Found 826 results
510(k) Data Aggregation
(111 days)
Glucose HK Gen.3; ONLINE DAT Methadone II; cobas pro integrated solutions
Regulation Number: 21 CFR 862.1345
(photometric, discrete), for Clinical Use |
| Product Codes, Regulation Numbers | • CFR, 21 CFR 862.1345
cobas pro integrated solutions is an automated analyzer, intended for running qualitative, semiquantitative, and quantitative clinical chemistry and immunochemistry assays as well as ion-selective measurements.
Glucose HK Gen.3 is an in vitro test for the quantitative determination of glucose in human serum, plasma, urine and CSF on cobas c systems. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and pancreatic islet cell tumors.
Methadone II (MDN2) is an in vitro diagnostic test for the qualitative and semiquantitative detection of methadone in human urine on cobas c systems at a cutoff concentration of 300 ng/mL. Semiquantitative test results may be obtained that permit laboratories to assess assay performance as part of a quality control program. Semiquantitative assays are intended to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as gas chromatography/mass spectrometry (GC-MS).
Methadone II provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. GC-MS is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.
The cobas pro integrated solutions consists of a high throughput core unit (sample supply unit and sample buffer unit) and can be configured with different analytical units for immunology, clinical chemistry, and electrolyte testing. The cobas c 703 analytical unit being added to cobas pro integrated solutions is a new clinical chemistry analyzer intended for the in-vitro quantitative/qualitative determination of analytes in body fluids.
Glucose is phosphorylated by hexokinase (HK) in the presence of adenosine triphosphate (ATP) and magnesium ions to produce glucose-6-phosphate (G-6-P) and adenosine diphosphate (ADP). Glucose-6-phosphate dehydrogenase (G-6-PDH) specifically oxidizes G-6-P to 6-phosphogluconate with the concurrent reduction of nicotinamide adenine dinucleotide (NAD) to nicotinamide adenine dinucleotide reduced (NADH). One micromole of NADH is produced for each micromole of glucose consumed. The NADH produced absorbs light at 340 nm and can be detected spectrophotometrically as an increased absorbance.
The Methadone assay is based on the kinetic interaction of microparticles in a solution (KIMS) as measured by changes in light transmission. In the absence of sample drug, soluble drug conjugates bind to antibody-bound microparticles, causing the formation of particle aggregates. As the aggregation reaction proceeds in the absence of sample drug, the absorbance increases. When a urine sample contains the drug in question, this drug competes with the drug derivative conjugate for microparticle‑bound antibody. Antibody bound to sample drug is no longer available to promote particle aggregation, and subsequent particle lattice formation is inhibited. The presence of sample drug diminishes the increasing absorbance in proportion to the concentration of drug in the sample. Sample drug content is determined relative to the value obtained for a known cutoff concentration of drug.
N/A
Ask a specific question about this device
(111 days)
Glucose HK Gen.3; ONLINE DAT Methadone II; cobas pro integrated solutions
Regulation Number: 21 CFR 862.1345
(photometric, discrete), for Clinical Use |
| Product Codes, Regulation Numbers | • CFR, 21 CFR 862.1345
cobas pro integrated solutions is an automated analyzer, intended for running qualitative, semiquantitative, and quantitative clinical chemistry and immunochemistry assays as well as ion-selective measurements.
Glucose HK Gen.3 is an in vitro test for the quantitative determination of glucose in human serum, plasma, urine and CSF on cobas c systems. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and pancreatic islet cell tumors.
Methadone II (MDN2) is an in vitro diagnostic test for the qualitative and semiquantitative detection of methadone in human urine on cobas c systems at a cutoff concentration of 300 ng/mL. Semiquantitative test results may be obtained that permit laboratories to assess assay performance as part of a quality control program. Semiquantitative assays are intended to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as gas chromatography/mass spectrometry (GC-MS).
Methadone II provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. GC-MS is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.
cobas pro integrated solutions is an automated analyzer, intended for running qualitative, semiquantitative, and quantitative clinical chemistry and immunochemistry assays as well as ion-selective measurements.
The cobas pro integrated solutions consists of a high throughput core unit (sample supply unit and sample buffer unit) and can be configured with different analytical units for immunology, clinical chemistry, and electrolyte testing. The cobas c 703 analytical unit being added to cobas pro integrated solutions is a new clinical chemistry analyzer intended for the in-vitro quantitative/qualitative determination of analytes in body fluids.
Glucose is phosphorylated by hexokinase (HK) in the presence of adenosine triphosphate (ATP) and magnesium ions to produce glucose-6-phosphate (G-6-P) and adenosine diphosphate (ADP). Glucose-6-phosphate dehydrogenase (G-6-PDH) specifically oxidizes G-6-P to 6-phosphogluconate with the concurrent reduction of nicotinamide adenine dinucleotide (NAD) to nicotinamide adenine dinucleotide reduced (NADH). One micromole of NADH is produced for each micromole of glucose consumed. The NADH produced absorbs light at 340 nm and can be detected spectrophotometrically as an increased absorbance.
The Methadone assay is based on the kinetic interaction of microparticles in a solution (KIMS) as measured by changes in light transmission. In the absence of sample drug, soluble drug conjugates bind to antibody-bound microparticles, causing the formation of particle aggregates. As the aggregation reaction proceeds in the absence of sample drug, the absorbance increases. When a urine sample contains the drug in question, this drug competes with the drug derivative conjugate for microparticle‑bound antibody. Antibody bound to sample drug is no longer available to promote particle aggregation, and subsequent particle lattice formation is inhibited. The presence of sample drug diminishes the increasing absorbance in proportion to the concentration of drug in the sample. Sample drug content is determined relative to the value obtained for a known cutoff concentration of drug.
N/A
Ask a specific question about this device
(269 days)
| Classification Name | Regulation Number | Product Code / Class |
|---|---|---|
| Glucose test system | 862.1345 | |
| Measured Analytes | GLU, TP, ALB | ALB, GLU |
| Product Code/ Regulation | GLU: CFR / 862.1345 | |
| TP: CEK / 862.1635ALB: CIX / 862.1035 | ALB: CIX / 862.1035GLU: CFR / 862.1345 | TP: CEK |
The Core Metabolic panel is part of the TruWellness Panel™ and is intended for use on the TruVerus™. The Core Metabolic panel (part of the TruWellness Panel™) is an in vitro diagnostic device and intended to be used for the quantitative determination of Glucose (GLU), Total Protein (TP), and Albumin (ALB) in lithium-heparinized venous whole blood in clinical laboratory or point-of-care settings.
The Core Metabolic panel (part of the TruWellness Panel™) is an in vitro diagnostic test system that aids the physician in the diagnosis and treatment of the following disorders in adults 18 years of age or older:
- Glucose (GLU): Carbohydrate metabolism disorders, including diabetes mellitus, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
- Total Protein (TP): Variety of diseases involving the liver, kidney, or bone marrow, as well as other metabolic and nutritional disorders.
- Albumin (ALB): Numerous diseases involving primarily the liver and kidneys.
The TruSystem is an automated, integrated in vitro diagnostic platform consisting of the TruVerus™ and the TruWellness Panel™, a Single-Use Consumable Kit that includes a Disc and a Support Pack. Designed for point-of-care and clinical laboratory use, the system enables the simultaneous measurement of clinical chemistry, immunoassay, and hematology parameters from a lithium-heparinized venous whole blood sample in a single run. The TruSystem delivers quantitative results for routine clinical chemistry and immunoassay analytes as well as a complete blood count (CBC) with a 3-part differential, all without the need for specialized operating skills, external calibration, or complex infrastructure.
The Single-Use Consumable Kit (TruWellness Panel™) houses all the components needed to process as well as analyze samples on the TruVerus™, including dried reagents, internal process control solutions, barcodes that manage the identity of the kit lot (e.g., Disc and Support Pack ID), calibration information, dilution buffers, and single-use plastic pipette tips. It also serves as a waste container which the user discards at the end of the run.
The Core Metabolic panel is part of the TruWellness Panel™ and is intended for use on the TruVerus™. The Core Metabolic panel (part of the TruWellness Panel™) is intended for in vitro quantitative determination of Glucose (GLU), Total Protein (TP), and Albumin (ALB) in lithium-heparinized venous whole blood samples.
N/A
Ask a specific question about this device
(87 days)
Longford,
Ireland
Re: K252357
Trade/Device Name: Glucose2
Regulation Number: 21 CFR 862.1345
Longford,
Ireland
Re: K252357
Trade/Device Name: Glucose2
Regulation Number: 21 CFR 862.1345
Classification:** Class II
Classification Name: Glucose Test System
Regulation Number: 21 CFR §862.1345
The Glucose2 assay is used for the quantitation of glucose in human serum, plasma, urine, or cerebrospinal fluid (CSF) on the ARCHITECT c System.
Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
The Glucose2 assay is an automated clinical chemistry assay.
Glucose is phosphorylated by hexokinase in the presence of adenosine triphosphate (ATP) and magnesium ions to produce glucose-6-phosphate (G-6-P) and adenosine diphosphate (ADP). Glucose-6-phosphate dehydrogenase (G-6-PDH) specifically oxidizes G-6-P to 6-phosphogluconate with the concurrent reduction of nicotinamide adenine dinucleotide phosphate (NADP) to its reduced form (NADPH). One micromole of NADPH is produced for each micromole of glucose consumed. The NADPH produced absorbs light at 340 nm and can be detected spectrophotometrically as an increased absorbance.
Methodology: Hexokinase/G6PDH
N/A
Ask a specific question about this device
(270 days)
System; On Call® Sure Sync GK Blood Glucose and Ketone Monitoring System
Regulation Number: 21 CFR 862.1345
The On Call® Sure GK Blood Glucose & Ketone Monitoring System is comprised of the On Call® Sure GK Blood Glucose & Ketone Meter, the On Call® Sure Blood Glucose Test Strips, and the On Call® Sure Blood Ketone Test Strips.
The On Call® Sure GK Blood Glucose & Ketone Monitoring System is intended to quantitatively measure blood glucose or blood ketone in fresh capillary whole blood from the fingertip. The system is intended for self-testing by people with diabetes at home as an aid in monitoring the effectiveness of diabetes control program. The system is for in vitro diagnostic use. It should only be used by a single patient and should not be shared. It is not intended for diagnosis or screening of diabetes or for neonatal use.
The On Call® Sure Sync GK Blood Glucose & Ketone Monitoring System is comprised of the On Call® Sure Sync GK Blood Glucose & Ketone Meter, the On Call® Sure Blood Glucose Test Strips, and the On Call® Sure Blood Ketone Test Strips.
The On Call® Sure Sync GK Blood Glucose & Ketone Monitoring System is intended to quantitatively measure blood glucose or blood ketone in fresh capillary whole blood from the fingertip. The system is intended for self-testing by people with diabetes at home as an aid in monitoring the effectiveness of diabetes control program. The system is for in vitro diagnostic use. It should only be used by a single patient and should not be shared. It is not intended for diagnosis or screening of diabetes or for neonatal use.
Not Found
N/A
Ask a specific question about this device
(126 days)
| OUG: 880.6310 Medicaldevice data systemNBW: 862.1345
The iCare App is intended for use in the home and clinical settings as and their healthcare professionals to view test results which are measured by iHealth devices to better manage user's health and get feedback from their professional care team.
The iCare App can also connect to medical devices and or non-medical devices and get data from devices during measurement or from the data stored in memory of the device for enhanced data managements. Data can be transmitted, displayed, and stored in the App.
The iCare APP is a mobile application on both Android and iOS platforms.iCare allows users to better manage their own health by enabling them to measure their vital signs, access their results and relevant health information with just their smart device and internet connection, and receive feedback from their professional care team.
iCare includes a patient darshboard featuring the Home, Health, Plus, Education, and Profile tabs. Accessory devices can be connected to the system to allow for collection of blood sugar, blood pressure, blood oxygen, and/or weight measurements. The patient darshboard functionality includes the ability to start measuring, allows users to view and track measurements, and export testing schedules for blood sugar, blood pressure, blood oxygen, and weight measurements; send messages to their professional care team; view previous appointment history information; view medication instructions; add entries to the food diary and review feedback from their registered dietician; set timers; and access articles and videos about health knowledge.
The provided text is a 510(k) Summary for the iCare App, focusing on its substantial equivalence to a predicate device. It primarily details regulatory information, device description, and non-clinical test summaries. It does not contain information about a study that proves the device meets specific performance acceptance criteria for a medical diagnostic or screening function.
The iCare App is classified as a "Medical Device Data System" (MDDS) that transmits, displays, and stores data from connected medical devices. Its function is to aid users and healthcare professionals in viewing test results for health management. It explicitly states: "Both devices make no interpretation, evaluation, medical judgments, or recommendations for treatment." This means the app itself doesn't perform diagnostic functions that would require specific performance metrics like sensitivity, specificity, or AUC against a ground truth.
Therefore, many of the requested criteria, such as acceptance criteria for diagnostic performance, a test set, expert ground truth establishment, MRMC studies, or standalone algorithm performance, are not applicable or not provided in this document because the device is a data management system, not a diagnostic algorithm.
Here's a breakdown of the applicable information based on the provided text:
1. A table of acceptance criteria and the reported device performance
The document does not present a table of quantitative performance acceptance criteria for diagnostic accuracy, sensitivity, or specificity, because the iCare App is an MDDS for data management, not a diagnostic tool. Instead, acceptance criteria are implied through the successful completion of non-clinical tests that demonstrate the basic functionality, safety, and effectiveness for its intended use as a data display and storage system.
| Test Category | Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|---|
| Software Verification & Validation | Compliance with FDA guidance for "moderate" level of concern software; no minor injury to patient/operator due to failure or latent flaw. | "Software verification and validation has been performed according to FDA guidance... The iCare App software was considered a 'moderate' level of concern...". All tests passed. |
| Wireless Coexistence Test | Ability to be used in intended environments without harmful interference. | "Wireless coexistence test has been performed to verify that the subject device can be used in intended environments." All tests passed. |
| Cybersecurity | Adherence to FDA guidance for cybersecurity; appropriate risk-based assessment and testing. | "Cybersecurity activities were conducted in accordance with FDA Guidance... The iCare App underwent appropriate risk-based cybersecurity assessment and testing..." All tests passed. |
| Usability Testing | Safe and effective use by lay users with provided labeling. | "Usability testing was conducted in accordance with FDA guidance... The test result demonstrates that the iCare App can be used by lay users with only provided labeling, the device is safe and effective for the intended use." All tests passed. |
2. Sample sized used for the test set and the data provenance
- Sample Size for Test Set: Not applicable for diagnostic performance as the device is not a diagnostic algorithm. The document mentions non-clinical testing (software, wireless, cybersecurity, usability) but does not specify "test set" sizes in the context of clinical data for diagnostic performance.
- Data Provenance: Not applicable in the context of clinical diagnostic data. The document focuses on the technical aspects of the software.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- Number of Experts: Not applicable, as the device does not perform diagnostic interpretations requiring expert-established ground truth for clinical cases.
- Qualifications of Experts: Not applicable.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Adjudication Method: Not applicable, as there is no clinical test set requiring ground truth adjudication for diagnostic performance.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- MRMC Study: No, an MRMC study was not done. The iCare App is an MDDS and does not involve AI assistance for human readers in a diagnostic capacity. It makes "no interpretation, evaluation, medical judgments, or recommendations for treatment."
- Effect Size: Not applicable.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Standalone Performance: No, a standalone performance study in the context of diagnostic accuracy was not done. The device's function is data transmission, display, and storage, not diagnostic algorithm performance.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- Type of Ground Truth: Not applicable for clinical diagnostic performance. For the software verification and validation, the "ground truth" would be the successful execution against specified requirements and accepted software engineering practices and FDA guidance.
8. The sample size for the training set
- Training Set Sample Size: Not applicable. This document does not describe a machine learning model that was trained on a dataset. The iCare App is a software application for data management, not an AI/ML algorithm requiring a training set of clinical data for diagnostic purposes.
9. How the ground truth for the training set was established
- Ground Truth Establishment for Training Set: Not applicable, as there is no mention of a training set for an AI/ML algorithm.
Ask a specific question about this device
(125 days)
Trade/Device Name: TeleRPM Gen2 Blood Glucose Monitoring System Regulation Number: 21 CFR 862.1345
2.3 Classification Name:
Classification Name: Glucose Test System Regulation Number: 21 CFR 862.1345
TeleRPM Gen2 Blood Glucose Monitoring System is comprised of the TeleRPM Gen2 Blood Glucose Meter and the TeleRPM Blood Glucose Test Strips. TeleRPM Gen2 Blood Glucose Monitoring System is intended to quantitatively measure the glucose concentration in fresh capillary whole blood samples drawn from the fingertips. It is intended for use by persons with diabetes at home as an aid to monitor the effectiveness of diabetes control. It is not intended for neonatal use or for the diagnosis of or screening for diabetes. This system is intended for self-testing outside the body (in vitro diagnostic use), and should only be used by a single person and should not be shared.
TeleRPM Gen2 Blood Glucose Monitoring System consists of TeleRPM Gen2 Blood Glucose Meter and the TeleRPM Blood Glucose Test Strips.
TeleRPM Control Solution, TeleRPM Lancing Device, TeleRPM Lancets are required for use but not included in meter box or test strips box and should be purchased separately. The TeleRPM Control Solution is for use with the above meter and test strip as a quality control check to verify that the meter and test strip are working together properly, and that the test is performing correctly. TeleRPM Lancing Device and TeleRPM Lancets are used for puncturing fingertip and then user can perform qlucose test with blood sample.
TeleRPM Gen2 Blood Glucose Monitoring System is designed to quantitatively measure the glucose concentration in fresh capillary whole blood from the fingertip. The glucose measurement is achieved by using the amperometric detection method. The test is based on measurement of electrical current caused by the reaction of the glucose with the reagents on the electrode of the test strip. The blood sample is pulled into the tip of the test strip through capillary action. Glucose in the sample reacts with glucose oxidase and the mediator. Electrons are generated, producing a current that is positive correlation to the glucose concentration in the sample. After the reaction time, the glucose concentration value is reported in plasma equivalents and is displayed on meter screen.
The provided text primarily focuses on the substantial equivalence determination for the TeleRPM Gen2 Blood Glucose Monitoring System to a predicate device. While it mentions the general types of studies conducted (robustness, precision, linearity, user evaluation, interference, stability, flex studies, software, cybersecurity controls, and a clinical usability study), it does not provide detailed acceptance criteria or numerical performance data as requested for several of your points.
Based on the information available:
1. A table of acceptance criteria and the reported device performance:
The document broadly states that the device "met the FDA SMBG OTC Guidance and industry standards" and that "these devices performed as intended and met associated guidance documents and industry standards." Specific numerical acceptance criteria and reported device performance for each study (precision, linearity, interference, etc.) are not detailed in the provided summary. For the user evaluation (clinical study), it states "the clinical performance met the FDA SMBG OTC Guidance."
| Acceptance Criteria (General) | Reported Device Performance (General) |
|---|---|
| Met FDA SMBG OTC Guidance and Industry Standards | Performed as intended, met FDA SMBG OTC Guidance and industry standards. |
| User evaluation criteria met | Inexperienced lay persons able to obtain blood glucose readings, understand labeling, use system, interpret results and error messages. No adverse effects or complications. |
2. Sample size used for the test set and the data provenance:
- User Evaluation (Clinical Study): The document mentions "All participants" were able to understand the labeling, use the system, and interpret results. However, the exact sample size for the clinical usability evaluation is not specified.
- Provenance: Not explicitly stated, but the company is located in Zhongshan, Guangdong, China. The testing location isn't specified, but it's reasonable to infer the studies were conducted by or on behalf of the manufacturer, likely in China or a region where they operate.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- This information is not provided in the summary. For a blood glucose monitoring system, the "ground truth" for glucose levels would typically be established by a laboratory reference method, not by experts adjudicating results.
- For the usability evaluation, the "ground truth" is about successful interaction with the device, which is assessed through user performance and observation, not expert consensus on a measurement.
4. Adjudication method for the test set:
- This information is not provided. For analytical performance, laboratory reference methods are used, not typically expert adjudication. For usability, the success of user interaction is observed and recorded.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- This is not applicable to this device. A "Blood Glucose Monitoring System" measures blood glucose; it is not an AI-assisted diagnostic imaging device that involves "human readers." Therefore, an MRMC study comparing human readers with and without AI assistance was not performed.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- The device performs a direct measurement of blood glucose. Its core function is a "standalone" algorithmic interpretation of the electrochemical reaction to display a glucose reading. This is its fundamental operation. There isn't a separate "human-in-the-loop" component in the direct glucose measurement process that would necessitate a distinction here.
7. The type of ground truth used:
- For analytical performance (precision, linearity, interference), the ground truth for blood glucose concentration would be established using a laboratory reference method (e.g., YSI analyzer). This is standard for blood glucose meter validation.
- For the usability evaluation, the "ground truth" assesses whether users can successfully operate the device and interpret results, which is based on direct observation and participant feedback.
8. The sample size for the training set:
- This device is a physical blood glucose meter and test strips relying on electrochemical principles, not a machine learning or AI model that requires a "training set" in the computational sense. Therefore, the concept of a training set sample size is not applicable here.
9. How the ground truth for the training set was established:
- As above, the concept of a training set is not applicable.
Ask a specific question about this device
(273 days)
| - 21 CFR 862.1145, Calcium test system |
| | - 21 CFR 862.1345
| | - 21 CFR 862.1145, Calcium test system |
| | - 21 CFR 862.1345
The ABL90 FLEX PLUS System is an in vitro diagnostic, portable, automated analyzer that quantitatively measures electrolytes (cK+, cNa+, cCa2+), glucose, and lactate in heparinized arterial and venous whole blood.
The ABL90 FLEX PLUS System is intended for use by trained technologists, nurses, physicians and therapists. It is intended for use in a laboratory environment, near patient, or point-of-care setting. These tests are only performed under a physician's order.
Potassium (cK+): Potassium measurements are used to monitor electrolyte balance in the diagnosis and treatment of disease conditions characterized by low or high blood potassium levels.
Sodium (cNa+): Sodium measurements are used in the diagnosis and treatment of aldosteronism, diabetes insipidus, adrenal hypertension, Addison's disease, delydration, inappropriate antidiuretic secretion, or other diseases involving electrolyte imbalance.
Calcium (cCa2+): Calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany.
Glucose (cGlu): Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Lactate (cLac): The lactate measure the concentration of lactate. Lactate measurements are used to evaluate the acid-base status and are used in the diagnosis and treatment of lactic acidity of the blood).
The ABL90 FLEX PLUS System consists of the ABL90 FLEX PLUS analyzer, sensor cassette and solution pack consumables, and related accessories for the analyzers. The ABL90 FLEX PLUS is a portable, automated system intended for in vitro testing of samples of balanced heparinized whole blood for electrolytes (cK+, cNa*, cCa²), glucose, and lactate. The ABL90 FLEX PLUS System has an automated sample inlet mechanism, which can collect blood through two different measuring modes: the S65 syringe mode and the SP65 short probe mode.
The provided text is a 510(k) Summary for the ABL90 FLEX PLUS System, an in vitro diagnostic device. This document focuses on demonstrating substantial equivalence to a legally marketed predicate device (ABL90 FLEX) rather than proving the device meets specific acceptance criteria as might be defined for a novel AI/ML device.
Therefore, much of the requested information regarding acceptance criteria for AI/ML performance, study design (test set, ground truth establishment, expert adjudication, MRMC studies, standalone performance, training set details) is not applicable to this type of device and its regulatory submission.
The document primarily proves the analytical performance of the new device is comparable to the predicate device through various analytical studies.
Here's a breakdown of the applicable information based on the provided text, and an explanation of why other requested information is not present:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly present "acceptance criteria" in a pass/fail table for each performance metric in the way it might for a novel AI/ML device. Instead, it presents analytical performance data (linearity, precision, detection, method comparison, interference) which is implicitly compared against pre-defined internal specifications or what is considered acceptable for the similar predicate device. The goal is to show the new device performs equivalently to the predicate.
Below is a summary of the reported device performance from the tables in the document. The "Acceptance Criteria" column cannot be fully populated as precise numerical thresholds are not explicitly stated as "acceptance criteria" in this 510(k) summary, but are rather implied by the successful demonstration of performance often within CLSI guidelines and comparable to the predicate.
| Parameter (Unit) | Test Category | Reported Performance (Range / Values) | Implicit Acceptance Criteria (based on predicate equivalence and CLSI) |
|---|---|---|---|
| cCa2+ (mg/dL) | Linearity | Slope: 0.883, Intercept: 0.445, R^2: 1.000 | R^2 near 1.0, slope near 1.0, intercept near 0, demonstrating linearity over the reportable range. |
| LoQ | 1.26 | Established lower limit of reliable quantitation. | |
| Precision (QC) | Repeatability SD: 0.003-0.014, CV%: 0.1-0.3 | Low SD and CV%, demonstrating consistent results. | |
| Precision (Blood) | Repeatability SD: 0.003-0.022, CV%: 0.06-0.45 | Low SD and CV%, demonstrating consistent results within biological samples. | |
| Method Comp. (Bias at MD) | S65: 0.001-0.003, SP65: 0.003-0.009 | Low bias compared to the predicate device, indicating equivalent measurements. | |
| cK+ (mEq/L) | Linearity | Slope: 1.001, Intercept: 0.027, R^2: 1.000 | R^2 near 1.0, slope near 1.0, intercept near 0, demonstrating linearity over the reportable range. |
| LoQ | 1.6 | Established lower limit of reliable quantitation. | |
| Precision (QC) | Repeatability SD: 0.00-0.01, CV%: 0.1-0.2 | Low SD and CV%, demonstrating consistent results. | |
| Precision (Blood) | Repeatability SD: 0.007-0.026, CV%: 0.14-0.96 | Low SD and CV%, demonstrating consistent results within biological samples. | |
| Method Comp. (Bias at MD) | S65: 0.002-0.004, SP65: 0.004-0.008 | Low bias compared to the predicate device, indicating equivalent measurements. | |
| cNa+ (mEq/L) | Linearity | Slope: 1.001, Intercept: -0.642, R^2: 1.000 | R^2 near 1.0, slope near 1.0, intercept near 0, demonstrating linearity over the reportable range. |
| LoQ | 99 | Established lower limit of reliable quantitation. | |
| Precision (QC) | Repeatability SD: 0.1-0.2, CV%: 0.1 | Low SD and CV%, demonstrating consistent results. | |
| Precision (Blood) | Repeatability SD: 0.061-0.194, CV%: 0.05-0.14 | Low SD and CV%, demonstrating consistent results within biological samples. | |
| Method Comp. (Bias at MD) | S65: 0.265-0.290, SP65: 0.221-0.259 | Low bias compared to the predicate device, indicating equivalent measurements. | |
| cGlu (mg/dL) | Linearity | Slope: 1.032, Intercept: -1.073, R^2: 1.000 | R^2 near 1.0, slope near 1.0, intercept near 0, demonstrating linearity over the reportable range. |
| LoD/LoQ | LoD: 5, LoQ: 5 | Established lower limits of detection and reliable quantitation. | |
| Precision (QC) | Repeatability SD: 0.3-1.3, CV%: 0.5-1.1 | Low SD and CV%, demonstrating consistent results. | |
| Precision (Blood) | Repeatability SD: 0.207-2.221, CV%: 0.35-0.85 | Low SD and CV%, demonstrating consistent results within biological samples. | |
| Method Comp. (Bias at MD) | S65: -0.460 to -2.028, SP65: -0.663 to -2.045 | Low bias compared to the predicate device, indicating equivalent measurements. | |
| cLac (mg/dL) | Linearity | Slope: 0.971, Intercept: -0.433, R^2: 1.000 | R^2 near 1.0, slope near 1.0, intercept near 0, demonstrating linearity over the reportable range. |
| LoD/LoQ | LoD: -0.3, LoQ: 2 | Established lower limits of detection and reliable quantitation. (Note: Negative LoD likely a calculation artifact near zero) | |
| Precision (QC) | Repeatability SD: 0.2-0.3, CV%: 0.3-1.1 | Low SD and CV%, demonstrating consistent results. | |
| Precision (Blood) | Repeatability SD: 0.177-0.379, CV%: 0.75-2.25 | Low SD and CV%, demonstrating consistent results within biological samples. | |
| Method Comp. (Bias at MD) | S65: -0.116 to 0.013, SP65: -0.156 to -0.169 | Low bias compared to the predicate device, indicating equivalent measurements. |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Test Set (for performance validation):
- Linearity: The specific number of samples tested for linearity is not explicitly stated as 'N' values in Table 1 but ranges presented (e.g., 1.896-11.146 for cCa2+) imply a sufficient number of points across the range were used.
- Detection (LoB, LoD, LoQ): Not explicitly stated as 'N' values in Table 2.
- Precision (using stable, aqueous ampoule-based QC material): Varies per parameter/level, but generally 243-244 replicates (N) per parameter/level.
- Precision (using blood): Varies per parameter/mode/interval, ranging from 2 to 202 replicates (N).
- Method Comparison:
- Arterial blood (S65 mode): 221-225 samples (N) across parameters.
- Arterial blood (SP65 mode): 214-218 samples (N) across parameters.
- Venous blood (S65 mode): 231-234 samples (N) across parameters.
- Venous blood (SP65 mode): 219-225 samples (N) across parameters.
- Combined (S65 mode): 436-441 samples (N) for combined arterial/venous.
- Combined (SP65 mode): 420-425 samples (N) for combined arterial/venous.
- Interference: "Large panel of likely interferents" for paired-difference study; dose-response studies for significant interferents. Specific sample sizes for each interferent are not detailed in the summary.
- Data Provenance: The document states that precision studies using QC material were conducted at "three external sites." Method comparison and precision studies using blood were conducted using both arterial and venous blood, and in both sample collection modes. The country of origin for the data (patients or samples) is not specified in this summary. The studies are described as "analytical performance testing," implying they are prospective or controlled laboratory studies rather than retrospective analysis of existing clinical data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not Applicable: This device is an in vitro diagnostic (IVD) analyzer that quantitatively measures analytes. Its performance is evaluated against reference measurement procedures or highly controlled materials, not by expert interpretation of images or clinical cases requiring expert consensus or qualifications. Ground truth is established by the reference method itself or the known concentration of QC materials.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not Applicable: As this is an IVD device measuring quantitative analytes, there is no expert adjudication process in this context, unlike an AI/ML device interpreting medical images. Performance is determined by comparison to reference methods or statistical analysis against known values.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not Applicable: This is an IVD analyzer, not an AI/ML device that assists human readers. Therefore, an MRMC study is not relevant to its regulatory approval process.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Partially Applicable (in a different sense): The ABL90 FLEX PLUS System is a standalone automated analyzer. Its performance is measured directly (algorithm only, if you consider the device's internal measurement algorithm) against reference methods or known concentrations, without a human-in-the-loop interpretation being the primary output that's being evaluated for accuracy. The results presented (linearity, precision, method comparison) are representative of its standalone performance.
7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)
- Quantitative Reference Methods / Known Concentrations:
- Linearity/Detection: Ground truth is established by preparing samples with known, precise concentrations across the measurement range, or by the inherent properties of the measurement system for LoB/LoD/LoQ.
- Precision: Ground truth is the expected value of the quality control (QC) materials or the prepared blood samples, or simply the reproducibility of measurements on the same sample.
- Method Comparison: Ground truth is the measurement from the legally marketed predicate device (ABL90 FLEX, specifically "ABL90 FLEX PLUS analyzer as it was designed at the time of the clearance of K160153") that the new device is being compared against. This device itself serves as the "reference method" for substantial equivalence.
- Interference: Ground truth is the expected measurement of known samples, with and without the interferent, using a reference method, to identify if the interferent causes a clinically significant deviation.
8. The sample size for the training set
- Not Applicable (in the AI/ML sense): This document describes the analytical validation of a traditional IVD device, not an AI/ML algorithm. There is no "training set" in the machine learning sense for this type of submission. The device is a physical instrument with established chemical/electrochemical measurement principles.
9. How the ground truth for the training set was established
- Not Applicable: As there is no "training set" in the AI/ML context, this question is not relevant. The device's internal parameters and calibration would be established through a manufacturing and calibration process, not through a "training" phase with a ground truth dataset in the way an AI model is trained.
Ask a specific question about this device
(255 days)
Name:Classification Name: | AIMBite block with gas sampling and oxygen delivery21 CFR §862.1345
K140473)Endoscopic bite block and nasal oxygen cannula withCO2 monitoring accessory21 CFR §862.1345
AIM is a bite block intended for use in patients 18 years and older who require supplemental oxygen and CO2 monitoring during procedures where the patient is expected to be minimally or moderately sedated. AIM is not indicated for use during procedures that are expected to require deep sedation.
AIM is a single-use, non-sterile bite block with integrated oxygen (O2) delivery and expired gas sampling tubing for patients undergoing procedures where supplemental oxygen and expired gas sampling is required expired. When paired with an oxygen supply and a capnography monitor, AIM can be left in place after the procedure to deliver oxygen and monitor CO2 levels.
AIM consists of a bite block, an attached oxygen delivery line and an attached CO2 sampling line. It delivers oxygen and samples exhaled CO2 in the oropharynx.
The provided text describes a 510(k) summary for a medical device named AIM, which is a bite block with integrated oxygen delivery and expired gas sampling tubing. The summary compares AIM to a predicate device, DualGuard™ (K140473), to demonstrate substantial equivalence.
Here's an analysis of the acceptance criteria and study proving the device meets these criteria, based on the provided document:
Acceptance Criteria and Device Performance Study for AIM
1. Table of Acceptance Criteria and the Reported Device Performance
The document describes performance tests by comparing the AIM device to its predicate, DualGuard™. The acceptance criteria appear to be equivalent or better performance than the predicate device.
| Acceptance Criteria (Internal/Predetermined) | Reported Device Performance |
|---|---|
| Biocompatibility: Meet ISO 10993 standards (ISO 10993-5:2009, ISO 10993-23:2021, ISO 10993-10:2021, ISO 18562-2:2017, ISO 18562-3:2017) for surface contact, skin and mucosal, externally communicating tissue, limited use (<24hr). | Met the specified ISO standards for biocompatibility. |
| Performance Tests | |
| * % Fraction of Inspired O2 (FiO2) | AIM maintained FiO2 equivalent to the predicate device in all simulated conditions (at several respiratory rates, tidal volumes, and oxygen flow rates). |
| * % End Tidal CO2 (EtCO2) | AIM indicated EtCO2 is equivalent to the predicate device in all simulated conditions (under two simulated conditions at multiple simulated EtCO2 values). |
| * CO2 Waveforms | AIM captured CO2 waveforms as well as or better than the predicated device in all simulated conditions (under several simulated respiratory conditions, oxygen flow rates, and simulated EtCO2 values). |
| Mechanical Tests: | |
| * Bite Force Finite Element Analysis (FEA) | Minimal deformation observed. |
| * Dislodgement Force | Comparison to predicate (implies equivalence or better). |
| * Axial Separation Force (ISO 80369-2) | Comparison to predicate (implies equivalence or better) and compliance with ISO 80369-2. |
| Accelerated Aging: Comparison of pre and post-aging performance (ISO 80369-2, ISO 594-2, ISO 11607-1). | All samples passed the performance tests at least as well as the predicate device after accelerated aging. Compliance with ISO 80369-2, ISO 594-2, ISO 11607-1. |
| Ship Testing: ISTA 3A | Passed ISTA 3A. |
Note: The primary acceptance criterion for many of the performance and mechanical tests is "Comparison to predicate," implying that the AIM device must perform equivalently or better than the legally marketed predicate device.
2. Sample Size Used for the Test Set and the Data Provenance
- Sample Size: The document does not specify the exact sample sizes (number of devices tested) for the performance, mechanical, or aging tests. It mentions "All AIM samples passed the non-clinical tests, and all samples passed the performance tests..." which suggests that all units tested within the sample passed, but the size of "all samples" is not quantified.
- Data Provenance: The study appears to be retrospective in the sense that it's a submission for regulatory clearance based on historical testing. The data originates from simulated conditions (e.g., "simulated oxygen delivery," "simulated EtCO2 values," "simulated respiratory conditions"). The geographic origin of the data/testing is not explicitly stated, but as it's an FDA submission, the testing would typically be conducted under recognized standards, often implicitly in the US or by labs adhering to US-acceptable standards.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts
Not applicable. This device is not an AI/ML device that generates diagnostic outputs requiring expert interpretation for ground truth. The acceptance criteria relate to physical performance (FiO2, EtCO2, mechanical properties) measured by instrumentation under simulated conditions, compared against a predicate device. There is no mention of human expert adjudication for ground truth.
4. Adjudication Method (e.g., 2+1, 3+1, none) for the Test Set
Not applicable. As noted above, the ground truth is based on physical measurements and comparisons to a predicate device, not human interpretation or consensus.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done
No, an MRMC study was not done. This type of study is typically performed for AI-assisted diagnostic devices where human readers interpret medical images or data. The AIM device is a physical medical device for oxygen delivery and CO2 monitoring, not an AI diagnostic tool.
6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done
Yes, in a sense, the performance tests for FiO2, EtCO2, and CO2 waveforms were evaluative of the device's standalone performance under simulated conditions, without direct human intervention as part of the measurement process itself. The "algorithm" here refers to the physical design and function of the device rather than a computational algorithm.
7. The Type of Ground Truth Used
The ground truth for the performance tests (FiO2, EtCO2, CO2 Waveforms) was established by:
- Instrumental measurement under simulated conditions: The values of FiO2 and EtCO2, and the characteristics of CO2 waveforms, were presumably measured by calibrated instruments under controlled, simulated physiological conditions.
- Comparative equivalence to a legally marketed predicate device: The performance of the AIM device was directly compared to the performance of the DualGuard™ predicate device under the same simulated conditions. The predicate device's performance effectively served as a reference or "ground truth" for acceptable performance.
8. The Sample Size for the Training Set
Not applicable. The AIM device is a physical medical device, not an AI/ML model that requires a training set of data.
9. How the Ground Truth for the Training Set was Established
Not applicable, as there is no training set for this type of device.
Ask a specific question about this device
(925 days)
/Device Name: XPER Technology PREMIUM Pro Blood Glucose Monitoring System Regulation Number: 21 CFR 862.1345
|
| Regulation Number | 21 CFR §862.1345
The XPER Technology PREMIUM Pro Blood Glucose Monitoring System is intended for point-of-care, in vitro diagnostic, multiple-patient use for the quantitative determination of glucose in fresh capillary whole blood samples from the fingertips in endocrinology clinic laboratories and physician office laboratories.
The system should only be used with single-use, auto-disabling lancing devices when performing a fresh capillary whole blood sample from the fingertip.
The system is not intended for the screening or diagnosis of diabetes mellitus but is indicated for use in determining dysglycemia.
The system is not intended for use on patients receiving intensive medical intervention/therapy.
The system is not intended for use in acute care, nursing facilities, skilled nursing facilities or hospital settings. The system is not intended for use on neonates.
The XPER Technology PREMIUM Pro Blood Glucose Monitoring System consists of the XPER Technology PREMIUM Pro Blood Glucose Meter, the XPER Technology PREMIUM Pro Blood Glucose Test Strips, and the TaiDoc Blood Glucose Control Solutions. This system is a multiplepatient use for the quantitative determination of glucose in capillary whole blood samples from the fingertips in endocrinology clinic laboratories and physician office laboratories as an aid in monitoring the effectiveness of glucose control. The TaiDoc Blood Glucose Control Solutions are used to check that the meter and test strips are working together properly.
Here's a breakdown of the acceptance criteria and study details for the XPER Technology PREMIUM Pro Blood Glucose Monitoring System, based on the provided FDA 510(k) summary:
Acceptance Criteria and Device Performance
The acceptance criteria are implicitly defined by the statement "all test results were within acceptance criteria" in various sections. The reported performance is the achievement of these criteria. The specific numerical targets for each criterion are not always explicitly stated (e.g., for precision, linearity, hematocrit, interference), but the document confirms that the device met these criteria.
Table of Acceptance Criteria and Reported Device Performance
| Test Type | Acceptance Criteria (Implicitly Met) | Reported Device Performance |
|---|---|---|
| Precision | All test results were within acceptance criteria for within-run and intermediate precision across the glucose measuring range (10-800 mg/dL). | All precision data met the acceptance criteria. |
| Linearity | All test results were within acceptance criteria for linearity across the claimed measuring range 10-800 mg/dL. | All linearity data met the acceptance criteria. Meters accurately displayed "Lo" for glucose < 10 mg/dL and "Hi" for glucose > 800 mg/dL. |
| Hematocrit | All test results were within acceptance criteria, demonstrating that the claimed hematocrit range 10% - 70% doesn't affect performance. | All hematocrit data met the acceptance criteria. |
| Interference | Interference data showed the highest concentration with no interference for tested substances, with "maximum test concentration of each interfering substances tested" being within acceptance. (Specific bias limits not provided, but the statement indicates meeting the criteria.) | Demonstrated that most substances at expected concentrations do not interfere. Noted exceptions: Xylose can give falsely elevated results, and Pralidoxime Iodide >5 mg/dL may affect results. |
| Disinfection | Reduction of duck hepatitis B virus within acceptance criteria, and complete inactivation achieved using Clorox Healthcare™ Bleach Germicidal Wipe (EPA No.: 67619-12). | The specified wipe effectively eliminated duck hepatitis B virus from the device surface. |
| Robustness | All test results were within acceptance criteria over 27,500 cleaning and disinfection cycles. Indicates the device must maintain intended use performance. | Device maintained intended use performance after 27,500 cycles of cleaning/disinfection. |
| Flex Studies | All test results were within acceptance criteria, demonstrating risks of imprecisions are effectively mitigated under normal use for various stress conditions. | All flex study results met acceptance criteria, mitigating imprecision risks under stress conditions. |
| Stability | Protocols and acceptance criteria acceptable to support labeling claims: open vial stable after first opening; closed vials stable for 12 months at 2-30°C and 10-90% RH. | Test strips meet labeling claims for open and closed vial stability. |
| Clinical Accuracy (Capillary Blood) | For glucose < 75 mg/dL: - Within ±5 mg/dL: High percentage of samples expected. - Within ±10 mg/dL: Very high percentage of samples expected. - Within ±12 mg/dL: Extremely high percentage of samples expected. - Within ±15 mg/dL: All samples expected. For glucose ≥ 75 mg/dL: - Within ±5 %: High percentage of samples expected. - Within ±10 %: Very high percentage of samples expected. - Within ±12 %: Extremely high percentage of samples expected. - Within ±15 %: Nearly all samples expected. - Within ±20 %: All samples expected. | For glucose < 75 mg/dL (n=13): - Within ±5 mg/dL: 61.5% (8/13) - Within ±10 mg/dL: 92.3% (12/13) - Within ±12 mg/dL: 100% (13/13) - Within ±15 mg/dL: 100% (13/13) - Exceeds ±15 mg/dL: 0% (0/13) For glucose ≥ 75 mg/dL (n=401): - Within ±5 %: 55.9% (224/401) - Within ±10 %: 88.8% (356/401) - Within ±12 %: 97.5% (391/401) - Within ±15 %: 99.5% (399/401) - Within ±20 %: 100% (401/401) - Exceeds ±20 %: 0% (0/401) |
| Accuracy at Extreme Blood Glucose Values | Performance consistent with YSI-2300 for samples < 80 mg/dL and > 300 mg/dL. (Implied specific criteria for agreement at these extremes). | 50 samples with glucose <80 mg/dL and 50 samples with glucose >300 mg/dL were tested against YSI-2300, and results indicate acceptable accuracy. Numerical details for "accuracy" at extremes are not explicitly provided, only that it "was performed" and presumably met criteria. |
| Usability | Usability results indicate the device is easy to use and the labeling is easy to understand. | Operators confirmed ease of use and understandability of the device and labeling through questionnaires. |
Study Details
This document describes a medical device, a Blood Glucose Monitoring System, which does not utilize AI or involve human readers for image interpretation. Therefore, questions related to AI models, human reader improvement with AI assistance, expert adjudication for ground truth related to image analysis, or MRMC studies are not applicable to this device.
Here's the relevant information based on the provided text:
-
Sample sizes used for the test set and data provenance:
- Clinical Accuracy (Capillary Blood): 414 patients.
- Provenance: Clinical study conducted at 3 U.S. sites and 6 Taiwan sites. Data is prospective as it was collected during an active clinical study with patients.
- Accuracy at Extremes: 100 samples (50 for <80 mg/dL, 50 for >300 mg/dL).
- Provenance: "Blood samples were collected and allowed to glycolyze or were spiked with high concentration glucose solution". Implies a laboratory-controlled, prospective or specially prepared sample set.
- Clinical Accuracy (Capillary Blood): 414 patients.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not Applicable in the traditional sense of 'experts' interpreting data. For blood glucose monitoring systems, the "ground truth" is established by a highly accurate reference method.
- The ground truth for the clinical and extreme glucose value studies was the YSI Model 2300 Glucose Analyzer. This is a laboratory-based, well-established, and highly accurate reference method for glucose measurement, not human experts.
-
Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- Not Applicable. As ground truth is established by an automated reference analyzer (YSI-2300), there is no human interpretation or adjudication process involved in setting the ground truth for glucose values.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not Applicable. This is a blood glucose monitoring system, not an AI-assisted diagnostic imaging device. There are no human readers or AI assistance in the interpretation of results from this device in the same way as an imaging study.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Yes, in essence. Blood glucose meters are inherently "standalone" in their function of measuring glucose. The clinical study directly compares the device's numerical output (algorithm's result) to the YSI-2300 reference method without human interpretation of the device's output influencing the direct comparison. The "human-in-the-loop" for this device is the user taking and reading the measurement, but the performance evaluation is on the accuracy of the numerical reading itself, independent of user interpretation for the primary outcome.
-
The type of ground truth used:
- Reference Method: The YSI Model 2300 Glucose Analyzer, a laboratory-based, highly accurate method for quantitative glucose determination. Comparisons are quantitative.
-
The sample size for the training set:
- Not directly applicable/not explicitly stated in terms of an "AI training set." This device is a traditional electrochemical biosensor, not an AI/machine learning model that undergoes a distinct "training" phase with a large dataset in the way a deep learning algorithm would. The development and calibration of such a device involve extensive laboratory testing and optimization, which could be considered an analogous "training" or development process for its internal algorithms, but it's not described as a separate, quantifiable "training set" with ground truth in the context of AI regulatory submissions.
-
How the ground truth for the training set was established:
- Not applicable in the AI context. For this type of device, ground truth for development/calibration (analogous to "training") would be established through a combination of:
- Highly controlled laboratory experiments using reference solutions of known glucose concentrations.
- Comparisons to established reference methods (like YSI-2300) with well-characterized samples.
- Controlled studies to characterize and mitigate interferences (e.g., hematocrit, medications).
- Not applicable in the AI context. For this type of device, ground truth for development/calibration (analogous to "training") would be established through a combination of: