Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K250450

    Validate with FDA (Live)

    Date Cleared
    2025-09-19

    (213 days)

    Product Code
    Regulation Number
    888.3353
    Age Range
    All
    Predicate For
    N/A
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    • Versafitcup CC TRIO

    The Versafitcup CC Trio and the Versafitcup CC Trio No-Hole are designed for cementless use in total hip arthroplasty in primary or revision surgery.

    The patient should be skeletally mature.

    The patient's condition should be due to one or more of:

    • Severely painful and/or disabled joint: as a result of osteoarthritis, post-traumatic arthritis, rheumatoid arthritis or psoriatic arthritis, congenital hip dysplasia, ankylosing spondylitis.
    • Avascular necrosis of the femoral head.
    • Acute traumatic fracture of the femoral head or neck.
    • Failure of previous hip surgery: joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement where sufficient bone stock is present.

    • Quadra-P and Quadra-P Collared

    The hip prosthesis Quadra-P and Quadra-P Collared are designed for cementless use in total or partial hip arthroplasty in primary or revision surgery. The Quadra-P Cemented is designed for cemented use in total or partial hip arthroplasty in primary or revision surgery. Hip Replacement is indicated in the following cases:

    • Severely painful and/or disabled joint as a result of arthritis, traumatic arthritis, rheumatoid polyarthritis, or congenital hip dysplasia.
    • Avascular necrosis of the femoral head.
    • Acute traumatic fracture of the femoral head or neck.
    • Failure of previous hip surgery: joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement.

    • AMIStem-H, AMIStem-H Collared, AMIStem-H Proximal Coating, AMIStem-P and AMIStem-P Collared

    The hip prosthesis AMIStem-H, AMIStem-H collared, AMIStem-H Proximal Coating, AMIStem-P and AMIStem-P Collared are designed for cementless use in total or partial hip arthroplasty in primary or revision surgery. The AMIStem-C is designed for cemented use in total or partial hip arthroplasty in primary or revision surgery.

    Hip Replacement is indicated in the following cases:

    • Severely painful and/or disabled joint as a result of arthritis, traumatic arthritis, rheumatoid polyarthritis, or congenital hip dysplasia.
    • Avascular necrosis of the femoral head.
    • Acute traumatic fracture of the femoral head or neck.
    • Failure of previous hip surgery: joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement.

    • MiniMAX

    The MiniMax is designed for cementless use in total or partial hip arthroplasty in primary or revision surgery. Hip Replacement is indicated in the following cases:

    • Severely painful and/or disabled joint as a result of arthritis, traumatic arthritis, rheumatoid polyarthritis, or congenital hip dysplasia.
    • Avascular necrosis of the femoral head.
    • Acute traumatic fracture of the femoral head or neck.
    • Failure of previous hip surgery: joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement.

    • Quadra-H and Quadra-R

    The hip prosthesis Quadra -S, Quadra-H and Quadra-R is designed for cementless use in total or partial hip arthroplasty in primary or revision surgery. The hip prosthesis Quadra-C is designed for cemented use in total or partial hip arthroplasty in primary or revision surgery. Hip Replacement is indicated in the following cases:

    • Severely painful and/or disabled joint as a result of arthritis, traumatic arthritis, rheumatoid polyarthritis, or congenital hip dysplasia.
    • Avascular necrosis of the femoral head.
    • Acute traumatic fracture of the femoral head or neck.
    • Failure of previous hip surgery: joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement.

    The Quadra-C size 0 implant should not be implanted in patients with a mass of 65 kg or greater.

    • SMS and SMS Collared

    The hip prosthesis SMS and SMS Collared are designed for cementless use in total or partial hip arthroplasty in primary or revision surgery. Hip Replacement is indicated in the following cases:

    • Severely painful and/or disabled joint as a result of arthritis, traumatic arthritis, rheumatoid polyarthritis, or congenital hip dysplasia.
    • Avascular necrosis of the femoral head.
    • Acute traumatic fracture of the femoral head or neck.
    • Failure of previous hip surgery: joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement.

    • Mpres

    The Mpres stem is a cementless neck preserving stem designed for use in total or partial hip arthroplasty for primary or revision surgery.

    Total Hip Arthroplasty with the Mpres stem is indicated in the following cases:

    • Severely painful and/or disabled joint as a result of arthrosis, traumatic arthritis, rheumatoid polyarthritis or congenital hip dysplasia.
    • Avascular necrosis of the femoral head.
    • Acute traumatic fracture of the femoral head.
    • Failure of previous hip surgery:
      • Conservative hip surgery,
      • Internal fixation,
      • Arthrodesis,
      • Hip resurfacing replacement.

    Partial hip arthroplasty with the Mpres stem is indicated in the following cases:

    • Acute traumatic fracture of the femoral head.
    • Avascular necrosis of the femoral head.
    • Primary pathology involving the femoral head but with a non-deformed acetabulum.

    • Mpact Acetabular Shell

    The Mpact implants are designed for cementless use in total hip arthroplasty in primary or revision surgery.

    The patient should be skeletally mature.

    The patient's condition should be due to one or more of:

    • Severely painful and/or disabled joint: as a result of osteoarthritis, post-traumatic arthritis, rheumatoid arthritis or psoriatic arthritis, congenital hip dysplasia, ankylosing spondylitis.
    • Avascular necrosis of the femoral head.
    • Acute traumatic fracture of the femoral head or neck.
    • Failure of previous hip surgery: joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement where sufficient bone stock is present.
    Device Description

    The aim of this submission is to seek clearance for the addition of a new coating supplier for the following coatings:

    • Titanium + Hydroxyapatite coating;
    • Hydroxyapatite coating;
    • Porous Titanium coating for Mpact devices; and
    • Titanium coating for Mpact T sizes devices.

    The already FDA cleared devices affected by the change are listed in Table 1 below.

    510kProduct nameSubstrate materialCoating
    K103352 K122911Versafitcup CC TRIOTi-6Al-4V (ASTM F136)Titanium + Hydroxyapatite
    K181254 K192827 K202730Quadra-P Quadra-P Collared Quadra-P Short NeckTi-6Al-7Nb (ISO 5832-11)Titanium + Hydroxyapatite
    K161635 K173794AMIStem-H Proximal CoatingTi-6Al-7Nb (ISO 5832-11)Titanium + Hydroxyapatite
    K173794 K192126AMIStem-P AMIStem-P Collared AMIStem-P Short NeckTi-6Al-7Nb (ISO 5832-11)Titanium + Hydroxyapatite
    K170845 K192352MiniMAXTi-6Al-7Nb (ISO 5832-11)Titanium + Hydroxyapatite
    K181693 K201673SMSTi-6Al-7Nb (ISO 5832-11)Titanium + Hydroxyapatite
    K203041SMS CollaredTi-6Al-7Nb (ISO 5832-11)Titanium + Hydroxyapatite
    K210263MpresTi-6Al-7Nb (ISO 5832-11)Titanium + Hydroxyapatite
    K093944 K121011AMIStem-H AMIStem-H CollaredTi-6Al-7Nb (ISO 5832-11)Hydroxyapatite
    K082792 K093944 K121011Quadra-H Quadra-H Short NeckTi-6Al-7Nb (ISO 5832-11)Hydroxyapatite
    K082792Quadra-RTi-6Al-7Nb (ISO 5832-11)Hydroxyapatite
    K103721 K122641 K132879 K230011Mpact Acetabular ShellTi-6Al-4V (ASTM F136)Porous Titanium Porous Titanium / Titanium
    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K210062

    Validate with FDA (Live)

    Device Name
    Mectaplug PE II
    Date Cleared
    2021-04-08

    (87 days)

    Product Code
    Regulation Number
    878.3300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K800144, K791125, K192827, K131310

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    MectaPlug PE II is an intramedullary cement restrictor, designed to seal the intramedullary canal prior to cementation in total or partial hip arthroplasty, for primary or revision surgery.

    Total hip arthroplasty is indicated in the following cases:

    • Severely painful and/or disabled joint as a result of arthrosis, traumatic arthritis, rheumatoid polyarthritis or congenital hip dysplasia
    • Avascular necrosis of the femoral head
    • Acute traumatic fracture of the femoral head or neck
    • Failure of previous hip surgery: joint reconstruction, internal fixation, arthrodesis, partial hip arthroplasty, hip resurfacing replacement or total hip arthroplasty

    Partial hip arthroplasty is indicated in the following cases:

    • Acute traumatic fracture of the femoral head or neck
    • Non-union of femoral neck fracture
    • Avascular necrosis of the femoral head
    • Primary pathology involving the femoral head, but with a non-deformed acetabulum
    Device Description

    MectaPlug PE II is an intramedullary cement restrictor (plug), used to seal the intramedullary canal prior to cementation in total or partial hip arthroplasty. MectaPlug PE II prevents the cement from flowing down the diaphysis, thereby facilitating cement pressurization, MectaPlug PE II is composed of UHMWPE (ISO 5834-2) with a radiopaque stainless steel wire (ISO 5832-1).

    MectaPlug PE II is positioned distal to the stem tip preventing cement from being displaced distally in the canal and increasing the intramedullary pressure at the bone-cement interface during insertion of the stem. A radio marker is included in the tip of the Medacta cement restrictor.

    MectaPlug PE II has 6 discs divided in several leaves by oblique cuts and a central peg. The MectaPlug PE II device is available in the following sizes: S (Ø 8 to Ø 11), M (Ø 12 to Ø 15), L (Ø 16 to Ø 20), XL (Ø 21 to Ø 25), XXL (Ø 26 to Ø 30).

    MectaPlug PE II is made in UHMWPE according to ISO 5834-2 Implants for surgery -- Ultrahigh-molecular-weight polyethylene Moulded forms and with a radiopaque wire in AISI 316LVM according to ISO 5832-1 Implants for surgery - Metallic materials - Part 1: Wrought stainless steel.

    Mectaplug PE II implants are provided sterile and in single-use packages.

    AI/ML Overview

    The provided document describes a medical device called "MectaPlug PE II," an intramedullary cement restrictor. It does not contain information about acceptance criteria for a study proving device performance, nor does it detail a study that would meet the typical requirements listed in your prompt (e.g., sample size for test/training sets, expert involvement, MRMC studies, ground truth establishment, etc.).

    The document primarily focuses on:

    • Regulatory Clearance (510(k) summary): Seeking FDA clearance for marketing by demonstrating substantial equivalence to predicate devices.
    • Device Description: What the device is, its components, and sizes.
    • Indications for Use: When and how the device is intended to be used.
    • Comparison to Predicate Devices: Highlighting similarities and differences.
    • Non-Clinical Studies: Listing various risk assessments and material characterization tests.
    • Absence of Clinical Studies: Explicitly stating that no clinical studies were conducted.

    Therefore, I cannot provide a table of acceptance criteria and reported device performance from this document in the manner you requested, as the document states, “No clinical studies were conducted.” The "Performance Data" section solely lists non-clinical mechanical studies and material characterization, which are generally not presented with acceptance criteria in the format of clinical performance metrics like sensitivity, specificity, or reader improvement.

    To directly answer your numbered points based only on the provided text:

    1. A table of acceptance criteria and the reported device performance: Not available. The document lists non-clinical studies (e.g., "Risk that the cement restrictors are inadequate for femoral canal shape," "Risk of breakage of highly stressed parts") but does not provide specific quantitative acceptance criteria or reported performance outcomes from these studies. It focuses on the fact that these studies were conducted.
    2. Sample sizes used for the test set and the data provenance: Not applicable, as no clinical studies with test sets are reported.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable, as no clinical studies with test sets are reported.
    4. Adjudication method for the test set: Not applicable, as no clinical studies with test sets are reported.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is a physical medical implant (cement restrictor), not an AI/imaging diagnostic device. No MRMC study was conducted, and the document explicitly states, "No clinical studies were conducted."
    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done: Not applicable. This device is a physical medical implant, not an algorithm.
    7. The type of ground truth used: Not applicable, as no clinical studies are reported. The non-clinical tests would rely on engineering specifications and material standards rather than "ground truth" as it pertains to clinical outcomes or expert consensus on clinical data.
    8. The sample size for the training set: Not applicable, as no clinical studies or AI algorithm development are reported.
    9. How the ground truth for the training set was established: Not applicable, as no clinical studies or AI algorithm development are reported.
    Ask a Question

    Ask a specific question about this device

    K Number
    K203041

    Validate with FDA (Live)

    Date Cleared
    2020-12-02

    (57 days)

    Product Code
    Regulation Number
    888.3353
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K192827

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SMS system is designed for cementless use in total or partial hip arthroplasty, for primary or revision surgery. Hip replacement is indicated in the following cases:

    • . Severely painful and/or disabled joint as a result of arthrosis, traumatic arthritis, rheumatoid polyarthritis or congenital hip dysplasia
    • · Avascular necrosis of the femoral head
    • · Acute traumatic fracture of the femoral head or neck
    • · Failure of previous hip surgery: joint reconstruction, internal fixation, arthroplasty, hip resurfacing replacement or total hip arthroplasty
    Device Description

    The SMS femoral stem is a cementless bone preserving short stem designed for proximal fixation in total or partial hip arthroplasty for primary or revision surgery.
    The SMS implants are comprised of the following products:

    • SMS Cementless Solid Standard Stem (available in 15 sizes from size 1 to 15); and ●
    • SMS Cementless Solid Lateralized Stem (available in 15 sizes from size 1 to 15). .
      Both are available on the US market via the clearance - K181693 and K201673.
      SMS collared solid version, has been introduced as a product range extension to the stems currently marketed as the SMS product line (K181693 and K201673). Quadra P is the collared stem currently marked in US K192827.
      The SMS implants are line extensions to Medacta's Total Hip Prosthesis - AMIStem-H, Quadra-S and Quadra-H Femoral Stems (K093944), AMIStem and Quadra - Line Extension (K121011), AMIStem-P, AMIStem-P Collared and AMIStem-H Proximal Coating Femoral Stems (K173794), Quadra-H and Quadra-R Femoral Stems (K082792), AMIStem-H Proximal Coating (K161635), MiniMAX (K170845), and SMS (K181693 and K201673).
      The SMS implants are part of the Medacta Total Hip Prosthesis system. The Medacta Total Hip Prosthesis system consists of femoral stems, modular femoral heads and acetabular components. The acetabular components consist of metal cups and liners made of ultra-high molecular weight polyethylene (UHMWPE) or Highcross highly crosslinked ultra-high molecular weight polyethylene (HXUHMWPE). Acetabular components include the Mpact DM (K143453), VersafitCup (K083116 and K092265), VersafitCup CC Trio (K103352), Mpact (K103721 and K132879), Mpact 3D Metal (K171966) and Medacta Bipolar Head (K091967).
      The SMS stems can be combined with the CoCr ball heads (K072857, K080885 and K103721), Endo Head (K111145) or with the MectaCer BIOLOX® Forte (K073337), MectaCer BIOLOX® Delta Femoral Heads (K112115) or MectaCer BIOLOX® Option Heads (K131518).
    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and supporting studies for the SMS Cementless Stem.

    It's important to note that this document is a 510(k) summary for a medical device (hip implant), not an AI/ML powered device. Therefore, the questions related to AI/ML specific criteria (such as effect size of human readers with AI assistance, standalone algorithm performance, ground truth for training set, training set size) are not applicable to this submission.

    Acceptance Criteria and Reported Device Performance

    The provided text does not explicitly list "acceptance criteria" in a quantitative manner as one might find for an AI/ML device (e.g., target specificity/sensitivity). Instead, the criteria are implicitly derived from the performance tests conducted to demonstrate substantial equivalence to predicate devices. The device's performance is demonstrated through satisfying these tests.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Implied)Reported Device Performance
    Mechanical Performance:
    - Range of Motion (ROM) (per EN ISO 21535:2009)Performance confirmed; testing leveraged from predicate device.
    - Fatigue Testing (Stem Endurance) (per ISO 7206-4:2010 + A1:2016)Performance confirmed; testing leveraged from predicate device.
    - Fatigue Testing (Head and Neck Region Endurance) (per ISO 7206-6:2013)Performance confirmed; testing leveraged from predicate device.
    - Pull-off Force Testing (Taper Connections) (per ASTM F2009-00)Performance confirmed; testing leveraged from predicate device.
    Coating Performance:
    - Hydroxyapatite properties (per ISO 13779-1:2008)Performance confirmed; testing leveraged from predicate device.
    - Tension Testing of Calcium Phosphate and Metal Coatings (per ASTMPerformance confirmed; testing leveraged from predicate device.
    F1147-99)
    Biocompatibility/Safety:
    - Pyrogenicity (per European Pharmacopoeia §2.6.14 / USP <85> & <151>)Bacterial Endotoxin Test (LAL) and Pyrogen test conducted. Device is not labeled as non-pyrogenic or pyrogen-free. Performance confirmed; testing leveraged from predicate device.
    - Biocompatibility (general)Performance confirmed; testing leveraged from predicate device.
    Other Characteristics:
    - CCD angleSimilar to predicate.
    - Cementless applicationSimilar to predicate.
    - Material of constructionSimilar to predicate.
    - Coatings and coating compositionSimilar to predicate.
    - Device usage, sterility, shelf life, packaging, sizes, stem lengthsSimilar to predicate.
    Substantial Equivalence:The introduction of the SMS collared version does not create a new worst case; therefore, performance testing from the predicate device was leveraged. The SMS Collared implants are as safe and effective as the predicate devices (SMS solid stem, K181693 and K201673; and Quadra P collared, K192827).

    Study Details:

    2. Sample size used for the test set and the data provenance

    • Sample Size for Test Set: Not explicitly stated in terms of number of devices. The "tests" here refer to mechanical and material property tests (e.g., fatigue testing, pull-off force, coating adherence). These tests are typically performed on a statistically relevant number of samples of the device and its components, but the exact count isn't in this summary.
    • Data Provenance: The data is derived from non-clinical studies (laboratory testing) and in comparison to predicate devices. There is no mention of human subject data, so concepts like "country of origin" or "retrospective/prospective" don't directly apply in the clinical sense. These are engineering specifications leveraged from previously cleared devices.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • This question is not applicable. For this type of device (hip implant), ground truth isn't established by human experts reviewing medical images or patient data in the typical sense of AI/ML device evaluation. The "ground truth" or standard for performance is established by recognized international and national standards (e.g., ISO, ASTM) for the mechanical and material properties of orthopedic implants.

    4. Adjudication method for the test set

    • This question is not applicable. Adjudication methods like 2+1 or 3+1 are used to resolve disagreements among human reviewers (experts) in establishing ground truth for diagnostic or prognostic studies, which is not the nature of the evaluation for this hip implant. The "adjudication" for mechanical tests would be adherence to the specified test methods and criteria outlined in the standards.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No MRMC study was done. This study concerns a physical orthopedic implant, not an AI-powered diagnostic or assistive tool for human readers. Therefore, there is no "effect size of human readers improving with AI assistance."

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This device is a physical hip implant, not an algorithm.

    7. The type of ground truth used

    • The "ground truth" for the performance evaluation of this device is based on established engineering and material standards (ISO 21535, ISO 7206, ASTM F2009, ASTM F1147, ISO 13779, European Pharmacopoeia, USP). The device is deemed safe and effective and "substantially equivalent" if it meets the performance requirements stipulated by these standards, generally by demonstrating that it performs as well as, or better than, the predicate devices for which these criteria were previously established.

    8. The sample size for the training set

    • Not applicable. This submission is for a physical medical device, not an AI/ML model that requires a training set. The "training" for this device would refer to the design and manufacturing process, which adheres to quality systems regulations.

    9. How the ground truth for the training set was established

    • Not applicable. As above, there is no "training set" in the context of an AI/ML algorithm. The design and manufacturing of the device follow established engineering principles and standards.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1