    510k Summary Text (Full-text Search) :

    Massachusetts 01730

    Re: K213464

    Trade/Device Name: HemosIL Liquid Anti-Xa Regulation Number: 21 CFR 864.7525
    Description | Class | ProductCode | Panel | | | |
    | Heparin | 864.7525

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    HemosIL Liquid Anti-Xa is an automated chromogenic assay for in vitro diagnostic use by laboratory professionals in clinical laboratories. The assay provides quantitative results on 3.2% citrated human plasma for the following analytes based on the calibrators used:

    • · When used with HemosIL Heparin Calibrators: Quantitative determination of unfractionated heparin (UFH) and low molecular weight heparin (LMWH) activity on the ACL TOP Family and ACL TOP Family 50 Series.
    • · When used with HemosIL Apixaban Calibrators:

    Quantitative determination of apixaban on the ACL TOP Family 50 Series through measurement of Factor Xa activity, which is inversely proportional to the apixaban level. With HemosIL Apixaban Calibrators, the assay is intended to measure apixaban concentrations in patients on apixaban therapy in the following situations where measurement of apixaban levels could be useful to have as additional information:

    • Patients at risk for major bleeding
    • Patients experiencing a bleeding episode

    The assay is not a stand-alone test and the results should be used in conjunction with other clinical and laboratory findings.

    For use in adult population. For prescription use only.

    Device Description

    HemosIL Liquid Anti-Xa is a one stage chromogenic assay based on a synthetic chromogenic substrate and on Factor Xa inactivation. The assay provides quantitative results on 3.2% citrated human plasma for the following analytes based on the calibrators used:

    . When used with HemosIL Heparin Calibrators:

    Heparin levels in patient plasma are measured automatically on ACL TOP Family and ACL TOP Family 50 Series when this assay is calibrated with HemosIL Heparin Calibrators.

    Heparin is analyzed as a complex with antithrombin present in the sample. The concentration of this complex is dependent on the availability of the patient's endogenous antithrombin. When the heparinantithrombin complex is formed, two competing reactions take place.

      1. Factor Xa is neutralized by heparin-antithrombin complex.
      1. Residual Factor Xa is quantified with a synthetic chromogenic substrate. The paranitroaniline released is monitored kinetically at 405 nm and is inversely proportional to the heparin level in the sample.

    In order to reduce the influence from heparin antagonists, such as platelet factor 4 (PF4), dextran sulfate is included in the reaction mixture.

    When used with HemosIL Apixaban Calibrators: .

    Apixaban levels in patient plasma are measured automatically on ACL TOP Family and ACL TOP Family 50 Series when this assay is calibrated with HemosIL Apixaban Calibrators.

    Apixaban directly inhibits Factor Xa activity independent of the antithrombin present. The Factor Xa activity measured by the assay is exogenous. Factor Xa is neutralized directly by apixaban.

    Residual Factor Xa is quantified with a synthetic chromogenic substrate. The paranitroaniline released is monitored kinetically at 405 nm and is inversely proportional to the apixaban level in the sample.

    Measurement of apixaban concentration is recommended by the International Society of Thrombosis and Hemostasis Subcommittee on Control of Anticoagulation in certain clinical scenarios including bleeding episodes, perioperative management, and suspicion of overdose.

    AI/ML Overview

    The provided document is a 510(k) summary for a medical device (HemosIL Liquid Anti-Xa), primarily focused on a specific change: modifying the labeled on-board instrument stability claim from 7 days to 4 days and removing claims for a particular instrument family (ACL Elite/Elite Pro). This type of submission (Special 510(k)) indicates that the core device and its fundamental performance characteristics are already established and the submission is for a minor modification.

    Therefore, the document does not contain the information typically found in a clinical study report that would establish the initial acceptance criteria and prove the device meets them from scratch. It refers to a guideline (CLSI EP25-A) for the testing conducted for the stability claim change, but doesn't detail the acceptance criteria for the entire device's performance (e.g., accuracy, precision, linearity, etc., across its full analytical range).

    The information requested in the prompt (sample size for test set, data provenance, number/qualifications of experts, adjudication method, MRMC study, standalone performance, ground truth establishment for training and test sets) is characteristic of studies for diagnostic devices, particularly AI/software-as-a-medical-device (SaMD) products, where performance is often evaluated against human expert consensus or clinical outcomes. The HemosIL Liquid Anti-Xa is an in vitro diagnostic (IVD) assay, not an AI/SaMD product that requires human expert review of images or signals for ground truth. Its performance validation relies on analytical studies (e.g., accuracy against reference methods, precision, linearity, interference studies).

    Based on the provided document, I cannot fulfill most of the requested information because it is not contained within this 510(k) summary.

    Specifically, the document:

    • Does not provide a full table of acceptance criteria for the device's overall performance (only discusses a change in stability claim).
    • Does not discuss aspects like sample size for test sets in the context of clinical performance evaluation against a human-established ground truth, data provenance for such sets, expert numbers/qualifications, or adjudication methods, as these are typically not relevant for an IVD reagent's analytical performance assessment in the same way they are for image-based AI diagnostics.
    • Does not mention any multi-reader multi-case (MRMC) comparative effectiveness study, as it's not an AI-assisted diagnostic tool.
    • Does not discuss standalone algorithm performance, as it's a reagent for an automated instrument.
    • Does not specify the type of ground truth used in the context of human expert review. For an IVD, "ground truth" would typically refer to reference method results or clinical diagnosis established through established laboratory and clinical procedures.
    • Does not provide information on training set sample size or how ground truth was established for a training set, as it is not an AI/machine learning device that undergoes a training phase in the typical sense.

    The only relevant information that can be extracted regarding a "study" is for the change being submitted:

    1. Acceptance Criteria and Device Performance (for On-Board Stability Change):

    The document states the study was conducted "based on testing to the current CLSI EP25-A guideline" to support the change in on-board stability from 7 days to 4 days. While it doesn't explicitly list the numerical acceptance criteria for this specific study, the implication is that the data collected over 4 days met the performance specifications (e.g., accuracy, precision) as defined by the CLSI EP25-A guideline for reagent stability. The reported performance is that the device meets the 4-day stability claim, leading to the regulatory approval.

    Acceptance Criteria (Implied by CLSI EP25-A for Stability)Reported Device Performance (within 4 days)
    Performance (e.g., accuracy, precision) within specifications for the stated analyte range.Met for 4 days at 15-25°C

    2. Sample Size and Data Provenance for Stability Test: While the specific sample size for the stability study is not detailed, it would involve multiple replicates of control and/or patient samples measured over the 4-day period on the specified instruments (ACL TOP Family and ACL TOP Family 50 Series) stored at 15-25°C. Data provenance would be from internal laboratory studies of Instrumentation Laboratory Co. (the manufacturer). This would be a prospective analytical study designed to assess reagent stability under defined conditions.

    3. Number of Experts and Qualifications / Adjudication Method / MRMC Study / Standalone Performance: Not applicable for this type of IVD reagent validation. There are no human experts involved in establishing ground truth for analytical performance, nor is there a multi-reader study or standalone algorithm.

    4. Type of Ground Truth Used: For analytical performance like stability, the "ground truth" would be the initial performance of the freshly prepared reagent or its performance compared to a reference method, against which subsequent measurements over time are compared to assess degradation. This is an analytical ground truth, not a clinical ground truth established by human experts.

    5. Training Set Sample Size and Ground Truth Establishment: Not applicable. This is an IVD reagent and not an AI/ML device that requires a "training set" in the sense of machine learning. The design and parameters of the assay are established through chemical and biological research and development, not data-driven machine learning.

    Ask a Question

    Ask a specific question about this device

    K Number
    K162540

    Validate with FDA (Live)

    Date Cleared
    2017-08-22

    (344 days)

    Product Code
    Regulation Number
    864.7525
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Assay INNOVANCE® Heparin Calibrator INNOVANCE® Heparin UF and LMW Controls Regulation Number: 21 CFR 864.7525
    |
    | Classification Name: | Heparin assay (21 CFR 864.7525

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    INNOVANCE® Heparin Assay
    In vitro diagnostic automated chromogenic assay for the quantitative determination of unfractionated heparin (UFI) and low molecular weight heparin (LMWH) activity in human plasma collected from venous blood samples in 3.2% sodium citrate tubes on the BCS® XP System in the clinical laboratory. For use with plasma from patients undergoing heparin anticoagulant therapy with either UFH or LMWH. The performance of this device has not been established in neonate and pediatric patient populations.

    INNOVANCE® Heparin Calibrator
    For calibration of the INNOVANCE® Heparin assay for the quantitative determination of the activity of unfractionated heparin (UFH) and low molecular weight heparin (LMWH) in citrated human plasma.

    INNOVANCE® Heparin UF and LMW Controls
    For quality control of the INNOVANCE® Heparin assay for the quantitative determination of the activity of unfractionated heparin (UFH) and low molecular weight heparin (LMWH) in citrated human plasma.

    Device Description

    The INNOVANCE® Heparin assay is a one stage chromogenic assay. The reagent kit consists of two components. One component (INNOVANCE Heparin Reagent) contains Coagulation Factor Xa (Xa), the other (INNOVANCE Heparin Substrate) a chromogenic substrate specific for Xa. Upon mixing of INNOVANCE Heparin Reagent and INNOVANCE Heparin Substrate, Xa converts the chromogenic substrate into two products, one of them is paranitroaniline. The formation of paranitroaniline can be quantified by the coaqulation analyzer employing light absorption at a specific wavelength (405 nm). In the presence of a heparin containing sample the formation of paranitroaniline will be reduced in a time dependent manner. This is due to inhibition of Xa by the heparin/antithrombin (AT) complex. This complex is formed in the patient's plasma and competes with the substrate conversion by Xa. The concentration of the complex is not only dependent on the concentration of heparin but also on the availability of the patient's endogenous antithrombin. By comparison to a reference curve the heparin activity of the sample can be quantified. To reduce the influence from heparin antagonists, such as platelet factor 4 (PF4), dextran sulfate is included in the reaction mixture.

    The INNOVANCE® Heparin Calibrator consists of 5 calibrator levels. INNOVANCE® Heparin Calibrator 1 represents plasma containing no heparin. INNOVANCE® Heparin Calibrator 2, 3, 4 and 5 contain defined activities of LMWH and are calibrated against the World Health Orqanization (WHO) International Standards for UFH and LMWH. The calibrator levels are used to establish a reference curve (calibration curve) which then can be employed to quantify the heparin activity of UFH and LMWH containing plasmas.

    The INNOVANCE® Heparin Controls consist of plasmas containing defined activities of either UFH or LMWH. Recovery of these controls within their assigned ranges indicates proper functionality of the assay system.

    AI/ML Overview

    This document describes the performance data for the INNOVANCE® Heparin Assay, INNOVANCE® Heparin Calibrator, and INNOVANCE® Heparin UF and LMW Controls. The information provided focuses on the analytical performance of the device rather than a clinical study involving human readers and AI assistance. Therefore, some requested information related to human expert involvement, MRMC studies, and AI-specific ground truth establishment cannot be extracted from this document.

    Here's an analysis of the acceptance criteria and study proving the device meets them, based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document provides extensive data on linearity, specificity (interference), reproducibility, and repeatability. Specific acceptance criteria are mentioned as "pre-established acceptance criteria" or inferred from the "no interference up to..." statements. Since the exact numerical acceptance criteria are not explicitly listed in a consolidated table, I will present the reported performance and note that the studies "met pre-established acceptance criteria" or "were found to cause no interference."

    Acceptance Criteria and Reported Device Performance for INNOVANCE® Heparin Assay System

    Performance MetricAcceptance Criteria (Implicit/Explicit)Reported Device Performance
    Measuring Range (Linearity & LoQ)- Established Limit of Quantitation (LoQ)- Assay range established as 0.10 - 1.50 IU/mL.
    - Linearity verified across specified range- Linearity study performed with 21 different dilutions for UFH and LMWH (0.00 to 1.90 IU/mL).
    Specificity (Interference)- No interference up to specified concentrations for various endogenous and exogenous substances.No interference observed up to:
    - Ascorbic Acid: 176 mg/dL
    - Bilirubin (unconjugated): 60 mg/dL
    - Bilirubin (conjugated): 40 mg/dL
    - Hemoglobin: 347 mg/dL
    - Platelet Factor 4: 0.7 µg/mL
    - Triglycerides: 807 mg/dL
    - Rheumatoid Factor: 220 IU/mL
    - Apixaban: 4.1 ng/mL
    - Danaparoid Sodium: 0.03 IU/mL
    - Fondaparinux: 35.5 ng/mL
    - Rivaroxaban: 7.1 ng/mL
    Reproducibility & Repeatability (Reagent)- Meets CLSI EP05-A2 guidelines.- Reproducibility (Table 1): Total %CVs for LMW controls ranged from 2.97% to 4.40%, for UFH controls from 3.29% to 7.30%. For LMWH plasma pools, total %CVs ranged from 3.47% to 21.15%. For UFH plasma pools, total %CVs ranged from 4.30% to 20.60%.
    - Repeatability (Table 2): Total %CVs for LMW controls ranged from 1.85% to 3.12%, for UFH controls from 3.58% to 3.71%. For LMWH plasma pools, total %CVs ranged from 2.83% to 9.30%. For UFH plasma pools, total %CVs ranged from 3.21% to 7.58%.
    Reproducibility & Repeatability (Calibrator)- Total %CV for reproducibility (Table 3) and repeatability (Table 4) within specified limits.- Reproducibility (Table 3): Total %CVs for calibrator levels ranged from 2.23% to 3.36%.
    - Repeatability (Table 4): Total %CVs for calibrator levels ranged from 2.53% to 3.74%.
    Reproducibility & Repeatability (Control)- Total %CV for reproducibility (Table 5) and repeatability (Table 6) within specified limits.- Reproducibility (Table 5): Total %CVs ranged from 2.97% to 7.30%.
    - Repeatability (Table 6): Total %CVs ranged from 2.45% to 3.67%.
    Frozen vs. Fresh Method Comparison- Passing-Bablok regression slope, intercept, and correlation coefficient meet pre-established criteria.- Slope: 1.00, Intercept: -0.01 IU/mL, r: 0.996, r2: 0.993. Predicted bias for MDP (0.3, 0.4 IU/mL) was -0.01 IU/mL. Percent bias for MDP (0.6, 0.7, 1.0 IU/mL) ranged from -1.00% to -1.67%. All acceptance criteria were met.
    Method Comparison (vs. predicate device)- Passing-Bablok regression slope, intercept, and correlation coefficient meet pre-established criteria.- UFH samples: Slope: 0.93, Intercept: 0.03, Correlation Coefficient (r): 0.98.
    - LMWH samples: Slope: 1.06, Intercept: -0.01, Correlation Coefficient (r): 0.99.
    - Results met the pre-established acceptance criteria.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    • Measuring Range (Linearity): 21 different dilutions per heparin type (UFH and LMWH). Uses plasma pools spiked with standards.
    • Specificity (Interference): Base pools of UFH and LMWH were prepared. Specific sample count not provided, but implies multiple samples to establish "no interference up to..." levels.
    • Reproducibility and Repeatability:
      • Reagent (reproducibility): 240 measurements (N=240) per sample type (LMW/UF controls, LMWH/UFH plasma pools).
        • Conducted at three external sites and one internal site (Siemens Marburg site prepared pools). Data provided is multicenter.
        • Plasma pools were prepared at Siemens Marburg and sent frozen to external sites, implying retrospective analysis of these prepared samples.
      • Reagent (repeatability): 240 measurements (N=240) per sample type.
        • Conducted at one site (Bad Oeynhausen).
      • Calibrator (reproducibility): 120 measurements (N=120) per calibrator level.
        • Conducted at three internal sites.
      • Calibrator (repeatability): 240 measurements (N=240) per calibrator level.
        • Conducted at one site.
      • Controls (reproducibility): 240 measurements (N=240) per control level.
        • Conducted at three sites.
      • Controls (repeatability): 240 measurements (N=240) per control level.
        • Conducted at one site.
    • Frozen vs. Fresh Method Comparison: 69 samples (N=69).
      • Conducted at one study site in Germany.
      • This study involved both fresh samples (measured within 4 hours) and frozen aliquots of the same samples (measured after at least one week of -70°C storage), making it a comparison of sample handling rather than purely retrospective vs. prospective patient data collection.
    • Method Comparison (vs. predicate device):
      • UFH samples: 165 samples (N=165).
      • LMWH samples: 155 samples (N=155).
      • Conducted at three sites, including one external US site. All sites used the same protocol.
      • Frozen samples were thawed and measured, indicating a retrospective use of collected samples.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This document describes the analytical performance of an in-vitro diagnostic device for quantifying heparin activity. The "ground truth" in this context is established by reference methods, international standards (WHO International Standards for UFH and LMWH), and analytical accuracy/precision measurements, not by human expert interpretations of images or clinical outcomes. Therefore, there are no human experts involved in establishing the "ground truth" as described for medical imaging AI devices.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This is an analytical performance study for an in-vitro diagnostic assay. Adjudication methods like 2+1 or 3+1 are typical for subjective human interpretation tasks, such as radiology reads, which are not relevant here.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This document is about an in-vitro diagnostic device, not an AI-assisted diagnostic tool that involves human readers interpreting cases.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, the studies presented (linearity, specificity, reproducibility, repeatability, frozen vs. fresh comparison, method comparison) are effectively "standalone algorithm" performance evaluations, as they assess the device's analytical measurement capabilities directly, without human interpretation of results as a variable. The "algorithm" here refers to the biochemical reactions and instrument logic that produce the quantitative heparin activity measurement.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The ground truth for this device is based on:

    • International Standards: The INNOVANCE® Heparin Calibrators and Controls are traceable to the World Health Organization (WHO) International Standards for UFH and LMWH. This is the primary reference for accuracy and calibration.
    • Reference Methods/Predicate Devices: The method comparison studies demonstrate equivalence to an established, legally marketed predicate device (Coamatic Heparin assay on the IL ACL TOP®), which itself would have been validated against accepted reference methods and standards.
    • Known Concentrations: Linearity and interference studies use samples spiked with known concentrations of UFH, LMWH, and potential interferents.
    • Statistical Analysis: CLSI guidelines (e.g., EP06-A, EP07-A2, EP05-A2, EP09-A3) are used for statistical evaluation of analytical performance, ensuring robust and validated methodologies for assessing precision and accuracy.

    8. The sample size for the training set

    This document does not specify a "training set" in the context of machine learning. The device is a traditional in-vitro diagnostic assay, not an AI/ML algorithm that undergoes a distinct training phase on a large dataset. The "training" for such devices typically involves the development and optimization of the reagent formulations, instrument parameters, and calibration procedures, which is an engineering and chemistry development process rather than a data-driven training of a model.

    9. How the ground truth for the training set was established

    As there is no "training set" in the AI/ML sense, this question is not applicable. The development and calibration of the assay components (reagents, calibrators, controls) are established through rigorous analytical chemistry and metrology principles, tracing back to international reference standards and validated against known concentrations and predicate devices.

    Ask a Question

    Ask a specific question about this device

    K Number
    K111822

    Validate with FDA (Live)

    Date Cleared
    2011-10-26

    (120 days)

    Product Code
    Regulation Number
    864.7525
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    -------|
    | STA® - Liquid Anti-Xa | Heparin Assay | Heparin Assay21 CFR 864.7525
    Liquid Anti-Xa STA®-Multi Hep Calibrator STA®-Quality HNF/UFH STA®-HBPM/LMWH Regulation Number: 21 CFR § 864.7525

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The STA® - Liquid Anti-Xa kits are intended for use with STA-R®, STA Compact® and STA Satellite® analyzers, for the quantitative determination of the plasma levels of unfractionated (UFH) and low molecular weight (LMWH) heparins by measuring their anti-Xa activity on antithrombin in a competitive assay using a synthetic chromogenic substrate.

    The STA® - Multi Hep Calibrator is a set of calibrator plasmas intended for use with STA-R®, STA Compact®, and STA Satellite® analyzers, for the calibration of heparin (UFH and LMWH) activity assay by measuring the anti-Xa activity.

    The STA® - Quality HNF/UFH kit is a set of two plasmas intended for the quality control of unfractionated heparin (UFH) activity assay by measuring the anti-Xa activity performed on STA-R®, STA Compact®, and STA Satellite® analyzers.

    The STA® - Quality HBPM/LMWH kit is a set of two plasmas intended for the quality control of low molecular weight heparin (LMWH) activity assay by measuring the anti-Xa activity performed on STA-R®, STA Compact, and STA Satellite® analyzers.

    Device Description

    The STA® - Liquid Anti-Xa is a chromogenic assay technique used for determination of the level of UFH and LMWH that have high affinity for antithrombin by measuring their anti-Xa activity. The method is a one-step reaction based on a similar principle: as soon as factor Xa is added to the plasma-substrate mixture, two reactions take place simultaneously, namely, hydrolysis of the substrate by factor Xa and inhibition of factor Xa by the heparin-antithrombin complex. After the necessary period of time for the competitive reaction to reach equilibrium, the quantity of paranitroaniline that is released is inversely proportional to the concentration of heparin present in the test medium.

    The STA® - Multi Hep Calibrator reagents are lyophilized human plasmas at five different heparin concentrations. They are used to create the calibration curve on STA-R®, STA Compact®, and STA Satellite® analyzers performing the chromogenic method for heparin (UFH and LMWH) using STA® - Liquid Anti-Xa.

    The STA® - Quality HNF/UFH reagents are lyophilized human plasmas at two different UFH concentrations. They are used for the quality control of UFH activity assay by measuring the anti-Xa activity using the chromogenic method STA® - Liquid Anti-Xa performed on STA-R®, STA Compact®, and STA Satellite® analyzers.

    The STA® - Quality HBPM/LMWH reagents are lyophilized human plasmas at two different LMWH concentrations. They are used for the quality control of LMWH activity assay by measuring the anti-Xa activity using the chromogenic methods, STA® - Liquid Anti-Xa and STA® - Rotachrom® Heparin, performed on STA-R®, STA Compact, and STA Satellite® analyzers.

    AI/ML Overview

    The provided document describes the 510(k) summary for several in vitro diagnostic devices for measuring heparin activity. The document focuses on demonstrating substantial equivalence to predicate devices through performance characteristics, rather than establishing acceptance criteria and proving them with a specific study with a defined ground truth, as would be common for AI/ML device submissions.

    Therefore, many of the requested elements (e.g., test set sample size, data provenance, number of experts for ground truth, adjudication method, MRMC study, training set details) are not applicable or not explicitly detailed in this type of submission.

    However, I can extract information related to performance characteristics that serve as "acceptance criteria" for the device, and the studies performed to demonstrate these.

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document describes performance characteristics of the STA® - Liquid Anti-Xa device (and its associated calibrator and controls) through precision and detection limit/working range studies. The "acceptance criteria" are implied by the ranges and values obtained, showing performance similar or superior to predicate devices or within acceptable analytical limits for IVD assays.

    Performance CharacteristicAcceptance Criteria (Implied)Reported Device PerformanceComments
    Precision (Hybrid Calibration - 5-point UFH/LMWH)Coefficient of Variation (CV%) values within acceptable analytical limits for heparin assays. Specific numerical criteria are not explicitly stated, but common industry standards for IVD precision apply.UFH Samples:- Sample 1 (mean 0.21 IU/mL): Repeatability CV 6.2%, Within-lab CV 9.9%- Sample 2 (mean 0.55 IU/mL): Repeatability CV 3.1%, Within-lab CV 6.6%- Sample 3 (mean 0.97 IU/mL): Repeatability CV 3.4%, Within-lab CV 5.5%LMWH Samples:- Sample 4 (mean 0.86 IU/mL): Repeatability CV 3.1%, Within-lab CV 4.8%- Sample 5 (mean 1.48 IU/mL): Repeatability CV 3.0%, Within-lab CV 5.1%- Sample 6 (mean 1.75 IU/mL): Repeatability CV 2.9%, Within-lab CV 5.0%Performed according to CLSI guideline EP5-A2 over 22 days, 2 runs/day.
    Precision (Dedicated Calibration - 3-point UFH)CV% values within acceptable analytical limits.UFH Samples:- Sample 1 (mean 0.22 IU/mL): Repeatability CV 5.6%, Within-lab CV 9.2%- Sample 2 (mean 0.55 IU/mL): Repeatability CV 3.0%, Within-lab CV 6.1%- Sample 3 (mean 0.97 IU/mL): Repeatability CV 3.5%, Within-lab CV 5.1%Similar study design to hybrid calibration.
    Precision (Dedicated Calibration - 3-point LMWH)CV% values within acceptable analytical limits.LMWH Samples:- Sample 4 (mean 0.86 IU/mL): Repeatability CV 3.2%, Within-lab CV 5.2%- Sample 5 (mean 1.48 IU/mL): Repeatability CV 3.1%, Within-lab CV 5.3%- Sample 6 (mean 1.75 IU/mL): Repeatability CV 2.8%, Within-lab CV 5.1%Similar study design to hybrid calibration.
    Detection Limit (UFH/LMWH 5-point calibration)Detection threshold should be clinically relevant for heparin monitoring.0.10 IU/mL (UFH and LMWH)Assessed according to CLSI guideline EP17-A.
    Linearity Range (UFH/LMWH 5-point calibration)Linearity range should cover the clinically relevant range for heparin monitoring.UFH: up to 1.10 IU/mL; LMWH: up to 2.00 anti-Xa IU/mLAssessed according to CLSI guideline EP6-A.
    Detection Limit (UFH 3-point calibration)Detection threshold should be clinically relevant.0.10 IU/mLAssessed according to CLSI guideline EP17-A.
    Linearity Range (UFH 3-point calibration)Linearity range should cover the clinically relevant range.up to 1.10 IU/mLAssessed according to CLSI guideline EP6-A.
    Detection Limit (LMWH 3-point calibration)Detection threshold should be clinically relevant.0.10 anti-Xa IU/mLAssessed according to CLSI guideline EP17-A.
    Linearity Range (LMWH 3-point calibration)Linearity range should cover the clinically relevant range.up to 2.00 anti-Xa IU/mLAssessed according to CLSI guideline EP6-A.
    Interfering SubstancesNo significant interference from common endogenous substances at clinically relevant concentrations.Insensitive to: hemoglobin (up to 1.5 g/l), conjugated bilirubin (up to 288 mg/l), unconjugated bilirubin (up to 138 mg/l), triglycerides (up to 6.9 g/l).Performed according to CLSI guideline EP7-A2.

    2. Sample size used for the test set and the data provenance:

    • Test set sample size: For precision studies, 6 heparin samples (3 UFH, 3 LMWH) were used for each calibration type (hybrid and dedicated). For the detection limit and linearity studies, specific sample numbers are not provided but are implicit in the CLSI guidelines (EP17-A and EP6-A) which typically involve multiple replicates across different concentrations. For interfering substances, the number of samples is not explicitly stated, but the study tested specific concentrations of key interferents.
    • Data provenance: Not explicitly stated, but given this is an in vitro diagnostic device for global markets, the samples would likely be prepared laboratory controls and patient plasma samples (for linearity, detection limit, and interference studies), not geographically defined "countries of origin" in the same way as imaging data. The studies are prospective in the sense that they are designed experiments to validate performance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable (N/A) in the context of this IVD device. The "ground truth" for these types of assays is established by the known concentrations of calibrators, controls, and spiked samples, or by reference methods, not by expert interpretation.

    4. Adjudication method for the test set:

    • N/A. Adjudication is not relevant for analytical performance studies of quantitative IVD assays.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This is an in vitro diagnostic device, not an AI/ML-driven interpretive device typically associated with MRMC studies.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • Yes, this is a standalone device. The performance characteristics (precision, detection limits, linearity, interference) are determined for the algorithm/device only. Human involvement is in operating the analyzer and interpreting the numerical results, but the analytical performance itself is inherent to the device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For precision, linearity, and detection limit, the ground truth is established by the known concentrations of the calibrators and control plasmas, or by accepted reference methods for heparin activity (e.g., anti-Xa activity).
    • For interfering substances, the ground truth is the addition of specific interfering agents at known concentrations to samples, and then assessing if the device's measurement of heparin is accurately maintained.

    8. The sample size for the training set:

    • Not applicable (N/A) in the AI/ML sense. This is a traditional IVD device using established chromogenic assay principles, not an AI/ML system that requires a "training set" to learn. The method relies on biochemical reactions and quantitative measurement.

    9. How the ground truth for the training set was established:

    • N/A for the same reason as point 8.
    Ask a Question

    Ask a specific question about this device

    K Number
    K092937

    Validate with FDA (Live)

    Date Cleared
    2010-08-02

    (312 days)

    Product Code
    Regulation Number
    862.1150
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    AUG 0 2 2010

    Re: K092937

    Trade/Device Name: STA® - Hybrid Hep Calibrator Regulation Number: 21 CFR 864.7525

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The STA® - Hybrid Hep Calibrator is a set of calibrator plasmas intended for use with analyzers of the STA® line suitable to these reagents, for the calibration of heparin (UFH and LMWH) activity assay by measuring the anti-Xa activity using the chromogenic method, STA® - Rotachrom® Heparin.

    Device Description

    The STA® Hybrid Hep Calibrator is a set of lyophilized human plasmas used to create the calibration curve on the STA® line of IVD instruments performing the chromogenic method for heparin (UFH and LMWH) assays. Each STA® Hybrid Hep Calibrator available contains: 4 x 1-ml vials of Reagent 1: STA® Hybrid Hep Calibrator O lyophilized human plasma free of heparin. 4 x 1-ml vials of Reagent 2: STA® Hybrid Hep Calibrator 3 lyophilized human plasma containing a well-defined quantity of UFH. 4 x 1-ml vials of Reagent 3: STA® Hybrid Hep Calibrator 6 lyophilized human plasma containing a well-defined quantity of UFH that is greater than that of Reagent 2. 4 x 1-ml vials of Reagent 4: STA® - Hybrid Hep Calibrator 9 lyophilized human plasma containing a well-defined quantity of LMWH. 4 x 1-ml vials of Reagent 5: STA® Hybrid Hep Calibrator 18 lyophilized human plasma containing a well-defined quantity of LMWH that is greater than that of Reagent 4.

    AI/ML Overview

    The provided text describes a 510(k) summary for the STA® - Hybrid Hep Calibrator, a medical device. This submission seeks to bundle two previously cleared devices (STA® - Hepanorm® H and STA® - Calibrator HBPM/LMWH Kit) into a single kit. The document focuses on demonstrating substantial equivalence to these predicate devices for regulatory purposes.

    Therefore, the document does not contain information about:

    • Specific acceptance criteria for device performance (e.g., accuracy, precision) as typically defined in a clinical study.
    • A formal study that proves the device meets specific acceptance criteria in terms of measured performance metrics.
    • Sample sizes for test sets, data provenance, number of experts for ground truth, adjudication methods, MRMC comparative effectiveness studies, standalone performance, or grand truth types.
    • Training set sample sizes or how their ground truth was established, as this is a calibrator device and not an AI/ML diagnostic system.

    The document states that the STA® - Hybrid Hep Calibrator and its predicate devices are "identical products regarding indication/intended use, formulation or materials of construction and design, technology, and safety." The primary difference is the bundling of two existing calibrators into one kit. The substantial equivalence argument relies on the fact that no new questions of safety, effectiveness, or technology are raised due to this bundling.

    Essentially, the "study" proving the device meets acceptance criteria is implied by its substantial equivalence to previously cleared devices. The acceptance criteria, in this regulatory context, are primarily that the bundled calibrator performs equivalently to the two separate predicate calibrators, which have already been deemed safe and effective.

    Summary of available information related to device performance and regulatory acceptance:

    Acceptance Criteria (Implied by Substantial Equivalence)Reported Device Performance (Implied by Substantial Equivalence)
    Performs calibration for heparin (UFH and LMWH) activity assays by measuring anti-Xa activity using the chromogenic method (STA® - Rotachrom® Heparin).The STA® - Hybrid Hep Calibrator is "identical" to the predicate devices (STA® - Calibrator HBPM/LMWH Kit and STA® - Hepanorm® H) regarding indication/intended use, technology, and principles of operation.
    Operates with STA® line of IVD instruments.Designed to operate utilizing the STA® product line of IVD coagulation analyzers for creating calibration curves for heparin assays.
    Ensures safety and effectiveness for its intended use.Yields "no new questions in safety, effectiveness, or technology" compared to the predicate devices.

    Information not available in the provided text:

    • Sample sized used for the test set and the data provenance: Not applicable in this context as this is a calibrator, not a diagnostic algorithm. Performance is assessed through equivalence to existing calibrators.
    • Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable.
    • Adjudication method for the test set: Not applicable.
    • If a multi reader multi case (MRMC) comparative effectiveness study was done: Not applicable.
    • If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable as this is a calibrator device, not an algorithm.
    • The type of ground truth used: For calibrators, ground truth typically refers to the accurately assigned values of the analytes (UFH and LMWH in this case) within the calibrator plasmas. The text states the plasmas contain "well-defined quantity of UFH" and "well-defined quantity of LMWH", implying these values are established and verified during manufacturing of the calibrator materials.
    • The sample size for the training set: Not applicable as this is not an AI/ML device.
    • How the ground truth for the training set was established: Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K090209

    Validate with FDA (Live)

    Date Cleared
    2009-06-02

    (125 days)

    Product Code
    Regulation Number
    864.7525
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Controls

    Device Regulatory Information:

    | Heparin Assay: | Class II | Product Code: KFF | 21 CFR 864.7525
    Calibrators, HemosIL LMW Heparin Controls and HemosIL UF Heparin Controls

    Regulation Number: 21 CFR §864.7525

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    • HemosIL Liquid Heparin: Automated chromogenic assay for the quantitative determination of . unfractionated heparin (UFH) and low molecular weight heparin (LMWH) activity in human citrated plasma on IL Coagulation Systems (ACL TOP® Family, ACL™ ELITE/ELITE PRO®/8/9/10000 and ACL Futura/ACL Advance Systems).
    • HemosIL Heparin Calibrators: For the calibration of the HemosIL Liquid Heparin assay on IL . Coagulation Systems (ACL TOP® Family, ACL™ ELITE/ ELITE PRO®/8/9/10000 and ACL Futura/ACL Advance Systems).
    • HemosIL LMW Heparin Controls (Assayed): For the quality control of the HemosIL Liquid . Heparin assay when testing for low molecular weight heparin (LMW) on IL Coagulation Systems (ACL TOP® Family, ACL" ELITE/ ELITE PRO®8/9/10000 and ACL Futura/ACL Advance Systems).
    • · HemosIL UF Heparin Controls (Assayed): For the quality control of the HemosIL Liquid Heparin assay when testing for unfractionated heparin (UFH) on IL Coagulation Systems (ACL TOP® Family, ACL™ ELITE/ ELITE PRO®/8/9/10000 and ACL Futura/ACL Advance Systems).
      For in vitro diagnostic use.
    Device Description

    HemosIL Liquid Heparin .
    One stage chromogenic assay based on a synthetic chromogenic substrate and on Factor Xa inactivation. Heparin levels in patient plasma are measured automatically on IL Coagulation Systems.
    Heparin is analyzed as a complex with antithrombin present in the sample. The concentration of this complex is dependent on the availability of the patient's endogenous antithrombin. When the Heparin - antithrombin complex is formed, two competing reactions take place.

      1. Factor Xa is neutralized by heparin-antithrombin complex.
      1. Residual Factor Xa is quantified with a synthetic chromogenic substrate. The paranitroaniline released is monitored kinetically at 405 nm and is inversely proportional to the heparin level in the sample.
    • HemosIL Heparin Calibrators .
      Lyophilized calibrators prepared from human citrated plasma by means of a dedicated process at three different heparin concentrations: 0, 0.8 and 2.0 IU/mL and are traceable to the WHO International Standards for LMW and UF Heparin.
    • HemosIL LMW Heparin Controls (Assayed) .
      Lyophilized controls prepared from human citrated plasma by means of a dedicated process at two different LMW heparin concentrations (low and high) for the assessment of precision and accuracy of the Liquid Heparin assay when testing for low molecular weight heparin.
    • HemosIL UF Heparin Controls (Assayed) .
      Lyophilized controls prepared from human citrated plasma by means of a dedicated process at two different UF heparin concentrations (low and high) for the assessment of precision and accuracy of the Liquid Heparin assay when testing for unfractionated heparin.
    AI/ML Overview

    HemosIL Liquid Heparin Assay, Controls, and Calibrators - Acceptance Criteria & Study Summary

    This document describes the acceptance criteria and supporting studies for the HemosIL Liquid Heparin Assay, HemosIL Heparin Calibrators, HemosIL LMW Heparin Controls, and HemosIL UF Heparin Controls, as extracted from the provided 510(k) summary (K090209).

    1. Acceptance Criteria and Reported Device Performance

    The 510(k) submission primarily demonstrates substantial equivalence through method comparison studies and precision assessments. The acceptance criteria for substantial equivalence are implicitly defined by comparing the new device's performance to that of predicate devices and demonstrating acceptable precision.

    The following table summarizes the reported device performance, particularly focusing on precision and method comparison results, which serve as the evidence for meeting implied acceptance criteria for assay performance characteristics relevant to quantitative diagnostic devices. Specific, explicit acceptance criteria (e.g., "CV % must be < X%") are not directly stated in the provided text but are inferred from the presentation of the precision data and the successful substantial equivalence determination by the FDA. The method comparison data implies acceptance if the slope and correlation coefficient (r) are sufficiently close to 1, demonstrating agreement with the predicate.

    Table 1: Acceptance Criteria (Implied) and Reported Device Performance

    Performance CharacteristicAcceptance Criteria (Implied)Reported Device Performance
    PrecisionCV % (Within run) and CV % (Total) for UFH and LMWH controls should be within acceptable limits for a clinical diagnostic assay.ACL 8/9/1000/ELITE/ELITE PRO: - UFH Low: Mean 0.39 IU/mL, C.V.% (Within run) 4.0%, C.V.% (Total) 6.7% - UFH High: Mean 0.69 IU/mL, C.V.% (Within run) 1.0%, C.V.% (Total) 3.7% - LMWH Low: Mean 0.49 IU/mL, C.V.% (Within run) 5.5%, C.V.% (Total) 5.7% - LMWH High: Mean 1.31 IU/mL, C.V.% (Within run) 3.3%, C.V.% (Total) 4.0% ACL Futura/ACL Advance: - UFH Low: Mean 0.41 IU/mL, C.V.% (Within run) 4.4%, C.V.% (Total) 4.4% - UFH High: Mean 0.69 IU/mL, C.V.% (Within run) 1.3%, C.V.% (Total) 1.7% - LMWH Low: Mean 0.56 IU/mL, C.V.% (Within run) 2.5%, C.V.% (Total) 3.5% - LMWH High: Mean 1.37 IU/mL, C.V.% (Within run) 1.1%, C.V.% (Total) 1.6% ACL TOP Family: - UFH Low: Mean 0.41 IU/mL, C.V.% (Within run) 2.4%, C.V.% (Total) 3.3% - UFH High: Mean 0.68 IU/mL, C.V.% (Within run) 1.0%, C.V.% (Total) 1.6% - LMWH Low: Mean 0.55 IU/mL, C.V.% (Within run) 3.5%, C.V.% (Total) 4.5% - LMWH High: Mean 1.35 IU/mL, C.V.% (Within run) 1.9%, C.V.% (Total) 2.5%
    Method Comparison (Slope)Slope should be close to 1.0, indicating good agreement with the predicate device.In-house Study: - ACL ELITE: 0.894 - ACL Advance: 1.067 - ACL TOP: 0.946 Field Sites Study: - ACL ELITE: 1.032 - ACL Advance: 1.007 - ACL TOP: 0.952
    Method Comparison (r)Correlation coefficient (r) should be close to 1.0, indicating a strong linear relationship with the predicate device.In-house Study: - ACL ELITE: 0.907 - ACL Advance: 0.946 - ACL TOP: 0.958 Field Sites Study: - ACL ELITE: 0.949 - ACL Advance: 0.957 - ACL TOP: 0.978
    LinearityDevice should be linear up to 2.0 IU/mL for UFH and LMWH activity across all specified IL Coagulation Systems.Reported linearity for the new device is Up to 2.0 IU/mL for ACL 8/9/1000/ELITE/ELITE PRO, ACL Futura/ACL Advance, and ACL TOP Family. This is an improvement over the predicate device, which had linearity up to 1.0 IU/mL or 1.1 IU/mL depending on the system.

    2. Sample Sizes and Data Provenance for the Test Set

    The studies conducted for performance assessment include precision and method comparison.

    • Precision: "Precision was assessed over multiple runs using the two levels of both UFH and LMWH Controls on representative IL Coagulation Systems." The exact number of samples (individual test results) for precision is not explicitly stated, but it involved multiple runs of two levels of controls for both UFH and LMWH.
    • Method Comparison - In-house:
      • Sample sizes (n):
        • ACL ELITE: 124 samples
        • ACL Advance: 152 samples
        • ACL TOP: 148 samples
      • Data Provenance: "samples from patients undergoing heparin therapy." The country of origin is not specified, but the "in-house" designation suggests it was likely from a single internal site, possibly in the USA (where the applicant is located). The data is retrospective in the sense that these are patient samples, but the comparison itself is a prospective analysis of these samples using both devices.
    • Method Comparison - Field Sites:
      • Sample sizes (n):
        • ACL ELITE: 114 samples
        • ACL Advance: 111 samples
        • ACL TOP: 81 samples
      • Data Provenance: "samples from patients undergoing heparin therapy." Three field site studies were performed. The country of origin is not specified, but multi-site studies often involve diverse populations. The data is retrospective in the sense that these are patient samples, but the comparison itself is a prospective analysis of these samples using both devices.

    3. Number of Experts and Qualifications for Ground Truth

    The provided summary does not explicitly mention the use of experts to establish a "ground truth" in the typical sense (e.g., radiological reads). For this type of in vitro diagnostic device (quantitative assay for heparin activity), the "ground truth" for the method comparison studies is typically established by the predicate device's results. The predicate device (HemosIL Heparin) is considered the reference standard against which the new device's performance is compared. Therefore, no external experts were used in this manner.

    4. Adjudication Method for the Test Set

    Since the ground truth for this type of quantitative assay is established by the reading of the predicate device, there is no mention of an adjudication method involving multiple human readers as described (e.g., 2+1, 3+1). The comparison is directly between the numerical outputs of the new device and the predicate device.

    5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study was mentioned. This type of study is more common for imaging or interpretive diagnostic devices where human readers provide interpretations. For quantitative assays like the HemosIL Liquid Heparin, the focus is on the agreement between the new test results and a reference method/device, and the precision of the new test, rather than human reader improvement with AI assistance.

    6. Standalone Performance Study (Algorithm Only)

    The entire study focuses on the standalone performance of the HemosIL Liquid Heparin assay (an algorithm-driven instrument-based test) compared to its predicate and its internal precision. There is no human-in-the-loop performance described, as it is an automated chromogenic assay run on IL Coagulation Systems. Therefore, yes, a standalone (algorithm only) performance study was conducted.

    7. Type of Ground Truth Used

    The ground truth for the method comparison studies was established by the predicate device (HemosIL Heparin). For precision studies, the ground truth for the controls is their assigned nominal values, and the study assesses the reproducibility of the new device's measurements around these values.

    8. Sample Size for the Training Set

    The provided 510(k) summary does not mention a "training set" in the context of machine learning or AI development. This device is an automated chromogenic assay; its performance is based on chemical reactions and optical detection, not a machine learning model that requires training data in the typical sense. Data used for initial development or optimization of the assay formulation would not be referred to as a "training set" in this context.

    9. How Ground Truth for the Training Set Was Established

    As there is no "training set" in the context of an AI/ML algorithm for this device, this question is not applicable. The assay's development would involve standard analytical chemistry and immunoassay development pipelines, with "ground truth" being established through reference materials, spiking studies, and established analytical methods.

    Ask a Question

    Ask a specific question about this device

    K Number
    K050367

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2006-05-24

    (464 days)

    Product Code
    Regulation Number
    864.7525
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    45040

    Re: K050367

    Trade/Device Name: BIOPHEN Heparin 3, BIOPHEN Heparin 6 Regulation Number: 21 CFR 864.7525

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Biophen Heparin is an in vitro diagnostic test for the quantitative determination of unfractionated heparin (UFH) activity in human citrated plasma using automated or manual methods. The nopant of UFH is determined from the Anti factor Xa (anti-FXa) activity expressed by the [AT* Heparin] complex formed in plasma. Heparin is used for curative or preventive indications. Measuring Heparin concentration in patient's plasma allows monitoring therapy and adjusting drug dosage.

    Device Description

    Not Found

    AI/ML Overview

    The provided text is a 510(k) premarket notification letter from the FDA for a heparin assay device. It does not contain the detailed study information regarding acceptance criteria, device performance, sample sizes, ground truth establishment, or multi-reader multi-case studies typically associated with AI/ML device evaluations. Such information is usually found in the 510(k) summary or full submission, not in the FDA's decision letter.

    Therefore, I cannot fulfill your request for a table of acceptance criteria, device performance, sample sizes, expert qualifications, adjudication methods, MRMC studies, standalone performance, ground truth types, or training set details from the given text.

    The document primarily states that the device is substantially equivalent to a predicate device for the quantitative determination of unfractionated heparin (UFH) activity in human citrated plasma.

    Ask a Question

    Ask a specific question about this device

    K Number
    K013305

    Validate with FDA (Live)

    Date Cleared
    2002-08-23

    (323 days)

    Product Code
    Regulation Number
    864.7525
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Coag Enoxaparin (ENOX) Test Card

    Common Name: ENOX Test Card

    Classification Name: CFR Section 21 CFR 864.7525
    27560

    Re: K013305

    Trade/Device Name: TAS ™ Enoxaparin (ENOX)™ Test Card Regulation Number: 21 CFR 864.7525

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Rapidpoint™Coag Enoxaparin Test card (ENOX) is a qualitative test intended for exclusive use with the Rapidpoint Coag analyzer to detect the anticoagulant effects, ≥ 1.0 IU/ml, of the low molecular weight heparin (LMWH), Lovenox®/Clexane® (enoxaparin sodium)¹, in arterial citrated whole blood from patients with unstable angina (UA)/non-ST segment elevation myocardial infarction (NSTEMI) who may transition to percutaneous intervention (PCI). The ENOX test is intended for use at either the point of care or in the central laboratory. The device does not discriminate between values of enoxaparin below 1.0 IU/ml and the absence of drug.

    The test provides information on the patient's citrated arterial whole blood response to enoxaparin by measurement of the clotting time using a factor Xa activated clotting method and should be interpreted in conjunction with other clinical data available to the clinician

    Device Description

    The RapidPoint Enoxaparin (ENOX) test is a one-step dry coagulation method performed on the RapidPoint Coag analyzer. All of the components necessary to perform the assay, with the exception of patient sample, are included in the reaction chamber of the test card.

    In the ENOX test, factor X is rapidly converted to factor Xa by a specific factor X activator initiating the clotting process. Enoxaparin, from the patient's blood, complexes with antithrombin (AT), to inhibit factor Xa and lengthen the clotting time. Reported clotting times in excess of the assay cut-off indicate an enoxaparin concentration greater than or equal to 1.0 International Units per milliliter. The results generated by the ENOX test are indicative of the anticoagulant effect produced by enoxaparin in citrated arterial whole blood.

    The test card formulation contains purified Factor Xa activator, calcium, phospholipid and stabilizers. Paramagnetic iron oxide particles (PIOP) are included to provide an optical detection mechanism in the presence of patient sample.

    AI/ML Overview

    Here's a summary of the acceptance criteria and study details for the Thrombolytic Assessment System (TAS) or Rapidpoint™ Coag Enoxaparin (ENOX) Test Card.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implicitly defined by the clinical study's results in relation to the intended use. The device is intended to detect enoxaparin levels ≥1.0 IU/mL. The study uses a clotting time cutoff of ≥ 260 seconds to indicate this.

    Acceptance Criteria (Implied)Reported Device Performance (Clinical Study)
    Detect enoxaparin ≥1.0 IU/mL (indicated by clotting time ≥ 260 seconds)Sensitivity: 96.3% (95% CI: 89.4-99.2)
    Discriminate enoxaparin <1.0 IU/mL or absence of drug (indicated by clotting time < 260 seconds)Specificity: 25% (95% CI: 3.2-65.1)

    2. Sample Size and Data Provenance for the Test Set

    • Sample Size (Clinical Study): 88 patients
    • Data Provenance: Prospective, from "6 sites in the United States."
    • Sample Type: Arterial citrated whole blood samples, collected at baseline (no drug) and at peak values (within 15 minutes of dosing) after enoxaparin administration. The samples were tested in duplicate.

    3. Number of Experts and Qualifications for Ground Truth of the Test Set

    The document does not explicitly state the number of experts used to establish ground truth for the test set.

    • Ground Truth Method: The ground truth for the clinical study was established by comparing the ENOX card clot times to "central laboratory measured plasma anti-Xa levels." This implies that the central lab results, likely interpreted by qualified laboratory personnel, served as the reference standard. The qualifications of those performing the anti-Xa level measurements are not specified, but it's assumed to be standard clinical laboratory practice.

    4. Adjudication Method for the Test Set

    The document does not mention an adjudication method for the test set. The comparison was made directly between the device's output and the central lab's anti-Xa results.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not performed. This device is an in-vitro diagnostic (IVD) assay designed to provide a direct measurement, not an imaging or diagnostic aid that human readers would interpret to improve their performance with or without AI assistance. Therefore, there is no effect size reported for human readers improving with or without AI.

    6. Standalone (Algorithm Only) Performance

    Yes, the provided data describes the standalone performance of the algorithm (the ENOX Test Card used with the RapidPoint Coag analyzer). The sensitivity and specificity reported are for the device itself, without human interpretation influencing the final "positive" or "negative" determination based on the 260-second cutoff.

    7. Type of Ground Truth Used

    The type of ground truth used for the clinical study was: central laboratory measured plasma anti-Xa levels.

    8. Sample Size for the Training Set

    The document does not explicitly state the sample size for a "training set." The performance data presented focuses on the clinical validation of the device. However, "In-House Data" for reproducibility used unspecified numbers of "citrated plasma with and without the addition of 1.0 (anti-Xa) IU/ml enoxaparin" and "venous citrated whole blood from single donors." These in-house studies likely contributed to the optimization of the assay (e.g., establishing the 260-second cutoff), which could be considered an internal development/training phase.

    9. How the Ground Truth for the Training Set Was Established

    For the in-house reproducibility studies (which might loosely correspond to an internal development/training phase):

    • Ground Truth Method: Ground truth appears to have been established by directly spiking known concentrations of enoxaparin (0.0 IU/ml and 1.0 IU/ml) into citrated plasma or venous whole blood. This provides a clear, controlled, and known ground truth for evaluating the device's basic performance and reproducibility at specific enoxaparin levels.
    Ask a Question

    Ask a specific question about this device

    K Number
    K013637

    Validate with FDA (Live)

    Date Cleared
    2002-04-16

    (162 days)

    Product Code
    Regulation Number
    864.7525
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    2002

    Name of the Device:

    ACTICHROME® Heparin (anti-fIIa) Product No. 820

    Classification:

    864.7525
    Re: K013637

    Trade/Device Name: ACTICHROME® Heparin Assay (anti-fIIa) Regulation Number: 21 CFR § 864.7525

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ACTICHROME® Heparin (anti-fila) is a chromogenic assay intended for the quantitative determination of therapeutic heparin in human plasma by measurement of factor IIa (thrombin) inhibition. The Electra 900C® was used to determine performance data. This kit is for in vitro diagnostic use.

    Device Description

    ACTICHROME® Heparin (anti-fIIa) Product No. 820

    AI/ML Overview

    The provided text describes a 510(k) summary for the ACTICHROME® Heparin (anti-fIIa) assay. It focuses on demonstrating substantial equivalence to a predicate device, primarily through method comparison and precision studies.

    Here's an analysis based on your requested information:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't explicitly state "acceptance criteria" in a numerical or categorical format for demonstrating substantial equivalence. Instead, it relies on a strong "positive correlation" with the predicate device and acceptable precision (CV%).

    Performance MetricAcceptance Criteria (Implied/General)Reported Device Performance (ACTICHROME Heparin (anti-fIIa))
    Method ComparisonStrong positive correlation (e.g., R-value > 0.9) and regression equation demonstrating agreement with predicate.Lot 010 (N=88): R = 0.967, Y = 0.845X + 0.027
    Lot 010 (N=32): R = 0.917, Y = 0.764X + 0.032
    Sy.x (ng/ml) = 0.03 (N=88), 0.05 (N=32)
    Intra-Assay Precision (CV%)Low variability (e.g., typically <10% for clinical assays, depending on concentration)Hepanorm Control 6: 4.7%
    Hepanorm Control 3: 10.8%
    Spiked Plasma (0.5 U/ml): 3.8%
    Spiked Plasma (0.25 U/ml): 9.5%
    Inter-Assay Precision (CV%)Low variability (e.g., typically <15% for clinical assays, depending on concentration)Hepanorm Control 6: 7.6%
    Hepanorm Control 3: 9.6%
    Spiked Plasma (0.5 U/ml): 8.5%
    Spiked Plasma (0.25 U/ml): 9.4%

    Study Proving Device Meets Criteria:

    The studies described in the "Summary of Performance Data" are intended to prove the device meets these (implied) substantial equivalence criteria. These include:

    • Method Comparison Study: Compared the ACTICHROME Heparin (anti-fIIa) assay results against the predicate device (Spectrolyse® Heparin (anti-IIa)).
    • Precision Studies: Evaluated intra-assay and inter-assay variability using control samples.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Method Comparison Test Set:

      • Sample Sizes: 88 samples (first regression) and 32 samples (second regression). It's unclear if these are independent sets or subsets.
      • Data Provenance: The document does not explicitly state the country of origin or whether the data was retrospective or prospective. It refers to "human plasma," which implies the samples were derived from human subjects. The fact that the studies were performed as part of a 510(k) submission prior to market launch suggests they were prospective in nature for validation purposes.
    • Precision Test Set:

      • Hepanorm Controls: N=80 per control (for intra-assay and inter-assay with 20 runs).
      • Spiked Plasma Controls: N=40 per control (for intra-assay and inter-assay with 10 runs).
      • Data Provenance: Not explicitly stated regarding country of origin or retrospective/prospective. Given the nature of precision studies for assay validation, these would typically be performed prospectively in a laboratory setting.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    Not applicable. This device is an in vitro diagnostic (IVD) assay for quantitative determination. The "ground truth" for method comparison is the measurement obtained from the predicate device, and for precision, it's the statistical variation around the measured value of control samples. There are no human experts involved in establishing a "ground truth" in the way radiologists or pathologists establish it for imaging or histology.

    4. Adjudication Method for the Test Set:

    Not applicable. As noted above, this is an IVD assay, not a device requiring human interpretation adjudicated by experts. The "adjudication" is based on the comparison of quantitative results.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done:

    No, an MRMC comparative effectiveness study was not done. This type of study is typically conducted for diagnostic imaging or screening devices where human readers interpret results, and the impact of an AI algorithm on their performance is evaluated. The ACTICHROME Heparin (anti-fIIa) assay is a laboratory-based quantitative measurement device.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done:

    Yes, the studies presented are essentially "standalone" performance evaluations of the ACTICHROME Heparin (anti-fIIa) assay. The performance data (method comparison, precision) reflects the assay's output directly, without a human in the loop interpreting the result to reach a diagnosis or making a decision. The Electra 900C® (an analyzer) was used to determine the performance data, indicating automated measurement.

    7. The Type of Ground Truth Used:

    • For Method Comparison: The "ground truth" for comparison was the results obtained from the predicate device, Spectrolyse® Heparin (anti-IIa). This is a form of reference method comparison.
    • For Precision: The "ground truth" implicitly used for evaluating precision are the expected values and inherent variability of the control samples themselves.

    8. The Sample Size for the Training Set:

    Not applicable. This document describes an IVD assay that is likely based on established chromogenic assay principles, rather than a machine learning model that requires a "training set" in the computational sense. The "development" of the assay likely involved optimizing reagents and protocols, but not training an algorithm on a 'dataset'.

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable, as there is no "training set" in the context of a machine learning algorithm. The development of the assay would have involved standard chemical and biological assay development practices, not machine learning ground truth establishment.

    Ask a Question

    Ask a specific question about this device

    K Number
    K013318

    Validate with FDA (Live)

    Date Cleared
    2002-03-08

    (154 days)

    Product Code
    Regulation Number
    864.7525
    Age Range
    0 - 21
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Device:

    ACTICHROME® Heparin (anti-fXa) Product No. 832

    Classification:

    Heparin Assay, Class II 864.7525
    CT 06830

    K013318 Re:

    Trade/Device Name: ACTICHROME® Heparin (anti-fXa) Regulation Number: 21 CFR 864.7525

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ACTICHROME® Heparin (anti-fXa) is a chromogenic assay intended for the quantitative determination of unfractionated and low molecular weight heparins in human plasma. The assay measures the inhibition of factor Xa (fXa) activity by the various heparins. The Electra 900C® was used to determine performance data. This kit is for in vitro diagnostic use.

    Device Description

    ACTICHROME® Heparin (anti-fXa) is a chromogenic assay intended for the quantitative determination of unfractionated and low molecular weight heparins in human plasma. The assay measures the inhibition of factor Xa (fXa) activity by the various heparins.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the ACTICHROME® Heparin (anti-fXa) device:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided text does not explicitly state pre-defined acceptance criteria (e.g., "The device must achieve an R-value of at least X"). Instead, it presents performance data and implies that equivalence to the predicate device is the overarching acceptance criterion. Thus, the "acceptance criteria" are inferred from the demonstrated performance in comparison to the predicate.

    Acceptance Criteria (Inferred from Predicate Equivalence)Reported Device Performance (ACTICHROME Heparin (anti-fXa))
    Method Comparison vs. Predicate Device
    Positive correlation with predicate device (implied R > 0.8)Unfractionated Heparin (Lot 007): R = 0.913, Y = 0.843X + 0.003, Sy.x = 0.07
    Unfractionated Heparin (Lot 010): R = 0.859, Y = 0.765X - 0.009, Sy.x = 0.08
    LMW Heparin (Lot 010): R = 0.895, Y = 1.363X - 0.139, Sy.x = 0.14
    Precision (Intra-Assay CV%)
    Maintain low intra-assay variability (implied < 10%)Lot 007, Control 6: 3.3%
    Lot 007, Control 3: 5.6%
    Lot 010, 0.5 U/ml UFH: 2.8%
    Lot 010, 0.25 U/ml UFH: 3.6%
    Lot 010, 0.5 U/ml LMWH: 2.9%
    Lot 010, 0.25 U/ml LMWH: 5.1%
    Precision (Inter-Assay CV%)
    Maintain low inter-assay variability (implied < 15%)Lot 007, Control 6: 4.1%
    Lot 007, Control 3: 4.8%
    Lot 010, 0.5 U/ml UFH: 4.6%
    Lot 010, 0.25 U/ml UFH: 10.6%
    Lot 010, 0.5 U/ml LMWH: 6.2%
    Lot 010, 0.25 U/ml LMWH: 9.3%
    Substantial EquivalenceDeclared "substantially equivalent to the commercially available ACTICHROMIS Hopanin (and Tra) in Is ourin (Xa), manufactured by Biopool International, Ventura, CA, in performance and intended use."

    2. Sample Sizes Used for the Test Set and Data Provenance

    The text describes method comparison studies and precision studies which serve as the test set for device performance.

    • Method Comparison (Test Set):

      • Lot 007 (unfractionated Heparin): N = 92 samples
      • Lot 010 (unfractionated Heparin): N = 40 samples
      • Lot 010 (LMW Heparin): N = 42 samples
      • Data Provenance: Not explicitly stated (e.g., country of origin, retrospective/prospective). It mentions "human plasma" but no further detail.
    • Precision (Test Set Control Samples):

      • Lot 007: 2 control samples run in quadruplicate over 20 runs (N = 80 per control, total 160 measurements).
      • Lot 010: 2 control samples run in quadruplicate over 10 runs (N = 40 per control, total 80 measurements).
      • Data Provenance: Not explicitly stated. The samples are described as "control samples" and "Control Plasma spiked with..."

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    • This device is a chromogenic assay for quantitative determination of heparins. The "ground truth" for such assays is typically established by comparing against a reference method (the predicate device in this case) and assessing analytical performance like precision and accuracy, not by human expert consensus on interpretations. Therefore, the concept of "experts establishing ground truth" as it applies to image analysis or diagnostic interpretation by radiologists, for example, is not relevant here.

    4. Adjudication Method for the Test Set

    • Not applicable. This is a quantitative assay whose performance is assessed against a predicate device and through precision studies, not requiring human adjudication of results in the way a diagnostic interpretation might.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No, an MRMC comparative effectiveness study was not done. This type of study assesses human reader performance, often with and without AI assistance, which is not relevant for a quantitative diagnostic assay like the ACTICHROME® Heparin (anti-fXa) kit.

    6. Standalone (Algorithm Only) Performance Study

    • Yes, the performance data presented (method comparison and precision) represent the standalone performance of the ACTICHROME® Heparin (anti-fXa) assay. It measures the assay's capability to quantify heparin levels. There is no "human-in-the-loop" aspect to its intended use or performance evaluation described.

    7. Type of Ground Truth Used

    • The "ground truth" for this study is essentially the measurement provided by the predicate device for the method comparison, and known concentrations in control samples for the precision studies. It is based on established analytical methods and reference materials, rather than expert consensus, pathology, or outcomes data.

    8. Sample Size for the Training Set

    • The text describes performance studies (method comparison and precision), not a machine learning study with explicit training and test sets in the AI sense. Therefore, there is no specified "training set" for the algorithm itself (as a traditional assay does not have an algorithm that is "trained"). The phrase "training set" typically refers to data used to develop and optimize an AI model. This device is a chromogenic assay kit.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable as there is no "training set" in the context of an AI algorithm. The device is a chemical assay, and its development would involve standard analytical chemistry and immunoassay development processes to optimize its reagents and protocols.
    Ask a Question

    Ask a specific question about this device

    K Number
    K983178

    Validate with FDA (Live)

    Device Name
    COAMATIC HEPARIN
    Date Cleared
    1998-11-03

    (53 days)

    Product Code
    Regulation Number
    864.7525
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    September 10, 1998

    Name of the device:

    Coamatic® Heparin

    Classification name(s):

    | | 864.7525

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Coamatic® Heparin is an in vitro diagnostic test for the quantitative determination of unfractionated heparin (UFH) and low molecular weight heparin (LMWH) activity in human citrated plasma using automated and microplate methods. The amount of UFH or LMWH is determined from the anti-FXa activity expressed by the [AT*Heparin] complex formed in plasma.

    Device Description

    Coamatic® Heparin is an in vitro diagnostic test for the quantitative determination of unfractionated heparin (UFH) and low molecular weight heparin (LMWH) activity in human citrated plasma using automated and microplate methods.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Coamatic® Heparin device, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are not explicitly stated as numerical targets in the provided document. Instead, the study aims to demonstrate "substantial equivalence in its performance" to the predicate device (IL Test™ Heparin). This is primarily assessed through correlation studies, where a slope close to 1, an intercept close to 0, and a high correlation coefficient (r) indicate substantial equivalence. Precision is also evaluated using %CV.

    Therefore, the acceptance criteria are inferred from what constitutes "substantially equivalent performance" in such comparative studies.

    Performance MetricAcceptance Criteria (Inferred)Reported Device Performance
    Comparative Studies (vs. Predicate IL Test™ Heparin)
    SlopeClose to 1.00 (e.g., 0.90 to 1.10)Range: 0.91 to 1.04
    InterceptClose to 0.00 (e.g., -0.10 to 0.10)Range: -0.01 to 0.06
    Correlation (r)High (e.g., ≥ 0.95)Range: 0.97 to 0.99
    Precision (Microplate Method)
    Within Run %CV (UFH)Low (e.g., < 5%)2.8%, 3.4%
    Between Run %CV (UFH)Low (e.g., < 5%)1.2%, 1.5%
    Total %CV (UFH)Low (e.g., < 5%)2.8%, 3.7%
    Within Run %CV (LMWH)Low (e.g., < 5%)3.6%, 2.4%
    Between Run %CV (LMWH)Low (e.g., < 5%)2.8%, 2.3%
    Total %CV (LMWH)Low (e.g., < 5%)4.4%, 3.2%

    Note: The exact numerical acceptance criteria (e.g., specific ranges for slope, intercept, and r) are not explicitly stated but are common expectations for demonstrating substantial equivalence for such devices. The reported performance consistently falls within generally accepted ranges for good correlation and precision in clinical assays.

    2. Sample Sizes Used for the Test Set and Data Provenance

    The "test set" in this context refers to the samples used in the comparative studies against the predicate device and the precision studies.

    • Comparative Studies (against IL Test™ Heparin):

      • Patient Samples (UFH and LMWH treated):
        • Cobas Mira: 87 samples
        • Microplate: 70 samples
        • ACL 300: 62 samples
        • ACL Futura: 113 samples
        • MLA Electra: 80 samples
      • Normal Plasmas and Pooled Plasmas Spiked with UFH:
        • Sysmex 6000: 30 samples
        • AMAX: 30 samples
        • Hitachi 911: 30 samples
        • Hitachi 917: 30 samples
    • Precision Data (Microplate Method):

      • The sample size for the precision studies is not explicitly stated as a number of individual patient samples. The %CVs are reported at specific mean concentrations (0.7 IU/mL UFH, 0.4 IU/mL UFH, 0.7 IU/mL LMWH, 0.4 IU/mL LMWH), implying replicate measurements were performed to determine within-run, between-run, and total precision. The "n" in the comparative table refers to the number of samples compared across methods, not necessarily the number of replicates for precision.
    • Data Provenance: The document does not specify the country of origin of the data. It states samples were from "patients treated with UFH and LMWH" and "normal plasmas and pooled plasmas spiked with UHF heparin," implying these were likely clinical or laboratory samples. The study appears to be retrospective in the sense that existing samples (from patients or prepared in a lab) were used, rather than prospectively enrolling and following patients solely for this study.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This type of device (in vitro diagnostic for quantitative determination of substances) does not typically involve human expert "ground truth" establishment in the same way an image analysis AI might.

    • Number of Experts: Not applicable.
    • Qualifications of Experts: Not applicable.

    For quantitative assays like this, the "ground truth" is established by a reference method or a predicate device. In this study, the predicate device, IL Test™ Heparin, run on an ACL 300, served as the de facto reference standard against which the new Coamatic® Heparin device was compared. The performance of the predicate device itself would have been validated against an even more fundamental reference method or through extensive clinical use.

    4. Adjudication Method for the Test Set

    Not applicable. This is a quantitative laboratory assay for which adjudication by experts is not a standard practice. The "result" is the numerical concentration reported by the instrument. Discrepancies would be resolved through re-testing, calibration checks, or investigation of assay interference, not through expert consensus.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, an MRMC comparative effectiveness study was not done. These studies are typically relevant for diagnostic imaging or pathology, where human readers interpret and classify cases. This device is an in vitro diagnostic assay that produces a quantitative numerical output, not an image or case requiring human interpretation in the same way.

    6. Standalone (Algorithm Only) Performance

    Yes, the studies presented in the "Summary of Performance Data" are essentially standalone performance studies for the Coamatic® Heparin device. The device is a laboratory assay kit which, when run on various automated analyzers (e.g., Cobas Mira, Microplate, ACL 300, ACL Futura, MLA Electra, Sysmex 6000, AMAX, Hitachi 911, Hitachi 917), produces quantitative results. The performance metrics (slope, intercept, correlation against the predicate, and precision) represent the device's inherent capability to measure heparin directly. There is no human-in-the-loop component being evaluated in these specific performance summaries.

    7. Type of Ground Truth Used

    The ground truth used for the comparative studies was the measurements obtained from the predicate device, IL Test™ Heparin, particularly when run on an ACL 300. This approach is standard for demonstrating substantial equivalence for new in vitro diagnostic devices against already-cleared predicate devices. For precision, the ground truth is statistical variation around a known concentration of analyte.

    8. Sample Size for the Training Set

    The document does not mention a "training set" in the context of machine learning or algorithm development. This device is a biochemical assay kit, not an AI/ML algorithm that requires training data in that sense. The methods are established based on known chemical reactions and principles.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no mention of a machine learning "training set" for this device. The principles of the assay are based on established scientific understanding of Factor Xa inactivation using a synthetic chromogenic substrate.