Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K061970
    Manufacturer
    Date Cleared
    2006-08-31

    (50 days)

    Product Code
    Regulation Number
    866.5510
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    JHR

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Immunoassay for the in vitro quantitative determination of immunoglobulin E in human serum and plasma. Determination of total IgE is useful as an aid in the diagnosis of allergic diseases. The electrochemiluminescence immunoassay "ECLIA" is intended for use on the Roche Elecsys 2010 and MODULAR ANALYTICS E170 (Elecsys module) immunoassay analyzers.

    Device Description

    The Elecsys IgE II immunoassay is a two step sandwich immunoassay with streptavidin microparticles and electrochemiluminescence detection. Results are determined using a calibration curve that is generated specifically on each instrument by a 2 point calibration and a master curve provided with the reagent bar code.

    AI/ML Overview

    The given document is a 510(k) summary for the Elecsys IgE II Immunoassay. It describes the device, its intended use, and compares it to a predicate device. However, it does not contain specific acceptance criteria or a dedicated study section detailing how the device met those criteria in the format typically used for AI/ML device evaluations.

    The document focuses on demonstrating substantial equivalence to a previously approved device (Elecsys IgE Immunoassay, K984326, K961481/A003) for the purpose of 510(k) clearance. This means the assessment is primarily a device comparison rather than a standalone performance study against pre-defined acceptance criteria for a novel device.

    Therefore, many of the requested details about acceptance criteria, study design, sample sizes, ground truth establishment, expert involvement, and MRMC studies are not present in this type of submission.

    Here's a breakdown based on the information available:


    1. Table of Acceptance Criteria and Reported Device Performance

    As this is a substantial equivalence submission for an in vitro diagnostic (IVD) immunoassay, the "acceptance criteria" are implicitly met by demonstrating comparable performance to the predicate device across various analytical characteristics. There isn't a single table of "acceptance criteria" with pass/fail thresholds in the typical sense for AI/ML performance. Instead, the document presents comparative performance data to show the new device is "substantially equivalent" to the predicate.

    The table below summarizes the key analytical performance characteristics presented in the document, comparing the Elecsys IgE II (Modified Device) to the predicate Elecsys IgE. The "acceptance criteria" here are effectively that the new device's performance should be comparable enough to the predicate to be considered substantially equivalent.

    Performance CharacteristicPredicate Device (Elecsys IgE)Elecsys IgE II (Modified Device)"Acceptance Criteria" (Implicit for Substantial Equivalence)Outcome
    Intended UseImmunoassay for IgE quantification in human serum and plasma; aid in diagnosis of allergic diseases.SameFunctionally equivalent to predicate.Met
    AnalyzersRoche Elecsys 1010/2010 and MODULAR ANALYTICS E170Roche Elecsys 2010 and MODULAR ANALYTICS E170Compatible with similar clinical analyzer platforms.Met
    TraceabilityStandardized against 2nd IRP WHO Reference Standard 75/502SameSame standardization as predicate.Met
    Assay ProtocolSandwichSameSame assay principle.Met
    Sample TypeSerum and plasmaSameSame clinical utility.Met
    Calibrator, Calibration Verification & ControlsIgE CalSet, IgE CalCheck, PreciControl UniversalSameUses same calibration materials and controls.Met
    Measuring Range0.10 - 2500 IU/mLSameSame analytical range.Met
    Analytical Sensitivity0.10 IU/mLSameSame lower limit of detection.Met
    Functional Sensitivity0.50 IU/mlSameSame functional lower limit.Met
    CompositionR1: 2.4 mg/L AB-Bi, buffer, preservative; R2: 4.8 mg/L AB-Ru, buffer, preservative; M: 0.72 mg/ml streptavidin-coated microparticles, preservativeR1: 2.5 mg/L (mono-Bi)AB-Bi, buffer, preservative; R2: 5.5 mg/L AB-Ru, buffer, preservative; M: 0.72 mg/ml streptavidin-coated microparticles, preservativeMinor compositional changes that do not alter fundamental performance.Met
    Dilution Recommendation> 60 IU/mL> 125 IU/mLAcceptable adjustment based on method validation.Met
    Interferences (Bilirubin)No affect up to 37 mg/dlSameComparable resistance to common interferents.Met
    Interferences (Hemoglobin)No affect up to 1.1 g/dlNo affect up to 0.1 g/dlWhile the value is different, the FDA found it acceptable for substantial equivalence.Met
    Interferences (Triglycerides)No affect up to 2200 mg/dlSameComparable resistance to common interferents.Met
    Interferences (Biotin)No affect up to 100 ng/mlSameComparable resistance to common interferents.Met
    Precision (Elecsys 2010 Intra-assay CV)HS1 3.6% @ 5.18 IU/mL; HS2 3.2% @ 398 IU/mL; HS3 2.4% @ 1010 IU/mLHS1 4.1% @ 32.7 IU/mL; HS2 2.4% @ 265 IU/mL; HS3 2.6% @ 1295 IU/mLComparable precision performance.Met
    Precision (Elecsys 2010 Total CV)HS1 4.2% @ 5.18 IU/mL; HS2 3.9% @ 398 IU/mL; HS3 3.1% @ 1010 IU/mLHS1 5.1% @ 32.7 IU/mL; HS2 3.8% @ 265 IU/mL; HS3 3.9% @ 1295 IU/mLComparable precision performance.Met
    Precision (E170 Intra-assay CV)HS1 2.3% @ 3.36 IU/mL; HS2 2.2% @ 457 IU/mL; HS3 2.6% @ 1128 IU/mLHS1 1.4% @ 4.4 IU/mL; HS2 0.7% @ 261 IU/mL; HS3 1.0% @ 1018 IU/mLComparable precision performance.Met
    Precision (E170 Total CV)HS1 4.3% @ 3.31 IU/mL; HS2 3.8% @ 443 IU/mL; HS3 6.5% @ 1215 IU/mLHS1 2.7% @ 30.2 IU/mL; HS2 2.8% @ 245 IU/mL; HS3 3.4% @ 1207 IU/mLComparable precision performance.Met

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set (Clinical Performance): The document does not explicitly present a "test set" in the context of clinical performance data in the way an AI/ML device would. Instead, it compares analytical performance characteristics (sensitivity, measuring range, precision, interference) of the new device against the predicate.
      • For precision studies, the data points represent multiple measurements ("HS1", "HS2", "HS3" at different IgE concentrations) for intra-assay and total variation. The exact number of individual patient samples or replicates used to derive these CVs is not specified in the summary, but typical precision studies involve multiple runs over several days.
      • For interference studies, it indicates "No affect up to," implying specific concentrations of potential interferents were tested. The number of samples or replicates used for these tests is not specified.
    • Data Provenance: Not explicitly stated, but for IVD assays, analytical performance studies are typically conducted by the manufacturer in a controlled laboratory setting. It's likely a mix of internal validation studies, and possibly some external site testing, but specific countries or retrospective/prospective nature is not detailed.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable for this type of IVD immunoassay submission. The "ground truth" for analytical performance studies is based on established laboratory measurement techniques, reference standards (like WHO Reference Standard 75/502 for IgE), and accepted statistical methods for evaluating precision, sensitivity, etc. There are no "experts" establishing a subjective ground truth for image interpretation or diagnosis in this context.


    4. Adjudication Method for the Test Set

    Not applicable. As described above, this is an analytical performance assessment, not a clinical adjudication of diagnostic outcomes by human readers.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is an in vitro diagnostic immunoassay, not an AI/ML diagnostic aid intended to assist human readers (e.g., radiologists interpreting images). Therefore, an MRMC study is not relevant.


    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. The device is a laboratory instrument and reagent system that quantitatively measures IgE. Its performance is inherent to the assay and instrument, and it operates without human intervention in the result generation itself, but a lab technician operates the instrument. It is not an "algorithm" in the AI/ML sense.


    7. The Type of Ground Truth Used

    The ground truth for the performance characteristics presented (e.g., measuring range, analytical sensitivity, functional sensitivity, precision) is based on:

    • Reference Standards: Specifically, the 2nd IRP WHO Reference Standard 75/502 for IgE.
    • Defined Analytical Methods: The quantitative measurement of IgE following standardized laboratory procedures.
    • Statistical Analysis: Metrics like Coefficient of Variation (CV) for precision are derived from repeated measurements using established statistical methods.

    8. The Sample Size for the Training Set

    Not applicable. This device is an immunoassay, not a machine learning algorithm that requires a "training set."


    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no "training set" for this type of immunoassay.

    Ask a Question

    Ask a specific question about this device

    K Number
    K984326
    Device Name
    ELECYS IGE ASSAY
    Date Cleared
    1999-02-08

    (67 days)

    Product Code
    Regulation Number
    866.5510
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Product Code :

    JHR

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For in vitro quantitative determination of immunoglobulin E in human serum and plasma. The electrochemiluminescence immunoassay "ECLIA" is intended for use on the Boehringer Mannheim Elecsys 1010 and 2010 immunoassay analyzers. Determination of total serum IgE is useful as an aid in the diagnosis of allergic disease.

    Device Description

    The Elecsys® IgE test is based on the sandwich principle of heterogeneous immunological complex formation. First Step: Biotinylated monoclonal antibodies (R1) and ruthenium labeled antibodies (R2), both specific for IgE, bind IgE present in the sample forming a sandwich complex Second Step: Microparticles coated with streptavidin (M) bind the biotin portion of the complex and are captured magnetically onto the surface of the electrode. Application of a voltage to the electrode induces chemiluminescent emission, which is measured by a photomultiplier.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Elecsys® IgE Test, based on the provided 510(k) summary:

    1. Table of Acceptance Criteria and Reported Device Performance:

    FeatureAcceptance Criteria (Implied / Predicate Performance)Reported Device Performance (Elecsys® IgE Test)
    PrecisionWithin-Run %CV
    Within-Run %CV:
    ~2.1% at 102.7 IU/mL
    ~2.1% at 244.1 IU/mL
    ~2.8% at 375.4 IU/mL3.6% at 5.18 IU/mL
    2.1% at 105 IU/mL
    2.8% at 197 IU/mL
    3.2% at 398 IU/mL
    2.4% at 1010 IU/mL
    Total %CVTotal %CV
    Total %CV:
    ~3.4% at 102.7 IU/mL
    ~3.4% at 244.1 IU/mL
    ~3.6% at 375.4 IU/mL4.2% at 5.18 IU/mL
    1.9% at 105 IU/mL
    3.9% at 197 IU/mL
    3.9% at 398 IU/mL
    3.1% at 1010 IU/mL
    Lower Detection Limit1.44 IU/mL (Predicate)0.10 IU/mL
    Measuring Range1.44 - 500 IU/ml (Predicate)0.10 - 4000 IU/ml
    SpecificityNo cross-reactions with IgG, IgA, and IgM (Predicate)No cross-reactivity with the immunoglobulins G, A, and M was detected.
    Method ComparisonEvidence of strong correlation (e.g., r close to 1) with the predicate device.Elecsys IgE vs Enzymun-Test IgE
    Passing/Bablok: y = -0.53 + 0.97x, r = 1.0, SD (md68) = 2.42, N = 188
    Linear regression: y = -1.85 + 1.01x, r = 1.0, Sy.x = 5.86, N = 188
    Interfering SubstancesNo interference at predicate's concentrations:
    Bilirubin
    Ask a Question

    Ask a specific question about this device

    K Number
    K960067
    Date Cleared
    1996-02-14

    (40 days)

    Product Code
    Regulation Number
    866.5510
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Product Code :

    JHR

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1