K Number
K131310

Validate with FDA (Live)

Date Cleared
2013-08-22

(107 days)

Product Code
Regulation Number
888.3560
Age Range
All
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The GMK® Total Knee System is designed for cemented use in total knee arthroplasty, if there is evidence of sufficient sound bone to seat and support the components. This knee replacement system is indicated in the following cases: - Severely painful and/or disabled joint as a result of arthritis, traumatic arthritis, rheumatoid arthritis or polyarthritis. - Avascular necrosis of femoral condyle. - Post traumatic loss of joint configuration. - Primary implantation failure. Tibial augments are to be attached to the tibial baseplate with both the fixing cylinders and bone cement. If a semi-constrained insert is used, it is mandatory to implant an extension stem both on the tibial and the femoral components.

Device Description

The GMK Full PE Tibial Components are symmetric and come in ultracongruent and posterior-stabilized designs in sizes 1-6 with thicknesses of 10, 12, 14, and 17mm and have an axial rotation of +/- 10°. The GMK Full PE Tibial Components are made from UHMWPE (ISO 5834 -2) Type 1. There are two radiopaque wires made of AISI 316 LVM (ISO 5832-1) to check the final implant position during radiography. The GMK Full PE Tibial Components can be used in place of the metal backed solution (K090988), where a component entirely made of UHMWPE replaces the metal backed solution (metallic tibial tray + UHMWPE tibial insert, K090988).

AI/ML Overview

The Medacta International GMK Full PE Tibial Components is a medical device and the provided text is a 510(k) summary for its clearance. This type of submission relies on demonstrating substantial equivalence to existing predicate devices rather than conducting extensive clinical studies with human participants that would typically involve acceptance criteria, ground truth establishment, or multi-reader multi-case studies as might be found for AI/ML devices.

Therefore, the information you've requested regarding acceptance criteria, study design, expert involvement, and statistical analyses of human vs. AI performance is not directly applicable or present in this 510(k) summary for a physical medical device.

However, I can extract the closest equivalent information for the "Performance Testing" section provided:


1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategoryReported Device Performance
BreakageEquivalent to cleared devices; capable of withstanding expected in vivo loading without failure.
Implant detachmentEquivalent to cleared devices; capable of withstanding expected in vivo loading without failure.
Excessive wear of the articulating surfaceEquivalent to cleared devices; capable of withstanding expected in vivo loading without failure.
Femoral luxation or subluxation due to insufficient constraintsEquivalent to cleared devices; capable of withstanding expected in vivo loading without failure.
Insufficient mobility and/or Range of Motion (ROM) according to ASTM 2083Equivalent to cleared devices; capable of withstanding expected in vivo loading without failure.

2. Sample size used for the test set and the data provenance

  • Test Set Sample Size: Not specified. The testing involves mechanical data and likely uses a defined number of device components for each test.
  • Data Provenance: The data is generated from "mechanical data" testing, which implies laboratory-based tests rather than patient data. Country of origin for data is not specified, but the applicant (Medacta International SA) is in Switzerland, and the contact person (Medacta USA) is in the US. The nature of the tests (e.g., ASTM standards) suggests standardized laboratory conditions. It is retrospective in the sense that it's performed on manufactured devices for regulatory submission, not a prospective clinical trial.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Number of Experts: Not applicable. The "ground truth" for mechanical performance tests is typically established by engineering standards and validated testing protocols, not by expert consensus on clinical findings.
  • Qualifications of Experts: Not applicable in the context of clinical expert review. The "experts" would be engineers and material scientists interpreting the test results against established safety and performance benchmarks for orthopedic implants.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Adjudication Method: Not applicable. This concept is relevant for clinical expert review of patient data, not mechanical performance testing. Test results are typically compared directly against pre-defined acceptance limits derived from engineering standards or predicate device performance.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • MRMC Study: No, an MRMC study was not done. This type of study is specific to evaluating diagnostic algorithms, often AI-powered, where human readers interpret medical images. The device in question is a physical knee prosthesis.

6. If a standalone (i.e. algorithm only, without human-in-the-loop performance) was done

  • Standalone Performance: Not applicable. This device is a physical implant, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • Type of Ground Truth: For mechanical testing, the "ground truth" is defined by engineering standards and specifications (e.g., ASTM standards like ASTM 2083 mentioned for ROM), and the performance of previously cleared predicate devices. The tests aim to demonstrate that the device performs equivalently or better than established benchmarks in terms of mechanical integrity and wear.

8. The sample size for the training set

  • Training Set Sample Size: Not applicable. This is not an AI/ML device requiring a training set. The design of the device would be based on engineering principles and knowledge of biomechanics, rather than a data training set.

9. How the ground truth for the training set was established

  • Ground Truth Establishment for Training Set: Not applicable, as there is no training set for this physical device. The design and validation are based on engineering principles, material science, and testing against established performance metrics and predicate device data.

{0}------------------------------------------------

K131310 (1/3)

Image /page/0/Picture/1 description: The image shows the logo for Medacta International. The logo consists of the word "Medacta" in a sans-serif font, with two triangles above the "M". Below the word "Medacta" is the word "International" in a smaller font. To the right of the word "International" is a symbol that looks like a plus sign.

510(k) Summary

Applicant/Sponsor: Medacta International SA Strada Regina 6874 Castel San Pietro (CH) Switzerland Phone (+41) 91 696 60 60 Fax (+41) 91 696 60 66

AUG 2 2 2013

Contact Person: Mr. Adam Gross Director of Regulatory, Quality and Compliance Medacta USA 4725 Calle Quetzal, Unit B Camarillo, CA, 93012 Phone: (805) 322-3289 Fax: (805) 437-7553 Email: AGross@medacta.us.com

Date Prepared: April 25, 2013

DEVICE INFORMATION

Trade/Proprietary Name: GMK Full PE Tibial Components Common Name: Total Knee Prosthesis Classification Name: Knee joint patellofemorotibial polymer/metal/polymer semiconstrained cemented prosthesis

21 CFR 888.3560 Class II Device Product Codes: JWH

Predicate Devices:
510(k)Product510(k) HolderClearance Date
K951987Genesis II CRSmith and Nephew, Inc.08/22/1995
K953274Genesis II PSSmith and Nephew, Inc.02/05/1996
K090988, K113571,K120790, K122232GMK Total KneeSystemMedacta International7/10/2009,1/27/2012,6/8/2012, 9/28/2012
K081023Evolis Total KneeSystemMedacta International10/22/2008
K991581, K042271NexGen LPSZimmer7/30/1999, 10/13/2004
K070214Natural KneeZimmer3/16/2007

{1}------------------------------------------------

Product Description

The GMK Full PE Tibial Components are symmetric and come in ultracongruent and posterior-stabilized designs in sizes 1-6 with thicknesses of 10, 12, 14, and 17mm and have an axial rotation of +/- 10°. The GMK Full PE Tibial Components are made from UHMWPE (ISO 5834 -2) Type 1. There are two radiopaque wires made of AISI 316 LVM (ISO 5832-1) to check the final implant position during radiography. The GMK Full PE Tibial Components can be used in place of the metal backed solution (K090988), where a component entirely made of UHMWPE replaces the metal backed solution (metallic tibial tray + UHMWPE tibial insert, K090988).

Indications for Use

The GMK® Total Knee System is designed for cemented use in total knee arthroplasty. if there is evidence of sufficient sound bone to seat and support the components.

This knee replacement system is indicated in the following cases:

  • Severely painful and/or disabled joint as a result of arthritis, traumatic arthritis, . rheumatoid arthritis or polyarthritis.
  • Avascular necrosis of femoral condyle. .
  • Post traumatic loss of joint confiquration. .
  • . Primary implantation failure.

Tibial augments are to be attached to the tibial baseplate with both the fixing cylinders and bone cement.

If a semi-constrained insert is used, it is mandatory to implant an extension stem both on the tibial and the femoral components.

Comparison to Predicate Devices

The indications for use, design features, and materials of the subject device are substantially equivalent to those of the predicate devices. The design features of both the GMK Full PE Tibial Components and the predicates come in ultracongruent and posterior-stabilized designs in similar sizes and thicknesses. The GMK Full PE Tibial Components and the predicates are both made of UHMWPE (ISO 5834 -2) Type 1 and can be used in place of the metal backed solution (metallic tibial tray + UHMWPE tibial insert).

The fundamental scientific technology of the modified devices has not changed relative to the predicate devices. The safety and effectiveness of the GMK Full PE Tibial Components are adequately supported by the substantial equivalence information. materials information, and analysis data provided within this Premarket Notification.

{2}------------------------------------------------

Performance Testing

The GMK Full PE Tibial Components was tested for the following compared to the predicate devices:

  • Breakage .
  • Implant detachment .
  • Excessive wear of the articulating surface .
  • Femoral luxation or subluxation due to insufficient constraints .
  • Insufficient mobility and/or Range of Motion (ROM) according to ASTM 2083 .

A review of the mechanical data indicates that the GMK Full PE Tibial Components is equivalent to devices currently cleared for use and is capable of withstanding expected in vivo loading without failure.

Conclusion:

Based on the above information, the GMK Full PE Tibial Components can be considered as substantially equivalent to its predicate devices.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the symbol. The logo is black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 22, 2013

Medacta International SA % Mr. Adam Gross Director of Regulatory, Quality and Compliance Medacta USA 4725 Calle Quetzal, Unit B Camarillo, California 93012

Re: K131310

Trade/Device Name: GMK Full PE Tibial Components Regulation Number: 21 CFR 888.3560 Regulation Name: Knee joint patellofemorotibial polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: JWH Dated: July 2, 2013 Received: July 3, 2013

Dear Mr. Gross:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{4}------------------------------------------------

Page 2 - Mr. Adam Gross

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resources/or You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours,

Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

510(k) Number (if known): K131310

Device Name: GMK Full PE Tibial Components

Indications for Use:

The GMK® Total Knee System is designed for cemented use in total knee arthroplasty, if there is evidence of sufficient sound bone to seat and support the components.

This knee replacement system is indicated in the following cases:

  • Severely painful and/or disabled joint as a result of arthritis, traumatic arthritis, ● rheumatoid arthritis or polyarthritis.
  • Avascular necrosis of femoral condyle. .
  • Post traumatic loss of joint configuration. .
  • Primary implantation failure. .

Tibial augments are to be attached to the tibial baseplate with both the fixing cylinders and bone cement.

If a semi-constrained insert is used, it is mandatory to implant an extension stem both on the tibial and the femoral components.

Prescription Use × (21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

్లో ప్రాథ్program - 1 - 1 - 1 - 3 - 3 - 3 - 3 - 3 - 3 - 3 - 3 - 3 - 3 - 3 - 3 - 3 - 3 - 3 - 3 - 3 - 3 - 3 - 3 - 3 - 3 - 3 - 3 - 3 - 3 - 3 - 3 - 3 - 3 - 3 - 3 - 3 - 3 - 3 - 3 - 3 - 3 --------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ev L. Hanley PhD1:("gs"ាន បើក្រោយប្រើប:::::נייני נו
n of Ortheedie Devices1.9Divi1(o)n0and and the market of the states of the states of the states of the states of the states of the states of the station of the states of the station of the station of the stati

GMK Full PE Tibial Components 510(k) May 3, 2013

Section 4 - Page 2 of 2

§ 888.3560 Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis.

(a)
Identification. A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).(b)
Classification. Class II.