Search Filters

Search Results

Found 9 results

510(k) Data Aggregation

    K Number
    K162441
    Manufacturer
    Date Cleared
    2017-04-24

    (236 days)

    Product Code
    Regulation Number
    880.5970
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PowerPICC Provena Catheters with SOLO² Valve Technology are indicated for short or long-term peripheral access to the central venous system for intravenous therapy, blood sampling, power injection of contrast media, and allows for central venous pressure monitoring. For central venous pressure monitoring, it is recommended that a catheter lumen of 20 gauge or larger be used.

    Device Description

    Bard Access Systems, Inc.'s PowerPICC Provena Catheters with SOLO² Valve Technology are sterile, single use devices designed to provide access to the patient's vascular system. The devices are intended for long or short-term use, as clinically indicated, to sample blood and administer fluids intravenously. The catheters are capable of central venous pressure monitoring, and can withstand power injection of contrast media. The catheters are peripherally inserted central catheters (PICC) and utilize the same placement technique as the predicate device. The subject devices include a silicone valve on the proximal end. The subject devices included in this notification are of varying French size and catheter configuration types, as summarized in the table below.

    Summary of Subject Devices
    Catheter Configuration | French size (Number of Lumens)
    PowerPICC Provena Catheters with SOLO² Valve Technology | 3 French Single Lumen (SL)
    PowerPICC Provena Catheters with SOLO² Valve Technology | 4 French Dual Lumen (DL)

    The following device descriptors apply to all French sizes and configurations of the subject catheters:
    ● Catheters are open-ended, radiopaque polyurethane;
    ● Catheters incorporate a silicone valve on the proximal end;
    ● Catheters have a reverse taper design;
    ● Catheter shaft tubing is marked with depth indicators, with "0" indicated to serve as a reference for the catheter insertion point;
    ● Purple colorant is included in the catheter material to provide the catheter with an appearance that allows the end user to differentiate Bard's power injectable catheters from other manufacturers' power injectable catheters;
    ● Catheter extension legs, luer hubs, and junction are printed with markings to identify the catheter as PowerPICC Provena Catheters with SOLO² Valve Technology, and include information to facilitate proper use of the device.

    The subject devices are provided sterile in basic interventional radiology (IR) kits, as well as basic, full, and max barrier nursing PICC kits with legally marketed components to assist in the placement procedure. These kits are available in both standard and small patient versions.

    AI/ML Overview

    The provided text describes specific performance tests conducted for the PowerPICC Provena Catheters with SOLO² Valve Technology to demonstrate substantial equivalence to predicate devices, rather than an AI/ML medical device. Therefore, the information provided does not align with the request to describe acceptance criteria and a study proving an AI/ML device meets those criteria.

    However, based on the provided text, I can infer the "acceptance criteria" from the "Risk Acceptability Criteria (Acceptance Criteria of Test)" column and the reported performance from the "Verification / Validation Method" column. The study described is a series of in-house and standard-based performance tests, not an MRMC or standalone AI study.

    Here's an interpretation based on the provided content, acknowledging its limitations regarding AI/ML:

    This document describes the testing and verification of a medical catheter, not an AI/ML-driven device. Therefore, many of the requested criteria specific to AI/ML systems (sample size for test/training sets, data provenance, expert ground truth, MRMC studies, effect size, standalone performance) are not applicable or cannot be extracted from the provided text.

    However, I can extract the acceptance criteria and stated performance for the catheter device as presented:


    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Risk Acceptability Criteria)Reported Device Performance (Verification / Validation Method)
    Mechanical Hemolysis Test: Testing to determine the hemolytic properties when blood is aspirated through the catheter assembly. • Bard internal standards and proceduresTesting to determine the hemolytic properties when blood is aspirated through the catheter assembly. (Implies satisfactory performance as per the concluding statement: "The subject devices met all predetermined acceptance criteria...")
    Dimensional Test: Test to measure OD and ID for single lumen catheters and OD and lumen area for dual lumen catheters to ensure compliance with dimensional specification. • Bard internal standards and procedures and ISO 10555-1:2013 - Sterile Single-Use Intravascular Catheters - Part 1: General requirementsTest to measure OD and ID for single lumen catheters and OD and lumen area for dual lumen catheters to ensure compliance with dimensional specification. (Implies satisfactory performance as per the concluding statement: "The subject devices met all predetermined acceptance criteria...")
    Catheter Collapse Test: Test to measure the flow rate of aspiration and demonstrate that the catheter will not collapse under a vacuum. • Bard internal standards and procedures and FDA guidance, Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters (March 16, 1995)Test to measure the flow rate of aspiration and demonstrate that the catheter will not collapse under a vacuum. (Implies satisfactory performance as per the concluding statement: "The subject devices met all predetermined acceptance criteria...")
    Luer to Extension Leg Tensile Test: Test to demonstrate the peak tensile force of each test piece exceeds the minimum peak tensile force. • ISO 10555-1:2013 – Sterile Single-Use Intravascular Catheters - Part 1: General requirementsTest to demonstrate the peak tensile force of each test piece exceeds the minimum peak tensile force. (Implies satisfactory performance as per the concluding statement: "The subject devices met all predetermined acceptance criteria...")
    Priming Volume: Test to measure the volume required to prime a full length catheter. • FDA guidance, Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters (March 16, 1995)Test to measure the volume required to prime a full length catheter. (Implies satisfactory performance as per the concluding statement: "The subject devices met all predetermined acceptance criteria...")
    Pump Flow: Test to determine the maximum pressure generated by the catheter when infusing water through it at a maximum pump flow rate. • Bard internal standards and proceduresTest to determine the maximum pressure generated by the catheter when infusing water through it at a maximum pump flow rate. (Implies satisfactory performance as per the concluding statement: "The subject devices met all predetermined acceptance criteria...")
    Gravity Flow: Test to measure the gravity flow rate. • Bard internal standards and procedures and FDA guidance, Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters (March 16, 1995) • ISO 10555-1:2013 – Sterile Single-Use Intravascular Catheters - Part 1: General requirementsTest to measure the gravity flow rate. (Implies satisfactory performance as per the concluding statement: "The subject devices met all predetermined acceptance criteria...")

    The document states: "The subject devices met all predetermined acceptance criteria derived from the above listed references and demonstrated substantial equivalence as compared to the cited predicate device."


    AI/ML Specific Questions (Cannot be answered from the provided text as it's not an AI/ML device)

    1. Sample sizes used for the test set and data provenance: Not applicable. This is a physical device.
    2. Number of experts used to establish the ground truth for the test set and qualifications: Not applicable. Ground truth for a physical device is established through physical and mechanical testing against standards.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not applicable.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Ground truth for this physical device is established through meeting material, mechanical, and functional performance specifications defined by internal standards, ISO standards (e.g., ISO 10555-1:2013), and FDA guidance documents.
    7. The sample size for the training set: Not applicable. This is a physical device, not an AI model.
    8. How the ground truth for the training set was established: Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K162443
    Manufacturer
    Date Cleared
    2016-10-25

    (55 days)

    Product Code
    Regulation Number
    880.5970
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PowerPICC Provena Catheters are indicated for short or long-term peripheral access to the central venous system for intravenous therapy, blood sampling, power injection of contrast media, and allows for central venous pressure monitoring. For central venous pressure monitoring, it is recommended that a catheter lumen of 20 gauge or larger be used.

    Device Description

    Bard Access Systems, Inc.'s PowerPICC Provena Catheters are sterile, single use devices designed to provide access to the patient's vascular system. The devices are intended for short- or long-term use (>30 days) to sample blood and administer fluids intravenously. The catheters are capable of central venous pressure monitoring, and can withstand power injection of contrast media. The catheters are peripherally inserted central catheters (PICC) and utilize the same placement technique as the predicate devices.

    The subject devices included in this notification are of varying French size and catheter configuration types, as summarized in the table below.

    Summary of Subject Devices
    French size (Number of Lumens)
    Catheter Configuration
    PowerPICC Provena Catheters
    3 French Single Lumen (SL)
    4 French Dual Lumen (DL)

    The following device descriptors apply to all French sizes and configurations of the subject catheters:

    • Catheters are open-ended, radiopaque polyurethane;
    • Catheters have a reverse taper design;
    • Catheter shaft tubing is marked with depth indicators, with "0" indicated to serve as a reference for the catheter insertion point;
    • Purple colorant is included in the catheter material to provide the catheter with an appearance that allows the end user to differentiate Bard's power injectable catheters from other manufacturers' power injectable catheters; and
    • Catheter extension leg, luer hub, junction, and clamp ID tags are printed with markings to identify the catheter as PowerPICC Provena Catheters, and include information to facilitate proper use of the device.

    The subject devices are provided sterile in basic interventional radiology (IR) as well as basic, full, and max barrier nursing PICC kits with legally marketed components to assist in the placement procedure. These kits are available in both standard and small patient versions.

    AI/ML Overview

    The provided document describes the acceptance criteria and a detailed study for the C.R. Bard, Inc. PowerPICC Provena Catheters (K162443). This is a medical device, and the study involves a series of performance tests to demonstrate substantial equivalence to a predicate device, rather than a clinical study evaluating AI performance.

    Here's the information extracted from the document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document provides a comprehensive list of verification/validation methods and their corresponding risk acceptability criteria. The reported performance for all these tests is a general statement that the device met all predetermined acceptance criteria.

    Verification / Validation MethodRisk Acceptability Criteria (Acceptance Criteria of Test)Reported Device Performance
    Biocompatibility TestingTests to confirm that the catheter is free from biological hazard. Bard internal standards and procedures and ISO 10555-1:2013 - Sterile Single-Use Intravascular Catheters – 1: General requirements. Testing Performed includes: Sensitization, Cytotoxicity, Irritation/Intracutaneous Reactivity, Acute Systemic Injection, Material-mediated Pyrogenicity, Subacute 14 day IV toxicity (Saline), Subacute 14 day IP toxicity (Cottonseed oil), Ames Genotoxicity, Mouse Lymphoma, Rodent Blood Micronucleus, 2 week and 6 week implantation, Hemolysis, Complement Activation.Met all predetermined acceptance criteria
    Cantilever Stiffness TestTest to characterize the catheter shaft stiffness at the distal section using a cantilever bend technique. Bard internal standards and procedures.Met all predetermined acceptance criteria
    Clamp EngagementTest to confirm that the catheter assembly will not leak when the clamp is engaged. Bard internal standards and procedures and ISO 10555-1:2013 – Sterile Single-Use Intravascular Catheters – 1: General requirements.Met all predetermined acceptance criteria
    Mechanical Hemolysis TestTesting to determine the hemolytic properties when blood is aspirated through the catheter assembly. Bard internal standards and procedures.Met all predetermined acceptance criteria
    Leak TestTest to confirm that the catheter assembly will not leak when the distal end of the catheter is occluded. Bard internal standards and procedures and ISO 10555-1:2013 – Sterile Single-Use Intravascular Catheters – 1: General requirements.Met all predetermined acceptance criteria
    Dimensional TestTest to measure OD and ID for single lumen catheters and OD and lumen area for dual lumen catheters to ensure compliance with dimensional specification. Bard internal standards and procedures and ISO 10555-1:2013 - Sterile Single-Use Intravascular Catheters - 1: General requirements.Met all predetermined acceptance criteria
    Implantable LengthTest to measure useful length for catheters to ensure compliance with dimensional specification. Bard internal standards and procedures and ISO 10555-1:2013 – Sterile Single-Use Intravascular Catheters – 1: General requirements.Met all predetermined acceptance criteria
    Extension Leg LengthTest to measure and confirm extension leg length compliance with dimensional specification. Bard internal standards and procedures.Met all predetermined acceptance criteria
    Catheter Collapse TestTest to measure the flow rate of aspiration and demonstrate that the catheter will not collapse under a vacuum. Bard internal standards and procedures and ISO 10555-1:2013 – Sterile Single-Use Intravascular Catheters – 1: General requirements.Met all predetermined acceptance criteria
    Burst TestBurst pressure test to confirm the catheter burst pressure exceeds the peak pressure present in the catheter at maximum flow conditions when the distal end is occluded. Bard internal standards and procedures and ISO 10555-1:2013 - Sterile Single-Use Intravascular Catheters - Part 1: General requirements.Met all predetermined acceptance criteria
    OD SwellTest to confirm that the catheter does not swell beyond twice the size of the labeled OD during power injection. Bard internal standards and procedures.Met all predetermined acceptance criteria
    Hydraulic Catheter Burst TestBurst pressure test to confirm the catheter burst pressure exceeds the peak pressure present in the catheter at maximum flow conditions when the distal end is occluded. Bard internal standards and procedures and ISO 10555-1:2013 - Sterile Single-Use Intravascular Catheters - Part 1: General requirements.Met all predetermined acceptance criteria
    Extension Leg Burst TestBurst pressure test to confirm the extension leg burst pressure exceeds the peak pressure present in the catheter at maximum flow conditions when the extension leg is occluded. Bard internal standards and procedures and ISO 10555-1:2013 - Sterile Single-Use Intravascular Catheters - Part 1: General requirements.Met all predetermined acceptance criteria
    Power Injection ConditioningTest to confirm the catheter does not leak or burst as a result of power injections at maximum indicated flow rate. Bard internal standards and procedures and ISO 10555-1:2013 - Sterile Single-Use Intravascular Catheters - Part 1: General requirements.Met all predetermined acceptance criteria
    Tip Stability TestTest to confirm that the catheter tip remains in the same orientation during power injection (tip pointing in direction of venous flow) at the maximum indicated flow rate. Bard internal standards and procedures.Met all predetermined acceptance criteria
    Assembly Tensile TestTest to demonstrate the peak tensile force of each test piece exceeds the minimum peak tensile force.Met all predetermined acceptance criteria
    Shaft Tensile Test Luer to Extension Leg Tensile TestISO 10555-1:2013 – Sterile Single-Use Intravascular Catheters – Part 1: General requirements.Met all predetermined acceptance criteria
    Shaft Tensile TestTest to evaluate the maximum catheter strain and modulus at break. Bard internal standards and procedures and FDA guidance, Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters (March 16, 1995).Met all predetermined acceptance criteria
    RadiopacityTest to demonstrate catheter radio-detectability. Bard internal standards and procedures and ISO 10555-1:2013 - Sterile Single-Use Intravascular Catheters - Part 1: General requirements.Met all predetermined acceptance criteria
    Suture Wing Integrity TestTest to measure the maximum force a catheter junction suture wing can withstand prior to break. Bard internal standards and procedures and FDA guidance, Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters (March 16, 1995).Met all predetermined acceptance criteria
    Priming VolumeTest to measure the volume required to prime a full length catheter. FDA guidance, Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters (March 16, 1995).Met all predetermined acceptance criteria
    Gravity FlowTest to measure the gravity flow performance of a full length catheter. Bard internal standards and procedures and ISO 10555-1:2013 - Sterile Single-Use Intravascular Catheters - Part 1: General requirements.Met all predetermined acceptance criteria
    Pump FlowTest to determine the maximum pressure generated by the catheter when infusing water through it at a maximum pump flow rate. Bard internal standards and procedures.Met all predetermined acceptance criteria
    Kink Diameter TestTest to quantitatively evaluate kink characteristics associated with the catheter shafts. Bard internal standards and procedures.Met all predetermined acceptance criteria
    Stylet Drag TestTest to ensure that the stylets used to place the catheter can be removed without difficulty. Bard internal standards and procedures.Met all predetermined acceptance criteria
    StatLock CompatibilityTest to ensure the catheter is compatible with StatLock for catheter securement. Bard internal standards and procedures.Met all predetermined acceptance criteria
    Taper LengthTest to measure and confirm taper length compliance with dimensional specification. Bard internal standards and procedures.Met all predetermined acceptance criteria
    Catheter PrintingTest is to confirm ink on the catheter exhibits proper adherence. Bard internal standards and procedures.Met all predetermined acceptance criteria

    The document explicitly states: "The subject devices met all predetermined acceptance criteria derived from the above listed references and demonstrated substantial equivalence as compared to the cited predicate device."

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the exact sample sizes (number of devices tested) for each individual test. It generally references "Bard internal standards and procedures" and relevant ISO standards (e.g., ISO 10555-1:2013) for the testing methodology. These standards typically define acceptable sample sizes for device testing.

    The data provenance is implied to be from in-house testing conducted by Bard Access Systems, Inc. There is no mention of country of origin for data if it were to be from external sources, nor does it specify if the data is retrospective or prospective, as this is a device performance study rather than a clinical trial.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This type of information (number and qualifications of experts) is not relevant for this device performance study. The ground truth for these tests is established by objective, measurable engineering and material science standards and physical properties, not by expert interpretation.

    4. Adjudication Method for the Test Set

    Adjudication methods (like 2+1, 3+1, none) are typically used in clinical studies or studies involving human assessment of data. This document describes a battery of engineering and material science tests. Therefore, no adjudication method as described for clinical data interpretation is applicable or mentioned. The "adjudication" is inherent in the objective measurement against established criteria.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, a multi-reader multi-case comparative effectiveness study was not done. This document describes the safety and performance testing of a medical device (intravascular catheter), not an AI system. Therefore, there is no mention of human readers, AI assistance, or effect sizes related to AI.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    No, a standalone algorithm performance study was not done. This document pertains to the physical performance of a medical catheter, not an algorithm or AI system.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for the tests described is based on objective, measurable engineering and material properties, and established industry standards. Examples include:

    • Physical measurements: OD, ID, length, flow rates, burst pressure, tensile strength.
      • Reference: Dimensional Test, Implantable Length, Extension Leg Length, Burst Test, Hydraulic Catheter Burst Test, Extension Leg Burst Test, Power Injection Maximum Flow Rate, Assembly Tensile Test, Shaft Tensile Test, Suture Wing Integrity Test, Priming Volume, Gravity Flow, Pump Flow, Kink Diameter Test, Taper Length.
    • Chemical/Biological properties: Biocompatibility outcomes (e.g., absence of biological hazard, lack of sensitization, cytotoxicity, irritation, hemolysis).
      • Reference: Biocompatibility Testing, Mechanical Hemolysis Test.
    • Functional performance: Absence of leaks, tip stability during power injection, ability to remove stylets easily, adherence of printing.
      • Reference: Clamp Engagement, Leak Test, Catheter Collapse Test, Power Injection Conditioning, Tip Stability Test, Stylet Drag Test, Catheter Printing.
    • Compliance with published standards: ISO 10555-1:2013 and FDA guidance documents.

    8. The sample size for the training set

    This is not applicable. This is a device performance study, not an AI model development or validation study. There is no concept of a "training set" in this context.

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for an AI model in this document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K151985
    Manufacturer
    Date Cleared
    2016-06-14

    (333 days)

    Product Code
    Regulation Number
    880.5970
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PowerPICC® EtOH Catheter is indicated for short- or long-term peripheral access to the central venous system for intravenous therapy including prolonged exposure to intraluminal solutions containing up to 70% ethanol, power injection of contrast media, and allows for central venous pressure monitoring. For blood sampling, infusion, or therapy, use a 4 French or larger catheter. For central venous pressure monitoring, it is recommended that a catheter lumen of 20 gauge or larger be used.

    The PowerPICC SOLO® EtOH Catheter is indicated for short- or long-term peripheral access to the central venous system for intravenous therapy including prolonged exposure to intraluminal solutions containing up to 70% ethanol, power injection of contrast media, and allows for central venous pressure monitoring. For blood sampling, infusion, or therapy, use a 4 French or larger catheter. For central venous pressure monitoring, it is recommended that a catheter lumen of 20 gauge or larger be used.

    Device Description

    Bard Access Systems, Inc.'s PowerPICC® EtOH and PowerPICC SOLO® EtOH Catheters are sterile, single use devices designed to provide access to the patient's vascular system. The devices are intended for short- or long-term use (>30 days) to sample blood and administer fluids intravenously. The catheters are compatible with intraluminal solutions containing up to 70% ethanol, capable of central venous pressure monitoring, and can withstand power injection of contrast media. The catheters are peripherally inserted central catheters (PICC) and utilize the same placement technique as the predicate devices.

    The subject devices included in this notification are of varying French size and catheter configuration types, as summarized in the table below.

    Summary of Subject Devices
    Catheter Configuration | French size (Number of Lumens)
    PowerPICC® EtOH | 4F (Single Lumen (SL))
    | 5F (Dual Lumen (DL)) FT
    | 5F (Triple Lumen (TL))
    PowerPICC SOLO®2 EtOH | 4F (Single Lumen (SL))
    | 5F (Dual Lumen (DL)) FT
    | 5F (Triple Lumen (TL))

    The subject catheters will be packaged with legally marketed components used in the placement procedure.

    The following device descriptors apply to all French sizes and configurations of the subject Bard Access PowerPICC® EtOH and PowerPICC SOLO®2 EtOH Catheters:

    • The catheters are open-ended, radiopaque polyurethane;
    • The catheters have a reverse taper design;
    • Each catheter configuration has one power injectable lumen;
    • Catheter shaft tubing is marked with depth indicators, with "0" indicated to serve as a reference for the catheter insertion point;
    • Catheters are provided sterile in basic interventional radiology (IR), and basic, full, and max barrier nursing PICC configurations with legally marketed kit components;
    • Purple colorant is included in the catheter material to provide the catheter with an appearance that allows the end user to differentiate Bard's power injectable catheters from other manufacturers' power injectable catheters;
    • Yellow colorant was added to the catheter junction material to provide the catheter with an appearance that allows the end user to differentiate Bard's ethanol compatible catheters from other catheters not compatible with ethanol;
    • The catheter extension leg, luer hub, junction, and clamp ID tags are printed with markings to identify the catheter as PowerPICC® EtOH or PowerPICC SOLO®2 EtOH, and include information to facilitate proper use of the device; and
    • The PowerPICC® EtOH catheter clamp and PowerPICC SOLO®2 EtOH luer are labeled with EtOH to indicate ethanol compatibility.

    The following device descriptors apply to all French sizes of the subject Bard Access Systems PowerPICC® EtOH Catheters:

    • Yellow colorant was added to the catheter extension leg clamp material to provide the catheter with an appearance that allows the end user to differentiate Bard's ethanol compatible catheter from other catheters that are not compatible with ethanol.

    The following device descriptors apply to all French sizes of the subject Bard Access Systems PowerPICC SOLO®2 EtOH Catheters:

    • The PowerPICC SOLO®2 EtOH catheter is a clampless, proximally valved catheter, and
    • Yellow colorant was added to the lower portion of the luer hub to identify the catheter as a PowerPICC SOLO®2 EtOH catheter that is ethanol compatible.

    The following device descriptors apply to specific French sizes of the subject Device Bard Access Systems PowerPICC® EtOH and PowerPICC SOLO®2 EtOH Description Catheters:

    • The 5F TL catheter product labeling warns against power injection procedures through the two small lumens,
    • The 5F TL catheter's usable length is 50 cm,
    • The 5F DL FT and 4F SL catheter's usable length is 55 cm,
    • The 5 F DL FT catheter has a double taper design, and
    • The 5 F DL FT catheter product labeling warns against trimming the catheter in the "No Trim Zone" to maintain product functionality.
    AI/ML Overview

    The provided FDA submission for the PowerPICC® EtOH and PowerPICC SOLO®2 EtOH Catheters does not contain a table of acceptance criteria and reported device performance in the typical format one might expect for a detailed performance study. Instead, it states that "The subject devices met all predetermined acceptance criteria derived from the above listed references and demonstrated substantially equivalent performance as compared to the cited predicate devices."

    The submission describes the types of tests performed and the standards referenced, implying that the acceptance criteria are linked to these standards and the performance of the predicate device. However, it does not explicitly list the specific numerical acceptance criteria (e.g., "burst pressure > X psi") or the exact reported performance values.

    Here's a breakdown of the requested information based on the provided text:


    1. Table of Acceptance Criteria and Reported Device Performance

    As mentioned above, a direct table is not present. The document generally states that "The subject devices met all predetermined acceptance criteria... and demonstrated substantially equivalent performance." The acceptance criteria would have been derived from the referenced standards and safety/performance tests.

    Acceptance Criteria CategoryGeneral Description of Acceptance Criteria (Inferred from context)Reported Device Performance
    Material/Catheter PropertiesConformance to specifications for open-ended, radiopaque polyurethane, reverse taper design, depth indicators, colorants, markings.Met all predetermined acceptance criteria.
    Mechanical PerformanceDemonstrated appropriate assembly leak, tensile strength, elongation, modulus, fatigue, ISO Luer compliance, burst pressure, radiopacity, catheter leak under vacuum, and valve function.Met all predetermined acceptance criteria.
    FunctionalityDemonstrated required gravity flow, priming volume, power injection capability, central venous pressure monitoring, stylet compatibility, and ink adherence.Met all predetermined acceptance criteria.
    BiocompatibilityConformance to ISO 10993 series for cytotoxicity, pyrogenicity, subacute/subchronic toxicity, sensitization, irritation/intracutaneous reactivity, acute systemic toxicity, subchronic/subcutaneous implant toxicity, genotoxicity, implantation studies, hemolysis, prothrombin time, unactivated thromboplastin time, dog thrombogenicity, platelet and leukocyte counts, and compliment activation, chronic toxicity.Met all predetermined acceptance criteria.
    Sterilization & PackagingConformance to ISO 11135 and ISO 11607 standards for validation and routine control of ethylene oxide sterilization and packaging requirements.Met all predetermined acceptance criteria.
    Risk ManagementConducted in accordance with ISO 14971:2012.Successfully completed.

    2. Sample size used for the test set and the data provenance

    • Sample Size: Not explicitly stated for any of the tests. The document mentions "Verification and validation tests have been performed," but does not provide specific sample quantities (e.g., number of catheters tested for burst pressure).
    • Data Provenance: The tests are referred to as "in-house protocols" performed by Bard Access Systems. The country of origin is not specified but implicitly would be the USA given the company's address and FDA submission. The nature of these tests (retrospective or prospective) is not discussed but for device performance tests, they would inherently be prospective (i.e., conducted on newly manufactured devices).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This type of information (expert review for ground truth) is typically relevant for AI/ML device submissions, particularly those involving image analysis or diagnostic interpretation. For a physical medical device like a catheter, "ground truth" is established by direct physical, chemical, and biological testing against established standards and validated methods. Therefore, this section is not applicable in the context of this device and submission. The "experts" would be the engineers, scientists, and technicians conducting and evaluating the performance and safety tests, and their qualifications are not detailed in this submission.

    4. Adjudication method for the test set

    Similarly, adjudication methods (like 2+1 or 3+1 consensus) are typically for subjective interpretations, such as clinical reads in AI/ML performance studies. For physical device performance tests, the results are generally objective and quantitative (e.g., a burst pressure value). Therefore, an explicit adjudication method is not applicable to the performance and safety tests described here. Results are evaluated against predetermined acceptance criteria.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    An MRMC study is relevant for AI-assisted diagnostic devices. This submission pertains to a physical intravascular catheter. Therefore, an MRMC study was not conducted, and this section is not applicable.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This refers to the performance of an AI algorithm without human involvement. Since this is a physical medical device, not an AI/ML algorithm, a standalone algorithm performance study was not applicable and therefore not done.

    7. The type of ground truth used

    For this physical medical device, the "ground truth" for showing it meets acceptance criteria is established through:

    • Engineering Specifications: Conformance to dimensional tolerances, material composition, etc.
    • Validated Test Methods: The results of performance tests (e.g., burst pressure, flow rate, tensile strength) conducted according to recognized industry standards (ISO, ASTM) and internal protocols.
    • Biocompatibility Standards: Assessment against the criteria outlined in the ISO 10993 series for biological safety.
    • Predicate Device Performance: The primary comparison for substantial equivalence relies on demonstrating that the new device performs similarly to or better than legally marketed predicate devices, especially for existing functionalities. For new functionalities (like ethanol compatibility), dedicated tests (e.g., material degradation after ethanol exposure, though not explicitly detailed here) would establish its ground truth against the claim.

    8. The sample size for the training set

    This question relates to the development of AI/ML models. Since this submission is for a physical medical device and does not involve AI/ML, there is no training set in this context, and thus no sample size for a training set.

    9. How the ground truth for the training set was established

    Again, this is applicable to AI/ML device development. As there is no AI/ML component, there is no training set and therefore no ground truth established for a training set in this context.

    Ask a Question

    Ask a specific question about this device

    K Number
    K102159
    Manufacturer
    Date Cleared
    2010-11-17

    (107 days)

    Product Code
    Regulation Number
    880.5970
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PowerPICC® SV catheters are intended for short or long term peripheral access to the central venous system for intravenous therapy and blood sampling.
    The PowerPICC® SV catheter is indicated for short or long-term peripheral access to the central venous system for intravenous therapy, blood sampling, power injection of contrast media, and allows for central venous pressure monitoring. For central venous pressure monitoring, it is recommended that a catheter lumen of 20 gauge or larger be used.

    Device Description

    The subject PowerPICC® SV catheters are members of Bard Access Systems' PowerPICC® series of power injectable catheters. The PowerPICC® SV catheters are available in 3F single lumen and 4F dual lumen configurations. The catheter extension leg, luer hub and junction are printed with description markings to facilitate proper use of the device. The PowerPICC® SV catheters are provided in sterile kit configurations.

    AI/ML Overview

    The provided text does not contain information about an AI device or a study comparing human readers with and without AI assistance. The document is a 510(k) summary for a medical device called the PowerPICC® SV Catheter, a peripherally inserted central catheter (PICC). It details the device's characteristics, intended use, indications for use, and a summary of verification and validation activities demonstrating substantial equivalence to predicate devices, but does not involve AI, image analysis, or multi-reader studies.

    Therefore, I cannot fulfill most of your request regarding AI-related performance metrics and study details. However, I can extract the general acceptance criteria and the type of study conducted for the physical medical device as described.

    Here's the information that can be extracted from the provided text:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria CategoryReference/MethodReported Device Performance
    General Performance- Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters, March 16, 1995 - ISO 10555-1:2009, Sterile, single-use intravascular catheters, Part 1. General requirements - BS/EN/ISO 10555-3:1996/Cor 1:2002, Sterile, single-use intravascular catheters, Part 3. Central venous catheters"The subject PowerPICC® SV catheters met all predetermined acceptance criteria derived from the above mentioned references."
    Biocompatibility- AAMI/ANSI/ISO 10993-1:2009, Biological Evaluation of Medical Devices Part 1: Evaluation and Testing, and the FDA Modified ISO 10993 Test ProfileImplied to meet criteria, as stated "met all predetermined acceptance criteria."
    Sterilization- AAMI/ANSI/ISO 10993-7:2008, Biological Evaluation of Medical Devices Part 7: Ethylene Oxide Sterilization Results - AAMI/ANSI/ISO 11135:2007, Sterilization of Healthcare Products Ethylene oxide -- Part 1: Requirements for development, validation and routine control of a sterilization process for medical devicesImplied to meet criteria, as stated "met all predetermined acceptance criteria."
    Design ValidationIn-house protocols and design validation activities"Design validation was conducted on the subject PowerPICC® SV configuration and yielded acceptable results."
    Risk ManagementISO 14971:2007, Medical Devices - Risk Management for Medical Devices (Failure Modes and Effects Analysis - FMEA)"No new types of safety or efficacy questions were identified for the subject PowerPICC® SV catheters." (This indicates acceptable risk profile within regulatory norms)
    Flow Rate (Indications for Use)Not a test, but stated as part of the intended use parameters.3 F Single Lumen: 1 mL/sec 4 F Dual Lumen: 2.5 mL/sec

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document states, "Tests were performed on sterilized, finished devices." However, it does not specify the sample size for the test set or the data provenance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. This is a submission for a physical medical device, not an AI or diagnostic imaging device requiring expert ground truth for classification.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This is a submission for a physical medical device, not an AI or diagnostic imaging device requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC study was done, as this is not an AI device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an AI algorithm. The performance described is for the physical PowerPICC® SV Catheter itself.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For a physical device like a catheter, "ground truth" generally relates to adherence to physical specifications, material properties, sterility, and functional performance benchmarks (e.g., flow rates, tensile strength, biocompatibility). The document indicates performance was evaluated against "predetermined performance specifications" derived from various international standards (ISO, AAMI/ANSI) and FDA guidance documents. This can be considered the "ground truth" or benchmark against which the device's physical and functional attributes were measured.

    8. The sample size for the training set

    Not applicable. There is no "training set" in the context of this traditional medical device submission.

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K072230
    Device Name
    POWERPICC SOLO
    Manufacturer
    Date Cleared
    2007-10-05

    (56 days)

    Product Code
    Regulation Number
    880.5970
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PowerPICC SOLO™ catheters are intended for short or long term peripheral access to the central venous system for intravenous therapy and blood sampling.
    The PowerPICC SOLO™ catheter is indicated for short or long term peripheral access to the central venous system for intravenous therapy, power injection of contrast media and allows for central venous pressure monitoring. For blood sampling, infusion or therapy, use a 4 French or larger catheter. The maximum recommended infusion rate is 5ml/sec. For central venous pressure monitoring, it is recommended that catheter lumen of 20 gauge or larger be used.

    Device Description

    The PowerPICC SOLO™ catheter is a clampless proximally valved catheter. The PowerPICC SOLO™ catheters are open-ended radiopaque polyurethane. The PowerPICC SOLO™ catheters are offered in 4 Fr Single Lumen (SL), 5 Fr Single Lumen (SL), 5 Fr Dual Lumen (DL), 6 Fr Dual Lumen (DL), and 6 Fr Triple Lumen (TL) configurations. Catheter usable length is 55 cm. The catheter has a reverse taper design Catheter shaft tubing is marked with depth indicators, with "0" indicated to serve as a reference for the catheter insertion point. Catheters are provided sterile in basic and full PICC configurations with legally marketed kit components. Purple colorants were added to the catheter materials to provide the catheter with an appearance that allows the end user to differentiate Bard's power injectable catheters from other manufacturers' catheters. Lower portion of luer hub is blue to identify the catheter as PowerPICC SOLO™ catheter. The catheter extension leg, luer hub and junction were printed with markings to identify the catheter as PowerPICC SOLO™ and to include information to facilitate proper use of the device.

    AI/ML Overview

    This document describes the premarket notification (510(k)) for the PowerPICC SOLO™ Catheter, specifically addressing its substantial equivalence to previously cleared predicate devices. The information provided heavily focuses on regulatory compliance, device description, and safety testing rather than a clinical study evaluating specific performance metrics against acceptance criteria that are typically found in AI/ML-based device submissions.

    Therefore, many of the requested sections (e.g., acceptance criteria, test set details, expert ground truth, MRMC study, training set details) are not directly applicable or available from the provided text for this non-AI/ML device.

    Here's a breakdown of the available information:

    1. Table of Acceptance Criteria and Reported Device Performance

    This section is largely Not Applicable in the way it's typically understood for AI/ML device performance. For this medical device (a catheter), "acceptance criteria" and "performance" are framed in terms of meeting established industry standards and demonstrating substantial equivalence to predicate devices, rather than achieving specific numerical targets on a test set.

    Criteria CategoryAcceptance Criteria (Not explicit as numerical targets, but compliance with standards)Reported Device Performance and Study Type
    BiocompatibilityCompliance with ISO 10993-1: Evaluation and Testing and FDA Modified ISO 10993 Test Profile for externally communicating, blood contacting, long-term devices.Met. Biocompatibility requirements have been met. (Study type: Laboratory testing against ISO 10993 standards).
    Catheter PerformanceCompliance with various FDA guidance documents and international standards including: - FDA Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters, March 16, 1995 - BS/EN/ISO 10555-1: 1997, Sterile, single-use intravascular catheters, Part 1. General requirements - ISO 10555-1:2004, Sterile, single-use intravascular catheters, Part 1. General requirements, Amendment 2 - ASTM F640-79 (reapproved 2000), Standard Test Methods for Radiopacity of Plastics for Medical Use - BS/EN/ISO 10555-3:1997, Sterile, single-use intravascular catheters, Part 3. Central venous catheters - ISO 594-2: 1998, Conical fittings with a 6% (Luer) taper for syringes, needles and certain other medical equipment - Part 2: Lock Fittings - AAMI/ANSI/ISO 11135:1994, Medical Devices - Validation and Routine Control of Ethylene Oxide Sterilization - IEC 60601-2-34: 2000-10, Medical electrical equipment - Particular requirements for the safety, including essential performance, of invasive blood pressure monitoring equipment - AAMI TIR9: 1992, Evaluation of Clinical Systems for Invasive Blood Pressure Monitoring - ANSI/AAMI BP22: 1994, Blood Pressure TransducersMet. Performance testing was conducted in accordance with the listed documents and standards, yielding "acceptable safety & performance outcomes." (Study type: Laboratory testing against numerous specific standards).
    Substantial EquivalenceDemonstration of substantial equivalence in design, materials, sterilization, principles of operation, and indications for use to current commercially available/cited predicate devices.Achieved. The device "met the minimum requirements that are considered adequate for its intended use and is substantially equivalent in design, materials, sterilization, principles of operation and indications for use to current commercially available catheters/cited predicates." (Study type: Comparison to predicate devices and documentation of compliance with standards).

    2. Sample size used for the test set and data provenance

    Not Applicable (N/A) for this type of device submission. The provided text details engineering and biocompatibility testing against established standards and predicate devices, not a test set of data (e.g., medical images for an AI algorithm). The provenance of data refers to the source of materials or components used in the device, rather than patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not Applicable (N/A). This concept is for AI/ML models that rely on expert annotations for ground truth. This is a physical medical device.

    4. Adjudication method for the test set

    Not Applicable (N/A). Adjudication methods like 2+1 or 3+1 are used for resolving discrepancies in expert labeling for AI/ML ground truth.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not Applicable (N/A). This is not an AI-assisted device, and thus, no MRMC study involving human readers with/without AI assistance was conducted or reported.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not Applicable (N/A). This is a physical catheter, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    Not Applicable (N/A) in the context of clinical "ground truth" for diagnostic or prognostic endpoints. The "ground truth" for this device's performance is its adherence to established international and FDA medical device performance standards and its physical and material properties, as demonstrated through laboratory testing.

    8. The sample size for the training set

    Not Applicable (N/A). This device does not involve a "training set" in the context of machine learning.

    9. How the ground truth for the training set was established

    Not Applicable (N/A). As there is no training set for an AI/ML algorithm, this question is not relevant.

    Ask a Question

    Ask a specific question about this device

    K Number
    K053501
    Device Name
    POWERPICC
    Date Cleared
    2006-01-13

    (28 days)

    Product Code
    Regulation Number
    880.5970
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PowerPICC® Catheters are intended for short or long term peripheral access to the central venous system for intravenous therapy and blood sampling.

    The PowerPICC® catheter is indicated for short or long term peripheral access to the central venous system for intravenous therapy and power injection of contrast media. For blood sampling, infusion or therapy, use a 4 French or larger catheter. The maximum recommended infusion rate is 5ml/sec for power injection of contrast media. The maximum pressure of power injectors used with the PowerPICC catheter may not exceed 300 psi.

    Device Description

    The 6 FR TL PowerPICC® Catheters are open-ended radiopaque polyurethanc. . Catheter usable length is 55 cm. The catheter has a reverse taper design. Catheter shaft tubing is marked with depth indicators, with "0" indicated to serve as a reference for the catheter insertion point. Catheters are provided sterile in basic radiology and nursing PICC configurations. Purple colorants were added to the catheter materials to provide the catheter with an appearance that allows the end user to differentiate Bard's power injectable catheters from other manufacturers' power injectable catheters. The catheter has one power injectable lumen. The product labeling warns against power injection procedures through the two small lumens. The catheter extension legs, junction and clamp ID tags were printed with markings to identify the catheter as PowerPICC® and to include information to facilitate proper use of the device.

    AI/ML Overview

    This document is a 510(k) Summary for the 6 Fr TL PowerPICC® Catheter, indicating it's a premarket notification for a medical device. The focus of the summary is on demonstrating substantial equivalence to predicate devices rather than proving specific device performance against detailed acceptance criteria from a comprehensive study.

    Based on the provided text, the document states:

    • Acceptance Criteria and Reported Device Performance: General statements are made that "The 6 Fr TL PowerPICC® catheter meets all the prodetermined performance acceptance criteria of the testing performed" and that "Performance data gathered in design verification testing demonstrated that the 6 Fr TL PowerPICC® catheter is substantially equivalent to the predicate devices." However, specific quantitative acceptance criteria or detailed reported performance metrics are not provided in this summary. The summary refers to compliance with various ISO standards and FDA guidance documents, implying that performance was evaluated against the requirements outlined in those documents.

    A table of acceptance criteria and reported device performance cannot be generated from the given text because the specific criteria and corresponding performance metrics are not explicitly listed.

    Here's an analysis of the other requested information based on the text:

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample size used for the test set: Not specified in the provided text.
    • Data provenance: Not specified in the provided text. The testing was "in-house," implying it was conducted by Bard Access Systems, Inc.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • This information is not applicable and not provided. The study described is a device verification and validation study, not a study involving human interpretation of data where "ground truth" would be established by experts in the typical clinical sense (e.g., in an AI diagnostic device).

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable and not provided. This type of adjudication is relevant for studies involving human reviewers or AI interpretations, which is not the nature of this device's performance testing.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No, an MRMC comparative effectiveness study was not done. This document describes the 510(k) submission for a physical medical device (PICC catheter), not an AI-powered diagnostic tool. Therefore, the concept of human readers improving with or without AI assistance is not relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This device is a physical catheter, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • Not applicable in the conventional sense of establishing ground truth for diagnostic accuracy. The "ground truth" for this device would be established by direct physical measurements and engineering tests to ensure the catheter meets design specifications (e.g., flow rates, pressure resistance, material compatibility, strength). The text mentions performance data was gathered from "design verification testing" based on international standards (ISO) and FDA guidance for intravascular catheters. This implies the "ground truth" was derived from predefined engineering and material science specifications and test methodologies rather than clinical outcomes or expert consensus on clinical images/data.

    8. The sample size for the training set

    • Not applicable. This is a physical medical device, not an AI model that requires a training set.

    9. How the ground truth for the training set was established

    • Not applicable, as there is no training set for a physical device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K051672
    Device Name
    POWERPICC
    Date Cleared
    2005-11-23

    (153 days)

    Product Code
    Regulation Number
    880.5970
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PowerPICC® Catheters are intended for short or long term peripheral access to the central venous system for intravenous therapy and blood sampling.

    The Power PICC catheter is indicated for short or long term peripheral access to the central venous system for intravenous therapy and power injection of contrast media. For blood sampling, infusion or therapy, use a 4 French or larger catheter. The maximum recommended infusion rate is 5ml/sec for power injection of contrast media. The maximum pressure of power injectors used with the Power PICC catheter may not exceed 300 psi.

    Device Description

    The PowerPICC " Catheters are open-ended radiopaque polyurethane catheters. .
    . Catheter size is 5 Fr DI, with 55 cm usable length.
    . The catheter has a reverse taper design.
    . Catheter shaft tubing is marked with depth indicators, with '0' indicated to serve as a reference for the catheter insertion point.
    . Catheters are provided sterile in radiology and nursing PICC configurations.
    . Purple colorants were added to the catheter materials to provide the catheter with an appearance that allows the end user to differentiate the PowerPICC from other PICC catheters.
    The catheter extension leg, junction and clamp ID tag were printed with markings to identify . the catheter as PowerPICC and to include information to facilitate proper use of the device.

    AI/ML Overview

    The provided text describes a 510(k) submission for the 5 Fr DL PowerPICC® Catheter, comparing it to a predicate device, the 6 Fr DL PowerPICC® Catheter. This is a medical device submission, primarily focused on demonstrating substantial equivalence to a legally marketed predicate device rather than presenting a clinical study or specific acceptance criteria with performance data in the typical sense of an AI/software as a medical device (SaMD) study.

    Therefore, many of the requested points, especially those related to AI/SaMD specific studies (like MRMC studies, standalone algorithm performance, training set details), are not applicable to this 510(k) submission for a physical medical catheter.

    However, I can extract information related to the performance data and the conclusion of substantial equivalence.

    Here's a breakdown of the relevant information:

    Acceptance Criteria and Device Performance

    The document states that the new device (5 Fr DL PowerPICC® Catheter) underwent "Verification and validation testing...according to protocols based on the above referenced guidance document recommendations and standards, as well as in accordance with in-house protocols." It then concludes, "Performance data gathered in design verification testing demonstrated that the 5 Fr DL PowerPICC® catheter is substantially equivalent to the predicate 6 Fr DL PowerPICC catheter, and the risks associated with use of the new device were found acceptable when evaluated by FMEA."

    And, "The 5 Fr DL PowerPICC® catheter meets all the predetermined performance acceptance criteria of the testing performed..."

    Since this is a 510(k) for a physical medical device, the "acceptance criteria" are related to mechanical, material, and biocompatibility performance as outlined in the cited standards and guidance documents. The document does not list specific quantitative acceptance criteria values in a table, nor does it provide a table of reported device performance values. Instead, it provides a qualitative statement of meeting these criteria and substantial equivalence.

    Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria CategoryPredetermined Acceptance Criteria (Implicit from Standards/Guidance)Reported Device Performance
    Mechanical PerformanceMeet requirements of ISO 10555-1:1997, ISO 10555-3:1997 and internal protocols for catheter integrity, flow rates, pressure limits, etc."Performance data gathered in design verification testing demonstrated that the 5 Fr DL PowerPICC® catheter is substantially equivalent to the predicate 6 Fr DL PowerPICC catheter." "meets all the predetermined performance acceptance criteria of the testing performed"
    Material/Design equivalenceBe substantially equivalent to the predicate device (6 Fr DL PowerPICC® Catheter) in design, material, etc., or demonstrate safety/effectiveness of minor differences.Minor differences identified, but "the basic fundamental scientific technology of the catheter has not changed." "The risks associated with use of the new device were found acceptable when evaluated by FMEA."
    BiocompatibilityMeet requirements of ISO 10993 Biological Evaluation of Medical Devices Part-1 and FDA Modified ISO 10993 Test Profile for externally communicating, blood contacting, long-term devices."Biocompatibility requirements...were met. All materials used in the manufacture of the subject device were previously cleared for similar applications by Bard Access Systems."
    SterilizationMeet requirements of AAMI/ANSI ISO 11135:1994 Medical Devices -- Validation and Routine Control of Ethylene Oxide Sterilization.Implied by the statement that the device "meets all the predetermined performance acceptance criteria of the testing performed." (Though not explicitly detailed for sterilization in performance section, it's listed as a standard used).

    Study Details:

    1. Sample size used for the test set and the data provenance:

      • The document does not specify sample sizes for individual tests (e.g., how many catheters were tested for flow rate, burst pressure, etc.).
      • Data provenance is not explicitly mentioned (e.g., country of origin). The testing was done at Bard Access Systems, Inc.
      • The testing appears to be retrospective in the sense that it evaluates a manufactured device against established standards, rather than a prospective clinical trial.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • This question is not applicable in the context of this 510(k) for a physical medical device. "Ground truth" established by experts is typically for diagnostic devices or AI, where a human expert provides the definitive diagnosis or assessment. For a catheter, the "ground truth" is defined by the physical and material specifications and performance against industry standards.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • This question is not applicable. Adjudication methods like 2+1 or 3+1 are used in clinical trials, especially for evaluating diagnostic accuracy where multiple reviewers assess a case and consensus or a tie-breaker is needed. For product verification, performance is measured objectively against defined physical parameters.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No. This is not applicable. An MRMC study is relevant for diagnostic devices, particularly those involving human interpretation of images or other data, often with AI assistance. This submission is for a physical intravascular catheter.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • No. This is not applicable. This refers to AI algorithm performance.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The "ground truth" for this medical device submission is the engineering specifications, material properties, and performance requirements outlined in international standards (ISO, AAMI/ANSI) and FDA guidance documents, along with the demonstrated performance of the predicate device. Compliance is tested against these objective benchmarks.
    7. The sample size for the training set:

      • This is not applicable. There is no "training set" in the context of a physical medical device 510(k) submission. This term relates to machine learning models.
    8. How the ground truth for the training set was established:

      • This is not applicable, as there is no training set for this type of device submission.
    Ask a Question

    Ask a specific question about this device

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PowerPICC™ catheter is indicated for short or long term peripheral access to the central venous system for intravenous therapy, power injection of contrast media, and allows for central venous pressure monitoring. For blood sampling, infusion or therapy, use a 4 French or larger catheter. The maximum recommended infusion rate is 5mL/sec. The maximum pressure of power injectors used with the PowerPICC catheter may not exceed 300 psi. For central venous pressure monitoring, it is recommended that catheter lumen of 20 gauge or larger be used.

    The Poly Per-Q-Cath PICC is indicated for short or long term peripheral access to the central venous system for intravenous therapy, blood sampling and allows for central venous pressure monitoring. For blood therapy, it is recommended that a 4 French or larger catheter be used. For central venous pressure monitoring, it is recommended that catheter lumen of 20 gauge or larger be used.

    The Poly Per-O-Cath® Triple Lumen PICC is indicated for short or long term peripheral access to the central venous system for intravenous therapy, blood sampling and allows for central venous pressure monitoring. For blood therapy, it is recommended that a 4 French or larger catheter be used. For central venous pressure monitoring, it is recommended that catheter lumen of 20 gauge or larger be used.

    PowerHohn and PowerLine Catheters are indicated for short or long term access to the central venous system. They are designed for administering I.V. fluids, blood products, drugs and parenteral nutrition solutions, as well as blood withdrawal, power injection of contrast media and allow for central venous pressure monitoring. The maximum recommended infusion rate is 5mL/sec. The maximum pressure of power injectors used with the PowerHohn and PowerLine catheters may not exceed 300 psi. For central venous pressure monitoring, it is recommended that catheter lumen of 20 gauge or larger be used.

    Device Description

    Subject devices:

    • Catheters range in French size from 3-5 Fr SL; 4-6 Fr DL and 6 Fr TL .
    • Catheter usable length ranges from 40 -60 cm. .
    • Catheters are open-ended catheters extruded from polyurethane material containing barium ● sulfate for radiopacity.
    • The catheter extension legs are polyurethane extrusions. Extension legs are minimum 2.2 in. . in length to promote easy application of occlusive dressings. Each extension leg has a thumb clamp.
    • The luer hub base material is Isoplast polyurethane. .
    • The catheter has a reverse taper design .
    • The user is informed of the gage size in product labeling and it is printed on the luer hub. .
    • The catheter shaft tubing is marked with depth indicators, with "0" indicated to serve as a . reference for the catheter insertion point
    • Catheters are provided sterile and are packaged with legally marketed kit components that are . preferred by clinicians
    AI/ML Overview

    The provided document describes a 510(k) premarket notification for Central Venous Pressure Monitoring (CVPM) capability in various catheter devices, rather than a standalone device with extensive acceptance criteria and a study demonstrating its performance in a clinical setting against a ground truth.

    The document focuses on demonstrating substantial equivalence to existing predicate devices for an expanded indication for use: "allows for central venous pressure monitoring." The key point is that the catheters themselves are not new, but their indicated use now includes CVPM. Therefore, the "acceptance criteria" and "study" are geared towards proving that the existing or slightly modified catheters are suitable for this additional function, specifically focusing on the catheter lumen size for effective CVPM.

    Here's the breakdown of the information requested, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Criteria/StandardReported Device Performance
    CVPM FunctionalityCatheter lumen of 20 gauge or larger for central venous pressure monitoringSubject devices, except for 21 gauge catheter lumens, met the performance criteria of design verification. As a result, the recommendation of catheter lumen of 20 gauge or larger was added to the indications for use.
    Safety and EffectivenessNo change in design affecting safety or effectiveness compared to predicate devicesNo change in design that could affect the safety or effectiveness of the device.
    No new types of safety and effectiveness questions raised by new characteristicsNo new types of safety and effectiveness questions.
    Applicable Standards- IEC 60601-2-34:2000(E) Ed.2 (Medical electrical equipment - Particular requirements for the safety, including essential performance, of invasive blood pressure monitoring equipment)Testing was based on recognized and un-recognized standards. (Implied: device met these standards for the relevant aspects)
    - AAMI TIR: 1992 (Evaluation of Clinical Systems for Invasive Blood Pressure Monitoring)Testing was based on recognized and un-recognized standards. (Implied: device met these standards for the relevant aspects)
    - ANSI/AAMI BP22:1994 (Blood Pressure Transducers)Testing was based on recognized and un-recognized standards. (Implied: device met these standards for the relevant aspects)

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not describe a test set in the traditional sense of a clinical trial with a specific number of patients or samples. The "testing" mentioned is bench testing (laboratory studies). Therefore, there is no information on:

    • Sample size used for a clinical test set.
    • Country of origin of data for a clinical test set.
    • Retrospective or prospective nature of a clinical test set.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This information is not applicable as the document describes bench testing and design verification, not a clinical study involving human readers or expert consensus to establish a clinical "ground truth."

    4. Adjudication Method for the Test Set

    This information is not applicable for the same reasons as #3.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, and Effect Size

    No, an MRMC comparative effectiveness study was not done. The document focuses on the device's capability for CVPM, not on human reader performance with or without AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    No, a standalone algorithm performance study was not done. This submission is for physical catheter devices, not AI algorithms.

    7. The Type of Ground Truth Used

    The "ground truth" used for evaluating the central venous pressure monitoring capability was based on recognized and un-recognized standards for invasive blood pressure monitoring equipment and transducers, likely involving physical measurements and comparisons in a controlled bench setting. This is best described as bench test performance against established engineering or medical device performance standards. It is not clinical pathology, expert consensus on images, or patient outcomes data.

    8. The Sample Size for the Training Set

    This information is not applicable. The document describes physical medical devices (catheters) and their expanded indications, not a machine learning model that would require a "training set."

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable for the same reason as #8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K033389
    Manufacturer
    Date Cleared
    2004-03-19

    (148 days)

    Product Code
    Regulation Number
    880.5970
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PowerPICC catheter is indicated for short or long term peripheral access to the central venous system for intravenous therapy and power injection of contrast media. For blood sampling, infusion or therapy, use a 4 French or larger catheter. The maximum recommended infusion rate is 5cc/sec. The maximum pressure of power injectors used with the PowerPICC catheter may not exceed 300 psi.

    Device Description

    The PowerPICC Catheters are open-ended radiopaque polyurethane catheters. Catheter size is 5 Fr SL with 50 cm usable length. The catheter has a reverse taper design. Catheter shaft tubing is marked with depth indicators, with "0" indicated to serve as a reference for the catheter insertion point. Catheters are provided sterile in basic radiology PICC configurations. Purple colorants were added to the catheter materials to provide the catheter with an appearance that allows the end user to differentiate the PowerPICC from other PICC catheters. The catheter extension leg, junction and clamp ID tag were printed with markings to identify the catheter as Power PICC and to include information to facilitate proper use of the device.

    AI/ML Overview

    The provided text is a 510(k) premarket notification for a medical device (PowerPICC™ Catheters) and does not describe an AI/ML powered device or a study involving such a device. Therefore, I cannot extract the requested information about acceptance criteria and studies related to AI/ML technology from this document.

    The document discusses the substantial equivalence of the PowerPICC™ Catheter to a predicate device, focusing on its intended use, technological characteristics, and conformance to existing medical device regulations and standards. It mentions "verification testing" for performance but does not provide details on specific acceptance criteria, study methodologies, or AI/ML performance metrics.

    For me to provide the requested information, the input text would need to describe a medical device that utilizes AI/ML, along with a study detailing its performance and acceptance criteria.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1