "NO" and "YES" beneath it. "NO" is displayed over the left-arrow key and "YES" is displayed over the right-arrow key. Both arrow keys are located directly below the LCD. If the left-arrow key is pressed, for NO, the phrase "Will operate at previous settings" is displayed on the LCD. The operating parameters for the current therapy session remain as they were for the previous session. The THERAPY ON/OFF key, located to the left of the LCD, is then pressed to begin therapy.

{1}------------------------------------------------

If the right-arrow key is pressed for YES, indicating a new patient, the phrase "USING DEFAULTS" is displayed on the LCD. The user must specify whether default or user-selected parameters are to be used for the current therapy session. To use default parameters and begin operation, the THERAPY ON/OFF key is pressed and the unit will operate at default settings of 125mmHg with continuous therapy. To specify new parameters, the SELECT OPTIONS key, located to the right of the LCD, is pressed. The LCD displays the words "VAC TARGET =" followed by a number representing the current vacuum pressure setting. Pressing the right-arrow key increases the current setting in 25mmHg increments, while pressing the left-arrow key When the desired vacuum pressure is displayed, the decreases it in 25mmHg increments. SELECT OPTIONS key is again pressed to save the setting.

The LCD next displays a choice of two operation modes: 1) continuous and 2) intermittent. The word "CONTIN." appears over the left-arrow key and the word "INTERMIT." appears over the right-arrow key. The currently selected mode flashes on and off. To select continuous mode (if it is not already the currently selected mode), the user presses the left-arrow key and then the THERAPY ON/OFF key, to begin operation. If the continuous mode is already selected, the user simply presses the THERAPY ON/OFF key. To select the intermittent mode (if it is not already the currently selected mode), the user presses the right-arrow key and then the THERAPY ON/OFF key to being operation of therapy ON for 5 minutes and OFF for 2 minutes. The Ambulatory V.A.C. runs continuously until the THERAPY ON/OFF key is again pressed.

The Ambulatory V.A.C. pump has the following Alarm functions:

  • · Canister Full
  • · Dressing Leak
  • · Battery Low

Each alarm condition is accompanied by an appropriate LCD message and an audible alarm that can be silenced by correcting the condition or pressing the therapy button.

SINGLE USE DISPOSABLES

  • · wound fluid collection canister (50 ml) with integral filter and attached multilumen tube ( flexible medical tubing, meets USP Class VI criteria)
  • polyurethane foam pad .
  • plastic adhesive overlay .
  • · Flexible tube to dressing connector

The wound fluid collection canister is a disposable chamber made of translucent plastic used for receiving fluids emitted from a wound. The canister includes a plastic cap that firmly locates the canister inside the pump housing. The canister is also graduated to allow the user to easily view and measure the volume of wound fluids within the canister.

Page 2

510(k) Summary Ambulatory V.A.C. Date Prepared: April 24, 1997

ﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤ

{2}------------------------------------------------

The collection canister has two parts which locate on spigots inside the pump housing. One spigot is used for applying a vacuum down the central bore of the multilumen tube. The other spigot locates in the second port which in turn is connected to the outer lumen of the tube and is used for sensing negative pressure at the wound site. A 0.2 micron hydrophobic membrane filter is mounted over the canister outlet. Fluid is drawn up through the central bore of the tube and into the canister. When the canister becomes full, the face of the filter is occluded and the negative pressure is reduced at the wound site which triggers the canister full alarm.

The disposable wound pad is fabricated from either polyurethane or PVA foam. In order for wound fluids to be communicated into the canister, the pad must be secured over the wound. This is accomplished by applying a plastic adhesive drape over the wound pad that gently adheres to the skin surrounding the wound. With the wound area sealed, vacuum pressure is transmitted to the wound when the suction pump is activated.

INDICATIONS FOR USE

The Ambulatory V.A.C. is a compact, portable device indicated for patients who would benefit from a suction device, particularly as the device use may promote wound healing, including for patients who would benefit from vacuum-assisted drainage and removal of infectious material or other fluids from wounds under the influence of continuous and/or alternating suction pressures.

CONTRAINDICATIONS

The Ambulatory V.A.C. suction device is contraindicated for the following types of wounds:

  • · Fistulas, excluding blind or incomplete fistulas,
  • · Presence of necrotic tissue (including bone osteomyelitis),
  • · Skin Cancer.

PRECAUTIONS

  • · Patients on anticoagulant, or difficult hemostatis
    Page 3

510(k) Summary Ambulatory V.A.C. Prepared by: J. Harbour Kinetic Concepts, Inc. P.O. Box 659508 San Antonio, TX 78265

Date Prepared: April 24, 1997

.
A

1 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 -

{3}------------------------------------------------

DESCRIPTIONPump unit carried on belt withseparate battery packHand carry, portable pump unit
WEIGHT1.1 pounds10 pounds
POWERBattery powered or low voltageadapter, 115V/60Hz115V/60Hz
THERAPY SETTINGSRange 0-125mmHgRange 0-125mmHg
CONTINUOUS ANDINTERMITTENTTHERAPY SETTINGCustom settings range from 30secs. to 10 min. for ON/OFFtimesCustom settings range from 30secs. to 10 min for ON/OFFtimes
EXHAUST FILTERYESYES
SETTINGS LOCKOUTYESYES
PUMP TYPEOil-less DiaphragmOil-less Diaphragm
CANISTER CAPACITY50 ml300 ml
MICROPROCESSORCONTROLYESYES
NEGATIVE PRESSUREFEEDBACKYESYES

DIFFERENCES BETWEEN CURRENT AND PREDICATED DEVICE

Page 4

510(k) Summary Ambulatory V.A.C. Prepared by: J. Harbour Kinetic Concepts, Inc. P.O. Box 659508 San Antonio, TX 78265 Date Prepared: April 24, 1997

{4}------------------------------------------------

Ambulatory V.A.C. SPECIFICATIONS COMPARISON CHART

AmbuVACVACPLUSCARE E-VAC
MANUFACTURER/DISTRIBUTORKinetic Concepts, Inc.Kinetic Concepts, Inc.Aeros Instruments
MAX. PRESSURE380mm/15" Hg550mm/22" Hg
INDICATORGaugeGaugeGauge
WEIGHT2 lbs10 lbs9.7 lbs
FREE AIRDISPLACEMENT2 liters/min9 liters/min27 liters/min
CYCLEIntermittent andContinuous OptionIntermittent andContinuous OptionContinous
VACUUMVARIABILITYYes, ManualYes, ManualYes, Manual
TIME METERYesYesNo
EXHAUSTFILTERYesYesUnavailable
PUMP TYPEOil-less DiaphragmOil-less DiaphragmOil-less Diaphragm
POWER115V/60Hz115V/60Hz110V/60Hz
DIMENSIONS6"Wx4.3"Hx3"D11"Wx12"Hx7"D19"Wx11.5"Hx16.5"D
IN-TAKE HOSE0.25" ID0.25" IDUnavailable
ILLUMINATEDPOWER SWITCHYesYesYes

Page 5

510(k) Summary Ambulatory V.A.C. Prepared by: J. Harbour Kinetic Concepts, Inc. P.O. Box 659508 San Antonio, TX 78265 Date Prepared: April 24, 1997

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with its wings spread, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle. The eagle is composed of three curved lines that form the shape of the bird's body and wings. The text is in all caps and is evenly spaced around the circle.

Public Health Service

APR - 7 2009

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

KCI USA, Inc. % Ms. Christy Oviatt 6203 Farinon Drive San Antonio, Texas 78230

Re: K971548

Trade/Device Name: AmbuVAC Device Regulation Number: 21 CFR 878.4780 Regulation Name: Powered Suction Pump Regulatory Class: II Product Code: OMP Dated: April 24, 1997 Received: April 28, 1997

Dear Ms. Oviatt:

This letter corrects our substantially equivalent letter of May 22, 1997.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not

{6}------------------------------------------------

Page 2 - Ms. Christy Oviatt

limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to continue marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

for Rob Romer

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

                                  • -. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : :

: 【

Page / of /

510(k) Number (if known): K971548

Device Name: Ambulatory V.A.C.

Indications For Use:

・しと

The Ambulatory V.A.C. device is indicated for patients who whould benefit from a suction device, particularly as the device use may promote wound healing, including for patients who would benefit from vacuum-assisted drainage and removal of infectious material or other fluids from wounds under the influence of continuous and/or alternating suction pressures.

NOTE: The Indications for Use of the Ambulatory V.A.C. are identical to the predicate device, VAC PLUS, 510(k)945062.

IPLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of General Restorative Devices
510(k) NumberK971548

Prescription Use (Per 21 CFR 801. 109)

OR

Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________

(Optional Format 1-2-96)

§ 878.4780 Powered suction pump.

(a)
Identification. A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.(b)
Classification. Class II.