K Number
K023218

Validate with FDA (Live)

Manufacturer
Date Cleared
2003-03-24

(179 days)

Product Code
Regulation Number
868.5150
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Pajunk Plexolong sets are intended for delivery of continuous conduction anesthesia of peripheral nerves and plexus for up to 72 hours. Continuous delivery for up to 72 hours is accomplished using the Polyamide indwelling catheter. An electrical stimulus may be applied to the conduction needle via a cable and connectors to assist the physician pinpoint the area of application.

Device Description

The PAJUNK Plexolong sets contain single use sterile and non-pyrogenic needles with tubing. with or without a plastic cannula and catheters intended for delivery of continuous conduction anesthesia of peripheral nerves and plexus for up to 72 hours. Continuous delivery is accomplished using the catheter. An electrical stimulus may be applied to the conduction needle via a cable and connector to assist the physician pinpoint the area of application.

AI/ML Overview

This 510(k) summary describes a medical device, specifically an Anesthesia Conduction Kit. For this type of device, "acceptance criteria" are typically defined by adherence to recognized consensus standards and "device performance" is demonstrated through testing to ensure conformance to these standards, as opposed to clinical performance metrics for an AI-enabled device.

Let's break down the information requested based on the provided document:


Description of Acceptance Criteria and Study to Prove Device Meets Acceptance Criteria

The PAJUNK Plexolong Sets are designed to conform to recognized national and international consensus standards relevant to medical devices, particularly those for needles, cannulas, Luer fittings, sterilization, packaging, and risk analysis. The study demonstrating that the device meets these acceptance criteria involved performance testing to verify conformance to these applicable standards.


1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criterion (Standard)Reported Device Performance
Mechanical/Physical Conformance:
DIN 13090/ISO 594: Luer fittings w/wo locking featureConformance verified through testing.
DIN 13097 Part 1: Medical injection cannulaConformance verified through testing.
DIN 13097 Part 3: Medical cannulaConformance verified through testing.
DIN EN 1707: 6% Luer cone connections for injection cannula and particular medical equipmentConformance verified through testing.
BS 4843: Single entry IV cannulaConformance verified through testing.
Material Conformance:
DIN 17442/ISO 9626: Steel for medical instrumentsConformance verified through testing. PAJUNK Unipolar needles (including physical dimensions, coating, connector, tubing, metal, and plastics) were previously cleared under K000722, indicating material conformance. The Touhy needle's tip configuration is identical to a predicate cleared under K813186.
DIN EN/ISO 9626: Cannula tube of non-rusting steel (SS) for the manufacture of medical productsConformance verified through testing.
DIN 17440: Stainless SteelsConformance verified through testing.
Sterilization & Packaging Conformance:
DIN EN 550: Sterilization of med. Prod.; Validation & routine controls for sterilization with ETOConformance verified through testing. The contract sterilizer and sterilizing process are identical to those used for previously cleared Pajunk Plexolong sets (K013041), implying validated ETO sterilization.
DIN EN 556: Sterilization of medical products, requirements for medical products that are labeled "sterile"Conformance verified through testing.
PrEN 868-1: Packaging materials for the sterilization of packaged goods. Part 1: general requirements for the validation of the packaging of sterilized end-packaged productsConformance verified through testing. Packaging materials are the same as previously cleared Pajunk Plexolong sets, implying validated packaging for sterility.
DIN EN 868-2: Packaging materials for the sterilization of packaged goods. Part 2: sterilization packaging, requirements and tests.Conformance verified through testing.
Quality System & Risk Management Conformance:
DIN EN 724: Guidance on the application of EN29001 and EN46001 for non-active medical productsConformance verified through testing.
DIN EN 980: Graphic symbols for marking medical productsConformance verified through testing.
DIN EN 1441: Risk analysis for medical productsConformance verified through testing.
DIN en 30993-1: Biological evaluation of medical products - instructions for selection of testsConformance verified through testing.
DIN EN 46001: Particular requirements for medical productsConformance verified through testing.

Regarding AI/Machine Learning Specific Questions:

The provided document describes a conventional medical device (anesthesia conduction kit) and does not refer to any AI or machine learning components. Therefore, the following AI/ML-specific questions are not applicable to this submission.


2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Not applicable. This document describes a physical medical device, not an AI/ML algorithm or software that uses test data in the conventional sense. Performance testing was conducted on samples of the physical device according to industry standards.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Not applicable. Ground truth in the context of expert consensus is not relevant for this physical device. Performance is assessed against engineering and material specifications as defined by consensus standards.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Not applicable. This relates to AI/ML model performance evaluation, not a physical medical device.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not applicable. This device does not involve human "readers" or AI assistance.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not applicable. There is no algorithm or software component described.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • For this device, "ground truth" is established by the specifications outlined in the referenced consensus standards. Device components (e.g., Luer fittings, cannula dimensions, material properties, sterility assurance) are tested against these published, accepted industry standards.

8. The sample size for the training set

  • Not applicable. This device does not involve a training set for an AI/ML model.

9. How the ground truth for the training set was established

  • Not applicable. This device does not involve a training set for an AI/ML model.

{0}------------------------------------------------

510(k) Summary of Safety and Effectiveness

Submitter Information:

PAJUNK GmbH AM Holzplatz 5-7 78187 Geisingen Germany

USA Contact:Mr. Burk A. Brandt CE Consultancy, Inc. 5010 NW Crescent Vly Dr Corvallis, OR 97330 - USA

Phone: (541) 752-3953 Fax : (541) 752-3972

Device Name:

Trade Name: Plexolong Sets Common Name: Anesthesia Conduction Kit Classification Name: Anesthesia Conduction Kit (Reference, 21CFR, 868.5140, April 1, 2000)

Predicate Device:

These Plexus Anesthesia sets consist of a Paiunk Touby needle (with nerve stimulus connector and Luer lock), tubing (with Luer locks)}, catheter (with Teflon coated stylet), catheter adapter.and filter. The set, including the packaging is identical to the Plexolong Sets cleared for market under 510(k) number K013041. except for the Touhy needles with nerve stimulus connector. The Touhy needle is identical to the Unipolar needle, included in these sets, except for the needle tip. The Touhy needle tip configuration is same as the B Braun Touhy needle, cleared for market under 510(k) number K813186 as referenced in their submission number K013610.

The intended use statement, except for a reference to the Tuohy needle, is identical to the intended use statement in K013041. Except for the shape of the needle tip, the materials, construction and manufacturing processes are also identical to the cleared Pajunk Anesthesia Sets.

The contract sterilizer and sterilizing process are identical to those used for the Pajunk Plexolong sets. The packaging materials are also the same as those used to package the Pajunk previously cleared Plexolong sets.

Device Description:

The PAJUNK Plexolong sets contain single use sterile and non-pyrogenic needles with tubing. with or without a plastic cannula and catheters intended for delivery of continuous conduction anesthesia of peripheral nerves and plexus for up to 72 hours. Continuous delivery is accomplished using the catheter. An electrical stimulus may be applied to the conduction needle via a cable and connector to assist the physician pinpoint the area of application.

Intended Use:

The PAJUNK Plexolong sets consist of Pajunk Unipolar conduction needle with a plastic cannula placed over the needle, or a Tuohy conduction needle. They are intended for delivery of continuous conduction anesthesia of peripheral nerves and plexus for up to 72 hours. Continuous delivery for up to 72 hours is accomplished using the Polyamide indwelling catheter. An electrical stimulus may be applied to the conduction needle via a cable and connector to assist the physician pinpoint the area of application.

{1}------------------------------------------------

Warning:

The Pajunk GmbH needles and puncture sets are not intended for RF ablation or any other type of ablation procedure.

Technology Characteristics:

The PAJUNK Unipolar needles, which include the physical dimensions, coating, connector, tubing, metal and plastics, have been cleared under 510(k) number K000722. The Touhy needle is identical to the Unipolar needle, included in these sets, except for the shape of the needle tip. The Touhy needle tip configuration is same as the B Braun Touhy needle, cleared for market under 510(k) number K813186 as referenced in their submission number K013610. The Plexolong Sets are supplied in sealed polypropylene containers or polypropylene and styrol paper envelopes that are sealed to assure sterility.

{2}------------------------------------------------

Summary of Performance Testing:

The PAJUNK Plexolong Sets were designed to conform to the applicable sections of the following recognized consensus standards. The testing included verifying conformance to these standards.

StandardIssue DateTitle
DIN 13090/ISO 59408.1984Luer fittings w/wo locking feature
DIN 13097 Part 101.1980Medical injection cannula
DIN 13097 Part 311.1979Medical cannula
DIN 17442/ISO 962610.1977Steel for medical instruments
DIN EN 55007.1993Sterilization of med. Prod.; Validation & routine controls forsterilization with ETO
DIN EN 55601.1995Sterilization of medical products, requirements for medicalproducts that are labeled "sterile"
DIN EN 72412.1994Guidance on the application of EN29001 and EN46001 fornon-active medical products
PrEN 868-110.1996Packaging materials for the sterilization of packaged goods.Part 1: general requirements for the validation of thepackaging of sterilized end-packaged products
DIN EN 868-203.1993Packaging materials for the sterilization of packaged goods.Part 2: sterilization packaging, requirements and tests.
DIN EN 98008.1996Graphic symbols for marking medical products
DIN EN 144108.1994Risk analysis for medical products
DIN EN 170701.19976% Luer cone connections for injection cannula andparticular medical equipment
DIN EN/ISO 962606.1995Cannula tube of non-rusting steel (SS) for the manufacture ofmedical products
DIN en 30993-112.1994Biological evaluation of medical products - instructions forselection of tests
DIN EN 4600112.1993Particular requirements for medical products
DIN 1744009.1996Stainless Steels
BS 4843Single entry IV cannula

Conclusion:

The PAJUNK Plexolong Sets are as safe and effective as the predicate devices when used according to the instructions in the directions for use supplied with the devices.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract image of an eagle or bird-like figure, with stylized lines representing its body and wings.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAR 2 4 2003

PAJUNK GmbH C/O Mr. Burk A. Brandt CE Consultancy, Incorporated 5010 NW Crescent Valley Drive Corvallis, Oregon 97330

Re: K023218

Trade/Device Name: Pajunk Plexolong Sets Regulation Number: 868.5140 Regulation Name: Anesthesia Conduction Kit Regulatory Class: II Product Code: CAZ, BSP Dated: February 25, 2003 Received: February 27, 2003

Dear Mr. Brandt:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Mr. Brandt

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely vours.

Susan Runner

Susan Runner, DDS, MA Interim Director Division of Anesthesiology. General Hospital. Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

INDICATIONS FOR USE

510(k) Number: K023218________________________________________________________________________________________________________________________________________________

Device Name: Pajunk Plexolong Sets

Indications for use:

The Pajunk Plexolong sets are intended for delivery of continuous conduction anesthesia of peripheral nerves and plexus for up to 72 hours. Continuous delivery for up to 72 hours is accomplished using the Polyamide indwelling catheter. An electrical stimulus may be applied to the conduction needle via a cable and connectors to assist the physician pinpoint the area of application.

Warning:

The Pajunk GmbH needles and puncture sets are not intended for RF ablation or any other type of ablation procedure.

(Please do not write below this line - continue on another page if needed)

Concurrence of CDRH, Office of Device Evaluation

V Prescription Use or Over-The- Counter _________________________________________________________________________________________________________________________________________________________ (Per 21 CFR 801.109)

S.H. Ward

nesthesiology. General Hospital.

510(k) Number:

§ 868.5150 Anesthesia conduction needle.

(a)
Identification. An anesthesia conduction needle is a device used to inject local anesthetics into a patient to provide regional anesthesia.(b)
Classification. Class II (performance standards).