Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K251097

    Validate with FDA (Live)

    Device Name
    V-DAC Catheter
    Manufacturer
    Date Cleared
    2026-01-06

    (271 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K211654

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The V-DAC Catheter is indicated for use with compatible guide catheters in facilitating the insertion and guidance of microcatheters into a selected blood vessel in the peripheral and neurovascular systems.

    Device Description

    The V-DAC Catheter consists of 1) Distal Access Catheter and 2) Peel-Away Introducer Sheath.

    The V-DAC Catheter is a single lumen, coil-reinforced, flexible, variable stiffness composite catheter. The catheter distal shaft has an external hydrophilic coating aimed at reducing friction during use. The distal end of the catheter shaft is radiopaque for fluoroscopic visualization, and the proximal end contains a luer hub that allows the insertion and guidance of microcatheters into a selected blood vessel in the neurovascular system. Complete dimensions of the catheter are included on the device label.

    The Peel-Away Introducer is an accessory provided with the catheter to aid in the delivery of the catheter.

    The V-DAC Catheter and a Peel-Away Introducer are packaged together, they are provided sterile, non-pyrogenic, and are intended for single use only.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K242319

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-09-03

    (29 days)

    Product Code
    Regulation Number
    870.5150
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K211654

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    INDIGO Aspiration Catheters and Separators: As part of the INDIGO Aspiration System, the INDIGO Aspiration Catheters and Separators are indicated for the removal of fresh, soft emboli and thrombi from vessels of the peripheral arterial and venous systems, and for the treatment of pulmonary embolism.

    INDIGO Aspiration Tubing: As part of the INDIGO Aspiration System, the INDIGO Sterile Aspiration Tubing is indicated to connect the INDIGO Aspiration Catheters to the Penumbra Aspiration Pump.

    Penumbra Aspiration Pump: The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems.

    Device Description

    The INDIGO® Aspiration System is comprised of the several devices: INDIGO Aspiration Catheter, Penumbra Aspiration Pump, INDIGO Aspiration Pump Canister, INDIGO Aspiration Tubing, INDIGO Separator. The INDIGO Aspiration System is designed to remove thrombus from the vasculature using mechanical aspiration. The INDIGO Aspiration Catheter targets aspiration from the pump directly to the thrombus. The INDIGO Separator may be used to clear the lumen of the INDIGO Aspiration Catheter should it become blocked with thrombus. The INDIGO Aspiration Catheter is introduced through a guide catheter or vascular sheath into the peripheral vasculature and guided over a guidewire to the site of the primary occlusion. TraX may be used with its prepackaged Aspiration Catheter to facilitate access to the primary occlusion. The INDIGO Aspiration Catheter is used with the Penumbra Aspiration Pump to aspirate thrombus from an occluded vessel. As needed, an INDIGO Separator may be deployed from the INDIGO Aspiration Catheter to assist with thrombus removal. The INDIGO Separator is advanced and retracted through the INDIGO Aspiration Catheter at the proximal margin of the primary occlusion to facilitate clearing of the thrombus from the INDIGO Aspiration Catheter tip. The devices are visible under fluoroscopy. For the aspiration source, the INDIGO Aspiration Catheter is used in conjunction with the Penumbra Aspiration Pump, which is connected using the INDIGO Aspiration Tubing and the INDIGO Aspiration Pump Canister. The INDIGO Aspiration Catheter may be provided with a steam shaping mandrel, rotating hemostasis valve, Select Catheter, introducer, and TraX. TraX compatibility is limited to its prepackaged Aspiration Catheter. The INDIGO Separator may be provided with an introducer and torque device.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification from the FDA for the Penumbra, Inc. Indigo® Aspiration System - Aspiration Catheter 6X. It details the device, its intended use, and its substantial equivalence to predicate devices based on non-clinical (bench-top) performance testing.

    1. A table of acceptance criteria and the reported device performance

    The document does not provide a specific table of acceptance criteria with corresponding performance metrics like sensitivity, specificity, or AUC, as it is related to a physical medical device (catheter) rather than a diagnostic algorithm or AI. Instead, it lists various bench-top performance tests conducted. For each test, the implicit acceptance criterion is that the device met all requirements.

    Test CategoryReported Device Performance (Acceptance Criteria)
    Particulate TestingMet all requirements
    Dimensional/Visual InspectionMet all requirements
    Friction TestingMet all requirements
    Performance/Simulated Use TestingMet all requirements
    Markerband VisibilityMet all requirements
    Tensile TestingMet all requirements
    Pressure TestingMet all requirements
    Elongation TestingMet all requirements
    Torsion TestingMet all requirements
    Hub Air AspirationMet all requirements
    Corrosion TestingMet all requirements
    Biocompatibility (ISO 10993-1)Met all requirements
    Shelf-LifeEstablished for 12 months based on accelerated aging data

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document primarily describes bench-top testing for a physical device. It does not mention a "test set" in the context of data used for an algorithm or AI. Therefore, concepts like sample size for a test set, country of origin, retrospective or prospective data are not applicable. The data provenance would be from internal laboratory testing conducted by Penumbra, Inc.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable as the document describes bench-top performance testing of a physical medical device (catheter), not a diagnostic algorithm or AI that would require expert-established ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable for the same reason as point 3.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    An MRMC comparative effectiveness study was not done. The study described is for a physical medical device, not an AI or diagnostic tool that would involve human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    A standalone performance study in the context of an algorithm or AI was not done. The performance evaluation was of the physical device's mechanical, material, and functional properties on a test bench.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the bench-top tests, the "ground truth" would be defined by engineering specifications, material science standards, and functional performance benchmarks for the device. For example, a tensile test would measure the force required to break the catheter, with the "ground truth" being the specified minimum tensile strength that the device must meet according to its design requirements. Similarly, biocompatibility is assessed against ISO 10993-1 standards.

    8. The sample size for the training set

    This information is not applicable as the document describes bench-top performance testing of a physical medical device, not an AI or algorithm that uses a training set.

    9. How the ground truth for the training set was established

    This information is not applicable for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K222786

    Validate with FDA (Live)

    Date Cleared
    2023-08-25

    (344 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K193380, K211654

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    072 Aspiration Catheter: As part of the 072 Aspiration System, the 072 Aspiration Catheter with a compatible suction pump is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA therapy are candidates for treatment.

    Aspiration Tubing: As part of the 072 Aspiration System, the Aspiration Tubing is intended to connect the 072 Aspiration Catheter to a compatible suction pump.

    Device Description

    The 072 Aspiration Catheter is a single lumen, variable stiffness catheter. The catheter has a hydrophilic coating to reduce friction during use. The catheter includes radiopaque markers on the distal end for angiographic visualization and a Luer hub on the proximal end allowing attachments for flushing and aspiration. The Delivery Tool is an optional accessory for use with the 072 Aspiration Catheter and should be removed prior to aspiration. The 072 Aspiration Catheter, Delivery Tool. Rotating Hemostasis Valve, and Flow Switch are included in the package. The Aspiration Tubing is provided in a separate package. For the aspiration source, the 072 Aspiration Catheter is used in conjunction with a compatible suction pump with prespecified performance parameters that is connected using the Aspiration Tubing.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Q'Apel Medical, Inc. 072 Aspiration System, based on the provided document:


    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are generally phrased as meeting "predetermined acceptance criteria" or specific ISO standards. The reported device performance for all bench tests and animal studies is "Pass," indicating that all samples met the defined criteria.

    Note: For bench tests, specific numerical acceptance criteria are not detailed in the provided summary, only the general statement that "All samples met the predetermined acceptance criteria." For biocompatibility, the specific numerical acceptance criteria are listed.

    Test CategoryAcceptance Criteria (as stated)Reported Device Performance
    Bench Testing
    Visual Surface RequirementDevice meets visual surface requirements.Pass
    Packaging Visual InspectionPackaging meets visual inspection.Pass
    Dimensional/Visual InspectionDevice dimensions conform to specifications.Pass
    Liquid Leakage Under PressureCatheter joint strength meets freedom from leakage requirements of ISO 10555-1:2013, Annex C.Pass
    Hub Aspiration Air LeakageDevice passes hub aspiration air leakage test of ISO 10555-1:2013, Annex D.Pass
    Simulated UseEffectiveness of device at retrieval of soft and firm clots and mechanical integrity after multiple passes.Pass
    Flex FatigueMeets minimum value per specification for multiple passes in the simulated use model.Pass
    Track and Advance ForceTest specimens meet track and advance force criteria.Pass
    Tip DeflectionTest specimens meet tip deflection forces criteria and compare favorably to predicate catheters.Pass
    TorqueNumber of revolutions to failure of the Catheter in simulated anatomy meets criteria.Pass
    Tip Elongation and CompressionTest specimens meet tip elongation and compression criteria.Pass
    Peak TensileTensile strength of catheter sections and bonds meets criteria after simulated use.Pass
    Particulates, Coating IntegrityIntegrity of hydrophilic coating evaluated after multiple insertion/withdrawal cycles, and particulates measured during simulated use compared to reference device, meet criteria.Pass
    Flow RateFlow rate through a catheter meets ISO 10555-1, Annex E.Pass
    Aspiration Flow RateAspiration flow rate through the aspiration catheter meets criteria when connected to a constant vacuum source.Pass
    Kink ResistanceTest specimen segments formed into a defined bend diameter meet kink resistance criteria.Pass
    Corrosion ResistanceNo visible corrosion immediately after Corrosion Testing procedure based on ISO 10555-1, Annex A.Pass
    RadiopacityMarker band is fluoroscopically visible.Pass
    Burst Pressure-StaticTested per ISO 10555-1:2013, Annex F, after simulated use, meets criteria.Pass
    Burst Pressure-DynamicMinimum value per specification.Pass
    Connectors for IntravascularHubs tested per ISO 80369-7.Pass
    Lumen CollapseAspiration Catheter samples meet lumen patency under maximum applied vacuum pressures.Pass
    Manual Syringe Injection Peak PressureMeasure peak pressure during manual injection of contrast media with a syringe, meets criteria.Pass
    Aspiration Tubing Bench Testing
    Dimensional/Visual InspectionAspiration Tubing meets all dimensional and visual specifications.Pass
    Tensile StrengthAspiration Tubing meets existing tensile strength specifications.Pass
    Simulated Use TestAspiration Tubing passes testing specified in the simulated use test protocol.Pass
    Resistance to Collapse and Leakage at Maximum Aspiration PressuresAspiration Tubing resistance to collapse at maximum aspiration pressure meets testing specified in the test protocol and does not show signs of leakage at maximum aspiration pressure.Pass
    Flow Switch Functionality TestingFlow Control Switch completely and immediately stops fluid flow after a specified number of ON/OFF cycles.Pass
    Biocompatibility Testing (072 Aspiration Catheter and Delivery Tool)
    Cytotoxicity - MEM ElutionThe achievement of a numerical grade greater than 2 is considered a cytotoxic effect. (Implicit acceptance: numerical grade ≤ 2)Non-cytotoxic
    SensitizationMagnusson and Kligman grades of 1 or greater in the test group generally indicate sensitization, provided grades of less than 1 are seen in control animals. If grades of 1 or greater are noted in control animals, then the reactions of test animals which exceed the most severe reaction in control animals are presumed to be due to sensitization. (Implicit acceptance: no statistically significant sensitization compared to control)Non-sensitizing
    Irritation: Intracutaneous ReactivityThe requirements of the test were met if the final test article score was ≤ 1.0.Non-irritating
    Acute Systemic ToxicityIf using five animals per group, the test article meets the requirement of the test if: 1. Two or more animals from the test group die. 2. Animal behavior, such as convulsions or prostration, occurs in two or more animals from the test group. 3. A final (end of study) body weight loss > 10% occurs in three or more animals from the test group. (Implicit acceptance: none of these conditions occurred)Non-toxic
    Material Mediated PyrogenicityThe requirements of the test were met if no rabbit showed an individual rise in temperature of 0.5 °C or more above its respective baseline temperature throughout the duration of the test.Non-pyrogenic
    ASTM Hemolysis- DirectThe positive control's mean hemolytic index above the negative control must be ≥ 5% for the direct method. The negative control must display a mean hemolytic index of < 2% for the direct method. (Implicit acceptance: test article meets non-hemolytic criteria based on these controls)Non-hemolytic for the Direct Method
    ASTM Hemolysis- IndirectThe positive control's mean hemolytic index above the negative control must be ≥ 5% for the indirect method. The negative control must display a mean hemolytic index of < 2% for the indirect method. (Implicit acceptance: test article meets non-hemolytic criteria based on these controls)Non-hemolytic for the Indirect Method
    Hemocompatibility- Complement ActivationThe negative control (HDPE) concentration must not be significantly higher when compared to the NHS at 37 °C concentration or the final concentration must fall within ± 1 standard deviation of the mean in the test facility historical range for HDPE. The positive reference control (Latex) concentration must be statistically significant when compared to the NHS at 37 °C concentration or the final concentration must fall within ± 1 standard deviation of the mean in the test facility historical range for Latex. (Implicit acceptance: test article acts like negative control)Non-activator of the Complement System
    ThrombogenicityThe thrombogenic potential of a blood-contacting medical device must be comparable to a predicate device. (Implicit acceptance: comparable to predicate device)Non-thrombogenic
    Biocompatibility Testing (Aspiration Tubing)
    Cytotoxicity - MEM ElutionThe achievement of a numerical grade greater than 2 is considered a cytotoxic effect. (Implicit acceptance: numerical grade ≤ 2)Non-cytotoxic
    SensitizationTest group shall yield Grade < 1 score on Magnusson and Kligman scale (provided control group yields Grade < 1). (Implicit acceptance: Grade < 1)Non-sensitizer
    Irritation: Intracutaneous ReactivityThe test requirements are met if the difference between the test mean score and the control mean score is 1.0 or less and the test does not fail at any observation period. Differences of less than 0 are reported as 0. (Implicit acceptance: difference < 1.0)Non-irritant

    2. Sample Size Used for the Test Set and Data Provenance

    • Bench Testing: The document consistently states "All samples met the predetermined acceptance criteria" for each bench test. However, the exact numerical sample size for each specific bench test is not provided.
    • Animal Study: The study involved a "porcine model." The exact number of animals used is not specified. The study was conducted "according to Good Laboratory Practices (GLP) per 21 CFR Part 58," which indicates controlled experimental conditions.
    • Data Provenance:
      • Bench Testing: In-house testing by Q'Apel Medical, Inc.
      • Animal Study: Experimental, prospective study in a porcine (animal) model.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • Bench Testing: Not applicable in the context of expert ground truth. These tests likely involved technicians or engineers following standardized protocols.
    • Animal Study: The document doesn't specify the number or qualifications of experts (e.g., veterinarians, pathologists) involved in establishing ground truth (e.g., assessing vessel histology or clot aspiration results). It mentions that "Sub-chronic and chronic time points were assessed" and that "Clot aspiration and wedge assessment were comparable between the test and control catheters, and both were shown to be safe in porcine vessels via angiography and vessel histology." This implies expert assessment of the outcomes, but details are absent.

    4. Adjudication Method for the Test Set

    • Bench Testing: Not applicable. Bench tests typically rely on objective measurements against predefined specifications.
    • Animal Study: Not specified. While the results were declared "comparable" and "safe," it's not explicitly stated if multiple experts reviewed the outcomes and how any discrepancies were resolved.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    • No. A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The document explicitly states: "No clinical study was conducted as bench and animal testing was determined sufficient for verification and validation purposes and to support the substantial equivalence of the 072 Aspiration System." Therefore, there is no effect size reported for human readers with and without AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    • Not applicable. The device is a physical medical instrument (an aspiration catheter system), not an AI algorithm. Therefore, "standalone" performance in the context of an algorithm is not relevant. The performance reviewed is the device's physical function.

    7. The Type of Ground Truth Used

    • Bench Testing: Reference standards (e.g., ISO standards), design specifications, and comparison to a reference device (K211564 for simulated use particulates) were used to establish "ground truth" (i.e., expected performance).
    • Animal Study: Angiography and vessel histology were used as ground truth for assessing safety and effectiveness (clot aspiration, wedge assessment, tissue response) in the porcine model. These are objective measures collected during and after the animal procedures.

    8. The Sample Size for the Training Set

    • Not applicable. This device is a physical medical instrument, not a machine learning or AI algorithm. Therefore, there is no "training set" in the context of data used to train an AI model.

    9. How the Ground Truth for the Training Set was Established

    • Not applicable. As stated above, this is not an AI device, so there is no training set or ground truth for a training set.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1