K Number
K183539

Validate with FDA (Live)

Manufacturer
Date Cleared
2019-04-08

(109 days)

Product Code
Regulation Number
878.4635
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

This sunlamp product is intended exclusively for aestetic tanning of the human skin, for one person at a time, at the age of 18 or above.

Device Description

The primary technical components of sunlamp products are artificial sources of UV radiation as well as a mechanical structure with a defined active surface. UV lamps emit different UV-A and UV-B proportions of the UV radiation. The UV-A proportion primarily generates a superficial tan, which appears rapidly and is intensive but also fades more rapidly. UV-B radiation is primarily responsible for more long-term tanning results. The Ergoline Sunrise 7200 Hybrid Light Technology / Hybrid Technology provides tanning results by combining traditional UV light with red light in a synergistic combination of light spectra.

AI/ML Overview

The provided text is a 510(k) summary for a sunlamp product. This type of document is for demonstrating substantial equivalence to a predicate device, not for proving a device meets specific acceptance criteria through a clinical study in the way an AI diagnostic device would.

Therefore, many of the requested sections about clinical study details (such as sample size for test sets, data provenance, number of experts for ground truth, adjudication methods, MRMC studies, standalone performance, and training set details) are not applicable to this document as it describes a Class II sunlamp product, not an AI/ML-driven medical device.

The document primarily focuses on demonstrating that the new device, Ergoline Sunrise 7200 Hybrid Technology, is substantially equivalent to a legally marketed predicate device (Sunrise 6200 and Sunrise 7200, K173565) by comparing their technological characteristics, intended use, and compliance with relevant safety standards.

Here's an attempt to answer the questions based on the available information, noting the inapplicability of some points:


1. Table of Acceptance Criteria and Reported Device Performance

For this type of device, "acceptance criteria" and "reported device performance" are primarily demonstrated through compliance with established safety and performance standards rather than specific diagnostic metrics. The document asserts compliance with these standards.

Acceptance Criteria/StandardReported Device Performance/Compliance
Intended Use/Indications for Use (Substantial Equivalence)Identical: "This sunlamp product is intended exclusively for aesthetic tanning of the human skin, for one person at a time, at the age of 18 or above."
Number of body lamps (Comparison to Predicate)Similar (Proposed: 34 UV lamps and 14 UV+Red Light lamps; Predicate: 48 UV lamps). Safety confirmed by IEC 62471 tests. Functionality and safety not impacted.
Watts (Lamps)Identical (200W for both proposed and predicate UV lamps).
Lamp item description (Comparison to Predicate)Similar (Proposed: GENESIS Type U Hybrid Performance - 200 W (UV light) and GENESIS Type R Hybrid Performance - 200 W (Mix of UV+Red Light); Predicate: GENESIS VHP9 Smart Performance 200 W). Safety confirmed by IEC 62471 tests. Functionality and safety not impacted.
Number of Beauty Light LEDsDifferent (Proposed: 60 (3x20); Predicate: none). Safety confirmed by IEC 62471 tests. Functionality and safety not impacted.
Watts (Beauty Light LEDs)Different (Proposed: 45 (3x15); Predicate: none). No impact on safety.
Max exposure time [min]Identical (9 minutes).
Electrical requirementsIdentical (230V 3Ø or 230V 2Ø).
Total power consumption [Watts]Similar (Proposed: 10600W; Predicate: 9500W). Higher due to different electronic lamp ballasts, performance not impaired.
Rated overcurrent protection device (circuit breaker)Identical (40A / 3-pole 3Ø or 70A / 2-pole 2Ø).
Number of wiresIdentical (4 3Ø or 3 2Ø).
Irradiance ratio in accordance with 1040.20Fulfilled (for both proposed and predicate).
Electrical safety (IEC 60601-1)Identical (Compliant for both proposed and predicate).
Electromagnetic compatibility (IEC 60601-1-2)Identical (Compliant for both proposed and predicate).
Biocompatibility (ISO 10993-series)Proposed devices tested and in compliance.
U.S. performance standard 21CFR 1040.20Proposed devices tested and in compliance.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Not Applicable: This document describes a sunlamp product, not an AI/ML diagnostic or predictive device that typically uses "test sets" of data. The "testing" referred to is for compliance with electrical, mechanical, and safety standards like IEC 62471, IEC 60601-1, IEC 60601-1-2, ISO 10993, and 21 CFR 1040.20. These are engineering/device performance tests, not clinical studies with patient data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Not Applicable: As above, this is not an AI/ML diagnostic device requiring expert-established ground truth from clinical data. Compliance is assessed against technical standards.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Not Applicable: No clinical "test set" and thus no adjudication method for ground truth determination.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not Applicable: This is not an AI-assisted diagnostic device, so no MRMC study would be relevant.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not Applicable: This is a sunlamp product, not an algorithm, so "standalone performance" in this context is irrelevant.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • Not Applicable: The "ground truth" for this device's compliance is defined by the technical specifications and safety standards (e.g., measured irradiance levels, electrical safety parameters, biocompatibility test results). It is not based on clinical outcomes, pathology, or expert consensus on a diagnosis.

8. The sample size for the training set

  • Not Applicable: This is a physical sunlamp product and does not involve AI/ML models that require a training set.

9. How the ground truth for the training set was established

  • Not Applicable: As there is no training set, this question is not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym followed by the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of the square.

JK-Holding GmbH Juergen Gerstenmeier Executive Director Research & Development Koehlershohner Strabe 60 Windhagen, 53578 DE

April 8, 2019

Re: K183539

Trade/Device Name: Ergoline Sunrise 7200 Hybrid Technology Regulation Number: 21 CFR 878.4635 Regulation Name: Sunlamp Products And Ultraviolet Lamps Intended For Use In Sunlamp Products. Regulatory Class: Class II Product Code: LEJ Dated: March 5, 2019 Received: March 7, 2019

Dear Juergen Gerstenmeier:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K183539

Device Name Ergoline Sunrise 7200 Hybrid Technology

Indications for Use (Describe)

This sunlamp product is intended exclusively for aestetic tanning of the human skin, for one person at a time, at the age of 18 or above.

Type of Use (Select one or both, as applicable)

] Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Chapter 5 510(k) Summary

{4}------------------------------------------------

510(k) Summary of Safety and Effectiveness

In accordance with the requirements of the Safe Medical Device Act, JK-HOLDING GmbH herewith submits a Summary of Safety and Effectiveness.

This 510(k) summary for the Ergoline Sunrise 7200 Hybrid Technology and their further private label sunlamp products meets the requirements of 21 CFR 807.92

Submitter information:JK-HOLDING GmbHKoehlershohner Strasse 6053578 WindhagenGermanyPhone: +49 (0) 2224 / 818-0Fax: +49 (0) 2224 / 818-116
Facility:JK-Products GmbHRottbitzer Strasse 6953604 Bad HonnefGermanyEstablishment Registration Number: 9615023Owner/Operator Number: 9034303
Official Correspondent:Juergen GerstenmeierKoehlershohner Strasse 6053578 WindhagenGermanyPhone: +49 (0) 2224 / 818-330Email: juergen.gerstenmeier@de.jk-group.net
US Agent (Contact):Kevin HartJK PRODUCTS & SERVICES, INC.411 West Washington Ave Suite BJonesboro, AR 72401Phone: 870-2682862Fax: 870 9359046Email: Kevin.Hart@Jkamerica.com
Date Prepared:April 8th, 2019

{5}------------------------------------------------

Device(s) Identification:

Device trade name:Ergoline Sunrise 7200 Hybrid Technology
Common name:Sunlamp product

The following table presents the different variants of the Ergoline Sunrise 7200 Hybrid Light Technology / Hybrid Technology sunlamp product.

All variants use the same basic device technology and differ only in effect lighting, comfort features and private labelling. All variants use the hybrid light technology.

Model-No.:Variant Name:
JK 158 / 48-0 HTErgoline Sunrise 7200 Hybrid Light Technology Planet Fitness Sound/Bluetooth
JK 158 / 48-0 HT VSErgoline Sunrise 7200 Hybrid Light Technology Planet Fitness Sound /Bluetooth / Vibra Shape
JK 158 / 48-0 HTErgoline Sunrise 7200 Hybrid Technology / Sound/ Bluetooth
JK 158 / 48-0 HT VSErgoline Sunrise 7200 Hybrid Technology Sound / Bluetooth / Vibra Shape

The basic products can be upgraded with conversion kits, consisting of stickers with the private label naming, to Soltron, Palm Beach Tan or further private label sunlamp products, The conversion kit only serves to label the basic product as a Soltron, Palm Beach Tan or further private label sunlamp product. It does not alter any technical specification or aspect of the devices.

Classification of the device:

Device Classification Name:Booth, Sun Tan
Product Code:LEJ C.F.R.
Section:878.4635
Classification Panel:General & Plastic Surgery
Device Class:Class II

Device Description:

The primary technical components of sunlamp products are artificial sources of UV radiation as well as a mechanical structure with a defined active surface.

UV lamps emit different UV-A and UV-B proportions of the UV radiation. The UV-A proportion primarily generates a superficial tan, which appears rapidly and is intensive but also fades more rapidly. UV-B radiation is primarily responsible for more long-term tanning results.

The Ergoline Sunrise 7200 Hybrid Light Technology / Hybrid Technology provides tanning results by combining traditional UV light with red light in a synergistic combination of light spectra.

Indications for use:

This sunlamp product is intended exclusively for aesthetic tanning of the human skin, for one person at a time, at the age of 18 or above.

{6}------------------------------------------------

Legally Marketed Predicate device:

Device Name:Sunrise 6200 and Sunrise 7200
Model No.:JK 158 / 48-0 and JK 158 / 48-0 VS
510k number:K173565
Device Classification Name:Booth, Sun Tan
Product Code:LEJ
C.F.R. Section.:878.4635
Classification Panel:General & Plastic Surgery
Device Class:Class II

Comparison to predicate device:

The following table presents the comparison of technological characteristics, functions and parameters of the identified predicate device and the proposed devices.

ParameterProposed deviceJK 158 / 48-0 HTJK 158 / 48-0 HT VSPredicate deviceJK 158 / 48-0JK 158 / 48-0 VSEvaluation
IntendedUse/Indications for UseThis sunlamp productis intendedexclusively foraesthetic tanning ofthe human skin, forone person at a time,at the age of 18 orabove.This sunlamp productis intendedexclusively forcosmetic tanning ofthe human skin, forone person at a time,at the age of 18 orabove.Identical
Number of body lamps48 (34 UV lamps and14 UV+Red Lightlamps)48 (all UV lamps)Similar:The irradiancecharacteristics arecomparable. The safetyhas been confirmedadditionally bycorresponding tests inaccordance with IEC62471.In terms of functionalityand safety there is noimpact due to thisdifference.
Watts200200Identical
ParameterProposed deviceJK 158 / 48-0 HTJK 158 / 48-0 HT VSPredicate deviceJK 158 / 48-0JK 158 / 48-0 VSEvaluation
Lamp item descriptionGENESIS Type UHybrid Performance -200 W(UV light) andGENESIS Type RHybrid Performance -200 W(Mix of UV+Red Light)GENESIS VHP9Smart Performance200 WSimilar:The irradiancecharacteristics arecomparable. The safetyhas been confirmedadditionally bycorresponding tests inaccordance with IEC62471.In terms of functionalityand safety there is noimpact due to thisdifference
Number of Beauty LightLEDs60 (3x20)noneThe irradiancecharacteristics arecomparable. The safetyhas been confirmedadditionally bycorresponding tests inaccordance with IEC62471.In terms of functionalityand safety there is noimpact due to thisdifference.
Watts45 (3x15)noneNo impact (see above)
Max exposure time[min]99Identical
Electrical requirements230V 3Ø or 230V 2Ø230V 3Ø or 230V 2ØIdentical
Total powerconsumption [Watts]106009500Similar:The power is higherdue to the use ofdifferent electroniclamp ballasts.Performance is notimpaired.
ParameterProposed deviceJK 158 / 48-0 HTJK 158 / 48-0 HT VSPredicate deviceJK 158 / 48-0JK 158 / 48-0 VSEvaluation
Rated overcurrentprotection device(circuit breaker)40A / 3-pole 3Ø or70A / 2-pole 2Ø40A / 3-pole 3Ø or70A / 2-pole 2Øidentical.
Number of wires4 3Ø or 3 2Ø4 3Ø or 3 2Øidentical
Irradiance ratio inaccordance with1040.20fulfilledfulfilledidentical
Electrical safetyIEC 60601-1IEC 60601-1identical
ElectromagneticcompatibilityIEC 60601-1-2IEC 60601-1-2identical

{7}------------------------------------------------

{8}------------------------------------------------

Summary of performance testing:

The proposed devices have been tested in respect to biocompatibility in accordance with the ISO 10993-series, electrical and mechanical safety in accordance with IEC 60601-1 and EMC in accordance with IEC 60601-1-2.

The proposed devices are in compliance with U.S. performance standard 21CFR 1040.20

Conclusion:

JK Holding GmbH believes that the proposed devices are substantially equivalent to the legally marketed predicate device. Also, all Special Controls are fulfilled. The proposed devices do not introduce new indications for use, have the same technological characteristics and do not introduce new potential hazards or safety risks.

§ 878.4635 Sunlamp products and ultraviolet lamps intended for use in sunlamp products.

(a)
Identification. A sunlamp product is any device designed to incorporate one or more ultraviolet (UV) lamps intended for irradiation of any part of the living human body, by UV radiation with wavelengths in air between 200 and 400 nanometers, to induce skin tanning. This definition includes tanning beds and tanning booths. A UV lamp intended for use in sunlamp products is any lamp that produces UV radiation in the wavelength interval of 200 to 400 nanometers in air.(b)
Classification. Class II (special controls). The special controls for sunlamp products and UV lamps intended for use in sunlamp products are:(1) Conduct performance testing that demonstrates the following:
(i) Device meets appropriate output performance specifications such as wavelengths, energy density, and lamp life; and
(ii) Device's safety features, such as timers to limit UV exposure and alarms, function properly.
(2) Demonstrate that device is mechanically safe to prevent user injury.
(3) Demonstrate software verification, validation, and hazard analysis.
(4) Demonstrate that device is biocompatible.
(5) Demonstrate that device is electrically safe and electromagnetically compatible in its intended use environment.
(6)
Labeling —(i)Sunlamp products. (A) The warning statement below must appear on all sunlamp products and must be placed in a black box. This statement must be permanently affixed or inscribed on the product when fully assembled for use so as to be legible and readily accessible to view by the person who will be exposed to UV radiation immediately before the use of the product. It shall be of sufficient durability to remain legible throughout the expected lifetime of the product. It shall appear on a part or panel displayed prominently under normal conditions of use so that it is readily accessible to view whether the tanning bed canopy (or tanning booth door) is open or closed when the person who will be exposed approaches the equipment and the text shall be at least 10 millimeters (height). Labeling on the device must include the following statement: Attention: This sunlamp product should not be used on persons under the age of 18 years. (B) Manufacturers shall provide validated instructions on cleaning and disinfection of sunlamp products between uses in the user instructions.
(ii)
Sunlamp products and UV lamps intended for use in sunlamp products. Manufacturers of sunlamp products and UV lamps intended for use in sunlamp products shall provide or cause to be provided in the user instructions, as well as all consumer-directed catalogs, specification sheets, descriptive brochures, and Web pages in which sunlamp products or UV lamps intended for use in sunlamp products are offered for sale, the following contraindication and warning statements:(A) “Contraindication: This product is contraindicated for use on persons under the age of 18 years.”
(B) “Contraindication: This product must not be used if skin lesions or open wounds are present.”
(C) “Warning: This product should not be used on individuals who have had skin cancer or have a family history of skin cancer.”
(D) “Warning: Persons repeatedly exposed to UV radiation should be regularly evaluated for skin cancer.”
(c)
Performance standard. Sunlamp products and UV lamps intended for use in sunlamp products are subject to the electronic product performance standard at § 1040.20 of this chapter.