Intended Use

The IPF% (Immature Platelet Fraction) on the Sysmex® XE-2100 for in Vitro diagnostics is used to enumerate the immature platelet fraction.

Device Description

The Sysmex XE-2100 uses a gating system in the optical (fluorescence) reticulocyte/platelet channel to quantify the immature platelets. (Note: XE IPF master with XE pro is required to obtain results described.)

AI/ML Overview

The provided text describes the submission of a 510(k) premarket notification for the "IPF (Immature Platelet Fraction) parameter on the Sysmex® XE-2100, Automated Hematology Analyzer." However, it does not contain specific acceptance criteria, detailed study designs, sample sizes for training or test sets, or information about expert panel ground truth adjudication.

The document states that "Studies were performed to evaluate the equivalency of the IPF parameter to the predicate method. Results indicate equivalent performance," and "The performance data demonstrated substantial equivalence." This suggests a comparison study was conducted, but the specifics are not included in the provided text.

Therefore, many of the requested details cannot be extracted from this document. I will provide what can be gleaned from the text and explicitly state where information is missing.

1. A table of acceptance criteria and the reported device performance

  • Acceptance Criteria: Not explicitly stated in the provided text.
  • Reported Device Performance: The document only broadly states "Results indicate equivalent performance" and "The performance data demonstrated substantial equivalence" to the predicate device (FACSCalibur flow cytometric method for reticulated platelet). No specific performance metrics (e.g., accuracy, precision, correlation coefficients) are provided.

2. Sample sized used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

  • Sample Size for Test Set: Not specified in the provided text.
  • Data Provenance: Not specified in the provided text.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

  • Number of Experts: Not specified in the provided text.
  • Qualifications of Experts: Not specified in the provided text.
  • This information would typically be relevant for studies where expert review is the ground truth, which is often the case for image-based diagnostics. For a blood analyzer, the "ground truth" is typically established by a reference method or validated gold standard, not necessarily by expert consensus in the same way.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

  • Not specified in the provided text. Given the nature of a blood analyzer, an adjudication method (like for image interpretation) is unlikely to be directly applicable in the same way. Instead, the comparison would be against an established reference method.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • MRMC Study: No, an MRMC study is not mentioned or implied. This type of study is relevant for AI systems assisting human interpretation, which is not the function of this automated hematology analyzer.
  • Effect Size: Not applicable.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Yes, the device (Sysmex® XE-2100 Automated Hematology Analyzer) measures the IPF parameter directly. The study would have evaluated the performance of this automated algorithm (the device itself) in a standalone manner against a predicate method. The text states "Studies were performed to evaluate the equivalency of the IPF parameter to the predicate method."

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • The "ground truth" for the performance study was the predicate method: "reticulated platelet on the FACSCalibur flow cytometric method." The study aimed to demonstrate substantial equivalence to this existing, legally marketed method.

8. The sample size for the training set

  • Sample Size for Training Set: Not applicable. As this device is an automated hematology analyzer, it is not an AI/Machine Learning system that typically requires a large 'training set' in the conventional sense. Its functionality is based on established laboratory principles (gating system in the optical fluorescence reticulocyte/platelet channel). Thus, there's no "training set" as understood in machine learning contexts.

9. How the ground truth for the training set was established

  • Not applicable, as there is no training set in the machine learning sense for this device.

{0}------------------------------------------------

AUG 8 - 2005

510(k) SUMMARY OF SAFETY AND EFFECTIVENESS

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is: 1005 || 99

1. Submitted by:Sysmex America, Inc.One Nelson C. White ParkwayMundelein, IL 60060Phone: (847) 996-4675 FAX: (847) 996-4655Contact person: Nina GamperlingDate prepared: May 5, 2005
2. Name of Device:Trade or proprietary name: IPF (Immature Platelet Fraction)parameter on the Sysmex® XE-2100, Automated HematologyAnalyzer.Common name: IPF parameterClassification name: IPF parameterAutomated Differential Cell Counter, Sysmex® XE-2100(21 CFR 864.5220)
3. Predicate Device:The IPF parameter on the Sysmex® XE-2100 is substantiallyequivalent to the reticulated platelet on the FACSCalibur flowcytometric method.
4. Device Description:The Sysmex XE-2100 uses a gating system in the optical(fluorescence) reticulocyte/platelet channel to quantify theimmature platelets. (Note: XE IPF master with XE pro isrequired to obtain results described.)
5. Intended Use:The IPF% (Immature Platelet Fraction) on the Sysmex® XE-2100for in Vitro diagnostics is used to enumerate the immature plateletfraction.
6. Substantialequivalence-similaritiesand differencesThe following table compares the IPF parameter with predicatemethod.

{1}------------------------------------------------

510(k) SUMMARY OF SAFETY AND EFFECTIVENESS (continued)

FACSCaliburIPF%parameter onXE-Series
PredicateNew method
Intended UseUsing surrogatemarkers and gatingstrategies toidentify reticulatedplatelets.For in Vitrodiagnostics use toenumerate theimmature plateletfraction.
MethodologyEnumeration ofreticulated platelets(immatureplatelets) by flowcytometryEnumeration ofthe IPF%(immatureplatelet fraction)by the XE-2100,automatedhematologyanalyzer
Type ofAnticoagulantEDTAEDTA
Specimen TypePeripheral bloodPeripheral blood

Comparison Table to Predicate Method

7. Clinical Performance Data:Studies were performed to evaluate the equivalency of the IPF parameter to the predicate method. Results indicate equivalent performance.
8. Conclusions:The performance data demonstrated substantial equivalence.

:

:

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its wings and body. The eagle is facing to the left. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Nina M. Gamperling, MBA, MT (ASCP), RAC Manager, Regulatory Affairs Sysmex America, Inc. One Nelson C. White Parkway Mundelein, Illinois 60060

AUG 8 - 2005

K051199 Re: Trade/Device Name: Immature Platelet Fraction (IPF) on Sysmex® XE-2100, Automated Hematology Analyzer Regulation Number: 21 CFR § 864.5220 Regulation Name: Automated differential cell counter Regulatory Class: II Product Code: GKZ Dated: May 5, 2005 Received: May 10, 2005

Dear Ms. Gamperling:

We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your betermined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encreations of the enactment date of the Medical Device Amendments, or to conimer of the have been reclassified in accordance with the provisions of the Federal Food, Drug, de rices that have been that do not require approval of a premarket approval application (PMA). and coometre for , roy ( roy ) re- ) subject to the general controls provisions of the Act. The r ou may, diereleve, missas of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean r that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must corply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice orecreate as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

{3}------------------------------------------------

Page 2 -

If you desire specific information about the application of labeling requirements to your device, If you desire specific information actuation of your device, please contact the Office of In of questions on the promotion and Safety at (240) 276-0484. Also, please note the VIGO Duelood De Misbranding by reference to premarket notification" (21GFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the I ou may outliner general international and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html

Sincerely yours,

loburt Beckerh

Robert L. Becker, Jr., MD, PK.D Director Division of Immunology and Hematology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

INDICATIONS FOR USE STATEMENT

510(k) Number (if known):K051199
------------------------------------

Device Name: IPF (Immature Platelet Fraction) parameter on the Sysmex XE-2100™, Automated Hematology Analyzers ﺘﺤﺘﻬﺎ ﺗﺴﺘﺨﺪ ﺗﺴﺘﺨ

Indications For Use:

The IPF% (Immature Platelet Fraction) on the Sysmex® XE-2100 for in Vitro diagnostics is used to enumerate the immature platelet fraction.

(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CHRD, Office of Device Evaluation (ODE)

Prescription Use

OR

Over-The-Counter Use

Tobal Rscaf
Division Sign-Off

ffice of In Vitro Diagnostic Device luation and Safety

510(k) K031199

IPF on Sysmex XE-2100 Series 510(k) FDA Submission

11

§ 864.5220 Automated differential cell counter.

(a)
Identification. An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.(b)
Classification. Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”