K Number
K053501

Validate with FDA (Live)

Device Name
POWERPICC
Date Cleared
2006-01-13

(28 days)

Product Code
Regulation Number
880.5970
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The PowerPICC® Catheters are intended for short or long term peripheral access to the central venous system for intravenous therapy and blood sampling.

The PowerPICC® catheter is indicated for short or long term peripheral access to the central venous system for intravenous therapy and power injection of contrast media. For blood sampling, infusion or therapy, use a 4 French or larger catheter. The maximum recommended infusion rate is 5ml/sec for power injection of contrast media. The maximum pressure of power injectors used with the PowerPICC catheter may not exceed 300 psi.

Device Description

The 6 FR TL PowerPICC® Catheters are open-ended radiopaque polyurethanc. . Catheter usable length is 55 cm. The catheter has a reverse taper design. Catheter shaft tubing is marked with depth indicators, with "0" indicated to serve as a reference for the catheter insertion point. Catheters are provided sterile in basic radiology and nursing PICC configurations. Purple colorants were added to the catheter materials to provide the catheter with an appearance that allows the end user to differentiate Bard's power injectable catheters from other manufacturers' power injectable catheters. The catheter has one power injectable lumen. The product labeling warns against power injection procedures through the two small lumens. The catheter extension legs, junction and clamp ID tags were printed with markings to identify the catheter as PowerPICC® and to include information to facilitate proper use of the device.

AI/ML Overview

This document is a 510(k) Summary for the 6 Fr TL PowerPICC® Catheter, indicating it's a premarket notification for a medical device. The focus of the summary is on demonstrating substantial equivalence to predicate devices rather than proving specific device performance against detailed acceptance criteria from a comprehensive study.

Based on the provided text, the document states:

  • Acceptance Criteria and Reported Device Performance: General statements are made that "The 6 Fr TL PowerPICC® catheter meets all the prodetermined performance acceptance criteria of the testing performed" and that "Performance data gathered in design verification testing demonstrated that the 6 Fr TL PowerPICC® catheter is substantially equivalent to the predicate devices." However, specific quantitative acceptance criteria or detailed reported performance metrics are not provided in this summary. The summary refers to compliance with various ISO standards and FDA guidance documents, implying that performance was evaluated against the requirements outlined in those documents.

A table of acceptance criteria and reported device performance cannot be generated from the given text because the specific criteria and corresponding performance metrics are not explicitly listed.

Here's an analysis of the other requested information based on the text:

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Sample size used for the test set: Not specified in the provided text.
  • Data provenance: Not specified in the provided text. The testing was "in-house," implying it was conducted by Bard Access Systems, Inc.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • This information is not applicable and not provided. The study described is a device verification and validation study, not a study involving human interpretation of data where "ground truth" would be established by experts in the typical clinical sense (e.g., in an AI diagnostic device).

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Not applicable and not provided. This type of adjudication is relevant for studies involving human reviewers or AI interpretations, which is not the nature of this device's performance testing.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No, an MRMC comparative effectiveness study was not done. This document describes the 510(k) submission for a physical medical device (PICC catheter), not an AI-powered diagnostic tool. Therefore, the concept of human readers improving with or without AI assistance is not relevant.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not applicable. This device is a physical catheter, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • Not applicable in the conventional sense of establishing ground truth for diagnostic accuracy. The "ground truth" for this device would be established by direct physical measurements and engineering tests to ensure the catheter meets design specifications (e.g., flow rates, pressure resistance, material compatibility, strength). The text mentions performance data was gathered from "design verification testing" based on international standards (ISO) and FDA guidance for intravascular catheters. This implies the "ground truth" was derived from predefined engineering and material science specifications and test methodologies rather than clinical outcomes or expert consensus on clinical images/data.

8. The sample size for the training set

  • Not applicable. This is a physical medical device, not an AI model that requires a training set.

9. How the ground truth for the training set was established

  • Not applicable, as there is no training set for a physical device.

{0}------------------------------------------------

6 Fr TL PowerPICC®

53501 1 of 3

Section 6

510(k) Summary

6 Fr TL PowerPICC® Catheter

510(k) Summary of Safety and Effectiveness Information 21CFR 807.92

6.1 Submitter Information

Submitter Name:Bard Access Systems, Inc. (BAS)[Subsidiary of C.R. Bard, Inc.]
Address:5425 W. Amelia Earhart DriveSalt Lake City, UT 84116
Telephone Number:(801) 595-0700, Ext. 7136
Fax Number:(801) 595-5425
Contact Person:Lynn M. Kirchoff
Date of Preparation:December 15, 2005

6.2 Device Name

Device Name:6 Fr TL PowerPICC ® Catheter
Trade Name:6 Fr TL PowerPICC ® Catheter
Common/Usual Name:Peripherally Inserted Central Catheter (PICC)
Classification Panel:General Hospital
Classification Name:80LJS - Long Term Intravascular Catheter21 CFR 880.5970, Class IIPeripherally Inserted Central Catheter (PICC)

6.3 Predicate Device Name(s)

Device Name:6 Fr TL Poly Per-Q-Cath ® PICC Catheter
Trade Name:6 Fr TL Poly Per-Q-Cath ® PICC Catheter
Common/Usual Name:Peripherally Inserted Central Catheter (PICC)
Classification Name:Long Term Intravascular Catheter (80 LJS)
Premarket Notification:K043502, concurrence date-January 14, 2005
Device Name:5 Fr DL PowerPICC® Catheter
Trade Name:5 Fr DL PowerPICC® Catheter
Common/Usual Name:Peripherally Inserted Central Catheter (PICC)
Classification Name:Long Term Intravascular Catheter (80 LJS)
Premarket Notification:K051672, concurrence date-November 23, 2005

6.4 Device Description

  • The 6 FR TL PowerPICC® Catheters are open-ended radiopaque polyurethanc. .
  • . Catheter usable length is 55 cm.
  • . The catheter has a reverse taper design
  • Catheter shaft tubing is marked with depth indicators, with "0" indicated to serve as a reference ◆ for the catheter insertion point.
  • . Catheters are provided sterile in basic radiology and nursing PICC configurations.

Image /page/0/Picture/19 description: The image shows the number 00046. The numbers are in a bold, sans-serif font. The number 46 is also written above the last two digits, but it is smaller in size.

{1}------------------------------------------------

KOS350)

2 of

  • Purple colorants were added to the catheter materials to provide the catheter with an appearance . that allows the end user to differentiate Bard's power injectable catheters from other manufacturers' power injectable catheters.
  • . The catheter has one power injectable lumen.
  • The product labeling warns against power injection procedures through the two small lumens. .
  • The catheter extension legs, junction and clamp ID tags were printed with markings to identify the � catheter as PowerPICC® and to include information to facilitate proper use of the device.

6.5 Intended Use

The PowerPICC® Catheters are intended for short or long term peripheral access to the central venous system for intravenous therapy and blood sampling.

The intended use has not changed.

6.6 Indications for Use

The indications for use have not changed from the predicate 5 Fr DL PowerPICC® catheter (K051672).

The PowerPICC® catheter is indicated for short or long term peripheral access to the central venous system for intravenous therapy and power injection of contrast media. For blood sampling, infusion or therapy, use a 4 French or larger catheter. The maximum recommended infusion rate is 5ml/sec for power injection of contrast media. The maximum pressure of power injectors used with the PowerPICC catheter may not exceed 300 psi.

6.7 Summary of Technological Characteristics in Relation to the Predicate Device

Does the new device have the same technological characteristics, e.g. design, material, etc.?

Not in all regards. The 6 Fr TL PowerPICC® catheter has some minor differences from the predicate devices. However, the basic fundamental scientific technology of the catheter has not changed.

Could the new characteristics affect safety or effectiveness?

Yes. The new characteristics could affect safety or effectiveness of the device.

Do the new characteristics raise new types of safety and effectiveness questions?

No. A failure modes and effects analysis (FMEA) of the subject device was conducted in accordance with an internal protocol based on ISO 14971:2000, Medical Devices - Rich Management for Medical Devices, to assure that risks posed by the subject device are acceptable. The analysis did not raise any new types of safety or effectiveness questions.

Do accepted scientific methods exist for assessing effects of the new characteristics?

Yes. The following FDA guidance documents and international standards were used to evaluate the device's performance:

  • Guidance on Premarket Notification [510(k)] Suhmission for Short-Term and Long-Term . Intravascular Catheters, March 16, 1995
  • . ISO 10555-1:1997, Sterile, Single-Use Intravascular Catheters, General requirements

{2}------------------------------------------------

053051 3 of 3

  • ISO 10555-1:1997, Sterile, Single-Use Intravascular Catheters, General requirements, . Amendment I
  • ISO 10555-3:1997, Sterile, Single-Use Intravascular Catheters, Central venous catheters .
  • . AAMI/ANSI/ISO 11135:1994, Medical Devices - Validation and Routine Control of Ethylene Oxide Sterilization

Design validation was performed to meet the recommendations of the FDA guidance document, Design Control Guidance for Medical Device Manufacturers, dated March 11, 1997.

Biocompatibility requirements of ISO 10993 Biological Evaluation of Medical Devices Part-1 : Evaluation and Testing and the FDA Modified ISO 10993 Test Profile for externally communicating, blood contacting, long-term devices were met. All materials used in the manufacture of the subject device were previously cleared for similar applications by Bard Access Systems.

Are performance data available to assess effects of new characteristics?

Yes. Verification and validation testing was performed according to protocols based on the above referenced guidance document recommendations and standards, as well as in accordance with inhouse protocols.

Do performance data demonstrate equivalence?

Yes. Performance data gathered in design verification testing demonstrated that the 6 Fr TL PowerPICC® catheter is substantially equivalent to the predicate devices, and the risks associated with use of the new device were found acceptable when evaluated by FMEA.

6.8 Conclusion

The 6 Fr TL PowerPICC* catheter meets all the prodetermined performance acceptance criteria of the testing performed and, based on FDA's decision tree, is substantially equivalent to the predicate devices the 6 Fr TL Poly Per-Q-Cath" PICC catheter, K043502, concurrence date, November 23, 2005.

Image /page/2/Picture/13 description: The image shows the number 48 at the top and the number 000048 below it. The number 000048 is written in a bold font. The number 48 at the top is smaller than the number 000048 below it.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH AND HUMAN SERVICES. USA" around the perimeter. Inside the circle is an abstract image of an eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JAN 1 3 2006

Ms. Lynn M. Kirchoff Regulatory Affairs Specialist Bard Access Systems, Incorporated 5425 West Amelia Earhart Drive Salt Lake City, Utah 84116

Re: K053501

Trade/Device Name: 6 Fr TL PowerPICC Catheter Regulation Number: 880.5970 Regulation Name: Percutaneous, Implanted Long-Term Intravascular Cathcter Regulatory Class: II Product Code: LJS Dated: December 15, 2005 Received: December 16, 2005

Dear Ms. Kirchoff:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Ms. Kirchoff

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-frec number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Clus

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

6 Fr TL PowerPICC®

K053501 1 of 1

Section 1.2

Indications for Use Statement

510(k) Number (if known):

Device Name: 6 Fr TL PowerPICC® Catheter

Indications For Use:

The PowerPICC® catheter is indicated for short or long term peripheral access to the central venous system for intravenous therapy and power injection of contrast media. For blood sampling, infusion or therapy, use a 4 French or larger catheter. The maximum recommended infusion rate is 5mVsec for power injection of contrast media. The maximum pressure of power injectors used with the PowerPICC catheter may not exceed 300 psi.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use_ (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Anthen D. Anz

Schooly, Correstal I Dirist Liarter

K453501

§ 880.5970 Percutaneous, implanted, long-term intravascular catheter.

(a)
Identification. A percutaneous, implanted, long-term intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings, such as luer hubs, and accessories that facilitate the placement of the device. The device allows for repeated access to the vascular system for long-term use of 30 days or more, and it is intended for administration of fluids, medications, and nutrients; the sampling of blood; and monitoring blood pressure and temperature. The device may be constructed of metal, rubber, plastic, composite materials, or any combination of these materials and may be of single or multiple lumen design.(b)
Classification. Class II (special controls) Guidance Document: “Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters.”