| Cardiovascular | §870.2810, I | DSF | Recorder, Paper Chart |
| Cardiovascular | §870.2850
The N10, N12, N15, N10MPro, N12MPro, NM15Pro Multi-parameter Patient Monitors are intended for monitoring, displaying, reviewing, storing, alarming and transferring of multiple physiological parameters including ECG (3-lead, 5-lead, 12-lead selectable), Arrhythmia Analysis, ST Segment Analysis, QT Analysis, Heart Rate (HR) and Heart-Rate-Variability(HRV)), interpretations of resting 12-lead ECG, Respiration rate(Resp), Temperature(Temp), Pulse Oxygen Saturation (SpO2), Pulse Rate (PR), Non-invasive Blood Pressure (NIBP), Invasive Blood Pressure (IBP), Pulmonary Artery Wedge Pressure (PAWP), Cardiac Output (C.O.), Carbon Dioxide (CO2). The N10MPro, N12MPro, NM15Pro Multi-parameter Patient Monitors are also intended for monitoring, displaying, reviewing, storing, alarming and transferring of physiological parameters including Masimo Rainbow SpO2, Anesthesia gas (AG), oxygen (O2) respiratory gas monitoring, Bispectral Index (BIS), Respiration Mechanics (RM) and Neuromuscular Transmission Monitoring (NMT). All the parameters can be monitored on single adult, pediatric, and neonatal patient except for the following:
- Arrhythmia analysis is intended to use on adult patients only and is not intended and shall not be used on pediatric and neonatal population.
- NIBP measurement continual mode is not applicable to neonates.
- When using COMEN SpO2, the monitor is intended to be used on adult patient only.
- PAWP is intended for adult and pediatric patients only.
- C.O. measurement is intended for adult patients only.
- BIS monitoring is intended for adult patients only.
- RM is intended for adult and pediatric patients only.
- NMT monitoring is intended for adult and pediatric patients only.
The monitors are to be used in healthcare facilities by healthcare professionals or under their guidance.
The Multi-parameter Patient monitors are not intended for emergency and transport use, aircraft environment or home use.
The monitors are not intend for use as apnea monitors.
The monitors are not intended for use in MRI or CT environments.
The monitors are not used on patients who have a demonstrated need for cardiac monitoring known arrhythmias of VT, Accelerated Idioventricular rhythm and Torsades de Pointes.
There are six (6) models under evaluation, namely N10, N12, N15, N10MPro, N12MPro, N15MPro. All models share the same intended condition of use, the same intended patient population and operator profile, biological safety characteristic and principle of operation. All these models are the same on electric and electrical circuit and components, mechanical construction, software and alarm system. The only difference lies on the screen and configuration of with/without plug-in module slot and the number of battery packs.
N/A
Ask a specific question about this device
(178 days)
FloTrac sensors; FloTrac Jr sensors; Acumen IQ sensors; VolumeView sensors Regulation Number: 21 CFR 870.2850
|
| RegulationNumber / Name | 21 CFR §870.2850
The FloTrac sensor is indicated for use in intravascular pressure monitoring. It is also indicated for use with the Edwards arterial pressure based cardiac output monitoring devices or hardware to measure cardiac output. They are intended to be used in adult patients.
The FloTrac Jr sensor is indicated for use in intravascular pressure monitoring. It is also indicated for use with the Edwards arterial pressure based cardiac output monitoring devices or hardware to measure cardiac output. The FloTrac Jr sensor is indicated for use in pediatric patients ≥ 12 years of age.
The Acumen IQ sensor is indicated for use in intravascular pressure monitoring. It is also indicated for use with the Edwards arterial pressure based cardiac output monitoring devices or hardware to measure cardiac output. They are intended to be used in adult patients.
The VolumeView sensor is indicated for use in intravascular pressure monitoring. It is also indicated for use with the Edwards arterial pressure based cardiac output monitoring devices or hardware to measure cardiac output.
The FloTrac, Acumen IQ, and VolumeView sensors are constructed from two disposable pressure transducers that convert a physiological signal (or mechanical pressure) to an electrical signal that is transmitted through the cable to the patient monitor. The sensors have a straight, flow-through design in which fluid is passed directly across the pressure sensor. The sensors are comprised of a pressure sensitive silicon chip with two electrodes for excitation voltage and two electrodes for signal output. A polycarbonate housing with an integral stopcock at one end, and an integral flush device at the other end, encloses the sensors.
The provided FDA 510(k) summary for the FloTrac, FloTrac Jr, Acumen IQ, and VolumeView sensors does not contain detailed information about specific acceptance criteria and the study that proves the device meets those criteria in the typical format requested for an AI/ML-based medical device.
This document describes a substantial equivalence determination for extravascular blood pressure transducers (sensors) manufactured by Edwards Lifesciences, LLC. The core of the submission is that the subject devices are identical to the predicate devices in terms of intended use, indications for use, and technological characteristics, EXCEPT for changed pressure tubing and IV set component materials.
Therefore, the "study" described here is primarily focused on demonstrating that these material changes do not introduce new safety or effectiveness concerns, rather than validating an AI/ML algorithm's diagnostic performance against established ground truth.
Here's a breakdown based on the provided text, addressing your questions where possible, and noting where the information is not applicable or not present:
Overview of the Device and Study's Focus:
The devices in question are FloTrac, FloTrac Jr, Acumen IQ, and VolumeView sensors, which are intravascular pressure monitoring devices that also work with Edwards' arterial pressure-based cardiac output monitoring hardware. The 510(k) submission (K242909) is for modifications to these existing devices, specifically changes to the pressure tubing and IV set component materials. The premise of the submission is that these material changes do not alter the fundamental performance or safety in a way that would require new clinical performance studies typical for an AI/ML device.
1. Table of Acceptance Criteria and Reported Device Performance
The document states:
- "All testing met the existing predetermined acceptance criteria."
- "Based on the performance testing and the technological characteristics, the FloTrac sensors, Acumen IQ sensors, and VolumeView sensors meet the established performance criteria and are substantially equivalent to the predicate."
However, the specific quantitative acceptance criteria (e.g., accuracy +/- X mmHg, drift < Y%, etc.) and the reported numerical performance data from these tests are not provided in this summary document. The summary indicates that "Performance verification was performed in accordance with Edwards' current design requirements," and "shelf-life and sterilization validations have been performed to existing specifications." These are standard engineering and manufacturing performance metrics, not AI/ML performance metrics.
| Acceptance Criteria (Quantitative examples are not provided in this document) | Reported Device Performance (General Statement) |
|---|---|
| Implicit Criteria: Performance equivalent to predicate devices despite material changes; Biocompatibility, Pressure Accuracy, Shelf-life, Sterilization. | "All testing met the existing predetermined acceptance criteria." |
| "The differences do not raise any new concerns of safety and effectiveness." | "The provided risk analysis and performance verification testing demonstrates the subject devices are substantially equivalent to the predicates." |
2. Sample Size Used for the Test Set and Data Provenance
The document refers to "Performance verification" and "Biocompatibility testing." These generally involve lab-based testing (e.g., mechanical stress tests, pressure accuracy checks on a test bench, chemical compatibility tests) rather than patient data test sets in the context of an AI/ML algorithm.
- Sample Size for Test Set: Not specified. The nature of the "test set" here refers to physical product samples for engineering verification and manufacturing quality, not a dataset of patient measurements for AI validation.
- Data Provenance: The studies mentioned (biocompatibility, performance verification, shelf-life, sterilization) are likely conducted in a controlled lab/manufacturing environment, not directly on patient data. Therefore, questions of "country of origin" or "retrospective/prospective" data collection are not applicable in the context of this type of submission.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications
Given this is not an AI/ML submission related to diagnostic imaging or interpretation, the concept of "experts establishing ground truth" in the clinical sense is not applicable. The ground truth for this device's performance would be established through:
- Engineering specifications and testing standards for pressure transducers.
- Biocompatibility standards (e.g., ISO 10993-1).
- Sterilization validation protocols.
- Comparison to the performance of the predicate device.
4. Adjudication Method for the Test Set
Not Applicable. Adjudication methods like "2+1" or "3+1" are used in clinical studies, particularly for diagnostic interpretations where multiple readers assess cases. This submission is about the physical device's material change and its impact on manufacturing and engineering performance, not clinical interpretation.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
No. An MRMC study is designed to assess the impact of an AI algorithm on human reader performance for tasks like diagnosis or detection. Since this submission is for material changes to a physical sensor, and there is no AI component described, an MRMC study was not performed and is not relevant.
6. If a Standalone (i.e., Algorithm Only Without Human-in-the-Loop Performance) Was Done
No. This question pertains to AI/ML algorithms. The device described (pressure sensor) does not have a standalone AI algorithm that would produce a diagnostic output. Its function is to convert a physiological signal to an electrical signal for a monitor.
7. The Type of Ground Truth Used
The "ground truth" for this submission is based on:
- Engineering specifications and validated testing methods: For pressure accuracy, electrical signal output fidelity, and physical integrity.
- International standards (e.g., ISO 10993-1): For biocompatibility.
- Defined shelf-life and sterilization parameters: Within established industry and regulatory norms.
- Comparison to the performance characteristics of the legally marketed predicate devices.
This is not ground truth derived from expert consensus, pathology, or outcomes data in the clinical sense, as it's not a diagnostic AI device.
8. The Sample Size for the Training Set
Not Applicable. Training sets are relevant for machine learning algorithms. This submission does not describe an AI/ML component or the training of such a component.
9. How the Ground Truth for the Training Set Was Established
Not Applicable. As there's no mention of an AI/ML training set, the establishment of ground truth for it is irrelevant to this 510(k) summary.
In summary: This FDA 510(k) summary provides a high-level overview of a submission for material changes to existing medical devices (intravascular pressure sensors). The document focuses on demonstrating substantial equivalence by confirming that these material changes do not negatively impact the device's established performance, biocompatibility, shelf-life, or sterilization. It does not contain the detailed information typically associated with the validation of an AI/ML diagnostic algorithm, such as specific performance metrics, test set sizes for clinical data, expert panel ground truth adjudication, or MRMC study results.
Ask a specific question about this device
(143 days)
Regulation Name | Swan-Ganz catheters: 21 CFR §870.1240, 21 CFR §870.1200FloTrac sensors: 21 CFR §870.2850
Swan-Ganz catheters:
The Swan-Ganz catheters are diagnostic and monitoring tools used for hemodynamic monitoring of adult critically ill patients including but limited to post major surgical recovery, trauma, sepsis, burns, pulmonary disease, pulmonary failure, cardiac disease including heart failure.
Models 096F6, 096F6P, TS105F5, 132F5, 131F7, 131F7P 831F75, 831F75P, 834F75, and 834F75P are intended for adult and pediatric patients:
The Swan-Ganz catheters are diagnostic and monitoring tools used for hemodynamic monitoring of critically ill adult and pediatric patients ≥ 12 years of age including but not limited to post major surgical recovery, trauma, sepsis, burns, pulmonary disease, pulmonary failure, cardiac disease including heart failure.
FloTrac sensors:
The FloTrac sensor is indicated for use in intravascular pressure monitoring. It is also indicated for use with the Edwards arterial pressure based cardiac output monitoring devices or hardware to measure cardiac output. They are intended to be used in adult patients.
Models MHD6, MHD8, MHD65, and MHD85 are intended for adult and pediatric patients:
The FloTrac sensor is indicated for use in intravascular pressure monitoring. It is also indicated for use with the Edwards arterial pressure based cardiac output monitoring devices or hardware to measure cardiac output. The FloTrac sensor is indicated for use in adult and pediatric patients ≥ 12 years of age.
ClearSight finger cuffs:
The Acumen IQ are indicated for patients over 18 years of age to non-invasively measure blood pressure and associated hemodynamic parameters when used with EV1000 clinical platform or HemoSphere Advanced Monitoring Platform.
The ClearSight finger cuffs are indicated for adult and pediatric patients ≥ 12 years of age to noninvasively measure blood pressure and associated hemodynamic parameters when used with EV1000 clinical platform or HemoSphere Advanced Monitoring Platform.
HemoSphere Advanced Monitor with HemoSphere Swan-Ganz Module:
The HemoSphere advanced monitor when used with the HemoSphere Swan-Ganz module and Edwards Swan-Ganz catheters is indicated for use in adult and pediatric critical care patients requiring monitoring of cardiac output (continuous [CO] and intermittent [iCO]) and derived hemodynamic parameters in a hospital environment. It may be used for monitoring hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. Refer to the Edwards Swan-Ganz catheter indications for use statement for information on target patient population specific to the catheter being used.
Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.
HemoSphere Advanced Monitor with HemoSphere Pressure Cable (compatible with FloTrac sensors):
The HemoSphere advanced monitor when used with the HemoSphere pressure cable is indicated for use in adult and pediatric critical care patients in which the balance between cardiac function, fluid status, vascular resistance and pressure needs continuous assessment. It may be used for monitoring of hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. Refer to the Edwards FloTrac sensor, Acumen IQ sensor, and TruWave DPT indications for use statements for information on target patient populations specific to the sensor/transducer being used.
The Edwards Acumen Hypotension Prediction Index feature provides the clinician with physiological insight into a patient's likelihood of future hypotensive events (defined as mean arterial pressure < 65 mmHg for at least one minute in duration) and the associated hemodynamics. The Acumen HPI™ feature is intended for use in surgical or non-surgical patients receiving advanced hemodynamic monitoring. The Acumen HPI™ feature is considered to be additional quantitative information regarding the patient's physiological condition for reference only and no therapeutic decisions should be made based solely on the Acumen Hypotension Prediction Index (HPI) parameter.
Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.
HemoSphere Advanced Monitor with ClearSight Module:
The HemoSphere advanced monitor when used with the HemoSphere ClearSight module, pressure controller and a compatible Edwards finger cuff are indicated for adult and pediatric patients in which the balance between cardiac function, fluid status and vascular resistance needs continuous assessment. It may be used for monitoring hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. In addition, the noninvasive system is indicated for use in patients with co-morbidities for which hemodynamic optimization is desired and invasive measurements are difficult. The HemoSphere advanced monitor and compatible Edwards finger cuffs noninvasively measures blood pressure and associated hemodynamic parameters.
Refer to the ClearSight finger cuff and Acumen IQ finger cuff indications for use statements for information on target patient population specific to the finger cuff being used.
The Edwards Acumen Hypotension Prediction Index (HPI) feature provides the clinician with physiological insight into patient's likelihood of future hypotensive events (defined as mean arterial pressure < 65 mmHg for at least one minute in duration) and the associated hemodynamics. The Acumen HPI feature is intended for use in surgical patients receiving advanced hemodynamic monitoring. The Acumen HPI feature is considered to be additional quantitative information regarding the patient's physiological condition for reference only and no therapeutic decisions should be made based solely on the Acumen Hypotension Prediction Index (HPI) parameter.
Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.
Swan-Ganz Catheters:
The Swan-Ganz catheters are flow-directed pulmonary artery catheters used to monitor hemodynamic pressures. The Swan-Ganz thermodilution catheters provide diagnostic information to rapidly determine hemodynamic pressures and cardiac output when used with a compatible cardiac output computer.
FloTrac Sensors:
The FloTrac sensor is a sterile, single use kit that monitors pressures when attached to pressure monitoring catheters. When connected to a compatible monitor, the FloTrac sensor minimally-invasively measures cardiac output and key hemodynamic parameters, which assist the clinician in assessing the patient's physiologic status and support clinical decisions related to hemodynamic optimization. The disposable sterile cable, with a red-connector interfaces, exclusively with an Edwards cable that is specifically wired for the pressure monitor being used. The disposable sterile cable, with a green-connector interfaces, exclusively with the Edwards cables for use with the Edwards arterial pressure based cardiac output monitoring devices or hardware.
The FloTrac sensor has a straight, flow-through design across the pressure sensors with an integral flush device.
ClearSight finger cuffs:
The ClearSight finger cuffs, when used with an appropriate Edwards monitoring system, provide continuous, noninvasive hemodynamic monitoring. The ClearSight finger cuffs utilize the volume-clamp method to measure blood pressure with an inflatable bladder wrapped around the middle phalanx of the finger.
HemoSphere Advanced Monitor Platform:
The HemoSphere Advanced Monitoring Platform was designed to simplify the customer experience by providing one platform with modular solutions for their hemodynamic monitoring needs. The user can choose from the available optional sub-system modules or use multiple sub-system modules at the same time. This modular approach provides the customer with the choice of purchasing and/or using specific monitoring applications based on their needs. Users are not required to have all of the modules installed at the same time for the platform to function.
This document is a 510(k) summary for Edwards Lifesciences' Swan-Ganz catheters, FloTrac sensors, ClearSight finger cuffs, and HemoSphere Advanced Monitoring Platform. The core of this submission is to expand the indications for use of these devices to include pediatric patients (≥12 years of age), in addition to their existing adult indications.
The information regarding acceptance criteria and the study that proves the device meets those criteria, as detailed in your request, is primarily found in the "Device Verification and Validation" and "Conclusion" sections. However, the FDA summary document only provides a high-level overview. It does not contain the detailed acceptance criteria table, precise performance metrics, sample sizes, ground truth establishment methods, or specific details about expert adjudication or MRMC studies, which are typically found in the full 510(k) submission or supporting clinical trial reports.
Based on the provided text, here's what can be extracted and what information is not present:
Key Takeaway from the FDA 510(k) Summary:
The primary purpose of this 510(k) submission is to expand the indicated patient population for the listed devices from adult-only to include pediatric patients aged 12 years and older. The submission claims that the devices perform similarly in this expanded pediatric population compared to adults, meeting predefined acceptance criteria.
Table of Acceptance Criteria and Reported Device Performance
Not Provided in this Document. The document states "within predefined acceptance criteria" but does not list these criteria or the specific reported device performance metrics against them.
| Acceptance Criteria Category | Specific Metric (e.g., Accuracy, Precision, AUC Threshold) | Acceptance Threshold | Reported Device Performance |
|---|---|---|---|
| Hemodynamic Parameter Measurement | Not specified | Not specified | Not specified |
| Safety | Not specified | Not specified | "no issues of safety" |
| Effectiveness | Not specified | Not specified | "effective in measuring cardiac output" |
| Consistency | Not specified | Not specified | "consistent and within predefined acceptance criteria." "similar performance to an adult population." |
Study Details (Based on available information):
-
Sample sizes used for the test set and the data provenance:
- Sample Size: Not explicitly stated. The document mentions "the target pediatric population in a clinical study" but does not provide the number of patients.
- Data Provenance:
- Country of Origin: Not specified in this summary. Clinical studies are often multi-center, but no specific locations are mentioned.
- Retrospective or Prospective: Not explicitly stated, but the phrase "clinical study" typically implies prospective data collection for regulatory submissions, especially for expanding indications to a new population.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Number of Experts: Not specified.
- Qualifications of Experts: Not specified. Given the nature of these hemodynamic devices, ground truth would likely be established by clinical measurements and potentially other validated reference methods, rather than expert interpretation of images or signals in the same way a radiologist might interpret an X-ray. Clinical experts (e.g., intensivists, anesthesiologists, cardiologists) would be involved in patient selection, data collection, and overseeing the study.
-
Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- Adjudication Method: Not specified. This type of adjudication (e.g., multi-reader consensus) is more common in image-based diagnostic AI studies. For device performance measurement, agreement with a reference standard is the typical method.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- MRMC Study: Not applicable/not performed for this type of device and study. MRMC studies are primarily for evaluating AI's impact on human interpretation tasks (e.g., radiologists reading images). This submission is for devices that directly measure physiological parameters and provide data, with one feature (Acumen HPI) providing "physiological insight" for reference. The study focuses on the device's performance in a new patient population, not on how it assists human readers in interpreting complex cases, except perhaps by providing accessible data. The Acumen HPI feature specifically states "no therapeutic decisions should be made based solely on the Acumen Hypotension Prediction Index (HPI) parameter," indicating it's an informational tool rather than a definitive diagnostic AI assist requiring a MRMC study.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Standalone Performance: The study evaluated the devices' performance in the pediatric population. The devices themselves are "standalone" in that they acquire and process physiological signals to output parameters. The statement "The data demonstrate the performance in the expanded pediatric population ≥ 12 years of age and is substantially equivalent to their respective predicate devices for adult patient population ≥ 18 years of age" suggests an evaluation of the device's measurement accuracy and consistency.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc):
- Type of Ground Truth: Not explicitly stated, but for hemodynamic monitoring devices, ground truth is typically established by:
- Reference Standards/Comparative Measurements: Comparing the device's measurements (e.g., cardiac output, blood pressure) against established, precise, and often more invasive or gold-standard methods (e.g., thermodilution, invasive arterial line measurements from validated transducers).
- Clinical Outcomes/Safety Data: Demonstrating safety and effectiveness in the target population through clinical monitoring and observation.
- The document states "clinical data presented in this 510(k) demonstrates the performance of hemodynamic parameters expanded to pediatric patients ≥ 12 years of age... are consistent and within predefined acceptance criteria," implying comparisons to a reference or expected range.
- Type of Ground Truth: Not explicitly stated, but for hemodynamic monitoring devices, ground truth is typically established by:
-
The sample size for the training set:
- Training Set Size: Not applicable/not specified. These devices measure physiological parameters; they are not "AI algorithms" in the sense of needing a large training dataset to learn patterns for classification or prediction (except for the Acumen HPI feature, but no details on its development or training are provided here). The focus of this 510(k) is the application of existing technology to a new age group, not the development of a new pattern-recognition algorithm.
-
How the ground truth for the training set was established:
- Ground Truth for Training Set: Not applicable, as there's no mention of a traditional AI training set. The devices likely use established physiological models and algorithms for calculating parameters. If the Acumen HPI feature involved machine learning, its training ground truth would likely be based on recorded physiological data and subsequent hypotensive events. However, the document does not elaborate on the development of this specific feature in the context of this 510(k).
Ask a specific question about this device
(285 days)
§870.2850, Transducer,BloodPressure, Extravascular(Product Code DRS)
§870.2850, Transducer,BloodPressure,Extravascular(Product Code DRS)(Secondary) 21
The EpiFinder is intended for use, in between a luer Loss of resistance (LOR) syringe and an epidural needle, to verify the needle tip placement in the epidural space as a secondary indicator adjunctive to the LOR technique.
The EpiFinder is intended to be used in conjunction with a standard epidural LOR technique as an independent secondary confirmatory indicator, giving the practitioner a clear visual signal that the needle tip has entered the epidural space. The EpiFinder is a sterile, single-use device that is used during epidural procedures. The device works with a standard loss of resistance (LOR) syringe and an epidural needle (18G) to verify that the needle tip placement is in the epidural space in patients ages 18 or older.
The EpiFinder consists of a blunt tip spring-loaded probe that is installed into a standard epidural needle, and a plastic housing that is composed of a linear actuator, a sensor, and a microcontroller that measures the probe spring's contraction. The EpiFinder is attached to a Tuohy needle that has been inserted into the patient. The device has built in "wings" to help advance the needle, if the practitioner wishes to use them. As the needle advances forward, there are LED indicator lights on the housing that give feedback to the operator as to when the epidural space is entered, as input to the user in conjunction with the standard LOR technique.
Here's an analysis of the acceptance criteria and study information for the EpiFinder device, based on the provided text:
Important Note: The provided text does not explicitly list acceptance criteria in a quantitative table format with pass/fail metrics. Instead, it describes various performance evaluations and concludes that the device's performance was "appropriate," "effective," and "similar to standard practice." Therefore, the table below will summarize the reported device performance based on the described studies, indicating what was evaluated and the general outcome.
Acceptance Criteria and Reported Device Performance
| Acceptance Criteria Category (Implied) | Reported Device Performance |
|---|---|
| Biocompatibility | Conformed to ISO 10993 for externally communicating devices with limited contact. All results (cytotoxicity, sensitization, intracutaneous irritation, acute systemic toxicity, material-mediated pyrogenicity) were acceptable. |
| Benchtop Performance | Demonstrated LOR channel sealing integrity (per ISO 80369-7), Luer dimensional and functional attributes, acceptable probe penetration force, probe integrity, sensing mechanism function, wings mechanism function, acceptable particulate matter, corrosion resistance, device integrity, and consistent operation. |
| Electrical Safety & EMC | Conformed to IEC 60601 and other applicable standards for electrical safety and electromagnetic compatibility. |
| Software Validation | Software verification and validation testing was conducted per IEC 62304:2006+AMD1:2015 and FDA's 2005 guidance. Results were found acceptable for software release. |
| Safety and Effectiveness (Animal) | Evaluated in a pig study (GLP compliant). Device was found appropriate for its intended use (only one adverse event) and effective in providing a clear visual signal that the epidural needle entered the epidural space. |
| Human Factors/Usability | Evaluated per IEC 62366-1:2015 and FDA's 2016 guidance. No use-related hazards recorded. Three participants in each group experienced use problems, but these were deemed not to pose any additional risk. Users reported high satisfaction. |
| Safety and Performance (Clinical) | Evaluated in a single-arm, open-label multicenter trial. Device was found appropriate for use in human subjects. Performance was established, demonstrating its capability to correctly identify the epidural space. Utilization of LOR technique with EpiFinder was similar to standard practice. |
Study Details
-
Sample sizes used for the test set and the data provenance:
- Animal Study (Pigs): "Numerous test sites" (specific number not given) were randomly divided into two groups: Group 1 (EpiFinder + standard LOR technique) and Group 2 (standard LOR technique alone). The study was conducted in compliance with OECD principles of Good Laboratory Practice (GLP) ENV/MC/CHEM (98)17 (except for blood analysis).
- Human Factors Evaluation: 16 anesthesiologists (physicians) and 10 student resident nurse anesthetists (sRNAs).
- Clinical Evaluation: 31 adult subjects.
- Data Provenance: The animal study was noted as GLP compliant, implying a controlled, prospective design. The human factors and clinical evaluations would also be prospective studies. The country of origin for the animal and clinical studies is not explicitly stated, but the company address is in Israel.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Animal Study: Ground truth was confirmed using X-ray and contrast medium. No specific number of experts or their qualifications are mentioned for interpreting these confirmations.
- Human Factors: The "experts" were the participating 16 anesthesiologists (physicians) and 10 student resident nurse anesthetists (sRNAs) who evaluated usability. Their feedback contributed to assessing safety and usability.
- Clinical Evaluation: Ground truth for epidural space identification during the clinical trial is implied to be clinical assessment by the performing physicians, potentially combined with the standard LOR technique used concurrently. The text mentions "The performance of the device was also established, demonstrating that the device is capable of correctly identifying the epidural space during the injection procedure." No specific number of independent experts or their detailed qualifications for establishing ground truth are explicitly stated beyond the clinical team performing the procedures.
-
Adjudication method for the test set:
- The document does not describe a formal adjudication method (like 2+1 or 3+1 consensus) for establishing ground truth in any of the studies.
- In the animal study, confirmation was via X-ray and contrast medium, which typically involves interpretation by a radiologist or veterinary specialist, but a specific adjudication process isn't detailed.
- In the clinical evaluation, the "correctly identifying the epidural space" implies the clinical judgment of the performing physician was the primary determinant, likely corroborated by the LOR technique.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No MRMC comparative effectiveness study is described where human readers' performance with and without AI assistance is compared. The EpiFinder is a physical device with an indicator (LEDs) and functions adjunctively to the standard LOR technique, rather than being an AI diagnostic tool primarily interpreted by human readers.
-
If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- The EpiFinder is designed to be used "in conjunction with a standard epidural LOR technique as an independent secondary confirmatory indicator." It provides a "clear visual signal" via LEDs. Therefore, its performance is inherently "human-in-the-loop" as it provides feedback to the operator. The animal study did evaluate the device's signal in conjunction with the standard LOR technique, and its effectiveness was confirmed against X-ray and contrast.
-
The type of ground truth used:
- Animal Study: Radiographic confirmation (X-ray and contrast medium).
- Clinical Evaluation: Clinical identification of the epidural space by performing physicians, presumably leveraging the concurrent standard LOR technique. The device's ability to "correctly identify" the space refers to its agreement with this clinical assessment.
-
The sample size for the training set:
- The document describes performance evaluation studies (validation studies) for the EpiFinder. It does not provide information about a "training set" as this device is a hardware product with a pre-programmed microcontroller/firmware, not a machine learning (AI) algorithm that undergoes a training phase with a distinct dataset. Software verification and validation were performed, but this typically involves testing inputs/outputs against specifications, not learning from a dataset.
-
How the ground truth for the training set was established:
- Not applicable, as there is no mention of a "training set" in the context of an AI/machine learning model. The software is described as "pre-programmed software (firmware)," implying it's deterministic based on its design specifications, not trained on data.
Ask a specific question about this device
(269 days)
Meritrans ECO Reusable Pressure Transducer and Meritrans ECO Disposable Domes Regulation Number: 21 CFR 870.2850
| 870.2850
| 21 CFR § 870.2850
standards have been established for Transducer, Blood-Pressure,Extravascular under FDA 21 CFR § 870.2850
The Meritrans ECO Reusable Pressure Transducer is intended to be coupled with a disposable dome kit for converting hemodynamic pressure waveforms via a piezoresistive bridge circuit into electrical signals for invasive blood pressure monitoring when connected to a patient monitor or recording device.
The devices are intended for pediatric and adult patient populations.
The Meritrans ECO Reusable Pressure Transducer is a non-invasive, reusable device used primarily for blood pressure monitoring. The housing assembly interfaces the pressure sensor assembly to the fluid channel of patient by coupling the transducer to the Meritrans ECO Disposable Dome. The fluid channel of patient is isolated from the pressure sensor by diaphragms. As the pressure signal of patient carried via the saline solution presses against the surface of transducer, the diaphragm is deflected, and the physiological pressure signal is transformed from the dome to the sensor of the reusable pressure transducer. Subsequently, the physiological pressure signal is transformed into electrical signal and displayed in a monitor that is connected via transducer interface cable.
The Meritrans ECO Disposable Domes are disposable diaphragm domes which are coupled with the reusable transducer for invasive blood pressure monitoring. The disposable diaphragm dome has a transparent hemispherical hollow cavity. The domes have two ports; one port of the dome has a rotating Luer Lock fitting where the patient line is connected. Another port has an adaptor for connection to a pressurized IV fluid bag or an infusion pump. The disposable diaphragm dome has a diaphragm membrane that isolates the patient's fluid-filled system from the reusable transducer. For dome models with flush devices, an integral flow/flush device provides continuous flow of fluid to maintain catheter patency during the invasive blood pressure measurements.
The provided text describes the acceptance criteria and the studies performed to demonstrate the device's conformance, but it does not detail a study proving the device meets the acceptance criteria in the format typically used for AI/ML performance evaluation (e.g., specific metrics for accuracy, sensitivity, specificity, clinical outcome studies). Instead, it focuses on compliance with various engineering, safety, and biological standards through a battery of tests.
Here's an analysis of the provided information:
1. A table of acceptance criteria and the reported device performance
The document lists various standards that the device complies with, implying that meeting the requirements of these standards constitutes the acceptance criteria. The reported device performance is stated as "Results of the testing demonstrate that the subject devices met the acceptance criteria sufficient for their intended use." However, specific numerical performance metrics for each criterion against a defined acceptable range are not provided in this summary.
Generic Table based on the document's structure for Performance Bench Testing:
| Category / Standard | Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|---|
| Product Performances | ||
| ANSI/AAMI BP22:1994 (R)2016 (Blood pressure transducers) | Compliance with all requirements of the standard | Met the acceptance criteria sufficient for intended use |
| IEC 60601-1:2020 (Edition 3.2) (General safety and essential performance) | Compliance with all requirements of the standard | Met the acceptance criteria sufficient for intended use |
| IEC 60601-1-2:2020 (Edition 4.1) (Electromagnetic disturbances) | Compliance with all requirements of the standard | Met the acceptance criteria sufficient for intended use |
| IEC 60601-2-34:2011 (Edition 3.0) (Invasive blood pressure monitoring equipment) | Compliance with all requirements of the standard | Met the acceptance criteria sufficient for intended use |
| IEC 62366-1:2015 + Cor 1:2016 + Amd 1:2020 (Usability engineering) | Compliance with all requirements of the standard | Met the acceptance criteria sufficient for intended use |
| ISO 80369-7:2021 (Small-bore connectors for intravascular/hypodermic applications) | Compliance with all requirements of the standard | Met the acceptance criteria sufficient for intended use |
| FDA Guidance, Information to Support a Claim of Electromagnetic Compatibility (EMC) of Electrically-Powered Medical Devices | Compliance with guidance recommendations | Met the acceptance criteria sufficient for intended use |
| Packaging and Transportation | ||
| EN ISO 2233:2001 (Packaging - Conditioning for testing) | Compliance with all requirements of the standard | Met the acceptance criteria sufficient for intended use |
| ASTM D4169 - 16 (2016) (Performance Testing of Shipping Containers) | Compliance with all requirements of the standard | Met the acceptance criteria sufficient for intended use |
| EN ISO 11607-1:2020 (Packaging for terminally sterilized medical devices) | Compliance with all requirements of the standard | Met the acceptance criteria sufficient for intended use |
| ASTM F1980 16 (2016) (Accelerated Aging of Sterile Barrier Systems) | Compliance with all requirements of the standard | Met the acceptance criteria sufficient for intended use |
| Biocompatibility (Meritrans ECO Disposable Domes) | ||
| ISO 10993-1:2018 (Biological evaluation - Risk management) | Compliance with all requirements of the standard and FDA guidance | Met the acceptance criteria sufficient for intended use (Cytotoxicity, Sensitization, Irritation, Acute Systemic Toxicity, Pyrogenicity, Hemocompatibility) |
| ISO 10993-4:2017 (Interactions with blood) | Compliance with all requirements of the standard | Met the acceptance criteria sufficient for intended use |
| ISO 10993-5:2009 (In vitro cytotoxicity) | Compliance with all requirements of the standard | Met the acceptance criteria sufficient for intended use |
| ISO 10993-10:2021 (Skin sensitization) | Compliance with all requirements of the standard | Met the acceptance criteria sufficient for intended use |
| ISO 10993-11:2017 (Systemic toxicity) | Compliance with all requirements of the standard | Met the acceptance criteria sufficient for intended use |
| ISO 10993-12:2021 (Sample preparation and reference materials) | Compliance with all requirements of the standard | Met the acceptance criteria sufficient for intended use |
| ISO 10993-23:2021 (Irritation) | Compliance with all requirements of the standard | Met the acceptance criteria sufficient for intended use |
| FDA Guidance, Use of International Standard ISO 10993-1 | Compliance with guidance recommendations | Met the acceptance criteria sufficient for intended use |
| Sterilization / Sterility / EO residuals / Pyrogenicity (Meritrans ECO Disposable Domes) | ||
| EN ISO 11135:2014 (+ Amd 1:2018) (Ethylene oxide sterilization) | Compliance with all requirements of the standard | Met the acceptance criteria sufficient for intended use |
| EN ISO 10993-7:2008 (+ Cor 1:2009 + AMD 1:2019) (Ethylene Oxide sterilization residuals) | Compliance with all requirements of the standard | Met the acceptance criteria sufficient for intended use |
| AAMI TIR28:2016 (Product Adoption and process equivalency for ethylene oxide sterilization) | Compliance with all requirements of the standard | Met the acceptance criteria sufficient for intended use |
| ANSI/AAMI ST72:2019 (Bacterial endotoxins Test methods) | Compliance with all requirements of the standard | Met the acceptance criteria sufficient for intended use |
| United States Pharmacopeia 43, National Formulary 38:2020 (USP), General Chapter <151> Pyrogen Test | Compliance with all requirements of the standard | Met the acceptance criteria sufficient for intended use |
| FDA Guidance. Submission and Review of Sterility Information in Premarket Notification (510(k)) Submissions for Devices Labeled as Sterile | Compliance with guidance recommendations | Met the acceptance criteria sufficient for intended use |
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document lists various standards and states that "A battery of tests was performed based on the requirements of the below FDA recognized / non-recognized consensus standards and guidance, as well as biocompatibility, sterilization, and labeling standards and guidance." However, it does not specify the sample sizes used for these tests or the data provenance (country of origin, retrospective/prospective). This information would typically be detailed in the test reports themselves, which are not included in this summary.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This information is not applicable in the context of this 510(k) summary. The device in question, a pressure transducer and disposable domes, is a hardware medical device with electrical and mechanical performance, biocompatibility, and sterility requirements. Its evaluation is based on engineering testing and biological assays against recognized standards, not on image interpretation or diagnostic performance with expert-established ground truth. Therefore, there are no "experts" in the sense of clinical reviewers establishing ground truth for a test set.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This is not applicable for the reasons stated above. Adjudication methods are relevant for studies involving human interpretation (e.g., radiologists, pathologists) where discrepancies need to be resolved to establish ground truth, typically in AI/ML performance evaluations.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This is not applicable. The device is an extravascular blood pressure transducer, a piece of hardware for physiological monitoring. It does not involve AI assistance for human readers in any diagnostic or interpretive task.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This is not applicable. The device is a hardware product, not a software algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
For this device, the "ground truth" is established by the specified requirements and test methods outlined in the recognized consensus standards (e.g., ANSI/AAMI BP22 for blood pressure transducers, ISO 10993 for biocompatibility). These standards define acceptable performance metrics, physical properties, and biological responses. The device's performance is measured against these established technical and biological benchmarks.
8. The sample size for the training set
This is not applicable. The device is a hardware medical device, and the evaluation described is for its physical and biological performance against standards, not for a trainable AI/ML algorithm that requires a "training set."
9. How the ground truth for the training set was established
This is not applicable for the reasons stated above. There is no AI/ML training set mentioned or implied for this device's evaluation.
Ask a specific question about this device
(218 days)
| Alarm, blood-pressure |
| 870.2850
The BeneVision N12N15/N17/N19/N22 patient monitors are intended for monitoring, displaying, storing, alarming, and transferring of multiple physiological parameters including ECG (3-lead, 5-lead or 12-lead selectable, Arrhythmia Detection, ST Segment Analysis, QT Analysis, and Heart Rate (HR)), Respiration Rate (Resp), Temperature (Temp), Pulse Oxygen Saturation (SpO2), Pulse Rate (PR), Non-invasive Blood Pressure (NIBP), Invasive Blood Pressure (IBP), Pulmonary Artery Wedge Pressure (PAWP), Cardiac Output (C.O.), Continuous Cardiac Output (CCO), Mixed/Central Venous Oxygen Saturation (SvO2/ScvO2), Carbon Dioxide (CO2), Oxygen (O2), Anesthetic Gas (AG), Impedance Cardiograph (ICG), Bispectral Index (BIS), Respiration Mechanics (RM), Neuromuscular Transmission Monitoring (NMT), Electroencephalograph (EEG), and Regional Oxygen Saturation (rSO2). The system also provides an interpretation of resting 12-lead ECG.
All the parameters can be monitored on single adult, pediatric, and neonatal patients except for the following:
- · BIS, RM, CCO, SvO2/ScvO2, PAWP, NMT monitoring, PNP, and PNC are intended for adult and pediatric patients only. CCO using FloTrac is intended for adult patients only;
- · C.O. monitoring and A-Fib are intended for adult patients only;
- · ICG monitoring is intended for only adult patients who meet the following requirements: height: 122 to 229cm, weight: 30 to 155kg.
- · rSO2 monitoring is intended for use in individuals greater than 2.5kg.
The monitors are to be used in healthcare facilities by clinical professionals or under their guidance. They should only be used by persons who have received adequate training in their use. The BeneVision N12/N15/N17/N19/N22 monitors are not intended for helicopter transport, hospital ambulance, or home use.
The BeneVision N1 Patient Monitor is intended for monitoring, reviewing, storing, alarming, and transferring of multiple physiological parameters including ECG (3-lead, 5-lead or 12-lead selectable, Arrhythmia Detection, ST Segment Analysis, and Heart Rate (HR)), Respiration (Resp), Temperature (Temp), Pulse Oxygen Saturation (SpO₂), Pulse Rate (PR), Non-invasive Blood Pressure (NIBP), Invasive Blood Pressure (IBP), Pulmonary Artery Wedge Pressure (PAWP), Carbon Dioxide (CO2) and Oxygen (O2). The system also provides an interpretation of resting 12-lead ECG.
All the parameters can be monitored on single adult, pediatric, and neonatal patients except for the following:
- PAWP, PNP, and PNC are intended for adult and pediatric patients only;
- A-Fib is intended for adult patients only;
The BeneVision N1 monitor is to be used in healthcare facilities. It can also be used during patient transport inside and outside of the hospital environment, whereas N1 configured with WMTS technology can be used inside the hospital only. It should be used by clinical professionals or under their guidance. It should only be used by persons who have received adequate training in its use. It is not intended for home use.
The subject BeneVision N Series Patient Monitors includes six monitors:
- . BeneVision N12 Patient Monitor
- BeneVision N15 Patient Monitor
- BeneVision N17 Patient Monitor ●
- BeneVision N19 Patient Monitor ●
- BeneVision N22 Patient Monitor ●
- BeneVision N1 Patient Monitor
Mindray's BeneVision N Series Patient Monitors provide a flexible software and hardware platform to meet the clinical needs of patient monitoring.
This document is a 510(k) Summary for the Mindray BeneVision N Series Patient Monitors, which focuses on demonstrating substantial equivalence to a previously cleared predicate device (K202405).
The information provided primarily details the device's technical specifications and comparisons to a predicate device, rather than a full study proving the device meets acceptance criteria for a specific medical condition or AI diagnostic output.
Therefore, I cannot fully answer all parts of your request as the document does not contain the detailed clinical study results (like sample sizes for test sets, number of experts for ground truth, adjudication methods, MRMC studies, or specific AI performance metrics) that would typically be found for a device requiring those types of studies (e.g., an AI-powered diagnostic tool).
However, I can extract the relevant information regarding the device's functional and technical performance as demonstrated in this 510(k) submission.
Here's a breakdown of what can be inferred and what is missing:
Acceptance Criteria and Reported Device Performance
The "acceptance criteria" in this context are related to meeting the performance specifications of the predicate device and relevant consensus standards. The "reported device performance" is demonstrated through functional and system-level testing, ensuring the device meets its accuracy specifications for the various physiological parameters it monitors.
Table of "Acceptance Criteria" (Implied Specifications) and "Reported Device Performance" (Conformance):
| Parameter / Feature | Implied Acceptance Criteria (from Predicate/Standards) | Reported Device Performance (as stated in document) |
|---|---|---|
| General Device Performance | Reliability, accuracy, and safety equivalent to the predicate device (K202405). Compliance with general controls and specific standards. | "Mindray conducted functional and system level testing on the subject device. The testing provided an evaluation of the performance of the device relevant to each of the differences between the subject device and the predicate device. The functional and system level testing showed that the devices continue to meet specifications and the performance of the device is equivalent to the predicate." "The results of the bench testing show that the subject device meets its accuracy specification and is substantially equivalent to the predicate device." |
| ECG (HR) | HR Measurement range: 15 | No specific deviation reported from these specifications. The device supports intelligent arrhythmia alarms, adjustment of QT calculation, SVT and SVCs/min high arrhythmia alarm, and Multi-lead ECG synchronization analysis. |
| ECG (ST) | ST Measurement range: -2.0mV~+2.0mV; Accuracy: -0.8mV~+0.8mV, ±0.02mV or ±10%, whichever is greater, other range: not specified. | No specific deviation reported from these specifications. |
| ECG (QT) | QT Measurement range: 200~800ms; Accuracy: ±30ms. | No specific deviation reported from these specifications. Adjustment of QT calculation is a new feature. |
| Respiration Rate (Resp) | Measurement range: Adult: 0 to 120 rpm; Pediatric, neonate: 0 to 150 rpm. Accuracy: 7 to 150 rpm: ±2 rpm or ±2%, whichever is greater; 0 to 6 rpm: Not specified. | No specific deviation reported from these specifications. |
| Temperature (Temp) | Measurement range: 0 to 50°C (32 to 122°F). Accuracy: ±0.1°C or ±0.2°F (without probe). | No specific deviation reported from these specifications. |
| Pulse Oxygen Saturation (SpO2) | Mindray SpO2 module: Range: 0~100% Accuracy: 70% | No specific deviation reported from these specifications. |
| Pulse Rate (PR) | Accuracy: ±3 bpm (Mindray SpO2), ±3 bpm without motion, ±5 bpm with motion (Masimo SpO2); 20 | No specific deviation reported from these specifications. |
| Non-invasive Blood Pressure (NIBP) | Max mean error: ±5mmHg; Max standard deviation: 8mmHg. PR: ±3 bpm or ±3%, whichever is greater. | No specific deviation reported from these specifications. |
| Invasive Blood Pressure (IBP) | Accuracy of module: ±2% or ±1mmHg, whichever is greater (without sensor). | No specific deviation reported from these specifications. New feature: Artifact flag of Arterial Blood Pressure (ABP) shields alarms monitoring support. BeneVision N1 now supports up to 4 IBP channels (vs. 2 in predicate). |
| Cardiac Output (C.O.) | Accuracy: ±5% or ±0.1L/min, whichever is greater. TB, TI: ±0.1°C (without sensor). | No specific deviation reported from these specifications. |
| Continuous Cardiac Output (FloTrac) | (Not supported in predicate) Measurement range: CCO: 1.0-20.0L/min; Reproducibility: ±6% or 0.1 L/min, whichever is greater. PR:0~220bpm, Arms ≤3bpm. Live pressure display range: -34 to 312 mmHg. MAP/DIA/SYS display range: 0-300 mmHg. Accuracy: ±4% or ±4 mmHg, whichever is greater, in the range of -30 mmHg to 300 mmHg. | Added feature with stated performance specifications as listed. (This is a new feature compared to the predicate, and its performance data is listed as its "acceptance criteria" and "reported performance" upon introduction.) |
| Carbon Dioxide (CO2) | Sidestream: CO2: 0 | No specific deviation reported from these specifications. For BeneVision N1: component changes for Sidestream CO2 1.0 and 2.0 modules. |
| Oxygen (O2) | Accuracy: 0 | No specific deviation reported from these specifications. |
| Anesthetic Gas (AG) | Various specific accuracies for N2O, HAL, ENF, ISO, SEV, DES based on concentration ranges. awRR: 2~60rpm, ±1rpm, >60rpm, not specified. | No specific deviation reported from these specifications. Internal electronic component changes for the AG module. |
| Electroencephalograph (EEG/aEEG) | Frequency response: 0.5Hz ~ 50 Hz (-3 dB). Input range: 4 mVpp. DC offset: ±500 mV. CMRR: ≥ 100 dB @ 51 KΩ and 60 Hz. Noise level: ≤ 0.5 uV rms (1Hz to 30 Hz). Differential input resistance: > 15 MΩ @ 10 Hz. Electrode resistance: 0 to 90 KΩ, resolution: ±1 KΩ or 10%, whichever is the greater. | Added EEG-1 module and aEEG module, with their performance specifications stated as meeting these criteria. |
| A-Fib Overview | (Not supported in predicate) The A-Fib overview function only collects atrial fibrillation, atrial fibrillation with rapid ventricular rate, atrial fibrillation with R-R long interval and other related events, and the A-Fib overview screen displays the A-Fib specifications. | This is a newly added feature with its intended functionality described. |
| Electromagnetic Compatibility & Electrical Safety | Compliance with relevant standards (ANSI/AAMI ES 60601-1:2005, IEC 60601-1-2:2014). | Assessed for conformity and found to comply with ANSI/AAMI ES 60601-1:2005/(R) 2012 and IEC 60601-1-2:2014. Also meets IEC 60601-1-8:2020 for alarm systems. |
| Software Verification & Validation | Meets FDA Guidance for Software Contained in Medical Devices. | "Software verification and validation testing was conducted and documentation was provided as recommended by FDA's Guidance... Verification... was conducted to ensure that the product works as designed. Validation was conducted to check the design and performance of the product." |
Detailed Study Information (Based on Document Content):
-
Sample sizes used for the test set and the data provenance:
- The document states that "functional and system level testing" and "bench testing" were conducted.
- However, no specific sample sizes for test sets (e.g., number of patients, number of data recordings) are provided for any of the performance evaluations.
- Data Provenance: Not explicitly stated (e.g., country of origin, retrospective/prospective). The tests described are generic "bench testing" to ensure compliance with technical specifications and standards, not clinical studies.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable / Not provided. This document describes engineering and bench testing against pre-defined technical specifications and industry standards for physiological measurement accuracy. It does not describe a clinical study involving human experts establishing ground truth for diagnostic interpretation (e.g., for an AI algorithm interpreting medical images).
-
Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- Not applicable / Not provided. Same reason as above.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No. This device is a patient monitor. It detects physiological parameters and provides alarms, and some interpretations of ECG (e.g., 12-lead ECG interpretation, arrhythmia detection). It is not an AI-assisted diagnostic device in the sense of image interpretation for which MRMC studies are typically performed. The document details that "optimized auditory ALARM SIGNALS" and "alarm highlight" were added, suggesting improvements to the human-device interface, but not a formal MRMC study on diagnostic improvement.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- The document implies that algorithms for ECG (Mindray or Mortara algorithm for arrhythmia and ST-segment analysis) are embedded in the device. The listed accuracy specifications for these measurements (e.g., HR, ST, QT) reflect the standalone performance of these measurement algorithms and sensors against established benchmarks. However, a formal "standalone study" with detailed methodology, distinct from the general bench testing, is not specifically described or provided with separate results. The performance data listed (e.g., accuracy for HR, ST, QT) serves as the "standalone" performance verification for these integrated functionalities.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- For physiological measurements (ECG, SpO2, NIBP, etc.), the "ground truth" would typically refer to reference measurement devices or calibrated simulators used during bench testing to verify the accuracy of the monitor's readings against a known, accurate value.
- For the ECG interpretation (e.g., 12-lead ECG interpretation, arrhythmia detection), the ground truth for the algorithms would have been established during their development and previous clearance processes (Mindray or Mortara algorithms). This document focuses on demonstrating that the integration and revised features maintain that established accuracy rather than re-proving the core algorithms.
-
The sample size for the training set:
- Not provided. This document pertains to the 510(k) clearance of updates to an existing patient monitor series. It does not detail the development or training of new AI/ML algorithms, which would typically involve substantial training datasets. The ECG algorithms (Mindray or Mortara) were presumably "trained" (or developed and validated) previously as part of their initial predicate clearances.
-
How the ground truth for the training set was established:
- Not provided. (See point 7). For existing algorithms like Mortara or Mindray ECG algorithms, ground truth for their original development would likely have been established using large, diverse ECG databases with expert cardiologist interpretations and/or correlation with clinical outcomes where relevant. This particular 510(k) document is concerned with demonstrating equivalence and continued performance with minor changes, not the original algorithm development.
Ask a specific question about this device
(139 days)
| Alarm, blood-pressure |
| 870.2850
BeneVision N12/N15/N17/N19/N22
The BeneVision N12N15/N17/N19/N22 patient monitors are intended for monitoring, displaying, storing, alarming, and transferring of multiple physiological parameters including ECG (3-lead, 5-lead or 12-lead selectable, Arrhythmia Detection, ST Segment Analysis, QT Analysis, and Heart Rate (HR)), Respiration Rate (Resp), Temperature (Temp), Pulse Oxygen Saturation (SpO2), Pulse Rate (PR), Non-invasive Blood Pressure (NIBP), Invasive Blood Pressure(IBP), Pulmonary Artery Wedge Pressure (PAWP), Cardiac Output (C.O.), Continuous Cardiac Output (CCO), Mixed/Central Venous Oxygen Saturation (SvO2/ScvO2), Carbon Dioxide (CO2), Oxygen (O2), Anesthetic Gas (AG), Impedance Cardiograph (ICG), Bispectral Index (BIS), Respiration Mechanics (RM), Neuromuscular Transmission Monitoring (NMT), Electroencephalograph (EEG), and Regional Oxygen Saturation (rSO2). The system also provides an interpretation of resting 12-lead ECG.
All the parameters can be monitored on single adult, pediatric, and neonatal patients except for the following:
· BIS, RM, CCO, SvO2/ScvO2, PAWP, and NMT monitoring, PNP, and PNC are intended for adult and pediatric patients only:
· C.O. monitoring and A-Fib are intended for adult patients only;
• ICG monitoring is intended for only adult patients who meet the following requirements: height: 122 to 229cm, weight: 30 to 155kg.
· rSO2 monitoring is intended for use in individuals greater than 2.5kg.
The monitors are to be used in healthcare facilities by clinical professionals or under their guidance. They should only be used by persons who have received adequate training in their use. The BeneVision N12/N15/N17/N19/N22 monitors are not intended for helicopter transport. hospital ambulance, or home use.
BeneVision N1 Patient Monitor:
The BeneVision N1 Patient Monitor is intended for monitoring, displaying, storing , alarming, and transferring of multiple physiological parameters including ECG (3-lead, 5-lead or 12-lead selectable, Arrhythmia Detection, ST Segment Analysis, and Heart Rate (HR)), Respiration (Resp), Temperature (Temp), Pulse Oxygen Saturation (SpO2), Pulse Rate (PR), Non-invasive Blood Pressure (NIBP), Invasive Blood Pressure (IBP) , Pulmonary Artery Wedge Pressure (PAWP), Carbon Dioxide (CO2) and Oxygen (O2). The system also provides an interpretation of resting 12-lead ECG.
All the parameters can be monitored on single adult, pediatric, and neonatal patients except for the following:
- PAWP, PNP, and PNC are intended for adult and pediatric patients only;
- A-Fib is intended for adult patients only;
The BeneVision N1 monitor is to be used in healthcare facilities. It can also be used during patient transport inside and outside of the hospital environment. It should be used by clinical professionals or under their guidance. It should only be used by persons who have received adequate training in its use. It is not intended for home use.
The subject BeneVision N Series Patient Monitors includes six monitors:
- BeneVision N12 Patient Monitor
- BeneVision N15 Patient Monitor
- BeneVision N17 Patient Monitor ●
- . BeneVision N19 Patient Monitor
- . BeneVision N22 Patient Monitor
- BeneVision N1 Patient Monitor ●
Mindray's BeneVision N Series Patient Monitors provide a flexible software and hardware platform to meet the clinical needs of patient monitoring.
The provided document describes the BeneVision N Series Patient Monitors and their clearance by the FDA based on substantial equivalence to predicate devices, particularly focusing on changes and new features. The document details the device's intended use, technological comparisons, and performance data from various tests. However, it does not explicitly provide a table of acceptance criteria and reported device performance in the format requested. Instead, it states that "the results of the bench testing show that the subject device meets its accuracy specification and is substantially equivalent to the predicate device."
Specifically regarding "Arrhythmia Detector And Alarm (Including ST-Segment Measurement And Alarm)" which is the relevant regulation name for the product code MHX (used by the BeneVision N Series Patient Monitors), the document states:
"Add arrhythmia detection for neonatal patients when used with the MPM 3.0 module." (Page 9, Table 2 ECG section; Page 22, Table 3 ECG section).
And references "K200015 - ePM series Patient Monitors...: provided as reference devices for ECG algorithm supporting arrhythmia detection in neonate that has been added to the subject BeneVision N Series Patient Monitors." (Page 6, Section 4).
Given the information in the document, here's a structured response:
Acceptance Criteria and Device Performance Study for Arrhythmia Detection in Neonates
The provided FDA 510(k) summary (K202405) for the BeneVision N Series Patient Monitors primarily focuses on demonstrating substantial equivalence to a predicate device (K192972) and a reference device (K200015) for the new feature of neonatal arrhythmia detection. While it explicitly states that the device meets its accuracy specifications, it does not provide a specific table of acceptance criteria or detailed reported performance metrics for neonatal arrhythmia detection. Instead, it relies on general statements about meeting specifications and compliance with standards.
However, based on the document's information and common regulatory expectations for such devices, we can infer the type of acceptance criteria and the nature of the study.
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly provide a table for the acceptance criteria and reported device performance for neonatal arrhythmia detection. It states that the device "meets its accuracy specification." For arrhythmia detection systems, typical performance metrics and acceptance criteria, often derived from standards like IEC 60601-2-27, would generally include:
| Metric | Acceptance Criteria (Inferred from regulatory standards and typical arrhythmia detector performance) | Reported Device Performance (Implicitly stated as "meets accuracy specifications" and "substantially equivalent") |
|---|---|---|
| Arrhythmia Detection Accuracy (Sensitivity/Specificity) | e.g., Sensitivity ≥ 90%, Specificity ≥ 95% for various arrhythmia types (e.g., asystole, bradycardia, tachycardia, PVCs) in applicable age groups (neonates) | Met (stated as "meets accuracy specification") |
| False Alarm Rate | e.g., False QRS detection rate < 1%, False alarm rate per hour < 0.2 | Met (stated as "meets accuracy specification") |
| Response Time to Alarms | e.g., Alarm generation within X seconds of event detection | Met (stated as "meets accuracy specification") |
| Heart Rate Accuracy | ±1 bpm or ±1%, whichever is greater (for neonates, 15-350 bpm range) | Met (as per device specifications in Table 2 & 3 ECG section) |
| Pace Detection | Accurate detection of paced beats; no false QRS detection due to pacing spikes | Met (stated as "pace detection enhancement" and "meets accuracy specification") |
| PVCs/min Range (Neonates) | [0, 350] | Met (as per device specifications in Table 2 & 3 ECG section) |
2. Sample Size for Test Set and Data Provenance
The document does not specify the exact sample size used for the test set for the neonatal arrhythmia detection algorithm. It broadly states that "Mindray conducted functional and system level testing on the subject device." The data provenance is also not explicitly stated in terms of country of origin or whether it was retrospective or prospective data. However, given that it's a 510(k) for a patient monitor and a feature addition (neonatal arrhythmia detection), it's highly likely that a combination of retrospective ECG databases (e.g., MIT-BIH Arrhythmia Database, AHA database, or internal proprietary databases containing neonatal ECGs) and/or prospective testing on simulated or recorded neonatal ECGs would have been used.
3. Number of Experts and Qualifications for Ground Truth
The document does not specify the number or qualifications of experts used to establish the ground truth for the test set related to neonatal arrhythmia detection. In studies for arrhythmia detection, ground truth is typically established by certified cardiologists, electrophysiologists, or other medical specialists trained in ECG interpretation, often with years of experience.
4. Adjudication Method
The document does not specify any adjudication method (e.g., 2+1, 3+1, none) used for establishing the ground truth or evaluating the test set results. For arrhythmia detection, a common method involves multiple experts reviewing the same data, with discrepancies resolved by a senior expert or consensus voting (e.g., 2 out of 3 agreement).
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
The document does not mention a Multi-Reader Multi-Case (MRMC) comparative effectiveness study directly related to the new neonatal arrhythmia detection feature or the overall performance of the patient monitor with human readers. The submission focuses on the performance of the device itself and its substantial equivalence to predicate devices, not on the improvement of human reader performance with AI assistance. Therefore, no effect size of human reader improvement with AI vs. without AI assistance is reported.
6. Standalone Performance Study
Yes, a standalone performance study was done for the algorithm. The document explicitly states: "Mindray conducted functional and system level testing to validate the performance of the devices. The results of the bench testing show that the subject device meets its accuracy specification and is substantially equivalent to the predicate device." This indicates that the algorithm's performance was evaluated independently of human-in-the-loop interaction.
7. Type of Ground Truth Used
The document does not explicitly state the type of ground truth used for the testing. For arrhythmia detection algorithms, the ground truth is typically established through expert consensus on annotated ECG waveforms, possibly validated against physiological events or other clinical data where available.
8. Sample Size for the Training Set
The document does not provide any information regarding the sample size for the training set used to develop or refine the neonatal arrhythmia detection algorithm.
9. How Ground Truth for the Training Set Was Established
Similarly, the document does not describe how the ground truth for the training set was established. Analogous to the test set, it would typically involve manual annotation of ECG data by experts, but this information is not present in the provided text.
Ask a specific question about this device
(401 days)
| |
| §870.2850
The monitors are indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients.
The monitors are intended to be used for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults, pediatrics, and neonates. The monitors are intended for use by trained healthcare professionals in a hospital environment.
The monitors are additionally intended for use in transport situations within hospital environments.
The monitors are only for use on one patient at a time. They are not intended for home use. Not therapeutic devices. The monitors are for prescription use only.
The ECG measurement is intended to be used for diagnostic recording of rhythm and detailed morphology of complex cardiac complexes (according to AAMI EC 11).
ST segment monitoring is intended for use with adult patients only and is not clinically validated for use with neonatal and pediatric patients.
The transcutaneous gas measurement (tcGas) with the M1018A plug-in module is restricted to neonatal patients only.
BIS is intended for use under the direct supervision of a licensed health care practitioner or by personnel trained in its proper use. It is intended for use on adult and pediatric patients within a hospital or medical facility providing patient care to monitor the state of the brain by data acquisition of EEG signals. The BIS may be used as an aid in monitoring the effects of certain anesthetic agents. Use of BIS monitoring to help guide anesthetic administration may be associated with the reduction of the incidence of awareness with recall in adults during general anesthesia and sedation.
The SSC Sepsis Protocol, in the ProtocolWatch clinical decision support tool, is intended for use with adult patients only.
The Integrated Pulmonary Index (IPI) is intended for use with adult and pediatric (1 to 12 years) patients only. The IPI is an adjunct to and not intended to replace vital sign monitoring.
The derived measurement Pulse Pressure Variation (PPV) is intended for use with sedated patients receiving controlled mechanical ventilation and mainly free from cardiac arrhythmia. The PPV measurement has been validated only for adult patients.
The IntelliVue NMT Module is intended to be used as an objective neuromuscular transmission monitor, using accelerometry for measuring the muscle contraction following an electrical stimulation of a peripheral nerve. The NMT Module is intended to be used with adult and pediatric patients.
The Masimo rainbow SET measurement is indicated for the noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, carboxyhemoglobin saturation (SpCO), methemoglobin saturation (SpMet), total hemoglobin concentration (SpHb), and/or respiratory rate (RRac). The Masimo rainbow SET measurement is indicated for use during both no motion and motion conditions, and for patients who are well or poorly perfused.
The noninvasive Masimo O3 Regional Oximeter System and accessories are intended for use as an adjunct monitor of absolute and trended regional hemoglobin oxygen saturation of blood (rSO2) in the cerebral region under the sensors. The Masimo O3 Regional Oximeter System and accessories are indicated for use on adults ≥40 kg and on pediatrics ≥5 kg and <40 kg, in healthcare environments.
The IntelliVue Patient Monitors MX500 and MX550 acquire multiple physiological patient signals, display measurement values, waves and trends, generate physiological and technical alarms, provide data recording and support patient data management. The monitors support multiple non-invasive and invasive measurements such as ECG, arrhythmia, ST, QT, SpO2, respiration rate, pulse rate, invasive and noninvasive blood pressure, temperature, CO2, tcpO2/ tcpCO2, C.O., CCO, intravascular SO2, SvO2, ScvO2, spirometry, EEG, BIS, NMT, and gas analysis.
The monitors offer a monitoring solution optimized for the surgical, cardiac, medical and neonatal care environments. They are located at the patient bedside vicinity and can also be used during patient transport inside hospitals.
The monitors have a color display with touchscreen as a primary input device. They also support a specialized remote control, keyboard and pointing devices such as a mouse. External displays can be connected to a built-in video port to provide an adaptive duplicate image of the primary display.
The monitors can interact with several compatible external measuring and auxiliary devices locally at the bedside or in transport situations and with the Central Station via LAN or wireless link.
Here's a breakdown of the acceptance criteria and the study information for the Philips IntelliVue Patient Monitors MX500 and MX550 based on the provided FDA 510(k) summary:
This submission is a 510(k) for modifications to an existing device, the IntelliVue Patient Monitors MX500 and MX550. The acceptance criteria and supporting studies primarily focus on demonstrating that the modifications do not introduce new questions of safety or effectiveness and that the modified device remains substantially equivalent to the predicate device.
1. Table of Acceptance Criteria and Reported Device Performance
Since this is an update to an existing device focusing on integrating new measurement modules (Masimo O3, IRMA CO2, ISA CO2) and updating EMC standards, the "acceptance criteria" are primarily related to meeting established performance standards for patient monitors and demonstrating that the new modules function as intended without compromising the overall device safety or efficacy. The document doesn't present specific numerical acceptance criteria for all physiological parameters of the overall monitor, but rather confirms compliance with recognized standards.
| Acceptance Criteria Category | Details/Standard Adherence | Reported Device Performance |
|---|---|---|
| General Safety and Performance | Demonstrated compliance with harmonized standards. | Passed all safety tests for demonstrated compliance with harmonized standards. |
| Software Life Cycle Processes | IEC 62304:2015 Edition 1.1 (consolidated version) - Medical device software - software life cycle processes. | Complies. Software Verification and Validation testing was conducted, and documentation was provided as recommended by FDA guidance. Software Level of Concern determined to be Major. |
| Basic Safety and Essential Performance | AAMI/ANSI ES60601-1:2005/(R)2012 and A1:2012 (Ed. 3.1) - Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance. | Complies. Electrical safety testing conducted. |
| Electromagnetic Compatibility (EMC) | IEC 60601-1-2:2014 (Ed. 4.0) - Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Disturbances - Requirements And Tests. | Complies. EMC testing conducted. Note: The predicate device conformed to Ed. 3; the subject device updates to Ed. 4. |
| Alarm Systems | IEC 60601-1-8:2012 (Ed. 2.1) - Medical Electrical Equipment - Part 1-8: General requirements for basic safety and essential performance - Collateral standard: General requirements, tests and guidance for alarm systems. | Device passed tests for this standard. |
| Respiratory Gas Monitors | ISO 80601-2-55:2011 (Ed. 1) - Medical Electrical Equipment - Part 2-55: Particular Requirements For The Basic Safety And Essential Performance Of Respiratory Gas Monitors. | Device passed tests for this standard. |
| System Level Device Specifications | Bench testing for system level device specifications. | Verified through bench testing. |
| Mechanical & Electrical Specifications | Bench testing for mechanical and electrical specifications. | Verified through bench testing. |
| Packaging Integrity | Bench testing for packaging integrity. | Verified through bench testing. |
| Shelf Life | Not applicable, as devices do not contain aging components. | No shelf life specified. |
| Sterility | Not applicable, as devices are not sterile. | Sterilization not evaluated. |
| Biocompatibility | Not applicable, as devices do not have patient contact. | Biocompatibility requirements are not applicable. Legally marketed medical accessories remain unchanged, so biocompatibility aspects are not affected. |
| Functionality of Added Masimo O3 Module | Intended use as an adjunct monitor of absolute and trended regional hemoglobin oxygen saturation of blood (rSO2) in the cerebral region for adults ≥40 kg and pediatrics ≥5 kg and <40 kg. (Reference Device: K162603) | The added Masimo O3 module's intended use fully aligns with the reference device, and its integration does not alter the substantial equivalence of the overall device. Received physiological data and forward it into existing SW infrastructure for presentation, data storage, user interface and alarming. |
| Functionality of Added Masimo IRMA/ISA CO2 | Intended use as a gas analyzer for Carbon Dioxide. (Reference Devices: K123043, K171121) | The added IRMA/ISA CO2 measurements align with the reference devices and their integration does not alter the substantial equivalence of the overall device. Received physiological data and forward it into existing SW infrastructure for presentation, data storage, user interface and alarming. |
2. Sample Size Used for the Test Set and Data Provenance
The document explicitly states: "The IntelliVue Patient Monitors MX500 and MX550, like the predicate device, did not require clinical trials."
Therefore, there is no "test set" in the traditional sense of a clinical study involving patients or human readers. The evaluation was based on non-clinical bench testing and compliance with harmonized standards.
- Sample Size for Test Set: Not applicable as no clinical test set was used.
- Data Provenance: Not applicable. The data is from in-house bench testing and compliance evaluations against recognized standards.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications
Not applicable. As no clinical trials or human-in-the-loop performance studies were conducted for this submission, there was no need for experts to establish ground truth in the context of a test set involving patient data or diagnostic interpretations. The ground truth for the functionality of the integrated Masimo modules is established by their prior clearance as standalone devices (reference devices K162603, K123043, K171121).
4. Adjudication Method for the Test Set
Not applicable, as no clinical test set requiring adjudication was used.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The submission states, "The IntelliVue Patient Monitors MX500 and MX550, like the predicate device, did not require clinical trials." This indicates that there was no study involving human readers or AI assistance.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
While there isn't an "algorithm only" study defined in the context of AI, the performance of the integrated components (Masimo O3, IRMA CO2, ISA CO2 modules) in receiving and forwarding physiological data was evaluated. This can be seen as a form of standalone testing of the integration without human interpretation. The original Masimo devices (reference devices) would have undergone their own standalone validation studies prior to their clearances (K162603, K123043, K171121), which are implicitly relied upon here. This submission focuses on the integration into the Philips monitor.
7. Type of Ground Truth Used
For the integrated measurement modules:
- Previous FDA clearances of reference devices: The performance and safety of the Masimo O3 Regional Oximeter System, Masimo Infrared Mainstream Gas Analyzer CO2, and Masimo Infrared Sidestream Gas Analyzer CO2 were established during their original 510(k) clearances (K162603, K123043, K171121). This forms the "ground truth" for their individual measurement capabilities.
- Compliance with harmonized standards: For the overall monitor and its specific functions (e.g., electrical safety, EMC, alarm systems), the ground truth is defined by meeting the requirements of internationally recognized standards (e.g., IEC 60601 series, ISO 80601-2-55).
- Bench testing results: In-house bench testing validated the system-level specifications, mechanical/electrical aspects, and software functionality according to pre-defined internal test protocols and specifications derived from design requirements.
8. Sample Size for the Training Set
Not applicable. This device is a patient monitor with physiological measurement capabilities, not an AI/machine learning device that requires a training set in the typical sense. Its functionality is based on established engineering principles and validated measurement technologies.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no training set for this device.
Ask a specific question about this device
(196 days)
| Alarm, blood-pressure |
| 870.2850
The BeneVision N12N15/N17/N19/N22 patient monitors are intended for monitoring, displaying, storing, storing, alarming, and transferring of multiple physiological parameters including ECG (3-lead, 5-lead or 12-lead selectable, Arrhythmia Detection, ST Segment Analysis, QT Analysis, and Heart Rate (HR)), Respiration Rate (Resp), Temperature (Temp), Pulse Oxygen Saturation (SpO2), Pulse Rate (PR), Non-invasive Blood Pressure (NIBP), Invasive Blood Pressure(IBP), Pulmonary Artery Wedge Pressure (PAWP), Cardiac Output (C.O.), Continuous Cardiac Output (CCO), Mixed/Central Venous Oxygen Saturation (SvO2/ScvO2), Carbon Dioxide (CO2), Oxygen (O2), Anesthetic Gas (AG), Impedance Cardiograph (ICG), Bispectral Index (BIS), Respiration Mechanics (RM), Neuromuscular Transmission Monitoring (NMT), Electroencephalograph (EEG), and Regional Oxygen Saturation (rSO2). The system also provides an interpretation of resting 12-lead ECG.
All the parameters can be monitored on single adult, pediatric, and neonatal patients except for the following: · The arrhythmia detection, BIS, RM, CCO, SvO2/ScvO2, PAWP, and NMT monitoring are intended for adult and pediatric patients only:
· C.O. monitoring is intended for adult patients only;
• ICG monitoring is intended for only adult patients who meet the following requirements: height: 122 to 229cm, weight: 30 to 155kg.
· rSO2 monitoring is intended for use in individuals greater than 2.5kg.
The monitors are to be used in healthcare facilities by clinical professionals or under their guidance. They should only be used by persons who have received adequate training in their use. The BeneVision N12/N15/N17/N19/N22 monitors are not intended for helicopter transport. hospital ambulance, or home use.
BeneVision N1 Patient Monitor:
The BeneVision N1 Patient Monitor is intended for monitoring, reviewing, storing , alarming, and transferring of multiple physiological parameters including ECG (3-lead, 5-lead or 12-lead selectable, Arrhythmia Detection, ST Segment Analysis, and Heart Rate (HR)), Respiration (Resp), Temperature (Temp), Pulse Oxygen Saturation (SpO2), Pulse Rate (PR), Non-invasive Blood Pressure (NIBP), Invasive Blood Pressure (IBP) , Pulmonary Artery Wedge Pressure (PAWP), Carbon Dioxide (CO2) and Oxygen (O2). The system also provides an interpretation of resting 12-lead ECG.
All the parameters can be monitored on single adult, pediatric, and neonatal patients except for the following:
• The arrhythmia detection and PAWP is intended for adult and pediatric patients only
The BeneVision N1 monitor is to be used in healthcare facilities. It can also be used during patient transport inside and outside of the hospital environment. It should be used by clinical professionals or under their guidance. It should only be used by persons who have received adequate training in its use. It is not intended for home use.
The subject BeneVision N Series Patient Monitors includes six monitors: BeneVision N12 Patient Monitor, BeneVision N15 Patient Monitor, BeneVision N17 Patient Monitor, BeneVision N19 Patient Monitor, BeneVision N22 Patient Monitor, BeneVision N1 Patient Monitor. Mindray's BeneVision N Series Patient Monitors provide a flexible software and hardware platform to meet the clinical needs of patient monitoring.
The provided text is a 510(k) Summary for the Mindray BeneVision N Series Patient Monitors. It mainly focuses on demonstrating substantial equivalence to a predicate device rather than presenting a detailed study with acceptance criteria and specific performance metrics in the format requested.
The document discusses functional and system level testing and bench testing to validate performance and ensure specifications are met, but it does not provide a table of acceptance criteria alongside reported device performance data, nor does it detail the specifics of such a study.
Here's a breakdown of the requested information based on the provided text, highlighting what is not available:
-
A table of acceptance criteria and the reported device performance
- Not Available. The document states that "functional and system level testing showed that the devices continue to meet specifications and the performance of the device is equivalent to the predicate" and "the subject device meets its accuracy specification." However, it does not provide the specific numerical acceptance criteria or the reported performance data in a table. It only lists measurement ranges and accuracies for various parameters (ECG, Respiration Rate, Temperature, SpO2, Pulse Rate, NIBP, IBP, C.O., CCO, SvO2/ScvO2, CO2, O2, AG, ICG, RM, NMT, EEG, rSO2) under the "Technological Comparison" tables (Table 2 and Table 3). These are presented as specifications rather than acceptance criteria for a study.
-
Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Not Available. The document mentions "functional and system level testing" and "bench testing" but provides no details on the sample size of data used for these tests, nor its provenance (country, retrospective/prospective nature).
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not Applicable / Not Available. This device is a patient monitor, and its performance is assessed against established physical measurement standards and algorithms, not typically against expert interpretation of data for ground truth in the way an AI diagnostic algorithm might be (e.g., radiologists for medical images). The document does not mention any expert review process for determining ground truth.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not Applicable / Not Available. As mentioned above, the assessment appears to be against physical specifications and consensus standards, not an expert-adjudicated ground truth.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No, not done. The document does not mention an MRMC study or any assessment of human reader improvement with AI assistance. This device is a patient monitor providing physiological parameters, not an AI-assisted diagnostic tool for human interpretation.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- N/A (Implicitly standalone for most parameters against physical standards). The performance evaluation described largely implies standalone testing against specifications and standards. For example, for ECG, SpO2, NIBP, etc., the device's internal algorithms process signals and provide readings, which are then assessed for accuracy. The "reporting" is of the device's ability to measure parameters correctly, not its ability to assist a human.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- Physical Standards and Reference Devices/Methods. The ground truth for the measurements would be established by validated physical standards, calibrated equipment, and comparison to reference methods (e.g., for invasive blood pressure, cardiac output). The document implies validation against these types of established standards and specifications.
-
The sample size for the training set
- Not Applicable / Not Available. The document does not describe the development or training of AI algorithms that would require a "training set" in the context of machine learning. The patient monitor's algorithms are based on established physiological signal processing, not deep learning models.
-
How the ground truth for the training set was established
- Not Applicable. See point 8.
Ask a specific question about this device
(131 days)
| Alarm, blood-pressure |
| 870.2850
The BeneVision N12/N15/N17/N19/N22 patient monitors are intended for monitoring, displaying, reviewing, storing, and transferring of multiple physiological parameters including ECG (3-lead, 5-lead, 6-lead or 12-lead selectable, Arrhythmia Detection, ST Segment Analysis, QT Analysis, and Heart Rate (HR)), Respiration Rate (Resp), Temperature (Temp), Pulse Oxygen Saturation (SpO2), Pulse Rate (PR), Non-invasive Blood Pressure (NIBP), Invasive Blood Pressure(IBP), Pulmonary Artery Wedge Pressure (PAWP), Cardiac Output (C.O.), Continuous Cardiac Output (CCO), Mixed/Central Venous Oxygen Saturation (SvO2/ScvO2), Carbon Dioxide (CO2), Oxygen (O2), Anesthetic Gas (AG), Impedance Cardiograph (ICG), Respiration Mechanics (RM), Neuromuscular Transmission Monitoring (NMT), Electroencephalograph (EEG), and Regional Oxygen Saturation (rSO2). The system also provides an interpretation of resting 12-lead ECG.
All the parameters can be monitored on single adult, pediatric, and neonatal patients except for the following:
- The arrhythmia detection, RM, CCO, SvO2/ScvO2, PAWP, and NMT monitoring are intended for adult and pediatric patients only;
- C.O. monitoring is intended for adult patients only;
- ICG monitoring is intended for only adult patients who meet the following requirements: height: 122 to 229cm, weight: 30 to 155kg.
- rSO2 monitoring is intended for use in individuals greater than 2.5kg.
The monitors are to be used in healthcare facilities by clinical professionals or under their guidance. They should only be used by persons who have received adequate training in their use. The BeneVision N12/N15/N17/N19/N22 monitors are not intended for helicopter transport, hospital ambulance, or home use.
The BeneVision N1 Patient Monitor is intended for monitoring, displaying, storing, alarming, and transferring of multiple physiological parameters including ECG (3-lead, 5-lead, 6-lead or 12-lead selectable, Arrhythmia Detection, ST Segment Analysis, QT Analysis, and Heart Rate (HR)), Respiration (Resp), Temperature (Temp), Pulse Oxygen Saturation (SpO2), Pulse Rate (PR), Non-invasive Blood Pressure (NIBP), Invasive Blood Pressure (IBP) , Pulmonary Artery Wedge Pressure (PAWP), Carbon Dioxide (CO2) and Oxygen (O2). The system also provides an interpretation of resting 12-lead ECG.
All the parameters can be monitored on single adult, pediatric, and neonatal patients except for the following:
- The arrhythmia detection and PAWP is intended for adult and pediatric patients only
The BeneVision N1 monitor is to be used in healthcare facilities. It can also be used during patient transport inside and outside of the hospital environment. It should be used by clinical professionals or under their guidance. It should only be used by persons who have received adequate training in its use. It is not intended for home use.
The subject BeneVision N Series Patient Monitors includes six monitors:
- BeneVision N12 Patient Monitor
- BeneVision N15 Patient Monitor
- BeneVision N17 Patient Monitor
- BeneVision N19 Patient Monitor
- BeneVision N22 Patient Monitor
- BeneVision N1 Patient Monitor
The BeneVision N Series Patient Monitors are Mindray's new generation monitoring product family with ergonomic and flexible design in platform of both software and hardware to meet the clinical needs of monitoring.
The provided document is a 510(k) Summary for the Mindray BeneVision N Series Patient Monitors. It focuses on demonstrating substantial equivalence to predicate devices rather than proving the device meets specific acceptance criteria through a dedicated study with statistical endpoints.
Therefore, many of the requested elements for a detailed study description (e.g., sample size for test/training sets, data provenance, number/qualifications of experts, adjudication methods, MRMC studies, standalone performance with specific metrics, and ground truth establishment for training data) are not present in the provided text.
The document primarily highlights changes from predicate devices and states that functional and system-level testing, along with compliance with consensus standards, demonstrate equivalence.
Here's a summary of the available information:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly present a table of acceptance criteria with corresponding device performance metrics in the format typically seen for a new device's efficacy study. Instead, it compares the specifications of the subject device with those of predicate devices, implicitly indicating that the subject device's performance meets or exceeds the predicate's established performance or relevant cleared standards.
| Feature Area / Parameter | Acceptance Criteria (Implicitly, equivalent to/better than Predicate or standard) | Reported Device Performance (Subject BeneVision Devices) |
|---|---|---|
| Display | - | N22: 22" 16801050 pixelsN19: 19" 16801050 pixelsN17: 18.5" 19201080 pixelsN15: 15.6" 19201080 pixelsN12: 12.1" 1280800 pixelsN1: 5.5" 7201280 pixels |
| Wireless | 2.4GHz/5GHz dual band module (Passport 12m) | 2.4GHz/5GHz dual band module (All BeneVision N Series) |
| Data Storage | Compact Flash (Passport 17m)SD card (T1) | Solid State Hard Drive (SSD) (N22, N19)Embedded Multi Media Card (eMMC) (N17, N15, N12, N1) |
| Alarm System | Yellow/red alarm lamp (Predicate) | Cyan, yellow, or red alarm lamp; Supports Alarm Volume Escalation (Feature cleared in K161531) |
| ECG - 6-lead ECG | Not supported by predicate | Supported (Feature cleared in K162607) |
| ECG - Intelligent Arrhythmia Alarm | Not supported by predicate | Supported (Feature cleared in K161531) |
| ECG - ST Segment Analysis (Pediatric/Neonate) | Only adult (Predicate) | Pediatric and neonate supported (Feature cleared in K131414) |
| SpO2 - Masimo SpO2 module in MPM 3.0 | Not supported by predicate | Supported (Feature cleared in K053269) |
| CO2 - Sidestream CO2 2.0 Module | Not supported by predicate (Type 1.0 supported) | Supported (Feature cleared in K170712)CO2 measurement range: 0-150mmHg (wider than predicate)AwRR measurement range: 0-150rpm (wider than predicate)AwRR accuracy improved |
| NMT Module | Not applicable (Predicate) | Supported (Feature cleared in K170876) |
| EEG Module | Not applicable (Predicate) | Supported (Feature cleared in K161531) |
| rSO2 Module | Not applicable (Predicate) | Supported (Feature cleared in K082327) |
| Gas Recycling (AG module) | Not supported by predicate | Supported (Feature cleared in K171292) |
| Early Warning Score (EWS) | Not applicable (Predicate) | Supported (Feature cleared in K170712) |
| Helicopter/ambulance transport (N1) | Not applicable (Predicate) | Supported for ECG, RESP, Temp, SpO2, PR, NIBP, IBP (Feature cleared in K161531) |
| NIBP Measurement Range | Adult: 40-270 (Systolic), 10-210 (Diastolic), 20-230 (Mean)Pediatric: 40-200 (Systolic), 10-150 (Diastolic), 20-165 (Mean)Neonate: 40-135 (Systolic), 10-100 (Diastolic), 20-110 (Mean) | Adult: 25-290 (Systolic), 10-250 (Diastolic), 15-260 (Mean)Pediatric: 25-240 (Systolic), 10-200 (Diastolic), 15-215 (Mean)Neonate: 25-140 (Systolic), 10-115 (Diastolic), 15-125 (Mean) |
| NIBP Accuracy | Max mean error: ±5 mmHg; Max standard deviation: 8 mmHg (Predicate) | Max mean error: ±5 mmHg; Max standard deviation: 8 mmHg (Same as Predicate) |
| IBP Measurement Range | -50 to 300 mmHg (Predicate) | -50 to 300 mmHg (Same as Predicate) |
| IBP Accuracy | ±2% or ±1 mmHg, whichever is greater (without sensor) (Predicate) | ±2% or ±1 mmHg, whichever is greater (without sensor) (Same as Predicate) |
| Cardiac Output Measurement Range | 0.1 to 20 L/min (C.O.); 23 to 43 °C (TB); 0 to 27 °C (TI) (Predicate) | Same as Predicate |
| Cardiac Output Accuracy | ±5% or ±0.1 L/min (C.O.); ±0.1 °C (TB, TI) (Predicate) | Same as Predicate |
2. Sample size used for the test set and the data provenance
The document does not specify sample sizes for test sets. The testing mentioned is referred to as "functional and system level testing" and "bench testing." It also states Mindray conducted "clinical testing to demonstrate that the Mindray and Nellcor SpO2 modules meet relevant consensus standards."
There is no mention of data provenance (e.g., country of origin of data, retrospective or prospective).
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. The document does not describe the use of experts to establish ground truth for testing. The evaluation focused on meeting specifications and consensus standards, and demonstrating equivalence to predicate devices.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
Not applicable. There is no mention of adjudication methods.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is a patient monitor, not an AI-assisted diagnostic device, and no MRMC studies are mentioned.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
The document describes performance in terms of functionality and adherence to technical specifications and consensus standards, not in terms of "algorithm-only" performance as would be relevant for an AI device. The tests performed are for the integrated device.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The term "ground truth" is not used. The performance evaluation is based on meeting engineering specifications, comparing against predicate device performance, and compliance with recognized consensus standards (e.g., IEC, ISO, AAMI standards for physiological measurement accuracy).
8. The sample size for the training set
Not applicable. A "training set" is relevant for machine learning algorithms. This document describes a patient monitor, and no machine learning model training is discussed.
9. How the ground truth for the training set was established
Not applicable, as no training set for a machine learning model is mentioned.