Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K202453

    Validate with FDA (Live)

    Date Cleared
    2021-03-24

    (209 days)

    Regulation Number
    862.3100
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K182654, K181968, K153646

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    SAFECARE® Multi-Drug Urine Test Dip Card is competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Oxazepam, Methamphetamine, Morphine, Oxycodone, Secobarbital, Buprenorphine, Methylenedioxy-methamphetamine, Phencyclidine, Methadone, Nortriptyline d-Propoxyphene and 2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine (EDDP) in human urine at the cutoff concentrations of:

    Drug(Identifier)Cut-off level
    Amphetamine500ng/mL
    Oxazepam300 ng/mL
    Cocaine150ng/mL
    Marijuana50 ng/mL
    Methamphetamine500ng/mL
    Morphine300ng/mL
    Oxycodone100 ng/mL
    Secobarbital300 ng/mL
    Buprenorphine10 ng/mL
    Methylenedioxy-methamphetamine500 ng/mL
    Phencyclidine25 ng/mL
    Methadone300 ng/mL
    Nortriptyline1000 ng/mL
    d-Propoxyphene300 ng/mL
    2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine300 ng/mL

    Configuration of SAFECARE® Multi-Drug Urine Test Dip Card can consist of any combination of the above listed drug analytes.

    The test may yield positive results for the prescription drugs Buprenorphine, Nortriptyline, Oxazepam, Secobarbital, Propoxyphene and Oxycodone when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. The test provides only preliminary test results. A more specific alternative chemical must be used in order to obtain a confirmed analytical result. GC/MS or LC/MS is the preferred confirmatory method.

    The tests are intended for over-the-counter use.

    SAFECARE® Multi-Drug Urine Test Cup is competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine. Oxazepam. Marijuana. Methamphetamine, Morphine. Oxycodone, Secobarbital, Buprenorphine, Methylenedioxy-methamphetamine, Phencyclidine, Methadone, Nortriptyline d-Propoxyphene and 2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine (EDDP) in human urine at the cutoff concentrations of:

    Drug(Identifier)Cut-off level
    Amphetamine500ng/mL
    Oxazepam300 ng/mL
    Cocaine150ng/mL
    Marijuana50 ng/mL
    Methamphetamine500ng/mL
    Morphine300ng/mL
    Oxycodone100 ng/mL
    Secobarbital300 ng/mL
    Buprenorphine10 ng/mL
    Methylenedioxy-methamphetamine500 ng/mL
    Phencyclidine25 ng/mL
    Methadone300 ng/mL
    Nortriptyline1000 ng/mL
    d-Propoxyphene300 ng/mL
    2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine300 ng/mL

    Configuration of SAFECARE® Multi-Drug Urine Test Cup can consist of any combination of the above listed drug analytes.

    The test may yield positive results for the prescription drugs Buprenorphine, Nortriptyline, Oxazepam, Secobarbital, Propoxyphene and Oxycodone when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. The test provides only preliminary test results. A more specific alternative chemical must be used in order to obtain a confirmed analytical result. GC/MS or LC/MS is the preferred confirmatory method.

    The tests are intended for over-the-counter use.

    Device Description

    The SAFECARE® Dip Card Tests and SAFECARE® Cup Tests are immunochromatographic assays that use a lateral flow system for the qualitative detection of Amphetamine, Oxazepam, Cocaine, Cannabinoids, Methamphetamine, Morphine, Secobarbital, Methadone, Methylenedioxymethamphetamine, Oxycodone, Buprenorphine, Phencyclidine, Nortriptyline, Propoxyphen and 2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine (target analytes) in human urine. The products are single-use in vitro diagnostic devices, which come in the formats of Dip Cards or Cups. Each test kit contains a Test Device (in one of the two formats), a package insert and a urine cup for sample collection. Each test device is sealed with a desiccant in an aluminum pouch.

    AI/ML Overview

    The provided document describes the performance characteristics and studies for the SAFECARE® Multi-Drug Urine Test Dip Card and SAFECARE® Multi-Drug Urine Test Cup. It does not describe an AI/ML device but rather an in-vitro diagnostic device (IVD) for drug screening. Therefore, several of the requested categories for AI/ML device evaluation are not applicable (e.g., number of experts, adjudication method, MRMC study, standalone performance, training set).

    Here's the information extracted from the document, tailored to the nature of the device:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for qualitative drug tests are typically defined by precision around the cutoff concentration. The device is expected to consistently classify samples below the cutoff as negative and above the cutoff as positive. For samples near the cutoff, some variability in classification is expected.

    Test Parameter / Drug (Cut-off)Acceptance Criteria (Implicit from study design)Reported Device Performance (Precision Study - Example for Dip Card, Lot 1)
    PrecisionSamples < -25% Cut-off: 100% negative calls. Samples > +25% Cut-off: 100% positive calls. Samples within +/-25% of Cut-off: Expected variability.Amphetamine 500: -100% Cut-off: 50-/0+ (100% negative) -75% Cut-off: 50-/0+ (100% negative) -50% Cut-off: 50-/0+ (100% negative) -25% Cut-off: 50-/0+ (100% negative) Cut-off: 24-/26+ (48% negative, 52% positive) +25% Cut-off: 50+/0- (100% positive) +50% Cut-off: 50+/0- (100% positive) +75% Cut-off: 50+/0- (100% positive) +100% Cut-off: 50+/0- (100% positive) Cocaine 150: -100% Cut-off: 50-/0+ (100% negative) -75% Cut-off: 50-/0+ (100% negative) -50% Cut-off: 50-/0+ (100% negative) -25% Cut-off: 50-/0+ (100% negative) Cut-off: 24-/26+ (48% negative, 52% positive) +25% Cut-off: 50+/0- (100% positive) +50% Cut-off: 50+/0- (100% positive) +75% Cut-off: 50+/0- (100% positive) +100% Cut-off: 50+/0- (100% positive) Methamphetamine 500: -100% Cut-off: 50-/0+ (100% negative) -75% Cut-off: 50-/0+ (100% negative) -50% Cut-off: 50-/0+ (100% negative) -25% Cut-off: 50-/0+ (100% negative) Cut-off: 24-/26+ (48% negative, 52% positive) +25% Cut-off: 50+/0- (100% positive) +50% Cut-off: 50+/0- (100% positive) +75% Cut-off: 50+/0- (100% positive) +100% Cut-off: 50+/0- (100% positive) Morphine 300: -100% Cut-off: 50-/0+ (100% negative) -75% Cut-off: 50-/0+ (100% negative) -50% Cut-off: 50-/0+ (100% negative) -25% Cut-off: 50-/0+ (100% negative) Cut-off: 24-/26+ (48% negative, 52% positive) +25% Cut-off: 50+/0- (100% positive) +50% Cut-off: 50+/0- (100% positive) +75% Cut-off: 50+/0- (100% positive) +100% Cut-off: 50+/0- (100% positive)
    Lay-User Study (Accuracy)For samples far from cutoff (e.g., -50% and +50%), approximately 100% correct results. For samples near cutoff (e.g., -25% and +25%), high accuracy (e.g., >90%).AMP500: -50% Cutoff: 100% correct (0 Pos, 170 Neg)+50% Cutoff: 100% correct (40 Pos, 0 Neg)-25% Cutoff: 90% correct (2 Pos, 18 Neg)+25% Cutoff: 95% correct (19 Pos, 1 Neg)(Similar results reported for COC150, THC, BAR, BZO, MET500, MTD, MOP300, MDMA, OXY, BUP, PCP, TCA, PPX, EDDP)

    2. Sample sizes used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Precision Study (Test Set):

      • For each drug (Amphetamine, Cocaine, Methamphetamine, Morphine), 9 concentrations were tested around the cutoff (-100%, -75%, -50%, -25%, Cut-off, +25%, +50%, +75%, +100%).
      • For each concentration, tests were performed two runs per day for 25 days, using 3 different lots of the device. This equates to 50 replicates per concentration per lot, totaling 450 tests per lot per drug.
      • Total replicates for 4 drugs (AMP, COC, MET, MOP): 9 concentrations x 50 replicates/concentration x 3 lots = 1350 tests per drug. (Some data for other drugs were referenced from prior 510(k)s: K182654, K181968, K153646, K201120).
      • Data Provenance: The document states "in-house" for comparison studies and "urine samples were prepared by spiking drug in negative samples" for precision studies. This suggests a controlled laboratory setting (likely prospective, artificial samples). The document does not specify the country of origin of the data.
    • Method Comparison Study (Clinical/Test Set):

      • 80 "unaltered clinical samples" were used for each drug. These samples were split into categories: 10 negative, 10 low negative, 20 near cutoff negative, 20 near cutoff positive, 20 high positive for each drug.
      • Total samples per device type (Dip Card or Cup) for 4 drugs mentioned: 80 samples x 4 drugs = 320 samples per device type.
      • Data Provenance: "in-house" and "unaltered clinical samples," implying real-world samples but processed within the manufacturer's lab. The document does not specify the country of origin or whether these clinical samples were retrospective or prospectively collected for the study.
    • Lay-User Study (Test Set):

      • 310 lay persons participated for each device format (Dip Card and Cup).
      • Urine samples were prepared at 7 concentrations: negative, +/-75%, +/-50%, +/-25% of the cutoff.
      • Total samples: For each drug, the number of samples varied across concentrations. For example, for AMP500, 20 samples at -100% cutoff, 20 at -75%, 170 at -50%, 20 at -25%, 20 at +25%, 40 at +50%, 20 at +75%. This totals 310 samples per drug per device format.
      • Data Provenance: "at three intended user sites." Samples were "prepared by spiking drugs into drug free-pooled urine specimens," making them artificial but intended to mimic a range of concentrations. This suggests a prospective study design, mimicking real-world use conditions but with controlled samples.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Precision Study: Ground truth was established by preparing urine samples with known drug concentrations confirmed by LC/MS. No human experts were involved in establishing ground truth, as it was an analytical study.
    • Method Comparison Study: The ground truth for clinical samples was established by LC/MS results. No mention of human experts for ground truth.
    • Lay-User Study: Ground truth was established by LC/MS results of the prepared spiked urine samples.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. This device is a qualitative diagnostic test read directly by users, not an AI/ML imaging device requiring expert adjudication. In the method comparison study, three laboratory assistants "ran" the samples, implying they performed the test, but the ground truth was LC/MS, not their consensus.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This is not an AI/ML device, and no MRMC study comparing human readers with and without AI assistance was mentioned. The lay-user study evaluated the device's performance with lay users, not an "AI assistance" scenario.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. The device is a physical, lateral flow immunochromatographic assay. Its performance inherently involves a human interpreting the result line, even if it's a simple positive/negative visual interpretation. It is not an algorithm-only device. The precision and method comparison studies evaluate the device's analytical performance, which is its inherent "standalone" capability.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • LC/MS (Liquid Chromatography-Mass Spectrometry) was used as the ground truth method to confirm drug concentrations in both spiked samples (for precision and lay-user studies) and clinical samples (for method comparison studies). This is a highly accurate and quantitative analytical method.

    8. The sample size for the training set

    • Not applicable. This is not an AI/ML device that requires a "training set" in the machine learning sense. The device is based on immunoassay principles.

    9. How the ground truth for the training set was established

    • Not applicable, as there is no training set for an AI/ML device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K201120

    Validate with FDA (Live)

    Date Cleared
    2020-05-27

    (30 days)

    Regulation Number
    862.3100
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K181968, K153646

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    SAFECARE® Multi-Drug Urine Test Dip Card is competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Oxazepam, Methamphetamine, Morphine, Oxycodone, Secobarbital, Buprenorphine, Methylenedioxy-methamphetamine, Phencyclidine, Methadone, Nortriptyline d-Propoxyphene and 2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine (EDDP) in human urine at the cutoff concentrations listed. The test may yield positive results for the prescription drugs Buprenorphine, Oxazepam, Secobarbital, Propoxyphene and Oxycodone when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. The test provides only preliminary test results. A more specific alternative chemical must be used in order to obtain a confirmed analytical result. GC/MS or LC/MS is the preferred confirmatory method. The tests are intended for over-the-counter use.

    SAFECARE® Multi-Drug Urine Test Cup is competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Oxazepam, Marijuana, Methamphetamine, Morphine, Oxycodone, Secobarbital, Buprenorphine, Methylenedioxy-methamphetamine, Phencyclidine, Methadone, Nortriptyline d-Propoxyphene and 2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine (EDDP) in human urine at the cutoff concentrations listed. The test may yield positive results for the prescription drugs Buprenorphine, Nortriptyline, Oxazepam, Secobarbital, Propoxyphene and Oxycodone when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS or LC/MS is the preferred confirmatory method. The tests are intended for over-the-counter use.

    Device Description

    The SAFECARE® Dip Card Tests and SAFECARE® Cup Tests are immunochromatographic assays that use a lateral flow system for the qualitative detection of Amphetamine, Oxazepam, Cocaine, Cannabinoids, Methamphetamine, Morphine, Secobarbital, Methadone, Methylenedioxymethamphetamine, Oxycodone, Buprenorphine, Phencyclidine, Nortriptyline, Propoxyphen and 2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine (target analytes) in human urine. The products are single-use in vitro diagnostic devices, which come in the formats of Dip Cards or Cups. Each test kit contains a Test Device (in one of the two formats), a package insert and a urine cup for sample collection. Each test device is sealed with a desiccant in an aluminum pouch.

    AI/ML Overview

    The provided text describes the performance characteristics of the SAFECARE® Multi-Drug Urine Test Dip Card and SAFECARE® Multi-Drug Urine Test Cup. The document details analytical performance (precision, linearity, stability, interference, specificity, effect of specific gravity and pH) and comparison studies (method comparison and lay-user study).

    Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for each analytical performance characteristic are implicitly defined by the successful results reported in the study. The study aims to demonstrate that the device performs as expected according to its intended use and analytical specifications.

    • Precision:

      • Acceptance Criteria for -100% to -25% Cut-off: All samples should test negative.
      • Acceptance Criteria for +25% to +100% Cut-off: All samples should test positive.
      • Acceptance Criteria for Cut-off (nominal concentration): Approximately 50% positive and 50% negative results (reflecting the nature of the cut-off).
      • Reported Device Performance (for 2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine as an example):
        • Dip Card (Lot 1, 2, 3):
          • -100% Cut-off to -25% Cut-off: 50-/0+ (All negative)
          • +25% Cut-off to +100% Cut-off: 50+/0- (All positive)
          • Cut-off: 24-/26+, 25-/25+, 26-/24+ (Split between positive and negative, as expected)
        • Cup (Lot 1, 2, 3):
          • -100% Cut-off to -25% Cut-off: 50-/0+ (All negative)
          • +25% Cut-off to +100% Cut-off: 50+/0- (All positive)
          • Cut-off: 27-/23+, 24-/26+, 26-/24+ (Split between positive and negative, as expected)
    • Interference:

      • Acceptance Criteria: No interference from common physiological or pathological substances at specified concentrations.
      • Reported Device Performance: No interference observed for a comprehensive list of compounds at 100ug/mL (except albumin at 100mg/dL and ethanol at 1% volume).
    • Specificity (Cross-Reactivity):

      • Acceptance Criteria: Related compounds should either not cross-react or cross-react at concentrations significantly higher than the cut-off, as defined by medical relevance.
      • Reported Device Performance (for 2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine as an example):
        • Methadone: Positive at 300000 ng/mL (0.1% cross-reactivity)
        • EMDP: Positive at 300000 ng/mL (0.1% cross-reactivity)
        • Doxylamine, Disopyramide, LAAM HCl, Alpha Methadol: Positive at >100,000 ng/mL (<0.3% cross-reactivity)
    • Effect of Urine Specific Gravity and Urine pH:

      • Acceptance Criteria: Reliable results (positive above cut-off, negative below cut-off) across a specified range of specific gravity and pH.
      • Reported Device Performance: All samples at and above +25% Cut-Off were positive, and all samples at and below -25% Cut-Off were negative, for specific gravity from 1.000 to 1.035 and pH 4 to 9.
    • Method Comparison Study (Clinical Samples):

      • Acceptance Criteria: High agreement (minimal discordant results) between the device results and LC/MS (Liquid Chromatography/Mass Spectrometry) results, especially for samples near the cut-off.
      • Reported Device Performance (for 2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine as an example):
        • Dip Card: High concordance with LC/MS, with few discordant results (e.g., Viewer A: 1 false positive at 291 ng/mL; Viewer B: 1 false negative at 309 ng/mL).
        • Cup: High concordance with LC/MS, with few discordant results (e.g., Viewer A: 1 false positive at 285 ng/mL; Viewer B: 1 false negative at 372 ng/mL).
    • Lay-user Study:

      • Acceptance Criteria: High percentage of correct results for samples well below and well above the cut-off, and reasonable performance for samples near the cut-off, when interpreted by lay users. Ease of understanding instructions.
      • Reported Device Performance (for specific drugs like Amphetamine (AMP), Cocaine (COC), THC, BAR, BZO, MET, MTD, MDMA, OXY, OPI, BUP, PCP, TCA, PPX, EDDP as examples):
        • For samples at -100% to -50% Cut-off and +50% to +75% Cut-off: 100% correct results across all drugs.
        • For samples at -25% Cut-off: generally 90-95% correct negative results (some false positives).
        • For samples at +25% Cut-off: generally 85-95% correct positive results (some false negatives).
        • All lay users indicated the device instructions were easy to follow (Flesch-Kincaid Grade Level 7).

    2. Sample Sizes Used for the Test Set and the Data Provenance

    • Precision Study: For each drug and concentration, two runs per day for 25 days per device (Dip Card and Cup), implying a total of 50 tests per concentration per lot. With 3 lots, this is 150 tests per concentration per format. There were 9 concentrations tested. So 150*9 = 1350 tests per drug per format.
    • Interference Study: Not explicitly stated, but tests were performed on "three batches of each device" for various interfering substances spiked into drug-free and drug-spiked urine samples.
    • Specificity Study: Not explicitly stated how many tests per compound, but "three batches of each device" were used.
    • Effect of Urine Specific Gravity and Urine pH Study: "three lots of each device" were used for samples spiked with target drugs at +/- 25% Cut-Off levels across the range of specific gravity and pH.
    • Method Comparison Study: 80 "unaltered clinical samples" for each drug (40 negative and 40 positive relative to cut-off). This was conducted "in-house." The provenance of the clinical samples is not specified beyond being "clinical samples."
    • Lay-user Study: 310 lay persons for each device format. The samples were prepared by spiking drugs into "drug free-pooled urine specimens." The concentrations were confirmed by LC/MS. Thus, the samples were prepared and not naturally occurring clinical samples.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    • For the Precision, Interference, Specificity, and Effect of Urine Specific Gravity and Urine pH Studies: The ground truth was established by precise laboratory preparation of samples where drug concentrations were known and confirmed by LC/MS. There were no human experts classifying these samples for ground truth.
    • For the Method Comparison Study: The ground truth was established by LC/MS results. LC/MS is a highly accurate and definitive analytical method. The text does not mention human experts establishing ground truth for these samples; it was based on the objective chemical analysis. The "three laboratory assistants" were operators performing the device tests, not establishing ground truth.
    • For the Lay-user Study: The ground truth was the known drug concentration of the prepared urine samples, confirmed by LC/MS.

    4. Adjudication Method for the Test Set

    • Precision, Interference, Specificity, and Effect of Urine Specific Gravity and Urine pH Studies: No human adjudication was mentioned as the ground truth was based on precisely prepared samples with known concentrations.
    • Method Comparison Study: The device results were compared directly to LC/MS results. There was no adjudication mentioned involving human experts to reconcile discrepancies between device and LC/MS. Discrepancies were simply reported as "discordant results."
    • Lay-user Study: Comparisons were made against the known LC/MS confirmed concentrations of the prepared samples. No human adjudication was mentioned.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No MRMC comparative effectiveness study was done. This device is a rapid diagnostic test (lateral flow immunoassay) for drug detection, where the "reading" is a visual interpretation of lines by a user (or laboratory assistant in the method comparison study). It does not appear to involve AI or sophisticated image analysis where human readers would be "assisted" by AI. The "lay-user study" involved multiple readers (310 lay persons), but it was designed to assess the ease of use and accuracy of the device itself by untrained individuals, not to compare human performance with and without AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    • This device is a lateral flow immunoassay intended for visual interpretation. It does not have an "algorithm only" component in the sense of a software-based AI system. Its performance inherently involves a "human-in-the-loop" to view and interpret the test lines.

    7. The Type of Ground Truth Used

    • The primary ground truth used throughout the studies (precision, interference, specificity, method comparison, and lay-user study) was LC/MS (Liquid Chromatography/Mass Spectrometry), which is a definitive analytical method for confirming drug concentrations in urine samples. For analytical studies, ground truth was also established by careful preparation of samples with known drug concentrations.

    8. The Sample Size for the Training Set

    • This document describes performance studies for an immunochromatographic assay, which is a chemical/biological test, not a machine learning or AI model. Therefore, there is no "training set" in the context of AI model development. The assays are "trained" during their manufacturing and chemical formulation processes to react at specific cut-off levels.

    9. How the Ground Truth for the Training Set Was Established

    • As stated above, this device is not an AI/ML product and thus does not have a "training set" with ground truth in that context. The "training" of the assay refers to its inherent chemical design to react to certain drug concentrations, which relies on established analytical chemistry principles and known drug properties.
    Ask a Question

    Ask a specific question about this device

    K Number
    K182654

    Validate with FDA (Live)

    Date Cleared
    2018-11-19

    (55 days)

    Regulation Number
    862.3610
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K181968, K153646

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    SAFECARE® Multi-Drug Urine Test Dip Card is competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Oxazepam, Marijuana, Methamphetamine, Morphine, Oxycodone, Secobarbital, Buprenorphine, Methylenedioxymethamine, Phencyclidine, Methadone, Nortriotyline and d-Propoxyphene in human urine at the cutoff concentrations of:

    Drug(Identifier)Cut-off level
    Amphetamine1000 ng/mL
    Oxazepam300 ng/mL
    Cocaine300 ng/mL
    Marijuana50 ng/mL
    Methamphetamine1000 ng/mL
    Morphine2000 ng/mL
    Oxycodone100 ng/mL
    Secobarbital300 ng/mL
    Buprenorphine10 ng/mL
    Methylenedioxymethamphetamine500 ng/mL
    Phencyclidine25 ng/mL
    Methadone300 ng/mL
    Nortriptyline1000 ng/mL
    d-Propoxyphene300 ng/mL

    Configuration of SAFECARE® Multi-Drug Urine Test Dip Card can consist of any combination of the above listed drug analytes.

    The test may yield positive results for the prescription drugs Buprenorphine. Oxazepam, Secobarbital. Propoxyphene and Oxycodone when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. The test provides only preliminary test results. A more specific alternative chemical must be used in order to obtain a confirmed analytical result. GCMS or LC/MS is the preferred confirmatory method.

    The tests are intended for over-the-counter use.

    SAFECARE® Multi-Drug Urine Test Cup is competitive binding, lateral flow immunichromatographic assays for qualitative and simultaneous detection of Amphetanine, Oxazepam, Methamphetamine, Morghine, Oxycodone, Secobarbital, Buprenorphine, Methylenedioxymethamphetamine, Phencyclidine, Methadone, Nortriptyline and d-Propoxyphene in human urine at the cutoff concentrations of:

    Drug(Identifier)Cut-off level
    Amphetamine1000 ng/mL
    Oxazepam300 ng/mL
    Cocaine300 ng/mL
    Marijuana50 ng/mL
    Methamphetamine1000 ng/mL
    Morphine2000 ng/mL
    Oxycodone100 ng/mL
    Secobarbital300 ng/mL
    Buprenorphine10 ng/mL
    Methylenedioxymethamphetamine500 ng/mL
    Phencyclidine25 ng/mL
    Methadone300 ng/mL
    Nortriptyline1000 ng/mL
    d-Propoxyphene300 ng/mL

    Configuration of SAFECARE® Multi-Drug Urine Test Cup contination of the above listed drug analytes.

    The test may yield positive results for the prescription drugs Buprenorphine. Oxazenam, Secobarbital, Propoxyphene and Oxycodone when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GCMS or LC/MS is the preferred confirmatory method.

    The tests are intended for over-the-counter use.

    Device Description

    The SAFECARE Dip Card Tests and SAFECARE Cup Tests are immunochromatographic assays that use a lateral flow system for the qualitative detection of Amphetamine, Oxazepam, Cocaine, Cannabinoids, Methamphetamine, Morphine, Secobarbital, Methadone, Methylenedioxymethamphetamine, Oxycodone, Buprenorphine, Phencyclidine, Nortriptyline and Propoxyphene (target analytes) in human urine. The products are single-use in vitro diagnostic devices, which come in the formats of Dip Cards or Cups. Each test kit contains a Test Device (in one of the two formats), a package insert and a urine cup for sample collection. Each test device is sealed with a desiccant in an aluminum pouch.

    AI/ML Overview

    The document describes the performance characteristics and studies for the SAFECARE Multi-Drug Urine Test Dip Card and SAFECARE Multi-Drug Urine Test Cup.

    Here's an analysis of the acceptance criteria and study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state "acceptance criteria" in a separate section. However, the performance data presented implies a very high standard for accuracy, particularly around the established cutoff concentrations. The "precision" studies demonstrate the device's ability to consistently produce correct results for samples at varying concentrations relative to the cutoff. The "comparison studies" compare the device's qualitative results to a quantitative gold standard (LC/MS) to establish concordance. The "lay-user study" assesses the ability of untrained individuals to correctly use and interpret the device.

    Based on the provided data, the implied acceptance criteria for qualitative immunoassay devices for drug testing would typically be:

    • 100% agreement for samples significantly below the cutoff (Negative).
    • 100% agreement for samples significantly above the cutoff (Positive).
    • High percentage agreement for samples near the cutoff (typically >90%).

    For the Precision Study (Analytical Performance - "Test Set" Data):

    The table below summarizes the reported performance for Methylenedioxymethamphetamine (MDMA), Oxycodone, Buprenorphine, Phencyclidine, Nortriptyline, and Propoxyphene (results from 3 lots, 50 tests per concentration per lot, per device type). The format is (Number of Negative Results)/(Number of Positive Results).

    Drug (Identifier)Cut-off LevelConcentrationReported Performance (Typical across 3 lots)Implied Acceptance Criteria Met?
    Methylenedioxymethamphetamine (MDMA) (Dip Card & Cup)500 ng/mL-100% Cut-off50-/0+Yes (100% negative)
    -75% Cut-off50-/0+Yes (100% negative)
    -50% Cut-off50-/0+Yes (100% negative)
    -25% Cut-off50-/0+Yes (100% negative)
    Cut-off25-/25+ or 26-/24+Yes (near 50/50 split at cutoff)
    +25% Cut-off50+/0-Yes (100% positive)
    +50% Cut-off50+/0-Yes (100% positive)
    +75% Cut-off50+/0-Yes (100% positive)
    +100% Cut-off50+/0-Yes (100% positive)
    Oxycodone (Dip Card & Cup)100 ng/mL-100% Cut-off50-/0+Yes (100% negative)
    -75% Cut-off50-/0+Yes (100% negative)
    -50% Cut-off50-/0+Yes (100% negative)
    -25% Cut-off50-/0+Yes (100% negative)
    Cut-off24-/26+ or 26-/24+ or 25-/25+Yes (near 50/50 split at cutoff)
    +25% Cut-off50+/0-Yes (100% positive)
    +50% Cut-off50+/0-Yes (100% positive)
    +75% Cut-off50+/0-Yes (100% positive)
    +100% Cut-off50+/0-Yes (100% positive)
    Buprenorphine (Dip Card & Cup)10 ng/mL-100% Cut-off50-/0+Yes (100% negative)
    -75% Cut-off50-/0+Yes (100% negative)
    -50% Cut-off50-/0+Yes (100% negative)
    -25% Cut-off50-/0+Yes (100% negative)
    Cut-off27-/23+, 25-/25+, 24-/26+, 26-/24+Yes (near 50/50 split at cutoff)
    +25% Cut-off50+/0-Yes (100% positive)
    +50% Cut-off50+/0-Yes (100% positive)
    +75% Cut-off50+/0-Yes (100% positive)
    +100% Cut-off50+/0-Yes (100% positive)
    Phencyclidine (Dip Card & Cup)25 ng/mL-100% Cut-off50-/0+Yes (100% negative)
    -75% Cut-off50-/0+Yes (100% negative)
    -50% Cut-off50-/0+Yes (100% negative)
    -25% Cut-off50-/0+Yes (100% negative)
    Cut-off26-/24+, 25-/25+, 24-/26+Yes (near 50/50 split at cutoff)
    +25% Cut-off50+/0-Yes (100% positive)
    +50% Cut-off50+/0-Yes (100% positive)
    +75% Cut-off50+/0-Yes (100% positive)
    +100% Cut-off50+/0-Yes (100% positive)
    Nortriptyline (Dip Card & Cup)1000 ng/mL-100% Cut-off50-/0+Yes (100% negative)
    -75% Cut-off50-/0+Yes (100% negative)
    -50% Cut-off50-/0+Yes (100% negative)
    -25% Cut-off50-/0+Yes (100% negative)
    Cut-off25-/25+, 26-/24+, 24-/26+Yes (near 50/50 split at cutoff)
    +25% Cut-off50+/0-Yes (100% positive)
    +50% Cut-off50+/0-Yes (100% positive)
    +75% Cut-off50+/0-Yes (100% positive)
    +100% Cut-off50+/0-Yes (100% positive)
    Propoxyphene (Dip Card & Cup)300 ng/mL-100% Cut-off50-/0+Yes (100% negative)
    -75% Cut-off50-/0+Yes (100% negative)
    -50% Cut-off50-/0+Yes (100% negative)
    -25% Cut-off50-/0+Yes (100% negative)
    Cut-off23-/27+, 26-/24+, 24-/26+, 25-/25+Yes (near 50/50 split at cutoff)
    +25% Cut-off50+/0-Yes (100% positive)
    +50% Cut-off50+/0-Yes (100% positive)
    +75% Cut-off50+/0-Yes (100% positive)
    +100% Cut-off50+/0-Yes (100% positive)

    For the Lay-user Study:

    The acceptance criteria for the lay-user study would typically be a high percentage of correct results, especially for samples clearly negative or positive, and a reasonable proportion around the cutoff. The document shows the "percentage of correct results (%)" for each concentration level. The implied acceptance criteria for OTC urine drug screens are usually very high concordance (e.g., >90%) with the ground truth for samples away from the cutoff, and some level of agreement for near-cutoff samples (where variability is expected). The device performance meets these generally implied high standards, especially for samples clearly below or above the cutoff (100% for most categories). For samples near +/-25% of cutoff, the accuracy ranges from 85-95%, which is typical for qualitative tests and acceptable for over-the-counter use where confirmation is recommended for positive results.

    2. Sample Size Used for the Test Set and Data Provenance

    • Precision Studies (Test Set):
      • For each drug, device type (Dip Card/Cup), and each of the 9 concentration levels: 50 tests were performed per concentration.
      • This was repeated for 3 different lots for each device type.
      • Therefore, for a single drug and device type, the test set size for precision was: 9 concentrations * 50 tests/concentration * 3 lots = 1350 tests.
      • Data Provenance: The samples were prepared by "spiking drug in negative samples" and confirmed by LC/MS. This suggests controlled, prospective laboratory-prepared samples rather than retrospective clinical samples from specific countries. The testing was "in-house."
    • Comparison Studies (Test Set):
      • For each drug and device type (Dip Card/Cup): 80 unaltered clinical samples (40 negative and 40 positive) were used.
      • Data Provenance: "unaltered clinical samples." The document does not specify the country of origin, but given the manufacturer (China) and submission to US FDA, the samples could be from various global sources or acquired for the study. It's a retrospective analysis against independently confirmed samples.
    • Lay-user Study (Test Set):
      • For each device format (presumably Dip Card and Cup, though only combined results are shown in the given table): 300 lay persons participated.
      • The total number of samples processed for each drug across the different concentration levels in the lay-user study tables is: 20 + 20 + 160 + 20 + 20 + 40 + 20 = 300 samples. (It seems 300 samples were tested for each drug type, probably with each of the 300 lay users testing one sample)
      • Data Provenance: Urine samples were "prepared at the following concentrations; negative, +/-75%, +/-50%, +/-25% of the cutoff by spiking drugs into drug free-pooled urine specimens." These were laboratory-prepared samples. The study involved "three intended user sites."

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • Precision Studies: Ground truth was established by LC/MS (Liquid Chromatography-Mass Spectrometry), which is a definitive quantitative analytical method. It's an instrumental method, so human "experts" in interpretation are typically not specified in the same way as for imaging or clinical diagnosis. The samples were "blindly labeled by the person who prepared the samples," which indicates blinding to the LC/MS results during the device testing.
    • Comparison Studies: Ground truth was established by LC/MS results. The document states, "The samples were blind labeled and compared to LC/MS results." These are definitive chemical analysis results.
    • Lay-user Study: Ground truth for the samples was established by LC/MS confirmation of the spiked drug concentrations.

    4. Adjudication Method for the Test Set

    • Precision Studies: No explicit human adjudication method mentioned, as the device results are quantitative categories (positive/negative) and compared to definitive LC/MS values. The results are reported as counts (e.g., 25-/25+).
    • Comparison Studies: The device results were read by "three laboratory assistants." For discordant results, the individual viewer's result is listed against the LC/MS truth, but no specific human adjudication process (e.g., 2+1 majority voting) is described for resolving discrepancies among the assistants or between the assistants and LC/MS. The LC/MS is the definitive ground truth here.
    • Lay-user Study: The results are presented as aggregated counts of positive/negative readings by the lay users compared to the known ground truth (LC/MS confirmed concentrations). No human adjudication process is mentioned among the lay users.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No, an MRMC comparative effectiveness study was not done. This type of study is relevant for AI-assisted diagnostic devices where human readers interpret images or complex data, and the AI serves as an aid.
    • This device is a qualitative, over-the-counter urine drug test. Its performance is evaluated against chemical analytical ground truth, and its usability is assessed with lay users, not by evaluating how it assists expert readers.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    • This device is a physical, qualitative immunoassay test kit, not a standalone algorithm/AI for analysis. The "performance" of the device itself (the reaction on the dip card/cup) is evaluated.
    • However, the Comparison Studies and Precision Studies could be considered the "standalone" performance of the device when read by trained laboratory personnel (3 laboratory assistants), as it's the device's output itself being compared to LC/MS.
    • The Lay-user study evaluates the device performance when interpreted by the intended "human-in-the-loop" (a lay user).

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    • The primary ground truth used for all performance studies (Precision, Comparison, Lay-user) is quantitative chemical analysis by LC/MS (Liquid Chromatography-Mass Spectrometry). This is a highly accurate and widely accepted method for confirming drug concentrations in urine.
    • For the precision and lay-user studies, urine samples were spiked with known concentrations of drugs, and these concentrations were confirmed by LC/MS.
    • For the comparison studies, "unaltered clinical samples" were used, with LC/MS providing the definitive truth for the presence and concentration of drugs.

    8. The Sample Size for the Training Set

    • This document describes premarket notification (510(k)) for a drug test kit, which is a physical diagnostic device based on immunochromatography, not an AI/Machine Learning algorithm.
    • Therefore, there is no "training set" in the context of machine learning model development. The device's performance relies on its physical and chemical design, not on a trained algorithm.

    9. How the Ground Truth for the Training Set Was Established

    • As there is no AI/ML algorithm or training set, this question is not applicable.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1