K Number
K151400

Validate with FDA (Live)

Manufacturer
Date Cleared
2015-09-16

(113 days)

Product Code
Regulation Number
878.4635
Age Range
All
Reference & Predicate Devices
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

This sunlamp product is intended exclusively for cosmetic tanning of the human skin, for one person at a time, at the age of 18 or above.

Device Description

The primary technical components of sunlamp products are artificial sources of UV radiation and a mechanical structure. Different UV sources intensities with characteristic UV-A and UV-B proportions result in cosmetic tanning of the human skin. The UV-A proportion primarily generates a superficial tan, which appears rapidly and is intensive but also fades more rapidly. The UV-B radiation is primarily responsible for more longterm tanning results.

AI/ML Overview

I am sorry, but the provided text does not contain the information required to answer your request. The document is a 510(k) premarket notification for sunlamp products, which primarily focuses on substantial equivalence to predicate devices and regulatory compliance rather than outlining detailed acceptance criteria, specific study designs, or performance metrics typically associated with medical device efficacy studies.

Specifically, the document does not include:

  1. A table of acceptance criteria and reported device performance for an AI/algorithm.
  2. Sample sizes for test sets, data provenance, or details on training sets.
  3. Information about experts establishing ground truth or adjudication methods.
  4. Details of multi-reader multi-case comparative effectiveness studies or standalone algorithm performance.
  5. The type of ground truth used (e.g., pathology, outcomes data).

The document is about sunlamp products and their regulatory clearance, not about an AI or algorithmic device.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized design of three human profiles facing right, layered on top of each other. The profiles are connected by a curved line that forms the shape of a bird. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the logo.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 16, 2015

JK HOLDING GmbH Juergen Gerstenmeier Official Correspondent Koehlershohner Strasse 60 53578 Windhagen Germany

Re: K151400

Trade/Device Name: Ergoline, Sun Angel, Soltron, Sundash and Palm Beach Tan Sunlamp Products Regulation Number: 21 CFR 878.4635 Regulation Name: Ultraviolet lamp for tanning Regulatory Class: Class II Product Code: LEJ Dated: August 14, 2015 Received: August 18, 2015

Dear Gerstenmeier:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{1}------------------------------------------------

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Joshua C. Nipper -S

Binita S. Ashar, M.D., M.B.A., F.A.C.S.

For Director

Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K151400

Device Name

Ergoline, Sun Angel, Soltron, Sundash and Palm Beach Tan Sunlamp Products

Indications for Use (Describe)

This sunlamp product is intended exclusively for cosmetic tanning of the human skin, for one person at a time, at the age of 18 or above.

Type of Use (Select one or both, as applicable)

| | Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary of Safety and Effectiveness

In accordance with the requirements of the Safe Medical Device Act, JK Holding GmbH herewith submits a Summary of Safety and Effectiveness. This 510(k) summary for the Ergoline, Soltron, Sun Angel, Sundash and Palm Beach Tan private label sunlamp products meets the requirements of 21 CFR 807.92

Submitter information:JK HOLDING GmbHKoehlershohner Strasse 6053578 WindhagenPhone: +49 (0) 2224 / 818-0Fax: +49 (0) 2224 / 818-116
Establishment Registration Number: 9615023Owner/Operator Number: 9034303
Official Correspondent (Contact):Jürgen GerstenmeierKoehlershohner Strasse 6053578 WindhagenGermanyPhone: +49 (0) 2224 / 818-330Email: juergen.gerstenmeier@de.jk-group.net
US Agent:Lateef KhanJK PRODUCTS & SERVICES, Inc.#1 Walter Kratz DriveJonesboro, AR 72401Phone: 870-2682700Fax: 870-9359046E-Mail: Ikhan@jkamerica.com

Date Prepared:

August 12th, 2015

{4}------------------------------------------------

Device(s) Identification:

Device trade name:Ergoline sunlamp products
Common name:Ergoline sunlamp products

The following table presents the Ergoline sunlamp products in comparison to Soltron private label devices.

The basic products can be upgraded with the conversion kit to Soltron private label sunlamp products.

The Ergoline, Soltron, Sun Angel, Sundash or Palm Beach Tan private label sunlamp products contain the same basic products, which are the technical fundament for different product lines. The conversion kit only serves to label the basic product as a Soltron or Palm Beach Tan private label sunlamp product. It does not alter any technical specification or aspect of the devices.

Model-No.:Ergoline sunlamp products:Conversion Kit /Private Label:
JK 103 / 34-3Ergoline/PASSION 34/3 Super PowerXS-40 KISS
JK 103 / 40-3Ergoline/PASSION 40/3 Turbo Power(WHITE)S-45 FLIRT
JK 103 / 40-3Ergoline/PASSION 40/3 Turbo Power(BLACK)S-45 FLIRT
JK 103 / 40-3 ACErgoline/PASSION PLUS 40/3 Turbo Power(WHITE)S-45 FLIRT+AC
JK 103 / 40-3 ACErgoline/PASSION PLUS 40/3 Turbo Power(BLACK)S-45 FLIRT+AC
JK 98 / 48-0Ergoline/SUNRISE 480 Turbo PowerV-55 UNIVERSE
JK 98 / 48-0Ergoline/SUNRISE 480 Turbo Power(Sound & Ambient Light)V-55 UNIVERSE
JK 70 / 0-30Ergoline/ Open Sun 1050XXL-95 SHARK
JK 155 / 46-3 ACErgoline/AFFINITY 700 Advanced PerformanceL-70 BUTTERFLY
JK 155 / 46-3Ergoline/AFFINITY 700 Advanced PerformanceL-70 BUTTERFLY
JK 155 / 50-4 TTErgoline/AFFINITY 900 Advanced PerformanceXL-85 WILDCAT
JK 116 / 48-0Ergoline/ ESSENCE 48 SMART POWER(Grey)V-50 SPACE CAB
JK 127 / 48-4 TTErgoline/Prestige 1000 Intelligent PerformanceXXL-115 STARLIGHT
JK 127 / 52-4 TTErgoline/Prestige 1400 Intelligent PerformanceXXL-125 EAGLE
JK 126 / 38-3 ACErgoline/Inspiration 500 Smart PerformanceM-55 RACER
JK 126 / 42-4 ACErgoline/Inspiration 600 Dynamic PerformanceM-60 JUMP
JK 92 / 43-3Ergoline/AFFINITY 500 Turbo Power-
JK 92 / 46-3-ACErgoline/AFFINITY 600 Turbo PowerAir Condition-
JK 92 / 51-4 TTErgoline/AFFINITY 800 Turbo PowerTemptronic-
JK 116 / 48-0Ergoline/ ESSENCE 48 SMART POWER(Red)-
JK 116 / 48-0Ergoline/ ESSENCE 48 SMART POWER(PBT)-
JK 127 / SS 48-4-TTErgoline/Sun Select 1000 Intelligent Performance-

{5}------------------------------------------------

Model-No.:Ergoline sunlamp products:Conversion Kit /Private Label:
JK 127 / SS 52-4 TTErgoline/Sun Select 1400 Intelligent Performance-
JK 103 / 34-0Ergoline / Passion 34 Super Power(20 min)-
JK 103 / 34-0Ergoline / Passion 34 Super Power(15 min)-
JK 103 / 32-0Ergoline/ Flair 32 Super Power(20 min)-
JK 103 / 32-0Ergoline/ Flair 32 Super Power(15 min)-
JK 103 / 32-1Ergoline/ Flair 32/1 Super Power(20 min)-
JK 103 / 32-1Ergoline/ Flair 32/1 Super Power(15 min)-
JK 103 / 32-3Ergoline/ Flair 32/3 Super Power(15 min)-
JK 103 / 32-3Ergoline/ Flair 32/3 Super Power(20 min)-

Device trade name: Sun Angel sunlamp products Common name: Sun Angel sunlamp products

The following table presents the Sun Angel sunlamp products (no conversion kits).

Model-No.:Sun Angel sunlamp products:
JK 103 / SA 40-3 ACSUNANGEL 400
JK 127 / SA 52-4 TTSUN ANGEL DUO 1400 (12 min)
JK 127 / SA 52-4 TTSUN ANGEL DUO 1400 (15 min)
JK 126 / SA 42-4 ACSUN ANGEL DUO 600 (20 min)
JK 126 / SA 42-4 ACSUN ANGEL DUO 600 (15 min)
JK 100 / SA 52-4 TTSun Angel S 52 (20 min)
JK 100 / SA 52-4 TTSun Angel S 52 (15 min)
JK 104 / SA 46-3 TTSun Angel S 46
JK 132 / SA 52-4 TTSun Angel Duo S 52 (20 min)
JK 132 / SA 52-4 TTSun Angel Duo S 52 (15 min)

{6}------------------------------------------------

Device trade name: Soltron sunlamp products Common name: Soltron sunlamp products

The following table presents the Soltron sunlamp products (no conversion kits).

Model-No.:Soltron sunlamp products:
JK 103 / 32-0Soltron/XXS-30 Super Power (Mango Tango)
JK 79 / 52-4 TTSoltron/XXL-110 DYNAMIC POWER (Blueberry Blue)
JK 79 / 48-4 TTSoltron/XL-80 DYNAMIC POWER (Magic Melon)
JK 103 / 34-3Soltron/Coco Cabana XXS-35

Device trade name: Sundash sunlamp products Common name: Sundash sunlamp products

The following table presents the Sundash sunlamp products in comparison to Soltron private label devices.

The basic products can be upgraded with the conversion kit to Soltron private label sunlamp products.

The Ergoline, Soltron, Sun Angel, Sundash or Palm Beach Tan private label sunlamp products contain the same basic products, which are the technical fundament for different product lines. The conversion kit only serves to label the basic product as a Soltron or Palm Beach Tan private label sunlamp product. It does not alter any technical specification or aspect of the devices.

Model-No.:Ergoline sunlamp products:Conversion Kit /Private Label:
JK 103 / 32-0SUNDASH 32 Super PowerXXS-35 SURF
JK 103 / 32-1SUNDASH 32/1 Super Power(15 min)XXS-35 SURF
JK 103 / 32-1SUNDASH 32/1 Super Power(20 min)XXS-35 SURF

Device trade name: Palm Beach Tan sunlamp products Common name: Palm Beach Tan sunlamp products

The following table presents the Palm Beach Tan sunlamp products (no conversion kits).

Model-No.:Ergoline sunlamp products:
JK 103 / 32-0Palm Beach Tan / SIGNATURE

{7}------------------------------------------------

Classification of the device:

Device Classification Name:Booth, Sun Tan
Product Code:LEJ
C.F.R. Section.:878.4635
Classification Panel:General & Plastic Surgery
Device Class:Class II

Device Description:

The primary technical components of sunlamp products are artificial sources of UV radiation and a mechanical structure.

Different UV sources intensities with characteristic UV-A and UV-B proportions result in cosmetic tanning of the human skin.

The UV-A proportion primarily generates a superficial tan, which appears rapidly and is intensive but also fades more rapidly. The UV-B radiation is primarily responsible for more longterm tanning results.

Intended Use:

This sunlamp product is intended exclusively for cosmetic tanning of the human skin, for one person at a time, at the age of 18 or above.

Predicate devices:

All proposed devices are currently listed as Class I devices (510(k) exempt sunlamp product) and are legally offered for sale before September 2, 2014. In accordance with the final order these devices are used as predicate devices for substantial equivalence purposes.

{8}------------------------------------------------

Comparison to predicate devices:

The following table presents the technological characteristics and parameters of the identified predicate devices and the proposed devices.

Model No.Conversion KitBrand and type nameBody LampsWattsHigh Pressure FaceWattsUV-B facial lampsWattsShoulder LampsWattsMax. exposure time [min]
JK 103 / 32-0Soltron/XXS-35 SURFSUNDASH 32 Super Power32120N/AN/AN/AN/AN/AN/A20
JK 103 / 32-0Palm Beach Tan - SIGNITUREPalm Beach Tan / SIGNATURE32120N/AN/AN/AN/AN/AN/A20
JK 103 / 32-0Soltron/XXS-30 Super Power32120N/AN/AN/AN/AN/AN/A15
JK 103 / 32-1Soltron/XXS-35 SURFSUNDASH 32/1 Super Power321201400N/AN/AN/AN/A15
JK 103 / 32-1Soltron/XXS-35 SURFSUNDASH 32/1 Super Power321201400N/AN/AN/AN/A20
JK 103 / 34-3Soltron/XS-40 KISSErgoline/PASSION 34/3 Super Power34120340028N/AN/A15
JK 103 / 40-3Soltron/S-45 FLIRTErgoline/PASSION 40/3 Turbo Power(white)40160/120340028N/AN/A15
JK 103 / 40-3Soltron/S-45 FLIRTErgoline/PASSION 40/3 Turbo Power(black)40160/120340028N/AN/A15
JK 103 / 40-3 ACSoltron/S-45 FLIRT+ACErgoline/PASSION PLUS 40/3 Turbo Power(white)40160340028N/AN/A12
JK 103 / 40-3 ACSoltron/S-45 FLIRT+ACErgoline/PASSION PLUS 40/3 Turbo Power(black)40160340028N/AN/A12
JK 103 / SA 40-3 ACSUNANGEL 40040160340028N/AN/A15
JK 116 / 48-0Soltron/V-50 SPACE CABErgoline/ ESSENCE 48 SMART POWER48200N/AN/AN/AN/AN/AN/A9
Model No.Conversion KitBrand and type nameBody LampsWattsHigh Pressure FaceWattsUV-B facial lampsWattsShoulder LampsWattsMax. exposure time [min]
JK 116 / 48-0-Ergoline/ ESSENCE 48 SMART POWER48200N/AN/AN/AN/AN/AN/A9
JK 126 / 38-3 ACSoltron/M-55 RACERErgoline/Inspiration 500 SmartPerformance38160/180340028N/AN/A12
JK 126 / 42-4 ACSoltron/M-60 JUMPErgoline/Inspiration 600 DynamicPerformance42200440038N/AN/A12
JK 126 / SA 42-4 AC-SUN ANGEL DUO 60042200440038N/AN/A20
JK 126 / SA 42-4 AC-SUN ANGEL DUO 60042200440038N/AN/A15
JK 127 / 48-4 TTSoltron/XXL-115 STARLIGHTErgoline/Prestige 1000 IntelligentPerformance48200450038224010
JK 127 / 52-4 TTSoltron/XXL-125 EAGLEErgoline/Prestige 1400 IntelligentPerformance52200452038224010
JK 127 / SA 52-4 TT-SUN ANGEL DUO 140052200452038224012
JK 127 / SA 52-4 TT-SUN ANGEL DUO 140052200452038224015
JK 127 / SS 48-4-TT-Ergoline/Sun Select 1000 IntelligentPerformance48200450038224010
JK 127 / SS 52-4 TT-Ergoline/Sun Select 1400 IntelligentPerformance52200452038224010
JK 155 / 46-3Soltron/L-70 BUTTERFLYErgoline/AFFINITY 700 AdvancedPerformance4616035002872512
Model No.Conversion KitBrand and type nameBody LampsWattsHigh PressuraceNattsB facial lamNattsoulder LampsNattsMax. exposur time [min]
JK 155 / 46-3 ACSoltron/L-70 BUTTERFLYErgoline/AFFINITY 700 AdvancedPerformance4616035002872512
JK 155 / 50-4 TTSoltron/XL-85 WILDCATErgoline/AFFINITY 900 AdvancedPerformance5016045203872512
JK 70 / 0-30Soltron/XXL-95 SHARKErgoline/ Open Sun 1050275803640N/AN/AN/AN/A12
JK 79 / 48-4 TTSoltron/ XL-80 DYNAMIC POWER482004520N/AN/A224010
JK 79 / 52-4 TTSoltron/ XXL-110 DYNAMIC POWER52200पी52038224010
JK 92 / 43-3Ergoline/AFFINITY 500 Turbo Power431603500N/AN/A72512
JK 92 / 46-3-AC-Ergoline/AFFINITY 600 Turbo Power AirCondition461603500N/AN/A72512
JK 92 / 51-4 TTErgoline/AFFINITY 800 Turbo PowerTemptronic501604520N/AN/A72512
JK 98 / 48-0Soltron/V-55 UNIVERSEErgoline/SUNRISE 480 Turbo Power48200N/AN/AN/AN/AN/AN/Aರಿ
JK 98 / 48-0Soltron/V-55 UNIVERSEErgoline/SUNRISE 480 Turbo Power48200N/AN/AN/AN/AN/AN/Aರಿ
JK 116 / 48-0ESSENCE 48 10 MN PBT48160N/AN/AN/AN/AN/AN/A10
JK 100 / SA 52-4 TTSun Angel S52 20 min(with Sensor 3)5220052038224020
JK 100 / SA 52-4 TTSun Angel S52 15 min(with Sensor 3)52200520382240ਹੈ ਟ
JK 104 / SA 46-3 TTSun Angel S46 20 min(with Sensor 3)4620035202872520
JK 132 / SA 52-4 TTSun Angel Duo S52 20 min(with Sensor 3)5220052038224020
Model No.Conversion KitBrand and type nameBody LampsWattsHigh Pressure FaceWattsUV-B facial lampsWattsShoulder LampsWattsMax. exposure time [min]
JK 132 / SA 52-4 TT-Sun Angel Duo S52 15 min(with Sensor 3)52200452038224015
JK 103 / 34-0-Passion 34/0 20 min34120N/AN/AN/AN/AN/AN/A20
JK 103 / 34-0-Passion 34/0 15min34120N/AN/AN/AN/AN/AN/A15
JK 103 / 32-0-Flair 32/0 20 min32120N/AN/AN/AN/AN/AN/A20
JK 103 / 32-0-Flair 32/0 15 min32120N/AN/AN/AN/AN/AN/A15
JK 103 / 32-1-Flair 32/1 20 min321201400N/AN/AN/AN/A20
JK 103 / 32-1-Flair 32/1 15 min321201400N/AN/AN/AN/A15
JK 103 / 32-3-Flair 32/3 15 min321203400N/AN/AN/AN/A15
JK 103 / 32-3-Flair 32/3 20 min321203400N/AN/AN/AN/A20
JK 103 / 34-3-Coco Cabana XXS-35 34/3 15min34120340028N/AN/A15

{9}------------------------------------------------

K151400

Page 7 of 10

{10}------------------------------------------------

K151400

Page 8 of 10

{11}------------------------------------------------

K151400 Page 9 of 10

{12}------------------------------------------------

Summary of performed tests:

The proposed devices have been tested in respect to biocompatibility in accordance with the ISO 10993-series, electrical and mechanical safety in accordance with IEC 60601-1 (IEC 60335-2-27 and UL482) and EMC in accordance with IEC 60601-1-2 (CISPR 14-1 and 14-2).

The spectral emission of all sunlamp products was measured. The test procedure for measuring the spectral emission is in accordance with IEC 60335-2-27.

The irradiance ratio limits of all proposed sunlamp products are in accordance with 21 CFR 1040.20 (c) (1). The maximum timer intervals and exposure schedules have been determined according to FDA's "policy on maximum timer interval and exposure schedule for sunlamp products"

Conclusion:

JK Holding GmbH believes that the Ergoline, Soltron, Sun Angel, Sundash and Palm Beach Tan private label sunlamp products are substantially equivalent to currently legally marketed devices. Also all Special Controls as demanded by the Final Order are fulfilled. These devices do not introduce new indications for use, have the same technological characteristics and do not introduce new potential hazards or safety risks.

§ 878.4635 Sunlamp products and ultraviolet lamps intended for use in sunlamp products.

(a)
Identification. A sunlamp product is any device designed to incorporate one or more ultraviolet (UV) lamps intended for irradiation of any part of the living human body, by UV radiation with wavelengths in air between 200 and 400 nanometers, to induce skin tanning. This definition includes tanning beds and tanning booths. A UV lamp intended for use in sunlamp products is any lamp that produces UV radiation in the wavelength interval of 200 to 400 nanometers in air.(b)
Classification. Class II (special controls). The special controls for sunlamp products and UV lamps intended for use in sunlamp products are:(1) Conduct performance testing that demonstrates the following:
(i) Device meets appropriate output performance specifications such as wavelengths, energy density, and lamp life; and
(ii) Device's safety features, such as timers to limit UV exposure and alarms, function properly.
(2) Demonstrate that device is mechanically safe to prevent user injury.
(3) Demonstrate software verification, validation, and hazard analysis.
(4) Demonstrate that device is biocompatible.
(5) Demonstrate that device is electrically safe and electromagnetically compatible in its intended use environment.
(6)
Labeling —(i)Sunlamp products. (A) The warning statement below must appear on all sunlamp products and must be placed in a black box. This statement must be permanently affixed or inscribed on the product when fully assembled for use so as to be legible and readily accessible to view by the person who will be exposed to UV radiation immediately before the use of the product. It shall be of sufficient durability to remain legible throughout the expected lifetime of the product. It shall appear on a part or panel displayed prominently under normal conditions of use so that it is readily accessible to view whether the tanning bed canopy (or tanning booth door) is open or closed when the person who will be exposed approaches the equipment and the text shall be at least 10 millimeters (height). Labeling on the device must include the following statement: Attention: This sunlamp product should not be used on persons under the age of 18 years. (B) Manufacturers shall provide validated instructions on cleaning and disinfection of sunlamp products between uses in the user instructions.
(ii)
Sunlamp products and UV lamps intended for use in sunlamp products. Manufacturers of sunlamp products and UV lamps intended for use in sunlamp products shall provide or cause to be provided in the user instructions, as well as all consumer-directed catalogs, specification sheets, descriptive brochures, and Web pages in which sunlamp products or UV lamps intended for use in sunlamp products are offered for sale, the following contraindication and warning statements:(A) “Contraindication: This product is contraindicated for use on persons under the age of 18 years.”
(B) “Contraindication: This product must not be used if skin lesions or open wounds are present.”
(C) “Warning: This product should not be used on individuals who have had skin cancer or have a family history of skin cancer.”
(D) “Warning: Persons repeatedly exposed to UV radiation should be regularly evaluated for skin cancer.”
(c)
Performance standard. Sunlamp products and UV lamps intended for use in sunlamp products are subject to the electronic product performance standard at § 1040.20 of this chapter.