K Number
K251281

Validate with FDA (Live)

Date Cleared
2026-01-21

(272 days)

Product Code
Regulation Number
862.1225
Age Range
18 - 120
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Nova Max Creat eGFR Monitoring System is comprised of the Nova Max Creat eGFR Monitor and the Nova Max Creat eGFR Test Strips.

The Nova Max Creat eGFR Monitoring System is intended for in vitro diagnostic use for the quantitative measurement of creatinine and estimation of glomerular filtration rate (eGFR) in fresh capillary whole blood obtained from the fingertip of adult patients aged 18 and above. The system is intended for single patient home use by prescription only and should not be shared. It is intended for use by patients as an aid to monitor kidney function on the order of a treating healthcare professional.

Device Description

The Nova Max Creat eGFR Monitoring System consists of Nova Max Create eGFR Monitor and Nova Max Create eGFR Creatinine-Test Strips. The Nova Max Creat eGFR test strips, Nova Max Creat eGFR control solution (Level 1 and Level 2), and a single-use, disposable, 21 gauge safety lancet can be purchased separately.

Nova Max Creat eGFR Monitor
The monitor is intended to be used in conjunction with the Nova Max Creat eGFR Creatinine-Test Strips to measure creatinine and calculate estimated glomerular filtration rate (eGFR) using capillary whole blood obtained from the fingertip. The hand-held monitor is lightweight, portable, utilizes a 2.8" color touchscreen display and has capability for storage (400 test samples and/or quality control samples) and review of data. The monitor is powered by a rechargeable, 3.7 V Li polymer battery.

Nova Max Creat eGFR Creatinine-Test Strips
The Nova Max Creat eGFR Creatinine-Test Strip is designed with an electrode that measures creatinine levels. Creatinine in the capillary whole blood sample mixes with the reagents in the test strip to produce an electric current. The amount of current that is produced depends on how much creatinine is in the sample. A digital readout is displayed on the monitor in 30 seconds, the eGFR is calculated using patient information and the measured creatinine level. The test strips are for single use only.

Nova Max Creat eGFR Creatinine Control Solutions
The Nova Max Creat eGFR Creatinine Control Solution is used as a quality control check to ensure that the Nova Max Creat eGFR Monitor and the Nova Max Creat eGFR Creatinine-Test Strips are working properly as a system. There are two levels of Nova Max Creat eGFR Creatinine Control Solution. Each level of control solution has a known creatinine concentration that reacts with the reagents in the test strip.

AI/ML Overview

N/A

U.S. Food & Drug Administration FDA Clearance Letter

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.08.02

January 21, 2026

Nova Biomedical Corporation
Mariya Cesnulevicius
Regulatory Affairs Manager
200 Prospect St.
Waltham, MA 02454

Re: K251281
Trade/Device Name: Nova Max Creat eGFR Monitoring System
Regulation Number: 21 CFR 862.1225
Regulation Name: Creatinine Test System
Regulatory Class: Class II
Product Code: SHB
Dated: December 22, 2025
Received: December 22, 2025

Dear Mariya Cesnulevicius:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2

K251281 - Mariya Cesnulevicius Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

Page 3

K251281 - Mariya Cesnulevicius Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

PAULA V. CAPOSINO -S

Paula Caposino, Ph.D.
Deputy Director
Division of Chemistry and Toxicology Devices
OHT7: Office of In Vitro Diagnostics
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known): K251281

Device Name: Nova Max Creat eGFR Monitoring System

Indications for Use (Describe)

The Nova Max Creat eGFR Monitoring System is comprised of the Nova Max Creat eGFR Monitor and the Nova Max Creat eGFR Test Strips.

The Nova Max Creat eGFR Monitoring System is intended for in vitro diagnostic use for the quantitative measurement of creatinine and estimation of glomerular filtration rate (eGFR) in fresh capillary whole blood obtained from the fingertip of adult patients aged 18 and above. The system is intended for single patient home use by prescription only and should not be shared. It is intended for use by patients as an aid to monitor kidney function on the order of a treating healthcare professional.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF

Page 5

510(k) Summary

510(k) Owner: Nova Biomedical Corporation
Establishment Registration Number: 1219029
510(k) Number: K251281
Address: 200 Prospect Street
Waltham, MA, 02454 USA
Phone: 781-894-0800
Fax Number: 781-899-6259

Primary Contact Person
Mariya Cesnulevicius
Regulatory Affairs Manager

Secondary Contact Person
Robert Zinck,
Director of Regulatory and Clinical Affairs

Date Prepared
January 21, 2026

Trade Name
Nova Max Creat eGFR Monitoring System

Common Name
Creatinine Test System

Classification Name

Classification NameProduct CodeRegulation NumberClassification510(k) Review Panel
Creatinine test systemSHB862.1225Class IIClinical Chemistry

Predicate Device
K171059, Nova StatSensor Creatinine Hospital Meter System

Page 1 of 7

Page 6

Device Description

The Nova Max Creat eGFR Monitoring System consists of Nova Max Create eGFR Monitor and Nova Max Create eGFR Creatinine-Test Strips. The Nova Max Creat eGFR test strips, Nova Max Creat eGFR control solution (Level 1 and Level 2), and a single-use, disposable, 21 gauge safety lancet can be purchased separately.

Nova Max Creat eGFR Monitor
The monitor is intended to be used in conjunction with the Nova Max Creat eGFR Creatinine-Test Strips to measure creatinine and calculate estimated glomerular filtration rate (eGFR) using capillary whole blood obtained from the fingertip. The hand-held monitor is lightweight, portable, utilizes a 2.8" color touchscreen display and has capability for storage (400 test samples and/or quality control samples) and review of data. The monitor is powered by a rechargeable, 3.7 V Li polymer battery.

Nova Max Creat eGFR Creatinine-Test Strips
The Nova Max Creat eGFR Creatinine-Test Strip is designed with an electrode that measures creatinine levels. Creatinine in the capillary whole blood sample mixes with the reagents in the test strip to produce an electric current. The amount of current that is produced depends on how much creatinine is in the sample. A digital readout is displayed on the monitor in 30 seconds, the eGFR is calculated using patient information and the measured creatinine level. The test strips are for single use only.

Nova Max Creat eGFR Creatinine Control Solutions
The Nova Max Creat eGFR Creatinine Control Solution is used as a quality control check to ensure that the Nova Max Creat eGFR Monitor and the Nova Max Creat eGFR Creatinine-Test Strips are working properly as a system. There are two levels of Nova Max Creat eGFR Creatinine Control Solution. Each level of control solution has a known creatinine concentration that reacts with the reagents in the test strip.

Indications for Use

The Nova Max Creat eGFR Monitoring System is comprised of the Nova Max Creat eGFR Monitor and the Nova Max Creat eGFR Test Strips.

The Nova Max Creat eGFR Monitoring System is intended for in vitro diagnostic use for the quantitative measurement of creatinine and estimation of glomerular filtration rate (eGFR) in fresh capillary whole blood obtained from the fingertip of adult patients aged 18 and above. The system is intended for single patient home use by prescription only and should not be shared. It is intended for use by patients as an aid to monitor kidney function on the order of a treating healthcare professional.

Summary of Performance Testing

Analytical bench and clinical testing were conducted to demonstrate that the Nova Max Creat eGFR Monitoring System achieves its intended purpose at the intended use setting.

Linearity Testing
Testing was performed to validate the Nova Max Creat eGFR Monitoring System linearity. Ten (10) specimens were tested on the Nova Max Creat eGFR Monitoring System. The same specimens were analyzed on the reference method. The results met the acceptance criteria, validating the linearity of the Nova Max Creat eGFR Monitoring System across the measurement range of 0.30 mg/dL to 7.00 mg/dL.

Analytical Specificity
The analytical specificity for the Nova Max Creat eGFR Monitoring System was assessed by testing the effect of various exogenous and endogenous substances on the measurement of creatinine by the system. Whole blood samples were spiked with forty-six (46) potential interferents and the creatinine concentration was measured with the Nova Max Creat eGFR Monitoring System.

Page 2 of 7

Page 7

The test substance was considered to be an interfering substance if the absolute difference between the test value and the control value was greater than 10% for creatinine concentration higher than 2 mg/dL, and greater than ± 0.2 mg/dL if the creatinine concentration is less than 2 mg/dL.

Based upon the results, all potential interfering substances tested met the acceptance criteria and were demonstrated not to have clinical interference. No clinical interference was identified up to the test concentrations reported in Table 1.

Table 1, Interference Test Result Summary

#InterferentInterference concentration#InterferentInterference concentration
1Acetaldehyde0.2 mg/dL24Glucose1000 mg/dL
2Acetaminophen15.6 mg/dL25High Hct and Low Hct20 – 70 %
3N-Acetyl-L-Cysteine15 mg/dL26Ibuprofen22 mg/dL
4Ascorbic Acid5.25 mg/dL27Paracetamol-4-acetamidopenol30 mg/dL
5β-Hydroxybutyric Acid65 mg/dL28Pyruvate5 mg/dL
6Potassium bicarbonate294 mg/dL29Lithium Salicylate3 mg/dL
7Bilirubin (unconjugated)40 mg/dL30Potassium thiocyanate6 mg/dL
8Calcium Chloride20 mg/dL31Urea120 mg/dL
9Creatine5 mg/dL32Farxiga (dapagliflozin)0.6 mg/dL
10Dopamine Hydrochloride2 mg/dL33Kerendia (finerenone)1.2 mg/dL
11Formaldehyde0.4 mg/dL34Jardiance (empagliflozin)1.5 mg/dL
12Glycolic Acid76 mg/dL35Lithium Heparin1.2 mg/dL
13Hydroxyurea4 mg/dL36Lithium Bromide325.69 mg/dL
14Lithium Lactate90 mg/dL37Hemoglobin2600 mg/dL
15Methyldopa5 mg/dL38Cholesterol1000 mg/dL
16pH8.039L-Dopa0.3 mg/dL
17Sodium Thiosulfate265 mg/dL40Maltose360 mg/dL
18Triglyceride1500 mg/dL41Intralipid2000 mg/dL
19Uric Acid23.5 mg/dL42Lactic Acid108.1 mg/dL
20Acetoacetate20 mg/dL43Salicylic Acid60 mg/dL
21Acetylsalicylic Acid3 mg/dL44Semaglutide (Ozempic or Wegovy)0.144 mg/dL
22Chlorpromazine HCl0.4 mg/dL45Tirzepatide (Mounjaro)0.9 mg/dL
23Ethanol600 mg/dL46Bilirubin (conjugated)40 mg/dL

Limits of Detection and Quantification
Testing was conducted to assess the analytical sensitivity of the Nova Max Creat eGFR Monitoring System. Blank whole blood samples were tested to estimate the limit of the blank (LoB). This value was used to select a low level of creatinine concentration to determine the LoD and LoQ. Low level samples were prepared with creatinine concentration close to the estimated low value for creatinine concentration. These samples were tested repeatedly on the Nova Max Creat eGFR Monitoring System. Samples were also tested on the reference analyzer.

The study yielded the following results for creatinine:
Limit of Blank (LoB) = 0.078 mg/dL
Limit of Detection (LoD) = 0.174 mg/dL
Limit of Quantification (LoQ) = 0.174 mg/dL

Page 3 of 7

Page 8

Traceability
Nova Max Creat eGFR Creatinine Control Materials are traceable to NIST Standard Reference Material 967b

Precision
Total Imprecision of the Nova Max Creat eGFR Monitoring System was assessed using methods described in CLSI "Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guidelines – Second Edition", CLSI EP5-A3 as guidance.

Day-to-Day Imprecision
To investigate day-to-day precision five (5) levels of Nova Max Creatinine Aqueous Solutions were used. Two (2) Nova Max Creat eGFR Monitors and three (3) Nova Max Creat eGFR Creatinine-Test Strip lots were tested in the study.

Each level of the aqueous solutions was analyzed in duplicates, twice a day over twenty (20) days. The results (presented in Table 2) met the acceptance criteria indicating that the day-to-day precision of the Creatinine assay on the Nova Max Creat eGFR Monitoring Systems is clinically acceptable.

Table 2, Day-to-Day Imprecision Summary

Aqueous SolutionTest Strip Lot 1Test Strip Lot 2Test Strip Lot 3
Level 1 Mean, mg/dL0.560.570.58
SD0.060.060.06
Level 2 Mean, mg/dL1.331.351.37
SD0.060.080.08
Level 3 Mean, mg/dL2.582.652.66
CV (%)5.004.644.93
Level 4 Mean, mg/dL4.134.144.09
CV (%)4.894.984.98
Level 5 Mean, mg/dL6.035.976.04
CV (%)4.854.994.87

Within-run Imprecision
To investigate the precision using whole blood, five (5) blood specimens with creatine concentrations spanning the measurement range were tested. Five (5) Nova Max Creat eGFR Monitors and three (3) Nova Max Creat eGFR Creatinine-Test Strip lots were used in the study. The results (presented in Table 3) met the acceptance criteria, indicating that the Nova Max Creat eGFR System report clinically acceptable precision.

Table 3, Within-run Imprecision Summary

Whole Blood SpecimensTest Strip Lot 1Test Strip Lot 2Test Strip Lot 3
Level 1 Mean, mg/dL0.470.470.49
SD0.050.060.06
Level 2 Mean, mg/dL0.820.830.83
SD0.060.060.06
Level 3 Mean, mg/dL1.681.681.66
SD0.090.100.10
Level 4 Mean, mg/dL2.812.882.83
CV (%)4.6%4.4%4.8%

Page 4 of 7

Page 9

| Level 5 Mean, mg/dL | 5.83 | 5.88 | 5.97 |
| CV (%) | 4.7% | 5.0% | 4.9% |

Flex Studies
Nova Biomedical conducted the following flex studies to validate the robustness of the Nova Max Creat eGFR Monitoring System:

  1. Short sample
  2. Used Nova Max Creat eGFR Creatinine-Test Strips
  3. Incorrect test strips
  4. Moving during analysis
  5. Atypical positions
  6. Removing Nova Max Creat eGFR Creatinine-Test Strips during the analysis
  7. Double dosing
  8. Delayed analysis
  9. Flicking
  10. Damaged Strip
  11. Incorrect Nova Max Creat eGFR Creatinine Control Solutions storage
  12. Uncapped Nova Max Creat eGFR Creatinine Control Solutions
  13. Extreme Environmental Conditions
  14. Improper sampling

The results of the Flex Studies were determined to meet the Acceptance Criteria and error codes were correctly generated (where applicable) as specified in the Instructions for Use Manual.

Clinical Performance
Nova Biomedical conducted a clinical study in the hands of lay users. The clinical study was conducted across two (2) sites with a total of 532 subjects. Measurements on fingerstick samples taken by lay users using the Nova Max Creat eGFR Monitoring System were compared to venous plasma creatinine results measured by laboratory operators using an IDMS traceable comparative method.

The Pearson correlation coefficient (r) between the fingerstick blood readings obtained by lay users and the corresponding comparative method readings obtained by laboratory operators was calculated using least squares linear regression and is displayed in Table 4.

Table 4, Creatinine Linear Regression Analysis Summary

| Slope: | 0.9832 |
| Y-Intercept: | 0.0073 |
| Correlation Coefficient (r): | 0.9951 |

The bias between the capillary whole blood creatinine readings from the Nova Max Creat eGFR Monitoring System and the venous plasma creatinine readings measured by the comparative method was evaluated and shown in Table 5.

Table 5, Within Clinical Study Accuracy (Bias/Bias%)

Less than 2 mg/dLGreater than 2 mg/dL
SD ≤ 0.2 mg/dLCV ≤ 10%
97.87%95.36%

Page 5 of 7

Page 10

The results showed that the Nova Max system's measurements were in strong agreement with the comparative method, meeting the clinical accuracy acceptance criteria in the hands of lay users.

Table 6: Comparison of Predicate and Proposed Devices

Predicate Device: Nova StatSensor Creatinine Hospital Meter SystemSubject Device: Nova Max Creat GFR Monitoring System
Intended Use and Indications for UseThe StatSensor Creatinine Meter is intended for in vitro diagnostic use by health care professionals and for Point-Of-Care usage for the quantitative measurement of creatinine in capillary, venous, and arterial whole blood. Creatinine measurements are used in the diagnosis and treatment of renal diseases and in monitoring renal dialysis. Not for use in neonates.The Nova Max Creat eGFR Monitoring System is comprised of the Nova Max Creat eGFR Monitor and the Nova Max Creat eGFR Test Strips. The Nova Max Creat eGFR Monitoring System is intended for in vitro diagnostic use for the quantitative measurement of creatinine and estimation of glomerular filtration rate (eGFR) in fresh capillary whole blood obtained from the fingertip of adult patients aged 18 and above. The system is intended for single patient home use by prescription only and should not be shared. It is intended for use by patients as an aid to monitor kidney function on the order of a treating healthcare professional.
Intended Use EnvironmentPrescription use, point-of-care settingsHome use by prescription
Hand-held meterYesSame
Test Measured/Reported ParametersCreatinine and eGFRCreatinine and eGFR
Operating PrincipleEnzyme, AmperometrySame
eGFR equationsCKD-EPI (2009), MDRD and Cockcroft-Gault equationCKD-EPI 2021 equation
Sample TypeCapillary, Venous and Arterial whole blood specimenFingertip capillary whole blood specimen
Sample Volume1.2 µLSame
Sample ApplicationTest strip capillary drawSame
Measurement Range for Creatinine0.3 - 12.0 mg/dL (1 significant digit)0.30 - 7.00 mg/dL (2 significant digits)
Test Range for eGFR calculated using the CKD-EPI equation15 - 90 mL/min/1.73 m²Same
Analysis Time30 secondsSame
Hematocrit Range30% to 60%20% to 70%
Quality ControlThree LevelsTwo Levels
LinearityFive LevelsNone
CalibrationAutomatic, No Calibration CodeSame

Page 6 of 7

Page 11

Predicate Device: Nova StatSensor Creatinine Hospital Meter SystemSubject Device: Nova Max Creat GFR Monitoring System
Location of the test strip portBottom of the meterSame
Physical dimensions of the meter153 mm (6.0 in) x 82.5 mm (3.25 in) x 46 mm (1.8 in)95.25 mm (3.75 in) x 61.98 mm (2.44 in) x 18.80 mm (0.74 in)
Weight360 grams90 grams
Menu navigationColor touch screen displaySame
Power/Energy Source3.7 V Li-Polymer battery (Rechargeable/Replaceable)3.7 V Li-Polymer battery (Rechargeable/Non-Replaceable)
Battery ChargingCharged via a Desk-mount Docking/Charging StationCharged like a cell phone, using an external power supply connected to an electric outlet, and a charging cable connecting the power supply to the monitor.
Meter Memory1000 patient results, 200 QC results, 4000 Operators400 Patient and/or QC Results
Data TransferVia Docking Station Ethernet ConnectivityNo data transfer is available
Operating Temperature Range59°F to 104°F (15°C to 40°C)Same
Operating Humidity Range10% to 90% relative humiditySame
Operating Altitude RangeUp to 15,000 feet (4572 meters)Up to 12,000 feet (3658 meters)
Test Strip Storage Temperature (Closed Vial)35.6°F to 46.4°F (2°C to 8°C)Same
Test Strip Shelf Life24 monthsSame
Control Solutions Storage Temperature (Closed Vial)35.6°F to 46.4°F (2°C to 8°C)Same
Control Solutions Shelf Life24 monthsSame

Conclusion

Based on the performance data provided with this submission, it can be concluded that the Nova Max Creat eGFR Monitoring System has been demonstrated to be safe, and effective for its intended use for monitoring of kidney health in the intended use setting. The similarities in the intended use and technological characteristics between the proposed Nova Max Creat eGFR Monitoring System and the FDA cleared predicate device, StatSensor Creatinine Hospital Meter System, support that the two devices are substantially equivalent.

Page 7 of 7

§ 862.1225 Creatinine test system.

(a)
Identification. A creatinine test system is a device intended to measure creatinine levels in plasma and urine. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.(b)
Classification. Class II.