510(k) Data Aggregation
(122 days)
92677
Re: K252960
Trade/Device Name: SOLIUS PRO UVB Light Panel
Regulation Number: 21 CFR 878.4635
Ultraviolet Light Therapy Device
Classification Name: Ultraviolet lamp
Regulation Number: 21 CFR 878.4635
PREDICATE DEVICE
Name of Device: Sperti Sun Lamp Model P-164
Regulation Number: 21 CFR 878.4635
SOLIUS PRO is an over-the-counter (OTC) light panel intended to stimulate production of Vitamin D in people 22 years and older.
The SOLIUS PRO is a portable, over-the-counter (OTC) photobiomodulation light panel designed for home or facility use. The device stimulates vitamin D production through controlled exposure of a targeted ultraviolet B (UVB) spectrum ranging from 280-315 nm. Specialized UVB LEDs deliver a controlled UVB emission aligned with the action spectrum for vitamin D synthesis while incorporating multiple technological and software safety features.
The SOLIUS PRO provides a uniform light output across an area of exposed skin on the torso to optimize vitamin D production while minimizing unnecessary exposure to the face and eyes. A height-adjustable hanging system positions the panel at shoulder level, and an integrated distance sensor ensures the user maintains the optimal distance for both safety and effectiveness.
Before the first therapy session, the Personalized Dosing System uses a built-in optical sensor to quantifiably measure the Individual Typology Angle degree of the user's skin and a user response to a skin history question. The system calculates a personalized dose below 1 Minimal Erythema Dose (MED) that is delivered for a precisely controlled exposure time to stimulate vitamin D production and mitigate the risk of erythema.
Therapy sessions are initiated and monitored through a mobile application. The mobile application enforces the use of the provided protective eyewear by performing a facial scan before a session can begin. Once therapy begins, the system provides real-time visual and audio feedback from the distance sensor to assist user positioning. The mobile application records dosing information and user feedback and manages session scheduling and dose adjustments to ensure safe operation.
N/A
Ask a specific question about this device
(265 days)
Way Antioch, California 94531
Re: K221523
Trade/Device Name: Tanning Lamp Regulation Number: 21 CFR 878.4635
|
| RegulationNumber | 21 CFR 878.4635
| 21 CFR 878.4635
The subject device emits UV-B and UV-A radiation and fulfill the requirements of 21 CFR 878.4635 and
These devices are UV-B and UV-A lamps intended to provide ultraviolet radiation to tan the skin.
The Unilam Tanning lamps are high-pressure metal-halide lamps, each comprised of a quartz glass bulb filled with metal halide, mercury, and argon gas, and equipped with 2 electrodes and an electrical connector, to provide ultraviolet light in order to tan the human body.
The provided document is a 510(k) summary for a "Tanning Lamp" and focuses on demonstrating substantial equivalence to a predicate device. It does not describe a study involving algorithms, AI, or human readers. Therefore, most of the requested information regarding acceptance criteria, study design for AI models, human expert involvement, and ground truth establishment for such studies is not available in the document.
However, I can extract the information related to the device's performance criteria mentioned in the document as part of its substantial equivalence claim.
Here's the relevant information based on the document:
1. Table of acceptance criteria and the reported device performance
| Acceptance Criteria (Regulatory Standard) | Reported Device Performance |
|---|---|
| Requirements of 21 CFR 878.4635 (Sunlamp Products and Ultraviolet Lamps) | The subject device emits UV-B and UV-A radiation and fulfills the requirements of 21 CFR 878.4635. |
| Performance standard 21 CFR 1040.20 (Sunlamp Products) | The subject device fulfills the requirements of 21 CFR 1040.20. |
| Ratio of UV-C (200-260nm) to UV-B (260-320nm) < 0.003 (per 21 CFR 1040.20(C)(1)) at any distance and direction from the product or lamp | The ratio of UV-C (200-260mm) to UV-B (260-320nm) meets the requirement (<0.003) per 21 CFR 1040.20(C)(1) at any distance and direction from the product or lamp. Additionally, UVC is cut by the quartz bulb (Technical Characteristic). |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- This information is not applicable and not provided. The study performed was a non-clinical performance test of the device's UV emission characteristics against regulatory standards, not involving a test set of data in the context of AI/algorithm performance.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not applicable. This document describes performance testing of a physical lamp against regulatory standards, not the evaluation of an AI model with expert-annotated ground truth.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not applicable.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not applicable. This is not an AI/software device.
6. If a standalone (i.e. algorithm only, without human-in-the-loop performance) was done
- Not applicable. This is not an AI/software device.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- The "ground truth" in this context is the established regulatory standards and performance requirements specified in 21 CFR 878.4635 and 21 CFR 1040.20 for sunlamp products. The device's performance was measured and compared directly to these objective, quantitative standards.
8. The sample size for the training set
- Not applicable. This document does not describe the development or training of an AI model.
9. How the ground truth for the training set was established
- Not applicable.
Ask a specific question about this device
(129 days)
Planet Fitness 46/3 PLT by Ergoline, Planet Fitness 42/3 PLT by Ergoline Regulation Number: 21 CFR 878.4635
Section.: | 878.4635 |
| Classification Panel: | General & Plastic
Section.: | 878.4635 |
| Classification Panel: |
Section.: 878.4635 General & Plastic Surgery Classification Panel: Device Class: Class II
{10}------
Section.: | 878.4635 |
Device Class:
Class II
{17}---------------
This sunlamp product is intended exclusively for aesthetic tanning of the human skin, for one person at a time, at the age of 18 or above.
The primary technical components of sunlamp products are artificial sources of UV radiation as well as a mechanical structure with a defined active surface. UV lamps emit different UV-A and UV-B proportions of the UV radiation. The UV-A proportion primarily generates a superficial tan, which appears rapidly and is intensive but also fades more rapidly. The UV-B radiation is primarily responsible for more long-term tanning results. Using traditional UV light with the combination of red and blue light the user indulges in a new luxurious tanning experience. Optimal tanning results are achieved due to the synergistic combination of light spectra.
The provided text is a 510(k) Summary of Safety and Effectiveness for sunlamp products. It details comparisons between proposed devices and legally marketed predicate devices to demonstrate substantial equivalence, rather than providing an acceptance criteria table and a study proving a device meets those criteria in the context of an AI/algorithm-driven medical device.
The document does not contain information relevant to AI/algorithm performance, ground truth establishment, sample sizes for training/test sets, or expert review processes for such devices. Therefore, I cannot generate the requested table and study description based on the provided text.
The text focuses on hardware comparison and compliance with established performance standards for sunlamp products (e.g., IEC 60601-1 for electrical and mechanical safety, IEC 60601-1-2 for EMC, and U.S. performance standard 21CFR 1040.20 for irradiance ratio).
To summarize why I cannot fulfill your request:
- No AI/Algorithm Component: The submitted document describes sunlamp products, which are physical devices. There is no mention of any AI component, algorithm, or software that would require performance evaluation metrics like sensitivity, specificity, or AUC, nor would it involve concepts like "ground truth" derived from expert consensus or pathology for diagnostic purposes.
- Performance Testing Focus: The "Summary of performance testing" section clearly states that the devices were tested for "biocompatibility," "electrical and mechanical safety," and "EMC," and compliance with "U.S. performance standard 21CFR 1040.20" (which pertains to sunlamp products and their UV emissions). These are physical device performance criteria, not AI/algorithm performance.
- Comparative Approach: The entire document is structured as a comparison (substantial equivalence) to predicate devices, demonstrating that new models are similar in technology, intended use, and safety to already cleared devices, rather than a novel device proving its effectiveness against a set of performance criteria through a clinical study.
Therefore, the requested information (Acceptance Criteria Table, Sample Sizes for Test/Training, Expert Ground Truth, Adjudication, MRMC study, Standalone performance, Type of Ground Truth, Training Set Size, Training Ground Truth) is not present in the provided document because it describes a different type of medical device lacking an AI/algorithm component.
Ask a specific question about this device
(340 days)
Niguel, California 92677
Re: K210004
Trade/Device Name: i8 Tanning Booth Regulation Number: 21 CFR 878.4635
Classification Name: | Booth, Sun Tan |
| Product Code: | LEJ (21 CFR §878.4635
|
| ---------------------- | ---------------- |
|---|
LEJ (21 CFR §878.4635
The i8 tanning booth is an over-the-counter tanning device emitting ultraviolet light in the UVB and UVA region of the spectrum for the tanning of the human skin of an adult person.
Ultrasun International B.V. i8 Sun Tanning Booth is a tanning device consisting of a mechanical structure equipped with artificial light sources using ultraviolet light technology, which produces ultraviolet radiation within the UVA and UVB ranges. The UV light is intended to induce skin tanning. The user stands within the unit.
This document is a 510(k) Premarket Notification from the FDA for a tanning booth device. It outlines the device's classification, indications for use, and a comparison to a predicate device.
However, the provided text does not contain any information about acceptance criteria or a study proving the device meets those criteria in the context of typical medical device performance evaluation (e.g., diagnostic accuracy, clinical efficacy). Instead, it focuses on the substantial equivalence of an "i8 Tanning Booth" to a predicate device, primarily based on:
- Technological characteristics: Number of lamps, wattage, maximum exposure time.
- Non-clinical performance data: Electrical and mechanical safety testing (IEC 60601-1), electromagnetic compatibility testing (IEC 60601-1-2), and software development (IEC 62304).
These are engineering and safety standards, not clinical performance metrics or studies involving patient data/diagnoses that would typically have acceptance criteria related to accuracy, sensitivity, specificity, or human reader improvement.
Therefore, I cannot provide the requested information from the given text as it does not exist within the document for this specific device.
If this were a typical AI/ML medical device submission document, it would include sections detailing clinical study design, performance metrics, and results. This document is for a Class II medical device product code LEJ (Sunlamp Products And Ultraviolet Lamps Intended For Use In Sunlamp Products), which generally follows a different regulatory pathway focused on safety and functional equivalence rather than clinical diagnostic performance.
Ask a specific question about this device
(105 days)
7000 alpha hybridSun, KBL 7900 alpha hybridSun, KBL 8000 alpha hybridSun Regulation Number: 21 CFR 878.4635
|
| Product Code | LEJ (21CFR 878.4635
Classification Name | Booth, Sun Tan |
| Product Code | LEJ (21CFR 878.4635
) |
levice n Tan FR 878.4635) General & Plastic Surgery Class II - Special Controls K151962
KBL Sunlamp Hybrid Products are over-the-counter tanning devices emitting ultraviolet light in the UVA and UVB region of the spectrum intended for tanning of the human skin of an adult person.
KBL Sunlamp Hybrid Products are whole-body tanning devices basically consisting of a mechanical structure equipped with artificial light sources using the hybrid light technology. The hybrid light technology produces ultraviolet radiation and radiation in the visual range of the light spectrum. The UV light is intended for irradiation of any part of the living human body to induce skin tanning. The user is lying on a bench.
The provided document is a 510(k) Pre-market Notification for KBL GmbH sunlamp products, specifically the KBL 7000 alpha hybridSun, KBL 7900 alpha hybridSun, and KBL 8000 alpha hybridSun. This document focuses on demonstrating substantial equivalence to predicate devices rather than proving a device meets specific performance acceptance criteria for a novel AI/software medical device.
Therefore, the information requested in your prompt regarding acceptance criteria, study details, sample sizes, ground truth establishment, expert adjudication, and MRMC studies, which are typically found in the context of validating performance for diagnostic or therapeutic AI/software, is not present in this document.
This document details the following for sunlamp products:
- Device Type: Sunlamp products and ultraviolet lamps intended for use in sunlamp products.
- Regulation Number: 21 CFR 878.4635
- Regulation Name: Sunlamp products and ultraviolet lamps intended for use in sunlamp products
- Regulatory Class: Class II
- Product Code: LEJ
- Indications for Use: "KBL Sunlamp Hybrid Products are over-the-counter tanning devices emitting ultraviolet light in the UVA and UVB region of the spectrum intended for tanning of the human skin of an adult person." This is identical to the predicate device.
- Comparison to Predicate: The document largely focuses on comparing the technological characteristics (number of lamps, wattage, dimensions, etc.) of the proposed devices to predicate devices (KBL 6800 alpha and KBL 7900 alpha). The evaluation states that differences are "Equivalent" or "Identical" and do not impact functionality or safety.
- Performance Data: Non-clinical performance tests were conducted to ensure compliance with relevant standards for sunlamp products, including:
- Biocompatibility testing (DIN EN ISO 10993 series)
- Electrical and mechanical safety testing (IEC 60601-1, IEC 60335-1, UL482)
- Electromagnetic compatibility testing (IEC 60601-1-2)
- Spectral emission measurement (IEC 60335-2-27)
- Software verification and validation (FDA Guidance for Software in Medical Devices)
- Performance Standards testing (21 CFR 1040.20)
- Irradiance ratio limits (21 CFR 1040.20(c)(1))
- Maximum timer intervals and exposure schedules (FDA policy)
Key takeaway: This 510(k) submission is for physical sunlamp devices, not an AI/software medical device. The "performance data" refers to compliance with safety and electrical standards for sunlamps, not the type of diagnostic performance (e.g., sensitivity, specificity) that requires the detailed study design elements you asked for. Therefore, I cannot extract the requested information from this document.
Ask a specific question about this device
(98 days)
AC / Sound, Ergoline Planet Fitness 42/4 Hybrid Light Technology / Sound Regulation Number: 21 CFR 878.4635
Section.: | 878.4635 |
| Classification Panel: | General & Plastic
Section.: | 878.4635 |
| Classification Panel: |
This sunlamp product is intended exclusively for aesthetic tanning of the human skin, for one person at a time, at the age of 18 or above.
The primary technical components of sunlamp products are artificial sources of UV radiation as well as a mechanical structure with a defined active surface.
UV lamps emit different UV-A and UV-B proportions of the UV radiation. The UV-A proportion primarily generates a superficial tan, which appears rapidly and is intensive but also fades more rapidly. The UV-B radiation is primarily responsible for more long-term tanning results. Traditional UV light with the addition of (beauty) red light results in a natural looking tan. Optimal tanning results are achieved due to the synergistic combination of light spectra.
The provided text is a 510(k) summary for sunlamp products (Ergoline Inspiration 550 Hybrid Technology and Ergoline Planet Fitness 42/4 Hybrid Light Technology). It primarily focuses on demonstrating substantial equivalence to a predicate device, rather than proving that the device meets specific performance criteria through a clinical study or complex AI validation.
Therefore, many of the requested details, such as those related to AI algorithm performance (e.g., sample size for test/training sets, expert ground truth, MRMC study, standalone performance, validation of ground truth), are not present in this document because they are not applicable to the type of device and regulatory submission described.
The "performance testing" mentioned in the document refers to compliance with established electrical, mechanical, and safety standards for sunlamp products, not an AI-driven diagnostic or treatment device.
Here's an analysis based on the available information and an explanation of why other requested information is not present:
Device Description: Sunlamp products (Ergoline Inspiration 550 Hybrid Technology, Ergoline Planet Fitness 42/4 Hybrid Light Technology, and their variants).
Intended Use: Exclusively for aesthetic tanning of human skin, for one person at a time, at the age of 18 or above.
Regulatory Class: Class II
Predicate Device: Ergoline Inspiration 600 Dynamic Performance (K151400)
1. Table of Acceptance Criteria and Reported Device Performance:
The "acceptance criteria" for this type of device, as demonstrated in a 510(k), are primarily related to meeting established safety and performance standards for sunlamp products and demonstrating substantial equivalence to a predicate device. There aren't "performance metrics" in the typical sense of accuracy, sensitivity, or specificity as one would see for a diagnostic AI.
| Acceptance Criteria (as implied by 510(k) process for this device type) | Reported Device Performance (from "Summary of performance testing" and comparison table) |
|---|---|
| Compliance with U.S. Performance Standard 21 CFR 1040.20 | The proposed devices are in compliance with U.S. performance standard 21CFR 1040.20. |
| Biocompatibility | Tested in accordance with ISO 10993-series. (Implied compliance, details not given) |
| Electrical and Mechanical Safety | Tested in accordance with IEC 60601-1. (Implied compliance, details not given) |
| Electromagnetic Compatibility (EMC) | Tested in accordance with IEC 60601-1-2. (Implied compliance, details not given) |
| Maintenance of Irradiance Ratio per 1040.20 | Fulfilled (for both proposed and predicate devices). |
| Max Exposure Time | 12 minutes (Identical to predicate). |
| Identical Intended Use/Indications for Use as Predicate | Identical. |
| No New Potential Hazards or Safety Risks Introduced | Device comparison and testing imply no new hazards. |
| All Special Controls Fulfilled | All Special Controls are fulfilled. |
2. Sample size used for the test set and the data provenance:
- Not applicable / Not provided. This document describes a submission for a sunlamp and does not involve AI or a complex test set of patient data. The "testing" refers to lab-based compliance testing of the physical device.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable / Not provided. No patient data test set or ground truth establishment by experts for performance evaluation (e.g., diagnostic accuracy) is described.
4. Adjudication method for the test set:
- Not applicable / Not provided. No test set requiring expert adjudication for AI performance.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No. This is a sunlamp product, not an AI-assisted diagnostic device. MRMC studies are not relevant here.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- Not applicable. This relates to an AI algorithm, which is not part of this device or submission type.
7. The type of ground truth used:
- Not applicable. No ground truth in the context of diagnostic accuracy is relevant. The "ground truth" for this device would be its adherence to specified physical and electrical parameters and safety standards, as verified through laboratory testing against those standards.
8. The sample size for the training set:
- Not applicable / Not provided. No AI training set is mentioned as part of this device submission.
9. How the ground truth for the training set was established:
- Not applicable / Not provided. No AI training set.
Summary of the Study (as described in the document):
The "study" presented in this 510(k) summary is not a clinical study in the traditional sense, nor is it an AI performance validation study. Instead, it is a design verification and validation process to demonstrate that the new sunlamp devices are substantially equivalent to a previously cleared predicate device and that they comply with applicable U.S. and international safety and performance standards.
The core of the "proof" the device meets acceptance criteria lies in:
- Comparison to Predicate Device: Showing that the proposed devices have identical intended use and similar technological characteristics, with any differences not raising new questions of safety or effectiveness. Key differences (e.g., number of lamps, wattage) are addressed by stating that irradiance characteristics are comparable and safety has been confirmed by corresponding tests.
- Compliance with Recognized Standards: Stating that the devices have been tested and comply with:
- U.S. Performance Standard 21 CFR 1040.20 (Sunlamp Products)
- ISO 10993-series (Biocompatibility)
- IEC 60601-1 (Electrical and Mechanical Safety)
- IEC 60601-1-2 (Electromagnetic Compatibility)
The document concludes that the proposed devices are substantially equivalent to the predicate, do not introduce new indications or hazards, and fulfill all Special Controls.
Ask a specific question about this device
(109 days)
Re: K183539
Trade/Device Name: Ergoline Sunrise 7200 Hybrid Technology Regulation Number: 21 CFR 878.4635
|
| Section: | 878.4635 |
| Classification Panel:
Section.: | 878.4635 |
| Classification Panel: | General &
This sunlamp product is intended exclusively for aestetic tanning of the human skin, for one person at a time, at the age of 18 or above.
The primary technical components of sunlamp products are artificial sources of UV radiation as well as a mechanical structure with a defined active surface. UV lamps emit different UV-A and UV-B proportions of the UV radiation. The UV-A proportion primarily generates a superficial tan, which appears rapidly and is intensive but also fades more rapidly. UV-B radiation is primarily responsible for more long-term tanning results. The Ergoline Sunrise 7200 Hybrid Light Technology / Hybrid Technology provides tanning results by combining traditional UV light with red light in a synergistic combination of light spectra.
The provided text is a 510(k) summary for a sunlamp product. This type of document is for demonstrating substantial equivalence to a predicate device, not for proving a device meets specific acceptance criteria through a clinical study in the way an AI diagnostic device would.
Therefore, many of the requested sections about clinical study details (such as sample size for test sets, data provenance, number of experts for ground truth, adjudication methods, MRMC studies, standalone performance, and training set details) are not applicable to this document as it describes a Class II sunlamp product, not an AI/ML-driven medical device.
The document primarily focuses on demonstrating that the new device, Ergoline Sunrise 7200 Hybrid Technology, is substantially equivalent to a legally marketed predicate device (Sunrise 6200 and Sunrise 7200, K173565) by comparing their technological characteristics, intended use, and compliance with relevant safety standards.
Here's an attempt to answer the questions based on the available information, noting the inapplicability of some points:
1. Table of Acceptance Criteria and Reported Device Performance
For this type of device, "acceptance criteria" and "reported device performance" are primarily demonstrated through compliance with established safety and performance standards rather than specific diagnostic metrics. The document asserts compliance with these standards.
| Acceptance Criteria/Standard | Reported Device Performance/Compliance |
|---|---|
| Intended Use/Indications for Use (Substantial Equivalence) | Identical: "This sunlamp product is intended exclusively for aesthetic tanning of the human skin, for one person at a time, at the age of 18 or above." |
| Number of body lamps (Comparison to Predicate) | Similar (Proposed: 34 UV lamps and 14 UV+Red Light lamps; Predicate: 48 UV lamps). Safety confirmed by IEC 62471 tests. Functionality and safety not impacted. |
| Watts (Lamps) | Identical (200W for both proposed and predicate UV lamps). |
| Lamp item description (Comparison to Predicate) | Similar (Proposed: GENESIS Type U Hybrid Performance - 200 W (UV light) and GENESIS Type R Hybrid Performance - 200 W (Mix of UV+Red Light); Predicate: GENESIS VHP9 Smart Performance 200 W). Safety confirmed by IEC 62471 tests. Functionality and safety not impacted. |
| Number of Beauty Light LEDs | Different (Proposed: 60 (3x20); Predicate: none). Safety confirmed by IEC 62471 tests. Functionality and safety not impacted. |
| Watts (Beauty Light LEDs) | Different (Proposed: 45 (3x15); Predicate: none). No impact on safety. |
| Max exposure time [min] | Identical (9 minutes). |
| Electrical requirements | Identical (230V 3Ø or 230V 2Ø). |
| Total power consumption [Watts] | Similar (Proposed: 10600W; Predicate: 9500W). Higher due to different electronic lamp ballasts, performance not impaired. |
| Rated overcurrent protection device (circuit breaker) | Identical (40A / 3-pole 3Ø or 70A / 2-pole 2Ø). |
| Number of wires | Identical (4 3Ø or 3 2Ø). |
| Irradiance ratio in accordance with 1040.20 | Fulfilled (for both proposed and predicate). |
| Electrical safety (IEC 60601-1) | Identical (Compliant for both proposed and predicate). |
| Electromagnetic compatibility (IEC 60601-1-2) | Identical (Compliant for both proposed and predicate). |
| Biocompatibility (ISO 10993-series) | Proposed devices tested and in compliance. |
| U.S. performance standard 21CFR 1040.20 | Proposed devices tested and in compliance. |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Not Applicable: This document describes a sunlamp product, not an AI/ML diagnostic or predictive device that typically uses "test sets" of data. The "testing" referred to is for compliance with electrical, mechanical, and safety standards like IEC 62471, IEC 60601-1, IEC 60601-1-2, ISO 10993, and 21 CFR 1040.20. These are engineering/device performance tests, not clinical studies with patient data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not Applicable: As above, this is not an AI/ML diagnostic device requiring expert-established ground truth from clinical data. Compliance is assessed against technical standards.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not Applicable: No clinical "test set" and thus no adjudication method for ground truth determination.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not Applicable: This is not an AI-assisted diagnostic device, so no MRMC study would be relevant.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not Applicable: This is a sunlamp product, not an algorithm, so "standalone performance" in this context is irrelevant.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- Not Applicable: The "ground truth" for this device's compliance is defined by the technical specifications and safety standards (e.g., measured irradiance levels, electrical safety parameters, biocompatibility test results). It is not based on clinical outcomes, pathology, or expert consensus on a diagnosis.
8. The sample size for the training set
- Not Applicable: This is a physical sunlamp product and does not involve AI/ML models that require a training set.
9. How the ground truth for the training set was established
- Not Applicable: As there is no training set, this question is not applicable.
Ask a specific question about this device
(224 days)
Re: K180555
Trade/Device Name: Ergoline Prestige 1600 Hybrid Performance Regulation Number: 21 CFR 878.4635
Section.: | 878.4635 |
| Classification Panel: | General & Plastic
Section.: | 878.4635 |
| Classification Panel:
This sunlamp product is intended exclusively for aesthetic tanning of the human skin, for one person at a time, at the age of 18 or above.
The primary technical components of sunlamp products are artificial sources of UV radiation and a mechanical structure. Different UV sources intensities with characteristic UV-A and UV-B proportions result in cosmetic tanning of the human skin. The UV-A proportion primarily generates a superficial tan, which appears rapidly and is intensive but also fades more rapidly. The UV-B radiation is primarily responsible for more longterm tanning results.
This document is a 510(k) summary for the Ergoline Prestige 1600 Hybrid Performance sunlamp product. It compares the proposed device to a predicate device, the Ergoline/Prestige 1400 Intelligent Performance, to demonstrate substantial equivalence.
Here's an analysis based on your request:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria for this device are primarily based on demonstrating substantial equivalence to a legally marketed predicate device (Ergoline/Prestige 1400 Intelligent Performance) and compliance with relevant safety and performance standards. The "Evaluation" column in the comparison table serves as the reported device performance against these criteria.
| Acceptance Criteria / Parameter | Reported Device Performance (Ergoline Prestige 1600 Hybrid Performance) |
|---|---|
| Intended Use | "This sunlamp product is intended exclusively for cosmetic tanning of the human skin, for one person at a time, at the age of 18 or above." (Identical to Predicate) |
| Number of body lamps | 52 (38 UV lamps, 14 UV+red light lamps) - Similar: The irradiance characteristics are still comparable as provided under section 18. The safety has been confirmed additionally by corresponding tests in accordance with IEC 62471. In terms of functionality and safety there is no impact due to this difference. |
| Watts (Body lamps) | 200 (Identical to Predicate) |
| Lamp item description (Body lamps) | GENESIS Type U Hybrid Performance - 200 W (UV light), GENESIS Type R Hybrid Performance - 200 W (red light) - Similar: The irradiance characteristics are still comparable as provided under section 18. The safety has been confirmed additionally by corresponding tests in accordance with IEC 62471. In terms of functionality and safety there is no impact due to this difference. |
| Number of High Pressure Facial lamps | 4 (Identical to Predicate) |
| Watts (High Pressure Facial lamps) | 520 (Identical to Predicate) |
| Lamp item description (High Pressure Facial lamps) | Ergoline Ultra 520W (Identical to Predicate) |
| Filter (High Pressure Facial lamps) | Ultra Performance 412 (Identical to Predicate) |
| Number of UV-B Facial lamps | 3 (Identical to Predicate) |
| Watts (UV-B Facial lamps) | 8 (Identical to Predicate) |
| Lamp item description (UV-B Facial lamps) | Genesis Type R Hybrid 8W (Similar to Predicate) |
| Number of Shoulder lamps | 2 (Identical to Predicate) |
| Watts (Shoulder lamps) | 240 (Identical to Predicate) |
| Lamp item description (Shoulder lamps) | Ergoline Ultra 250W (Identical to Predicate) |
| Filter (Shoulder lamps) | Ultra Performance 912 (Identical to Predicate) |
| Max exposure time [min] | 10 (Identical to Predicate) |
| Electrical requirements | 230V 3Ø or 230V 2Ø (Identical to Predicate) |
| Total power consumption [watts] | 17,600 W - Similar: less power consumption due to the use of electronic ballasts instead of magnetic ballasts. Performance is not impaired. |
| Rated overcurrent protection device (circuit breaker) | 70A / 3-pole 3Ø or 100A / 2-pole 2Ø (Identical to Predicate) |
| Number of wires | 4 3Ø or 3 2Ø (Identical to Predicate) |
| Irradiance ratio in accordance with 1040.20 (max. 0.003) | Fulfilled - Identical (refer to section 18 for bench test reports) |
| Electrical safety | IEC 60601-1 (Identical to Predicate) |
| Electromagnetic compatibility | IEC 60601-1-2 (Identical to Predicate) |
| Biocompatibility | Tested in accordance with ISO 10993-series |
| Compliance with U.S. performance standard 21CFR 1040.20 | In compliance |
| Does not introduce new indications for use | True (Identical to Predicate) |
| Has the same technological characteristics | Primarily true, with justified minor differences that do not impact safety or effectiveness. |
| Does not introduce new potential hazards or safety risks | True (Based on testing and comparison to Predicate) |
2. Sample size used for the test set and the data provenance
The document does not specify a "test set" in the context of clinical data for algorithmic performance. This device is a physical sunlamp, and the comparison is primarily based on technical specifications and compliance with established standards rather than diagnostic performance using a data set.
The testing mentioned (e.g., biocompatibility) refers to laboratory and bench testing of the components and the device itself, not a clinical study on human subjects with a test set of cases.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. This is not a device that generates diagnostic outputs requiring expert ground truth establishment.
4. Adjudication method for the test set
Not applicable.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is not an AI/CADe/CADx device where human reader performance with or without AI assistance would be relevant.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
Not applicable. This is not an algorithm-based device.
7. The type of ground truth used
For the safety and performance evaluations, the "ground truth" is adherence to established international and national standards (e.g., IEC 60601-1, IEC 60601-1-2, ISO 10993-series, 21 CFR 1040.20) and the technical specifications of a predicate device. This is a form of benchmarking against established standards and predicate device characteristics.
8. The sample size for the training set
Not applicable. This device does not involve machine learning or a training set.
9. How the ground truth for the training set was established
Not applicable.
Ask a specific question about this device
(98 days)
Trade/Device Name: ProSun sunlamp products, Luxura sunlamp products Regulation Number: 21 CFR 878.4635
ProSun sunlamp products Common Name: Prosun sunlamp products Classification Name: Booth, Sun Tan (21 CFR 878.4635
Luxura sunlamp products |
| Classification Name: | Booth, Sun Tan (21 CFR 878.4635
1.5.2 - DEVICE CLASSIFICATION
Classification Name: Booth, Sun Tan (21 CFR 878.4635
Tanning Bed Systems are intended to provide tanning of the human skin.
The primary technical components of sunlamp products are whole body tanning devices consisting of a mechanical frame with Ultraviolet (UV) lamps providing an artificial source of UV radiation. The UV light is intended for irradiation of the human body to induce skin tanning. The user of the proposed devices device either lies in the horizontal tanning bed or stands in the vertical tanning booth.
Skin tanning varies with skin type and as the UV sources intensities with characteristic UV-A and UV-B proportions result in cosmetic tanning of the human skin.
The UV-A proportion primarily generates a superficial tan, which appears rapidly and is intensive but fades more rapidly. The UV-B radiation is primarily responsible for more longterm tanning results.
This document describes ProSun International LLC's submission for their sunlamp products, establishing substantial equivalence to previously marketed predicate devices. This is a 510(k) submission for tanning bed systems, which are classified as Class II medical devices. The document does not describe an AI/ML powered device, therefore the typical acceptance criteria and study data for AI/ML devices involving human readers, test sets, ground truth, and training sets are not applicable.
Instead, the acceptance criteria for these devices revolve around compliance with established performance standards and safety regulations for sunlamp products.
Here's a breakdown of the requested information, adapted to the context of this device:
1. A table of acceptance criteria and the reported device performance
| Acceptance Criteria (FDA Performance Standards & Guidance) | Reported Device Performance (Summary of Performed Tests) |
|---|---|
| Biocompatibility: In accordance with DIN EN ISO 10993 | Non-clinically tested in accordance with DIN EN ISO 10993. |
| Electrical and Mechanical Safety: In accordance with IEC 60601-1 (IEC 60335-2-27 and UL 482) | Non-clinically tested in accordance with IEC 60601-1 (IEC 60335-2-27 and UL 482). |
| EMC (Electromagnetic Compatibility): In accordance with IEC 60601-1-2 (CISPR 14-1 and 14-2) | Non-clinically tested in accordance with IEC 60601-1-2 (CISPR 14-1 and 14-2). |
| Spectral Emissions: Measured in accordance with IEC 6033-2-27 and 21 CFR 1002, 1010-1050 | Spectral emissions of all sunlamp products were measured (test procedure in accordance with IEC 6033-2-27 and 21 CFR 1002, 1010-1050). |
| Irradiance Ratio Limits: In accordance with 21 CFR 1040.2(c)(1) | Irradiance ratio limits of all proposed sunlamp products are in accordance with 21 CFR 1040.2(c)(1). |
| Software Verification & Validation: According to FDA's "Guidance for the Content of Premarket Submission for Software Contained in Medical Devices" (if applicable) | Non-clinically tested for software verification & validation testing according to FDA's "Guidance for the Content of Premarket Submission for Software Contained in Medical Devices" (implying compliance, as the devices are presented as being substantially equivalent). |
| Maximum Timer Intervals and Exposure Schedules: Determined according to FDA's "Policy on maximum timer interval and exposure schedule for sunlamp products." | The maximum timer intervals and exposure schedules have been determined according to FDA's "Policy on maximum timer interval and exposure schedule for sunlamp products." (The table in section 1.5.6 on page 6 details the "Max Exposure Min." for each model, ranging from 9 to 20 minutes, which represents the device's adherence to these schedules). |
| Performance Standards Testing: In accordance with 21 CFR 1040.20 | Performance Standards testing in accordance with 21 CFR 1040.20. |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
This information is not applicable as this is not an AI/ML study involving a test set of data. The devices themselves are the "test set" and their performance is evaluated against regulatory standards. The testing described is primarily non-clinical, evaluating the physical device characteristics and compliance with electrical, mechanical, and radiation emission standards.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This information is not applicable. The "ground truth" for sunlamp products relies on established engineering and medical device safety standards and regulations (e.g., IEC, UL, 21 CFR). Expert review would be conducted by regulatory bodies and internal engineering teams to ensure compliance, rather than clinical experts establishing a ground truth for a diagnostic AI.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This information is not applicable. This is not a study requiring adjudication of interpretations from multiple readers.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This information is not applicable. This submission is for sunlamp products, not a device with AI assistance for human readers.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This information is not applicable. These are physical sunlamp devices, not software algorithms.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The "ground truth" in this context is the set of established safety and performance standards, regulations, and internationally recognized testing procedures for sunlamp products. This includes:
- 21 CFR 878.4635 (Regulation for Ultraviolet Lamp for Tanning)
- 21 CFR 1040.20 (Performance standards for sunlamp products)
- 21 CFR 1002, 1010-1050 (Electronic product radiation control)
- IEC 60601-1 (Medical electrical equipment - General requirements for basic safety and essential performance)
- IEC 60335-2-27 (Household and similar electrical appliances - Safety - Part 2-27: Particular requirements for appliances for skin exposure to ultraviolet and infrared radiation)
- UL 482 (Standard for Portable Sun/Health Lamps)
- DIN EN ISO 10993 (Biological evaluation of medical devices)
- CISPR 14-1 and 14-2 (Electromagnetic compatibility for household appliances, electric tools and similar apparatus)
- FDA Guidance documents (e.g., "Guidance for the Content of Premarket Submission for Software Contained in Medical Devices" and "Policy on maximum timer interval and exposure schedule for sunlamp products")
8. The sample size for the training set
This information is not applicable. This is not an AI/ML device that requires a training set.
9. How the ground truth for the training set was established
This information is not applicable. There is no training set for this device.
Ask a specific question about this device
(1093 days)
52241
Re: K151721
Trade/Device Name: Fiji Sun Fluorescent Tanning Unit Regulation Number: 21 CFR 878.4635
FijiSun Fluorescent Tanning Unit (FIJI Sun)
- 2.4 Classification: Class II
- 2.5 Classification Number: 878.4635
The Fiji Sun Fluorescent Tanning Unit device (FIJI Sun) is intended to provide ultraviolet light for the purpose of stimulating a tanning response in the skin.
The FIJI Sun Fluorescent Tanning Unit (FIJI SUN) is a small tabletop sunlamp product which uses four Sperti brand ultraviolet lamps (model number: FRT20TI21BUHO) manufactured for KBD. The FIJI Sun Fluorescent Tanning Unit uses a mechanical timer with a maximum timer interval of 15 minutes and a minimum timer interval of one minute. The maximum timer interval error as a percent of that interval is +/- 6%. Output performance testing and FDA performance standards were used to determine the maximum timer interval and the recommended exposure schedule which is on the device labeling and in the user instructions. Two pair of protective eye wear are supplied with the device. The spectral transmittance to the eye of the protective eyewear is 0.0008 in the wavelength range from 200 nm to 320 nm and 0.0004 in the wavelength range from 320nm to 400 nm and is sufficient over the wavelengths greater that 400nm to enable the user to see clearly enough to reset the timer. In order for the product to fulfill its function it emits UV light In the wavelength range from 290 nm to 400 nm and the emission level is one SED (100 J/m²). The minimum use distance specified in the product labeling and the user instruction is 18 inches. Adequate directions for use, warnings and contraindications are included in user instructions provided with the device and in the device labeling. There is no software contained in the device. The device is sold as Over the Counter (OTC).
This document describes a 510(k) premarket notification for the "Fiji Sun Fluorescent Tanning Unit (FIJI Sun)." This device is a tanning unit, not an AI/ML powered medical device, therefore the typical acceptance criteria and study information related to AI/ML software (like sensitivity, specificity, AUC, sample sizes for test/training sets, expert ground truth, adjudication methods, MRMC studies, standalone performance, etc.) are not applicable to this submission.
The "acceptance criteria" and "device performance" in this context refer to compliance with established safety and performance standards for sunlamp products.
Here's a summary of the relevant information provided:
Acceptance Criteria and Reported Device Performance
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Timer Interval | Maximum timer interval of 15 minutes, minimum of 1 minute. |
| Timer Interval Error | +/- 6% of the maximum timer interval. |
| UV Wavelength Range | Emits UV light in the wavelength range from 290 nm to 400 nm. |
| Emission Level | Emission level is one SED (100 J/m²). |
| Minimum Use Distance | 18 inches specified in product labeling and user instructions. |
| Protective Eyewear Transmittance | Spectral transmittance to the eye: 0.0008 in the wavelength range from 200 nm to 320 nm. |
| Spectral transmittance to the eye: 0.0004 in the wavelength range from 320 nm to 400 nm. | |
| Protective Eyewear Visibility | Sufficient over wavelengths greater than 400 nm to enable the user to see clearly enough to reset the timer. |
| Electrical Safety Standard | Conformance to IEC/EN 60601-1:2005 Edition 3/(R)2012 And A1:2012 (Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance). |
| Radiation Control Standard | Compliance with performance standards under Section 1040.20 of the Federal Food, Drug, and Cosmetics Act (specifically pertaining to radiation limits). |
Study Information (Applicable to Non-AI/ML Device)
- Sample size used for the test set and the data provenance: No "test set" in the context of an AI/ML study. The device itself (the Fiji Sun Fluorescent Tanning Unit) underwent performance testing to ensure compliance with relevant standards. The data provenance is internal testing performed by the manufacturer KBD, Inc.
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for a tanning unit involves physical measurements and compliance with regulatory limits, not expert interpretations of medical images or data.
- Adjudication method for the test set: Not applicable.
- If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI/ML device.
- If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is not an AI/ML device.
- The type of ground truth used (expert consensus, pathology, outcomes data, etc): The "ground truth" here is objective physical measurements and established regulatory limits for UV emission, timer accuracy, and protective eyewear transmittance as defined by the Federal Food, Drug, and Cosmetics Act (specifically 21 CFR 1040.20) and international electrical safety standards (IEC/EN 60601-1).
- The sample size for the training set: Not applicable. This is not an AI/ML device.
- How the ground truth for the training set was established: Not applicable. This is not an AI/ML device.
In summary, the "study" demonstrating the device meets "acceptance criteria" involved specific performance testing of the physical device to ensure it conforms to established regulatory standards for sunlamp products, particularly regarding UV radiation emission, timer accuracy, and protective eyewear safety. The submission claims "Specific performance testing (spectral analysis) was done on these devices to measure irradiance" and "Specific performance testing was done on the included protective eyewear (goggles)."