Search Results
Found 19 results
510(k) Data Aggregation
(132 days)
The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. The Direct Screw to MUA may be used with single-stage or two-stage procedures, for screw-retained multi-unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading.
The digitally designed personalized copings to be used with the Direct Connection Screw to Multi-Unit Abutment are to be:
- Sent to Straumann for manufacture at a validated milling center, or
- Manufactured following the Neodent InLab Validated Workflow.
The Neodent InLab Validated Workflow is indicated for the design and fabrication of screw-retained multi-unit restorations for the restoration of partially or fully edentulous mandibles and maxillae. The system integrates multiple components: scan files from intra-oral scanners, CAD software, CAM software, milling machines and associated tooling and accessories.
The purpose of this premarket notification is to obtain market clearance to expand the options for fabrication of digitally designed personalized copings for use with the Neodent Direct Screw to multiunit abutment (MUA) from a "validated milling center" (cleared in K242686), to include a digital dentistry workflow.
The Neodent InLab Validated Workflow is integrated by multiple components of digital dentistry workflow: scan files from intra-oral scanners, CAD software, CAM software, milling machines and associated tooling and accessories. This new workflow employs optical impression files that document the topographical characteristics of teeth and traditional dental impressions, which are transferred to the CAD software (CARES® Visual, 3Shape or Exocad) and allows the design of the desired restorations, according to the Direct Screw to MUA dimensions. The CAM software converts the digital restoration design into the tooling and tool path commands needed to fabricate the restoration by the indicated milling equipments (Roland and Zirkonzahn).
N/A
Ask a specific question about this device
(163 days)
GM Helix Implant:
The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. The GM Helix Implant System with 3.75 and 4.0 diameter and 18 mm length are indicated for surgical installation in the pterygoid region (upper jaw), for multiple unit restorations, when combined with 45° angulation abutment, in cases of severe jaw resorption, in order to restore patient esthetics and chewing function. Implants in the pterygoid indication should be used in splinted applications that utilize at least two implants.
GM Helix LG Implant:
The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. The Neodent GM Helix LG implants can be placed bicortically in cases of reduced bone density. The Neodent GM Helix LG implants are only indicated for multiple unit restorations in splinted applications that utilize at least two implants. The GM Helix LG Implant System is indicated for surgical installation in the pterygoid region (upper jaw), for multiple unit restorations, when combined with 45° angulation abutment, in cases of severe jaw resorption, in order to restore patient esthetics and chewing function.
GM Mini Conical Abutment 45°:
The Mini Conical Abutments 45° and 60° are indicated for use with Zygomatic Implants, in cases of severe jaw resorption, in order to restore patient aesthetics and chewing function. The Mini Conical Abutments 45° are also indicated for use with Pterygoid implants in cases of severe jaw resorption, in order to restore patient esthetics and chewing function. The Mini Conical Abutments may be used with single stage or two-stage procedures, for multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading.
GM Mini Conical Abutment 45° Slim:
The Mini Conical Abutments 45° and 52° are indicated for surgical procedures in Zygomatic bones, making possible the rehabilitation with screw-retained abutments over the implant, thus restoring the chewing function. The Mini Conical Abutments 45° are also indicated for use with Pterygoid implants in cases of severe jaw resorption, in order to restore patient esthetics and chewing function. The Mini Conical Abutments may be used in one- or two-stage procedures, multiple unit restorations, and immediate loading when there is primary stability and adequate occlusal load. Multiple rehabilitations may be splinted rigidly.
This premarket notification seeks to expand the indications for use of the subject implants and abutments, in order to include anchorage in the pterygoid region. These devices were previously cleared under 510(k) K163194, K190958, K190718 and K232099 for the functional and esthetic oral rehabilitation of the upper or lower jaw in edentulous or partially edentulous patients. The proposed implants and abutments remain with the same design and features already cleared in the original submission.
N/A
Ask a specific question about this device
(200 days)
Zirconia Implants:
The Neodent Implant System is intended to be surgically placed in the bone of upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage surgical procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with physiological occlusion loading. Multiple teeth applications can be rigidly splinted. The implants with length of 5 mm (short implants) may be used only with two-stage surgical procedures. The recommended healing time before loading is between 10 to 12 weeks.
Zi Transmucosal Cover Screw and Healing:
The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage surgical procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with physiological occlusal loading. Multiple teeth applications can be rigidly splinted.
Zi Transmucosal Provisional Coping:
The Neodent Implant System is intended for surgical procedures in maxilla or mandible, providing support for prosthetic devices such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage procedures, for single- or multi-unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading.
Zi Transmucosal Abutment Replacement Screw:
The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading.
Zi Transmucosal Universal Base:
The Universal Ceramic Base Zi Transmucosal 5.0 is an abutment placed over Neodent Zi Transmucosal 5.0 Ceramic Implant System in order to provide support for custom-made prosthetic restorations, such as copings or crowns. It may be used for cement or screw-retained single unit restorations. All digitally designed copings and/or crowns to be used with the Neodent Zirconia Base Abutment System are intended to be sent to Straumann for manufacture at a validated milling center.
Zirconia Base for Bridge:
The Zirconia Base for Bridge is an abutment placed over Neodent Zirconia Implants in order to provide support for custom-made prosthetic restorations. It may be used for cement or screw-retained multi-unit restorations. All digitally designed copings and/or crowns to be used with the Neodent Zirconia Base Abutment System are intended to be sent to Straumann for manufacture at a validated milling center.
Zirconia Base C:
The Zirconia Base C is an abutment placed over Neodent Zirconia Implants in order to provide support for customized prosthetic restorations, such as copings or crowns. It may be used for single-unit restorations that are screw- or cement-retained in esthetic areas over implants installed in the maxilla or mandible. All copings and/or crowns digitally designed for use with the Titanium Base C are to be designed using Sirona inLab software or Sirona CEREC Software and manufactured using a Sirona CEREC or inLab MC X or MC XL milling unit.
This premarket notification includes new ceramic devices into Neodent Implant System, which are compatible with Zirconia Implant System. The Zirconia Implants and Abutments proposed on this submission are similar to devices already cleared in previous submissions of Neodent Implant System – Zirconia Implant System, according to predicate devices described above. This submission intends to expand the portfolio with new solutions and diameter, in order to provide more treatment options to the customers.
The Zirconia Implants are manufactured in Zirconia Y-TZP and are available in Bone Level (BL) or Tissue Level (TL or Transmucosal) configurations. The Zirconia Implants (BL) are available in a diameter of 5.0 mm and lengths in a range of 8 to 13 mm. The Zi Transmucosal Implants (TL) are available in a diameter of 5.0 mm and lengths in a range of 5 to 11.5 mm.
The Zi Transmucosal Healing and Cover Screw are temporary abutments manufactured in PEEK and used during the healing phase. They are compatible with the Zi Transmucosal Implants Ø5.0. The Zi Transmucosal Healing Abutment is available in the heights of 2 and 3.5mm.
The Zi Transmucosal Provisional Coping is a temporary abutment made of polycarbonate (PC) and has a double function: used for molding procedures or production of provisional restoration.
The Zi Transmucosal Abutment Replacement Screw is a prosthetic component manufactured in titanium alloy and used to fix the fix the Zi Transmucosal Base to the Zi Transmucosal Implant.
The Zi Transmucosal Universal Base is a two-piece abutment of base and top-half prosthetic structure to provide support for customized single-unit restorations over Zi Transmucosal Implant (TL). The base is manufactured in Zirconia Y-ZTP and used with a patient-specific top-half prosthetic structure. The two-piece abutment has a cementable portion of 4mm and is available with gingival heights of 0.3, 1.0 and 1.5 mm. The top-half prosthetic structure to be used with Zi Transmucosal Universal Base must be designed and milled in a Straumann Validated Milling center, using the following restoration materials and dimensions:
Material: IPS e.max CAD HT, Associated Material 510(k): K132209, Minimum wall thickness: 0.9 mm, Maximum angulation: 30°
Material: IPS e.max CAD LT, Associated Material 510(k): K132209, Minimum wall thickness: 0.9 mm
Material: N!ce, Associated Material 510(k): K171773, Minimum wall thickness: 1.0 mm
Material: IVOCLAR Multilink cement, Associated Material 510(k): K130436, Minimum wall thickness: N/A
Material: Zirconia N!ce® LT, Associated Material 510(k): K222836, Minimum wall thickness: 0.4 mm
Material: Zirconia N!ce® HT, Associated Material 510(k): K222836, Minimum wall thickness: 0.4 mm
Material: Zirconia N!ce® XT, Associated Material 510(k): K222836, Minimum wall thickness: 0.4 mm
Material: PMMA N!ce, Associated Material 510(k): K071548, Minimum wall thickness: 0.7 mm
Material: Panavia—Kuraray Cement, Associated Material 510(k): K150704, Minimum wall thickness: N/A, Maximum angulation: N/A
The Zi Base for Bridge is a two-piece abutment of base and top-half prosthetic structure to provide support for customized multi-unit restorations over Zirconia Implants (BL). The base is manufactured in Zirconia Y-ZTP and used with a patient-specific top-half prosthetic structure. The two-piece abutment has a cementable portion of 4mm and is available with gingival heights of 1.5, 2.5 and 3.5 mm. The top-half prosthetic structure to be used with Zi Base for Bridge must be designed and milled in a Straumann Validated Milling center, using the following restoration materials and dimensions:
Material: Zirconia N!ce® LT, Associated Material 510(k): K222836, Minimum wall thickness: 0.4 mm, Maximum angulation: 30°
Material: Zirconia N!ce® HT, Associated Material 510(k): K222836, Minimum wall thickness: 0.4 mm
Material: Zirconia N!ce® XT, Associated Material 510(k): K222836, Minimum wall thickness: 0.4 mm
Material: PMMA N!ce, Associated Material 510(k): K071548, Minimum wall thickness: 0.7 mm
Material: Panavia—Kuraray Cement, Associated Material 510(k): K150704, Minimum wall thickness: N/A, Maximum angulation: N/A
The Zi Base C is a two-piece abutment of base and top-half prosthetic structure to provide support for customized single-unit restorations over Zirconia Implants (BL). The base is manufactured in Zirconia Y-ZTP and used with a patient-specific top-half prosthetic structure. The two-piece abutment has a cementable portion of 4mm and is available with gingival heights of 1.5, 2.5, 3.5 and 4.5 mm. The top-half prosthetic structure to be used with Zi Base C must be designed and milled in a Sirona InLab Validated Workflow, using the following restoration materials and dimensions:
Material: IPS e.max CAD, Associated Material 510(k): K132209, Minimum wall thickness: 0.9 mm, Maximum angulation: 20°
Material: IVOCLAR Multilink cement, Associated Material 510(k): K130436, Minimum wall thickness: N/A, Maximum angulation: N/A
All these abutments have an internal connection with the implants (ZiLock) and the prosthetic platform is identical for all subject devices described in this submission. They are intended for single use and provided sterile via Ethylene Oxide method, along with undergoing moist heat sterilization after end-user customization.
The provided FDA 510(k) clearance letter and its associated summary for the Neodent Implant System - Zirconia Implant System contain extensive information about the device, its intended use, and comparisons to predicate devices. However, it does not include specific acceptance criteria with numerical thresholds directly stated within the tables, nor does it detail a study that directly proves the device meets such criteria in terms of performance metrics like sensitivity, specificity, or image quality assessments.
Instead, the submission focuses on demonstrating substantial equivalence to predicate devices through various tests, implying that if the new device performs similarly to or better than previously cleared devices, it meets the necessary standards. The performance testing section describes the types of tests conducted (e.g., dynamic fatigue, torsion, insertion, pull-out, and software validation), but it does not present clear quantitative acceptance criteria or the specific performance results in a comparative table format.
Therefore, many of the requested fields cannot be directly extracted from the provided text as they pertain more to the performance evaluation of AI/software in interpreting medical images, which is not the primary focus of this dental implant submission.
Here's an attempt to answer the questions based on the available information, noting where information is not explicitly provided in the document:
Acceptance Criteria and Device Performance Study for Neodent Implant System - Zirconia Implant System
The FDA 510(k) summary for the Neodent Implant System - Zirconia Implant System focuses on demonstrating substantial equivalence to predicate devices through a combination of bench testing, software validation, MRI compatibility, biocompatibility, and sterilization validation. It does not present specific quantitative acceptance criteria or performance metrics directly from a comparative study in the way one might expect for an AI/software-based medical device (e.g., sensitivity, specificity thresholds). Instead, the "acceptance criteria" are implicitly met by demonstrating that the proposed devices perform at a level substantially equivalent to legally marketed predicate devices under standardized testing conditions.
1. A table of acceptance criteria and the reported device performance
As mentioned, explicit numerical acceptance criteria and reported device performance in a comparative table (e.g., for diagnostic accuracy) are not provided in this 510(k) summary. The summary indicates that tests were conducted according to relevant ISO standards and FDA guidance, and the results demonstrated that the subject devices exhibit a level of performance substantial equivalent to the predicate and reference devices.
Below is a conceptual table based on the types of tests mentioned, noting that specific numerical acceptance criteria and performance data are not detailed in the provided text.
| Acceptance Criteria Category (Implicit) | Standard/Guidance | Reported Device Performance (Summary) |
|---|---|---|
| Dynamic Fatigue Strength | ISO 14801, FDA Guidance (Class II Special Controls) | Demonstrated a level of performance substantial equivalent to predicate and reference devices in identical conditions. |
| Torsion Strength | Not specified | Adequate torsion strength in accordance with recommended IFU installation torque. |
| Insertion Torque | Not specified | Evaluated insertion torque in sawbones material (Bone type I, II, III, IV). (Specific values not given). |
| Implant Surface Area & Pull-Out Strength | Not specified | Greater surface area compared to reference devices; higher resistance values in Pull Out Test. |
| Software Validation (Sirona Digital Workflow) | Not specified | Accuracy requirement was met; critical design parameters (min wall thickness, max angulation) respected and monitored. |
| MRI Compatibility | K182620, FDA Guidance | MR conditional labeling from K182620 is applicable; safe for scanning under previously established parameters. |
| Biocompatibility | ISO 10993-1, ISO 10993-18, ISO 10993-5, FDA Guidance | Subject devices are equivalent in material and manufacturing processes to predicates; no new issues raised; no additional testing required. |
| Sterilization Validation (Ethylene Oxide) | ISO 11135:2014 | Validated to a Sterility Assurance Level (SAL) of 1x10⁻⁶; residuals below max allowable limits per ISO 10993-7. |
| Sterilization Validation (Moist Heat) | ISO 17665-1 | Validated using parameters described in IFU. |
| Endotoxin Test | ANSI/AAMI ST72:2011, ISO 11737-3, US Pharmacopeia chapter 85 | Results <0.05 EU/device. |
| Shelf Life | Not specified | Determined to be 5 years; packaging identical to predicates. |
2. Sample size used for the test set and the data provenance
The document details various bench tests (dynamic fatigue, torsion, insertion, pull-out) and validation studies (software, MRI, biocompatibility, sterilization, endotoxin, shelf-life). However, specific numerical sample sizes ("n=") for these individual tests are not provided in the summary. For example, for dynamic fatigue testing, it states "the results demonstrated that...". It doesn't specify how many implants or setups were tested.
- Data Provenance (Country of Origin): The submitter is JJGC Indústria e Comércio de Materiais Dentários S.A. (dba Neodent) based in Curitiba, Paraná, Brazil. The tests are generally conducted to international ISO standards and FDA guidance, but the location where the tests were performed is not explicitly stated for each test beyond the company's Brazilian location.
- Retrospective or Prospective: Not applicable in the context of bench testing and validation studies as described. These are laboratory-based tests rather than clinical studies on patient data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This information is not applicable and not provided in the context of this 510(k) submission. The device is a dental implant system (hardware), not an AI/software device that interprets medical images requiring expert-established ground truth for a test set. The validation processes involve standardized engineering and biological tests as opposed to expert review of clinical cases.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This information is not applicable and not provided. As explained above, the submission pertains to a hardware device, not an AI/software system requiring human adjudication for ground truth establishment.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This information is not applicable and not provided. This type of study is relevant for AI-assisted diagnostic tools, not for dental implant hardware.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This information is not applicable and not provided. This concept is specific to AI/software performance, not a physical dental implant system. The "software validation" mentioned is for the compatibility of milling units (CAD/CAM workflow), ensuring design parameters are respected, not for an AI algorithm interpreting dental images.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
For the bench tests, the "ground truth" is defined by compliance with established engineering standards (e.g., ISO 14801 for dynamic fatigue) and specifications (e.g., material properties, dimensions). For biological safety, it's compliance with ISO 10993. For the CAD/CAM software, it's the accurate adherence to design parameters. This isn't "ground truth" in the diagnostic sense, but rather adherence to predefined engineering and biocompatibility specifications.
- Clinical Literature Review: A clinical literature review was conducted to support the safety of Zi Transmucosal Implant less than 7mm of length, referencing animal studies and histomorphometry evaluation. This aligns with outcomes data or pathological/histological evidence from preclinical studies, suggesting "similar osseointegration outcomes" and "ability to promote bone formation around them in a manner equivalent to titanium implants."
8. The sample size for the training set
This information is not applicable and not provided. This concept is specific to machine learning/AI models, which are not the subject of this 510(k) submission.
9. How the ground truth for the training set was established
This information is not applicable and not provided. As above, this is relevant for AI/ML training data, not for a physical dental implant system.
Ask a specific question about this device
(88 days)
Custom Abutment AS Ti:
The Custom Abutment with Angled Screw Channel is a customized prosthetic abutment, manufactured in titanium alloy, placed onto dental implants to provide support for customized prosthetic restorations. All abutments are only intended to be digitally designed and manufactured using specifics CAD/CAM software according to digital dentistry workflow. Custom Abutments with Angled Screw Channel are indicated for screw-retained single restorations or cemented retained single or multiple restorations. All digitally designed abutments for use with the Custom Abutment Ti with Angled Screw Channel are intended to be sent to Straumann for manufacturing at a validated milling center.
ASC Screw:
The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading.
This premarket notification includes new digital abutments to the Neodent Implant System (GM, NGM and HS prosthetic interfaces). The abutments proposed on this submission are identical to devices already cleared in previous submissions of Neodent Implant System, according to predicate devices described above, being the only difference between them the introduction of a new angled channel solution. This submission intends to expand the portfolio of Neodent Implant System. The Custom Abutments AS Ti are composed of a unique body with two different regions: the upper region, which is the customizable portion, and the end region presents the prosthetic interface that fits with the implant, which does not allow customization. They must be sent to a Straumann Validated Milling
This document describes the Neodent Implant System - Custom Abutments and its FDA 510(k) clearance (K250614). The information provided is primarily focused on demonstrating substantial equivalence to a predicate device, rather than detailed performance study results with specific acceptance criteria and performance metrics for a novel AI/software device.
Based on the provided text, the device is a physical medical device (dental implant abutment) and not an AI/software device. Therefore, the questions related to AI/software performance studies, such as sample size for test sets, expert adjudication, MRMC studies, standalone algorithm performance, and training set details, are not applicable to this submission.
The "Performance Testing" section primarily describes bench testing (dynamic fatigue and torsion tests), biocompatibility testing, and sterilization validation, which are standard for physical medical devices.
Here's a breakdown of the available information:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly state "acceptance criteria" in the format typically seen for performance metrics of a diagnostic device. Instead, it describes general compliance with standards and the determination of "not a new worst case" for dynamic fatigue.
| Acceptance Criterion (Inferred from testing) | Reported Device Performance |
|---|---|
| Dynamic Fatigue Testing (ISO 14801 compliant, FDA Guidance) | Subject devices are not a new worst case compared to predicate devices. (No specific values reported in this document). |
| Torsion Testing | Results prove adequate torsion strength in accordance with recommended installation torque. (No specific values reported). |
| Biocompatibility (ISO 10993-1, ISO 10993-5, ISO 10993-18 compliant) | Subject devices are equivalent in material and manufacturing processes to predicate devices; no new issues raised, no additional testing required. |
| Sterilization Validation (ISO 17665-1 compliant) | Sterilization method (moist heat, SAL of 1x10-6) was validated and is identical to predicate devices. |
| MRI Compatibility | MR conditional labeling from predicate (K182620) is applicable to subject devices. |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Bench Testing (Dynamic Fatigue and Torsion): Not
explicitly stated. The document mentions "a worst case analysis determined that the subject devices are not a new worst case compared to those devices tested for the primary predicate." This implies that testing was performed on a sufficient number of samples to make this determination, likely following the requirements of ISO 14801. - Data Provenance: Not explicitly stated, but these are typically laboratory-based bench tests performed by the manufacturer or a contracted testing facility.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
- Not Applicable. This is a physical device, and the testing involves objective measurements (e.g., force, torque, material properties) rather than expert interpretation of data for ground truth establishment.
4. Adjudication Method for the Test Set
- Not Applicable. As above, the tests are objective physical measurements.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done
- Not Applicable. This is a physical device, not an AI-assisted diagnostic or decision support system.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done
- Not Applicable. This is a physical device. While it involves digital design using CAD/CAM software, the 'device' itself is the physical abutment, and its performance is evaluated through physical and material tests, not software algorithm performance.
7. The Type of Ground Truth Used
- Physical/Material Standards and Engineering Specifications. For dynamic fatigue and torsion tests, the "ground truth" or acceptance criteria are derived from established engineering principles, material properties (ASTM F136 for titanium alloy), and international standards (ISO 14801). For biocompatibility, it's compliance with ISO 10993 series.
8. The Sample Size for the Training Set
- Not Applicable. This is a physical device; there is no "training set" in the context of machine learning or AI models.
9. How the Ground Truth for the Training Set Was Established
- Not Applicable. As above, no training set for an AI/ML model for this physical device.
Ask a specific question about this device
(103 days)
Patient-specific abutment restorations, milled from Pre-milled Abutment Blanks (PMABs), are indicated for single tooth replacement and multiple tooth restorations. They are directly connected to various endosseous dental implant systems using a basal screw. Patient-specific abutment restorations milled from Pre-milled Abutment Blanks are to be digitally designed and milled using the Straumann InLab Validated Workflow. The Straumann InLab Validated Workflow is indicated for the design and fabrication of single or multiple-unit implant-borne prosthetics for the restoration of partially or fully edentulous mandibles and maxillae. The system integrates multiple components of the digital dentistry workflow: scan files from Intra-Oral Scanners or Extra-Oral Scanners. CAD software, CAM software, pre-milled abutment blanks, milling machines and associated tooling and accessories.
The Straumann InLab Validated Workflow is similar to the primary predicate K171649. It employs optical impression files that document the topographical characteristics of teeth, traditional dental impressions, or stone models. The 3Shape CAD software then allows the design of the desired restorations.
The CAM software converts the digital restoration design into the tooling and tool path commands needed to fabricate the restoration. When choosing the Straumann Validated workflow, the user will only see the available and cleared components which were tested and demonstrated as part of the validated workflow. The milling command file is encrypted prior to transfer to the Roland DWX-42W Plus milling System; this encryption ensures that files generated using other CAD or CAM software cannot be used with the Straumann InLab Validated Workflow. The user will then load the milling command file into the Roland DWX-42W Plus milling System where it is decoded. The user loads the appropriate dental material blank and initiates the milling operation.
This premarket notification includes restorations (one-piece metal patient-specific abutment restorations) manufactured from various Pre-milled Abutment Blanks (PMABs) from the Straumann Group companies : Institut Straumann AG and Neodent PMABs. The digital workflow using the Straumann InLab Validated Workflow includes the use of the following products: Dental Scan of the patient's situation, CAD Software, CAM Module, Milling System, Abutment Milling Blanks.
The provided text is a 510(k) Summary for the Straumann InLab Validated Workflow, a device for designing and manufacturing patient-specific dental abutments from pre-milled abutment blanks (PMABs). The document describes the device, its intended use, technological characteristics, and performance testing.
Here's a breakdown of the requested information based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly present a table of acceptance criteria and reported device performance in a consolidated format as typically seen in a clinical study report. However, it mentions key performance areas and states that requirements were met.
Acceptance Criteria (Implicit from tests performed):
- Sterilization Validation: Meet ISO 17665-1 and FDA guidance for reprocessing medical devices.
- Biocompatibility: In line with ISO 10993-1 and FDA guidance for biological evaluation.
- Electromagnetic Compatibility (EMC): Considered MR Conditional.
- Milling Accuracy: Dimensions of milled restoration are the same as the intended CAD design, met within tool's expected lifetime.
- Simulated Use Validation: Correct implementation of PMABs, design constraints, and workflow restrictions in software and libraries.
- Implant-Abutment Connection Protection: No damage to the connection geometry during milling.
- Dynamic Fatigue: Performance consistent with FDA guidance for dental implants/abutments.
Reported Device Performance:
| Performance Area | Reported Device Performance |
|---|---|
| Sterilization Validation | The recommended sterilization method (moist heat/steam) was validated according to ISO 17665-1 and applicable FDA guidance on reprocessing medical devices. Sterilization parameters are equivalent. |
| Biocompatibility Testing | Biological assessment performed per ISO 10993-1 and FDA guidance. Device materials and manufacturing processes are equivalent to predicate, raising no new biocompatibility issues. |
| Electromagnetic Compatibility | No significant changes from currently marketed predicate devices; no new EMC issues raised. Device can be considered MR Conditional. |
| Milling Accuracy | Black-box validation confirmed that dimensions of the milled restoration are the same as the intended CAD design. The accuracy requirement was met considering the tool's expected lifetime. |
| Simulated Use Validation | Confirmed that relevant PMABs, design constraints, and workflow restrictions are correctly implemented (adequately written and locked into compatible design software and available libraries). |
| Implant-Abutment Connection Protection | A simulated implant-abutment connection protection test was conducted to mitigate potential risk of damaging the implant-abutment connection geometry during milling. |
| Dynamic Fatigue | The device design and performance testing submitted or referenced were conducted according to FDA guidance for root-form endosseous dental implants and abutments. |
2. Sample Size Used for the Test Set and Data Provenance
The document does not specify the exact sample sizes used for the test sets in the performance studies (e.g., dynamic fatigue, milling accuracy, simulated use validation). It refers to these as "Performance Testing - Bench."
Data Provenance: The studies are described as "nonclinical tests" and "bench" testing, suggesting they were conducted in a laboratory or simulated environment, rather than involving patient data. The document does not specify country of origin for the data (beyond the applicant being Institut Straumann AG in Switzerland and JJGC Indústria e Comércio de Materiais Dentários AS in Brazil, with contact in Andover, MA, USA). The studies appear to be prospective, designed specifically for this submission.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
The document does not mention the use of experts to establish ground truth for the test set. The validation methods described (e.g., milling accuracy, simulated use) imply technical verification against design specifications and functional requirements rather than expert consensus on diagnostic or clinical outcomes.
4. Adjudication Method for the Test Set
No adjudication method is mentioned, as there is no indication of multiple reviewers or subjective assessments requiring adjudication in the described bench tests.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No MRMC comparative effectiveness study is mentioned. The device is a workflow for manufacturing dental restorations, not an AI diagnostic or assistive tool for human readers in the traditional sense of comparing human performance with and without AI assistance for interpretation.
6. Standalone Performance (Algorithm Only Without Human-in-the-Loop)
The milling accuracy test ("black-box validation confirming that the dimensions of the milled restoration are the same as the intended CAD design") and the simulated use validation implicitly test the algorithm and workflow in a standalone manner, separate from a human operator's interpretative judgment. The "Straumann InLab Validated Workflow" itself is largely an automated process once the design is finalized. It is designed to ensure that the digital design translates accurately into a physical product.
7. Type of Ground Truth Used
The ground truth used for these performance tests appears to be:
- Design Intent/Specifications: For milling accuracy, the ground truth is the CAD design itself. For simulated use, the ground truth refers to the correct implementation of design constraints and workflow restrictions within the software and libraries.
- Industry Standards and Regulatory Guidance: For sterilization, biocompatibility, and dynamic fatigue, the ground truth is defined by recognized international standards (ISO) and FDA guidance documents.
- Functional Requirements: For implant-abutment connection protection, the ground truth is the prevention of damage to the connection geometry.
8. Sample Size for the Training Set
The document does not describe the use of machine learning or AI models in a way that would involve a distinct "training set" in the context of typical software or diagnostic AI submissions. The workflow involves CAD/CAM software where parameters and constraints are defined by engineering and design principles rather than learned from a large training dataset.
9. How the Ground Truth for the Training Set Was Established
As no specific training set for a machine learning model is described, there is no information on how its ground truth would have been established. The workflow relies on validated engineering designs, material properties, and manufacturing processes.
Ask a specific question about this device
(89 days)
NeoConvert Solution: The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. The NeoConvert Solution may be used with single-stage or two-stage procedures, for temporary multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading.
Direct Screw to MUA: The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. The Direct Screw to MUA may be used with single-stage or two-stage procedures, for screw-retained multi-unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. All digitally designed personalized copings to be used with the Direct to Multi Unit Abutment Screw are intended to be sent to Straumann for manufacture at a validated milling center.
GM Attachment TiN 30°: The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. The GM Attachment TiN 30° may be used with single-stage or two-stage procedures, for multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading.
This premarket notification includes the following additions to the Neodent Implant System: NeoConvert Solution, Direct Screw to MUA, GM Attachment TiN 30°. The Neodent Multi-Unit Abutments proposed on this submission are similar to devices already cleared in previous submissions of Neodent Implant System, according to predicate devices described above. This submission intends to expand the portfolio of Neodent Abutments with these new solutions for multi-unit restorations, to provide more treatment options to customers. The NeoConvert Solution is a technique comprised by some abutments (not subject of this submission) used for converting removable temporary prosthesis into a screw-retained fixed temporary prosthesis. The Direct Screw to MUA is a prosthetic component designed to fix planned and digitally designed multi-unit milled restorations directly onto mini or micro abutments, eliminating the need for using the passive seating cylinder to restore chewing function. The GM Attachment TiN 30° is a prosthetic component used to stabilize implant-mucosa-supported removable full prosthesis (retained on the implant and supported on the mucosa), on implants installed in the maxilla or mandible. Although all these three new solutions are compatible with the same implants and components of Neodent Implant System and used for multiple unit rehabilitation, they are not used in conjunction with each other.
The provided text describes a 510(k) premarket notification for the "Neodent Implant System," focusing on the NeoConvert Solution, Direct Screw to MUA, and GM Attachment TiN 30°. This document details the device's technical characteristics, intended use, and its substantial equivalence to previously cleared predicate devices.
However, the provided text does not contain information about acceptance criteria and a study proving a device meets those criteria in the context of an AI/ML medical device submission. The document is a traditional 510(k) submission for dental implants and abutments, which are physical medical devices, not AI/ML software. Therefore, the specific requirements for an AI/ML device study (such as sample sizes for test and training sets, expert qualifications for ground truth, MRMC studies, or standalone performance) are not applicable or present in this document.
The "Performance Testing" section (pages 15-16) describes bench testing (dynamic fatigue, torsion, SEM images), MRI compatibility testing, biocompatibility testing, sterilization validation, and shelf-life testing for the physical dental implant components. These are standard performance tests for this type of physical device, not AI/ML algorithm validation.
Therefore, I cannot provide a table of acceptance criteria and a study proving the device meets acceptance criteria as typically outlined for an AI/ML medical device, as this information is not present in the provided document. The document focuses on demonstrating substantial equivalence of a physical dental device to predicate devices through engineering and material testing.
Ask a specific question about this device
(149 days)
The Custom Abutment Ti with Screw is a customized prosthetic abutment, manufactured in titanium alloy, placed onto dental implants to provide support for customized prosthetic restorations. All abutments are only intended to be digitally designed and manufactured using specifics CAD/CAM software according to digital dentistry workflow. Custom Abutments Ti with Screw are indicated for screw-retained single restorations or cemented-retained single or multiple restorations. All digitally designed abutments for use with the Custom Abutment Ti with Screw are intended to be sent to Straumann for manufacturing at a validated milling center.
The Custom Abutment with Angled Screw Channel is a customized prosthetic abutment, manufactured in titanium alloy, placed onto dental implants to provide support for customized prosthetic restorations. All abutments are only intended to be digitally designed and manufactured using specifics CAD/CAM software according to digital dentistry workflow. Custom Abutments with Angled Screw Channel are indicated for screw-retained single restorations or cemented-retained single or multiple restorations. All digitally designed abutments for use with the Custom Abutment Ti with Angled Screw Channel are intended to be sent to Straumann for manufacturing at a validated milling center.
The Custom Abutments subject to this submission are similar to the devices already cleared in previous Neodent and Nuvo Implant System submissions, per the predicate and reference devices described above. This submission intends to expand the portfolio of digital abutments for the Grand Morse (GM), Narrow Grand Morse (NGM) and Helix Short (HS) lines, to provide more treatment options for customers. These abutments are composed of a unique body with two regions: the upper region is the customized portion, while the end region presents the prosthetic interface that fits with the implant, which does not allow customization.
They are intended for single use and provided non-sterilization recommended before installation in the mouth.
All proposed abutments are delivered to the final user already customized in a validated milling center, with a straight channel solution. For the GM line, the abutments could also be milled with a new angled channel solution for the screw access. All subject devices must be used with their corresponding prosthetic interface.
The provided text describes the Neodent Implant System - Custom Abutments, a medical device. This product is not an AI/ML device, and therefore does not have acceptance criteria or a study that proves the device meets those criteria in the way requested in the prompt.
The document discusses the substantial equivalence of the Neodent Implant System - Custom Abutments to legally marketed predicate devices, which is the basis for its 510(k) clearance by the FDA. The performance testing section describes bench tests (dynamic fatigue and torsion tests), sterilization validation, MR compatibility testing, and biocompatibility testing. These tests evaluate the physical, chemical, and biological properties of the device, comparing them to established standards and predicate devices to demonstrate equivalence, not to meet a set of acceptance criteria for an AI/ML algorithm's performance.
Because the device is not an AI/ML device, the specific information requested about acceptance criteria for AI performance, sample sizes for test/training sets, data provenance, expert ground truth establishment, adjudication methods, MRMC studies, standalone performance, and how ground truth for training was established is not applicable and is not present in the provided text.
Ask a specific question about this device
(176 days)
The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage surgical procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with physiological occlusal loading. Multiple teeth applications can be rigidly splinted. Healing,
The Zirconia Base is an abutment placed over Neodent Zirconia Implants in order to provide support for custom-made prosthetic restorations, such as copings or crowns. It may be used for cement- or screw-retained single unit restorations. All digitally designed copings and/or crowns to be used with the Neodent Zirconia Base Abutment System are intended to be sent to Straumann for manufacture at a validated milling center.
The CR Abutment is an abutment placed over Neodent Zirconia Implants in order to provide support for prosthetic restorations, such as copings or crowns. It may be used for single unit restorations that are cement-retained in esthetical areas over implants installed in maxilla or mandible.
The PEEK CR Abutment is indicated to be used on Neodent Implants to provide support for prosthetic structures for up to 6 months. They can be used in single- or two-stage procedures and they are intended to be placed out of occlusion.
The Zirconia Abutments subject of this submission are similar to devices already approved in previous submissions of Zirconia Implant System, according to predicate devices described above. This submission intends to expand the portfolio of Zirconia Abutments with new dimensions of gingival height for Zirconia Healing Abutments, Zirconia Bases, CR Abutments for Zirconia and PEEK CR Abutments for Zirconia, to provide more treatment options to the customers. All these abutments have an internal connection with the implants (ZiLock) and the prosthetic platform is identical for all subject devices described in this submission. They are intended for single use and provided sterile via Ethylene Oxide method. The new transmucosal heights are 3.5mm and 4.5mm.
The restorative materials Polycon ae, N!ce and IPS e.max CAD LT were not presented in previous submissions of Zirconia Implant System and are being included in the scope of Zirconia Bases indication.
The provided document is a 510(k) summary for the Neodent Implant System - Zirconia Implant System. This type of FDA submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving a device meets specific acceptance criteria through a clinical study with defined performance metrics, sample sizes, and expert adjudication as might be seen for novel AI/software as a medical device (SaMD) products.
Therefore, the document does not contain the information requested regarding acceptance criteria related to a study proving performance, sample sizes for test sets, data provenance, number or qualifications of experts for ground truth, adjudication methods, MRMC studies, standalone algorithm performance, or ground truth establishment methods for training sets.
The performance testing section (pages 14-15) describes:
- Bench Testing:
- Dynamic fatigue testing: Conducted according to FDA guidance and ISO 14801. The results demonstrated the subject devices do not introduce a new worst case compared to predicate and reference devices.
- Torsion tests: Performed to evaluate screw strength, and results proved adequate torsion strength according to installation torque.
- Biocompatibility Testing:
- Assessment performed according to ISO 10993-1 and FDA guidance. No new issues were raised and no additional testing was required because the subject devices are identical in material and manufacturing processes to predicate and reference devices.
- Sterilization Validation and Packaging:
- Ethylene Oxide (EO) sterilization validated to SAL of 10-6 per ISO 11135:2014. EO residuals verified per ISO 10993-7. The packaging and sterilization are identical to predicate devices.
- Steam sterilization validated per ISO 17665-1, using parameters identical to predicate devices.
- Pyrogen limit specifications met via LAL Endotoxin Analysis (testing limit 20 EU/device) for blood contacting and implanted devices.
In essence, the study presented is a non-inferiority or equivalence-based assessment against previously cleared devices, focusing on engineering and biocompatibility characteristics, not clinical performance metrics or AI algorithm validation metrics.
Therefore, I cannot provide the requested table and details because the information is not present in the provided text. The document describes a traditional medical device (dental implants and abutments) clearance, not a software or AI device that would have the specific performance study criteria outlined in your request.
Ask a specific question about this device
(110 days)
Indications for Use for Zygoma-S GM Implant:
Zygomatic implants are indicated for intraoral surgical procedures in the zygoma region in cases of severe maxilla bone resorption, to restore the patient's chewing function and aesthetics. Zygomatic Implants may be used in one or two-stage procedures, multiple unit restorations, and immediate loading when there is primary stability and adequate occlusal load.
Indications for Use for GM Mini Conical Abutment 52° and 45°:
Product indicated for surgical procedures in zygomatic bones, making possible the rehabilitation with screw-retained abutments over the implant, thus restoring the chewing function. It may be used in one- or two-stage procedures, multiple unit restorations, and immediate loading when there is primary stability and adequate occlusal load. Multiple rehabilitations may be splinted rigidly.
Indications for Use for Coping for Removable Prosthesis:
The product, when used with non-zygomatic implants, is intended to be surgically placed in the bone of the upper or lower jaw, to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. It may be used with single stage or two-stage procedures, for multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with approprate occlusal loading. Multiple tooth applications may be rigidly splinted. When used with zygomatic implants, the product is indicated for surgical procedure only in upper jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. It may be used in two-stage procedures (delayed loading protocol) and for multiple unit restorations. Multiple rehabilitations may be splinted rigidly.
The subject devices implants are single use devices, provided sterile by Gamma Radiation, made of commercially pure Titanium grade 4 (ASTM F67). The Zygomatic Implant is a long implant with a nominal diameter of 3.50 and 3.75mm, conical apex with rounded tip, 3 helical chambers, trapezoidal thread and GM prosthetic interface. The cervical portion in both models has Ø 4.3 mm. The length varies from 30 to 55 mm. Zygomatic Implants are intended to provide support for a fixed prosthesis in patients with severe atrophy in the maxillary region. They are extra-long to allow bone anchorage in the zygomatic bone. The subject implants consider the Zygoma Anatomy-Guided approach (ZAGA) for planning the surgical steps, being indicated for extrasinus and extra maxillary installation technique (ZAGA-4).
The subject abutments are single use devices, provided sterile by Ethylene Oxide, made of Titanium alloy (Tl6Al4V-ELI). They are intermediate prosthetic components angled at 52 and 45 degrees with gingiva heights 1.5 and 2.5 mm, to be installed on the implant, offering a structure to support the screw-retained multiple prosthesis. The abutments present GM prosthetic interface and are available in the anti-rotational form in the implant to abutment interface, and rotational form in the abutment to prosthesis interface, presenting different transmucosal heights.
The subject coping is single use device, provided sterile by Ethylene Oxide, made of titanium alloy (Ti6Al4V-ELI). It presents a NEO torque interface for torque application. The subject device has diameter 4.80 mm and total height 4.60 mm. They are indicated for removable prostheses with fitting over Mini Conical Abutment installed in the maxilla or mandible and can be used in partial removable prostheses or over dentures. They are compatible with all previously cleared GM Mini Conical abutments.
The provided document is a 510(k) premarket notification for a medical device (Neodent Implant System - GM Zygomatic Implant System). It focuses on demonstrating substantial equivalence to predicate devices through comparisons of indications for use, technological characteristics, and performance data, primarily bench testing, sterilization validation, shelf life validation, and biocompatibility.
Crucially, this document does not contain information about an AI/ML algorithm or a study validating its performance, nor does it detail acceptance criteria for software or AI/ML components. The product being cleared is a set of dental implants and abutments. Therefore, I cannot extract the acceptance criteria and study details for an AI-powered device as requested.
The document discusses performance data in the context of the physical dental implants and abutments, which includes:
- Performance Testing – Bench: Dynamic fatigue testing (ISO 14801), insertion tests, torsion tests, and microscopic analysis of the implant body.
- Surface treatment: Discussion of the NeoPoros surface.
- MR Compatibility testing: Assessment of safety in MRI environments.
- Sterilization validation: Conformance to ISO standards for Gamma Irradiation (implants) and Ethylene Oxide (abutments).
- Shelf Life validation: Based on packaging integrity and physical stability.
- Biocompatibility: Testing per ISO 10993 standards.
- Clinical data: A review of published literature on zygomatic implants, including survival rates and observed complications (not a specific clinical study conducted for this device clearance).
If you are looking for information regarding acceptance criteria and study data for an AI/ML medical device, this document does not provide it, as it pertains to a different type of medical device.
Ask a specific question about this device
(200 days)
Indications for Use for Helix Short Implant:
The Neodent Implant System is recommended for surgical procedures on maxilla or mandible bones. It provides support for prosthetic components such as artificial teeth, thus restoring the chewing function. It may be used with single-stage or two-stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading.
Indications for Use for Mini Straight, Angled Abutment and screws:
The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading.
Indications for Use for Helix Short Attachment:
The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. It may be used with single stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading.
Indications for Use for Helix Short Healing Abutment:
This product is used for the maintenance of soft tissue, during the bone integration stage of Neodent implants to be rehabilitated with the late loading technique. The Healing Abutment may be used in the implant installation surgery or the reopening surgery (second surgical stage).
Indications for Use for Helix Short Cover Screw:
This product is used for the maintenance of soft tissue, during the bone integration stage of Neodent implants to be rehabilitated with the late loading technique. The Cover Screw must remain intragingival, preventing the growth of tissue over the implant platform. The Cover Screw takes place in the conical region of the implant interface.
Indications for Use for Temporary Abutment for Helix Short Implant
The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. It may be used with single stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. Temporary Abutments are installed over Helix Short Implants and to provide support for prosthetic structures for up to 6 months.
Indications for Use for Helix Short Titanium Base:
The Helix Short Implant Titanium base that is placed over Neodent dental implants to provide support for customized prosthetic restorations, such as copings and crowns. It is indicated for single- and multiple-structure restorations, screw- or cement-retained on implants installed in the maxilla or mandible. All digitally-designed copings and/or crowns to be used with the Neodent Titanium Base System must be sent to Straumann for manufacture at a validated milling center.
- All the subject devices are intended for single use;
- The subject implants are provided sterile via gamma irradiation and the subject abutments are provided sterile via Ethylene Oxide to an SAL of 1x10-9;
- The subject implants are manufactured of Commercially Pure Titanium (Grade 4), the subject abutments, prosthetic screws and abutment screws are manufactured of Titanium alloy Ti-6Al-4V ELI (ASTM F136) and the restorations placed onto Helix Short Titanium Bases are made of various top half materials.
- The subject implants have a wide cone prosthetic interface with an internal hexagon for short implants lengths available from 4.0 to 7.0; cylindrical shape with double threads and conical apex with three helical flutes for diameters from 3.75 to 7.0mm.
- The implants are provided in two different surfaces: Neoporos: Sand blasted and acid etched, and Acqua: Hydrophilic surface aggregated to the rough surface obtained by double treatment: abrasive blasting and acid subtraction.
- The Helix Short Implant Titanium Base abutments is used as two-piece abutment, where the base is premanufactured from titanium alloy and the top half is created via burn out coping or a digital workflow with CAD/CAM at a validated milling center. The final top half can be created from Ticon, Coron, Zerion LT, IPS e.max, or PMMA. The top half and base pieces are cemented together to form the final abutment.
Here's an analysis of the acceptance criteria and supporting studies for the Neodent Implant System - Helix Short Implant System, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The FDA 510(k) summary does not explicitly present a table of specific numerical acceptance criteria alongside reported performance for each test. Instead, it describes what was tested and states that the results "met the acceptance criteria" or demonstrated "equivalent performance." We can infer the general acceptance criterion from the study descriptions.
| Acceptance Criteria (Inferred) | Reported Device Performance |
|---|---|
| Mechanical Testing: | |
| Dynamic Fatigue Test (per ISO 14801): Equivalent fatigue strength to predicate devices. | The tested subject devices exhibit a level of performance equivalent to that reviewed for the predicate devices. |
| Torsion Test (static torsional loading): Meet acceptance criteria. | The results met the acceptance criteria. |
| Insertion Test: Evaluate insertion torque in various bone types. | Performed to evaluate the insertion torque of the Helix Short System when it is inserted in bones type I, II, III, and IV. (No explicit "met criteria" statement, but implied by successful evaluation). |
| Implant Surface Area Simulation and Pull Out Test: Demonstrate acceptable BIC rate and pull out resistance. | Results demonstrated that the subject devices, despite having a slightly smaller surface area than the reference devices, presented a higher BIC rate and pull out resistance upon placement. Thus, the subject devices exhibit a level of benchtop performance equivalent to that reviewed for the reference device. |
| MR Compatibility Testing: | |
| Not configure a new worst case compared to reference devices, allow safe scanning under established parameters. | An assessment was made to demonstrate that the subject devices do not configure a new worst case and can be represented by the previously conducted studies reviewed for reference devices, since both have the same raw material and similar dimensions. The subject devices are therefore MR conditional devices and a patient treated with the subject devices can be safely scanned observing the parameters previously established per reference devices. |
| Surface Treatment: | |
| Identical or similar chemical processes to predicate/reference devices to achieve roughness. | The surface treatments applied to subject devices are identical to these applied and previously evaluated for primary predicate devices. Although some reference devices have a surface with different nomenclature, the chemical processes applied to the surface are similar to create the roughness surface. Acqua implants are submitted to an additional step in order to increase their hydrophilicity, as already cleared for primary predicate device. |
| Sterilization Validation: | |
| Implants (Gamma Irradiation): Minimum Sterility Assurance Level (SAL) of 1 x 10^-6. | A minimum Sterility Assurance Level (SAL) of 1 x 10^-6 has been validated based on ISO 11137-1 and ISO 11137-2. |
| Abutments (Ethylene Oxide): Minimum Sterility Assurance Level (SAL) of 1 x 10^-6. | A minimum Sterility Assurance Level (SAL) of 1 x 10^-6 has been validated based on ISO 11135-1. |
| Endotoxin Testing (LAL): | |
| Meet pyrogen limit specifications (acceptance criteria for endotoxin levels). | Subject devices meet the acceptance criteria for endotoxin levels as well as the primary predicate and reference devices. |
| Shelf Life Validation: | |
| Packaging integrity maintained; established expiration date. | The shelf life for devices provided sterile is 5 years, except Helix Short Acqua Implants that have a shelf life of 4 years. The expiration date was determined considering product integrity and packaging tests. |
| Biocompatibility: | |
| No new issues compared to predicate/reference devices. | No new issues of biocompatibility are raised for the subject devices when compared to primary predicate and reference devices. Therefore, no additional biocompatibility testing was required. |
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state the numerical sample size for the test sets in the mechanical, MR compatibility, or other in-vitro studies. It refers to "tested subject devices" or "representative samples."
- Data Provenance: The studies described (mechanical testing, MR compatibility, surface treatment, sterilization, endotoxin, shelf life, biocompatibility) are all benchtop/laboratory studies (in vitro). There is no mention of human subject data, animal data, or clinical retrospective/prospective data for these specific performance criteria. The manufacturer is JJGC Indústria e Comércio de Materiais Dentários S.A. (dba Neodent) based in Curitiba, Paraná, Brazil.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
Not applicable. As these are benchtop/laboratory performance studies, "ground truth" is established by adherence to recognized standards (e.g., ISO 14801, ISO 11137, ISO 11135, ISO 10993) and scientific methods, not by expert consensus in the typical sense for clinical data.
4. Adjudication Method for the Test Set
Not applicable. Adjudication methods like 2+1 or 3+1 are typically used for interpreting clinical images or patient outcomes, not for objective benchtop performance data against engineering standards.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance
Not applicable. This device is an endosseous dental implant system, which is a physical medical device, not an AI-powered diagnostic or assistive tool. Therefore, MRMC studies involving human readers and AI assistance are not relevant.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
Not applicable, as this is a physical medical device, not an algorithm.
7. The Type of Ground Truth Used
The "ground truth" for the performance claims is based on:
- Adherence to International Standards: Performance measured against specifications outlined in standards like ISO 14801 (dynamic fatigue), ISO 11137 and ISO 11135 (sterilization), ISO 10993 series (biocompatibility).
- Benchtop Test Results: Direct measurements and observations from the mechanical, physical, chemical, and biological laboratory tests performed on the device.
- Comparison to Predicate Devices: Performance is frequently established by demonstrating equivalence or superiority to already legally marketed predicate devices, which serve as a benchmark.
8. The Sample Size for the Training Set
Not applicable. This is a physical medical device. The concepts of "training set" and "validation set" are primarily used for machine learning and AI, not for the regulatory approval of physical implants based on benchtop testing.
9. How the Ground Truth for the Training Set Was Established
Not applicable for the same reasons as point 8.
Ask a specific question about this device
Page 1 of 2