K Number
K043502

Validate with FDA (Live)

Manufacturer
Date Cleared
2005-01-14

(25 days)

Product Code
Regulation Number
880.5970
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Poly Per-Q-Cath® Triple Lumen PICC is indicated for short or long term peripheral access to the central venous system for intravenous therapy and blood sampling. For blood therapy, it is recommended that a 4 French or larger catheter be used.

Device Description

There are no new operating principles. The modified 6 Fr Poly Per-Q-Cath® TL PICC catheter relies on the same basic, fundamental scientific technology as the predicate 6 Fr DL Poly Per-Q Cath PICC. The devices serve as conduit for fluids and blood into. and out of the centry venous system.

AI/ML Overview

The provided text is a 510(k) Summary for a medical device called the "6 Fr Poly Per-Q-Cath®3 Triple Lumen PICC". It focuses on establishing substantial equivalence to a predicate device rather than presenting a performance study with detailed acceptance criteria and results in the format requested.

Therefore, many of the requested sections about acceptance criteria, study details, expert involvement, and ground truth cannot be fully extracted or are not explicitly stated in the document. The document primarily describes the process of demonstrating substantial equivalence through various tests and adherence to standards, but it doesn't provide a table of acceptance criteria with corresponding performance metrics like a standalone study report would.

Here's an attempt to answer as much as possible based on the provided text, with clear indications where information is not available:


Acceptance Criteria and Device Performance

1. A table of acceptance criteria and the reported device performance

The document does not provide a table with specific acceptance criteria and detailed reported device performance. It states that "The 6 Fr Poly Per-Q-Cath® TL PICC meets all the predetermined performance acceptance criteria of the testing performed" and that "Performance data gathered in design verification and validation testing demonstrated that the 6 Fr Poly Per-Q-Cath® TL PICC is substantially equivalent to the predicate 6 Fr DL Poly Per-Q- Cath PICC".

The document emphasizes adherence to guidance documents and international standards for evaluation, which implicitly define the acceptance criteria for various aspects of the device. These include:

  • Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters, March 16, 1995: This document would outline performance requirements for catheters.
  • ISO 10555-1:1997, Sterile, Single-Use Intravascular Catheters, General requirements.
  • ISO 10555-1:1997, Sterile, Single-Use Intravascular Catheters, General requirements, Amendment I.
  • ISO 10555-3:1997, Sterile, Single-Use Intravascular Catheters, Central venous catheters.
  • AAMI/ANSI/ISO 11135:1994, Medical Devices -- Validation and Routine Control of Ethylene Oxide Sterilization.
  • Design Control Guidance for Medical Device Manufacturers, dated March 11, 1997.
  • Biocompatibility requirements of ISO 10993 Biological Evaluation of Medical Devices Part-J : Evaluation and Testing and the FDA Modified ISO 10993 Test Profile for externally communicating, blood contacting, long-term devices.

The device performance is generally stated as having met these requirements and demonstrating substantial equivalence to the predicate device. Specific numerical results for performance metrics are not given.

Acceptance Criteria (Implied by Standards/Guidance)Reported Device Performance
Catheter Performance (e.g., flow rates, material strength, durability, patency)Met predetermined performance acceptance criteria.
Sterilization Effectiveness (Ethylene Oxide)Compliant with AAMI/ANSI/ISO 11135:1994.
BiocompatibilityMet ISO 10993 and FDA Modified ISO 10993 Test Profile.
Design ControlsAdhered to Design Control Guidance for Medical Device Manufacturers.
Safety and EffectivenessRisks associated with use found acceptable when evaluated by FMEA. No new types of safety and effectiveness questions. Substantially equivalent to predicate.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

This information is not provided in the document. The text refers to "design verification and validation testing" but does not detail sample sizes, whether a "test set" in the context of an AI/ML device was used, or the provenance of any data (country, retrospective/prospective). This is a physical medical device, not an AI/ML diagnostic tool, so these types of data considerations are typically not applicable in the same way.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not provided and is not applicable in the context of this device type and submission. Ground truth, in the AI/ML sense, is not relevant for the physical performance testing of a PICC catheter. Evaluation of risks was performed by "FMEA" (Failure Mode and Effects Analysis), which is a common engineering analysis, not typically involving "experts" in the sense of clinical ground truth establishment for diagnostic output.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not provided and is not applicable to the type of device and testing described.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No, a multi-reader, multi-case (MRMC) comparative effectiveness study was not done. This type of study is relevant for diagnostic devices, particularly those involving AI/ML interpretation of images or data that human readers would also evaluate. This submission is for a physical medical catheter, not an AI-powered diagnostic tool. Therefore, the concept of "human readers improve with AI vs without AI assistance" does not apply.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

No, a standalone (algorithm only) performance study was not done. This is not an AI/ML algorithm-based device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The concept of "ground truth" as typically applied to AI/ML diagnostic devices is not applicable to this physical catheter device. The "ground truth" for evaluating this device's performance would be the objective measurements and observations against established engineering specifications, safety standards, and performance test protocols (e.g., measuring flow rates, material tensile strength, sterility assurance levels, biocompatibility test results against defined toxicity thresholds).

8. The sample size for the training set

This information is not provided and is not applicable. This is not an AI/ML device that requires a training set.

9. How the ground truth for the training set was established

This information is not provided and is not applicable. This is not an AI/ML device that requires a training set or ground truth in that context.

{0}------------------------------------------------

6 Fr Poly Per-

Section 6

510(k) Summary

6 Fr Poly Per-Q-Cath 83 Triple Lumen PICC

510(k) Summary of Safety and Effectiveness Information 21CFR 807.92

6.1 Submitter Information

Submitter Name:Bard Access Systems, Inc. (BAS)
[Subsidiary of C.R. Bard, Inc.]
Address:5425 W. Amelia Earhart Drive
Salt Lake City, UT 84116
Telephone Number:(801) 595-0700, Ext. 7136
Fax Number:(801) 595-5425
Contact Person:I... M. Kimhoff
Date of Preparation:16 December, 2004

6.2 Device Name

Device Name:Poly Per-Q-Cath®3 Triple Lumen PICC Catheter
Trade Name:Poly Per-Q-Cath®3 Triple Lumen PICC Catheter
Common/Usual Name:Peripherally Inserted Central Catheter (PICC)
Classification Panel:General Hospital
Classification Name:80LJS - Long Term Intravascular Catheter21 CFR 880.5970, Class IIPeripherally Inserted Central Catheter (PICC)

6.3 Predicate Device Name

Device Name:Poly Per-Q-Cath® PICC Catheter
Trade Name:Poly Per-Q-Cath® PICC Catheter
Common/Usual Name:Peripherally Inserted Central Catheter (PICC)
Classification Name:Long Term Intravascular Catheter (80 LJS)
Premarket Notification:K034019, concurrence date -- January 21, 2004

6.4 Device Description

Principle of Operation

There are no new operating principles. The modified 6 Fr Poly Per-Q-Cath® TL PICC catheter relies on the same basic, fundamental scientific technology as the predicate 6 Fr DL Poly Per-Q Cath PICC. The devices serve as conduit for fluids and blood into. and out of the centry venous system.

6.5 Intended Use

The Triple Lumen PICC is intended for short or long term peripheral access to the central venous system for intravenous therapy and blood sampling.

The intended use has not changed.

Indications for Use 6.6

{1}------------------------------------------------

The indications for use have not changed from the predicate Poly Per-Q-Cath PICC (K034019).

The Poly Per-Q-Cath® Triple Lumen PICC is indicated for short or long term peripheral access to the central venous system for intravenous therapy and blood sampling. For blood therapy, it is recommended that a 4 French or larger catheter be used.

6.7 Summary of Technological Characteristics in Relation to the Predicate Device

Does the new device have the same technological characteristics, e.g. design, material, etc.?

Not in all regards. The 6 Fr Poly Per-Q-Cath® TL PICC has some minor differences from the predicate 6 Fr DI. Poly Per-Q-Cath PICC. However, the basic fundamental scientific technology of the catheter has not changed.

Could the new characteristics affect safety or effectiveness?

ires. The new characteristics could affect safety or effectiveness of the davice,

Do the new characteristics raise new types of safety and effectiveness questions?

No. There are no new types of issues of safety and effectiveness.

Do accepted scientific methods exist for assessing effects of the new characteristics?

Yes. The following FDA guidance documents and international standards were used to evaluate the device's performance:

  • Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term . Intravascular Catheters, March 16, 1995
  • ISO 10555-1:1997, Sterile, Single-Use Intravascular Catheters, General requirements .
  • ISO 10555-1:1997, Sterile, Single-Use Intravascular Catheters, General requirements, . Amendment |
  • . ISO 10555-3:1997, Sterile, Single-Use Intravascular Catheters, Central venous catheters
  • AAMI/ANSI/ISO 11135:1994, Medical Devices -- Validation and Routine Control of Ethylene . Oxide Sterilization

Design validation was performed to meet the recommendations of the FDA guidance document, Design Control Guidance for Medical Device Manufacturers, dated March 11, 1997.

Biocompatibility requirements of ISO 10993 Biological Evaluation of Medical Devices Part-J : Evaluation and Testing and the FDA Modified ISO 10993 Test Profile for externally communicating, blood contacting, long-term devices were met. All materials used in the manufacture of the subject device were previously cleared for similar applications by Bard Access Systems.

Are performance data available to assess effects of new characteristics?

Yes. Verification and validation testing was performed according to protocols based on the above referenced guidance document recommendations and standards, as well as in accordance with inhouse protocols.

Do performance data demonstrate equivalence?

Yes. Performance data gathered in design verification and validation testing demonstrated that the 6 Fr Poly Per-Q-Cath® TL PICC is substantially equivalent to the predicate 6 Fr DL Poly Per-Q-

Image /page/1/Picture/23 description: The image shows the number 000034. The number is written in a bold, sans-serif font. The numbers are all the same size and are evenly spaced. The background of the image is white.

{2}------------------------------------------------

KC4 3562

6 Fr Poly Per-Q-Cath 53 TL PIC Special 510(k)

Cath PICC, and the risks associated with use of the new device were found acceptable when evaluated by FMFA.

Conclusion 6.8

The 6 Fr Poly Per-Q-Cath** TI. PICC meets all the predetermined performance acceptance criteria of the testing performed and, based on FDA's decision tree, is substantially equivalement the predicate device 6 Fr DL Poly Per-Q-Cath PICC, K034019, cleared January quivalen

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is an abstract image of an eagle or other bird-like figure, represented by three stylized lines.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JAN 1 4 2005

Ms. Lynn M. Kirchoff Regulatory Affairs Specialist Bard Access Systems, Incorporated 5425 West Amelia Earhart Drive Salt Lake City, Utah 84116

Re: K043502

Trade/Device Name: 6 Fr Poly Per-Q-Cath®3 Triple Lumen PICC Regulation Number: 880.5970 Regulation Name: Percutaneous, Implanted Long-Term Intravascular Catheter Regulatory Class: II Product Code: LJS Dated: December 17, 2004 Received: December 20, 2004

Dear Ms. Kirchoff:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Ms. Kirchoff

Please be advised that FDA's issuance of a substantial equivalence determination does not I lease be auvisod that I DIT of issuaires on that your device complies with other requirements Incall that I DX mas made a astes and regulations administered by other Federal agencies. of the Act of ally I oderal base species and systements, including, but not limited to: registration 1 ou mast comply with and the rise ling (21 CFR Part 801); good manufacturing practice allu listing (21 CFR Part 807), labelity systems (QS) regulation (21 CFR Part 820); and if requirements as set form in the quarty is a control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) Fins letter will anow Jourse of Substantial equivalence of your device to a premarket notification. - The PDF results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), If you desire specific advice to: Joan ee at (301) 594-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the may obtain of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Chiu Lin, Ph.D.

Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

K04350d

..............................................................................................................................................................................

Indications for Use

510(k) Number (if known):

Device Name: __ 6 Fr Poly Per-Q-Cath®3-Triple Lumen PICC_

Indications for Use:

The Poly Per-Q-Cath® Triple Lumen PICC is indicated for short or long term peripheral access to the centra venous system for intravenous therapy and blood sampling. For blood therapy, it is recommended that a 4 French or larger catheter be used.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Anthony V. nutz

(Division Sign-Off) Division of Anesthesiology, General Hospital, Infection Control. Dental Devices

510(k) Number __ kg 4 35 42

§ 880.5970 Percutaneous, implanted, long-term intravascular catheter.

(a)
Identification. A percutaneous, implanted, long-term intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings, such as luer hubs, and accessories that facilitate the placement of the device. The device allows for repeated access to the vascular system for long-term use of 30 days or more, and it is intended for administration of fluids, medications, and nutrients; the sampling of blood; and monitoring blood pressure and temperature. The device may be constructed of metal, rubber, plastic, composite materials, or any combination of these materials and may be of single or multiple lumen design.(b)
Classification. Class II (special controls) Guidance Document: “Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters.”