K Number
K061970

Validate with FDA (Live)

Manufacturer
Date Cleared
2006-08-31

(50 days)

Product Code
Regulation Number
866.5510
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Immunoassay for the in vitro quantitative determination of immunoglobulin E in human serum and plasma. Determination of total IgE is useful as an aid in the diagnosis of allergic diseases. The electrochemiluminescence immunoassay "ECLIA" is intended for use on the Roche Elecsys 2010 and MODULAR ANALYTICS E170 (Elecsys module) immunoassay analyzers.

Device Description

The Elecsys IgE II immunoassay is a two step sandwich immunoassay with streptavidin microparticles and electrochemiluminescence detection. Results are determined using a calibration curve that is generated specifically on each instrument by a 2 point calibration and a master curve provided with the reagent bar code.

AI/ML Overview

The given document is a 510(k) summary for the Elecsys IgE II Immunoassay. It describes the device, its intended use, and compares it to a predicate device. However, it does not contain specific acceptance criteria or a dedicated study section detailing how the device met those criteria in the format typically used for AI/ML device evaluations.

The document focuses on demonstrating substantial equivalence to a previously approved device (Elecsys IgE Immunoassay, K984326, K961481/A003) for the purpose of 510(k) clearance. This means the assessment is primarily a device comparison rather than a standalone performance study against pre-defined acceptance criteria for a novel device.

Therefore, many of the requested details about acceptance criteria, study design, sample sizes, ground truth establishment, expert involvement, and MRMC studies are not present in this type of submission.

Here's a breakdown based on the information available:


1. Table of Acceptance Criteria and Reported Device Performance

As this is a substantial equivalence submission for an in vitro diagnostic (IVD) immunoassay, the "acceptance criteria" are implicitly met by demonstrating comparable performance to the predicate device across various analytical characteristics. There isn't a single table of "acceptance criteria" with pass/fail thresholds in the typical sense for AI/ML performance. Instead, the document presents comparative performance data to show the new device is "substantially equivalent" to the predicate.

The table below summarizes the key analytical performance characteristics presented in the document, comparing the Elecsys IgE II (Modified Device) to the predicate Elecsys IgE. The "acceptance criteria" here are effectively that the new device's performance should be comparable enough to the predicate to be considered substantially equivalent.

Performance CharacteristicPredicate Device (Elecsys IgE)Elecsys IgE II (Modified Device)"Acceptance Criteria" (Implicit for Substantial Equivalence)Outcome
Intended UseImmunoassay for IgE quantification in human serum and plasma; aid in diagnosis of allergic diseases.SameFunctionally equivalent to predicate.Met
AnalyzersRoche Elecsys 1010/2010 and MODULAR ANALYTICS E170Roche Elecsys 2010 and MODULAR ANALYTICS E170Compatible with similar clinical analyzer platforms.Met
TraceabilityStandardized against 2nd IRP WHO Reference Standard 75/502SameSame standardization as predicate.Met
Assay ProtocolSandwichSameSame assay principle.Met
Sample TypeSerum and plasmaSameSame clinical utility.Met
Calibrator, Calibration Verification & ControlsIgE CalSet, IgE CalCheck, PreciControl UniversalSameUses same calibration materials and controls.Met
Measuring Range0.10 - 2500 IU/mLSameSame analytical range.Met
Analytical Sensitivity0.10 IU/mLSameSame lower limit of detection.Met
Functional Sensitivity0.50 IU/mlSameSame functional lower limit.Met
CompositionR1: 2.4 mg/L AB-Bi, buffer, preservative; R2: 4.8 mg/L AB-Ru, buffer, preservative; M: 0.72 mg/ml streptavidin-coated microparticles, preservativeR1: 2.5 mg/L (mono-Bi)AB-Bi, buffer, preservative; R2: 5.5 mg/L AB-Ru, buffer, preservative; M: 0.72 mg/ml streptavidin-coated microparticles, preservativeMinor compositional changes that do not alter fundamental performance.Met
Dilution Recommendation> 60 IU/mL> 125 IU/mLAcceptable adjustment based on method validation.Met
Interferences (Bilirubin)No affect up to 37 mg/dlSameComparable resistance to common interferents.Met
Interferences (Hemoglobin)No affect up to 1.1 g/dlNo affect up to 0.1 g/dlWhile the value is different, the FDA found it acceptable for substantial equivalence.Met
Interferences (Triglycerides)No affect up to 2200 mg/dlSameComparable resistance to common interferents.Met
Interferences (Biotin)No affect up to 100 ng/mlSameComparable resistance to common interferents.Met
Precision (Elecsys 2010 Intra-assay CV)HS1 3.6% @ 5.18 IU/mL; HS2 3.2% @ 398 IU/mL; HS3 2.4% @ 1010 IU/mLHS1 4.1% @ 32.7 IU/mL; HS2 2.4% @ 265 IU/mL; HS3 2.6% @ 1295 IU/mLComparable precision performance.Met
Precision (Elecsys 2010 Total CV)HS1 4.2% @ 5.18 IU/mL; HS2 3.9% @ 398 IU/mL; HS3 3.1% @ 1010 IU/mLHS1 5.1% @ 32.7 IU/mL; HS2 3.8% @ 265 IU/mL; HS3 3.9% @ 1295 IU/mLComparable precision performance.Met
Precision (E170 Intra-assay CV)HS1 2.3% @ 3.36 IU/mL; HS2 2.2% @ 457 IU/mL; HS3 2.6% @ 1128 IU/mLHS1 1.4% @ 4.4 IU/mL; HS2 0.7% @ 261 IU/mL; HS3 1.0% @ 1018 IU/mLComparable precision performance.Met
Precision (E170 Total CV)HS1 4.3% @ 3.31 IU/mL; HS2 3.8% @ 443 IU/mL; HS3 6.5% @ 1215 IU/mLHS1 2.7% @ 30.2 IU/mL; HS2 2.8% @ 245 IU/mL; HS3 3.4% @ 1207 IU/mLComparable precision performance.Met

2. Sample Size Used for the Test Set and Data Provenance

  • Test Set (Clinical Performance): The document does not explicitly present a "test set" in the context of clinical performance data in the way an AI/ML device would. Instead, it compares analytical performance characteristics (sensitivity, measuring range, precision, interference) of the new device against the predicate.
    • For precision studies, the data points represent multiple measurements ("HS1", "HS2", "HS3" at different IgE concentrations) for intra-assay and total variation. The exact number of individual patient samples or replicates used to derive these CVs is not specified in the summary, but typical precision studies involve multiple runs over several days.
    • For interference studies, it indicates "No affect up to," implying specific concentrations of potential interferents were tested. The number of samples or replicates used for these tests is not specified.
  • Data Provenance: Not explicitly stated, but for IVD assays, analytical performance studies are typically conducted by the manufacturer in a controlled laboratory setting. It's likely a mix of internal validation studies, and possibly some external site testing, but specific countries or retrospective/prospective nature is not detailed.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

Not applicable for this type of IVD immunoassay submission. The "ground truth" for analytical performance studies is based on established laboratory measurement techniques, reference standards (like WHO Reference Standard 75/502 for IgE), and accepted statistical methods for evaluating precision, sensitivity, etc. There are no "experts" establishing a subjective ground truth for image interpretation or diagnosis in this context.


4. Adjudication Method for the Test Set

Not applicable. As described above, this is an analytical performance assessment, not a clinical adjudication of diagnostic outcomes by human readers.


5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is an in vitro diagnostic immunoassay, not an AI/ML diagnostic aid intended to assist human readers (e.g., radiologists interpreting images). Therefore, an MRMC study is not relevant.


6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

Not applicable. The device is a laboratory instrument and reagent system that quantitatively measures IgE. Its performance is inherent to the assay and instrument, and it operates without human intervention in the result generation itself, but a lab technician operates the instrument. It is not an "algorithm" in the AI/ML sense.


7. The Type of Ground Truth Used

The ground truth for the performance characteristics presented (e.g., measuring range, analytical sensitivity, functional sensitivity, precision) is based on:

  • Reference Standards: Specifically, the 2nd IRP WHO Reference Standard 75/502 for IgE.
  • Defined Analytical Methods: The quantitative measurement of IgE following standardized laboratory procedures.
  • Statistical Analysis: Metrics like Coefficient of Variation (CV) for precision are derived from repeated measurements using established statistical methods.

8. The Sample Size for the Training Set

Not applicable. This device is an immunoassay, not a machine learning algorithm that requires a "training set."


9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no "training set" for this type of immunoassay.

{0}------------------------------------------------

K061970

AUG 3 1 2006

510(k) Summary - Elecsys IgE II Immunoassay

IntroductionAccording to the requirements of 21 CFR 807.92, the following information provides sufficient detail to understand the basis for a determination of substantial equivalence
Submitter name, address, contactRoche Diagnostics9115 Hague RdIndianapolis IN 46250(317) 521-3544Contact person: Kay A. TaylorDate prepared: July 11, 2006
Device NameProprietary name: Elecsys IgE II immunoassayCommon name: IgE testClassification name: radioimmunoassay, immunoglobulins (d, e)
Device descriptionThe Elecsys IgE II immunoassay is a two step sandwich immunoassay with streptavidin microparticles and electrochemiluminescence detection. Results are determined using a calibration curve that is generated specifically on each instrument by a 2 point calibration and a master curve provided with the reagent bar code.
Intended Use / Indications for UseImmunoassay for the in vitro quantitative determination of immunoglobulin E in human serum and plasma. Determination of total IgE is useful as an aid in the diagnosis of allergic diseases.
Substantial equivalenceThe Elecsys IgE II immunoassay is substantially equivalent to another device legally marketed in the United States. Elecsys IgE II (modified) assay is equivalent to the Elecsys IgE immunoassay (K984326, K961481/A003). Both products are intended for use in the quantitative determination of IgE in serum and plasma.
Device ComparisonThe table below compares the device features of the Elecsys IgE II immunoassay (modified) and original (K984326, K961481/A003).

Confidential

:

.

{1}------------------------------------------------

510(k) Summary – Elecsys IgE II Immunoassay, continued

TopicElecsys IgE (K984326,K961481/A003)Elecsys IgE II (Modified Device)
Intended useImmunoassay for the in vitroquantitative determination ofimmunoglobulin E in human serumand plasma. Determination of totalIgE is useful as an aid in the diagnosisof allergic diseases.Same
AnalyzersRoche Elecsys 1010/2010 andMODULAR ANALYTICS E170(Elecsys module) immunoassayanalyzers.Roche Elecsys 2010 and MODULARANALYTICS E170 (Elecsys module)immunoassay analyzers.
TraceabilityAssay standardized against the 2nd IRPWHO Reference Standard 75/502Same
Assay ProtocolSandwichSame
Sample TypeSerum and plasmaSame
Calibrator,CalibrationVerification &ControlsIgE CalSetIgE CalCheckPreciControl UniversalSame
MeasuringRange0.10-2500 IU/mlSame
AnalyticalSensitivity0.10 IU/mlSame
FunctionalSensitivity0.50 IU/mlSame
CompositionR1: 2.4 mg/L AB-Bi, buffer,preservativeR2: 4.8 mg/L AB-Ru, buffer,preservativeM: 0.72 mg/ml streptavidin-coatedmicroparticles, preservativeR1: 2.5 mg/L (mono-Bi)AB-Bi,buffer, preservativeR2: 5.5 mg/L AB-Ru, buffer,preservativeM: 0.72 mg/ml streptavidin-coatedmicroparticles, preservative
DilutionRecommendationConcentration of diluted samples mustbe > 60 IU/mLConcentration of diluted samples mustbe > 125 IU/mL
InterferencesNo affect up to,bilirubin < 37 mg/dlhemoglobin < 1.1 g/dltriglycerides < 2200 mg/dlbiotin < 100 ng/mlNo affect up to,Bilirubin - Samehemoglobin < 0.1 g/dltriglycerides - Samebiotin - Same

{2}------------------------------------------------

510(k) Summary – Elecsys IgE II Immunoassay, continued

TopicElecsys IgE (K984326,K961481/A003)Elecsys IgE II (Modified Device)
PrecisionElecsys 1010 / 2010:Elecsys 2010:
Intra-assayIntra-assay:
HS1 3.6% CV @5.18 IU/mLHS1 4.1% CV @32.7 IU/mL
HS2 3.2% CV @398 IU/mLHS2 2.4% CV @265 IU/mL
HS3 2.4% CV @1010 IU/mLHS3 2.6% CV @1295 IU/mL
Total:Total:
HS1 4.2% CV @5.18 IU/mLHS1 5.1% CV @32.7 IU/mL
HS2 3.9% CV @398 IU/mLHS2 3.8% CV @265 IU/mL
HS3 3.1% CV @1010 IU/mLHS3 3.9% CV @1295 IU/mL
E170:E170:
Intra-assayIntra-assay:
HS1 2.3% CV @3.36 IU/mLHS1 1.4% CV @4.4 IU/mL
HS2 2.2% CV @457 IU/mLHS2 0.7% CV @261 IU/mL
HS3 2.6% CV @1128 IU/mLHS3 1.0% CV @1018 IU/mL
Total:Total:
HS1 4.3% CV @3.31 IU/mLHS1 2.7% CV @30.2 IU/mL
HS2 3.8% CV @443 IU/mLHS2 2.8% CV @245 IU/mL
HS3 6.5% CV @1215 IU/mLHS3 3.4% CV @1207 IU/mL

Confidential

·

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three lines forming its body and wings. The eagle is positioned to the right of a circular inscription that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA".

Public Health Service

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Ms. Kay A. Taylor, MT (ASCP), RAC Regulatory Affairs Principal US Regulatory Affairs Roche Diagnostics, Inc. 9115 Hague Road P.O. Box 50416 Indianapolis, Indiana 46250-0416

AUG 3 1 2006

K061970 Re:

Trade/Device Name: Elecsys IgE II Immunoassay Regulation Number: 21 CFR § 866.5510 Regulation Name: Immunoglobulins A, G, M, D and E Immunological Test System Regulatory Class: II Product Code: JHR Dated: August 22, 2006 Received: August 23, 2006

Dear Ms. Taylor:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for about to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean I hat FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice Or re rate overy, societing (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

{4}------------------------------------------------

Page 2 -

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html

Sincerely yours,

lobatz Beckerh

Robert L. Becker, Jr., MD, PK Director Division of Immunology and Hematology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

510(k) Number (if known):

19061970

Device Name: Elecsys IgE II Immunoassay

Indications For Use:

Immunoassay for the in vitro quantitative determination of immunoglobulin E in human serum and plasma.

Determination of total IgE is useful as an aid in the diagnosis of allergic diseases.

The electrochemiluminescence immunoassay "ECLIA" is intended for use on the Roche Elecsys 2010 and MODULAR ANALYTICS E170 (Elecsys module) immunoassay analyzers.

Peter R. Roper

Division Sign-Off

ce of In Vitro Diagnostic Device Evaluation and Safety

510(k) K061970

Prescription Use XXXX (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Confidential

§ 866.5510 Immunoglobulins A, G, M, D, and E immunological test system.

(a)
Identification. An immunoglobulins A, G, M, D, and E immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the immunoglobulins A, G, M, D, an E (serum antibodies) in serum. Measurement of these immunoglobulins aids in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents.(b)
Classification. Class II (performance standards).