K Number
K033758

Validate with FDA (Live)

Date Cleared
2003-12-19

(17 days)

Product Code
Regulation Number
870.1330
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

To facilitate the placement of devices used during diagnostic and interventional procedures.
To facilitate the placement of devices for diagnostic and interventional procedures.

Device Description

Nitinol or stainless steel core wire with stainless steel or platinum coil secured to the ground, flexible (distal) end of the core. Additionally, the proximal end of the coil is secured to the core either at the proximal end of the core (where the entire core is enclosed in the coil as in K779077 design) or at a design-driven location on the core (where a portion of the core is exposed as in K011084 design). A portion of the guidewire is coated with hydrophilic coating. The guidewires are bound by the following parameters: .014" - . 018" Outside Diameter: Lengths: 40cm - 150cm Tips: Straight or shaped with various tip flexibilities NOTE: None of these guidewires are for PTCA use.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly present a table of acceptance criteria or numerical performance metrics. Instead, it states that:

Acceptance Criteria (Implied)Reported Device Performance
Visual/Tactile Attributes"The results of these tests demonstrated the functionality and performance characteristics of these guidewires are comparable to the currently marketed devices."
Dimensional Attributes"The results of these tests demonstrated the functionality and performance characteristics of these guidewires are comparable to the currently marketed devices."
Mechanical Attributes"The results of these tests demonstrated the functionality and performance characteristics of these guidewires are comparable to the currently marketed devices."
Hydrophilic PropertiesMust meet established criteria for hydrophilic properties. (No specific metrics provided, but implied to have been met).
BiocompatibilityAcceptable, established by testing and market history of proven biocompatibility.
Design Control RequirementsConforms to 21 CFR 820.30, and all verification/validation demonstrated that predetermined acceptance criteria were met.
Quality SystemsFormal quality systems are in place to ensure equivalence to predicate and no adverse effects.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: The document states "Test pieces were tested and inspected according to established specific inspection criteria requirements..." but does not specify the sample size for these tests.
  • Data Provenance: The study was conducted by Lake Region Manufacturing, Inc. (LRM) to demonstrate equivalence to its own predicate devices. Therefore, the data provenance is internal to the company, and it's a prospective study as it was performed specifically for this 510(k) submission to show the new device meets requirements.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

This information is not provided in the document. The testing appears to be primarily engineering and physical performance tests, rather than a clinical study requiring expert human interpretation for ground truth.

4. Adjudication Method for the Test Set

This information is not applicable/not provided. The tests described are objective engineering and physical measurements rather than subjective assessments requiring adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance?

No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This device is a guidewire, and the submission is focused on demonstrating physical and performance equivalence, not on AI assistance for human readers.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

No, a standalone algorithm performance study was not done. This device is a physical medical device, not an algorithm or AI system.

7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

The "ground truth" for the tests described appears to be predetermined engineering specifications and established inspection criteria. For biocompatibility, it refers to testing results and market history of proven biocompatibility. There is no mention of expert consensus, pathology, or outcomes data being used for ground truth in this context.

8. The Sample Size for the Training Set

This information is not applicable/not provided. This is not a machine learning or AI-driven device, so there is no "training set."

9. How the Ground Truth for the Training Set Was Established

This information is not applicable/not provided as there is no training set for this device.

{0}------------------------------------------------

SECTION 2.0 - SUMMARY OF SAFETY AND EFFECTIVENESS

December 1, 2003

  • 2.1 General Information
    • Company Name, Address, and Telephone Number 2.1.1

Lake Region Manufacturing, Inc. (LRM) 340 Lake Hazeltine Drive Chaska, MN 55318 Telephone: (952) 448-5111 Fax: (952) 448-3441

Contact Name: Karen Mortensen Senior Regulatory Compliance Specialist

  • Device Trade Name/Proprietary Name 2.1.2
    LRM produces guidewires on an OEM basis for other manufacturers, kit assemblers, and distributors. Consequently there are a large number of trade and proprietary names not including or associated with LRM. LRM has no proprietary names of its own to be included with this submission.

  • Device Common Names/Usual Names and Classification Names 2.1.3
    These devices are commonly known as guides, guidewires, or spring guidewires. The current classification names, and product codes are Angiographic Guidewire (74HAP), Catheter Guidewire (74DQX), and Radiological Catheter Guidewire (74JAJ).

  • Establishment Registration Number: 2126666 2.1.4

  • Classification of Devices 2.1.5

The classification names listed above were originally classified as Class II devices by the Neurology (84HAD), Cardiovascular (74DQX) and Radiology (90JAJ) Review Panels, respectively.

  • Applicability of Performance Standards 2.1.6
    LRM has determined that no mandatory performance standards have been established for these devices under Section 514 of the Medical Amendments to Federal Food, Drug, and Cosmetic Act or by any subsequent regulatory action. LRM has also determined that there are no applicable voluntary standards.

{1}------------------------------------------------

  • Labels, Labeling, and Advertising 2.2
    LRM produces cardiovascular and vascular guidewires on an OEM basis for other manufacturers, kit assemblers, and distributors. There is no direct distribution by LRM. Changes to the customer controlled labels, labeling, or promotional material are at their discretion, including the resolution of any resulting regulatory obligations. A fraction of the total production bears LRM controlled labels and labeling.

  • 2.3 Statement of Availability
    This summary is being included in the Premarket Notification submission in lieu of a statement of availability.

  • 2.4 Device Description

    • Nitinol or stainless steel core wire with stainless steel or platinum coil 2.4.1 secured to the ground, flexible (distal) end of the core. Additionally, the proximal end of the coil is secured to the core either at the proximal end of the core (where the entire core is enclosed in the coil as in K779077 design) or at a design-driven location on the core (where a portion of the core is exposed as in K011084 design). A portion of the guidewire is coated with hydrophilic coating. The guidewires are bound by the following parameters:

.014" - . 018" Outside Diameter: Lengths: 40cm - 150cm Tips: Straight or shaped with various tip flexibilities

NOTE: None of these guidewires are for PTCA use.

Engincering Specifications 2.4.2

The design specifications are the same for the proposed device as they are for the LRM predicate devices {reference 510(k) K770977 and K011084]. The finished device with hydrophilic coating must meet the same basic design criteria and also meet the criteria established for the hydrophilic properties of the guidewire.

  • 2.5 Substantial Equivalence Data
    • 2.5.1 Background Information

The table below lists the differences between the predicate device and the proposed device. Testing was done to ensure the changes to the device met the predetermined acceptance criteria.

{2}------------------------------------------------

ItemProposed Device Differences from LRM Predicate cleared under510(k) K770977 and K011084
Raw MaterialsCore:No change to raw materialCoil:No change to raw materialCoating:Addition of hydrophilic coating as alternate coating(reference K991898 Microvena Corporation HyTekTMGuidewire for predicate of a legally marketed device withinthe same classification for the same intended use thatemploys the hydrophilic coating in its design.)
Assembly ProcessNo change to assembly processes
Physical CharacteristicsNo change
Labeling/IFUThe only change to the label or IFU will be wording specific to thehydrophilic properties of the guidewire.
Intended UseNo change to intended use
Anatomical SitesNo change
Target PopulationNo change
Performance TestingNo change
Safety CharacteristicsNo change
BiocompatibilityNo change
Risk AnalysisNo change

2.5.2 Non-Clinical Tests

In order to demonstrate equivalence of the proposed device, LRM performed testing to established requirements. Test pieces were tested and inspected according to established specific inspection criteria requirements for visual/tactile, dimensional and mechanical attributes. The results of these tests demonstrated the functionality and performance characteristics of these guidewires are comparable to the currently marketed devices.

Qualification and Biocompatibility Test Data 2.6

2.6.1

  • Design Control
    LRM is in conformance with the design control procedure requirements as specified in 21 CFR 820.30. Risk analysis was completed by means of a Failure Mode and Fffect Analysis (FMEA) and all verification and validation activities resulted in the ability to demonstrate that the predetermined acceptance criteria were met.

Material/Product/Process Qualification 2.6.2

LRM has formal quality systems in place to assure that each product manufactured with the hydrophilic coating remains equivalent to the predicate products, and that the changes will not have an adverse affect on the safe and effective use of the product. The quality systems include Engineering Change Order Review, Material Qualification, Product Qualification, and Process Qualification. These controls are applied to each product size/group.

{3}------------------------------------------------

  • Biocompatibility Testing 2.6.3
    Biocompatibility testing has been performed on the material components of this device. This testing, along with a market history of proven biocompatibility, establishes acceptable biocompatibility for this device.

  • 2.7 Packaging and Sterilization Information
    LRM produces guidewires on an OEM basis for other manufacturers, kit assemblers, and distributors. There is no direct distribution by LRM. A portion of the production is private label, sterile packaged to customer specifications, a fraction of that product is provided sterile to the customer.

The single packaged hydrophilic coated guidewire is placed in a dispenser and then into a Tyvek®/poly pouch. The packaged product may be packaged as five or ten pouchs in a shelf carton, which are typical packaging configurations.

There will be no changes to the sterilization process for the portion of packaged product shipped sterile to the customer. For the product that is shipped bulk, the packaging design and sterilization process parameters are the customer's responsibility. LRM will not recommend that its customers modify their packaging or sterilization procedures as a result of this submission.

  • 2.8 Intended Use Statement
    To facilitate the placement of devices used during diagnostic and interventional procedures. NOTE: The modification of this device does not alter its intended use.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an image of an eagle with its wings spread.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

DEC 1 9 2003

Lake Region Manafacturing, Inc. c/o Ms. Karen Mortensen Senior Regulatory Compliance Specialist 340 Lake Hazeltine Drive Chaska, MN 55318-1029

Re: K033758

Hydrophilic Guidewire Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter Guidewire Regulatory Class: Class II Product Code: DQX Dated: December 1, 2003 Received: December 2, 2003

Dear Ms. Mortensen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set

{5}------------------------------------------------

Page 2 – Ms. Karen Mortensen

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic form in the quality systems (QS) regular (Sections 531-542 of the Act); 21 CFR 1000-1050. product radiation control provisions (Decirons of I device as described in your Section 5 10(k)
This letter will allow you to begin marketing your device to locar device to a I his letter will anow you to begin manisang your and equivalence of your device to a legally premarket nothleadon. The PDA miding of backand is plan
marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific advice for your ac (301) 594-4648. Also, please note the regulation entitled, Colliact the Office of Compullier as (est notification" (21CFR Part 807.97). You may obtain Misoraliums by reference to premantonitiities under the Act from the Division of Small other general information on your response at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address

(901)445 657 fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Alison Hag Jr

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

510(k) Number (if known): Unknown

Device Name: Hydrophilic Guidewire

Indications for Use:

: "

To facilitate the placement of devices for diagnostic and interventional procedures.

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED

_- - '

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use X Or Over-The-Counter Use __
(PER 21 CFR 801.109)

(Division Sign-off) for BraPREMARKET NOTIFICATU
Division of Cardiovascular Devices
510(K) Number_K033758 (SM.K
9-1

§ 870.1330 Catheter guide wire.

(a)
Identification. A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.(b)
Classification. Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.