Search Filters

Search Results

Found 5 results

510(k) Data Aggregation

    K Number
    K201568

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-11-12

    (154 days)

    Product Code
    Regulation Number
    888.3070
    Age Range
    All
    Predicate For
    N/A
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Calvary Spine Pedicle Screw System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of thoracic, lumbar, and sacral spine (T1 to S1/ilium); degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, deformities or curvatures (i.e. scoliosis, kyphosis, and/or lordosis), spinal tumor, pseudarthrosis and failed previous fusion.

    When used for posterior, non-cervical (T1 to S1/ilum), pedicte screw fixation in pediatric patients, the Calvary Spine Pedicle Screw System implants are indicated as an adjunct to fusion to treat progressive spinal deformities (i.e., scoliosis, kyphosis, or lordosis) including idiosis, neuromuscular scoliosis, and congenital scoliosis. Additionally, the Calvary Spine Pedicle Screw System is intended to treat pediatric patients diagnosed with spondylolisthesis/spondylolysis, fracture caused by tumor and/or trauma, pseudarthrosis, and/or failed previous fusion.

    This system is intended to be used with autograft. Pediatric pedicle screw fixation is limited to a posterior approach.

    Device Description

    Calvary Spine Pedicle Screw System includes screws and connecting components in a wide variety of sizes and shapes, which can be locked in various configurations, each assembly being tailor-made. Calvary pedicular screws must be used with Ø 5.5mm straight and curved rods.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the Calvary Spine Pedicle Screw System:

    Based on the provided FDA 510(k) summary (K201568), there is no information about acceptance criteria or a study proving the device meets specific performance criteria in terms of an AI/algorithm-based medical device.

    The document describes a traditional medical device, specifically a pedicle screw system, which is a hardware implant used in spinal fusion surgery. The "acceptance criteria" discussed in this document refer to the regulatory requirements for establishing substantial equivalence to previously cleared predicate devices, primarily based on design, materials, and mechanical performance through non-clinical testing.

    Therefore, I cannot populate most of the requested fields because they are specifically tailored for studies validating AI/algorithm performance.

    However, I can extract the relevant information regarding the type of study and the basis for substantial equivalence:


    1. Table of Acceptance Criteria and Reported Device Performance

    As this is a traditional medical device and not an AI/algorithm, there are no specific performance metrics like sensitivity, specificity, or AUC to list as acceptance criteria for automated diagnosis or analysis. The "acceptance criteria" for this device are broadly the demonstration of substantial equivalence to predicate devices based on:

    Acceptance Criteria (Broad Interpretation for Traditional Device)Reported Device Performance (Summary)
    Intended Use EquivalenceEquivalent to predicate devices.
    Material EquivalenceUses same materials (Ti-6Al-4V ELI, CP Titanium, CoCrMo).
    Design EquivalenceEquivalent in shape, sizes, material, manufacturing process.
    Strength EquivalenceGreater or equivalent strength values compared to predicate devices.

    2. Sample Size Used for the Test Set and Data Provenance

    • Not applicable (N/A) for an AI/algorithm test set.
    • The "test set" in this context refers to the samples used in non-clinical mechanical testing, which are typically manufactured units of the device. No information on the specific number of units tested for each mechanical test is provided, but it would involve standard engineering sample sizes for material and mechanical property evaluations.
    • Data Provenance: The data provenance for mechanical testing would be from internal laboratory testing of the manufactured devices. No country of origin for data in the AI sense is relevant here, nor is retrospective or prospective human data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • N/A for an AI/algorithm test set.
    • Ground truth for mechanical properties is established through standard engineering and materials testing protocols (e.g., ASTM standards), not by human experts adjudicating images or clinical data.

    4. Adjudication Method for the Test Set

    • N/A for an AI/algorithm test set.
    • Mechanical test results are objective measurements against defined standards, not subject to human adjudication methods like 2+1 or 3+1.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    • No. An MRMC study was not done.
    • This device is not an AI/algorithm, therefore, there is no human-in-the-loop performance improvement with AI assistance to measure.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • No. A standalone algorithm performance study was not done.
    • This device is not an AI/algorithm.

    7. The Type of Ground Truth Used

    • Mechanical and Material Standards: The "ground truth" for this medical device's performance is compliance with established material specifications (e.g., ASTM F136, F67, F1582) and mechanical testing standards for spinal implants, demonstrating equivalent or superior strength to predicate devices.

    8. The Sample Size for the Training Set

    • N/A for an AI/algorithm training set.
    • This device does not involve a training set as it's not an AI/algorithm.

    9. How the Ground Truth for the Training Set Was Established

    • N/A for an AI/algorithm training set.
    • Not applicable as there is no training set.

    Summary of what was done (based on the provided text):

    The Calvary Spine Pedicle Screw System underwent non-clinical testing to demonstrate that it is substantially equivalent to legally marketed predicate devices. This equivalence was established based on:

    • Similar materials (Ti-6Al-4V ELI, CP Titanium, CoCrMo).
    • Similar intended use.
    • Similar design (shape, sizes, manufacturing process).
    • Comparable or superior strength values, presumably measured through mechanical testing (though specific tests and methodology are not detailed in this summary).

    The document explicitly states: "No clinical studies were performed." This is common for 510(k) clearances where substantial equivalence can be demonstrated through non-clinical data.

    Ask a Question

    Ask a specific question about this device

    K Number
    K201659

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-10-05

    (109 days)

    Product Code
    Regulation Number
    888.3070
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The KHEIRON Spinal Fixation System is intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of thoracic, lumbar, and sacral spine: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, deformities or curvatures (i.e. scoliosis, kyphosis, and/or lordosis), spinal stenosis, spinal tumor, pseudarthrosis and failed previous fusion.

    When used for posterior, non-cervical, pedicle screw fixation in pediatric patients, the KHEIRON System implants are indicated as an adjunct to fusion to treat progressive spinal deformities (i.e., scoliosis, kyphosis) including idiopathic scoliosis, neuromuscular scoliosis, and congenital scoliosis. Additionally, the KHEIRON Spinal Fixation System is intended to treat pediatic patients diagnosed with spondylolysis, fracture caused by tumor and/or trauma, pseudarthrosis, and/or failed previous fusion.

    This system is intended to be used with autograft. Pediatric pedicle screw fixation is limited to a posterior approach.

    Device Description

    The KHEIRON® Spinal Fixation System may be used for a variety of conditions that affect the thoracic and lumbar spine. In cases in which the posterior elements are fractured, the pedicle screw offers an excellent means of stabilizing a specific spinal segment. KHEIRON Spinal Fixation System includes screws and connecting components in a wide variety of sizes and shapes, which can be locked in various configurations, each assembly being tailor-made. KHEIRON pedicular screws must be used with Ø 5.5mm and Ø 6mm rods.

    AI/ML Overview

    The KHEIRON® Spinal Fixation System is a medical device and the provided document is a 510(k) summary for its premarket notification to the FDA. This summary focuses on demonstrating substantial equivalence to predicate devices rather than proving the device meets specific performance criteria through a standalone study of its capabilities in a clinical setting.

    Therefore, the requested information regarding acceptance criteria, study details, sample sizes, ground truth establishment, expert qualifications, and MRMC studies is not directly available in the provided text. The document states "No clinical studies were performed."

    However, the document does provide information on non-clinical testing to demonstrate the physical performance and safety of the device:

    1. A table of acceptance criteria and the reported device performance:

    Since no specific numerical acceptance criteria (e.g., minimum strength values) are given, and only a comparative statement is made regarding performance against predicates, a direct table of acceptance criteria and the reported device performance as typically expected for an AI/diagnostic device is not applicable here.

    Instead, the non-clinical test summary states:

    Acceptance Criteria CategoryReported Device Performance
    Strength (Mechanical)"The KHEIRON Spinal Fixation System has greater or equivalent strength values compared to other devices cleared for use in the thoracolumbosacral spine."

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):

    • Test Set Sample Size: Not specified as clinical studies were not performed. For non-clinical tests (mechanical tests), the sample sizes for the devices tested were not provided.
    • Data Provenance: Not applicable as no clinical data was used. Mechanical testing is typically done in a laboratory setting. No country of origin for the mechanical test data is mentioned.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):

    • Not applicable, as no clinical studies requiring expert consensus for ground truth were performed.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not applicable, as no clinical studies requiring ground truth adjudication were performed.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The document explicitly states, "No clinical studies were performed." This device is a mechanical spinal fixation system, not an AI or diagnostic tool that would typically involve human "readers."

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This device is a mechanical implant, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • For the non-clinical mechanical tests, the "ground truth" would be the engineering standards (e.g., ASTM F1717 and ASTM F1798) and the performance characteristics of the predicate devices. The device's strength was directly measured against these established benchmarks.

    8. The sample size for the training set:

    • Not applicable, as this is a mechanical spinal fixation system and not an AI/machine learning device that requires a training set.

    9. How the ground truth for the training set was established:

    • Not applicable, as this is a mechanical spinal fixation system and not an AI/machine learning device.

    Summary of Non-Clinical Testing:

    The KHEIRON® Spinal Fixation System underwent the following non-clinical mechanical tests:

    • Static and dynamic compression bending (according to ASTM F1717)
    • Static torsion (according to ASTM F1717)
    • Torsional and axial gripping capacity (according to ASTM F1798)

    The results of these evaluations indicated that the KHEIRON Spinal Fixation System is "substantially equivalent to legally marketed predicate devices" and specifically noted that its "strength values" were "greater or equivalent" compared to other cleared devices for thoracolumbosacral spine use.

    Ask a Question

    Ask a specific question about this device

    K Number
    K150294

    Validate with FDA (Live)

    Date Cleared
    2015-06-05

    (119 days)

    Product Code
    Regulation Number
    888.3070
    Age Range
    All
    Predicate For
    N/A
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DPZ Pedicular Fixation System is intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of thoracic, lumbar and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, deformities or curvatures (i.e., scoliosis, and/or lordosis), spinal tumor, pseudarthrosis and failed previous fusion.

    The DPZ Pedicular Fixation System is also intended for non-cervical pedicle screw fixation for the following indications: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion. It is also intended for the following indications: trauma (i.e., fracture or dislocation); spinal stenosis, deformities or curvatures (i.e., scoliosis, and/or lordosis), tumor, pseudarthrosis and failed previous fusion.

    Device Description

    The DPZ Pedicular Fixation System is a top loading, multiple component, posterior spinal fixation system which consists of pedicle screws, rods and cross links. All of the components are available in a variety of sizes to more closely match the patient's anatomy.

    AI/ML Overview

    The provided text describes the DPZ Pedicular Fixation System, its intended use, and the non-clinical tests conducted to demonstrate its equivalence to predicate devices. However, it explicitly states: "No clinical studies were performed."

    Therefore, I cannot provide information on acceptance criteria based on human performance or studies involving human readers/experts, as no such clinical data is presented in this document.

    Here's the information that can be extracted from the provided text regarding the device's acceptance criteria and the study that proves it meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Test Standard)Reported Device Performance (Result Summary)
    Static compression per ASTM F1717"The results of these evaluations indicate that the DPZ Pedicular Fixation System is equivalent to predicate devices."
    Dynamic compression per ASTM F1717"The results of these evaluations indicate that the DPZ Pedicular Fixation System is equivalent to predicate devices."
    Static torsion per ASTM F1717"The results of these evaluations indicate that the DPZ Pedicular Fixation System is equivalent to predicate devices."

    2. Sample size used for the test set and the data provenance:

    • Sample Size: Not specified for the non-clinical tests. The document only mentions "The following analyses were conducted."
    • Data Provenance: Not explicitly stated, but it refers to "non-clinical tests" which typically involve laboratory or bench testing of the device hardware.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. The "ground truth" for non-clinical mechanical testing is based on the specifications within the ASTM F1717 standard and the measured performance of the device compared to predicate devices. Human experts are not involved in establishing this ground truth for a mechanical test.

    4. Adjudication method for the test set:

    • Not applicable, as this refers to human reader review and consensus, which was not performed.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No MRMC study was done, as "No clinical studies were performed."

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is a medical device (pedicle fixation system), not an algorithm or AI. The standalone performance refers to the device's mechanical integrity under specified ASTM standards.

    7. The type of ground truth used:

    • The ground truth for the non-clinical tests was based on the performance requirements and methodologies outlined in the ASTM F1717 standard for spinal implant constructs. The device's performance was then compared to that of predicate devices to establish substantial equivalence.

    8. The sample size for the training set:

    • Not applicable. This is a non-AI/algorithm medical device. "Training set" refers to data used to train an AI model.

    9. How the ground truth for the training set was established:

    • Not applicable for the same reason as above.
    Ask a Question

    Ask a specific question about this device

    K Number
    K140678

    Validate with FDA (Live)

    Date Cleared
    2014-06-19

    (93 days)

    Product Code
    Regulation Number
    888.3070
    Age Range
    All
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K093077 / K100685, K050461 / K052761 / K060361 / K060748 / K071373, K113666, K940631 / 950099, K950697, K920116

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CosmoLock Pedicle Screw System is intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, deformities or curvatures (i.e. scoliosis, and/or lordosis), spinal tumor, pseudarthrosis and failed previous fusion.

    The CosmoLock Pedicle Screw System is also intended for non-cervical pedicle screw fixation for the following indications: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion. It is also intended for the following indications: trauma (i.e. fracture or dislocation); spinal stenosis; deformities or curvatures (i.e. scoliosis, and/or lordosis), tumor; pseudoarthrosis; and failed previous fusion.

    Device Description

    The CosmoLock Pedicle Screw System is a top loading, multiple component, posterior spinal fixation system which consists of pedicle screws, rods and cross links. All of the components are available in a variety of sizes to match more closely the patient's anatomy.

    Materials:

    Ti-6Al-4V per ASTM F136 CoCr per ASTM F1537

    AI/ML Overview

    This document is a 510(k) Summary for the CosmoLock Pedicle Screw System. It describes the device and its intended use and claims substantial equivalence to predicate devices. However, it does not contain the detailed information needed to answer many of your questions, as it explicitly states that no clinical studies were performed.

    Here's an analysis based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (Test)Reported Device Performance
    Static compression per ASTM F1717Equivalent to predicate devices (implied: met criteria)
    Dynamic compression per ASTM F1717Equivalent to predicate devices (implied: met criteria)
    Static torsion per ASTM F1717Equivalent to predicate devices (implied: met criteria)

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided in the document. The studies mentioned are non-clinical (mechanical tests), and details about sample sizes for these tests are not specified.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable/provided. No clinical studies were performed, and therefore no ground truth established by experts is relevant to the data presented.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable/provided. No clinical studies were performed.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This document states, "No clinical studies were performed."

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable. The device is a mechanical pedicle screw system, not an AI or algorithm-based device. No standalone performance study in this context was performed.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    The ground truth for the non-clinical tests would be the established mechanical testing standards (ASTM F1717) and the performance of the predicate devices. There is no biological or expert-defined "ground truth" in the typical medical imaging/diagnosis sense, as this is a mechanical implant.

    8. The sample size for the training set

    This information is not applicable/provided. No clinical studies were performed, and there is no mention of an algorithm or AI requiring a training set.

    9. How the ground truth for the training set was established

    This information is not applicable/provided. No training set or associated ground truth was established, as no clinical studies or AI development are described. The focus is on demonstrating mechanical equivalence to predicate devices.

    Ask a Question

    Ask a specific question about this device

    K Number
    K111370

    Validate with FDA (Live)

    Date Cleared
    2011-09-21

    (128 days)

    Product Code
    Regulation Number
    888.3070
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K000536, K920116

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ODALYS™ Pedicle Screw System is a pedicle screw system intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis).

    As a pedicle screw system ODALYS™ Pedicle Screw System is indicated for the treatment of severe spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (13 to sacrum) with removal of the implants after the attainment of a solid fusion.

    Device Description

    The ODALYS™ Pedicle Screw System has been developed with the objective of providing the surgeon with an adaptable fixation system in order to carry out dorsal stabilization of the spine simply, quickly and effectively.

    The system consists of a variety of color coded top loading pedicle screws. The pedicle screws are available in various lengths and diameters. The screw is connected to the rod via a rod connector. Two sizes of connectors are available, short and long. The long is used in cases of spondylolisthesis where the short connector would not be able to engage the rod. The rods are available in multiple lengths.

    Materials: The components are manufactured from titanium alloy (Ti 6Al 4V ELI) per ASTM F136.

    Function: The ODALYS™ Pedicle Screw System is a pedicle screw system intended to provide immobilization and stabilization of spinal segments until fusion takes place.

    AI/ML Overview

    The provided text describes the ODALYS™ Pedicle Screw System, a medical device, and its acceptance criteria as determined by the FDA's 510(k) clearance process.

    Here's an analysis of the acceptance criteria and the "study" (non-clinical tests) used to prove the device meets these criteria:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Implied by Predicate Equivalence)Reported Device Performance (ODALYS™ Pedicle Screw System)
    Mechanical Performance (Static Compression Resistance)Met or exceeded the performance of predicate devices: Moss Miami (K000536) and PWB Lumbosacral Spine System (K920116) as evidenced by ASTM F1717 testing.
    Mechanical Performance (Dynamic Compression Resistance)Met or exceeded the performance of predicate devices: Moss Miami (K000536) and PWB Lumbosacral Spine System (K920116) as evidenced by ASTM F1717 testing.
    Mechanical Performance (Static Torsion Resistance)Met or exceeded the performance of predicate devices: Moss Miami (K000536) and PWB Lumbosacral Spine System (K920116) as evidenced by ASTM F1717 testing.
    Material CompositionThe device is manufactured from titanium alloy (Ti 6Al 4V ELI) per ASTM F136, which is equivalent to materials used in predicate devices.
    Intended Use (Immobilization and stabilization of spinal segments for fusion)The ODALYS™ Pedicle Screw System is intended for immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion for specific conditions (degenerative spondylolisthesis, fracture, dislocation, scoliosis, kyphosis, spinal tumor, failed previous fusion), and for severe spondylolisthesis (Grades 3 and 4) of L5-S1. This matches the intended use of predicate devices.
    Design and Function (Top-loading pedicle screws, rod connectors, rods to provide spinal stabilization)The ODALYS™ Pedicle Screw System has a similar design (variety of color-coded top-loading pedicle screws, rod connectors, and rods) and function (immobilization and stabilization of spinal segments until fusion takes place) compared to predicate devices.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The document does not specify a "test set" in the context of human data. The testing performed was non-clinical, involving mechanical tests of the device. The sample size for these mechanical tests (e.g., number of screws or constructs tested for each configuration) is not provided in the summary.
    • Data Provenance: The data provenance is from non-clinical laboratory testing (mechanical testing per ASTM standards). There is no information regarding country of origin or whether it's retrospective or prospective, as it pertains to physical device testing, not human data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This section is not applicable. The "ground truth" for this device, in the context of the 510(k) submission, is its mechanical performance and material properties. This is established by adherence to recognized ASTM standards for medical devices, rather than expert consensus on a dataset. The results are compared to predicate devices.

    4. Adjudication Method for the Test Set

    This section is not applicable. There was no human "test set" requiring adjudication. The assessment was based on objective mechanical testing results comparing the device to accepted standards and predicate device performance.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, a Multi-Reader, Multi-Case (MRMC) comparative effectiveness study was not conducted. This type of study is relevant for diagnostic or AI-assisted devices that involve human interpretation of medical images or data. The ODALYS™ Pedicle Screw System is a physical implant, not an AI-based diagnostic tool.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    No, a standalone algorithm performance study was not done. The device is a physical implant; there is no AI algorithm associated with it.

    7. The Type of Ground Truth Used

    The "ground truth" for this device's acceptance is based on:

    • Compliance with recognized industry standards (ASTM F1717 and ASTM F136).
    • Substantial equivalence to legally marketed predicate devices in terms of intended use, design, materials, and function, as demonstrated by meeting or exceeding the mechanical performance and material specifications of those predicates.

    8. The Sample Size for the Training Set

    This section is not applicable. The ODALYS™ Pedicle Screw System is a physical pedicle screw system, not an AI or machine learning device that requires a "training set."

    9. How the Ground Truth for the Training Set was Established

    This section is not applicable for the same reason as point 8. There is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1