K Number
K984063

Validate with FDA (Live)

Manufacturer
Date Cleared
1999-02-09

(85 days)

Product Code
Regulation Number
880.5440
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Paragon Infusion Kit is intended to provide continuous infusion a local anesthetic directly into an intraoperative (soft tissue / body cavity) site for general surgery for postoperative pain management. Additional routes of administration include percutaneous, subcutaneous, intramuscular and epidural infusion. The Paragon pump is re-usable. The disposable Paragon administration set is single patient use only. No testing has been conducted to determine the efficacy of the Paragon for the delivery of blood, blood products, lipids or fat emulsions. The Paragon is not intended for the delivery of blood, blood products, lipids or fat emulsions. The Paragon is suitable for use as an ambulatory device and is intended for use in the hospital, home environment or alternative care sites.

Device Description

The Paragon Infusion Kit is identical to the I-Flow PainBuster Infusion Kit with the exception of the Paragon pump and administration set replacing the PainBuster pump. The kit is comprised of a Paragon pump and administration set (K923875) and various kit components such as catheter, needle, syringe, Y adapter, dressing, tape, gauze and carry case. The Paragon administration set is intended to attach to the kit catheter at the distal end of the set to provide continuous infusion of a local anesthetic directly into the intraoperative site for general surgery for postoperative pain management. The Paragon administration set is a disposable device intended for single patient use. The Paragon pump is reusable. The Paragon is suitable for use as an ambulatory device and is intended for use in the hospital, home environment or alternative care sites.

AI/ML Overview

The provided text describes the "Paragon Infusion Kit" and includes a summary of safety and effectiveness, as well as operational specifications and performance data. The device is an infusion pump kit intended for continuous infusion of local anesthetics for postoperative pain management.

Based on the document, here's an analysis of the acceptance criteria and study information:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria for Flow Rate Accuracy: ±10% at 95% confidence interval (as stated in Section 3.1)

ModelTargeted Flow Rate (ml/hr)Reported Average Flow Rate (ml/hr)Standard DeviationNumber of Samples (N)Achieved Accuracy (within ±10%)
100 ml x 0.5 ml/hr0.50.530.015Yes (0.45 to 0.55 ml/hr)
100 ml x 1.0 ml/hr1.00.990.035Yes (0.90 to 1.10 ml/hr)
100 ml x 2.0 ml/hr2.01.990.075Yes (1.80 to 2.20 ml/hr)
100 ml x 4.0 ml/hr4.04.140.105Yes (3.60 to 4.40 ml/hr)
100 ml x 10.0 ml/hr10.010.300.3810Yes (9.0 to 11.0 ml/hr)

Additional Acceptance Criteria / Operational Specifications (Section 3.1):

  • Priming/Residual Volume: < 5 ml
  • Operating Temperature: 31°C skin temperature (90°F)
  • Test Solution: 0.9% NaCl
  • Operating Pressure: 6.0 psi pressure source
  • Head Height: 0"

The document states: "All models produced an average flow rate within the ±10% accuracy claim." (Section 3.2)

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size:
    • For 100 ml x 0.5 ml/hr models: 5 pieces
    • For 100 ml x 1.0 ml/hr models: 5 pieces
    • For 100 ml x 2.0 ml/hr models: 5 pieces
    • For 100 ml x 4.0 ml/hr models: 5 pieces
    • For 100 ml x 10.0 ml/hr models: 10 pieces
  • Data Provenance: The document does not specify the country of origin for the data or whether the study was retrospective or prospective. It appears to be an internal laboratory performance test conducted under specified "standard operating conditions" (Section 3.1).

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

  • Not applicable. The "ground truth" for this performance study is a physical measurement of flow rate against a specified target. This is a laboratory test, not an assessment requiring expert consensus on clinical data.

4. Adjudication Method for the Test Set

  • Not applicable. This was a direct measurement of physical performance (flow rate), not a judgment-based assessment requiring adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

  • No. An MRMC study is relevant for diagnostic or clinical interpretation devices where human readers assess cases, often with and without AI assistance, to determine changes in diagnostic accuracy or efficiency. This document describes a medical device (infusion pump) and its physical performance, not a diagnostic algorithm.

6. If a Standalone Study (i.e., algorithm only without human-in-the-loop performance) Was Done

  • Yes, in spirit. The testing presented is entirely standalone performance for the device (infusion pump). There is no "algorithm" in the sense of AI or software interpreting data for diagnosis, but rather the mechanical performance of the pump itself. The measurements are taken directly from the device's output.

7. The Type of Ground Truth Used

  • Physical Measurement/Engineered Specification: The ground truth is the targeted flow rate (e.g., 0.5 ml/hr, 1.0 ml/hr), and the device's performance is measured against this specification under controlled laboratory conditions. The "Test Solution" was 0.9% NaCl, and "Operating Pressure" was 6.0 psi, along with other specified conditions (Section 3.1).

8. The Sample Size for the Training Set

  • Not applicable. The Paragon Infusion Kit is a mechanical device, not an AI or machine learning model. Therefore, there is no "training set" in the context of data-driven model development. The design and engineering of the pump and administration set are based on physical principles (e.g., Poiseuille's equation - Section 2.3.3.1).

9. How the Ground Truth for the Training Set Was Established

  • Not applicable. As explained above, this is a mechanical device and does not have a training set or ground truth established in the typical sense of AI/ML. Its performance is designed and validated through engineering principles and physical testing.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows a date of 2/9/99 at the top left. To the right of the date is a logo for I-FLOW, which is a company name listed below the logo. The company name is I-FLOW CORPORATION.

20202 Windrow Drive Lake Forest. CA 92630 949) 206-2700 800) 443-3569 Fax (949) 206-2600

K9840603

SUMMARY OF SAFETY AND EFFECTIVENESS

r

November 11, 1998

Trade Name: Paragon Infusion Kit

Common Name: Infusion Pump Kit

Classification Name: Pump, Infusion

All questions and/or comments concerning this document should be made to:

Robert J. Bard, Esq., R.A.C. Vice President of Regulatory and Legal Affairs

I-Flow Corporation 20202 Windrow Drive Lake Forest, CA 92630

Telephone: 949.206.2700 Fax: 949.206.2600

{1}------------------------------------------------

GENERAL INFORMATION 1.0

1.1 Purpose of Submission

  • This submission is intended to notify the Federal Food and Drug 1.1.1 Administration that I-Flow Corporation intends to market a new intended use for the Paragon Infusion System (K923875), originally identified as the SideKick 50 Plus and SideKick 100 Plus. The Paragon will be marketed as a kit, the Paragon Infusion Kit, including labeling changes and additional components.
  • Trade Name: Paragon Infusion Kit 1.1.2
  • 1.1.3 Common Name: Infusion Pump Kit
  • Classification Name: Pump, Infusion 1.1.4
  • Classification Panel: General Hospital and Personal Use Device 1.1.5

1.2 Statement of Equivalence

  • The Paragon Infusion Kit includes components that are legally marketed 1.2.1 (either pre-amendment devices or devices that have been granted permission to market via premarket notification regulation).
  • The Paragon pump and administration set are the same as used in K923875. 1.2.2
  • The Paragon Kit is substantially equivalent to the I-Flow Paragon Infusion 1.2.3 System (K923875), the I-Flow PainBuster Infusion Kit (K980558, K982946), the Sgarlato Pain Control Infusion Pump (PCIP) (K896422) and the I-Flow Homepump C-Series (K944692).

PHYSICAL SPECIFICATIONS AND DESCRIPTIONS 2.0

Description of the Paragon Infusion Kit 2.1

  • The Paragon Infusion Kit is identical to the I-Flow PainBuster Infusion Kit with 2.1.1 the exception of the Paragon pump and administration set replacing the PainBuster pump.
  • The kit is comprised of a Paragon pump and administration set (K923875) 2.1.2 and various kit components such as catheter, needle, syringe, Y adapter, dressing, tape, gauze and carry case.
    • The PainBuster kit contains all the above components except for 2.1.2.1 an elastomeric pump with integrated administration set instead of the Paragon pump and administration set.
  • The Paragon administration set is intended to attach to the kit catheter at the 2.1.3 distal end of the set to provide continuous infusion of a local anesthetic directly into the intraoperative site for general surgery for postoperative pain management.
  • 2.1.4 The Paragon administration set is a disposable device intended for single patient use. The Paragon pump is reusable.
  • 2.1.5 The Paragon is suitable for use as an ambulatory device and is intended for use in the hospital, home environment or alternative care sites.

{2}------------------------------------------------

Description of Paragon Pump 2.2

  • The Paragon pump consists of two cylindrical shells. The top half of the 2.2.1 pump has internal threads which mate to the external threads of the bottom half of the pump.
  • The top incorporates a pressure plate which applies a load to the pliable drug 2.2.2 bag. The load is applied to the drug bag by way of two opposing springs which act upon a scissors mechanism. The spring/scissors mechanism creates a near constant pressure in the drug bag.
  • The bottom half of the pump is slotted to allow for positioning of the 2.2.3 administration set.
  • When the top and bottom halves of the pump are fully threaded together, the 2.2.4 pressure plate contacts the drug bag and acts as the pressurizing element.
  • This premarket notification proposes no changes to the Paragon pump. 2.2.5

2.3 Description of Paragon Administration Set

  • The Paragon administration set consists of a PVC drug bag attached to the 2.3.1 administration line.
  • 2.3.2 Each administration set consists of fixed diameter flow control tubing or glass orifice.
  • The flow control tubing or glass orifice is cut to a specific length L. When the 2.3.3 PVC drug bag is pressured by the Paragon pump, the delivery times are defined by the inside diameter of the flow control orifice.
    • 2.3.3.1 The delivery time characteristic is derived from the flow rate of the device which is in turn approximated by Poiseulle's equation:

$$\mathcal{Q} = \frac{\Delta \mathcal{\rho} \pi D^4}{12 \mathfrak{B} \mu \mathcal{L}}$$

  • 2.3.3.2 Where () is the flow rate, & is the pressure drop across the orifice, D is the inside diameter of the flow controlling orifice, u is the dynamic viscosity of the fluid and L is the length of the orifice. The equation provides an approximation of the actual delivery time.

2.4 Product Configuration

  • 2.4.1 The Paragon pump:
    • PG100000P: 100 ml volume 2.4.1.1
  • The Paragon administration sets: 2.4.2
    • 2.4.2.1 PG100005: 100 ml volume, 0.5 ml/hr flow rate
    • 2.4.2.2 PG100010: 100 ml volume, 1.0 ml/hr flow rate
    • 2.4.2.3 PG100020: 100 ml volume, 2.0 ml/hr flow rate
    • PG100040: 100 ml volume, 4.0 ml/hr flow rate 2.4.2.4
    • 2.4.2.5 PG100100: 100 ml volume, 10.0 ml/hr flow rate
    • 2.4.2.6 PG100020Y: 100 ml volume, 2.0 ml/hr flow rate, dual orifice, dual catheter with Y adapter

{3}------------------------------------------------

  • 2.4.2.6.1 The dual orifice set consists of a standard Paragon set with dual orifice downstream from the Y adapter. Each orifice allows 2 ml/hr flow rate.
  • Each model consists of a kit with the following components: 2.4.3
    • 2.4.3.1 Paragon pump (optional).
      • The reusable Paragon pump may be packaged and 2.4.3.1.1 sold separately from the disposable kit components.
    • Paragon administration set. 2.4.3.2
    • 2.4.3.3 Catheter:
      • 18 to 22 G catheter, 11 to 40 in. length, polyamide, 2.4.3.3.1 nylon, FEP (fluorinated ethylene propylene) polymer, silicone, polyurethane or Teflon.
      • A catheter connector is included to connect the 2.4.3.3.2 catheter to the distal luer of the administration set.
      • The B. Braun Perifix® Epidural Catheter Set is an 2.4.3.3.3 example of the type of catheter that may be used with the Paragon Infusion Kit.
        • 2.4.3.3.3.1 Product code: EC20-0.
        • 510(k) number: K813186. 2.4.3.3.3.2
    • 2.4.3.4 Needle:
      • 2.4.3.4.1 14 to 18 G. 1 ½ to 3 ¼ in. length, stainless steel.
      • 2.4.3.4.2 The needle may be a catheter over needle as in the Angiocath™ example below.
      • 2.4.3.4.3 The Angiocath catheter introducer needle is an example of the type of catheter introducer needle that may be used with the Paragon Infusion Kit.
        • 2.4.3.4.3.1 Product code: 382258.
      • Syringe (optional): 2.4.3.5
        • 2.4.3.5.1 60 cc plastic, luer lock syringe.
        • The syringe is used to fill the Paragon drug bag with 2.4.3.5.2 medication.
        • The B-D 60 cc syringe is an example of the type of 2.4.3.5.3 syringe that may be included in the Paragon Kit.
          • 2.4.3.5.3.1 Product Code: 309663.
    • 2.4.3.6 Dressing (optional):
      • 2.4.3.6.1 The dressing is used to hold the catheter and/or flow restrictor in place.
      • The OpSite™ is an example of the type of dressing 2.4.3.6.2 that may be used in the Paragon Kit.
        • 2.4.3.6.2.1 Product Code: 4973.

{4}------------------------------------------------

  • Carry Case (optional): 2.4.3.7
    • 2.4.3.7.1 The carry case is used to hold the Paragon pump while delivering medication.

2.4.3.7.1.1 I-Flow part numbers 1400749, 1400752 or 1400758.

Antiseptic Skin Swabs (optional): 2.4.3.8

  • The antiseptic skin swabs are used to prep the skin 2.4.3.8.1 area of the patient prior to inserting the catheter.
  • 2.4.3.8.2 The Alcohol Prep Pads or lodophor PVP Scrub Swabsticks manufactured by Clinipad Corporation are examples of the type of antiseptic skin swabs that may be used in the Paragon Kit.
2.4.3.8.2.1Model number: CL0110.
------------------------------------

2.4.3.8.2.2 Model number: CL1244.

  • 2.4.3.9 Tape (optional):
    • 2.4.3.9.1 The tape may be used to the secure catheter, flow control tubing or gauze.
    • The Transpore™ Surgical Tape manufactured by 3M 2.4.3.9.2 is an example of the type of tape that may be used in the Paragon Kit.

2.4.3.9.2.1 Model number: 3M1527-1.

  • 2.4.3.10 Gauze (optional):
    • The gauze may be used to secure the catheter or 2.4.3.10.1 flow control tubing.
    • The Kling® Conforming Gauze manufactured by 2.4.3.10.2 Johnson and Johnson is an example of the type of gauze which may be used in the Paragon Kit.

2.4.3.10.2.1 Model number: JJ6923.

  • 2.4.3.11 Y Adapter (optional)
    • 2.4.3.11.1 The Y adapter is used for an additional catheter for a large wound or multiple wound sites.
    • 2.4.3.11.2 The Flexible Y Fitting by Qosina is an example of the type of Y Adapter that may be used in the Paragon Kit.

2.4.3.11.2.1 Model number: 81120.

{5}------------------------------------------------

Components and Materials 2.5

The pump and administration sets used in the Paragon Infusion Kit are currently available models of the Paragon Infusion System (K923875). No changes will be made to the Paragon pump or administration sets.

All kit components other than the pump and administration set are identical to those used in the PainBuster infusion system submitted under K980558, K982946.

The Paragon administration set is a disposable device intended for single patient use. The Paragon pump is reusable.

2.6 Power Requirements

  • The Paragon pump is a mechanical pump that utilizes spring energy for 2.6.1 power. No additional external power source is required.

OPERATIONAL SPECIFICATIONS AND DESCRIPTIONS 3.0

Standard Operating Conditions: 3.1

Priming/Residual Volume:< 5 ml
Operating Temperature:31°C skin temperature (90°F)
Test Solution:0.9% NaCl
Operating Pressure:6.0 psi pressure source
Head Height:0"
Accuracy:±10% at 95% confidence interval
  • 3.2 Flow Rate Performance Data: Testing occurred at standard operating conditions. All models produced an average flow rate within the ±10% accuracy claim.
100 ml x 0.5ml/hr100 ml x 1ml/hr100 ml x 2ml/hr100 ml x 4ml/hr100 ml x 10ml/hr
Average Flow Rate (ml/hr)0.530.991.994.1410.30
Std. Dev.0.010.030.070.100.38
N555510

100 ml x 0.5 ml/hr: A five (5) piece sample produced an average flow rate of 0.53 ml/hr. The resulting average is well within its ±10% accuracy claim. The fastest infusion had an average flow rate of 0.54 ml/hr and the slowest infusion had an average flow rate of 0.51 mi/hr.

100 ml x 1.0 ml/hr: A five (5) piece sample produced an average flow rate of 0.99 ml/hr. The resulting average is within its ±10% accuracy claim. The fastest infusion had an average flow rate of 1.04 ml/hr and the slowest infusion had an average flow rate of 0.96 ml/hr.

100 ml x 2.0 ml/hr: A five (5) piece sample produced an average flow rate of 1.99 ml/hr. The resulting average is within its ±10% accuracy claim. The fastest infusion had an average flow rate of 2.05 ml/hr and the slowest infusion had an average flow rate of 1.90 ml/hr.

100 ml x 4.0 ml/hr: A five (5) piece sample produced an average flow rate of 4.14 ml/hr. The resulting average is within its ±10% accuracy claim. The fastest infusion had an average flow rate of 4.24 ml/hr and the slowest infusion had an average flow rate of 3.99 ml/hr.

{6}------------------------------------------------

100 ml x 10.0 ml/hr: A ten (10) piece sample produced an average flow rate of 10.30 ml/hr. The resulting average is within its ±10% accuracy claim. The fastest infusion had an average flow rate of 10.80 ml/hr and the slowest infusion had an average flow rate of 9.65 ml/hr.

Back Pressure (Head Height) Comparison: Approximately 0.57 psi pressure difference results per 16" head height. Thus, a 10% flow rate change may occur for each 16" head height difference from nominal assuming a 6 psi pressure source.

  • Drug Delivery Comparison: Local anesthetics have densities similar to normal saline 3.3 (e.g. 1.002 to 1.005 for Ropivacaine HCI vs. 1.0045 for normal saline) and should not affect flow rate. Product labeling includes a statement as to delivery times and the possible deviation from nominal due to drug viscosity.

Safety / Alarm Functions 3.4

  • The Paragon pump and administration set provide a continuous fixed flow 3.4.1 and as such is not subject to fluid runaway conditions similar to that of some electronic pumps.
  • The Paragon pump will not be recommended for any application that exceeds 3.4.2 the minimum internal pressure of the system.
  • If for any reason the patient needs to stop his or her infusions, each 3.4.3 administration set is supplied with a pinch clamp to stop the infusion.
  • This device contains no alarms or indicators for flow other than visual. 3.4.4
  • 3.4.5 This device contains no alarms or indicators to detect air in line or an occlusion; however, each set may include an integrated air-eliminating filter.

BIOLOGICAL SPECIFICATIONS 4.0

  • Biological testing is in conformance with ISO 10993 Part 1 for all fluid path 4.1 components of the Paragon administration set.

5.0 CHEMICAL AND DRUG SPECIFICATIONS

  • 5.1 Compatibility
    • There are no specific drugs referenced in the labeling for the Paragon 5.1.1 Infusion Kit.
    • The Paragon Infusion Kit is intended for use with general local anesthetics 5.1.2 and epidural medications.
  • 5.2 Drug Stability
    • 5.2.1 There are no drugs included in the Paragon Infusion Kit.

6.0 INTENDED USE

  • 6.1 The Paragon Infusion Kit is intended to provide continuous infusion a local anesthetic directly into an intraoperative (soft tissue / body cavity) site for general surgery for postoperative pain management.
  • 6.2 Additional routes of administration include percutaneous, subcutaneous, intramuscular and epidural infusion.
  • The Paragon pump is re-usable. The disposable Paragon administration set is single 6.3 patient use only.

{7}------------------------------------------------

  • No testing has been conducted to determine the efficacy of the Paragon for the 6.4 delivery of blood, blood products, lipids or fat emulsions. The Paragon is not intended for the delivery of blood, blood products, lipids or fat emulsions.
  • The Paragon is suitable for use as an ambulatory device and is intended for use in the 6.5 hospital, home environment or alternative care sites.

7.0 PACKAGING

  • 7.1 The Paragon Kit consists of an inner pouch or tray with Tyvek lid stock surrounded by a header bag with an ETO Tyvek strip.
  • 7.2 The Paragon administration set may be packaged in either a Tyvek pouch or Form/Fill/Seal.
  • 7.3 The Paragon Kit components are placed in the inner tray or pouch.
  • Packaging is suitable for either radiation or ETO sterilization. 7.4

8.0 STERILIZATION INFORMATION

Note: The kit components of the Paragon Infusion Kit may be purchased non-sterile and packaged by I-Flow or sterile from the manufacture. The Paragon administration set and non-sterile purchased components shall be sterilized as follows:

  • The methods of sterilization are gamma radiation (Cobalt 60) or ETO gas. 8.1

9.0 COMPARISON TO LEGALLY MARKETED DEVICES

See Table 1 that follows this section for more specific information.

  • 9.1 Intended Use
    • 9.1.1 The Paragon Infusion Kit, the PainBuster Infusion Kit and the Sgarlato Pain Control Infusion Pump (PCIP) have the same intended use:
      • 9.1.1.1 To provide continuous infusion of a local anesthetic directly into the intraoperative site for general surgery for postoperative pain management.
    • 9.1.2 The predicate Paragon Infusion System and Homepump C-Series are intended for general infusion use, including chemotherapy and pain management.
  • 9.2 Device Descriptions
    • 9.2.1 The Paragon Infusion Kit
      • 9.2.1.1 The Paragon Infusion Kit is identical to the predicate PainBuster Infusion Kit with the exception of the Paragon pump and administration set replacing the PainBuster pump.
      • 9.2.1.2 The Paragon Infusion Kit uses the same mechanical, reusable Paragon pump as the predicate Paragon Infusion System.
      • 9.2.1.3 The Paragon Infusion Kit uses the same Paragon Administration Set as the Paragon Infusion System.
    • 9.2.2 The Paragon Infusion System
      • 9.2.2.1 The Paragon Infusion Kit uses the same pump and administration sets as the Paragon Infusion System (K923875).

{8}------------------------------------------------

  • The Sqarlato Pain Control Infusion Pump (PCIP) 9.2.3
    • The Sgarlato PCIP consists of a kit very similar to the Paragon 9.2.3.1 and PainBuster Infusion Kit. These kits consist of an infusion pump and administration set, catheter, needle, syringe, Y adapter, carry case, dressing, tape and gauze.
    • The Sgarlato kit uses a disposable, spring driven syringe pump 9.2.3.2 with integrated administration set.
  • The Homepump C-Series 9.2.4
    • 9.2.4.1 The Homepump C-Series consists of a disposable, elastomeric infusion pump with integrated administration set.
  • 9.2.5 Specifications
    • The Paragon Infusion Kit, PainBuster Infusion Kit and Sgarlato 9.2.5.1 PCIP have similar fill volumes and flow rates, see Table 1.
  • 9.2.6 Flow Control
    • The Paragon Infusion Kit and all its predicate devices use either a 9.2.6.1 glass orifice or PVC tubing to control the flow rate.
  • 9.2.7 Materials
    • The Paragon Infusion Kit uses the same Paragon Administration 9.2.7.1 Set as the Paragon Infusion System. All fluid path materials of the Paragon Administration Set are in conformance with ISO 10993 Part 1.
  • Based upon the data presented in this section 9.0 and Table 1, I-Flow 9.2.8 Corporation has determined that the Paragon Infusion Kit is substantially equivalent to the named predicate devices.

{9}------------------------------------------------

ComparisonElementParagon Infusion Kit(subject device)SE' Paragon Infusion System(K923875)SE' PainBuster Infusion Kit(K980558, K982946)SE¹ Sgarlato PCIP(K896422)SE' Homepump C-Series(K944692)
Intended UseTo provide continuous infusionof a local anesthetic directlyinto the intraoperative site forgeneral surgery forpostoperative painmanagement.General infusion use, includingchemotherapy and painmanagement.To provide continuous infusionof a local anesthetic directlyinto the intraoperative site forpostoperative painmanagement.To provide continuous infusionof a local anesthetic directlyinto the surgical wound site forpostoperative painmanagement.General infusion use, includingchemotherapy and painmanagement.
Routes ofAdministrationPercutaneous, subcutaneous,intramuscular and epiduralIntravenousPercutaneous andsubcutaneousPercutaneous, subcutaneousand epiduralIntravenous, intra-arterial,epidural or subcutaneous
ContraindicationsNot intended for intravenous orintra-arterial delivery. Notintended for delivery of blood,blood products, lipids or fatemulsions.Not intended for delivery ofblood, blood products, lipids orfat emulsions.Not intended for intravenous,intra-arterial or epiduraldelivery. Not intended fordelivery of blood, bloodproducts, lipids or fatemulsions.Not intended for rapid infusions.Not intended for intravenousinfusion.Not intended for delivery ofblood, blood products, lipids orfat emulsions.
ReuseCapabilityRe-usable pump, single patientuse disposable administrationsetRe-usable pump, single patientuse disposable administrationsetDisposable, single patient useDisposable, single patient useDisposable, single patient use
DescriptionSold empty and capable ofbeing filled via a fill port.Sold empty and capable ofbeing filled via a fill port.Sold empty and capable ofbeing filled via a fill port.Sold empty and capable ofbeing filled via a fill port.Sold empty and capable ofbeing filled via a fill port.
Fill Volumes100 ml100 ml50 to 270 ml50 to 100 ml50 to 500 ml
Flow Rates0.5, 1.0, 2.0, 4.0 or 10.0 ml/hr0.5 to 200 ml/hr0.5, 1.0, 2.0, 5.0 or 10.0 ml/hr0.5, 1.0 or 2.0 ml/hr0.5 to 500 ml/hr
Pump TypeMechanical springMechanical springElastomeric PumpSpring driven syringe pumpElastomeric Pump
PowerRequirementsNoneNoneNoneNoneNone
Pump MechanismConstant pressure is applied tothe fluid reservoir.Constant pressure is applied tothe fluid reservoir.Constant pressure is applied tothe fluid reservoir.Constant pressure is applied tothe fluid reservoir.Constant pressure is applied tothe fluid reservoir.
Pressure SourceMechanical spring energyMechanical spring energyStrain energy of elastomericmembranesMechanical spring energyStrain energy of elastomericmembranes
Fluid ReservoirPVC drug bagPVC drug bagThermoplastic (Krayton)elastomeric membranePolypropylene plastic syringeThermoplastic (Krayton)elastomeric membrane
Administration Set
Flow ControlConsistent flow rate throughoutthe entire course of therapy isachieved by the combination ofconstant pressure and flowcontrol tubing.Consistent flow rate throughoutthe entire course of therapy isachieved by the combination ofconstant pressure and flowcontrol tubing.Consistent flow rate throughoutthe entire course of therapy isachieved by the combination ofconstant pressure and flowcontrol tubing.Consistent flow rate throughoutthe entire course of therapy isachieved by the combination ofconstant pressure and flowcontrol tubing.Consistent flow rate throughoutthe entire course of therapy isachieved by the combination ofconstant pressure and flowcontrol tubing.
Safety / AlarmFunctionsFixed flow rate tubing preventsfluid runaway conditions. Eachadministration set is suppliedwith a clamp to stop theinfusion if necessary.Fixed flow rate tubing preventsfluid runaway conditions. Eachadministration set is suppliedwith a clamp to stop theinfusion if necessary.Fixed flow rate tubing preventsfluid runaway conditions. Eachadministration set is suppliedwith a clamp to stop theinfusion if necessary.Fixed flow rate tubing preventsfluid runaway conditions. Eachadministration set is suppliedwith a clamp to stop theinfusion if necessary.Fixed flow rate tubing preventsfluid runaway conditions. Eachadministration set is suppliedwith a clamp to stop theinfusion if necessary.

{10}------------------------------------------------

Image /page/10/Picture/1 description: The image is a black and white circular seal. The seal contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement around a central graphic. The graphic depicts a stylized image of an eagle or other bird with its wings spread, rendered in a simple, bold line style.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

FEB - 9 1999

Robert J. Bard, Esq., R.A.C. Vice President Regulatory and Legal Affairs I-Flow Corporation 20202 Window Drive Lake Forest, California 92630

K984063 Re : Paragon Basal/Bolus Administration Set Trade Name: Requlatory Class: II Product Code: FPA Dated: November 11, 1998 November 16, 1998 Received:

Dear Mr. Bard:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. ਰ substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) requlation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{11}------------------------------------------------

Page 2 - Mr. Bard

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA debtraouf substantial equivalence of your device to a legally marketed predicate device results in a classification for your markets produces permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact Also, please note the Office of Compliance at (301) 594-4639. the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Timothy A. Ulatowski

Director Division of Dental, Infection Control, and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{12}------------------------------------------------

510(k) Number (if known): K984063

Device Name: __ Paragon Basal/Bolus Administration Set

Indications for Use:

    1. The Paragon Basal/Bolus Administration Set is intended to provide a continuous, basal level infusion of medication and to allow patient controlled bolus delivery. The basal lover interest of the administration set enables fixed boluses of medication to be bolus componsible the patient or healthcare provider. The Paragon deliverou upon administration Set routes of administration include: intravenous, epidural, intramuscular and subcutaneous.
      (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUED ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)X
------------------------------------------------------------------------------

OR

Over-The-Counter Use
(Optional Format 1-2-96)
Signature
(Division Sign-Off)
Division of Dental, Infection Control,

and General Hospital Devices 510(k) Number_ 1

§ 880.5440 Intravascular administration set.

(a)
Identification. An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.(b)
Classification. Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.