    510k Summary Text (Full-text Search) :

    . § 862.1580)Product Code CEO | |
    | Predicate Devices: | Manufacturer for reagent

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ACE Albumin Reagent is intended for the quantitative determination of albumin concentration in serum and lithium heparin plasma using the ACE, ACE Alera, and ACE Axcel Clinical Chemistry Systems. Albumin measurements are used in the diagnosis and treatment of numerous diseases involving primarily the liver or kidneys. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.

    ACE Total Protein Reagent is intended for the quantitative determination of total protein concentration in serum and lithium heparin plasma using the ACE, ACE Alera, and ACE Axcel Clinical Chemistry Systems. Total protein measurements are used in the diagnosis and treatment of a variety of diseases involving the liver, kidney, or bone marrow as well as other metabolic or nutritional disorders. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.

    ACE Calcium-Arsenazo Reagent is intended for the quantitative determination of calcium concentration in serum and lithium heparin plasma using the ACE, ACE Alera, and ACE Axcel Clinical Chemistry Systems. Calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany (intermittent muscular contractions or spasms). This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.

    ACE Inorganic Phosphorus U.V. Reagent is intended for the quantitative determination of inorganic phosphorus concentration in serum and lithium heparin plasma using the ACE, ACE Alera, and ACE Axcel Clinical Chemistry Systems. Measurements of inorganic phosphorus are used in the diagnosis and treatment of various disorders, including parathyroid gland and kidney diseases and vitamin D imbalance. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.

    Device Description

    In the ACE Albumin Reagent assay, Bromcresol green binds specifically to albumin to form a green colored complex, which is measured bichromatically at 629 nm/692 nm. The intensity of color produced is directly proportional to the albumin concentration in the sample.

    In the ACE Total Protein Reagent assay, cupric ions react with the peptide bonds of proteins under alkaline conditions to form a violet colored complex, which is measured bichromatically at 544 nm/692 nm. The intensity of color produced is directly proportional to the total protein concentration in the sample.

    In the ACE Calcium-Arsenazo Reagent assay, calcium reacts with Arsenazo III in an acidic solution to form a blue-purple colored complex, which is measured bichromatically at 647 nm/692 nm. The intensity of color produced is directly proportional to the calcium concentration in the sample.

    In the ACE Inorganic Phosphorus U.V. Reagent assay, under acidic conditions, inorganic phosphorus in serum reacts with ammonium molybdate to form an unreduced phosphomolybdate complex, which absorbs strongly at 340 nm. The increase in absorbance, measured bichromatically at 340 nm/378 nm, is directly proportional to the amount of phosphorus in the sample.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and study information for the ACE Albumin Reagent, ACE Total Protein Reagent, ACE Calcium-Arsenazo Reagent, and ACE Inorganic Phosphorus U.V. Reagent, based on the provided text.


    1. Table of Acceptance Criteria and Reported Device Performance

    The provided documentation does not explicitly state formal "acceptance criteria" with specific thresholds for each performance metric. However, it presents detailed performance data, particularly precision (within-run and total %CV) and method comparison (regression analysis, correlation coefficient), comparing the new reagents on various ACE clinical chemistry systems (ACE, ACE Alera, ACE Axcel) against existing predicate devices and among themselves. The implied acceptance is that the new reagents perform comparably to, or as effectively as, the predicate devices and demonstrate acceptable precision and linearity for clinical use.

    Below is a summary of the reported device performance based on the "In-House Precision" and "In-House Matrix Comparison" tables. Since explicit acceptance criteria are not given, the performance data itself is presented as the evidence of meeting implied clinical utility and equivalence to predicate devices.

    ACE Albumin Reagent

    MetricAcceptance Criteria (Implied)Reported Performance (Range across ACE, Alera, Axcel systems)
    Precision (%CV)Clinically acceptableSerum: Within-Run: 0.5-1.6%, Total: 0.6-1.8%Plasma: Within-Run: 0.8-1.7%, Total: 1.1-1.7%
    Matrix Comparison (Serum vs. Plasma)Slope close to 1, Intercept close to 0, High CorrelationSlope: 0.956 - 1.002Intercept: -0.01 - 0.20Correlation: 0.9850 - 0.9905
    LinearityBroad clinical range, r^2 close to 1Linear to 7.6 g/dLy = 0.980x + 0.01, r^2 = 0.9982
    Detection Limits (ACE Alera)Low enough for clinical utilityLoB: 0.08 g/dL, LoD: 0.09 g/dL, LoQ: 0.09 g/dL
    Interferences (ACE Alera)No significant interference at clinically relevant levelsIcterus: 60 mg/dL, Hemolysis: 250 mg/dL, Lipemia: 1000 mg/dL, Ascorbic Acid: 6 mg/dL

    ACE Total Protein Reagent

    MetricAcceptance Criteria (Implied)Reported Performance (Range across ACE, Alera, Axcel systems)
    Precision (%CV)Clinically acceptableSerum: Within-Run: 0.7-1.3%, Total: 0.8-1.4%Plasma: Within-Run: 0.5-1.3%, Total: 0.7-1.4%
    Matrix Comparison (Serum vs. Plasma)Slope close to 1, Intercept close to 0, High CorrelationSlope: 0.994 - 1.001Intercept: 0.12 - 0.34Correlation: 0.9798 - 0.9885
    LinearityBroad clinical range, r^2 close to 1Linear to 15.1 g/dLy=0.991x + 0.04, r^2 = 0.9979
    Detection Limits (ACE Alera)Low enough for clinical utilityLoB: 0.08 g/dL, LoD: 0.13 g/dL, LoQ: 0.20 g/dL
    Interferences (ACE Alera)No significant interference at clinically relevant levelsIcterus: 56.8 mg/dL, Hemolysis: 250 mg/dL, Lipemia: 929 mg/dL, Ascorbic Acid: 6 mg/dL

    ACE Calcium-Arsenazo Reagent

    MetricAcceptance Criteria (Implied)Reported Performance (Range across ACE, Alera, Axcel systems)
    Precision (%CV)Clinically acceptableSerum: Within-Run: 0.7-1.6%, Total: 0.9-2.7%Plasma: Within-Run: 0.5-1.9%, Total: 1.1-2.0%
    Matrix Comparison (Serum vs. Plasma)Slope close to 1, Intercept close to 0, High CorrelationSlope: 0.978 - 1.008Intercept: -0.06 - 0.33Correlation: 0.9793 - 0.9911
    LinearityBroad clinical range, r^2 close to 1Linear to 16.5 mg/dLy=0.992x +0.27, r^2 = 0.9990
    Detection Limits (ACE Alera)Low enough for clinical utilityLoB: 0.09 mg/dL, LoD: 0.11 mg/dL, LoQ: 0.23 mg/dL
    Interferences (ACE Alera)No significant interference at clinically relevant levelsIcterus: 58.8 mg/dL, Hemolysis: 1000 mg/dL, Lipemia: 1000 mg/dL, Ascorbic Acid: 6 mg/dL

    ACE Inorganic Phosphorus U.V. Reagent

    MetricAcceptance Criteria (Implied)Reported Performance (Range across ACE, Alera, Axcel systems)
    Precision (%CV)Clinically acceptableSerum: Within-Run: 0.3-4.4%, Total: 0.5-5.0%Plasma: Within-Run: 0.9-5.1%, Total: 0.9-6.1%
    Matrix Comparison (Serum vs. Plasma)Slope close to 1, Intercept close to 0, High CorrelationSlope: 0.999 - 1.049Intercept: -0.28 - 0.04Correlation: 0.9927 - 0.9950
    LinearityBroad clinical range, r^2 close to 1Linear to 21 mg/dLy=1.001x +0.03, r^2 = 0.9995
    Detection Limits (ACE Alera)Low enough for clinical utilityLoB: 0.25 mg/dL, LoD: 0.35 mg/dL, LoQ: 0.35 mg/dL
    Interferences (ACE Alera)No significant interference at clinically relevant levelsIcterus: 11.5 mg/dL, Hemolysis: 250 mg/dL, Lipemia: 306 mg/dL, Ascorbic Acid: 6 mg/dL

    2. Sample Sizes Used for the Test Set and Data Provenance

    The studies mentioned are "In-House Precision," "In-House Matrix Comparison: Serum vs. Plasma," "POL - Precision," and "POL – Method Comparison."

    • In-House Precision (Serum vs. Plasma):
      • Sample Size: Not explicitly stated for each "low, mid, high" concentration level, but implies multiple replicates for each level tested across the three systems (ACE, Alera, Axcel). For example, the ACE Alera precision table (pg. 16) shows 3 levels (low, mid, high) for serum, with reported mean, within-run SD, and total SD. Typically, precision studies involve running samples multiple times a day over several days.
      • Data Provenance: "In-House" suggests it was conducted by Alfa Wassermann Diagnostic Technologies, LLC, likely at their own facilities. It is a prospective study as they are performing experiments to generate data.
    • In-House Matrix Comparison: Serum vs. Plasma:
      • Sample Size:
        • Albumin: ACE: 55 pairs, ACE Alera: 56 pairs, ACE Axcel: 56 pairs
        • Total Protein: ACE: 56 pairs, ACE Alera: 56 pairs, ACE Axcel: 81 pairs
        • Calcium-Arsenazo: ACE: 56 pairs, ACE Alera: 56 pairs, ACE Axcel: 81 pairs
        • Inorganic Phosphorus: ACE: 100 pairs, ACE Alera: 102 pairs, ACE Axcel: 56 pairs
      • Data Provenance: "In-House" suggests it was conducted by Alfa Wassermann Diagnostic Technologies, LLC, likely at their own facilities. The comparison between serum and plasma samples implies these were collected from human subjects. This is a prospective study.
    • POL (Physician Office Laboratory) - Precision:
      • Sample Size: For each reagent and each system (ACE and ACE Alera), there are 3 "samples" (representing different concentration levels) tested at 3 different POL sites. Each sample/site combination has "Within-Run" and "Total" precision reported, implying multiple replicates for each measurement.
      • Data Provenance: Conducted at "POL 1," "POL 2," and "POL 3" sites, indicating external collection and testing beyond the manufacturer's immediate facilities. This is a prospective study.
    • POL (Physician Office Laboratory) - Method Comparison:
      • Sample Size:
        • Albumin: 50 samples for each POL site (x3 POLs)
        • Total Protein: 51 samples for each POL site (x3 POLs)
        • Calcium-Arsenazo: 50 samples for each POL site (x3 POLs)
        • Inorganic Phosphorus: 50 samples for POL 1 & 3, 48 samples for POL 2
      • Data Provenance: Comparisons between "ACE In-House (x)" and "ACE POL (y)" or "ACE In-House (x)" and "ACE Alera POL (y)". This indicates the data for these studies was collected at both in-house facilities and external Physician Office Laboratories. This is a prospective study design, comparing results from different testing environments.
    • Detection Limits & Linearity (ACE Alera):
      • Sample Size: Not specified for these specific studies, but typically involves a series of diluted and concentrated samples to define the measuring range.
      • Data Provenance: In-House, prospective.
    • Interference (ACE Alera):
      • Sample Size: Not specified, but involves spiking samples with various interferents at different concentrations.
      • Data Provenance: In-House, prospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    For these types of in vitro diagnostic (IVD) assays, the "ground truth" is typically established by reference methods or validated comparative methods, often using certified calibrators and controls. The documentation does not mention the use of human experts to establish ground truth for the test set in the traditional sense of medical image interpretation (e.g., radiologists interpreting images). Instead, the studies rely on quantitative measurements and statistical comparisons with established methods (the predicate devices or in-house reference measurements) to demonstrate performance. Therefore, no information is provided on the number or qualifications of experts for ground truth establishment.


    4. Adjudication Method for the Test Set

    Not applicable. As described in point 3, the "ground truth" for these quantitative chemical assays is not established through expert consensus or adjudication in the way it would be for qualitative or interpretive diagnostic devices like medical imaging. Performance is evaluated by statistical comparison of numerical results.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    Not applicable. This device consists of chemical reagents for laboratory measurement, not an AI-assisted diagnostic tool interpreted by human readers. Therefore, an MRMC comparative effectiveness study involving human readers and AI is not relevant to this submission.


    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was Done

    The performance presented for these reagents is inherently "standalone" in the sense that it reflects the direct analytical performance of the assays on the specified automated clinical chemistry systems. The results are quantitative measurements produced by the device without human interpretation of raw data beyond reading the numerical output. The "without human-in-the-loop" aspect applies here as the device itself performs the measurement and outputs a numerical value of concentration. The method comparison studies demonstrate the standalone performance of the candidate devices compared to predicate devices.


    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    The ground truth for these assays is established through reference methods and comparison to legally marketed predicate devices.

    • For precision, the "ground truth" for each replicate is assumed to be the true concentration within the sample, and the study assesses the reproducibility of the device in measuring that concentration.
    • For method comparison studies (e.g., In-House vs. POL, or ACE vs. ACE Alera), one method's results (often the predicate or an established in-house method) serve as the comparative 'truth' to evaluate the new method's agreement. The reference method would itself be calibrated against known standards.
    • For linearity, samples of known, graded concentrations are used.
    • For detection limits, the ground truth involves samples with very low, known concentrations.

    These are established analytical chemistry principles rather than "expert consensus" or "pathology" in the diagnostic interpretation sense.


    8. The Sample Size for the Training Set

    The concept of a "training set" is primarily relevant for machine learning or AI algorithms which are iteratively developed and optimized using data. These reagents are chemical assays with a defined photometric measurement principle. While there is a development phase that involves optimizing reagent formulations and instrument parameters, there isn't a "training set" in the computational sense. The data presented here are from formal "verification and validation studies" to demonstrate performance characteristics (precision, linearity, accuracy/comparison, interference, detection limits).


    9. How the Ground Truth for the Training Set Was Established

    As noted in point 8, the concept of a "training set" is not directly applicable to these chemical reagents. The "ground truth" for establishing and validating the performance of such assays is based on:

    • Reference materials/calibrators: Solutions with precisely known concentrations of the analyte (albumin, total protein, calcium, phosphorus) traceable to international standards.
    • Validated comparison methods: Measurements made by existing, legally marketed predicate devices or other well-established and accurate laboratory methods.
    • Controlled spiking experiments: Adding known amounts of substance to samples to assess recovery, linearity, and interference.

    These methods establish the quantitative "truth" against which the performance of the new reagents is measured.

    Ask a Question

    Ask a specific question about this device

    K Number
    K113374

    Validate with FDA (Live)

    Date Cleared
    2012-07-13

    (240 days)

    Product Code
    Regulation Number
    862.1035
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    . § 862.1580

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ACE Albumin Reagent is intended for the quantitative determination of albumin concentration in serum using the ACE Axcel Clinical Chemistry System. Albumin measurements are used in the diagnosis and treatment of numerous diseases involving primarily the liver or kidneys. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.

    The ACE Total Protein Reagent is intended for the quantitative determination of total protein concentration in serum using the ACE Axcel Clinical Chemistry System. Total protein measurements are used in the diagnosis and treatment of a variety of diseases involving the liver, kidney, or bone marrow as well as other metabolic or nutritional disorders. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.

    The ACE Calcium-Arsenazo Reagent is intended for the quantitative determination of calcium concentration in serum using the ACE Axcel Clinical Chemistry System. Calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany (intermittent muscular contractions or spasms). This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.

    The ACE Inorganic Phosphorus U.V. Reagent is intended for the quantitative determination of inorganic phosphorus concentration in serum using the ACE Axcel Clinical Chemistry System. Measurements of inorganic phosphorus are used in the diagnosis and treatment of various disorders, including parathyroid gland and kidney diseases and vitamin D imbalance. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.

    Device Description

    In the ACE Albumin Reagent assay, Bromcresol green binds specifically to albumin to form a green colored complex, which is measured bichromatically at 629 nm/692 nm. The intensity of color produced is directly proportional to the albumin concentration in the sample.

    In the ACE Total Protein Reagent assay, cupric ions react with the peptide bonds of proteins under alkaline conditions to form a violet colored complex which is measured bichromatically at 544 nm/692 nm. The intensity of color produced is directly proportional to the total protein concentration in the sample.

    In the ACE Calcium-Arsenazo Reagent assay, calcium reacts with Arsenazo III in an acidic solution to form a blue-purple colored complex, which is measured bichromatically at 647 nm/692 nm. The intensity of color produced is directly proportional to the calcium concentration in the sample.

    In the ACE Inorganic Phosphorus U.V. Reagent assay, under acidic conditions, inorganic phosphorus in serum reacts with ammonium molybdate to form an unreduced phosphomolybdate complex, which absorbs strongly at 340 nm. The increase in absorbance, measured bichromatically at 340 nm/378 nm, is directly proportional to the amount of phosphorus in the sample.

    The ACE Albumin Reagent consists of a single reagent bottle. The reagent contains Bromcresol green and acetate buffer.

    The ACE Total Protein Reagent consists of a single reagent bottle. The reagent contains copper sulfate, sodium potassium tartrate, potassium iodide and sodium hydroxide.

    The ACE Calcium-Arsenazo Reagent consists of a single reagent bottle. The Reagent contains Arsenazo III.

    The ACE Inorganic Phosphorus U.V. Reagent consists of a single reagent bottle. The reagent contains ammonium molybdate and sulfuric acid.

    AI/ML Overview

    Acceptance Criteria and Device Performance Study for ACE Reagents

    The provided 510(k) summary (K113374) describes the performance of four reagents: ACE Albumin Reagent, ACE Total Protein Reagent, ACE Calcium-Arsenazo Reagent, and ACE Inorganic Phosphorus U.V. Reagent, when used with the Alfa Wassermann ACE Axcel Clinical Chemistry System. The study establishes the substantial equivalence of these devices to their predicate devices.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are not explicitly stated in numerical terms (e.g., "CV must be < X%"). Instead, the performance is reported and compared to the established performance of the predicate device (implied through correlation studies). The key performance metrics are precision (within-run CV and total CV), accuracy (correlation coefficient, standard error estimate, confidence intervals for slope and intercept from regression analysis against a predicate device), and detection limit.

    Note: For each analyte, the "acceptance criteria" can be inferred from the reported performance, which demonstrates that the new device performs comparably to the predicate device and within acceptable analytical limits for clinical use. The confidence intervals for the slope and intercept being close to 1 and 0 respectively for accuracy, and low CVs for precision, suggest good performance. A direct numerical acceptance criterion for each metric is not specifically stated in this summary but is inherent in the demonstration of substantial equivalence.

    Performance MetricACE Albumin ReagentACE Total Protein ReagentACE Calcium-Arsenazo ReagentACE Inorganic Phosphorus U.V. Reagent
    Precision
    Within-run CV (Lab)0.9 to 1.7% (at 4 levels)0.8 to 2.4% (at 4 levels)1.3 to 2.3% (at 4 levels)1.4 to 1.9% (at 4 levels)
    Total CV (Lab)1.2 to 2.0% (at 4 levels)1.0 to 2.9% (at 4 levels)1.4 to 2.3% (at 4 levels)1.5 to 2.5% (at 4 levels)
    Within-run CV (POL)0.0 to 1.6% (at 3 sites, 3 levels)0.7 to 1.3% (at 3 sites, 3 levels)0.8 to 1.4% (at 3 sites, 3 levels)0.6 to 3.2% (at 3 sites, 3 levels)
    Total CV (POL)0.0 to 2.3% (at 3 sites, 3 levels)0.8 to 1.6% (at 3 sites, 3 levels)1.1 to 2.9% (at 3 sites, 3 levels)1.0 to 3.9% (at 3 sites, 3 levels)
    Accuracy
    Correlation Study
    Sample Size118 samples121 samples111 samples110 samples
    Range0.4 to 6.4 g/dL0.4 to 13.5 g/dL0.7 to 14.5 mg/dL0.6 to 19.6 mg/dL
    Correlation Coeff.0.99590.99770.99350.9983
    Std Error Est.0.080.120.220.16
    CI Slope0.980 to 1.0130.978 to 1.0020.998 to 1.0420.994 to 1.017
    CI Intercept-0.04 to 0.10-0.12 to 0.06-0.50 to -0.08-0.06 to 0.05
    Patient Corr. (POL)
    Correlation Coeff.0.9894 to 0.9966 (3 sites)0.9932 to 0.9987 (3 sites)0.9895 to 0.9977 (3 sites)0.9977 to 0.9996 (3 sites)
    Std Error Est.0.08 to 0.13 (3 sites)0.09 to 0.24 (3 sites)0.16 to 0.23 (3 sites)0.11 to 0.19 (3 sites)
    CI Slope0.946 to 1.037 (3 sites)0.973 to 1.047 (3 sites)0.969 to 1.075 (3 sites)1.014 to 1.067 (3 sites)
    CI Intercept-0.14 to 0.39 (3 sites)-0.41 to 0.19 (3 sites)-0.43 to 0.42 (3 sites)-0.33 to 0.09 (3 sites)
    Detection Limit0.09 g/dL0.15 g/dL0.11 mg/dL0.07 mg/dL

    2. Sample Size Used for the Test Set and Data Provenance

    The studies are retrospective in nature, as they involve testing samples on both the new device (ACE Axcel System with new reagents) and a predicate device (Alfa Wassermann ACE Clinical Chemistry System).

    The data provenance is not explicitly stated regarding country of origin, but the testing was conducted at "three separate Physician Office Laboratory (POL) sites" in addition to the primary laboratory testing. This suggests the data is from a clinical laboratory environment.

    Test Set Sample Sizes:

    • ACE Albumin Reagent: 118 samples (correlation study)
    • ACE Total Protein Reagent: 121 samples (correlation study)
    • ACE Calcium-Arsenazo Reagent: 111 samples (correlation study)
    • ACE Inorganic Phosphorus U.V. Reagent: 110 samples (correlation study)

    These sample sizes were used for the accuracy/correlation studies comparing the new device to the predicate. The precision studies involved testing at 4 levels for 22 days in the main lab and at 3 levels over 5 days at the three POL sites, without specifying the exact number of individual samples for precision.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    This type of submission (IVD reagent for clinical chemistry) does not typically involve human experts establishing ground truth for individual measurements in the same way an imaging AI might. The "ground truth" for the test set is established by the predicate device, the Alfa Wassermann ACE Clinical Chemistry System. The performance of this predicate device is assumed to be accurate and clinically acceptable, serving as the comparative standard. Therefore, no external experts are explicitly mentioned for ground truth establishment.

    4. Adjudication Method for the Test Set

    Not applicable. This is a quantitative chemical assay, not an assessment requiring human interpretation or adjudication. The comparison is between quantitative measurements from two different analytical systems.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a clinical chemistry reagent, not an AI imaging or diagnostic algorithm that involves human readers.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Yes, the studies evaluate the standalone performance of the ACE Axcel Clinical Chemistry System using the new reagents. The measurements are automated, and the results are directly compared to those obtained from the predicate device. There is no "human-in-the-loop" aspect to the analytical process itself.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The ground truth for these studies is the quantitative measurements obtained from the legally marketed predicate device, the Alfa Wassermann ACE Clinical Chemistry System. This is a common approach for demonstrating substantial equivalence for in vitro diagnostic devices, where the performance of the new device is validated against an already approved method.

    8. The sample size for the training set

    Not applicable. These are chemical reagents and an associated analyzer system, not a machine learning or AI model that requires a "training set" in the conventional sense. The performance characteristics were determined through analytical studies.

    9. How the ground truth for the training set was established

    Not applicable, as there is no "training set" in the context of this device technology.

    Ask a Question

    Ask a specific question about this device

    K Number
    K103615

    Validate with FDA (Live)

    Date Cleared
    2011-09-08

    (273 days)

    Product Code
    Regulation Number
    862.1770
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    . § 862.1580)Product Code CEO |
    |

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ACE Urea Nitrogen Reagent is intended for the quantitative determination of urea nitrogen concentration in urine using the ACE and ACE Alera Clinical Chemistry Systems. Urea nitrogen measurements are used in the diagnosis and treatment of certain renal and metabolic diseases. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.

    The ACE Calcium-Arsenazo Reagent is intended for the quantitative determination of calcium in urine using the ACE and ACE Alera Clinical Chemistry Systems. Calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.

    The ACE Creatinine Reagent is intended for the quantitative determination of creatinine in urine using the ACE and ACE Alera Clinical Chemistry Systems. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis and as a calculation basis for measuring other urine analytes. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.

    The ACE Inorganic Phosphorus U.V. Reagent is intended for the quantitative determination of inorganic phosphorus concentration in urine using the ACE and ACE Alera Clinical Chemistry Systems. Measurements of inorganic phosphorus are used in the diagnosis and treatment of various disorders, including parathyroid gland and kidney diseases and vitamin D imbalance. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.

    Alfa Wassermann Diagnostic Technologies, LLC ACE Urine Standard is intended for the calibration of quantitative urine reagents on Alfa Wassermann Clinical Chemistry Systems. For in vitro diagnostic use only.

    Device Description

    In the ACE Urea Nitrogen Reagent assay, urea in urine is hydrolyzed to yield ammonia and carbon dioxide in the presence of urease. The ammonia formed then reacts with 2-oxoglutarate and NADH in the presence of glutamate dehydrogenase to yield glutamate and NAD. Two moles of NADH are oxidized for each mole of urea present. NADH absorbs strongly at 340 nm, whereas NAD+ does not. The initial rate of decrease in absorbance, monitored bichromatically at 340 nm/647 nm, is proportional to the urea concentration in the urine sample.

    In the ACE Calcium-Arsenazo Reagent assay, calcium reacts with Arsenazo III in an acidic solution to form a blue-purple colored complex, which is measured bichromatically at 647 nm/692 nm. The intensity of color produced is directly proportional to the calcium concentration in the urine sample.

    In the ACE Creatinine Reagent assay, creatinine reacts with picric acid in an alkaline medium to form a red-orange colored complex, which absorbs strongly at 505 nm. The rate of complex formation, determined by measuring the increase in absorbance bichromatically at 505 nm/573 nm during a fixed time interval, is directly proportional to the creatinine concentration in the urine sample.

    In the ACE Inorganic Phosphorus U.V. Reagent assay, under acidic conditions, inorganic phosphorus reacts with ammonium molybdate to form an unreduced phosphomolybdate complex, which absorbs strongly at 340 nm The increase in absorbance measured bichromatically at 340 nm/378 nm, is directly proportional to the amount of phosphorus in the urine sample.

    The ACE urine standard is a liquid, aqueous-based, ready-to-use calibration solution with known gravimetric concentrations of several analytes.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and supporting studies for the Alfa Wassermann ACE Reagents and Urine Standard, based on the provided 510(k) summary:

    Acceptance Criteria and Device Performance

    The acceptance criteria are not explicitly stated as distinct numerical targets in the provided document. Instead, the document presents performance data from studies and implies that these results demonstrate substantial equivalence to predicate devices, thus meeting acceptable performance standards. The reported device performance metrics are precision (within-run CV and total CV), accuracy (correlation coefficient, standard error estimate, confidence interval slope, and confidence interval intercept compared to predicate methods), and detection limit.

    Table of Acceptance Criteria (Implied) and Reported Device Performance for Urine Analytes

    AnalytePerformance MetricImplied Acceptance Criteria (e.g., comparable to predicate)Reported Device Performance (ACE System & ACE Alera System)
    Urea NitrogenPrecision (Within-run CV)Low CV (e.g., ≤ 5%)Lab: 1.1 - 2.3% (within-run), 2.0 - 3.5% (total CV)POL: 0.9 - 3.8% (within-run), 1.6 - 4.6% (total CV)
    Accuracy (Correlation Coefficient)High correlation (e.g., ≥ 0.97)ACE: 0.9769 (r)ACE Alera: 0.9805 (r)POLs: 0.9853 - 0.9895 (r)
    - Standard Error EstimateLow (e.g., < 100 mg/dL)ACE: 71.1ACE Alera: 66.2POLs: 61.5 - 75.1
    - Confidence Interval SlopeClose to 1 (e.g., 0.9-1.1)ACE: 0.871 - 0.945ACE Alera: 0.884 - 0.953POLs: 0.866 - 0.993
    - Confidence Interval InterceptClose to 0ACE: 7.9 - 68.9ACE Alera: -5.9 - 50.9POLs: -49.7 - 61.8
    Detection LimitLow (e.g., < 20 mg/dL)ACE: 10 mg/dLACE Alera: 14 mg/dL
    CalciumPrecision (Within-run CV)Low CV (e.g., ≤ 7%)Lab: 1.4 - 5.3% (within-run), 1.6 - 5.7% (total CV)POL: 0.8 - 6.9% (within-run), 0.8 - 7.5% (total CV)
    Accuracy (Correlation Coefficient)High correlation (e.g., ≥ 0.97)ACE: 0.9752 (r)ACE Alera: 0.9752 (r)POLs: 0.9669 - 0.9810 (r)
    - Standard Error EstimateLow (e.g., < 2 mg/dL)ACE: 1.35ACE Alera: 1.37POLs: 1.35 - 1.70
    - Confidence Interval SlopeClose to 1 (e.g., 0.9-1.1)ACE: 0.973 - 1.060ACE Alera: 0.979 - 1.066POLs: 0.935 - 1.125
    - Confidence Interval InterceptClose to 0ACE: -0.37 - 0.76ACE Alera: -0.46 - 0.66POLs: -1.64 - 1.24
    Detection LimitLow (e.g., < 0.5 mg/dL)ACE: 0.3 mg/dLACE Alera: 0.3 mg/dL
    CreatininePrecision (Within-run CV)Low CV (e.g., < 6%)Lab: 1.0 - 2.1% (within-run), 2.6 - 3.3% (total CV)POL: 0.8 - 3.5% (within-run), 1.5 - 5.8% (total CV)
    Accuracy (Correlation Coefficient)High correlation (e.g., ≥ 0.98)ACE: 0.9818 (r)ACE Alera: 0.9817 (r)POLs: 0.9882 - 0.9904 (r)
    - Standard Error EstimateLow (e.g., < 15 mg/dL)ACE: 11.40ACE Alera: 11.95POLs: 10.72 - 12.24
    - Confidence Interval SlopeClose to 1 (e.g., 0.9-1.1)ACE: 0.950 - 1.021ACE Alera: 0.992 - 1.066POLs: 0.928 - 1.101
    - Confidence Interval InterceptClose to 0ACE: 2.62 - 10.92ACE Alera: -3.25 - 5.44POLs: -6.84 - 9.66
    Detection LimitLow (e.g., < 2 mg/dL)ACE: 0.8 mg/dLACE Alera: 1.2 mg/dL
    Inorganic PPrecision (Within-run CV)Low CV (e.g., < 5%)Lab: 0.7 - 2.2% (within-run), 0.8 - 3.4% (total CV)POL: 0.4 - 3.3% (within-run), 0.6 - 4.6% (total CV)
    Accuracy (Correlation Coefficient)High correlation (e.g., ≥ 0.98)ACE: 0.9878 (r)ACE Alera: 0.9886 (r)POLs: 0.9809 - 0.9954 (r)
    - Standard Error EstimateLow (e.g., < 10 mg/dL)ACE: 5.21ACE Alera: 4.89POLs: 3.88 - 8.40
    - Confidence Interval SlopeClose to 1 (e.g., 0.9-1.1)ACE: 0.960 - 1.018ACE Alera: 0.940 - 0.995POLs: 0.885 - 1.030
    - Confidence Interval InterceptClose to 0ACE: 1.30 - 5.34ACE Alera: 2.06 - 5.91POLs: -1.13 - 7.46
    Detection LimitLow (e.g., < 2 mg/dL)ACE: 1.6 mg/dLACE Alera: 1.2 mg/dL

    Note: The "Implied Acceptance Criteria" are derived from typical expectations for in vitro diagnostic device performance and the fact that these results were used to demonstrate substantial equivalence to predicate devices.

    Study Details

    The provided document describes performance studies for the Alfa Wassermann ACE reagents for urine testing, including precision, accuracy, and detection limit data.

    1. Sample sizes used for the test set and the data provenance:

      • Urea Nitrogen:

        • Precision: 5 urea nitrogen levels tested for 22 days on 2 instruments. At four separate Physician Office Laboratory (POL) sites, 3 urea nitrogen levels were tested over 5 days.
        • Accuracy:
          • One correlation study: 112 samples (ACE System)
          • Another correlation study: 112 samples (ACE Alera System)
          • Patient correlation studies at four separate POL sites (sample size not specified per site, but implies multiple samples per site).
        • Data Provenance: Not explicitly stated, but clinical laboratories and Physician Office Laboratories (POLs) are mentioned, suggesting real-world clinical samples. The studies are referred to as "patient correlation studies," implying prospective collection or retrospective use of patient samples. The country of origin is not specified.
      • Calcium:

        • Precision: 4 calcium levels tested for 22 days on 2 instruments. At four separate POL sites, 3 calcium levels were tested over 5 days.
        • Accuracy:
          • One correlation study: 108 samples (ACE System)
          • Another correlation study: 109 samples (ACE Alera System)
          • Patient correlation studies at four separate POL sites (sample size not specified per site).
        • Data Provenance: Similar to Urea Nitrogen, derived from clinical relevance.
      • Creatinine:

        • Precision: 5 creatinine levels tested for 22 days on 2 instruments. At four separate POL sites, 3 creatinine levels were tested over 5 days.
        • Accuracy:
          • One correlation study: 113 samples (ACE System)
          • Another correlation study: 113 samples (ACE Alera System)
          • Patient correlation studies at four separate POL sites (sample size not specified per site).
        • Data Provenance: Similar to Urea Nitrogen.
      • Inorganic Phosphorus:

        • Precision: 5 phosphorus levels tested for 22 days on 2 instruments. At four separate POL sites, 3 phosphorus levels were tested over 5 days.
        • Accuracy:
          • One correlation study: 113 samples (ACE System)
          • Another correlation study: 112 samples (ACE Alera System)
          • Patient correlation studies at four separate POL sites (sample size not specified per site).
        • Data Provenance: Similar to Urea Nitrogen.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • This information is not provided in the document. For in vitro diagnostic devices like these reagents, the "ground truth" for accuracy studies is typically established by comparing the device's results to a "predicate method" (another legally marketed device) or a reference method, rather than expert consensus on individual cases. The document explicitly states comparison to a "predicate method (x)."
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Adjudication methods like "2+1" or "3+1" are relevant for subjective interpretations (e.g., image analysis, clinical diagnosis consensus) where human readers are making judgments. For quantitative chemical assays, this type of adjudication is generally not applicable. The comparison is objective, machine-to-machine, or machine-to-reference method.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No, MRMC studies are not applicable to these types of in vitro diagnostic reagents. These are chemical assays, not AI-powered diagnostic tools that assist human readers. The performance is assessed on the analytical accuracy and precision of the chemical reaction.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • Yes, implicitly. The performance data presented (precision, accuracy, detection limit) are measurements of the reagent system's analytical performance, which operates as a standalone automated assay on the ACE and ACE Alera Clinical Chemistry Systems. Human involvement is in operating the instrument and interpreting the numerical reports, but the core measurement is algorithm-only in the sense of the chemical reaction and photometric measurement.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The "ground truth" for accuracy was established by comparing the results of the new ACE reagents to existing predicate devices (Roche Diagnostics COBAS INTEGRA reagents, Siemens Healthcare Diagnostics ADVIA 1800/2400 reagent, Verichem Laboratories, Inc. standard) that are already legally marketed. This is a common approach for demonstrating substantial equivalence for IVD devices.
    7. The sample size for the training set:

      • This document describes performance verification studies, not the development or "training" of an AI algorithm or model. Therefore, there is no "training set" in the context of machine learning. The reagents themselves establish the analytical method.
    8. How the ground truth for the training set was established:

      • As there is no "training set" in the machine learning sense, this question is not applicable to the presented data. The reagents are chemical formulations, and their manufacturing consistency and analytical performance are verified against defined standards and predicate methods.
    Ask a Question

    Ask a specific question about this device

    K Number
    K100263

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2011-05-06

    (463 days)

    Product Code
    Regulation Number
    862.1580
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    PHOSPHORUSInorganic phosphorus, "PHOSPHORUS"Class IPhosphorus (inorganic) test system (Sec.862.1580

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ELITech Clinical Systems PHOSPHORUS reagent is for the quantitative in vitro diagnostic determination of inorganic phosphorus in human serum and plasma on the Vital Scientific Selectra/Flexor analyzers. It is not intended for use in Point of Care settings. Measurements of phosphorus (inorganic) are used in the diagnosis and treatment of various disorders, including parathyroid gland and kidney diseases, and vitamin D imbalance.

    ELITech Clinical Systems URIC ACID MONO SL reagent is for the quantitative in vitro diagnostic determination of uric acid in human serum and plasma on the Vital Scientific Selectra/Flexor analyzers. It is not intended for use in Point of Care settings. Measurements obtained by this device are used in the diagnosis and treatment of numerous renal and metabolic disorders, including renal failure, gout, leukemia, psoriasis, starvation or other wasting conditions, and of patients receiving cytotoxic drugs.

    ELITech Clinical Systems UREA UV SL reagent is for the quantitative in vitro diagnostic determination of urea nitrogen in human serum and plasma on the Vital Scientific Selectra/Flexor analyzers. It is not intended for use in Point of Care settings. Measurements obtained by this device are used in the diagnosis and treatment of certain renal and metabolic diseases.

    ELITech Clinical Systems ELICAL 2 is a multi-parametric calibrator for in vitro diagnostic use in the calibration of quantitative ELITech Clinical Systems methods on the Vital Scientific Selectra Junior Analyzer and the Vital Scientific Flexor Junior Analyzer.

    ELITech Clinical Systems ELITROL I is a multi-parametric control serum for in vitro diagnostic use in accuracy control of quantitative ELITech Clinical Systems methods on the Vital Scientific Selectra Junior Analyzer and the Vital Scientific Flexor Junior Analyzer. ELITech Clinical Systems ELITROL II is a multi-parametric control serum for in vitro diagnostic use in accuracy control of quantitative ELITech Clinical Systems methods on the Vital Scientific Selectra Junior analyzer and the Vital Scientific Flexor Junior analyzers.

    Device Description

    ELITech Clinical Systems ELICAL 2 is a lyophilized calibrator based on human serum containing constituents to ensure optimal calibration. ELICAL 2 is prepared exclusively from the blood of donors tested individually and found to be negative for HbsAg and to antibodies to HCV and HIV according to FDA-approved methods or methods in compliance with the European Directive 98/79/EC, Annex II, List A.

    The device for this submission is available as kit only. It consists of 2 reagents: Reagent 1 contains Tris buffer (pH 7.60), Adenosine diphosphate potassium salt (ADP), alpha-Ketoglutarate, Urease (jack bean), Glutamate dehydrogenase (GIDH) (bovine liver) and sodium azide. Reagent 2 contains NADH and sodium azide

    The device for this submission is available as kit only. It consists of 1 reagent R. Reagent R consists of sulfuric acid and ammonium molybdate.

    The device for this submission is available as kit only. It consists of 1 reagent R. Reagent R consists of Phosphate buffer (pH 7.0), N-Ethyl-N-(2-Hydroxy-3-Sulfopropyl) m-Toluidine (EHSPT), Ferrocyanide, Amino-4-antipyrine (4-AAP), Uricase (microorganisms), Peroxidase (horseradich) and sodium azide.

    ELITech Clinical Systems ELITROL I and ELITROL II are two level quality control products consisting of lyophilized human serum containing constituents at desired levels. Elitrol I and Elitrol II are prepared exclusively from the blood of donors tested individually and found to be negative for HbsAg and to antibodies to HCV and HIV according to FDA-approved methods or methods in compliance with the European Directive 98/79/EC, Annex II, List A.

    AI/ML Overview

    The provided text describes several in vitro diagnostic (IVD) devices: ELITech Clinical Systems ELICAL 2 (calibrator), ELITech Clinical Systems UREA UV SL (reagent), ELITech Clinical Systems PHOSPHORUS (reagent), ELITech Clinical Systems URIC ACID MONO SL (reagent), and ELITech Clinical Systems ELITROL I and ELITROL II (controls).

    The studies performed for these devices are primarily method comparison and precision studies, focusing on demonstrating substantial equivalence to predicate devices rather than proving a specific clinical utility through a multi-reader multi-case study or a standalone algorithm performance study.

    Here's an analysis based on your requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implicitly defined by demonstrating equivalence to the predicate devices. The performance metrics presented are directly compared.

    For ELITech Clinical Systems UREA UV SL (K100263):

    Performance MetricAcceptance Criteria (Predicate: ABX PENTRA UREA CP)Reported Device Performance (ELITech Clinical Systems UREA UV SL)
    Measuring Range1.03 to 140.3 mg/dL (Automatic post-dilution: 701.5 mg/dL)4.7 to 140.0 mg/dL (Extended measuring range: 140.0 to 670.0 mg/dL)
    Limit of Detection (LoD)Not explicitly stated0.3 mg/dL
    Limit of Quantification (LoQ)0.9 mg/dL2.3 mg/dL
    Precision
    Within Run (Level 7.3 mg/dL)2.27% (for 18.7 mg/dL)CV=2.1%
    Within Run (Level 29.2 mg/dL)1.66% (for 72.8 mg/dL)CV=0.8%
    Within Run (Level 72.4 mg/dL)2.76% (for 6.0 mg/dL) etc.CV=0.7%
    Total (Level 7.3 mg/dL)2.14% (for 18.5 mg/dL)CV=2.8%
    Total (Level 29.2 mg/dL)1.93% (for 71.7 mg/dL)CV=1.3%
    Total (Level 72.4 mg/dL)2.14% (for 19.2 mg/dL) etc.CV=1.6%
    Method Comparisony=0.99 x - 0.06 mg/dL, r²=0.996, range: 1.03 to 138.89 mg/dLy=0.991 x + 0.6 mg/dL, r=0.999, range: 4.4 to 139.8 mg/dL
    Interference (Hemoglobin)No significant influence up to 460 mg/dLNo significant interference up to 500 mg/dL
    Interference (Triglycerides)No significant influence up to 612.5 mg/dLNo significant interference up to 614 mg/dL triglyceride equivalent
    Interference (Total Bilirubin)No significant influence up to 22.23 mg/dLNo significant interference up to 30 mg/dL (unconjugated) / 29.5 mg/dL (conjugated)
    Calibration Frequency8 days7 days
    On-board Stability70 days14 days

    For ELITech Clinical Systems PHOSPHORUS (K100263):

    Performance MetricAcceptance Criteria (Predicate: ABX PENTRA PHOSPHORUS CP)Reported Device Performance (ELITech Clinical Systems PHOSPHORUS)
    Measuring Range0.30 to 24.18 mg/dL (Automatic post-dilution: 96.72 mg/dL)2.0 to 20.0 mg/dL
    Limit of Detection (LoD)0.28 mg/dL0.02 mg/dL
    Limit of Quantification (LoQ)Not explicitly stated1.00 mg/dL
    Precision
    Within Run (Level 2.37 mg/dL)1.25% (for 4.08 mg/dL)CV=1.1%
    Within Run (Level 4.80 mg/dL)0.77% (for 6.34 mg/dL)CV=1.5%
    Within Run (Level 9.55 mg/dL)2.48% (for 2.39 mg/dL) etc.CV=1.7%
    Total (Level 2.37 mg/dL)2.50% (for 4.01 mg/dL)CV=1.9%
    Total (Level 4.80 mg/dL)1.82% (for 6.35 mg/dL)CV=1.7%
    Total (Level 9.55 mg/dL)3.56% (for 2.50 mg/dL) etc.CV=2.2%
    Method Comparisony=1.04x + 0.15 mg/dL, r²=0.998, range: 0.30 to 24.08 mg/dLy=0.999 - 0.09 mg/dL, r=0.999, range: 2.02 to 20.08 mg/dL
    Interference (Hemoglobin)No significant influence up to 125 mg/dLNo significant interference up to 50 mg/dL
    Interference (Triglycerides)No significant influence up to 262.5 mg/dLNo significant interference up to 732 mg/dL
    Interference (Total Bilirubin)No significant influence up to 6 mg/dLNo significant interference up to 15 mg/dL (unconjugated) / 1.5 mg/dL (conjugated)
    Calibration Frequency34 days28 days
    On-board Stability70 days28 days

    For ELITech Clinical Systems URIC ACID MONO SL (K100263):

    Performance MetricAcceptance Criteria (Predicate: ABX PENTRA URIC ACID CP)Reported Device Performance (ELITech Clinical Systems URIC ACID MONO SL)
    Measuring Range0.18 to 25.00 mg/dL (Automatic post-dilution: 75.00 mg/dL)1.5 to 25.0 mg/dL (Extended measuring range: 25 to 78 mg/dL)
    Limit of Detection (LoD)0.19 mg/dL0.02 mg/dL
    Limit of Quantification (LoQ)Not explicitly stated0.50 mg/dL
    Precision
    Within Run (Level 2.49 mg/dL)0.45% (for 4.62 mg/dL)CV=0.8%
    Within Run (Level 5.19 mg/dL)0.34% (for 11.63 mg/dL)CV=1.3%
    Within Run (Level 7.63 mg/dL)1.24% (for 2.53 mg/dL) etc.CV=1.1%
    Total (Level 2.49 mg/dL)2.81% (for 4.64 mg/dL)CV=2.6%
    Total (Level 5.19 mg/dL)1.39% (for 11.73 mg/dL)CV=2.0%
    Total (Level 7.63 mg/dL)2.64% (for 4.67 mg/dL) etc.CV=2.1%
    Method Comparisony=0.95 x + 0.09 mg/dL, r²=0.996, range: 0.18 to 23.59 mg/dLy=1.015 x + 0.03 mg/dL, r=0.999, range: 1.49 to 24.40 mg/dL
    Interference (Hemoglobin)No significant influence up to 500 mg/dLNo significant interference up to 50 mg/dL
    Interference (Triglycerides)No significant influence up to 612.5 mg/dLNo significant interference up to 1070 mg/dL
    Interference (Total Bilirubin)No significant influence up to 36 mg/dLNo significant interference up to 30 mg/dL (unconjugated) / 14.8 mg/dL (conjugated)
    Calibration Frequency15 days28 days
    On-board Stability41 days28 days

    2. Sample size used for the test set and the data provenance

    • UREA UV SL: The method comparison study gives a range of 4.4 to 139.8 mg/dL for the device. The R-value of 0.999 implies a good correlation across the study range. While a specific number of samples isn't explicitly stated, the R-value is usually calculated from a sufficiently large set of patient samples spanning the analytical range. The data provenance is not stated (e.g., country of origin, retrospective/prospective), but it is implied to be clinical samples used for method comparison.
    • PHOSPHORUS: The method comparison study gives a range of 2.02 to 20.08 mg/dL for the device. The R-value of 0.999 implies a good correlation. Similar to UREA UV SL, the number of samples is not explicitly stated, and data provenance is not provided but implied to be clinical samples.
    • URIC ACID MONO SL: The method comparison study gives a range of 1.49 to 24.40 mg/dL for the device. The R-value of 0.999 implies a good correlation. The number of samples is not explicitly stated, and data provenance is not provided but implied to be clinical samples.

    For ELICAL 2, ELITROL I, ELITROL II: These are calibrators and controls; their performance is validated through stability studies and their intended use in calibrating and controlling assays. No "test set" in the sense of patient samples for diagnostic performance is applicable here. The device description for ELICAL 2, ELITROL I, and ELITROL II mentions they are "prepared exclusively from the blood of donors tested individually and found to be negative for HbsAg and to antibodies to HCV and HIV according to FDA-approved methods or methods in compliance with the European Directive 98/79/EC, Annex II, List A." This suggests human-derived materials.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. These are in vitro diagnostic devices for quantitative measurements, not imaging or subjective diagnostic interpretations requiring expert consensus for ground truth. "Ground truth" in this context refers to the measured values obtained by the predicate device or a reference method for the method comparison studies.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    Not applicable. This is not a study involving human interpretation or adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is an in vitro diagnostic device, not an AI-assisted diagnostic tool requiring human reader studies.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    This is an IVD reagent and calibrator kit, not an algorithm. The performance data presented (e.g., precision, method comparison, LoD, LoQ) represents the "standalone" analytical performance of the device on the specified analyzer (Selectra Junior/Flexor Junior).

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The ground truth for the method comparison studies is the quantitative result obtained from the predicate device (ABX PENTRA UREA CP, ABX PENTRA PHOSPHORUS CP, ABX PENTRA URIC ACID CP) or a reference method assumed to be the "truth" for those studies. For precision studies, there isn't an external "ground truth" but rather a measure of the device's own reproducibility.

    8. The sample size for the training set

    Not applicable. These are traditional IVD reagents and calibrators, not machine learning algorithms that require a "training set."

    9. How the ground truth for the training set was established

    Not applicable. As above, no training set for a machine learning algorithm is involved.

    Ask a Question

    Ask a specific question about this device

    K Number
    K072141

    Validate with FDA (Live)

    Date Cleared
    2008-06-24

    (327 days)

    Product Code
    Regulation Number
    862.1580
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    . § 862.1580, Product code CEO; 21C.F.R. § 862.1775, Product code KNK
    Phosphorous (IP) Reagent cartridge and S Test Uric Acid (UA) Reagent cartridge Regulation Number: 21 CFR 862.1580

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The S-Test Inorganic Phosphorous Reagent is intended for the quantitative determination of inorganic phosphorous concentration in serum or heparin plasma using the S40 Clinical Analyzer. Measurements of phosphorus (inorganic) are used in the diagnosis and treatment of various disorders, including parathyroid gland and kidney diseases, and vitamin D imbalance. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.

    The S-Test Uric Acid Reagent is intended for the quantitative determination of uric acid concentration in serum or heparin plasma using the S40 Clinical Analyzer. Uric acid measurements are used in the diagnosis and treatment of numerous renal and metabolic disorders, including renal failure, gout, leukemia, psoriasis, starvation or other wasting conditions, and of patients receiving cytotoxic drugs. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.

    Device Description

    The S-Test inorganic phosphorus (IP) reagent cartridge, used with the S40 Clinical Analyzer, is intended for quantitative in vitro diagnostic determination of IP in serum or heparin plasma based on a photometric test measuring the formation of molybdenum blue from IP and ammonium molybdate.

    The S-Test uric acid (UA) reagent cartridge, used with the S40 Clinical Analyzer, is intended for quantitative in vitro diagnostic determination of UA in serum or heparin plasma based on a photometric test measuring the formation of a reddish-purple pigment.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Alfa Wassermann S-Test IP and S-Test UA reagent cartridges, based on the provided 510(k) summary:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are not explicitly stated as distinct pass/fail thresholds in a separate section. Instead, the "Performance Data" section describes the results of various tests, and the "Conclusions" section states that the device is "safe and effective" and demonstrates "substantial equivalence to the predicate devices" based on this data. I will present the reported performance data, which implicitly serves as the criteria met for acceptance.

    S-Test Inorganic Phosphorus (IP) Reagent Cartridge

    Performance MetricAcceptance Criteria (Implicit from Results)Reported Device Performance
    Precision
    22-day testing (within-run CV)Low variability (e.g., CV < 5%)1.5% to 2.9% (at 3 IP levels)
    22-day testing (total CV)Low variability (e.g., CV < 5%)2.6% to 4.0% (at 3 IP levels)
    5-day POL & in-house (within-run CV)Low variability (e.g., CV < 5%)0.9% to 3.4%
    5-day POL & in-house (total CV)Low variability (e.g., CV < 5%)1.2% to 3.4%
    Accuracy (Correlation to Comparison Method)Strong correlation (e.g., r > 0.95), low error, slope near 1, intercept near 0
    Main Study (Correlation Coefficient)r > 0.950.976
    Main Study (Std. Error Estimate)Low error0.3
    Main Study (Confidence Interval Slope)Encompassing 11.026 to 1.150
    Main Study (Confidence Interval Intercept)Encompassing 0-0.03 to 0.34
    POL Studies (Correlation Coefficient)r > 0.950.992 to 0.998
    POL Studies (Std. Error Estimate)Low error0.10 to 0.21
    POL Studies (Confidence Interval Slope)Encompassing 11.002 to 1.173
    POL Studies (Confidence Interval Intercept)Encompassing 0-0.075 to 0.395
    SensitivityDetect low concentrations (e.g., clinically relevant levels)Detection limit: 1.2 mg/dL

    S-Test Uric Acid (UA) Reagent Cartridge

    Performance MetricAcceptance Criteria (Implicit from Results)Reported Device Performance
    Precision
    22-day testing (within-run CV)Low variability (e.g., CV < 5%)0.9% to 2.1% (at 3 UA levels)
    22-day testing (total CV)Low variability (e.g., CV < 5%)3.1% to 3.5% (at 3 UA levels)
    5-day POL & in-house (within-run CV)Low variability (e.g., CV < 5%)0.3% to 2.2%
    5-day POL & in-house (total CV)Low variability (e.g., CV < 5%)0.7% to 2.3%
    Accuracy (Correlation to Comparison Method)Strong correlation (e.g., r > 0.95), low error, slope near 1, intercept near 0
    Main Study (Correlation Coefficient)r > 0.950.974
    Main Study (Std. Error Estimate)Low error0.7
    Main Study (Confidence Interval Slope)Encompassing 11.003 to 1.087
    Main Study (Confidence Interval Intercept)Encompassing 0-1.03 to -0.49
    POL Studies (Correlation Coefficient)r > 0.950.967 to 0.995
    POL Studies (Std. Error Estimate)Low error0.32 to 0.69
    POL Studies (Confidence Interval Slope)Encompassing 10.968 to 1.202
    POL Studies (Confidence Interval Intercept)Encompassing 0-1.400 to 0.469
    SensitivityDetect low concentrations (e.g., clinically relevant levels)Detection limit: 1.4 mg/dL

    2. Sample Size Used for the Test Set and Data Provenance

    • S-Test IP:
      • Sample Size: 95 samples for the main correlation study. "Patient correlation studies at three separate POL sites" are also mentioned, implying additional samples, but the total number for these POL studies is not explicitly stated.
      • Data Provenance: Not explicitly stated, but "patient correlation studies at three separate POL sites" and "in-house" studies suggest data collected from different clinical settings and the manufacturer's laboratory, likely within the US given the context of a US FDA submission. It is reasonable to assume these were prospective or a mix of retrospective and prospective, as "patient samples" would either be freshly collected or archived. The summary does not specify.
    • S-Test UA:
      • Sample Size: 183 samples for the main correlation study. Similar to IP, "patient correlation studies at three separate POL sites" were also conducted, with the total sample size for these not specified individually.
      • Data Provenance: Similar to IP, likely from various clinical sites (POLs) and in-house, presuming US origin. The summary does not explicitly state if the data was retrospective or prospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the document. The "ground truth" for these types of diagnostic tests is typically established by comparison to a recognized predicate or reference method run on a different, established analyzer (as indicated by "comparison method" in the accuracy studies) rather than human expert interpretation of raw data.

    4. Adjudication Method for the Test Set

    This information is not applicable/not provided as the ground truth is established through a quantitative comparison method, not expert consensus requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    This information is not applicable. This is not an AI/imaging device that involves human readers or an MRMC study. It is an in vitro diagnostic device for chemical analysis.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    This information is not applicable. This describes the performance of a reagent cartridge on a clinical analyzer, which is inherently a "standalone" measurement system for the biochemical assay it performs. There is no "algorithm" in the AI sense, nor a human-in-the-loop component for interpretation of the primary measurement.

    7. The Type of Ground Truth Used

    The ground truth was established by a "comparison method" using predicate or reference devices/methods. This means the results from the S-Test cartridges on the S40 Clinical Analyzer were compared against measurements from other commercially available, legally marketed diagnostic systems (e.g., ACE plus ISE/Clinical Chemistry System, Olympus AU640 Clinical Chemistry Analyzer, Piccolo® xpress Chemistry Analyzer reagents or similar) that measure inorganic phosphorus and uric acid.

    8. The Sample Size for the Training Set

    This information is not provided. Given that this is a traditional in vitro diagnostic device measuring chemical concentrations, there isn't a "training set" in the machine learning sense. The device's performance characteristics (precision, accuracy, sensitivity) are inherently part of its design and manufacturing process, and validated through the performance studies described, rather than "trained" on data.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable as there is no "training set" in the context of this device. The reagents and analyzer are developed and manufactured to perform the specified chemical reactions, with their analytical performance validated through studies against established methods.

    Ask a Question

    Ask a specific question about this device

    K Number
    K070249

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2007-04-26

    (90 days)

    Product Code
    Regulation Number
    862.1145
    Age Range
    All
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Device Class Classification Name: Product Code:

    ABX PENTRA Phosphorus CP

    Phosphorus Class I §862.1580

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ABX PENTRA Calcium CP reagent with associated calibrators and controls are intended for use on ABX PENTRA 400 Clinical Chemistry Analyzer for quantitative in vitro diagnostic determination of calcium in human serum, plasma and urine based on a photometric test using orthocresolphtalein complexone. Calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany (intermittent muscular contractions or spasms).

    ABX PENTRA Creatinine CP reagent with associated calibrators and controls are intended for use on ABX PENTRA 400 Clinical Chemistry Analyzer for quantitative in vitro diagnostic determination of creatinine in human serum, plasma and urine based on a kinetic method using alkaline picrate (Jaffé method). Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.

    ABX PENTRA Phosphorus CP reagent with associated calibrators and controls are intended for use on ABX PENTRA 400 Clinical Chemistry Analyzer for quantitative in vitro diagnostic determination of phosphorus in human serum, plasma and urine based on a UV method using phosphomolybdate. Measurements of phosphorus (inorganic) are used in the diagnosis and treatment of various disorders, including parathyroid gland and kidney diseases, and vitamin D imbalance.

    ABX PENTRA Amylase CP reagent with associated calibrators and controls are intended for use on ABX PENTRA 400 Clinical Chemistry Analyzer for quantitative in vitro diagnostic determination of the activity of the enzyme amylase in human serum, plasma and urine based on an enzymatic photometric assay. Amylase measurements are used primarily for the diagnosis and treatment of pancreatitis (inflammation of the pancreas).

    The ABX PENTRA Urine Control L/H is for use in quality control by monitoring accuracy and precision.

    Device Description

    All the reagents, controls and calibrators included in this submission are for use on the ABX PENTRA 400 (K052007), which is a discrete photometric benchtop clinical chemistry analyzer.

    The ABX PENTRA Calcium CP is an in vitro diagnostic assay for the quantitative determination of calcium in human serum, plasma and urine based on a photometric test using orthocresolphtalein complexone. It is composed of a bi-reagent cassette, with 66 ml and 16.5 ml compartments. Reagents are chemical solutions with additives.

    The ABX PENTRA Creatinine CP is an in vitro diagnostic assay for the quantitative determination of creatinine in human serum, plasma and urine based on a kinetic method using alkaline picrate (Jaffé method). It is composed of a bi-reagent cassette, with two 28 ml compartments. Reagents are chemical solutions with additives.

    The ABX PENTRA Phosphorus CP is an in vitro diagnostic assay for the quantitative determination of phosphorus in human serum, plasma and urine based on a UV method using phosphomolybdate. It is composed of a mono-reagent cassette, with 29.5 ml compartment. The reagent is a chemical solution with additives.

    The ABX PENTRA Amylase CP is an in vitro diagnostic assay for the quantitative determination of alpha-amylase in human serum, plasma and urine based on an enzymatic photometric test. The assay is composed of a bi-reagent cassette, with 26 ml and 6.5 ml compartments. Reagents are chemical solutions with additives.

    The ABX PENTRA Urine Control L/H is a two-level (Low and High) quality control consisting of liquid solutions prepared from human urine with chemical additives and materials of biological origin added as required to obtain given component levels. The assigned values of the control components are given in the enclosed annexe, ensuring control of the appropriate HORIBA ABX methods on the ABX PENTRA 400 analyzer. Each control level is provided in one vial of 10 ml.

    AI/ML Overview

    The provided text describes the performance data for new urine sample indications for four reagents (ABX PENTRA Calcium CP, ABX PENTRA Creatinine CP, ABX PENTRA Phosphorus CP, and ABX PENTRA Amylase CP) when used with the ABX PENTRA 400 Clinical Chemistry Analyzer. The study aims to demonstrate substantial equivalence to predicate devices and establish performance characteristics for urine samples.

    Here's a breakdown of the requested information based on the provided text:

    Acceptance Criteria and Device Performance

    The acceptance criteria are implied by the reported performance characteristics for each reagent on urine samples. While explicit "acceptance criteria" are not listed as pass/fail thresholds, the reported values represent the achieved performance that allowed for the 510(k) clearance.

    ABX PENTRA Calcium CP (Urine)

    MetricAcceptance Criteria (Implied)Reported Device Performance
    Sample typeUrineUrine
    Detection limitNot explicitly defined0.12 mg/dl
    Accuracy and PrecisionNot explicitly definedCV Total < 3.61%
    Measuring rangeNot explicitly defined0.12 mg/dl – 24.06 mg/dl
    Upper linearity limitNot explicitly defined24.06 mg/dl (48.12 mg/dl with auto-dilution)
    Correlation (n=124)Not explicitly definedY = 1.13 x - 0.50, r² = 0.9904
    Calibration stabilityNot explicitly defined6 hours
    Reagent stability (closed)Not explicitly defined24 months at 2-8°C
    Reagent stability (on-board)Not explicitly defined14 days

    ABX PENTRA Creatinine CP (Urine)

    MetricAcceptance Criteria (Implied)Reported Device Performance
    Sample typeUrineUrine
    Detection limitNot explicitly defined1.3 mg/dl
    Accuracy and PrecisionNot explicitly definedCV Total < 4.60%
    Measuring rangeNot explicitly defined1.3 mg/dl - 316.4 mg/dl
    Upper linearity limitNot explicitly defined316.4 mg/dl (949.2 mg/dl with auto-dilution)
    Correlation (n=110)Not explicitly definedY = 0.99 x - 0.98, r² = 0.9953
    Calibration stabilityNot explicitly defined1 day
    Reagent stability (closed)Not explicitly defined36 months at 2-8°C
    Reagent stability (on-board)Not explicitly defined7 days

    ABX PENTRA Phosphorus CP (Urine)

    MetricAcceptance Criteria (Implied)Reported Device Performance
    Sample typeUrineUrine
    Detection limitNot explicitly defined1.28 mg/dl
    Accuracy and PrecisionNot explicitly definedCV Total < 5.95%
    Measuring rangeNot explicitly defined1.28 mg/dl - 198.4 mg/dl
    Upper linearity limitNot explicitly defined198.4 mg/dl (396.8 mg/dl with auto-dilution)
    Correlation (n=119)Not explicitly definedY = 1.07 x - 1.10, r² = 0.9892
    Calibration stabilityNot explicitly defined34 days
    Reagent stability (closed)Not explicitly defined36 months at 2-8°C
    Reagent stability (on-board)Not explicitly defined34 days

    ABX PENTRA Amylase CP (Urine)

    MetricAcceptance Criteria (Implied)Reported Device Performance
    Sample typeUrineUrine
    Detection limitNot explicitly defined4.92 U/l
    Accuracy and PrecisionNot explicitly definedCV Total < 6.03%
    Measuring rangeNot explicitly defined4.92 U/l – 2000 U/l
    Upper linearity limitNot explicitly defined2000 U/l (6000 U/l with auto-dilution)
    Correlation (n=121)Not explicitly definedY = 1.17 x + 21.90, r² = 0.9891
    Calibration stabilityNot explicitly defined8 days
    Reagent stability (closed)Not explicitly defined24 months at 2-8°C
    Reagent stability (on-board)Not explicitly defined42 days

    Study Details

    1. Sample sizes used for the test set and the data provenance:

      • ABX PENTRA Calcium CP: n=124 for correlation studies.
      • ABX PENTRA Creatinine CP: n=110 for correlation studies.
      • ABX PENTRA Phosphorus CP: n=119 for correlation studies.
      • ABX PENTRA Amylase CP: n=121 for correlation studies.

      The text does not specify the country of origin of the data or whether the data was retrospective or prospective. Given the company is based in France, it's possible the studies were conducted there. The context of a 510(k) submission generally implies prospective studies or well-controlled retrospective studies for performance validation.

    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
      The text does not provide information on the number or qualifications of experts used to establish ground truth.

    3. Adjudication method for the test set:
      The text does not specify any adjudication method. For chemical assays, ground truth is typically established by comparative methods (predicate device or accepted reference method), not expert adjudication.

    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
      This information is not applicable. The device described consists of chemical reagents and controls for a clinical chemistry analyzer, not an AI or imaging device with "human readers."

    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
      This information is not applicable as the device is a chemical reagent and not an algorithm. The reported performance characteristics (detection limit, accuracy, precision, measuring range, linearity, correlation, stability) are a standalone performance of the reagent on the ABX PENTRA 400 analyzer.

    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
      For the correlation studies, the ground truth was established by comparison (correlation) with another method (presumably the predicate device or a recognized reference method), as indicated by the "Y = mx + b" equations and correlation coefficients (r²). The text states: "Correlation (n=XXX) Y = (value) with a correlation coefficient r² = (value)." This implies a comparison against a reference method.

    7. The sample size for the training set:
      The text does not provide information about a "training set." For these types of diagnostic assays, methods are developed and optimized through R&D, and then validated with test sets. The concept of a distinct "training set" as used in machine learning is generally not directly applicable.

    8. How the ground truth for the training set was established:
      Not applicable, as a "training set" in the machine learning sense is not mentioned or implied for these chemical assays.

    Ask a Question

    Ask a specific question about this device

    K Number
    K060205

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2006-07-21

    (176 days)

    Regulation Number
    862.1050
    Age Range
    All
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Test System JGJ ; photometric method, magnesium

    ABX PENTRA Phosphorus CP

    Phosphorus Class II &862.1580

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    General Chemistries reagents, with associated calibrators and controls, are intended for use on ABX PENTRA 400 Clinical Chemistry Analyzer to measure a variety of analytes.

    ABX PENTRA ALP CP reagent with associated calibrators and controls are for quantitative in vitro diagnostic determination of alkaline phosphatase in human serum and plasma based on a kinetic photometric test using p-Nitrophenylphosphate. Measurements of alkaline phosphatase or its isoenzymes are used in the diagnosis and treatment of liver, bone, parathyroid, and intestinal diseases.

    ABX PENTRA Calcium CP reagent with associated calibrators and controls are for quantitative in vitro diagnostic determination of calcium in human serum and plasma based on a photometric test using orthocresolphtalein complexone. Calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany (intermittent muscular contractions or spasms).

    ABX PENTRA CO2 RTU reagent with associated calibrators and controls are for quantitative in vitro diagnostic determination of carbon dioxide in human serum and plasma based on an enzymatic test using phosphoenolpyruvate (PEP), phosphorenolpyruvate carboxylase (PEPC) and an analog of NADH. Bicarbonate/carbon measurements are used in the diagnosis and treatment of numerous potentially serious disorders associated with changes in body acid-base balance.

    ABX PENTRA Creatinine CP reagent with associated calibrators and controls are for quantitative in vitro diagnostic determination of creatinine in human serum and plasma based on a kinetic method using alkaline picrate (Jaffé method). Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.

    ABX PENTRA Iron CP reagent with associated calibrators and controls are for quantitative in vitro diagnostic determination of iron (non-heme) in human serum and plasma based on a photometric test (Ferene method). Iron (non-heme) measurements are used in the diagnosis and treatment of diseases such as iron deficiency anemia and hemochromatosis.

    ABX PENTRA Magnesium RTU reagent with associated calibrators and controls are for quantitative in vitro diagnostic determination of magnesium in human serum and plasma based on a photometric test using xylidyl blue. Magnesium measurements are used in the diagnosis and treatment of hypomagnesemia (abnormally low plasma levels of magnesium) and hypermagnesemia (abnormally high plasma levels of magnesium).

    ABX PENTRA Phosphorus CP reagent with associated calibrators and controls are for quantitative in vitro diagnostic determination of phosphorus in human serum and plasma based on a UV method using phosphomolybdate. Measurement of phosphorus (inorganic) are used in the diagnosis and treatment of various disorders, including parathyroid gland and kidney diseases, and vitamin D imbalance.

    ABX PENTRA Urea CP reagent with associated calibrators and controls are for quantitative in vitro diagnostic determination of urea nitrogen (an end-product of nitrogen metabolism) in human serum and plasma based on an enzymatic UV test using urease and glutamate dehydrogenase. Measurements obtained by this device are used in the diagnosis and treatment of certain renal and metabolic diseases.

    ABX PENTRA Uric Acid CP reagent with associated calibrators and controls are for quantitative in vitro diagnostic determination of uric acid in human serum and plasma based on the enzymatic determination of uric acid using a chromogenic system in the presence of peroxidase and uricase (Trinder method). Measurements obtained by this device are used in the diagnosis and treatment of numerous renal and metabolic disorders, including renal failure, gout, leukemia, psoriasis, starvation or other wasting conditions, and of patients receiving cytotoxic drugs.

    The ABX PENTRA CO2 Cal is a calibrator for use in the calibration of quantitative Horiba ABX PENTRA CO2 RTU method on Horiba ABX clinical chemistry analyzers as specified on the vial.

    The ABX PENTRA CO₂ Control is for use in quality control by monitoring accuracy and precision for the quantitative ABX PENTRA CO2 RTU method as specified in the enclosed annex.

    Device Description

    All the reagents, controls and calibrators included in this submission are for use on the ABX PENTRA 400 (K052007), which is a discrete photometric benchtop clinical chemistry analyzer.

    The ABX PENTRA 400 offers both Closed and Open channels for a multitude of parameters (clinical chemistry, DAT, TDM, plasma protein, hemostasis, optional ISE module).

    All reagents described in this submission are for the quantitative in-vitro determination of their respective parameters

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Horiba ABX PENTRA reagents, controls, and calibrators, based on the provided 510(k) summary:

    This submission is for multiple reagents, controls, and calibrators for the ABX PENTRA 400 clinical chemistry analyzer. The performance data is presented for each reagent individually. The acceptance criteria and reported performance will be summarized for each reagent as presented in the document.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are generally implied by the reported performance, which demonstrates that the devices "met all acceptance criteria" and are "substantially equivalent" to their respective predicate devices. Specific quantitative acceptance criteria are not explicitly stated as separate targets but are shown by the achieved performance values.

    ABX PENTRA ALP CP (Alkaline phosphatase)

    Acceptance Criteria (Implied)Reported Device Performance
    Detection limit met6 U/l
    Accuracy and Precision metCV Total < 4.36%
    Measuring range met6 U/l – 1500 U/l (Automatic post-dilution: 6000 U/l)
    Correlation metY = 1.07 x - 3.93 with r² = 0.998
    Calibration stability met7 hours
    Reagent stability metclosed: 18 months at 2-8°C, on-board: 29 days

    ABX PENTRA Calcium CP (Calcium)

    Acceptance Criteria (Implied)Reported Device Performance
    Detection limit met0.16 mg/dl
    Accuracy and Precision metCV Total < 1.67%
    Measuring range met0.16 mg/dl – 20.1 mg/dl (Automatic post-dilution: 40.2 mg/dl)
    Correlation metY = 1.15 x - 1.12 with r² = 0.950
    Calibration stability met6 hours
    Reagent stability metclosed: 24 months at 2-8°C, on-board: 14 days

    ABX PENTRA CO2 RTU (Carbon Dioxide)

    Acceptance Criteria (Implied)Reported Device Performance
    Detection limit met1.8 mmol/l
    Accuracy and Precision metCV Total < 7.7%
    Measuring range met1.8 mmol/l - 60.8 mmol/l (Automatic post-dilution: 121.6 mmol/l)
    Correlation metY = 0.85 x + 0.3 with r² = 0.976
    Calibration stability met1 day
    Reagent stability metclosed: 16 months at 2-8°C, on-board: 28 days

    ABX PENTRA Creatinine CP (Creatinine)

    Acceptance Criteria (Implied)Reported Device Performance
    Detection limit met0.11 mg/dl
    Accuracy and Precision metCV Total < 3.69%
    Measuring range met0.11 mg/dl - 15.8 mg/dl (Automatic post-dilution: 79 mg/dl)
    Correlation metY = 0.96 x - 0.12 with r² = 0.996
    Calibration stability met1 day
    Reagent stability metclosed: 36 months at 2-8°C, on-board: 7 days

    ABX PENTRA Iron CP (Iron)

    Acceptance Criteria (Implied)Reported Device Performance
    Detection limit met7.42 µg/dl
    Accuracy and Precision metCV Total < 3.61%
    Measuring range met7.42 µg/dl – 1004 µg/dl (Automatic post-dilution: 5020 µg/dl)
    Correlation metY = 1.13 x + 5.30 with r² = 0.997
    Calibration stability met8 days
    Reagent stability metclosed: 24 months at 2-8°C, on-board: 41 days

    ABX PENTRA Magnesium RTU (Magnesium)

    Acceptance Criteria (Implied)Reported Device Performance
    Detection limit met0.17 mg/dl
    Accuracy and Precision metCV Total < 3.19%
    Measuring range met0.17 mg/dl - 4.64 mg/dl (Automatic post-dilution: 13.92 mg/dl)
    Correlation metY = 1.23 x - 0.24 with r² = 0.971
    Calibration stability met2 days
    Reagent stability metclosed: 24 months at 2-8°C, on-board: 7 days

    ABX PENTRA Phosphorus CP (Phosphorus)

    Acceptance Criteria (Implied)Reported Device Performance
    Detection limit met0.28 mg/dl
    Accuracy and Precision metCV Total < 3.56%
    Measuring range met0.28 mg/dl – 24.18 mg/dl (Automatic post-dilution: 96.72 mg/dl)
    Correlation metY = 1.04 x + 0.19 with r² = 0.997
    Calibration stability met34 days
    Reagent stability metclosed: 36 months at 2-8°C, on-board: 34 days

    ABX PENTRA Urea CP (Urea)

    Acceptance Criteria (Implied)Reported Device Performance
    Detection limit metUrea: 1.86 mg/dl, BUN: 0.9 mg/dl
    Accuracy and Precision metCV Total < 2.76%
    Measuring range metUrea: 1.86 - 300 mg/dl (Auto post-dilution: 1500 mg/dl); BUN: 0.9 - 140.3 mg/dl (Auto post-dilution: 701.5 mg/dl)
    Correlation metUrea: Y = 1.01 x + 1.80 with r² = 0.9905; BUN: Y = 1.01 x + 0.81 with r² = 0.9905
    Calibration stability met8 days
    Reagent stability metclosed: 24 months at 2-8°C, on-board: 70 days

    ABX PENTRA Uric Acid CP (Uric Acid)

    Acceptance Criteria (Implied)Reported Device Performance
    Detection limit met0.19 mg/dl
    Accuracy and Precision metCV Total < 2.81%
    Measuring range met0.19 mg/dl - 25 mg/dl (Automatic post-dilution: 75 mg/dl)
    Correlation metY = 0.97 x - 0.13 with r² = 0.958
    Calibration stability met15 days
    Reagent stability metclosed: 36 months at 2-8°C, on-board: 41 days

    Controls and Calibrators Stability

    ProductAcceptance Criteria (Implied)Reported Device Performance
    ABX PENTRA CO2 CalStability metclosed: 15 months at 2-25°C, open: 3 months at 2-25°C
    ABX PENTRA MulticalStability metclosed: 24 months at 2-8°C, open: 8 hours at 15-25°C, 2 days at 2-8°C, 2 weeks at -25-(-15)°C (with exceptions for Bilirubins)
    ABX PENTRA CO2 ControlStability metclosed: 15 months at 2-25°C, open: 3 months at 2-25°C
    ABX PENTRA N ControlStability metclosed: 30 months at 2-8°C, open: 12 hours at 15-25°C, 5 days at 2-8°C, 1 month at -25-(-15)°C (with exceptions for Bilirubins)
    ABX PENTRA P ControlStability metclosed: 30 months at 2-8°C, open: 12 hours at 15-25°C, 5 days at 2-8°C, 1 month at -25-(-15)°C (with exceptions for Bilirubins)

    2. Sample Sizes Used for the Test Set and Data Provenance

    The sample sizes for the correlation studies (which serve as the primary test set for comparing to predicate devices) are as follows:

    • ABX PENTRA ALP CP: n=105
    • ABX PENTRA Calcium CP: n=95
    • ABX PENTRA CO2 RTU: n=97
    • ABX PENTRA Creatinine CP: n=95
    • ABX PENTRA Iron CP: n=98
    • ABX PENTRA Magnesium RTU: n=75
    • ABX PENTRA Phosphorus CP: n=105
    • ABX PENTRA Urea CP: n=108
    • ABX PENTRA Uric Acid CP: n=98

    The provenance of the data (country of origin, retrospective/prospective) is not explicitly stated in the provided summary. However, the company is based in France and the submission is for the US market. The samples are described generally as "human serum and plasma".

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    This information is not applicable as the devices are in vitro diagnostic reagents for quantitative measurement, not diagnostic imaging devices requiring expert interpretation for ground truth. The "ground truth" for correlation studies would typically be the results obtained from a legally marketed predicate device or a reference method.

    4. Adjudication Method for the Test Set

    This is not applicable for this type of in vitro diagnostic device study. Adjudication methods are typically employed in studies involving human interpretation or subjective assessments.

    5. If a Multi Reader Multi Case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable as the devices are in vitro diagnostic reagents and not AI-powered diagnostic tools for human interpretation.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    The entire performance data presented is for the standalone (algorithm only) performance of the reagents when used with the ABX PENTRA 400 analyzer. There is no human-in-the-loop component for these quantitative chemistry measurements.

    7. The type of ground truth used

    The "ground truth" for the performance studies, particularly the correlation data, is implicitly established by comparison to results from legally marketed predicate devices or established reference methods. This is indicated by the statement of "substantial equivalence" to named predicate devices and the correlation data that compares the results of the new device (Y) to an existing method (x).

    8. The Sample Size for the Training Set

    The document does not explicitly mention a "training set" in the context of machine learning or AI. For in vitro diagnostic devices like these, method development and optimization would precede the validation studies, but a distinct "training set" as understood in AI is not typically a component of the regulatory submission for this type of product. The performance data presented are for the validation of the finalized method.

    9. How the Ground Truth for the Training Set Was Established

    Given that a "training set" in the AI sense is not explicitly mentioned or applicable for these devices, the question of how its ground truth was established is not applicable. The development of the reagents involves standard chemical and analytical validation processes to ensure accuracy and precision, typically against reference standards and established methods, often during an internal R&D phase.

    Ask a Question

    Ask a specific question about this device

    K Number
    K061655

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2006-07-10

    (27 days)

    Regulation Number
    862.3030
    Age Range
    All
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    | 862.3240 Cholinesterase test system | Pseudocholinesterase test system |
    | 862.1580
    cartridge | Dimension® PHOSFlex® reagentcartridge | K861700 | I* | 862.1580

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Dimension Vista™ Acetaminophen (ACTM) Flex® reagent cartridge is a device intended to measure acetaminophen, an analgesic and antipyretic (fever reducing) drug, in serum and plasma. Measurements obtained by this device are used in the diagnosis and treatment of acetaminophen overdose.

    The Dimension Vista™ Amylase (AMY) Flex® reagent cartridge is a device intended to measure the activity of the enzyme amylase in serum, plasma and urine. Amylase measurements are used primarily for the diagnosis and treatment of pancreatitis (inflammation of the pancreas).

    The Dimension Vista™ Creatine Kinase (CK) Flex® reagent cartridge is a device intended to measure the activity of the enzyme creatine kinase in serum and plasma. Measurements of creatine kinase are used in the diagnosis and treatment of myocardial infarction and muscle diseases such as progressive Duchenne-type muscular dystrophy.

    The Dimension Vista™ Cholesterol (CHOL) Flex® reagent cartridge is a device intended to measure cholesterol in serum and plasma. Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders.

    The Dimension Vista™ Gamma-glutamyl transferase (GGT) Flex® reagent cartridge is a device intended to measure gamma-glutamyl transferase in human serum and plasma. Gamma-glutamyl transferase measurements are used in the diagnosis and treatment of liver diseases such as alcoholic cirrhosis and primary and secondary liver tumors.

    The Dimension Vista™ Glucose (GLU) Flex® reagent cartridge is a device intended to measure glucose in human serum, plasma, urine and cerebrospinal fluid. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal and idiopathic hypoglycemia, and pancreatic islet cell carcinoma.

    The Dimension Vista™ High-Density Lipoprotein Cholesterol (HDLC) Flex® reagent cartridge is intended to measure high-density lipoprotein cholesterol in serum and plasma. Measurements of high-density lipoprotein cholesterol are used in the diagnosis of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases.

    The Dimension Vista™ Low-Density Lipoprotein Cholesterol (LDLC) Flex® reagent cartridge is intended to measure low-density lipoprotein cholesterol in serum and plasma. Measurements of low-density lipoprotein cholesterol are used in the diagnosis of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases.

    The Dimension Vista™ Lidocaine (LIDO) Flex® reagent cartridge is a device intended to measure lidocaine, an antiarrythmic and anticonvulsant drug, in serum and plasma. Measurements obtained by this device are used in the diagnosis and treatment of lidocaine overdose or in monitoring levels of lidocaine to ensure appropriate therapy.

    The Dimension Vista™ Magnesium (MG) Flex® reagent cartridge is intended for the measurement of magnesium levels in serum and plasma. Magnesium measurements are used in the diagnosis and treatment of hypomagnesemia (abnormally low plasma levels of magnesium) and hypermagnesemia (abnormally high plasma levels of magnesium).

    The Dimension Vista™ Pseudocholinesterase (PCHE) Flex® reagent cartridge is a device intended to measure pseudocholinesterase activity in human serum and plasma. Measurements obtained by this device are used in the diagnosis and treatment of cholinesterase inhibition disorders (e.g., insecticide poisoning and succinylcholine poisoning).

    The Dimension Vista™ Phosphorus (PHOS) Flex® reagent cartridge is a device intended to measure inorganic phosphorus in serum, plasma, and urine. Measurements of phosphorus (inorganic) are used in the diagnosis and treatment of various disorders, including parathyroid gland and kidney diseases, and vitamin D imbalance.

    The Dimension Vista™ Procainamide (PROC) Flex® reagent cartridge is a device intended to measure procainamide in serum and plasma. Measurements obtained may be used in the diagnosis and treatment of procainamide overdose and in monitoring levels of procainamide to ensure appropriate therapy.

    The Dimension Vista™ Salicylate (SAL) Flex® reagent cartridge is a device intended to measure salicylates, a class of analgesic, antipyretic and anti-inflammatory drugs that includes aspirin, in human serum. Measurements obtained by this device are used in the diagnosis and treatment of salicylate overdose and in monitoring salicylate levels to ensure appropriate therapy.

    The Dimension Vista™ Thyroxine (T4) Flex® reagent cartridge is a device intended to measure total (free and protein bound) thyroxine (thyroid hormone) in serum and plasma. Measurements obtained by this device are used in the diagnosis and treatment of thyroid diseases.

    The Dimension Vista™ Tobramycin (TOBR) Flex® reagent cartridge is a device intended to measure tobramycin, an aminoglycoside antibiotic drug, in palsma and serum. Measurements obtained by this device are used in the diagnosis and treatment of tobramycin overdose and in monitoring levels of tobramycin to ensure appropriate therapy.

    The Dimension Vista™ Triglyceride (TRIG) Flex® reagent cartridge is a device intended to measure triglyceride (neutral fat) in serum and plasma. Measurements obtained by this device are used in the diagnosis and treatment of patients with diabetes mellitus, nephrosis, liver obstruction, other diseases involving lipid metabolism, or various endocrine disorders.

    The Dimension Vista™ Uric Acid (URCA) Flex® reagent cartridge is a device intended to measure uric acid in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of numerous renal and metabolic disorders, including renal failure, gout, leukemia, psoriasis, starvation or other wasting conditions, and of patients receiving cytotoxic drugs.

    The Dimension Vista™ Valproic Acid (VALP) Flex® reagent cartridge is a device intended to measure valproic acid, an anti-convulsant drug in serum and plasma. Measurements obtained may be used in the diagnosis and treatment of valproic acid overdose and in monitoring levels of valproic acid to ensure appropriate therapy.

    The Dimension Vista™ Vancomycin (VANC) Flex® reagent cartridge is a device intended to measure vancomycin, an antibiotic drug, in serum and plasma. Measurements obtained by this device are used in the diagnosis and treatment of vancomycin overdose and in monitoring the level of vancomycin to ensure appropriate therapy.

    Device Description

    Dade Behring Dimension Vista™ Flex® reagent cartridges are prepackaged in-vitro diagnostic test methods (assays) that are specifically designed to be used on the Vade Behring Dimension Vista™ Integrated system, a floor model, fully automated, microprocessor-controlled, integrated instrument system. The Dimension Vista™ system was previously cleared with seven associated test methods (K 051087). This Special 510(k) is submitted for a packaging modification to in-vitro diagnostic devices that have been cleared under the 510(k) process for use on Dimension® clinical chemistry systems. The packaging change is to allow use on the Dimension Vista™ system.

    The reagents contained in the Dimension Vista™ Flex® reagent cartridges are the same as those contained in the Flex® reagent cartridges manufactured for the Dimension® clinical chemistry systems, another family of Dade Behring analyzers. The packaging modification, does not affect the intended use of the devices, nor does it alter the fundamental scientific technology of the devices.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Dade Behring Dimension Vista™ Flex® reagent cartridges, based on the provided 510(k) summary:

    This device submission is a Special 510(k) for a packaging modification, meaning the core technology and reagents are the same as previously cleared devices. Therefore, the primary goal of the study is to demonstrate substantially equivalent performance after the packaging change, rather than to establish initial performance claims for a novel device.


    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of numerical acceptance criteria or specific performance metrics (e.g., accuracy, precision values) for each analyte. Instead, it relies on a comparative equivalency approach to a predicate device.

    The overarching acceptance criterion is "substantially equivalent performance" to the predicate Dimension® Flex® reagent cartridges.

    Acceptance CriterionReported Device Performance (Summary)
    Substantial Equivalence to Predicate Device"Comparative testing described in the protocol included in this submission demonstrates substantially equivalent performance."
    Same Intended Use and Indications for UseConfirmed; the packaging modification does not affect intended use or indications.
    Same Reagents and Fundamental Scientific TechnologyConfirmed; reagents are the same, and the fundamental scientific technology is unaltered.

    2. Sample Size Used for the Test Set and Data Provenance

    The document states: "Comparative testing described in the protocol included in this submission demonstrates substantially equivalent performance."

    • Sample Size for Test Set: This information is not explicitly stated in the provided summary. The summary refers to a "protocol included in this submission," which would contain these details.
    • Data Provenance: This information is not explicitly stated in the provided summary.
    • Retrospective or Prospective: This information is not explicitly stated. However, given the nature of in-vitro diagnostic testing for performance comparison, it would typically involve prospective testing on patient samples or spiked samples.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This is an in-vitro diagnostic device for quantitative measurement of analytes in human samples (serum, plasma, urine, CSF). The ground truth for such devices is established by:

    • Reference Methods: Highly accurate and precise laboratory methods, often gold standards like GC-MS, HPLC, or other well-validated enzymatic or spectrophotometric methods.
    • Certified Reference Materials (CRMs): Samples with known, certified concentrations of the analytes.

    Therefore, the concept of "experts" in the clinical imaging or diagnostic interpretation sense (e.g., radiologists) is not applicable here. The ground truth is laboratory-based and instrumental.


    4. Adjudication Method for the Test Set

    Not applicable for this type of in-vitro diagnostic device. Ground truth is established by reference methods or certified materials, not by expert consensus or adjudication.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    • No, an MRMC comparative effectiveness study was not done.
    • This device is an in-vitro diagnostic reagent cartridge, not an AI-powered diagnostic imaging tool or a system designed for human interpretation with or without AI assistance. The performance is measured instrumentally.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • Yes, the performance evaluated is inherently "standalone" in the context of an automated analytical instrument. The Flex® reagent cartridges are designed to be used on the Dimension Vista™ Integrated system, a "fully automated, microprocessor-controlled, integrated instrument system." The performance of the reagent (device) is measured by its output on this automated system.
    • There is no "human-in-the-loop" decision-making component for the measurement process itself, although clinical interpretation of the results by a healthcare professional is expected.

    7. The Type of Ground Truth Used

    The ground truth for this type of in-vitro diagnostic device would typically involve:

    • Reference Method Assays: Using established, highly accurate, and precise laboratory methods (e.g., a recognized primary reference measurement procedure or a well-characterized predicate device itself) to determine the true concentration of the analytes in the test samples.
    • Certified Reference Materials: Commercial or internal standards with known, traceable concentrations of the analytes.
    • Sample Matrix: Patient samples (serum, plasma, urine, CSF) with concentrations spanning the analytical range.

    The summary states "Comparative testing... demonstrates substantially equivalent performance." This strongly implies that the new device's measurements were compared against the measurements obtained by the predicate device on the same samples, which serves as the "reference" or "ground truth" for the equivalence claim.


    8. The Sample Size for the Training Set

    This device is a reagent cartridge for an in-vitro diagnostic test, not a machine learning or AI algorithm in the contemporary sense that requires a "training set" to learn. The reagents and their chemical reactions are based on established scientific principles.

    Therefore, the concept of a "training set" as understood in machine learning is not applicable to this device.


    9. How the Ground Truth for the Training Set Was Established

    As noted above, the concept of a "training set" is not applicable to this device. The ground truth for the performance evaluation (test set) would be established by reference methods or comparison to the predicate device, as described in point 7.

    Ask a Question

    Ask a specific question about this device

    K Number
    K041643

    Validate with FDA (Live)

    Date Cleared
    2004-08-18

    (62 days)

    Product Code
    Regulation Number
    862.1580
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Systems Phosphorus (PHS) Reagent

    3.2 Classification Name Phosphorus (inorganic) test system (21 CFR § 862.1580

    :

    Korrollo
    Trade/Device Name: SYNCHRON® Systems Phosphorus (PHS) Reagent Regulation Number: 21 CFR 862.1580

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    PHS reagent, in conjunction with SYNCHRON® Systems and SYNCHRON PTIS reagent, in ochjanoued for use in the quantitative determination of inorganic phosphorus in human serum, plasma, or urine.

    Measurements of phosphorus (inorganic) are used in the diagnosis and incasuromento of in disorders, including parathyroid gland and kidney diseases, and vitamin D imbalance.

    Device Description

    The SYNCHRON Systems PHS reagent is designed for optimal performance on the CX (CX4CE/4/4PRO, CX5CE/52/5PRO, CX7/7RTS/7A/7PRO, SYNCHRON CX9ALX/9PRO) and LX (LX20/PRO/LXi 725) Systems. The reagent kit contains two 300-test cartridges that are packaged separately from the associated calibrator.

    AI/ML Overview

    Here's an analysis of the provided 510(k) summary, specifically focusing on the acceptance criteria and the study that demonstrates the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    The 510(k) summary for the SYNCHRON® Systems Phosphorus (PHS) Reagent does not explicitly state pre-defined acceptance criteria with numerical targets (e.g., "Slope must be between 0.95 and 1.05"). Instead, it presents performance data (method comparison, linearity, and imprecision) and implies that these results demonstrate substantial equivalence to the predicate device.

    However, based on the common practices for in vitro diagnostic (IVD) device submissions and the provided predicate comparison, we can infer the implied acceptance criteria:

    SectionImplied Acceptance Criteria (Inferred)Reported Device Performance and How it Meets Criteria
    Method ComparisonThe candidate device (SYNCHRON® Systems PHS Reagent) should show strong correlation and agreement with the predicate device (Beckman Coulter SYNCHRON PO4 Assay on CX Systems) when measuring inorganic phosphorus in human serum and urine. This is typically assessed by: - Slope: Close to 1.0 - Intercept: Close to 0.0 - R (Correlation Coefficient): Close to 1.0 (indicating a strong linear relationship)Serum: - Slope: 0.997 (Very close to 1.0) - Intercept: -0.04 (Very close to 0.0) - R: 0.998 (Excellent linear correlation) Urine: - Slope: 0.976 (Close to 1.0) - Intercept: 0.25 (Reasonably close to 0.0) - R: 0.996 (Excellent linear correlation) Meets Criteria: The reported values for both serum and urine demonstrate excellent agreement and strong correlation with the predicate device, indicating substantial equivalence in measurement accuracy across relevant sample types.
    ImprecisionThe device should demonstrate acceptable within-run and total imprecision (reproducibility) for various concentrations of inorganic phosphorus in serum and urine. This is usually assessed by a low %CV (Coefficient of Variation) at different concentration levels. The acceptable %CV values would be determined by clinical relevance and comparison to the predicate or similar devices.Within-Run Imprecision: - Serum 1 (2.0 mg/dL): 2.4 %C.V. - Serum 2 (6.6 mg/dL): 1.4 %C.V. - Urine 1 (41.1 mg/dL): 1.0 %C.V. - Urine 2 (78.3 mg/dL): 1.2 %C.V. Total Imprecision: - Serum 1 (2.0 mg/dL): 2.7 %C.V. - Serum 2 (6.6 mg/dL): 1.5 %C.V. - Urine 1 (41.1 mg/dL): 1.5 %C.V. - Urine 2 (78.3 mg/dL): 1.6 %C.V. Meets Criteria: The %C.V. values for both within-run and total imprecision are consistently low (all below 3%), indicating excellent precision and reproducibility across the tested concentrations and sample types. This is generally considered acceptable for clinical chemistry assays.
    LinearityThe device should provide measurements that are linear across its claimed analytical measurement range. (While not explicitly detailed with data in the summary, linearity experiments are mentioned as being performed to demonstrate equivalence).Reported: "Equivalence is demonstrated through method comparison, linearity, and imprecision experiments." Meets Criteria: The submission states that linearity experiments were performed and contributed to the finding of substantial equivalence. While specific data isn't provided, its inclusion in the summary implies acceptable results.

    2. Sample Sizes Used for the Test Set and Data Provenance

    • Method Comparison Test Set:
      • Serum: 114 samples (n=114)
      • Urine: 81 samples (n=81)
    • Imprecision Test Set:
      • For each of the 4 samples (Serum 1, Serum 2, Urine 1, Urine 2), the "N" value listed is 80. This typically refers to the number of replicates or data points used in the imprecision study.
    • Data Provenance: Not explicitly stated. For 510(k) submissions, data is typically generated internally by the manufacturer or by a contracted lab. The document does not specify the country of origin or if the data was retrospective or prospective. Given the nature of a reagent performance study, it's most likely prospective data collected specifically for the submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable to this type of device (in vitro diagnostic reagent). For a phosphorus reagent, the "ground truth" is established by a reference method or a highly accurate clinical analyzer (in this case, the predicate device, Beckman Coulter SYNCHRON PO4 Assay on CX Systems). Expert consensus or human interpretation (like in imaging studies) is not used to establish the "ground truth" for quantitative chemical measurements.

    4. Adjudication Method for the Test Set

    This is not applicable. Adjudication methods (e.g., 2+1, 3+1) are used when human interpretation or expert opinion is involved in establishing a ground truth for qualitative or semi-quantitative results, particularly in areas like medical imaging. For a quantitative chemical assay, the comparison is directly between the candidate device's numerical output and the predicate device's numerical output.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, an MRMC study was not done. This type of study is relevant for devices where human readers interpret diagnostic information (e.g., radiologists reading images) and the AI aims to assist or replace that human interpretation. The SYNCHRON PHS Reagent is an automated in vitro diagnostic assay where the output is a numerical value, not an interpretation by a human reader.

    6. Standalone (Algorithm Only) Performance Study

    Yes, a standalone performance study was done. The entire performance summary (method comparison, linearity, and imprecision) is a demonstration of the SYNCHRON PHS Reagent's in vitro performance, independent of human interpretation or intervention beyond operating the analyzer. The results presented directly reflect the algorithm's (reagent's) ability to measure phosphorus.

    7. Type of Ground Truth Used

    The ground truth for the method comparison study was established by the predicate device, the Beckman Coulter SYNCHRON PO4 Assay on CX Systems. For imprecision studies, the ground truth is the "true" concentration of the sample, which is often determined through meticulous preparation of control materials or by a highly accurate reference method.

    8. Sample Size for the Training Set

    This information is not provided and is not typically relevant for a traditional chemical reagent assay like the SYNCHRON PHS Reagent. Chemical reagents are designed based on established chemical reactions and principles. They do not typically involve "training sets" in the machine learning sense. The performance data presented (method comparison, imprecision) serves as the validation of the reagent's performance.

    9. How the Ground Truth for the Training Set Was Established

    As explained in point 8, a "training set" in the machine learning sense is not applicable here. The development of a chemical reagent involves chemical formulation, optimization of reaction conditions, and calibration using known reference materials. The "ground truth" for calibration and optimization would be highly characterized reference materials with known concentrations of inorganic phosphorus.

    Ask a Question

    Ask a specific question about this device

    K Number
    K030015

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2003-03-07

    (64 days)

    Product Code
    Regulation Number
    862.1580
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    CA 92821

    K030015 Trade/Device Name: ATAC PAK Phosphorus Reagent Regulation Number: 21 CFR 862.1580

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ATAC PAK Phosphorus Reagent Kit is intended for use with the ATAC Calibrator and the ATAC 8000 Random Access Chemistry System as a system for the quantitative determination of inorganic phosphorus in serum and plasma. Phosphorus results are used for the diagnosis and treatment of various disorders, including parathyroid gland and kidney diseases, and vitamin D imbalance.

    Device Description

    The ATAC PAK Phosphorus Reagent determines phosphorus through its reaction with molybdate to form a phosphomolybdate complex. The resulting increase in absorbance at 340 nm is proportional to the phosphorus concentration in the sample.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study details for the ATAC PAK Phosphorus Reagent Kit, based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    Performance MetricAcceptance Criteria (Implied)Reported Device Performance
    Linearity RangeN/A (demonstrated over usable range)0.1 to 15 mg/dL
    Linearity Regression (Slope)Close to 1.000.991
    Linearity Regression (Intercept)Close to 0-0.05 mg/dL
    Linearity Regression (Correlation Coefficient, r)Close to 1.0001.000
    Linearity Regression (Standard Error of Y, SYX)N/A0.11 mg/dL
    Precision (Within-Run %CV)N/A (compared to predicate or industry standard, generally low)Serum 1: 3.2%
    Precision (Total %CV)N/A (compared to predicate or industry standard, generally low)Serum 1: 5.0%
    Method Comparison (Deming Regression Slope)Close to 1.001.001
    Method Comparison (Deming Regression Intercept)Close to 0-0.17 mg/dL
    Method Comparison (Standard Error of Regression, Syx)N/A0.17 mg/dL
    Detection LimitN/A (should be clinically acceptable for phosphorus)0.1 mg/dL
    Onboard Reagent StabilityTotal imprecision < 0.2 mg/dL or 4%Claimed and documented: < 0.2 mg/dL or 4% over 14 days
    Calibration StabilityTotal imprecision < 0.2 mg/dL or 4%Claimed and documented: < 0.2 mg/dL or 4% over 7 days

    Note on "Acceptance Criteria (Implied)": The document performs a substantial equivalence claim against an existing device (Beckman Synchron Phosphorus Reagent). Therefore, the "acceptance criteria" are not explicitly stated as hard numerical thresholds within this document, but rather are demonstrated by showing performance comparable to, or within acceptable limits for, clinical laboratory assays of this type, often implicitly benchmarked against the predicate device or established industry guidelines (like NCCLS EP3-T for precision). For regression analyses, slopes near 1, intercepts near 0, and high correlation coefficients (r-values) are generally considered good.

    2. Sample sized used for the test set and the data provenance

    • Linearity: 27 samples (implied from "n=27" associated with linearity regression, though the context indicates "% of the rod" which is a typo and likely meant "range" or "rod" as a unit for linearity standards). The samples used were "linearity standards that span the usable range."
    • Precision:
      • Serum 1: n = 60 replicates
      • Serum 2: n = 60 replicates
      • Serum 3: n = 60 replicates
        The samples were "commercially available control serum."
    • Method Comparison: n = 115 patient specimens. The specimens were "Mixed serum and plasma specimens, collected from adult patients." The data provenance is not explicitly stated as country of origin, but it is retrospective as the specimens were already collected.
    • Detection Limit: 30 replicates of a diluted serum pool.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not provided in the document. For an in vitro diagnostic (IVD) like a phosphorus reagent, "ground truth" is typically established by:

    • Reference methods for linearity standards.
    • Certified values for control sera used in precision studies.
    • The results from a well-established, legally marketed comparative method (predicate device) in method comparison studies.
      No human experts are explicitly mentioned as establishing ground truth in this context; rather, the "truth" is derived from established analytical methods and reference materials.

    4. Adjudication method for the test set

    Not applicable. The "adjudication method" (e.g., 2+1, 3+1) typically refers to a process where multiple human readers independently interpret data, and discrepancies are resolved by a consensus or a senior expert. This concept is relevant for imaging or diagnostic algorithms where human interpretation is the primary ground truth. For an IVD reagent, the performance is assessed against quantitative analytical metrics, not human adjudication of results.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No. This document describes a new reagent for quantitative chemical analysis, not an AI or imaging diagnostic device that would involve human readers or MRMC studies.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The entire study described is a standalone performance study of the ATAC PAK Phosphorus Reagent Kit on the ATAC 8000 Random Access Chemistry System. It assesses the analytical performance of the reagent (linearity, precision, method comparison, detection limit, stability) without human intervention in the result generation itself, beyond standard laboratory practices for sample handling and system operation. There is no human-in-the-loop component described for the device's performance.

    7. The type of ground truth used

    • Linearity: Based on "standard values" from "linearity standards that span the usable range." These standards would have known, certified phosphorus concentrations.
    • Precision: Based on "commercially available control serum" which have established target values and acceptable ranges.
    • Method Comparison: The "competitive reagent" (predicate device) serves as the "ground truth" or reference for comparison. The assumption is that the predicate device is accurate and provides clinically acceptable results.
    • Detection Limit: Derived from the statistical analysis of repetitive assays of a "diluted serum pool."
    • Stability: Assessed against the initial, presumably accurate, values of "serum controls" over time.

    In summary, the ground truth is primarily based on reference materials, certified controls, and comparison to an established predicate device.

    8. The sample size for the training set

    Not applicable. This device is an in vitro diagnostic reagent, not an AI or machine learning algorithm that requires a "training set." Its performance is based on chemical reactions and photometric measurements, validated through analytical studies.

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for this type of device.