K Number
K191881

Validate with FDA (Live)

Manufacturer
Date Cleared
2019-10-28

(105 days)

Product Code
Regulation Number
878.4635
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

KBL Sunlamp Hybrid Products are over-the-counter tanning devices emitting ultraviolet light in the UVA and UVB region of the spectrum intended for tanning of the human skin of an adult person.

Device Description

KBL Sunlamp Hybrid Products are whole-body tanning devices basically consisting of a mechanical structure equipped with artificial light sources using the hybrid light technology. The hybrid light technology produces ultraviolet radiation and radiation in the visual range of the light spectrum. The UV light is intended for irradiation of any part of the living human body to induce skin tanning. The user is lying on a bench.

AI/ML Overview

The provided document is a 510(k) Pre-market Notification for KBL GmbH sunlamp products, specifically the KBL 7000 alpha hybridSun, KBL 7900 alpha hybridSun, and KBL 8000 alpha hybridSun. This document focuses on demonstrating substantial equivalence to predicate devices rather than proving a device meets specific performance acceptance criteria for a novel AI/software medical device.

Therefore, the information requested in your prompt regarding acceptance criteria, study details, sample sizes, ground truth establishment, expert adjudication, and MRMC studies, which are typically found in the context of validating performance for diagnostic or therapeutic AI/software, is not present in this document.

This document details the following for sunlamp products:

  1. Device Type: Sunlamp products and ultraviolet lamps intended for use in sunlamp products.
  2. Regulation Number: 21 CFR 878.4635
  3. Regulation Name: Sunlamp products and ultraviolet lamps intended for use in sunlamp products
  4. Regulatory Class: Class II
  5. Product Code: LEJ
  6. Indications for Use: "KBL Sunlamp Hybrid Products are over-the-counter tanning devices emitting ultraviolet light in the UVA and UVB region of the spectrum intended for tanning of the human skin of an adult person." This is identical to the predicate device.
  7. Comparison to Predicate: The document largely focuses on comparing the technological characteristics (number of lamps, wattage, dimensions, etc.) of the proposed devices to predicate devices (KBL 6800 alpha and KBL 7900 alpha). The evaluation states that differences are "Equivalent" or "Identical" and do not impact functionality or safety.
  8. Performance Data: Non-clinical performance tests were conducted to ensure compliance with relevant standards for sunlamp products, including:
    • Biocompatibility testing (DIN EN ISO 10993 series)
    • Electrical and mechanical safety testing (IEC 60601-1, IEC 60335-1, UL482)
    • Electromagnetic compatibility testing (IEC 60601-1-2)
    • Spectral emission measurement (IEC 60335-2-27)
    • Software verification and validation (FDA Guidance for Software in Medical Devices)
    • Performance Standards testing (21 CFR 1040.20)
    • Irradiance ratio limits (21 CFR 1040.20(c)(1))
    • Maximum timer intervals and exposure schedules (FDA policy)

Key takeaway: This 510(k) submission is for physical sunlamp devices, not an AI/software medical device. The "performance data" refers to compliance with safety and electrical standards for sunlamps, not the type of diagnostic performance (e.g., sensitivity, specificity) that requires the detailed study design elements you asked for. Therefore, I cannot extract the requested information from this document.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The Department of Health & Human Services logo features a stylized human figure, while the FDA part includes the acronym "FDA" in a blue square, followed by "U.S. FOOD & DRUG" and "ADMINISTRATION" in blue text.

October 28, 2019

KBL GmbH Ralf de Andreis Director Operations Ringstrasse 24-26 Dernbach, 56307 De

Re: K191881

Trade/Device Name: KBL 7000 alpha hybridSun, KBL 7900 alpha hybridSun, KBL 8000 alpha hybridSun Regulation Number: 21 CFR 878.4635 Regulation Name: Sunlamp products and ultraviolet lamps intended for use in sunlamp products Regulatory Class: Class II Product Code: LEJ Dated: September 24, 2019 Received: September 26, 2019

Dear Ralf de Andreis:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

FOR (Neil Ogden) Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Office of Surgical & Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

510(k) Number (if known)

Device Name

Indications for Use (Describe)

Type of Use (Select one or both, as applicable)

_ Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for KBL GmbH, The Beauty & Health Company. The logo features a yellow circle with white lines on the left, followed by the letters "KBL" in gray. The words "GmbH" and "The Beauty & Health Company" are in a smaller font size and are located to the right and below the letters "KBL", respectively.

510(k) SUMMARY

(according to 21 CFR 807.92)

1. GENERAL INFORMATION

Submitter:KBL GmbHRingstrasse 24-26, 56307 Dernbach, GERMANYPhone: +49 (0) 2689 9426-0Fax: +49 (0) 2689 9426-66Establishment registration no. 3006116732
Official Correspondent:Ralf de AndreisRingstrasse 24-26, 56307 Dernbach, GERMANYPhone: +49 (0) 2689 9426-850Fax: +49 (0) 2689 9426-500E-Mail: randreis@kbl.de
Designated U.S. Agent (Contact):Shepard BentleyBentley Biomedical Consulting LLC28241 Crown Valley Parkway, Suite 510(k)Laguna Niguel, CA 92677, USAPhone: +1 949 374 9187E-Mail: sbentley@bentleybiomed.com
Preparation date:June 27, 2019

2. PROPOSED DEVICE(S)

Device NameKBL Sunlamp Hybrid Products
Device Trade Name(s)KBL 7000 alpha hybridSunKBL 7900 alpha hybridSunKBL 8000 alpha hybridSun
Device common nameTanning device
Device Classification NameBooth, Sun Tan
Product CodeLEJ (21CFR 878.4635)
Classification PanelGeneral & Plastic Surgery
Device ClassClass II - Special Controls

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for KBL GmbH, a beauty and health company. The logo features a yellow circle with white lines, followed by the letters "KBL" in gray. Below the letters is the phrase "The Beauty & Health Company" in a smaller font, also in gray.

3. PREDICATE DEVICE(S)

Device NameKBL Sunlamp Products
Device Trade Name(s)KBL 7900 alphaKBL 6800 alpha
Device common nameTanning device
Device Classification NameBooth, Sun Tan
Product CodeLEJ (21CFR 878.4635)

levice n Tan FR 878.4635) General & Plastic Surgery Class II - Special Controls K151962

4. DEVICE DESCRIPTION

Classification Panel

Device Class

510(k) No.

KBL Sunlamp Hybrid Products are whole-body tanning devices basically consisting of a mechanical structure equipped with artificial light sources using the hybrid light technology. The hybrid light technology produces ultraviolet radiation and radiation in the visual range of the light spectrum.

The UV light is intended for irradiation of any part of the living human body to induce skin tanning. The user is lying on a bench.

5. INDICATIONS FOR USE

The Indications for use for the Proposed Devices are identical to the Predicate Devices as shown in Table 1 below:

Table 1

Proposed DevicePredicate Device
KBL Sunlamp Hybrid Products are over-the-counter tanning devices emitting ultraviolet light in the UVA and UVB region of the spectrum intended for tanning of the human skin of an adult person.KBL Sunlamp Products are over-the-counter tanning devices emitting ultraviolet light in the UVA and UVB region of the spectrum intended for tanning of the human skin of an adult person.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image is a logo for KBL GmbH, The Beauty & Health Company. The logo features a yellow circle with lines emanating from the center, followed by the letters "KBL" in gray. The words "The Beauty & Health Company" are written in a smaller font below the letters, and the letters "GmbH" are written in a smaller font to the right of the letter "L".

6. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS

We believe that the Technological Characteristics of the Proposed Devices in comparison to the Predicate Devices show that they are substantially equivalent. See table 2, 3 and 4 below:

Table 2PROPOSED DEVICEPREDICATE DEVICEEVALUATION
KBL Sunlamp Hybrid ProductKBL Sunlamp Product
Device Trade NameKBL 7000 alpha hybridSunKBL 6800 alphaN/A
Number of LampsCanopy18x pureSunlight lamps /8x smartSunlight (uv+red light)lamps20x megaLine lamps /6x megaLine lamps /EquivalentThe irradiance characteristics are stillequivalent to the predicate device asprovided under Performance Testingtable (radiometric measurementresults). The safety was confirmedby appropriate tests according to IEC62471. In terms of funtionality and
Watt160160 / 80safety there is no impact due to thisdifference.
Number of LampsBench14x pureSunlight lamps /6x smartSunlight (uv+red light)lamps20x megaLine lampsEquivalentThe irradiance characteristics are stillequivalent to the predicate device asprovided under Performance Testingtable (radiometric measurementresults). The safety was confirmedby appropriate tests according to IEC62471. In terms of functionality andsafety there is no impact due to this
Watt180180difference.
Number ofFacial Lamps4x megaLine HP 8004x megaLine HP 800Identical
Watt600600Identical
Number ofShoulder Lamps2x ms 2502x ms 250Identical
Watt250250Identical
Number of LEDs(vis red light)40x beautyBooster40x Enhanced LEDIlluminationIdentical
Watt11Identical
Max. ExposureTime in min.1010Identical
Dimensions inInches(height x width xdepth)when closed~60,59"x ~93,86"x ~56,34"when open~80,24"x ~93,86"x ~56,34"when closed:~60,39"x ~93,86"x~56,34"when open:~76,46"x ~93,86"x~56,34"EquivalentThe dimensions are equivalent to thepredicate device. In terms offunctionality and safety there is noimpact due to this difference.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for KBL GmbH, a company focused on beauty and health. The logo features the letters "KBL" in a bold, sans-serif font, with "GmbH" in a smaller font to the right of the "L". Below the letters is a horizontal line, and below that is the text "The Beauty & Health Company" in a smaller, lighter font.

Table 3PROPOSED DEVICEKBL Sunlamp Hybrid ProductPREDICATE DEVICEKBL Sunlamp ProductEVALUATION
Device Trade nameKBL 7900 alpha hybridSunKBL 7900 alphaN/A
Number ofLamps Canopy18x pureSunlight lamps /8x smartSunlight (uv+red light)lamps20x megaLine lamps /6x megaLine lamps /EquivalentThe irradiance characteristics are stillequivalent to the predicate device asprovided under Performance Testingtable (radiometric measurementresults). The safety was confirmedby appropriate tests according to IEC62471. In terms of functionality andsafety there is no impact due to thisdifference.
Watt160160 / 80difference.
Number ofLamps Bench16x pureSunlight lamps /8x smartSunlight (uv+red light)lamps24x megaLine lampsEquivalentThe irradiance characteristics are stillequivalent to the predicate device asprovided under Performance Testingtable (radiometric measurementresults). The safety was confirmedby appropriate tests according to IEC62471. In terms of functionality andsafety there is no impact due to thisdifference.
Watt180180difference.
Number ofFacial Lamps4x megaLine HP 8004x megaLine HP 800Identical
Watt600600Identical
Number of LEDs(vis red light)40x beautyBooster40x Enhanced LEDIlluminationIdentical
Watt11Identical
Number ofShoulder Lamps2x ms 2502x ms 250Identical
Watt250250Identical
Max. ExposureTime in min.1010Identical
Dimensions in Inches(height x width xdepth)when closed67,44"x93,86"x56,34"when open82,16"x93,86"x56,34"when closed~67,44"x ~93,86"x~56,34"when open~82,16"x ~93,86"x~56,34"Identical
Table 4PROPOSED DEVICEKBL Sunlamp Hybrid ProductPREDICATE DEVICEKBL Sunlamp ProductEVALUATION
Device Trade nameKBL 8000 alpha hybridSunKBL 7900 alphaN/A
Number of LampsCanopy18x pureSunlight lamps /8x smartSunlight (uv+red light)lamps20x megaLine lamps /6x megaLine lamps /EquivalentThe irradiance characteristics are stillequivalent to the predicate device asprovided under Performance Testingtable (radiometric measurementresults). The safety was confirmed byappropriate tests according to IEC62471. In terms of functionality andsafety there is no impact due to this
Watt160160 / 80difference.
Number of LampsBench16x pureSunlight lamps /8x smartSunlight (uv+red light)lamps24x megaLine lampsEquivalentThe irradiance characteristics are stillequivalent to the predicate device asprovided under Performance Testingtable (radiometric measurementresults). The safety was confirmed byappropriate tests according to IEC62471. In terms of functionality andsafety there is no impact due to this
Watt180180difference.
Number of FacialLamps4x megaLine HP 8004x megaLine HP 800Identical
Watt600600Identical
Number of ShoulderLamps2x ms 2502x ms 250Identical
Watt250250Identical
Number of LEDs(vis red light)70x beautyBooster40x Enhanced LEDIlluminationEquivalentThe irradiance characteristics are stillequivalent to the predicate device asprovided under Performance Testingtable (radiometric measurementresults). The safety was confirmed byappropriate tests according to IEC62471. In terms of functionality andsafety there is no impact due to thisdifference.
Watt11Identical
Max. Exposure Timein min.1010Identical
Dimensions in Inches(height x width xdepth)when closed:67,44"x93,86"x~56,34"when open:~82,16"x ~93,86"x ~56,34"when closed~67,44"x~93,86"x ~56,34"when open: ~82,16"x~93.86"x ~56.34"Identical

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for KBL GmbH, The Beauty & Health Company. The logo features a yellow circle with white lines inside, followed by the letters "KBL" in gray. The words "The Beauty & Health Company" are written in gray below a yellow line, and the letters "GmbH" are written in gray to the right of the "L" in "KBL".

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for KBL GmbH, The Beauty & Health Company. The logo features a yellow circle with lines emanating from the center, followed by the letters "KBL" in gray. Below the letters is a yellow line, and below that is the text "The Beauty & Health Company" in a smaller, gray font.

7. PERFORMANCE DATA

The following non-clinical performance tests were performed for all proposed KBL sunlamp products:

  • Biocompatibility testing in accordance with DIN EN ISO 10993-1:2010-04; . DIN EN ISO 10993-5:2009-10; DIN EN ISO 10993-12:2012-10;
  • . Electrical and mechanical safety testing according to IEC 60601-1:2005 + Corr.:2006 + Corr.:2007 + A1:2012 (IEC 60335-1:2012, and UL482:2005/09/02 Ed.9 Rev. 2013/10/03)
  • Electromagnetic compatibility testing in compliance with IEC 60601-1-2:2014. .
  • . Spectral emission measurement based on the test procedure for measuring the spectral emission in accordance to IEC 60335-2-27:2013.
  • Software verification and validation testing according to FDA's Guidance for Industry and . FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices".
  • . Performance Standards testing in accordance with 21 CFR 1040.20
  • Irradiance ratio limits are in accordance with 21 CFR 1040.20(c)(1).
  • . Maximum timer intervals and exposure schedules have been determined according to FDA's "Policy on maximum timer interval and exposure schedule for sunlamp products".

8. CONCLUSION

Based on an analysis of the technological characteristics, non-clinical performance data and indications for use, KBL GmbH believes that the Proposed Devices are substantially equivalent to the legally marketed Predicate Devices and do not raise any new issues of safety and effectiveness.

End of 510(k) Summary

§ 878.4635 Sunlamp products and ultraviolet lamps intended for use in sunlamp products.

(a)
Identification. A sunlamp product is any device designed to incorporate one or more ultraviolet (UV) lamps intended for irradiation of any part of the living human body, by UV radiation with wavelengths in air between 200 and 400 nanometers, to induce skin tanning. This definition includes tanning beds and tanning booths. A UV lamp intended for use in sunlamp products is any lamp that produces UV radiation in the wavelength interval of 200 to 400 nanometers in air.(b)
Classification. Class II (special controls). The special controls for sunlamp products and UV lamps intended for use in sunlamp products are:(1) Conduct performance testing that demonstrates the following:
(i) Device meets appropriate output performance specifications such as wavelengths, energy density, and lamp life; and
(ii) Device's safety features, such as timers to limit UV exposure and alarms, function properly.
(2) Demonstrate that device is mechanically safe to prevent user injury.
(3) Demonstrate software verification, validation, and hazard analysis.
(4) Demonstrate that device is biocompatible.
(5) Demonstrate that device is electrically safe and electromagnetically compatible in its intended use environment.
(6)
Labeling —(i)Sunlamp products. (A) The warning statement below must appear on all sunlamp products and must be placed in a black box. This statement must be permanently affixed or inscribed on the product when fully assembled for use so as to be legible and readily accessible to view by the person who will be exposed to UV radiation immediately before the use of the product. It shall be of sufficient durability to remain legible throughout the expected lifetime of the product. It shall appear on a part or panel displayed prominently under normal conditions of use so that it is readily accessible to view whether the tanning bed canopy (or tanning booth door) is open or closed when the person who will be exposed approaches the equipment and the text shall be at least 10 millimeters (height). Labeling on the device must include the following statement: Attention: This sunlamp product should not be used on persons under the age of 18 years. (B) Manufacturers shall provide validated instructions on cleaning and disinfection of sunlamp products between uses in the user instructions.
(ii)
Sunlamp products and UV lamps intended for use in sunlamp products. Manufacturers of sunlamp products and UV lamps intended for use in sunlamp products shall provide or cause to be provided in the user instructions, as well as all consumer-directed catalogs, specification sheets, descriptive brochures, and Web pages in which sunlamp products or UV lamps intended for use in sunlamp products are offered for sale, the following contraindication and warning statements:(A) “Contraindication: This product is contraindicated for use on persons under the age of 18 years.”
(B) “Contraindication: This product must not be used if skin lesions or open wounds are present.”
(C) “Warning: This product should not be used on individuals who have had skin cancer or have a family history of skin cancer.”
(D) “Warning: Persons repeatedly exposed to UV radiation should be regularly evaluated for skin cancer.”
(c)
Performance standard. Sunlamp products and UV lamps intended for use in sunlamp products are subject to the electronic product performance standard at § 1040.20 of this chapter.