K Number
K013305

Validate with FDA (Live)

Date Cleared
2002-08-23

(323 days)

Product Code
Regulation Number
864.7525
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Rapidpoint™Coag Enoxaparin Test card (ENOX) is a qualitative test intended for exclusive use with the Rapidpoint Coag analyzer to detect the anticoagulant effects, ≥ 1.0 IU/ml, of the low molecular weight heparin (LMWH), Lovenox®/Clexane® (enoxaparin sodium)¹, in arterial citrated whole blood from patients with unstable angina (UA)/non-ST segment elevation myocardial infarction (NSTEMI) who may transition to percutaneous intervention (PCI). The ENOX test is intended for use at either the point of care or in the central laboratory. The device does not discriminate between values of enoxaparin below 1.0 IU/ml and the absence of drug.

The test provides information on the patient's citrated arterial whole blood response to enoxaparin by measurement of the clotting time using a factor Xa activated clotting method and should be interpreted in conjunction with other clinical data available to the clinician

Device Description

The RapidPoint Enoxaparin (ENOX) test is a one-step dry coagulation method performed on the RapidPoint Coag analyzer. All of the components necessary to perform the assay, with the exception of patient sample, are included in the reaction chamber of the test card.

In the ENOX test, factor X is rapidly converted to factor Xa by a specific factor X activator initiating the clotting process. Enoxaparin, from the patient's blood, complexes with antithrombin (AT), to inhibit factor Xa and lengthen the clotting time. Reported clotting times in excess of the assay cut-off indicate an enoxaparin concentration greater than or equal to 1.0 International Units per milliliter. The results generated by the ENOX test are indicative of the anticoagulant effect produced by enoxaparin in citrated arterial whole blood.

The test card formulation contains purified Factor Xa activator, calcium, phospholipid and stabilizers. Paramagnetic iron oxide particles (PIOP) are included to provide an optical detection mechanism in the presence of patient sample.

AI/ML Overview

Here's a summary of the acceptance criteria and study details for the Thrombolytic Assessment System (TAS) or Rapidpoint™ Coag Enoxaparin (ENOX) Test Card.

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are implicitly defined by the clinical study's results in relation to the intended use. The device is intended to detect enoxaparin levels ≥1.0 IU/mL. The study uses a clotting time cutoff of ≥ 260 seconds to indicate this.

Acceptance Criteria (Implied)Reported Device Performance (Clinical Study)
Detect enoxaparin ≥1.0 IU/mL (indicated by clotting time ≥ 260 seconds)Sensitivity: 96.3% (95% CI: 89.4-99.2)
Discriminate enoxaparin <1.0 IU/mL or absence of drug (indicated by clotting time < 260 seconds)Specificity: 25% (95% CI: 3.2-65.1)

2. Sample Size and Data Provenance for the Test Set

  • Sample Size (Clinical Study): 88 patients
  • Data Provenance: Prospective, from "6 sites in the United States."
  • Sample Type: Arterial citrated whole blood samples, collected at baseline (no drug) and at peak values (within 15 minutes of dosing) after enoxaparin administration. The samples were tested in duplicate.

3. Number of Experts and Qualifications for Ground Truth of the Test Set

The document does not explicitly state the number of experts used to establish ground truth for the test set.

  • Ground Truth Method: The ground truth for the clinical study was established by comparing the ENOX card clot times to "central laboratory measured plasma anti-Xa levels." This implies that the central lab results, likely interpreted by qualified laboratory personnel, served as the reference standard. The qualifications of those performing the anti-Xa level measurements are not specified, but it's assumed to be standard clinical laboratory practice.

4. Adjudication Method for the Test Set

The document does not mention an adjudication method for the test set. The comparison was made directly between the device's output and the central lab's anti-Xa results.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, a multi-reader multi-case (MRMC) comparative effectiveness study was not performed. This device is an in-vitro diagnostic (IVD) assay designed to provide a direct measurement, not an imaging or diagnostic aid that human readers would interpret to improve their performance with or without AI assistance. Therefore, there is no effect size reported for human readers improving with or without AI.

6. Standalone (Algorithm Only) Performance

Yes, the provided data describes the standalone performance of the algorithm (the ENOX Test Card used with the RapidPoint Coag analyzer). The sensitivity and specificity reported are for the device itself, without human interpretation influencing the final "positive" or "negative" determination based on the 260-second cutoff.

7. Type of Ground Truth Used

The type of ground truth used for the clinical study was: central laboratory measured plasma anti-Xa levels.

8. Sample Size for the Training Set

The document does not explicitly state the sample size for a "training set." The performance data presented focuses on the clinical validation of the device. However, "In-House Data" for reproducibility used unspecified numbers of "citrated plasma with and without the addition of 1.0 (anti-Xa) IU/ml enoxaparin" and "venous citrated whole blood from single donors." These in-house studies likely contributed to the optimization of the assay (e.g., establishing the 260-second cutoff), which could be considered an internal development/training phase.

9. How the Ground Truth for the Training Set Was Established

For the in-house reproducibility studies (which might loosely correspond to an internal development/training phase):

  • Ground Truth Method: Ground truth appears to have been established by directly spiking known concentrations of enoxaparin (0.0 IU/ml and 1.0 IU/ml) into citrated plasma or venous whole blood. This provides a clear, controlled, and known ground truth for evaluating the device's basic performance and reproducibility at specific enoxaparin levels.

{0}------------------------------------------------

510(k) Summary of Safety and Effectiveness Information

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and CFR 807.92.

The assigned 510(k) number is K013305

  • Patrick D. Mize, Ph.D. (author) Submitted by: Pharmanetics, Inc. (Cardiovascular Diagnostics, Inc.)
  • 9401 Globe Center Drive Address: Suite 140 Morrisville, NC 27560

Phone: 1-800-247-4234

Fax: 1-919-582-2601

  • Contact: Laura Nea Director Quality Assurance
    Date of Summary: August 22, 2002

{1}------------------------------------------------

Trade name: Thrombolytic Assessment System (TAS) or Rapidpoint™ Coag Enoxaparin (ENOX) Test Card

Common Name: ENOX Test Card

Classification Name: CFR Section 21 CFR 864.7525 Class II

Predicate Device: K925433 Stachrom® HEPARIN ASSAY KIT

Description of the Device:

The RapidPoint Enoxaparin (ENOX) test is a one-step dry coagulation method performed on the RapidPoint Coag analyzer. All of the components necessary to perform the assay, with the exception of patient sample, are included in the reaction chamber of the test card.

In the ENOX test, factor X is rapidly converted to factor Xa by a specific factor X activator initiating the clotting process. Enoxaparin, from the patient's blood, complexes with antithrombin (AT), to inhibit factor Xa and lengthen the clotting time. Reported clotting times in excess of the assay cut-off indicate an enoxaparin concentration greater than or equal to 1.0 International Units per milliliter. The results generated by the ENOX test are indicative of the anticoagulant effect produced by enoxaparin in citrated arterial whole blood.

The test card formulation contains purified Factor Xa activator, calcium, phospholipid and stabilizers. Paramagnetic iron oxide particles (PIOP) are included to provide an optical detection mechanism in the presence of patient sample.

Intended Use

The Rapidpoint™ Coag Enoxaparin Test card (ENOX) is a qualitative test intended for exclusive use with the Rapidpoint Coag analyzer to detect the anticoagulant effects, ≥1.0 IU/mL, of the low molecular weight heparin (LMWH), Lovenox®/Clexane® (enoxaparin sodium)', in arterial citrated whole blood from patients with unstable angina (UA)/non-ST segment elevation myocardial infarction (NSTEMI) who may transition to percutaneous intervention (PCI). The ENOX test is intended for use either at the point of care or in the central laboratory. The device does not discriminate between values of enoxaparin below 1.0 IU/ml and the absence of drug.

The test provides information on the patient's arterial citrated whole blood response to enoxaparin by measurement of the clotting time using a factor Xa activated clotting method and should be interpreted in conjunction with other clinical data available to the clinician.

{2}------------------------------------------------

The ENOX Test is optimized to detect the effects of enoxaparin only. The use of the ENOX test to monitor unfractionated heparin (UFH) or other low molecular weight heparins (LMWH) is contraindicated due to the different clot signatures or responses produced by each of these compounds. ENOX test results have only been validated for use in detecting the effect of enoxaparin in citrated arterial whole blood. Unfractionated heparin interferes with this test.

For in vitro diagnostic use only.

CharacteristicPredicate DeviceProposed Device
DeviceStachrom Heparin AssayRapidpoint ENOX Test
Intended UseDetermines heparin or LMWHconcentrationDetects arterial citrated wholeblood anticoagulant effect ofenoxaparin (a LMWH)
FormatChromogenic [colorimetric]assayDried reagents in reactionchamber on flat test card
ReactionTwo-stage chemicalOne-stage clotting
Sample typePlasmaArterial citrated whole blood
Reagent baseAntithrombin (AT), factor Xa,chromogenic substrateFactor X activator,phospholipid, calcium chloride,PIOP
ReactionHeparin in plasma complexeswith AT to inactivate addedfactor Xa. Remaining factor Xacatalyzes release of p-nitroaniline from the substrate;Color formation is inverselyproportional to heparin levelEnoxaparin (LMWH) in bloodcomplexes with AT toinactivate factor Xa. Factor Xactivator converts factor X tofactor Xa, initiating clottingprocess. Clot formation time isrelated to enoxaparin (LMWH)concentration.
InstrumentSpectrophotometerRapidPoint Coag (TAS)
Endpoint monitoredColorimetric reaction;Absorbance at 405 nmClot formation process.Clotting time in seconds
Test interpretationHeparin or LMWHconcentrationQualitative. Enoxaparin > 1.0IU/ml Anti Xa gives clot timesin arterial citrated whole blood> 260 Secs. Clot times < 260secs, equivocal or drugabsent.
Quality controlPlasma control run with eachset of assaysElectronic QC, and two levelsof control plasma

Technological characteristics of the Device compared to the Predicate Device

{3}------------------------------------------------

Summary of Performance Data:

TAS ENOX cards and the TAS (Rapidpoint Coag) analyzer were used to establish the performance characteristics of the system.

IN-HOUSE DATA

Reproducibility: Reproducibility studies performed using citrated plasma with and without the addition of 1.0 (anti-Xa) IU/ml enoxaparin produced the following ENOX test results.

Enoxaparin concentration (anti-Xa IU/ml)0.01.0
<260 Seconds100%3.5%
≥ 260 Seconds0%96.5%

Within Run reproducibility was also performed using venous citrated whole blood from single donors. A typical result for this study is presented below.

Enoxaparin Concentration (anti-Xa IU/ml)0.01.0
< 260 Second100%0%
≥ 260 Seconds0%100%

The ENOX test is optimized to detect the anticoagulant effect of enoxaparin. The ENOX test should not be used as a method to detect the effect of enoxaparin therapy when used in combination with other heparinoids, or direct thrombin inhibitors (e.g. lepirudin). For these drugs, the treating physician should refer to the pharmacokinetics section of the Prescribing Information on circulation clearance times. The performance characteristics of the ENOX test have not been evaluated with clinical samples containing these antithrombin agents. The use of plasma expanders (e.g. Isolyte) may cause a prolongation of clotting time to values outside the expected range, even in the absence of enoxaparin. The use of the ENOX test in cases where hemodilution is carried out is contraindicated.

The results of in vitro studies have demonstrated that lipemia does not interfere with the ENOX test results. Hematocrit values 20 - 50% do not cause significant changes in ENOX clotting times; hematocrit levels >55% may cause elongated clotting times and should be interpreted with caution.

Deficiencies in factors X, V, and prothrombin could result in a prolongation of the ENOX clotting test result even in the absence of enoxaparin in the sample. In the presence of enoxaparin, a deficiency in these factors would reflect both enoxaparin effect and prolongation caused by the deficient factor(s). Drugs that affect clotting factors in the common pathway (e.g. coumadin), including factors X, V, prothrombin and fibrinogen can affect ENOX clotting times. No testing has been performed on clinical samples with an International Normalized Ratio (INR) above 1.4.

{4}------------------------------------------------

CLINICAL Studies

Data were obtained from a clinical trial comparing arterial citrated whole blood clot times using the ENOX card versus central laboratory measured plasma anti-Xa levels, Citrated whole blood arterial samples were collected at baseline and at peak values (within 15 minutes of dosing) after enoxaparin administration among 88 patients. The trial was conducted at 6 sites in the United States and included 27 females (31%). The mean age of the population was 61.3 years, and the mean weight was 87.3 kg. The study cohort received enoxaparin for a primary indication of coronary artery disease. All patients received a weight-adjusted dose of 0.75mg/kg IV. Samples were tested in duplicate, and the first sample was used for final analysis. The relationship between clotting time cutoff (260 secs) and 1.0 IU/ml anti-Xa value from the clinical trial data is presented in the following table.

Number of Subjects Above and Below ENOX Cutoff of 260 seconds:

Time/anti-XaConcentrationanti-Xa ≥1.0 IUanti-Xa <1.0IU
≥ 260 secs776
< 260 secs32

The sensitivity and specificity point estimates are 96.3% (95% CI of 89.4-99.2) and 25% (95% CI of 3.2-65.1) respectively. In keeping with the qualitative nature of the claim. patients with clotting times < 260 seconds are considered to have equivocal results consistent with the absence of the drug or < 1.0 IU/ml drug levels. It is recommended that physicians continue to observe patients with equivocal clotting times, and conduct alternate tests as needed.

{5}------------------------------------------------

Image /page/5/Figure/1 description: The image shows frequency distributions for clot times. There are two distributions shown, one for baseline (no drug) and one for Enox present. The baseline distribution has N=88 and 87<260 secs, while the Enox present distribution has N=88 and 83>260 secs. The x-axis shows clot time, ranging from 0 to 600.

Frequency distribution of patients' ENOX card clotting times for citrated arterial whole blood samples, before and after the IV administration of Enoxaparin. The 88 patients were candidates for coronary intervention and were treated with 0.75mg/kg enoxaparin IV. Samples containing drug were obtained within 15 minutes of drug administration.

In summary, the test provides information on the patient's citrated arterial whole blood response to enoxaparin by measurement of the clotting time using a factor Xa activated clotting method. Results above the cutoff are consistent with values greater than or equal to 1.0 IU/ml of enoxaparin. The results should be interpreted in conjunction with other clinical information available to the clinician. Unexpected results should be verified with an alternative diagnostic method.

{6}------------------------------------------------

Statement of Substantial Equivalence

The TAS ENOX card has been shown to be substantially equivalent to the predicate device, STACHROM® Heparin Assay, in performance, intended use and safety and effectiveness, for the qualitative determination of > 1.0 IU/ml (anti –Xa activity) of enoxaparin in arterial citrated whole blood.

{7}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/7/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES (USA)" arranged in a circular pattern around the symbol. The logo is black and white.

Public Health Service

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

AUG 2 3 2002

Ms. Laura P. Nea Director, Quality Assurance PharmaNetics, Inc. Cardiovascular Diagnostics, Inc.(CVDI) 9401 Globe Center Drive - Suite #140 Morrisville, North Carolina 27560

Re: K013305

Trade/Device Name: TAS ™ Enoxaparin (ENOX)™ Test Card Regulation Number: 21 CFR 864.7525 Regulation Name: Heparin Assay Regulatory Class: Class II Product Code: KFF Dated: July 18, 2002 Received: July 19, 2002

Dear Ms. Nea:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{8}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Steven Butman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory-Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{9}------------------------------------------------

Page 1 of 1

510(k) Number (if known): K013305

Device Name: Enoxaparin (ENOX) Test Card

Indications for Use:

The Rapidpoint™Coag Enoxaparin Test card (ENOX) is a qualitative test intended for exclusive use with the Rapidpoint Coag analyzer to detect the anticoagulant effects, ≥ 1.0 IU/ml, of the low molecular weight heparin (LMWH), Lovenox®/Clexane® (enoxaparin sodium)¹, in arterial citrated whole blood from patients with unstable angina (UA)/non-ST segment elevation myocardial infarction (NSTEMI) who may transition to percutaneous intervention (PCI). The ENOX test is intended for use at either the point of care or in the central laboratory. The device does not discriminate between values of enoxaparin below 1.0 IU/ml and the absence of drug.

The test provides information on the patient's citrated arterial whole blood response to enoxaparin by measurement of the clotting time using a factor Xa activated clotting method and should be interpreted in conjunction with other clinical data available to the clinician

Jasshine Bautista
(Division/Sign-Off)

Wision of Clinical Laboratory Devices

510(k) Number K013305

(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

§ 864.7525 Heparin assay.

(a)
Identification. A heparin assay is a device used to determine the level of the anticoagulant heparin in the patient's circulation. These assays are quantitative clotting time procedures using the effect of heparin on activated coagulation factor X (Stuart factor) or procedures based on the neutralization of heparin by protamine sulfate (a protein that neutralizes heparin).(b)
Classification. Class II (performance standards).