K Number
K211564

Validate with FDA (Live)

Date Cleared
2021-10-19

(152 days)

Product Code
Regulation Number
870.1650
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Sterile High-pressure Angiographic Syringes for Single-use are intended for the injection of contrast media or saline: they shall be used with US legally marketed angiographic injectors.

Device Description

The subject devices are identical to all models of predicate devices of K192657. It includes disposable syringes, connection tube, J shape tube and spike. Labeling of subject device now includes pediatrics in the patient population.

AI/ML Overview

This document is a 510(k) summary for Sterile High-pressure Angiographic Syringes for Single-use. It primarily focuses on demonstrating substantial equivalence to a predicate device (K192657) and does not contain information about a study proving specific acceptance criteria in the context of device performance metrics like accuracy, sensitivity, or specificity for AI/ML technologies.

This type of submission typically demonstrates the device meets a set of performance standards or specifications, rather than a "study that proves the device meets the acceptance criteria" in the way a clinical trial or AI performance study would. The acceptance criteria here relate to physical, chemical, and biological properties, as well as compatibility.

Therefore, many of the requested details, such as sample size for test sets, data provenance, number of experts for ground truth, adjudication methods, MRMC studies, standalone performance studies, and training set details, are not applicable to this type of device submission.

Here's the available information based on the document:

1. Table of Acceptance Criteria and Reported Device Performance

The document lists various non-clinical tests and standards that the device complies with, along with some performance specifications.

ItemAcceptance Criteria (Standard/Test/Specification)Reported Device Performance
Product CodeDXTDXT
Regulation NumberCFR 870.1650CFR 870.1650
Indications for UseFor injection of contrast media or saline with US legally marketed angiographic injectors.Sterile High-pressure Angiographic Syringes for Single-use are intended for the injection of contrast media or saline; they shall be used with US legally marketed angiographic injectors.
Mode of operationPower-driven operation, single usePower-driven operation, single use
SterilityEO SterilizedEO Sterilized
Single UseYesYes
Max. Withstanding Pressure (Syringe)300psi, 400psi, 1200psi300psi, 400psi, 1200psi (per specific models, e.g., 100101 at 400psi, 300101 at 1200psi)
Max. Withstanding Pressure (Connection Tube)300psi, 400psi, 1200psi300psi, 400psi, 1200psi (per specific models)
Max. Withstanding Pressure (Spike)400 psi400 psi (for specific models)
Syringe PerformanceISO 7886ISO 7886
Luer ConnectorISO 594-1; ISO 594-2ISO 594-1; ISO 594-2
CompatibilityPass (with specific injectors outlined in Tables 1, 2, 3)Pass
BiocompatibilityNo Cytotoxicity, No Irritation, No Sensitization, No Pyrogen, No Acute Toxicity, No HemolysisNo Cytotoxicity, No Irritation, No Sensitization, No Pyrogen, No Acute Toxicity, No Hemolysis
Endotoxin Limit20 EU per device20 EU per device
EO/ECH Residue LimitLimited Contact: ≤ 24h, EO: 0.6mg/day, ECH: 1.28mg/day for 10kg patient (Children) as per ISO 10993-7EO/ECH residual test has been conducted on the models in this submission and test results have demonstrated EO/ECH residue level meet the residue limit for 10kg children group as per ISO 10993-7. (This is a specific test mentioned to justify the change in labeling regarding pediatric use compared to the predicate).
Sterilization ResidualsISO 10993-7:2008 (AMD1:2019)Complies with standard
Seal StrengthASTM F88/F88M-15Complies with standard
Seal Leak DetectionASTM F1929-15Complies with standard
Bacterial EndotoxinsUSP 41-NF36 2018 <85>Complies with standard
Sterile Hypodermic Syringes (Manual Use)ISO 7886-1:2017Complies with standard
Sterile Hypodermic Syringes (Power-driven)ISO 7886-2:1996Complies with standard
Conical fittings (Luer taper)ISO594-1:1986, ISO594-2:1998Complies with standards
In Vitro CytotoxicityISO 10993-5:2009Complies with standard (No Cytotoxicity)
Irritation and Skin SensitizationISO 10993-10:2010Complies with standard (No Irritation, No Sensitization)
Interactions with BloodISO 10993-4:2017Complies with standard (No Hemolysis)
Hemolytic PropertiesASTM F756:2017Complies with standard (No Hemolysis)
Pyrogen TestUSP 41 NF 36<151>Complies with standard (No Pyrogen)
Sterilization ProcessISO 11135:2014Complies with standard
Particulate MatterUSP <788>Complies with standard
Shelf life5 years5 years

2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

  • The document does not specify sample sizes for the individual non-clinical tests (e.g., number of syringes tested for pressure, number of samples for biocompatibility).
  • Data provenance: Not explicitly stated, but the submission is from Shenzhen Boon Medical Supply Co., Ltd. in China, suggesting the tests were likely conducted in China. The submission itself is dated August 13, 2021, and received August 23, 2021, implying these are retrospective test results for device validation.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

  • This is not applicable. The "ground truth" for this medical device (angiographic syringes) involves meeting engineering specifications and biological safety standards, not diagnostic interpretations that require expert consensus. The compliance is against defined standards (e.g., ISO, ASTM, USP) and internal specifications.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

  • This is not applicable for this type of device. Adjudication methods are typically used in clinical studies for diagnostic accuracy, where disagreements between readers or algorithms need resolution. Here, the tests yield objective pass/fail results against established criteria.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • This is not applicable. The device is a physical medical instrument (syringe), not an AI/ML diagnostic or assistive tool.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

  • This is not applicable. The device is a physical medical instrument (syringe), not an AI/ML algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • The "ground truth" in this context is defined by international and national standards (e.g., ISO 10993, ASTM F88, USP 41) and the device's own engineering specifications (e.g., maximum pressure resistance, volume, material properties). Compliance is measured against these objective criteria.

8. The sample size for the training set

  • This is not applicable. There is no AI/ML component, therefore no "training set."

9. How the ground truth for the training set was established

  • This is not applicable. There is no AI/ML component, therefore no "training set" ground truth.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath. The logo is simple and professional, and it is easily recognizable.

October 19, 2021

Shenzhen Boon Medical Supply Co., Ltd Baihan Feng Regulatory Affair Manager No.18 Jirong Road, Shenkeng, Henggang Street, Longgang District Shenzhen, Guangdong 518173 China

Re: K211564

Trade/Device Name: Sterile High-pressure Angiographic Syringes for Single-use Regulation Number: 21 CFR 870.1650 Regulation Name: Angiographic injector and syringe Regulatory Class: Class II Product Code: DXT Dated: August 13, 2021 Received: August 23, 2021

Dear Baihan Feng:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For CAPT Alan Stevens Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K211564

Device Name

Sterile High-pressure Angiographic Syringes for Single-use

Indications for Use (Describe)

Sterile High-pressure Angiographic Syringes for Single-use are intended for the injection of contrast media or saline: they shall be used with US legally marketed angiographic injectors.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.

{3}------------------------------------------------

K211564 510(k) Summary

This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR

  • Date of Preparation: 10/19/2021 1.
  • Sponsor Identification 2.

Shenzhen Boon Medical Supply Co., Ltd.

No.18 Jirong Road, Shenkeng, Henggang Street, Longgang District, Shenzhen, 518173, China

Establishment Registration Number: 3012395857

Contact Person: Baihan Feng Position: Regulatory Affair Manager Tel: +86-755-28638515 Fax: +86-755-28638033 Email: faguibu-feng@szboon.com

    1. Identification of Subject Device
      Trade Name: Sterile High-pressure Angiographic Syringes for Single-use;

Common Name: Disposable angiographic syringe

Product NameModels
Syringe100101, 100103, 100113, 200101, 200102, 300101,300105, 100104, 100114, 200103, 300103,100111,100121, 100124, 200104, 100108, 100118, 100129,100130,100109, 100119, 200107, 300102, 300108,300110, 300111, 300112
Connection Tube400101, 400102, 400103, 600101, 600102, 500105,500106, 500107, 500108, 400201, 400202, 400203,600201, 600202, 500101, 500102, 500103, 500104,500201, 500202, 500203, 500204, 500205, 500206,500207, 500208
J shape tube700103
Spike700101, 700102, 700104-1, 700104-2, 700105-1, 700105-2, 700106, 700107-1, 700107-2

{4}------------------------------------------------

Regulatory Information

Classification Name: Angiographic injector and syringe Classification: II Product Code: DXT; Regulation Number: 21 CFR 870.1650; Review Panel: Cardiovascular;

Indications for Use:

Sterile High-pressure Angiographic Syringes for Single-use are intended for the injection of contrast media or saline; they shall be used with US legally marketed angiographic injectors.

Device Description:

The subject devices are identical to all models of predicate devices of K192657. It includes disposable syringes, connection tube, J shape tube and spike. Labeling of subject device now includes pediatrics in the patient population.

  • Syringe: the syringe is intended to be used with an U.S. legally marketed angiography injector. A Compatibility is shown in Table 1.
Model(Syringe)Volume (ml)TypeResistant liquidleak pressure(psi)Injector
100101200mlSingle Shot400MCT & MCT plus CT, K924116Vistron CT, K991557EnVision CT, K934086
100103200mlSingle Shot400Stellant-S K182273
100113200/200mlDual Shots400Stellant-D, K182273
20010165/65mlDual Shots300Spectris, K935668
20010265/115mlDual Shots300Solaris MRI, K033247
300101150mlSingle Shot1200Mark V, K822536
300105130mlSingle Shot1200Mark III & Mark IV, K822536
100104200mlSingle Shot400CT 9000 & CT9000 ADV, K912944
100114200/200mlDual Shots400CT 9000 & CT9000 ADV, K912944
20010360/60mlDual Shots300Optistar LE, Elite, K073592
300103150mlSingle Shot1200ILLUMENA, K963071
100111200mlSingle Shot400Empower CT, K071378
100121200/200mlDual Shots400Empower CT, K071378

Table 1 Compatibility between Syringe and Injectors

{5}------------------------------------------------

10012460/100mlDual Shots400Dual Shot, K052633
20010460/60mlDual Shots300Sonic Shot, K091734
100108200mlSingle Shot400Dual shot CT, K062168
100118200/200mlDual Shots400Dual shot CT, K133189
100129125mlSingle Shot400CT Optione, K152361
100130125/125mlDual Shots400CT Optione, K152361
100109100mlSingle Shot400Dual shot, K062168
100119100/100mlDual Shots400Duat shot, K062168
200107100/100mlDual Shots300EZEM Empower, K062449
300102125mlSingle Shot1200120S, K092896
300108150mlSingle Shot1200Mark VII, K112086
300110150mlSingle Shot1200Rempress, K092896
300111150mlSingle Shot1200Angiomat 6000, K944875
300112200mlSingle Shot1200Illumena K963071
  • Connection tube: used to connect the syringe and the catheter. The tubes are also available in various configurations, which are straight tube (used with single shot syringe), type Y and type T tube (used with dual shot syringe). The pressure specification for connection tube is provided in Table 2 Pressure Specifications for Connection Tube.

ModelMaximumWithstandingPressure (psi)Type
400101400Straight
400102400Type Y
400103400Type T
600101400Type T
600102400Type Y
5001051200Straight
5001061200Straight
5001071200Straight
5001081200Straight
400201400Straight
400202400Type Y
400203400Type T
600201400Type T
600202400Type Y
5001011200Straight
5001021200Straight
5001031200Straight
5001041200Straight
5002011200Straight

Table 2 Pressure Specifications for Connection Tube

{6}------------------------------------------------

5002021200Straight
5002031200Straight
5002041200Straight
5002051200Straight
5002061200Straight
5002071200Straight
5002081200Straight
  • J shape tube: used to draw contrast media/saline into the syringe barrel before the syringe installed.

  • Spike: used to draw contrast media/saline into the syringe barrel before the syringe installed. The pressure specification for spike is provided in Table 3 Pressure Specifications for Spike

Table 3 Pressure Specifications for Spike
ModelMaximumWithstandingPressure (psi)Type
700101/Long spike
700102/Short spike
700104-1400 psiSingle Air ChamberTransfer Set
700104-2400 psiSingle Air ChamberTransfer Set withExtension Tube
700105-1400 psiDual Air ChamberTranfer Set
700105-2400 psiDual Air ChamberTranfer Set withextension tube
700106/Transfer Set With C-Clamp
700107-1/Transfer set with Claveconnector
700107-2/Clave connector Transferset with Check valve

Table 3 Pressure Specifications for Spike

    1. Identification of Predicate
      Device 510k Number: K192657

Product Name: Sterile High-pressure Angiographic Syringes for Single-use; Manufacturer: Shenzhen Boon Medical Supply Co., Ltd.

  • ર. Non-Clinical Test Conclusion

{7}------------------------------------------------

The non-clinical tests for the predicate device in K192657 are applicable for proposed device. The test results demonstrated that the proposed device complies with the following standards:

  • ISO 10993-7:2008 (AMD1:2019) Biological evaluation of medical devices-Part 7: Ethylene oxide > sterilization residuals;
  • ASTM F88/F88M-15 Standard test method for seal strength of flexible barrier materials; A
  • ASTM F1929-15 Standard Test Method for Detecting Seal Leaks in Porous Medical Package by A Dye Penetration:
  • USP 41-NF36 2018 <85> Bacterial Endotoxins Limit;

  • ISO 7886-1:2017 Sterile hypodermic syringes for single use-Part 1: Syringes for manual use;

  • ISO 7886-2:1996 Sterile hypodermic syringes for single use --Part 2: Syringes for use with power-> driven svringe pumps:
  • ISO594-1:1986 Conical fittings with a 6% (Luer) taper for syringes, needles and certain other medical equipment -Part 1: General Requirements;

  • ISO594-2:1998 Conical fittings with a 6% (Luer) taper for syringes, needles and certain other medical equipment -Part 2: Lock Fitting;

  • ISO 10993-5:2009 Biological evaluation of medical devices --Part 5: Tests for In Vitro cytotoxicity;

  • ISO 10993-10:2010 Biological evaluation of medical devices -Part 10: Tests for irritation and skin sensitization:

  • ISO 10993-4:2017, Biological Evaluation of Medical Device -Part 4: Selection of tests for interactions with blood.

  • ASTM F756:2017, Standard Practice for Assessment of Hemolytic Properties of Materials

  • USP 41 NF 36<151> Pyrogen Test

  • A ISO 11135:2014, Sterilization of health-care Products-Ethylene Oxide-Requirements for the development, validation and routine control of a sterilization process for medical devices
  • USP <788> Particular Matter in Injections >
  • Compatibility Test Report between injectors, syringe, connection tube and spike/J shape tube. The compatibility test demonstrated that each device meet performance under maximum sustained pressure specification.

    1. Clinical Test Conclusion No clinical study is included in this submission.
ItemSubject DevicePredicate DeviceK192657
Product CodeDXTDXT
Regulation NumberCFR 870.1650CFR 870.1650
Indications for UseSterile High-pressureAngiographic Syringes forSterile High-pressureAngiographic Syringes for
Single-use are intended forSingle-use are intended for
the injection of Contrastmedia or saline; they shall bethe injection of Contrastmedia or saline; they shall be
used with an US legallyused with an US legally
marketed angiographicinjectors.angiographicmarketed injectors.
Mode of operationPower-driven operation,single usePower-driven operation,single use
ConfigurationAngiographic SyringeAngiographic Syringe
Connecting tubeConnecting tube
J shape tube/SpikeJ shape tube/Spike
SterilityEO SterilizedEO Sterilized
Single UseYesYes
ModelSame Models
MaximumwithstandingpressureSyringe300psi, 400psi, 1200psi300psi, 400psi, 1200psi
Connection tube300psi, 400psi, 1200psi300psi, 400psi, 1200psi
J shape tubeNANA
Spike400 psi400 psi
SpecificationSyringe(Volume, ml)200, 150, 125, 130, 100,200/200, 60/100, 125/125,100/100, 65/65, 65/115,60/60, 50/50200, 150, 125, 130, 100,200/200, 60/100, 125/125,100/100, 65/65, 65/115,60/60, 50/50
Connectiontube (overallLength, mm)2002500, 15002500,1500, 1800, 2000, 2500, 500,750, 1000, 12002002500, 15002500,1500, 1800, 2000, 2500, 500,750, 1000, 1200
J shape tube(overallLength, mm)240240
Spike(overallLength, mm)58.8, 47.3, 1000, 2800, 1200,2900, 180, 260, 340, 420,500, 450, 550, 60058.8, 47.3, 1000, 2800, 1200,2900, 180, 260, 340, 420,500, 450, 550, 600
Performance
SyringeISO 7886ISO 7886
Luer connectorISO 594-1; ISO 594-2ISO 594-1; ISO 594-2
CompatibilityPassPass
Patient-Contact Material
SyringeBarrelPP (polypropylene) or PET(Polyethylene terephthalate)PP (polypropylene) or PET(Polyethylene terephthalate)
PistonPolyisoprene rubberPolyisoprene rubber
LubricantPolydimethylsiloxanePolydimethylsiloxane
ConnectiontubeTubingPVC (Polyvinylchloride) orPVC (Polyvinylchloride notmade with DEHP) or PUPVC (Polyvinylchloride) orPVC (Polyvinylchloride notmade with DEHP ) or PU
( Polyurethane )( Polyurethane )
LuerconnectorsPC (Polycarbonate)PC (Polycarbonate)
UV adhesiveUltraviolet adhesiveUltraviolet adhesive
SpikeClosure-piercingdeviceABS (acrylonitrile-butadiene-styrene)ABS (acrylonitrile-butadiene-styrene)
FiltermembranePP (polypropylene)PP (polypropylene)
TubePolyvinyl chloride (PVC)Polyvinyl chloride (PVC)
LuerConnectorPolycarbonate (PC)Polycarbonate (PC)
Protective capAcrylonitrile-butadiene-styrene (ABS)Acrylonitrile-butadiene-styrene (ABS)
J shape tubetubePE (Polyethylene)PE (Polyethylene)
BiocompatibilityNo CytotoxicityNo Cytotoxicity
No IrritationNo Irritation
No SensitizationNo Sensitization
No PyrogenNo Pyrogen
No Acute ToxicityNo Acute Toxicity
No HemolysisNo Hemolysis
Endotoxin Limit20 EU per device20 EU per device
EO/ECH residue limitLimited Contact: ≤ 24hEO: 0.6mg/ dayECH: 1.28mg /dayFor 10kg patient (Children)as per ISO 10993-7Limited Contact: ≤ 24hEO: 4 mg/dayECH: 9 mg/dayFor 70kg patient (Adult) ISO10993-7
PopulationNot intended for infant orneonatal useNot intended for pediatric orneonatal use
Shelf life5 years5 years

7. Substantial Equivalence Comparison

{8}------------------------------------------------

{9}------------------------------------------------

The only difference between the subject and the predicate is that the labeling states " Not intended for infant or neonatal use" in the subject device, while predicate device states "Not intended for pediatric or neonatal use." For this modification, EO/ECH residual test has been conducted on the models in this submission and test results have demonstrated EO/ECH residue level meet the residue limit for 10kg children group as per ISO 10993-7. Therefore, this difference will not impact the safety and effectiveness of the device.

Substantial Equivalence (SE) Conclusion: 8.

The differences between the predicate and the subject device do not raise any new or different questions of safety or effectiveness. The subject device is Substantially Equivalent (SE) to the predicate device.

§ 870.1650 Angiographic injector and syringe.

(a)
Identification. An angiographic injector and syringe is a device that consists of a syringe and a high-pressure injector which are used to inject contrast material into the heart, great vessels, and coronary arteries to study the heart and vessels by x-ray photography.(b)
Classification. Class II (special controls). The device, when it is a non-patient contacting balloon inflation syringe intended only to inflate/deflate balloon catheters and monitor pressure within the balloon, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.