Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K023218

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2003-03-24

    (179 days)

    Product Code
    Regulation Number
    868.5150
    Age Range
    All
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K813186, K013610, K000722

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Pajunk Plexolong sets are intended for delivery of continuous conduction anesthesia of peripheral nerves and plexus for up to 72 hours. Continuous delivery for up to 72 hours is accomplished using the Polyamide indwelling catheter. An electrical stimulus may be applied to the conduction needle via a cable and connectors to assist the physician pinpoint the area of application.

    Device Description

    The PAJUNK Plexolong sets contain single use sterile and non-pyrogenic needles with tubing. with or without a plastic cannula and catheters intended for delivery of continuous conduction anesthesia of peripheral nerves and plexus for up to 72 hours. Continuous delivery is accomplished using the catheter. An electrical stimulus may be applied to the conduction needle via a cable and connector to assist the physician pinpoint the area of application.

    AI/ML Overview

    This 510(k) summary describes a medical device, specifically an Anesthesia Conduction Kit. For this type of device, "acceptance criteria" are typically defined by adherence to recognized consensus standards and "device performance" is demonstrated through testing to ensure conformance to these standards, as opposed to clinical performance metrics for an AI-enabled device.

    Let's break down the information requested based on the provided document:


    Description of Acceptance Criteria and Study to Prove Device Meets Acceptance Criteria

    The PAJUNK Plexolong Sets are designed to conform to recognized national and international consensus standards relevant to medical devices, particularly those for needles, cannulas, Luer fittings, sterilization, packaging, and risk analysis. The study demonstrating that the device meets these acceptance criteria involved performance testing to verify conformance to these applicable standards.


    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criterion (Standard)Reported Device Performance
    Mechanical/Physical Conformance:
    DIN 13090/ISO 594: Luer fittings w/wo locking featureConformance verified through testing.
    DIN 13097 Part 1: Medical injection cannulaConformance verified through testing.
    DIN 13097 Part 3: Medical cannulaConformance verified through testing.
    DIN EN 1707: 6% Luer cone connections for injection cannula and particular medical equipmentConformance verified through testing.
    BS 4843: Single entry IV cannulaConformance verified through testing.
    Material Conformance:
    DIN 17442/ISO 9626: Steel for medical instrumentsConformance verified through testing. PAJUNK Unipolar needles (including physical dimensions, coating, connector, tubing, metal, and plastics) were previously cleared under K000722, indicating material conformance. The Touhy needle's tip configuration is identical to a predicate cleared under K813186.
    DIN EN/ISO 9626: Cannula tube of non-rusting steel (SS) for the manufacture of medical productsConformance verified through testing.
    DIN 17440: Stainless SteelsConformance verified through testing.
    Sterilization & Packaging Conformance:
    DIN EN 550: Sterilization of med. Prod.; Validation & routine controls for sterilization with ETOConformance verified through testing. The contract sterilizer and sterilizing process are identical to those used for previously cleared Pajunk Plexolong sets (K013041), implying validated ETO sterilization.
    DIN EN 556: Sterilization of medical products, requirements for medical products that are labeled "sterile"Conformance verified through testing.
    PrEN 868-1: Packaging materials for the sterilization of packaged goods. Part 1: general requirements for the validation of the packaging of sterilized end-packaged productsConformance verified through testing. Packaging materials are the same as previously cleared Pajunk Plexolong sets, implying validated packaging for sterility.
    DIN EN 868-2: Packaging materials for the sterilization of packaged goods. Part 2: sterilization packaging, requirements and tests.Conformance verified through testing.
    Quality System & Risk Management Conformance:
    DIN EN 724: Guidance on the application of EN29001 and EN46001 for non-active medical productsConformance verified through testing.
    DIN EN 980: Graphic symbols for marking medical productsConformance verified through testing.
    DIN EN 1441: Risk analysis for medical productsConformance verified through testing.
    DIN en 30993-1: Biological evaluation of medical products - instructions for selection of testsConformance verified through testing.
    DIN EN 46001: Particular requirements for medical productsConformance verified through testing.

    Regarding AI/Machine Learning Specific Questions:

    The provided document describes a conventional medical device (anesthesia conduction kit) and does not refer to any AI or machine learning components. Therefore, the following AI/ML-specific questions are not applicable to this submission.


    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not applicable. This document describes a physical medical device, not an AI/ML algorithm or software that uses test data in the conventional sense. Performance testing was conducted on samples of the physical device according to industry standards.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable. Ground truth in the context of expert consensus is not relevant for this physical device. Performance is assessed against engineering and material specifications as defined by consensus standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. This relates to AI/ML model performance evaluation, not a physical medical device.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This device does not involve human "readers" or AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. There is no algorithm or software component described.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • For this device, "ground truth" is established by the specifications outlined in the referenced consensus standards. Device components (e.g., Luer fittings, cannula dimensions, material properties, sterility assurance) are tested against these published, accepted industry standards.

    8. The sample size for the training set

    • Not applicable. This device does not involve a training set for an AI/ML model.

    9. How the ground truth for the training set was established

    • Not applicable. This device does not involve a training set for an AI/ML model.

    Ask a Question

    Ask a specific question about this device

    K Number
    K984063

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    1999-02-09

    (85 days)

    Product Code
    Regulation Number
    880.5440
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K813186

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Paragon Infusion Kit is intended to provide continuous infusion a local anesthetic directly into an intraoperative (soft tissue / body cavity) site for general surgery for postoperative pain management. Additional routes of administration include percutaneous, subcutaneous, intramuscular and epidural infusion. The Paragon pump is re-usable. The disposable Paragon administration set is single patient use only. No testing has been conducted to determine the efficacy of the Paragon for the delivery of blood, blood products, lipids or fat emulsions. The Paragon is not intended for the delivery of blood, blood products, lipids or fat emulsions. The Paragon is suitable for use as an ambulatory device and is intended for use in the hospital, home environment or alternative care sites.

    Device Description

    The Paragon Infusion Kit is identical to the I-Flow PainBuster Infusion Kit with the exception of the Paragon pump and administration set replacing the PainBuster pump. The kit is comprised of a Paragon pump and administration set (K923875) and various kit components such as catheter, needle, syringe, Y adapter, dressing, tape, gauze and carry case. The Paragon administration set is intended to attach to the kit catheter at the distal end of the set to provide continuous infusion of a local anesthetic directly into the intraoperative site for general surgery for postoperative pain management. The Paragon administration set is a disposable device intended for single patient use. The Paragon pump is reusable. The Paragon is suitable for use as an ambulatory device and is intended for use in the hospital, home environment or alternative care sites.

    AI/ML Overview

    The provided text describes the "Paragon Infusion Kit" and includes a summary of safety and effectiveness, as well as operational specifications and performance data. The device is an infusion pump kit intended for continuous infusion of local anesthetics for postoperative pain management.

    Based on the document, here's an analysis of the acceptance criteria and study information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria for Flow Rate Accuracy: ±10% at 95% confidence interval (as stated in Section 3.1)

    ModelTargeted Flow Rate (ml/hr)Reported Average Flow Rate (ml/hr)Standard DeviationNumber of Samples (N)Achieved Accuracy (within ±10%)
    100 ml x 0.5 ml/hr0.50.530.015Yes (0.45 to 0.55 ml/hr)
    100 ml x 1.0 ml/hr1.00.990.035Yes (0.90 to 1.10 ml/hr)
    100 ml x 2.0 ml/hr2.01.990.075Yes (1.80 to 2.20 ml/hr)
    100 ml x 4.0 ml/hr4.04.140.105Yes (3.60 to 4.40 ml/hr)
    100 ml x 10.0 ml/hr10.010.300.3810Yes (9.0 to 11.0 ml/hr)

    Additional Acceptance Criteria / Operational Specifications (Section 3.1):

    • Priming/Residual Volume: < 5 ml
    • Operating Temperature: 31°C skin temperature (90°F)
    • Test Solution: 0.9% NaCl
    • Operating Pressure: 6.0 psi pressure source
    • Head Height: 0"

    The document states: "All models produced an average flow rate within the ±10% accuracy claim." (Section 3.2)

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size:
      • For 100 ml x 0.5 ml/hr models: 5 pieces
      • For 100 ml x 1.0 ml/hr models: 5 pieces
      • For 100 ml x 2.0 ml/hr models: 5 pieces
      • For 100 ml x 4.0 ml/hr models: 5 pieces
      • For 100 ml x 10.0 ml/hr models: 10 pieces
    • Data Provenance: The document does not specify the country of origin for the data or whether the study was retrospective or prospective. It appears to be an internal laboratory performance test conducted under specified "standard operating conditions" (Section 3.1).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    • Not applicable. The "ground truth" for this performance study is a physical measurement of flow rate against a specified target. This is a laboratory test, not an assessment requiring expert consensus on clinical data.

    4. Adjudication Method for the Test Set

    • Not applicable. This was a direct measurement of physical performance (flow rate), not a judgment-based assessment requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    • No. An MRMC study is relevant for diagnostic or clinical interpretation devices where human readers assess cases, often with and without AI assistance, to determine changes in diagnostic accuracy or efficiency. This document describes a medical device (infusion pump) and its physical performance, not a diagnostic algorithm.

    6. If a Standalone Study (i.e., algorithm only without human-in-the-loop performance) Was Done

    • Yes, in spirit. The testing presented is entirely standalone performance for the device (infusion pump). There is no "algorithm" in the sense of AI or software interpreting data for diagnosis, but rather the mechanical performance of the pump itself. The measurements are taken directly from the device's output.

    7. The Type of Ground Truth Used

    • Physical Measurement/Engineered Specification: The ground truth is the targeted flow rate (e.g., 0.5 ml/hr, 1.0 ml/hr), and the device's performance is measured against this specification under controlled laboratory conditions. The "Test Solution" was 0.9% NaCl, and "Operating Pressure" was 6.0 psi, along with other specified conditions (Section 3.1).

    8. The Sample Size for the Training Set

    • Not applicable. The Paragon Infusion Kit is a mechanical device, not an AI or machine learning model. Therefore, there is no "training set" in the context of data-driven model development. The design and engineering of the pump and administration set are based on physical principles (e.g., Poiseuille's equation - Section 2.3.3.1).

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable. As explained above, this is a mechanical device and does not have a training set or ground truth established in the typical sense of AI/ML. Its performance is designed and validated through engineering principles and physical testing.
    Ask a Question

    Ask a specific question about this device

    K Number
    K980558

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    1998-05-28

    (104 days)

    Product Code
    Regulation Number
    880.5725
    Age Range
    All
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K813186

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    1. The PainBuster is intended to provide continuous infusion of a local anesthetic directly into the intraoperative (soft tissue / body cavity) site for postoperative pain management.
    2. The PainBuster is intended to deliver pain medication percutaneously via an administration set attached to a catheter.
    3. The PainBuster is single use only.
    4. The PainBuster is suitable for use as an ambulatory device and is intended for use in the home environment but not limited to use in the home environment.
    5. The PainBuster is not intended for epidural, subcutaneous or vascular drug delivery.
    6. The PainBuster is not intended for chemotherapy drugs.
    7. No testing has been conducted to determine the efficacy of the PainBuster for the delivery of blood, blood products or TPN. The PainBuster is not intended for the delivery of blood, blood products or TPN.
    Device Description

    The PainBuster Infusion System is a kit that is comprised of an elastomeric infusion pump, a catheter and a needle. The PainBuster pump is the Homepump C-Series with a new intended use. The pump design is identical to the original Homepump C-Series except as follows: The PainBuster pump utilizes the same soft PVC shell that the Homepump Eclipse uses. The two (2) outer natural latex bladders have been replaced by a single thicker natural latex bladder. The PainBuster pump is intended to be used with a catheter that is included with the kit.

    AI/ML Overview

    The provided document describes the I-Flow PainBuster Infusion System. This is a premarket notification (510(k)) submission, which typically focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than presenting detailed clinical studies for novel acceptance criteria.

    The acceptance criteria primarily relate to the device's operational specifications, particularly its flow rate accuracy. The study demonstrating that the device meets these criteria is an engineering performance test rather than a clinical study involving human patients or ground truth established by experts.

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Operational Specification)Reported Device Performance (PainBuster Pump)
    Accuracy: ±15% at 95% confidence interval (for flow rate)65ml x 0.5ml/hr model: Average flow rate 0.48 ml/hr (well within ±15% of 0.5 ml/hr nominal)
    125ml x 2.0ml/hr model: Average flow rate 1.98 ml/hr (well within ±15% of 2.0 ml/hr nominal)
    Priming Volume: less than 5.0 ml(Not explicitly stated in performance data, but listed as standard operating condition)
    Residual Volume: less than 5.0 ml(Not explicitly stated in performance data, but listed as standard operating condition)
    Operating Temperature: 31°C (skin temperature)Testing occurred at 31°C
    Test Solution: 0.9% NaClTesting occurred with 0.9% NaCl, and also compared against 5% Dextrose.
    Operating Pressure: 9 to 14 psi(Not explicitly tested for flow rate, but back pressure comparison tests were done on predicate device)
    Head Height: 16"Testing occurred at nominal head height of 16"

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Flow Rate Performance Data (Test Set):
      • 65ml x 0.5ml/hr model: 27 units
      • 125ml x 2.0ml/hr model: 26 units
    • Sample Size for Back Pressure Comparison (on Homepump Eclipse, a predicate device):
      • 65ml x 0.5ml/hr model: 15 units at 16" HH, 10 units at 42" HH, 10 units at 3.1 psi
      • 100ml x 2.0ml/hr model: 15 units at 16" HH, 15 units at 42" HH, 15 units at 3.1 psi
    • Sample Size for Drug Delivery Comparison (on Homepump Eclipse, a predicate device):
      • Saline: 30 units
      • 5% Dextrose: 15 units
    • Sample Size for Catheter/PICC Line Effects (on Homepump Eclipse, a predicate device):
      • PICC Line (Room Temp): Not specified per test, but maximum/minimum/std dev suggest multiple runs/samples
      • PICC Line (31°C): Not specified per test
      • Epidural Catheter (Room Temp): Not specified per test
      • Epidural Catheter (31°C): Not specified per test
    • Data Provenance: The data appears to be from prospective in-vitro engineering performance tests conducted by I-Flow Corporation. There is no information about the country of origin of the data, but it can be inferred that it was likely conducted in the US, given the company's address and the FDA submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. The "ground truth" here is the physical performance of the device (flow rate, residual volume, etc.) determined through engineering measurements, not expert assessment of a condition.

    4. Adjudication Method (e.g., 2+1, 3+1, none) for the Test Set

    Not applicable. This concept pertains to expert review of clinical cases. The tests described are objective engineering measurements.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not a diagnostic device involving human readers or AI.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    This refers to a standalone performance of the device itself (the infusion pump), which is what the engineering tests described in "Flow Rate Performance Data" and "Operational Specifications" represent. The device's performance is measured and compared against its defined accuracy criteria. Therefore, the standalone performance data is available.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" used is the objective physical measurement of the device's operational parameters (e.g., actual flow rate per hour, volume, pressure) against predetermined engineering specifications and industry standards.

    8. The Sample Size for the Training Set

    Not applicable. This is an elastomeric infusion pump, not an AI/ML algorithm that requires a training set. The device is designed based on engineering principles and tested for its inherent physical properties.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1