510(k) Data Aggregation

    K Number
    K252619

    Validate with FDA (Live)

    Device Name
    QSCHECK UISACR
    Manufacturer
    Date Cleared
    2026-02-20

    (185 days)

    Product Code
    Regulation Number
    862.1225
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Republic of Korea

    Re: K252619
    Trade/Device Name: QSCHECK UISACR
    Regulation Number: 21 CFR 862.1225
    Republic of Korea

    Re: K252619
    Trade/Device Name: QSCHECK UISACR
    Regulation Number: 21 CFR 862.1225
    Automated urinalysis system |
    | Regulation Number | 862.1225862.1645862.2900 |
    | Product

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    QSCHECK UISACR is intended for the in vitro semi-quantitative measurement of the following parameters: microalbumin and creatinine in urine specimens and for Albumin/Creatinine ratio (ACR) calculated using those values. The results may be used in conjunction with clinical evaluation as an aid in the assessment for kidney function. The system is intended for prescription use only, in clinical laboratory settings.

    Device Description

    QSCHECK UISACR consists of urine test strips and a smartphone application. In the test strips, the color of the reagent pad that contains enzymes and chemicals is changed according to the concentration of each biomarker in the urine. After the reaction on the strips, the degree of color change is photographed and analyzed by a smartphone application, QSCHECK-app, that runs on both Apple iPhone 14 (iOS v26) and Samsung Galaxy S23 (Android v16).

    The software of the QSCHECK-app. recognizes the location of the reagent pad through the QR code on the urine test strip and checks the degree of color development using RGB color codes. The RGB color codes are converted to Hue (H), Saturation (S), Value (V) color model, which is a method of expressing color or arranging colors according to this method. The results are displayed by matching colorimetric tables for each section that were previously classified according to HSV.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K251281

    Validate with FDA (Live)

    Date Cleared
    2026-01-21

    (272 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    18 - 120
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    K251281**
    Trade/Device Name: Nova Max Creat eGFR Monitoring System
    Regulation Number: 21 CFR 862.1225
    ----------|------------------|---------------|-------------------|
    | Creatinine test system | SHB | 862.1225

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Nova Max Creat eGFR Monitoring System is comprised of the Nova Max Creat eGFR Monitor and the Nova Max Creat eGFR Test Strips.

    The Nova Max Creat eGFR Monitoring System is intended for in vitro diagnostic use for the quantitative measurement of creatinine and estimation of glomerular filtration rate (eGFR) in fresh capillary whole blood obtained from the fingertip of adult patients aged 18 and above. The system is intended for single patient home use by prescription only and should not be shared. It is intended for use by patients as an aid to monitor kidney function on the order of a treating healthcare professional.

    Device Description

    The Nova Max Creat eGFR Monitoring System consists of Nova Max Create eGFR Monitor and Nova Max Create eGFR Creatinine-Test Strips. The Nova Max Creat eGFR test strips, Nova Max Creat eGFR control solution (Level 1 and Level 2), and a single-use, disposable, 21 gauge safety lancet can be purchased separately.

    Nova Max Creat eGFR Monitor
    The monitor is intended to be used in conjunction with the Nova Max Creat eGFR Creatinine-Test Strips to measure creatinine and calculate estimated glomerular filtration rate (eGFR) using capillary whole blood obtained from the fingertip. The hand-held monitor is lightweight, portable, utilizes a 2.8" color touchscreen display and has capability for storage (400 test samples and/or quality control samples) and review of data. The monitor is powered by a rechargeable, 3.7 V Li polymer battery.

    Nova Max Creat eGFR Creatinine-Test Strips
    The Nova Max Creat eGFR Creatinine-Test Strip is designed with an electrode that measures creatinine levels. Creatinine in the capillary whole blood sample mixes with the reagents in the test strip to produce an electric current. The amount of current that is produced depends on how much creatinine is in the sample. A digital readout is displayed on the monitor in 30 seconds, the eGFR is calculated using patient information and the measured creatinine level. The test strips are for single use only.

    Nova Max Creat eGFR Creatinine Control Solutions
    The Nova Max Creat eGFR Creatinine Control Solution is used as a quality control check to ensure that the Nova Max Creat eGFR Monitor and the Nova Max Creat eGFR Creatinine-Test Strips are working properly as a system. There are two levels of Nova Max Creat eGFR Creatinine Control Solution. Each level of control solution has a known creatinine concentration that reacts with the reagents in the test strip.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K252206

    Validate with FDA (Live)

    Date Cleared
    2026-01-09

    (179 days)

    Product Code
    Regulation Number
    862.1225
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Trade/Device Name:** Nova Allegro UACR Assay, Nova Allegro Analyzer
    Regulation Number: 21 CFR 862.1225
    Product Code |
    |---------------------|--------------|-------|--------------|
    | Creatinine Test system | 862.1225

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Nova Allegro urine albumin creatinine ratio (UACR) assay is intended for the quantitative determination of albumin, creatinine, and the albumin/creatinine ratio (UACR) in human urine. The measurement of urine albumin, creatinine, and albumin/creatinine ratio aids in the early diagnosis of nephropathy.

    The Nova Allegro Analyzer is intended for in vitro diagnostic use in clinical laboratory and near-patient testing (point-of-care) settings for the quantitative determination of Nova Allegro Assays using Nova Allegro Test Cartridges.

    Device Description

    Nova Allegro UACR Assay

    The Nova Allegro UACR Assay is a completely automated assay that functions with the Nova Allegro Analyzer for the determination of albumin and creatinine and the calculation of UACR in human urine. The Nova Allegro UACR Assays are performed sequentially using a single Allegro UACR test cartridge. The Allegro UACR Test Cartridge contains all of the reagents for measuring albumin and creatinine. Allegro UACR Controls are used to monitor the performance of the Nova Allegro UACR Test Cartridge and Allegro analyzer.

    Nova Allegro Analyzer

    The Nova Allegro Analyzer is a compact, point-of-care analyzer that features a clinically important menu of measured and calculated tests. All tests are measured with disposable, ready-to-use cartridges. The analyzer supports multiple wavelengths that are used to measure the assay of interest. The analyzer consists of the following key systems/components that the user interacts with:

    • Two analytical bays where the single use test cartridges are analyzed
    • Color Touchscreen Display
    • Barcode Scanner
    • Printer
    • Data Export Options
    • Ethernet Connection
    • USB Port
    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K251058

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-11-26

    (236 days)

    Product Code
    Regulation Number
    862.1225
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    K251058**
    Trade/Device Name: Tru Kidney Health Test Panel; Tru Analyzer
    Regulation Number: 21 CFR 862.1225

    Classification NameRegulation NumberProduct Code / Class
    Creatinine test system862.1225
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Tru Kidney Health Test Panel is part of the TruWellness Panel™ and is intended for use on the Tru Analyzer. The Tru Kidney Health Test Panel (part of the TruWellness Panel™) is an in vitro diagnostic device and intended to be used for the quantitative determination of Creatinine (CRE) and Blood Urea Nitrogen (BUN) in lithium-heparinized venous whole blood in clinical laboratory or point-of-care settings. From the CRE determination, Estimated Glomerular Filtration Rate (eGFR) is calculated by the analyzer.

    The Tru Kidney Health Test Panel (part of the TruWellness Panel™) is an in vitro diagnostic test system that aids the physician in diagnosing the following disorders in adults 18 years of age or older:

    • Creatinine (CRE): Renal disease and monitoring of renal dialysis.
    • Blood Urea Nitrogen (BUN): Certain renal and metabolic diseases.

    The Tru Analyzer is an automated, multi-assay integrated system, containing an absorbance reader, a bead scanner, and a cell imager for in vitro diagnostic use. This analyzer is intended for the quantitative determination of various analyte concentrations found in lithium-heparinized venous whole blood. It is for clinical laboratory or point-of-care use.

    Use only TruWellness Panel™ test kits with the Tru Analyzer.

    Device Description

    The TruSystem is an automated, integrated in vitro diagnostic platform consisting of the Tru Analyzer and the Tru Kidney Health Test Panel (part of the TruWellness Panel™), a Single-Use Consumable Kit that includes a Disc and a Support Pack. Designed for point-of-care and clinical laboratory use, the system enables the simultaneous measurement of clinical chemistry, immunoassay, and hematology parameters from a lithium-heparinized venous whole blood sample in a single run. The TruSystem delivers quantitative results for routine clinical chemistry and immunoassay analytes as well as a complete blood count (CBC) with a 3-part differential, all without the need for specialized operating skills, external calibration, or complex infrastructure.

    The Tru Analyzer is a benchtop instrument that fully automates sample processing, assay execution, and result reporting. Its touchscreen interface allows operators to control workflows, initiate tests, and review results with minimal training. The Single-Use Consumable Kit and blood sample are loaded into the analyzer drawer, where all necessary processing takes place. Internally, the Tru Analyzer integrates:

    • A high precision pipettor for automated sample and reagent handling,
    • An onboard centrifuge to separate whole blood into plasma,
    • A closed-loop thermal control system to maintain precise assay temperatures,
    • A motion control system to fully automate sample processing,
    • Dedicated detection modules for clinical chemistry, immunoassay, and hematology testing,
    • A high-definition camera used to collect assay readings, image capture, and instrument quality control checks,
    • An electronics board to manage individual module boards and associated firmware,
    • An integrated computer running the instrument software, accessible via the touchscreen user interface.

    The Tru Analyzer is factory-calibrated and continuously monitors its performance using optical sensing and electronic feedback mechanisms. Every time the analyzer is powered on or a Single-Use Consumable Kit (TruWellness Panel™) is loaded, the system runs an automated self-test to verify that it remains within calibration and is functioning properly. Internal quality control (QC) checks occur throughout the testing process to ensure the integrity of the analyzer, Disc, and Support Pack.

    If an internal QC check does not meet specifications, the instrument will automatically suppress result reporting to prevent the release of potentially inaccurate results. Any impacted tests will display associated codes in the footnotes of the patient report. The Tru Analyzer also applies result flags to highlight conditions that may indicate abnormalities in the patient sample or require attention. These flags appear in-line with impacted assay results for clear visibility. If a system operation failure occurs, the Tru Analyzer will halt the run, and no results will be generated. In these cases, the instrument displays an on-screen error message with a unique code and step-by-step instructions to guide resolution. These measures ensure the system provides accurate results, while requiring no special user knowledge, training, or on-site calibration.

    The Single-Use Consumable Kit (TruWellness Panel™) houses all the components needed to process and analyze samples on the Tru Analyzer, including dried reagents, internal process control solutions, barcodes that manage the identity of the kit lot (e.g., Disc and Support Pack ID), calibration information, dilution buffers, and single-use plastic pipette tips. It also serves as a waste container which the user discards at the end of the run.

    The Tru Kidney Health Test Panel is part of the TruWellness Panel™ and is intended for use on the Tru Analyzer. The Tru Kidney Health Test Panel (part of the TruWellness Panel™) provides in vitro quantitative determination of Creatinine (CRE), Blood Urea Nitrogen (BUN), and calculated estimated Glomerular Filtration Rate (eGFR) in lithium-heparinized venous whole blood samples.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K242585

    Validate with FDA (Live)

    Device Name
    Cystatin C
    Date Cleared
    2025-05-16

    (259 days)

    Product Code
    Regulation Number
    862.1225
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Milan (MI), 20152, Italy

    Re: K242585
    Trade/Device Name: Cystatin C
    Regulation Number: 21 CFR 862.1225
    Device Classification: Class II
    Regulation Description: Creatinine test system
    Regulation Number: 862.1225

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Cystatin C assay is an in vitro diagnostic test used in the quantitative immunoturbidimetric determination of cystatin C in human serum and plasma on the Alinity c system.

    Measurement of cystatin C aids in the diagnosis and treatment of renal diseases.

    For laboratory professional use only.

    Device Description

    The Cystatin C assay is an automated clinical chemistry assay.

    Cystatin C is a particle-enhanced turbidimetric immunoassay (PETIA) developed to accurately and reproducibly measure cystatin C levels in serum and plasma. Latex particles coated with anti-human cystatin C antibody agglutinate when mixed with sample containing human cystatin C. The change in absorbance due to agglutination of the reaction mixture is proportional to the quantity of human cystatin C in the sample.

    AI/ML Overview

    The provided text is a 510(k) clearance letter for an in vitro diagnostic test (Cystatin C assay), not an AI/ML-driven medical device. Therefore, the information requested for AI/ML device studies (such as MRMC studies, expert adjudication, training set ground truth, etc.) is not applicable to this document.

    However, I can extract and present the performance acceptance criteria and reported device performance from the provided document, even though they relate to a traditional laboratory assay rather than an AI/ML model.

    Here's an interpretation based on the provided text for a traditional in vitro diagnostic device:


    Based on the provided FDA 510(k) clearance letter for the Cystatin C Test System, the device is a quantitative immunoturbidimetric assay for the determination of cystatin C in human serum and plasma. The performance of this device was demonstrated through various nonclinical studies.

    1. Table of Acceptance Criteria and Reported Device Performance

    For in vitro diagnostic tests like the Cystatin C assay, "acceptance criteria" are typically defined by ranges or thresholds for performance characteristics such as precision, accuracy (bias), linearity, and limits of detection. The document describes the reported performance measured during the studies, and implicitly, these figures met the internal acceptance criteria set by the manufacturer (and approved by the FDA for clearance).

    Here's a summary of the reported device performance:

    Performance CharacteristicAcceptance Criteria (Implicit from Industry Standards/Predicate)Reported Device Performance (Cystatin C Assay)
    Analytical Measuring Interval (AMI)Defined range of accurate and precise measurement0.30 – 10.00 mg/L
    Extended Measuring Interval (EMI)Range accessible via dilution/spiking10.00 – 40.00 mg/L
    Reportable IntervalOverall range of reportable results0.30 – 40.00 mg/L
    Within-Laboratory Precision (Repeatability)Expected low variabilityControl Level 1: 0.81 mg/L Mean, 1.0%CV Control Level 2: 4.11 mg/L Mean, 0.6%CV Panel A: 0.49 mg/L Mean, 1.7%CV Panel B: 0.92 mg/L Mean, 0.8%CV Panel C: 5.89 mg/L Mean, 0.5%CV Panel D: 8.95 mg/L Mean, 0.8%CV
    Within-Laboratory Precision (Total)Expected low variability over timeControl Level 1: 1.7%CV Control Level 2: 1.0%CV Panel A: 1.8%CV Panel B: 0.9%CV Panel C: 0.6%CV Panel D: 1.0%CV
    Overall ReproducibilityExpected low variability across sites/lotsControl Level 1: 2.0%CV Control Level 2: 1.1%CV Panel 1: 5.4%CV Panel 2: 1.6%CV Panel 3: 1.6%CV
    Accuracy (Bias vs. Reference Material)Acceptable low bias to a certified referenceRanged from 1.3% to 1.8% across all reagent and calibrator lots relative to ERM-DA471/IFCC.
    LinearityDemonstrated proportional response across AMILinear across 0.30 to 10.00 mg/L (AMI).
    Limit of Blank (LoB)Lowest detectable signal for a blank sample0.03 mg/L
    Limit of Detection (LoD)Lowest concentration detectable with 95% probability0.05 mg/L
    Limit of Quantitation (LoQ)Lowest concentration meeting 25% total allowable error0.30 mg/L
    Hook Effect (Prozone Effect)No false low results at high concentrationsNo prozone effect observed up to 40.00 mg/L.
    Interference (Endogenous Substances)No significant interference at specified levelsNo significant interference observed for Bilirubin (60 mg/dL), Hemoglobin (1000 mg/dL), Total protein (10.2-11.8 g/dL), Triglycerides (1500 mg/dL). Some interference noted for high Rheumatoid factor and very high Total protein.
    Interference (Exogenous Substances)No significant interference at specified levelsNo significant interference observed for 16 common drugs (e.g., Acetaminophen, Ibuprofen, Ascorbic acid, Cyclosporin, etc.) at specified therapeutic/toxic levels.
    Correlation with Comparator AssayStrong correlation with predicate/comparable methodCorrelation Coefficient: 1.00 (Serum, n=161) vs. Comparator Cystatin C assay.
    Matrix ComparisonAcceptable for various specimen typesAcceptable for Serum, Serum separator, Dipotassium EDTA, Lithium heparin, Lithium heparin separator, Sodium heparin, Tripotassium EDTA plasma.
    Reference Interval (Expected Values)Established range for healthy individualsAdult: 0.59–1.28 mg/L (2.5th to 97.5th percentile)

    Study Details (Relevant to IVD, not AI/ML)

    The following points are addressed as much as possible for an in vitro diagnostic device, noting where the requested AI/ML specific information is not applicable.

    1. Sample sizes used for the test set and the data provenance:

      • Precision/Reproducibility:
        • Within-Laboratory Precision: 80 replicates per sample/control (2 replicates/day for 20 days) for 2 controls and 4 human serum panels.
        • Reproducibility: 240 replicates per sample/control (4 replicates, twice/day for 5 days at 3 sites) for 2 controls and 3 human serum panels.
      • Accuracy (Bias): Not specified as a separate sample size, but involved testing across 2 reagent lots and 2 calibrator lots.
      • Linearity: Not specified as a number of distinct samples, but assessed over a range of concentrations.
      • Lower Limits of Measurement (LoB, LoD, LoQ): ≥ 60 replicates of zero-analyte samples for LoB, and ≥ 60 replicates of low-analyte level samples for LoD/LoQ.
      • Analytical Specificity (Interference): Each substance tested at 2 analyte levels.
      • Method Comparison: 161 serum samples.
      • Matrix Comparison: Not specified as a specific number, but evaluated across various tube types.
      • High Dose Hook: Not specified as a specific number of samples, but tested up to 40.00 mg/L.
      • Expected Values (Reference Interval): 250 apparently healthy individuals (105 females, 145 males) with eGFR > 80, aged 18 to 69 years.

      Data Provenance: The document does not explicitly state the country of origin for the clinical samples. The studies are described as "nonclinical performance" studies, meaning they evaluate the analytical performance of the assay itself, rather than diagnostic accuracy in a clinical setting with patient outcomes. These are typically prospective studies conducted in a controlled lab environment.

    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not applicable for this in vitro diagnostic device. Ground truth for analytical performance studies of IVDs is established by the known concentration of analytes in reference materials, calibrators, and characterized control samples, or by comparison to a well-established reference or predicate method. It does not involve expert readers or their consensus.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not applicable for this in vitro diagnostic device. Adjudication methods are relevant for subjective image-based assessments or clinical diagnoses where human interpretation varies. This product is a quantitative chemical assay.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not applicable for this in vitro diagnostic device. MRMC studies are for evaluating the impact of AI on human reader performance, typically in radiology or similar interpretive fields. This device performs automated measurements.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Analogous concept applies. The performance data presented (precision, accuracy, linearity, limits of detection, etc.) are the standalone performance of the assay on the Alinity c system. It demonstrates the device's ability to measure Cystatin C without human interpretation of the analytical signal.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • For the analytical performance studies, the ground truth was established by:
        • Certified reference materials: Specifically, ERM-DA471/IFCC for assay standardization and accuracy validation.
        • Preparation of known concentrations: Through spiking or dilution of samples to evaluate linearity and limits of measurement.
        • Comparison to a validated comparator assay: For method comparison (Cystatin C on Alinity c vs a Comparator Cystatin C assay).
        • Characterized control materials: For precision and reproducibility studies.
        • Established healthy population studies: For determining the reference interval.
    7. The sample size for the training set:

      • Not applicable in the AI/ML sense. This is a traditional chemical assay, not an AI/ML model that undergoes a "training" phase with data. The assay's parameters are developed through conventional analytical chemistry principles and validated through the nonclinical performance studies detailed.
      • Development would involve formulation and optimization studies, but not "training data" in the AI sense.
    8. How the ground truth for the training set was established:

      • Not applicable for this in vitro diagnostic device. As noted above, there is no "training set" in the AI/ML context for a traditional immunoturbidimetric assay. The "ground truth" for developing such an assay would be based on fundamental chemical principles, known concentrations of analytes, and established analytical chemistry standards.
    Ask a Question

    Ask a specific question about this device

    K Number
    K242685

    Validate with FDA (Live)

    Date Cleared
    2024-12-04

    (89 days)

    Product Code
    Regulation Number
    862.1225
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    10591

    Re: K242685

    Trade/Device Name: Atellica® CH Creatinine 3 (Crea3) Regulation Number: 21 CFR 862.1225
    | CGX |
    | Regulation Number | 21 CFR 862.1225

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Atellica® CH Creatinine_3 (Crea3) assay is for in vitro diagnostic use in the quantitative determination of creatinine in human serum, plasma (lithium heparin, dipotassium EDTA, and sodium heparin), and urine using the Atellica® CH Analyzer. Such measurements are used in the diagnosis and treatment of renal diseases, and in monitoring renal dialysis.

    Device Description

    The Atellica CH Crea3 assay is based on the reaction of picrate with creatinine in an alkaline medium to produce a red chromophore creatinine picrate complex. The rate of complex formation is measured at 505/571 nm and is proportional to the creatinine concentration. The Atellica CH Crea3 assay is a modification of the Jaffe method, using rate blanking and intercept correction. Rate blanking is used to minimize bilirubin interference. Also, because non-specific serum/plasma protein interactions with this reagent have been found to produce a positive bias of approximately 0.3 mg/dL (26.5 µmol/L), serum/plasma measurements are automatically corrected by subtracting 0.3 mg/dL (26.5 µmol/L) from each result.

    AI/ML Overview

    The provided text describes the performance characteristics and studies for the Atellica® CH Creatinine_3 (Crea3) assay, a new in vitro diagnostic device for quantitative determination of creatinine. It compares this new device to a predicate device, the Atellica® CH Creatinine_2 (Crea_2) assay.

    Here's an analysis of the acceptance criteria and the study proving the device meets them, based on the provided text:

    Important Note: The document focuses on establishing substantial equivalence for an in vitro diagnostic (IVD) test, which primarily relies on analytical performance characteristics rather than clinical outcome studies or multi-reader multi-case (MRMC) comparative effectiveness studies typically seen with imaging AI devices. Therefore, some of your requested information (like number of experts for ground truth, adjudication methods, MRMC studies, and training set details for an AI model) are not directly applicable or provided in this type of submission.


    Acceptance Criteria and Reported Device Performance

    The acceptance criteria for this device are established through various analytical performance studies, primarily comparing it to a legally marketed predicate device (Atellica® CH Creatinine_2). The acceptance criteria are implicitly defined by the successful demonstration of equivalence or meeting pre-defined performance goals for each characteristic.

    Here's a table summarizing the acceptance criteria (inferred from the "designed to have" or "determined in accordance with" statements and the reported results meeting these) and the reported device performance:

    Performance CharacteristicAcceptance Criteria (Implicit)Reported Device Performance (Atellica® CH Creatinine_3 (Crea3))
    Detection CapabilityLoB: $\le$ LoD for serum and urine samples. LoD: $\le$ 0.15 mg/dL for serum/plasma; $\le$ 3.00 mg/dL for urine. LoQ: $\le$ 0.15 mg/dL for serum/plasma with $\le$ 0.10 mg/dL total analytical error; $\le$ 3.00 mg/dL for urine with $\le$ 1.50 mg/dL total analytical error.Serum/plasma: LoB: 0.05 mg/dL LoD: 0.10 mg/dL LoQ: 0.15 mg/dL Urine: LoB: 0.50 mg/dL LoD: 1.00 mg/dL LoQ: 3.00 mg/dL (All results meet the stated design goals/acceptance criteria).
    PrecisionDetermined in accordance with CLSI Document EP05-A3 (indicates adherence to specific statistical targets for repeatability and within-lab precision, implicitly accepted if within CLSI guidelines for the assay's use).Serum Samples (n=80 each): - Serum 1 (0.38 mg/dL): Repeatability SD 0.006, CV 1.6%; Within-Lab SD 0.012, CV 3.2% - Serum 2 (0.73 mg/dL): Repeatability SD 0.023, CV 3.2%; Within-Lab SD 0.029, CV 4.0% - Serum 3 (0.73 mg/dL): Repeatability SD 0.006, CV 0.8%; Within-Lab SD 0.019, CV 2.6% - Serum 4 (1.18 mg/dL): Repeatability SD 0.007, CV 0.6%; Within-Lab SD 0.019, CV 1.6% - Serum QC 1 (1.85 mg/dL): Repeatability SD 0.007, CV 0.4%; Within-Lab SD 0.024, CV 1.3% - Serum QC 2 (6.21 mg/dL): Repeatability SD 0.011, CV 0.2%; Within-Lab SD 0.067, CV 1.1% - Serum 5 (17.39 mg/dL): Repeatability SD 0.035, CV 0.2%; Within-Lab SD 0.189, CV 1.1% - Serum 6 (28.54 mg/dL): Repeatability SD 0.056, CV 0.2%; Within-Lab SD 0.317, CV 1.1% Urine Samples (n=80 each): - Urine 1 (56.74 mg/dL): Repeatability SD 0.102, CV 0.2%; Within-Lab SD 0.746, CV 1.3% - Urine 2 (135.80 mg/dL): Repeatability SD 0.206, CV 0.2%; Within-Lab SD 1.601, CV 1.2% - Urine QC 1 (195.79 mg/dL): Repeatability SD 0.253, CV 0.1%; Within-Lab SD 2.376, CV 1.2% (All results demonstrate low CVs, indicating good precision).
    ReproducibilityDetermined in accordance with CLSI Document EP05-A3 (implies meeting specific statistical targets for variability components across different days, lots, and instruments).Serum Samples (n=225 each): Overall CV (%) for reproducibility ranges from 1.0% to 5.0%. Urine Samples (n=225 each): Overall CV (%) for reproducibility ranges from 1.4% to 1.6%. (All results demonstrate good reproducibility across conditions).
    Assay ComparisonSerum: Correlation coefficient $\ge$ 0.950 and slope of 1.00 $\pm$ 0.05, compared to predicate (Atellica CH Creatinine 2), using Weighted Deming regression. Urine: Correlation coefficient $\ge$ 0.950 and slope of 0.000 $\pm$ 3.00, compared to predicate (Atellica CH Creatinine 2), using Weighted Deming regression.Serum (n=151): Regression equation y = 1.00x - 0.04 mg/dL, correlation coefficient (r) = 1.000. Sample range 0.44 to 28.64 mg/dL. Urine (n=113): Regression equation y = 1.00x + 0.14 mg/dL, correlation coefficient (r) = 1.000. Sample range 12.60 to 237.06 mg/dL. (Both serum and urine results meet the acceptance criteria for correlation and slope).
    Specimen EquivalenceDetermined using Weighted Deming regression (implicitly, the regression line should demonstrate equivalence, i.e., close to y=x, with high correlation coefficient).Sodium Heparin (n=50): y = 1.00x + 0.00 mg/dL, r=0.999. Lithium Heparin (n=50): y = 0.99x + 0.06 mg/dL, r=0.999. Dipotassium EDTA (n=50): y = 0.98x + 0.04 mg/dL, r=0.998. (All demonstrate strong equivalence to serum reference).
    Interferences (HIL)$\le$ 10% interference from hemoglobin, bilirubin, and lipemia. Bias > 10% or 0.15 mg/dL (whichever is greater for serum/plasma) is considered interference.Reported biases for Hemoglobin (1000 mg/dL), Conjugated Bilirubin (40-45 mg/dL), Unconjugated Bilirubin (45-60 mg/dL), and Lipemia (2250-3000 mg/dL) are all within the $\pm$10% or $\pm$0.15 mg/dL threshold for the tested analyte concentrations, demonstrating acceptable interference profiles.
    Interfering SubstancesBias $\le$ 10% or $\pm$0.15 mg/dL for Serum/plasma samples. Bias $\le$ 10% for Urine samples (for listed substances).Most tested substances (e.g., Acetaminophen, Ascorbic Acid, etc.) show negligible bias, meeting the criteria. Substances showing bias beyond acceptance criteria for Serum: - Cefoxitin: Significant interference (e.g., 243.6% and 947.9% bias at high concentrations). - Cephalothin: Shows significant bias (e.g., 44.0% bias at 180 mg/dL). - Glucose: Shows bias beyond 10% at higher concentrations (e.g., 11.5% at 500 mg/dL and 22.5% at 1000 mg/dL). - Total Protein: Shows bias beyond 0.15 mg/dL at 15 g/dL (0.45 mg/dL). - Acetohexamide: Shows bias beyond 10% at 2.0 mg/dL (10.4%). - Hydroxocobalamin (Cyanokit): Shows significant bias (e.g., 14.5% and 49.3% at higher concentrations). Substances showing bias beyond acceptance criteria for Urine: - Cefoxitin: Shows bias beyond 10% at higher concentrations (e.g., 11.3% and 15.4%). (The document explicitly lists these substances under "Interference beyond $\pm$10% for Serum" and "Interference beyond $\pm$10% for Urine," indicating that they failed the non-interference criteria at the tested concentrations. This is typical for IVD submissions, where known interferences are identified for labeling purposes).
    StandardizationThe assay shall be traceable to the reference material SRM967, from the National Institute of Standards and Technology (NIST).Statement confirms the assay is traceable to NIST SRM967.

    Study Details:

    1. Sample Size and Data Provenance:

      • Test Set Sample Sizes:
        • Detection Capability: Not explicitly stated as "sample size" but data points obtained according to CLSI EP17-A2.
        • Precision: 80 data points per serum/urine sample type (duplicate runs for 20 days, 2 runs/day).
        • Reproducibility: 225 data points per serum/urine sample type (n=5 in 1 run for 5 days using 3 instruments and 3 reagent lots).
        • Assay Comparison: 151 serum samples and 113 urine samples.
        • Specimen Equivalence: 50 samples for each plasma type (Sodium Heparin, Lithium Heparin, Dipotassium EDTA) compared to serum.
        • Interference (HIL & Non-Interfering Substances): Not explicitly stated as a total sample size, but experiments are designed to test specific analyte concentrations with and without interferents, following CLSI EP07-ED3.
      • Data Provenance: Not explicitly stated in terms of country of origin. Given the manufacturer (Siemens Healthcare Diagnostics Inc. in Tarrytown, New York, USA) and FDA submission, it's highly probable the studies were conducted in the US or in compliance with US regulatory standards. The studies described are retrospective in the sense that they use pre-collected or prepared samples to assess the analytical performance of the device under controlled conditions, not prospective in tracking patient outcomes in a clinical trial.
    2. Number of experts used to establish the ground truth for the test set and qualifications of those experts:

      • For an in vitro diagnostic (IVD) device measuring a quantitative analyte like creatinine, "ground truth" is typically established by reference methods or established laboratory standards and calibrators, not by human expert consensus or labeling of medical images.
      • The "ground truth" for creatinine concentration in this context is based on traceable reference materials (NIST SRM 967) and established laboratory measurement principles, and the performance is compared against a legally marketed predicate device.
      • Therefore, this question (relevant for AI/imaging devices) does not directly apply to this type of IVD submission.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not applicable. Adjudication methods are typically used in clinical trials or image labeling pipelines where there's human interpretation involved and a need to resolve disagreements among multiple readers; this is an analytical performance study of an IVD assay.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not applicable. This is not an AI/imaging device. It's an in vitro diagnostic assay.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • This is an automated IVD assay performed on the Atellica® CH Analyzer. Its intended use is quantitative determination of creatinine. Therefore, the performance described (precision, accuracy, interference, etc.) is its standalone performance without a human in the loop for the analytical measurement itself, though a human still interacts with the instrument and interprets the results in a clinical context.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The primary "ground truth" for the Atellica® CH Creatinine_3 assay's performance is traceability to NIST SRM 967 (a certified reference material for creatinine) and comparison to a legally marketed predicate device (Atellica® CH Creatinine_2) using method comparison validated against CLSI guidelines. This is a form of analytical reference standard and comparative performance to an established method.
    7. The sample size for the training set:

      • This device is an analytical chemistry assay, not a machine learning/AI algorithm that requires a "training set" in the computational sense. The "development" or "optimization" of the assay would involve various experimental data, but it's not codified as a "training set" for an algorithm.
    8. How the ground truth for the training set was established:

      • Not applicable, as there is no "training set" in the AI/ML context for this type of device. The assay development would rely on scientific principles of analytical chemistry, reagent formulation, and instrument calibration against known standards.
    Ask a Question

    Ask a specific question about this device

    K Number
    K221813

    Validate with FDA (Live)

    Date Cleared
    2024-11-19

    (881 days)

    Product Code
    Regulation Number
    862.1225
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    K221813

    Trade/Device Name: Nova Allegro UACR Assay, Nova Allegro Analyzer Regulation Number: 21 CFR 862.1225
    862.1645 | I | JIQ |
    | Creatinine Test System | 862.1225

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Nova Allegro urine albumin creatinine ratio (UACR) Assay is intended for the quantitative determination of albumin, creatinine, and the albumin/creatinine ratio (UACR) in human urine. The measurement of urine albumin, creatinine, and albumin/creatinine ratio aids in the early diagnosis of nephropathy.

    The Nova Allegro Analyzer is intended for in vitro diagnostic use in clinical laboratory and near-patient testing (point-of-care) settings for the quantitative determination of Nova Allegro Assays using Nova Allegro Test Cartridges.

    Device Description

    Nova Allegro UACR Assay: The Nova Allegro UACR Assay is a completely automated assay for the determination of albumin and creatinine in human urine. The results are used to calculate the UACR (Urine Albumin to Creatinine Ratio). Nova Allegro UACR Test Cartridges are the key element a user interacts with to determine the albumin and creatinine levels in a human urine sample. The main components of the Test Cartridge are the Capillary, the reaction chamber, and the barcode label.

    Nova Allegro Analyzer: The Nova Allegro Analyzer is a compact, point-of-care analyzer that features a clinically important menu of measured and calculated tests. All tests are measured with disposable, ready-to-use cartridges. The analyzer supports multiple wavelengths that are used to measure the assay of interest. The analyzer consists of the following key systems/components that the user interacts with:

    • Two analytical bays where the single use test cartridges are analyzed
    • Color Touchscreen Display
    • Barcode Scanner
    • Printer
    • Data Export Options
    • Ethernet Connection
    • USB Port
    AI/ML Overview

    The provided text is a 510(k) Summary for the Nova Allegro UACR Assay and Analyzer. It details the device's performance testing results. According to the prompt's request, I will extract relevant information to describe the acceptance criteria and the study proving the device meets these criteria.

    It's important to note that this document describes a medical device, specifically an in-vitro diagnostic test, not an AI/ML-based device. Therefore, many of the requested fields related to AI/ML specific quality evaluation criteria, such as "number of experts used to establish ground truth," "adjudication method," "MRMC study," "effect size of human readers improving with AI," "standalone performance," and details about training sets/ground truth for training, are not applicable in this context. The study performed is a performance validation of a diagnostic assay, comparing its results to a comparative method (CM), not a reader study.

    Here's a breakdown of the requested information based on the provided document:


    Device: Nova Allegro UACR Assay, Nova Allegro Analyzer
    Indications for Use: The Nova Allegro urine albumin creatinine ratio (UACR) Assay is intended for the quantitative determination of albumin, creatinine, and the albumin/creatinine ratio (UACR) in human urine. The measurement of urine albumin, creatinine, and albumin/creatinine ratio aids in the early diagnosis of nephropathy. The Nova Allegro Analyzer is intended for in vitro diagnostic use in clinical laboratory and near-patient testing (point-of-care) settings for the quantitative determination of Nova Allegro Assays using Nova Allegro Test Cartridges.


    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are generally implied by the successful conclusion of each test and the statement that the data "meets the acceptance criteria" or "demonstrated no significant interference." Specific quantitative acceptance criteria are sometimes stated (e.g., <10% difference for interference).

    Performance MetricAcceptance Criteria (Implied/Stated)Reported Device Performance
    LinearityData indicates the system reports linear albumin and creatinine test results across the reportable range when compared to the Comparative Method (CM). Expected to "meet the acceptance criteria."The linearity data met the acceptance criteria when compared to the CM. The data indicates that the Nova Allegro System reports linear albumin and creatinine test results using urine specimens across the reportable range.
    Hook EffectOccurrence above 5,000 mg/L (10 times higher than previously stated criteria in the document).The "Hook Effect" was determined to occur above 50,000 mg/L, meeting the acceptance criteria. This is 10 times higher than the acceptance criteria of 5,000 mg/L and approximately 167 times greater than the upper reportable measurement range for the assay.
    Detection LimitLOB, LOD, and LOQ for Albumin and Creatinine measurements must be below the lower limit of the claimed measurement range.The study demonstrates that the LOB, LOD, and LOQ results for Albumin and Creatinine measurements by the Nova Allegro UACR Assay are all below the lower limit of the claimed measurement range.
    InterferenceA test substance was considered to interfere if the absolute difference between the mean test value and the mean control value was greater than 10%.Most tested substances demonstrated no significant interference (greater than 10% effect). Specific interfering substances (Table 2) identified with their highest non-interfering concentrations. Urines with pH between 4.3 – 10.0 exhibited no interference (except low Albumin at pH 4.0). Specific gravity between 1.003 and 1.048 exhibited no clinical interference.
    Method ComparisonIndicated by "data supports the 510(k) substantial equivalence approval decisions." For albumin, creatinine, and UACR, linear regression analysis "r" values (correlation coefficients) close to 1 (e.g., ≥0.99) and slopes/intercepts indicative of close agreement with the CM.Albumin (ALL Combined): Slope = 0.983, Intercept = 0.858, r = 0.998. Sample range Allegro 5-300 mg/L.
    Creatinine (ALL Combined): Slope = 0.949, Intercept = 3.775, r = 0.997. Sample range Allegro 16-498 mg/dL.
    UACR (ALL Combined): Slope = 0.995, Intercept = 1.101, r = 0.997. Sample range Allegro 1-1592.
    Precision (Repeatability)Low %CV values (e.g., <5-10%) for mean concentrations across sites, indicating consistent results within a laboratory.Microalbumin: Overall %CV: Urine 1 (3.6%), Urine 2 (4.0%), Urine 3 (3.6%)
    Creatinine: Overall %CV: Urine 1 (4.0%), Urine 2 (3.1%), Urine 3 (3.4%)
    UACR: Overall %CV: Urine 1 (2.5%), Urine 2 (2.5%), Urine 3 (2.4%)
    Precision (Between Run/Operator)Low %CV values.Microalbumin: Overall %CV: Urine 1 (2.6%), Urine 2 (0.3%), Urine 3 (1.5%)
    Creatinine: Overall %CV: Urine 1 (2.3%), Urine 2 (0.5%), Urine 3 (2.7%) -----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
    UACR: Overall %CV: Urine 1 (1.2%), Urine 2 (0.6%), Urine 3 (1.8%)
    Precision (Between Lot/Day)Low %CV values.Microalbumin: Overall %CV: Urine 1 (1.3%), Urine 2 (1.3%), Urine 3 (0.5%)
    Creatinine: Overall %CV: Urine 1 (1.0%), Urine 2 (0.5%), Urine 3 (0.5%)
    UACR: Overall %CV: Urine 1 (0.5%), Urine 2 (1.1%), Urine 3 (0.2%)
    Precision (Within Laboratory)Low %CV values across sites.Microalbumin: Overall %CV: Urine 1 (4.6%), Urine 2 (4.1%), Urine 3 (4.2%) ---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
    Creatinine: Overall %CV: Urine 1 (4.8%), Urine 2 (3.4%), Urine 3 (4.5%)
    UACR: Overall %CV: Urine 1 (2.9%), Urine 2 (3.0%), Urine 3 (3.2%)
    Precision (Between Site)Low %CV values.Microalbumin: Overall %CV: Control 1 (1.0%), Control 2 (1.6%)
    Creatinine: Overall %CV: Control 1 (1.0%), Control 2 (0.9%) -----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
    UACR: Overall %CV: Control 1 (1.2%), Control 2 (1.2%)

    2. Sample Size and Data Provenance

    • Test Set Sample Sizes:

      • Linearity Testing: Not explicitly stated as a number of patient samples, but involved "UACR control solutions (Level 1 & Level 2)" and "two (2) urine pools containing low and very high albumin and creatinine concentrations."
      • Hook Effect: Involved "both pooled urine specimens and calibrator matrix."
      • Detection Limit: Sixty (60) replicate measurements for LOB, thirty-six (36) replicate measurements for LOQ.
      • Interference Testing: Ten (10) replicate UACR tests were performed on prepared urine specimens per interfering substance.
      • Method Comparison: A total of 535 samples within the measurement range for Albumin (41 altered) and 653 samples within the measurement range for Creatinine (0 altered). For UACR, 531 samples were tested (41 altered). (Note: the number of altered vs. unaltered samples is specified. The altered samples are used to ensure coverage of the analytical measurement range).
      • Precision: Urine samples (3 levels) were tested in replicates over multiple days/runs at 4 sites. Specific numbers of replicates are indicated by the CLSI guidance used, but not explicitly enumerated per site/level in the summary beyond general study setup.
    • Data Provenance:

      • Country of Origin: Not explicitly stated in the summary.
      • Retrospective or Prospective: The "Point-of-Care (POC) Method Comparison studies on fresh urine specimens were conducted within four (4) different POC sites." This implies a prospective collection of samples during the study.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    • Not applicable for this type of diagnostic device performance study. Ground truth is established by a "Comparative Method" (CM), which is another established, legally marketed, and often more robust laboratory assay.
    • The "CM method" for albumin was the "Siemens Dimension EXL 200 Integrated Chemistry System utilizing Dimension® Flex® reagent cartridge MALB, Dimension calibrator cartridge MALB CAL (Traceable to IFCC International Reference Preparation for Plasma Proteins)."
    • The "CM method" for creatinine was "Dimension Flex Reagent Cartridge CRE2, and Dimension CHEM 1 CAL (Traceable to NIST SRM 914 and verified with NIST SRM 967)." These are reference laboratory methods, not human expert interpretations.

    4. Adjudication Method for the Test Set

    • None. This is a quantitative chemical assay, not an interpretative AI/ML system requiring human adjudication of ground truth labels. Comparison is made against a validated laboratory reference method.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and what was the effect size of how much human readers improve with AI vs. without AI assistance

    • Not applicable. This is not an AI/ML-based device that assists human readers. It is a standalone diagnostic test for laboratory and point-of-care use.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Yes, this is a standalone device. The Nova Allegro Analyzer and UACR Assay provides quantitative results directly. Its performance is validated against a comparative method in a standalone manner. The "human-in-the-loop" here is the operator running the test, not an interpretive role.

    7. The Type of Ground Truth Used

    • Comparative Method / Reference Method: The ground truth for the "Method Comparison" study was established using the Siemens Dimension EXL 200 Integrated Chemistry System, which served as the reference standard (CM method). This system's calibration is traceable to international reference preparations (IFCC) and NIST standards.

    8. The Sample Size for the Training Set

    • Not applicable. As an in-vitro diagnostic device, this product does not involve machine learning or AI models that require a "training set." Its analytical performance is based on established chemical and immunoturbidimetric principles, validated through traditional analytical and clinical performance studies, not algorithmic training data.

    9. How the Ground Truth for the Training Set was Established

    • Not applicable. See point 8.
    Ask a Question

    Ask a specific question about this device

    K Number
    K232317

    Validate with FDA (Live)

    Date Cleared
    2024-04-26

    (267 days)

    Regulation Number
    864.6550
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    | Class I | Automated Urinalysis System | Clinical Chemistry |
    | 2 | 862.1225

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The UC-1800 Automatic Urine Analyzer is automated instrument which is intended for professional, in vitro diagnostic use only.

    Depending on the reagent strips being used, the instruments perform semi-quantitative detection of the following analytes in urine: ascorbic acid, microalbumin, leukocytes, creatinine, ketone, urobilinogen, bilirubin, glucose, protein, specific gravity, blood and pH in urine and for qualitative determination of nitrite in urine hydrometer (optional) can determine the color and turbidity of urine. Test results may provide information regarding the status of carbohydrate metabolism, kidney and liver function, acid-base balance and bacteriuria.

    The URIT 11FA urine reagent strips provide semi-quantitative tests for ascorbic acid, leukocytes, setone, urobilinogen, bilirubin, glucose, protein, specific gravity, blood and pH in urine and for qualitative determination of nitrite in urine. The URIT 11FA urine reagent strips are for use with the UC-1800 Automatic Urine Analyzer and are for professional, in vitro diagnostic use only. Test results may provide information regarding the status of carbohydrate metabolism, kidney and liver function, acid-base balance and bacteriuria.

    The URIT 12FA urine reagent strips provide semi-quantitative tests for microalbumin, leukocytes, creatinine, ketone, urobilinogen, bilirubin, glucose, protein, specific gravity, blood and pH in urine and for qualitative determination of nitrite in urine. The URIT 12FA urine reagent strips are for use with the UC-1800 Automatic Urine Analyzer and are for professional, in vitro diagnostic use only. Test results may provide information regarding the status of carbohydrate metabolism, kidney and liver function, acid-base balance and bacteriuria.

    Device Description

    UC-1800 Automatic Urine Analyzer is characterized by fully automated and simple operation. All you need to do is to set test strips and samples, press the START key, and the rest of operations are fully automated with UC-1800, which can measure samples continuously. For each measurement, the instrument automatically performs a series of operation: sample transmitting, sample aspirating, sample dropping, rinsing strip feeding and color identifying, etc. The instrument is used in conjunction with a serial of URIT urine test strips for measuring 15 parameters. Measure results are printed through either built-in printer or external printer.

    Urine Reagent Strips is used to determine the components to be measured in urine by dry chemistry method together with urine analyzer. Various components to be tested in the urine can result in changes to the colors of corresponding reagent blocks on the Urine Reagent Strips. The depth of reaction color is proportional to the corresponding component to be tested in the urine. Qualitative and semi-quantitative detection can be conducted to the contents of the corresponding detected components. As a reagent for the determination of multiple components in human urine and the most basic test item for clinical urine routine test), it is suitable for the screening test or auxiliary diagnosis for clinical diagnosis, without the specificity for diseases or indications, and urine dry chemistry test is a screening test and cannot be used as a single diagnostic method.

    AI/ML Overview

    The provided document describes the URIT UC-1800 Automatic Urine Analyzer and its associated reagent strips (URIT 11FA and 12FA Urine Reagent Strips). The information below summarizes the acceptance criteria and the studies performed to demonstrate the device meets these criteria.


    1. Table of Acceptance Criteria and Reported Device Performance

    The document presents the performance in terms of "Exact agreement" and "± 1 color block" agreement with expected values or between the proposed device and predicate devices. For most analytes, the criteria seem to be high exact agreement and 100% agreement within ±1 color block. Specific thresholds for acceptance were not explicitly stated as global criteria but are implied by the "Qualified" conclusions for individual tests. The comparison to predicates also uses agreement rates.

    Below is a summary of the reported device performance from the "Precision / Reproducibility" section (Tables 8, 9, 10, 11) for repeatability and "Comparison Studies" section (Tables a.1, a.2, a.3) for agreement with predicate devices. Given the extensive number of analytes and concentration levels, key representative results are presented.

    a. Repeatability (Within-Run Precision)
    Reported as "Exact agreement" and "± 1 color block". All tests for all expected values show 100% agreement for "± 1 color block". Exact agreement varies slightly, but most are 100% or very close.

    Test (Analyte)Expected Value (Example)Exact Agreement (Proposed Device: UC-1800 with 11FA/12FA strips)Conclusion
    Ascorbic acid (11FA)-(0) mg/dL100% (60/60)Qualified
    Nitrite (11FA/12FA)- (Negative)100% (60/60)Qualified
    Leukocyte (11FA/12FA)-(0) leu/μL100% (60/60)Qualified
    Ketone (11FA)-(0) mg/dL100% (60/60)Qualified
    Ketone (12FA)+1(15) mg/dL96.7% (58/60)Qualified
    Urobilinogen (11FA/12FA)Normal EU/dL100% (60/60)Qualified
    Bilirubin (11FA)+2(2.0) mg/dL96.7% (58/60)Qualified
    Bilirubin (12FA)+2(2.0) mg/dL91.7% (55/60)Qualified
    Glucose (11FA/12FA)All100% (60/60)Qualified
    Protein (11FA/12FA)+1(30) mg/dL93.3% (56/60)Qualified
    Specific Gravity (11FA)1.01096.7% (58/60)Qualified
    Specific Gravity (12FA)All100% (60/60)Qualified
    Blood (11FA)+1(25) Cell/μL98.3% (59/60)Qualified
    Blood (12FA)All100% (60/60)Qualified
    pH (11FA)7.096.7% (58/60)Qualified
    pH (12FA)All100% (60/60)Qualified
    Microalbumin (12FA)All100% (60/60)Qualified
    Creatinine (12FA)All100% (60/60)Qualified
    Turbidity (Physical Module)All100% (60/60)Qualified
    Color (Physical Module)All100% (60/60)Qualified

    b. Reproducibility (Between-Run Precision)
    Reported as "Exact agreement" and "± 1 color block". All tests for all expected values show 100% agreement for "± 1 color block". Exact agreement varies slightly, but most are 100% or very close.

    Test (Analyte)Expected Value (Example)Exact Agreement (Proposed Device: UC-1800 with 11FA/12FA strips)Conclusion
    Ascorbic acid (11FA)+2(50) mg/dL94.2% (113/120)Qualified
    Protein (11FA)+2(100) mg/dL99.2% (119/120)Qualified
    Blood (11FA)+2(80) Cell/μL95% (114/120)Qualified
    pH (11FA)7.099.2% (119/120)Qualified
    Nitrite (12FA)All100% (120/120)Qualified
    Blood (12FA)+2(80) Cell/μL95.8% (115/120)Qualified
    All other analytes (11FA/12FA)Most levels100% (120/120) or very closeQualified

    c. Comparison with Predicate Devices

    Analyte (Strip)Predicate DeviceAgreement TypeAgreement Rate (%) (Reported value from the table)Conclusion
    Ascorbic acid (11FA)Uritest-500B (K082811)Complete agreement97.60% to 100% across all levels (e.g., 99.00% for +1(25) level)Subst. Eq.
    Leukocyte (11FA/12FA)Uritest-500B (K082811)Complete agreement95.80% to 100% across all levels (e.g., 95.80% for +2(125))Subst. Eq.
    Ketone (11FA/12FA)Uritest-500B (K082811)Complete agreement97.40% to 100% across all levels (e.g., 97.40% for +1(15))Subst. Eq.
    Nitrite (11FA/12FA)Uritest-500B (K082811)Overall Agreement (OPA)99.40% (11FA); 98.70% (12FA)Subst. Eq.
    Urobilinogen (11FA/12FA)Uritest-500B (K082811)Complete agreement95.50% to 100% across all levels (e.g., 95.50% for +1(2.0) on 12FA)Subst. Eq.
    Bilirubin (11FA/12FA)Uritest-500B (K082811)Complete agreement95.20% to 100% across all levels (e.g., 95.20% for +2(2.0) on 12FA)Subst. Eq.
    Glucose (11FA/12FA)Uritest-500B (K082811)Complete agreement96.30% to 100% across all levels (e.g., 96.30% for +1(100))Subst. Eq.
    Protein (11FA/12FA)Uritest-500B (K082811)Complete agreement95.50% to 100% across all levels (e.g., 95.50% for +1(30) on 12FA)Subst. Eq.
    pH (11FA/12FA)Uritest-500B (K082811)Complete agreement97.80% to 100% across all levels (e.g., 97.80% for 6.5)Subst. Eq.
    Specific Gravity (11FA/12FA)Uritest-500B (K082811)Complete agreement97.80% to 100% across all levels (e.g., 97.80% for 1.010 on 12FA)Subst. Eq.
    Microalbumin (12FA)Mission® U120 Ultra Urine Analyzer (K142391)Complete agreement83.33% for 30mg/L, up to 98.21% for 150mg/LSubst. Eq.
    Creatinine (12FA)Mission® U120 Ultra Urine Analyzer (K142391)Complete agreement90.91% for 300mg/dL, up to 97.74% for 50mg/dLSubst. Eq.
    Color (Physical module)AUTION MAX AX-4030 Urinalysis System (K093098)Coincidence rate91.29% (Colorless) to 98.48% (Brown)Subst. Eq.
    Turbidity (Physical module)AUTION MAX AX-4030 Urinalysis System (K093098)Coincidence rate96.00% (Turbid) to 100% (Micro turbid)Subst. Eq.

    2. Sample Sizes Used for the Test Set and Data Provenance

    • Repeatability (within-run) & Reproducibility (between-run):

      • Sample Size:
        • Repeatability: 60 measurements per concentration level per analyte (20 replicates x 3 instruments).
        • Reproducibility: 120 measurements per concentration level per analyte (20 days x 2 runs/day x 1 time/run in 3 sites, with 1 instrument/site, 3 operators).
      • Data Provenance: The document does not explicitly state the country of origin or if the data was retrospective or prospective. It refers to "negative urines and spiked urines of known concentrations." This suggests controlled laboratory-prepared samples rather than direct patient samples.
    • Linearity/Assay Reportable Range:

      • Sample Size: 63 measurements per concentration level per analyte (reference solutions tested 21 times on 3 UC-1800 machines with 3 lots of reagent strips).
      • Data Provenance: Laboratory-prepared reference solutions, not clinical samples.
    • Analytical Sensitivity: No sample sizes mentioned, values are stated directly.

    • Critical Value: No sample sizes mentioned, values are stated directly.

    • Analytical Specificity (Interference, pH, Color, Specific Gravity):

      • Sample Size: For interference studies, samples were tested 5 times on 2 UC-1800 machines with 2 batch numbers of 11FA and 12FA reagent strips. This totals 20 measurements per condition (5 tests x 2 machines x 2 strips).
      • Data Provenance: Laboratory-prepared urine samples (negative samples prepared and spiked with interfering substances or adjusted for pH/color/SG).
    • Comparison Studies with Predicate Devices (Clinical Samples):

      • Uritest-500B (K082811): 1000 clinical urine samples.
      • Mission® U120 Ultra Urine Analyzer (K142391): 979 clinical urine samples.
      • AUTION MAX AX-4030 Urinalysis System (K093098):
        • Color: 1365 clinical urine samples.
        • Turbidity: 1000 clinical urine samples.
      • Data Provenance: The document explicitly states "clinical urine samples were collected." The country of origin is not specified but given the submitter's address (China), it is highly likely that these clinical samples were collected in China and were retrospective as the study compares against existing predicate devices.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not mention the use of experts to establish a "ground truth" for the test set in the traditional sense of medical image interpretation or clinical diagnosis. For chemical analyzers, the "ground truth" (or reference standard) is typically established by:

    • Known concentrations: For repeatability, reproducibility, linearity, and analytical sensitivity, samples are prepared with known concentrations of the analytes.
    • Reference methods: For the analytical specificity and comparison studies, the predicate devices themselves or established reference methods (e.g., 2,6-Dichlorophenolindophenol Titration Method for Ascorbic Acid, Lange method for Ketone, etc. as specified in Table 35 "Traceability") serve as the reference for comparison.

    Therefore, the concept of "number of experts" and their "qualifications" for ground truth establishment, as typically applied in AI/ML performance evaluation (e.g., for image interpretation), is not directly applicable here. The ground truth is analytical and based on laboratory standards and established measurement techniques.


    4. Adjudication Method for the Test Set

    Not applicable in the context of this analytical device. As explained above, the "ground truth" refers to known concentrations or results from predicate/reference methods. There isn't a subjective interpretation by multiple experts that would require an adjudication method like 2+1 or 3+1.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is an automated urine analyzer, not an AI/ML-driven diagnostic aid for human readers. It performs measurements to provide semi-quantitative results for various urine analytes. Therefore, there is no human-in-the-loop performance or comparison of human reader improvement with or without AI assistance.


    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Yes, the studies presented are all standalone performance evaluations of the device (UC-1800 Automatic Urine Analyzer with URIT 11FA/12FA Urine Reagent Strips) without human intervention in the measurement process. The device operates automatically to detect and report analyte levels. The comparison studies demonstrate its performance against existing predicate (standalone) analyzers.


    7. The Type of Ground Truth Used

    The ground truth used in the studies includes:

    • Known Concentrations: For analytical performance studies such as precision (repeatability and reproducibility), linearity, analytical sensitivity, and analytical specificity (interference testing), samples were prepared with known, precisely measured concentrations of the target analytes or interfering substances.
    • Predicate Device Measurements: For method comparison studies, the results obtained from the established predicate devices (Uritest-500B Urine Analyzer, Mission® U120 Ultra Urine Analyzer, and AUTION MAX AX-4030 Urinalysis System) served as the reference standard for comparison. These predicate devices also operate based on defined analytical principles.
    • Reference Methods: Table 35 details the "Reference Method" used for traceability for each analyte (e.g., 2,6-Dichlorophenolindophenol Titration Method for Ascorbic Acid, Glucose Oxidase Method for Glucose, Acidometer measurement for pH, etc.). This indicates fundamental analytical standards are the basis for the stated detection ranges and performance.

    8. The Sample Size for the Training Set

    The document does not explicitly mention a "training set" in the context of machine learning. The UC-1800 Automatic Urine Analyzer uses reflectance photometry and other physical principles (refractometer for specific gravity, light-scattering for turbidity, light-transmission for color) to generate results, not an AI algorithm that learns from a dataset in the conventional sense. The "training" of such a system typically involves calibrating optical sensors and algorithms to known standards, which is part of the engineering design and quality control processes.

    If interpretation of "training set" refers to the data used for the initial development and calibration of the device's measurement algorithms:

    • No specific sample size for a "training set" for an AI model is reported because the device does not employ machine learning that requires a distinct "training set" and "test set" in the typical AI/ML development lifecycle.
    • The system is calibrated using URIT urine control materials and calibration test strips (Table 9, "Calibration").

    9. How the Ground Truth for the Training Set Was Established

    As noted above, the device does not use an AI/ML model with a "training set" in the common understanding. The system's operational parameters and calibration are established using:

    • Reference materials and calibrators: The device is calibrated using "URIT urine control materials and calibration test strips" (Table 9, "Calibration"). These control materials and calibration strips would have their values established using highly accurate reference methods or certified reference materials, ensuring traceability to scientific standards (as indicated in Table 35 "Traceability").
    • Chemical principles: The underlying "ground truth" for the device's internal algorithms (i.e., how they convert optical signals to analyte concentrations) is based on established chemical reactions and physical measurement principles (e.g., reflectance photometry, refractometry, light scattering, dye-binding, enzymatic reactions). The "ground truth" for developing and fine-tuning these algorithms would be derived from rigorous scientific validation against these known chemical and physical properties.
    Ask a Question

    Ask a specific question about this device

    K Number
    K222921

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-09-08

    (347 days)

    Product Code
    Regulation Number
    862.1645
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Aviv, Israel 6777508

    Re: K222921

    Trade/Device Name: Minuteful-kidney test Regulation Number: 21 CFR 862.1225
    II: Creatinine

    21 CFR § 862.1645 Urinary protein or albumin (non-quantitative) test system 21 CFR § 862.1225

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Minuteful - kidney test is an in-vitro diagnostic, home-use urine analysis test system for the semi-quantitative measurement of albumin and creatinine in urine, as well as the presentation of their ratio, the albumin-creatinine ratio (ACR). The system consists of a smartphone application, proprietary Color-Board and an ACR Reagent Strip. The system is available for prescription-use only and is intended for people at risk of kidney disease. Results are used in conjunction with clinical evaluation as an aid in the assessment of kidney health.

    Device Description

    The Minuteful - kidney test is comprised of a kit and a smartphone application. It is intended for the semi-quantitative measurement of albumin and creatinine in urine, as well as the presentation of their ratio, the albumin-creatinine ratio (ACR). The Minuteful - kidney test is intended for prescription-use only, as a home-use device to aid in the assessment of kidney health. The results can be used together with clinical evaluation to guide patient care. The device's kit includes a urine receptacle, an ACR Reagent Strip, an absorbing (i.e. "blotting") pad, a proprietary Color-Board and a user manual. The device also consists of an easy-to-use smartphone application, image recognition algorithms, and a physician compendium. The software component of the Minuteful - kidney test consists of both an application (app) and a backend server. Both components encompass different computer vision and machine learning algorithmic components, performing the image analysis activities. The app instructs the user how to accurately administer the test. The Image Validation Transfer System (IVTS) component of the Minuteful - kidney test enables its usage across a wide range of smartphone types and operating systems, essentially making the test platform agnostic.

    AI/ML Overview

    The provided text describes the Minuteful-kidney test (K222921) and its substantial equivalence to a predicate device (K210069). However, it specifically states that "The rest of the analytical performance studies are still relevant for the modified version of the Minuteful - kidney test, and their summary is available in the predicate device documentation (K210069)." This means the detailed acceptance criteria and the comprehensive study demonstrating the device meets those criteria are not present in this document but are referenced as being in the predicate device's documentation.

    Therefore, I can report on the studies performed for K222921 to assert its substantial equivalence, but I cannot provide a table of acceptance criteria and reported device performance from this document for the overall device functionality as those details are in K210069. Nor can I provide information regarding sample sizes for test sets, expert qualifications, adjudication methods, MRMC studies, standalone performance, or ground truth details for K222921's overall performance since those are tied to the K210069 submission.

    The studies described in K222921 (the current device) are focused on demonstrating that changes made to the device in K222921 do not negatively impact performance, thus maintaining substantial equivalence to its predicate.

    Here's an analysis based solely on the provided text for K222921, noting the limitations:

    Acceptance Criteria and Device Performance (Limited to K222921 changes):

    Since the comprehensive performance data is referenced in K210069, the "acceptance criteria" discussed here are implicitly related to demonstrating that the modifications in K222921 (e.g., multilingual support, software enhancements) do not degrade the performance previously established for K210069. The studies conducted for K222921 focused on the robustness of the Image Validation Transfer System (IVTS) and the analytical limits of detection.

    Acceptance Criteria (Implied for K222921 changes)Reported Device Performance (K222921)
    Limit of Detection (LoD)Testing was conducted in accordance with CLSI document EP17-A2. (Specific LoD values are not provided in this document but are likely in K210069).
    Illumination ConditionsPerformance is "not impacted" by different lighting conditions (color temperatures, intensities, light sources) representative of home use, nor by different light color saturations and intensities at the edges of device boundary conditions.
    Physical ConditionsPerformance is "not impacted" by different distance and angle conditions at the edges of device boundary conditions.
    Multiple Shadow ConditionsPerformance is "not impacted" by different shadow configurations (intensity, coverage) at the edges of device boundary conditions.
    BlurrinessPerformance is "not impacted" by different levels of focus and motion blur in images at the edges of device boundary conditions.
    Misplaced Urine StickPerformance is "not impacted" by different urine test strip placements at the edges of device boundary conditions.
    Dirty Color-BoardPerformance is "not impacted" by different dirty substances covering parts of the Color-Board at the edges of device boundary conditions.
    Overall Equivalence to PredicateThe modified Minuteful-kidney test is concluded to be substantially equivalent to the predicate device (K210069), implying that the changes did not degrade its overall performance in terms of precision, interference, linearity, stability, and clinical performance, which are referenced back to the K210069 summary. The new IVTS system allows usage across a wide range of smartphone types and operating systems, making the test platform agnostic, without impacting performance in various challenging conditions.

    Study Details (for K222921 specific enhancements):

    1. Sample size used for the test set and the data provenance:

      • Limit of Detection (LoD): The document does not specify the sample size for the LoD study for K222921. It mentions the study was designed and executed according to CLSI document EP17-A2.
      • Illumination, Physical, Shadow, Blurriness, Misplaced Urine Stick, Dirty Color-Board Studies: The document refers to "Tested smartphones" and "different conditions," but specific numerical sample sizes (e.g., number of images, tests, or smartphones) are not provided. The data provenance is implied to be laboratory-controlled since these are experimental conditions, but no explicit country of origin or retrospective/prospective nature is stated for these new studies. The overall device is intended for home use, so these validations mimic adverse home conditions.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • This information is not provided in the K222921 document. These types of analytical studies typically rely on reference methods or scientifically established standards rather than expert consensus. For the clinical performance, the document refers to the predicate device K210069, where such details would likely be found if applicable.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not applicable/Not provided for these specific analytical studies. The assessment of whether performance was "not impacted" would likely come from statistical analysis against pre-defined thresholds.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No MRMC or comparative effectiveness study involving human readers with and without AI assistance is mentioned in the K222921 submission. This device is an in-vitro diagnostic home-use test system where the smartphone app performs the measurement, rather than assisting a human in interpreting diagnostic images. Thus, the concept of "human readers improve with AI" in a traditional MRMC sense does not directly apply to this device's function.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • The core of the described studies (LoD, Illumination, Physical, Shadow, Blurriness, Misplaced Urine Stick, Dirty Color-Board) are indeed standalone performance tests of the device's algorithmic capability to accurately read the test strip under various challenging conditions encountered in a home setting. The device is described as having "image recognition algorithms" and performing "image analysis activities."
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • For the analytical studies described (LoD, Illumination, etc.), the ground truth would be based on controlled experimental conditions and potentially quantitative reference measurements (e.g., known concentrations for LoD, controlled lighting conditions). The document does not specify the exact methods for establishing this ground truth but implies scientific rigor (e.g., "in accordance with guidance provided by the Clinical and Laboratory Standards Institute (CLSI) document EP17-A2"). For the clinical performance aspects, the document refers to K210069.
    7. The sample size for the training set:

      • Not provided in the K222921 document. Training set details would typically be part of the initial K210069 submission for the machine learning algorithms.
    8. How the ground truth for the training set was established:

      • Not provided in the K222921 document. This information would be found in the K210069 submission, likely involving laboratory-controlled tests with known analyte concentrations and reference methods for accurate measurement.
    Ask a Question

    Ask a specific question about this device

    K Number
    K223090

    Validate with FDA (Live)

    Date Cleared
    2023-01-27

    (119 days)

    Product Code
    Regulation Number
    862.1345
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Automated hematocrit instrument | II | GKF | 81 |
    | Creatinine | 862.1225

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The GEM Premier ChemSTAT is a portable critical care system for use by health care professionals to rapidly analyze lithium heparinized whole blood samples at the point of health care delivery in a clinical setting and in a central laboratory. The instrument provides quantitative measurements of sodium (Na+), Potassium (K+), Ionized Calcium (Ca++), Chloride (Cl-), Glucose (Glu), Lactate (Lac), Hematocrit (Hct), Creatinine (Crea), Blood Urea Nitrogen (BUN), Total Carbon Dioxide (tCO2), pH, and partial pressure of carbon dioxide (pCO2) from arterial and venous heparinized whole blood. These parameters, along with derived parameters, aid in the diagnosis of a patient's acid/base status, electrolyte and metabolite balance.

    Electrolytes in the human body have multiple roles. Nearly all metabolic processes depend on or vary with electrolytes:

    · Sodium (Na+) measurements are used in the diagnosis and treatment of aldosteronism, diabetes insipidus, adrenal hypertension, Addison's disease, dehydration, inappropriate antidiuretic secretion, or other diseases involving electrolyte imbalance.

    · Potassium (K+) measurements are used to monitor electrolyte balance in the diagnosis and treatment of disease conditions characterized by low or high blood potassium levels.

    · Ionized calcium (Ca++) measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany. · Chloride (Cl-) measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders, such as cystic fibrosis and diabetic acidosis.

    · Glucose (Glu) measurement is used in the diagnosis, monitoring and treatment of carbohydrate metabolism disturbances including diabetes mellitus, neonatal hypoglycemia, idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

    · Lactate (Lac) measurement is used to evaluate the acid-base status of patients suspected of having lactic acidosis, to monitor tissue hypoxia and strenuous physical exertion, and in the diagnosis of hyperlactatemia.

    · Hematocrit (Hct) measurements in whole blood of the packed red cell volume of a blood sample are used to distinguish normal from abnormal states, such as anemia and erythrocytosis (an increase in the number of red cells).

    · Creatinine (Crea) measurements are used in the diagnosis and treatment of renal diseases and in monitoring renal dialysis.

    · Blood Urea Nitrogen (BUN) or urea measurements are used for the diagnosis, monitoring, and treatment of certain renal and metabolic diseases.

    · Total carbon dioxide/tCO2 (also referred to as bicarbonate/HCO3-) is used in the diagnosis, monitoring, and treatment of numerous potentially serious disorders associated with changes in body acid-base balance.

    · pH and pCO2 measurements in whole blood are used in the diagnosis and treatment of life-threatening acid-base disturbances.

    Device Description

    The GEM Premier ChemSTAT system provides fast, accurate, quantitative measurements of Sodium (Na"), Potassium (K*), Ionized Calcium (Ca*), Chloride (Cl·), Glucose (Glu), Lactate (Lac), Hematocrit (Hct), Creatinine (Crea), Blood Urea Nitrogen (BUN), Total Carbon Dioxide (tCO2), pH, and partial pressure of carbon dioxide (pCO2) from arterial and venous lithium heparinized whole blood.

    AI/ML Overview

    The provided text describes a Special 510(k) submission for an upgrade to the operating system of the GEM Premier ChemSTAT device. The device itself is an in vitro diagnostic (IVD) system for quantitative measurements of various blood parameters. The submission focuses on the software upgrade rather than a change in the device's fundamental function or performance.

    Therefore, the "acceptance criteria" and "reported device performance" in this context refer to the successful verification and validation of the software upgrade and the continued adherence to the established performance of the unmodified device, as the indications for use and performance claims remain unchanged. The study proving this essentially consists of the software verification and validation activities.

    Here's the information extracted from the document, tailored to the context of a software upgrade:

    1. Table of Acceptance Criteria and Reported Device Performance

    Since this is a software upgrade with no changes to the performance claims of the device, the general acceptance criteria are that the upgraded software performs as intended without adversely affecting the device's established performance specifications. The reported device performance is that these criteria were met.

    Acceptance Criteria (Software Upgrade)Reported Device Performance (Software Upgrade)
    All identified risks associated with the design changes for the modified device are mitigated.Risk assessments were performed in compliance with ISO 14971:2019, and identified risks were mitigated.
    All software verification and validation activities are completed according to established plans and protocols.All verification and validation activities were performed in accordance with established plans and protocols and Design Control procedures.
    All acceptance criteria for software verification and validation are met.Testing verified all acceptance criteria were met.
    Cybersecurity vulnerabilities are identified, assessed, and compensating controls are implemented.Cybersecurity assessments were performed, vulnerabilities identified and assessed, and compensating controls implemented to mitigate threats and safeguard data.
    No changes to indications for use or intended use.No changes to indications for use or intended use.
    No changes to the fundamental scientific technology.No changes to the fundamental scientific technology.
    No changes to operating principle.No changes to operating principle.
    No changes to labeled performance claims.No changes to labeled performance claims.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not specify a "test set sample size" or "data provenance" in the traditional sense for evaluating diagnostic performance. The focus is on software verification and validation. Therefore, the "sample" for testing the software functionality would be the various test cases and scenarios designed to validate the operating system upgrade and its interaction with the GEM Premier ChemSTAT application software.

    The document states: "Performance data is limited to Software Verification and Validation as the scope of this Special 510(k) is specific to an operating system upgrade from Fedora 17 Linux to WindRiver LTS 18 Linux."

    Further details on the specific number of test cases, the nature of the data (e.g., simulated, actual runs on the device), or its origin are not provided in this summary.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable to a software operating system upgrade as described. "Ground truth" in the context of expert consensus is typically relevant for diagnostic performance studies where human interpretation or a gold standard reference is needed (e.g., pathology for an imaging device). Here, the "ground truth" is the proper functioning of the software and its integration with the hardware, which is evaluated through engineering and software testing.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable for a software operating system upgrade. Adjudication methods like 2+1 or 3+1 are used in clinical studies to resolve discrepancies in expert interpretation of diagnostic results. Software verification and validation typically rely on predefined test outcomes and engineering assessments.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable. An MRMC comparative effectiveness study is used to evaluate the impact of an AI algorithm on human reader performance, usually for diagnostic tasks. This submission is for a software operating system upgrade for an existing IVD device, not for a new AI algorithm.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The concept of "standalone performance" in the context of an algorithm's diagnostic capability (like an AI algorithm) is not directly applicable here. The device itself (GEM Premier ChemSTAT) operates to provide quantitative measurements. The software upgrade ensures the continued, correct operation of the device. The verification and validation activities demonstrate that the upgraded software performs its functions correctly as part of the overall device system.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For this software upgrade, the "ground truth" is the expected behavior and functionality of the software and the device. This is established through:

    • Functional specifications: The software is expected to perform according to its design specifications.
    • Risk analysis: The software should not introduce new risks or fail to mitigate existing ones.
    • Cybersecurity standards: The software should meet cybersecurity requirements.
    • Established device performance: The software upgrade should not negatively impact the established analytical and clinical performance of the GEM Premier ChemSTAT device (which relies on the physical and chemical principles of its measurements).

    The document explicitly states that the changes "do not introduce...changes to labeled performance claims." This implies that the performance of the device (e.g., accuracy, precision of Na+, K+, Glu measurements) remains the same as previously cleared, and the software upgrade was validated not to alter these.

    8. The sample size for the training set

    This information is not applicable. Training sets are used for machine learning models. This submission describes a conventional software operating system upgrade (Fedora 17 Linux to WindRiver LTS 18 Linux) for an existing IVD device, not the development or retraining of a machine learning algorithm.

    9. How the ground truth for the training set was established

    This information is not applicable, as there is no training set for a machine learning model; it is a software operating system upgrade.