Search Filters

Search Results

Found 4 results

510(k) Data Aggregation

    K Number
    K241177

    Validate with FDA (Live)

    Date Cleared
    2024-05-28

    (29 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K180715

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The React™ 71 Catheter is indicated for the introduction of interventional devices into the peripheral and neuro vasculature.

    Device Description

    The React™ 71 Catheter is a single lumen, flexible, variable stiffness composite catheter with a nitinol structure that is jacketed with various durable polymer outer layers. A lubricious, polytetrafluoroethylene liner and Engage™ inner layer is used to create a structure that has both proximal stiffness and distal flexibility. The React™ 71 Catheter is also designed with an encapsulated radiopaque distal platinumiridium marker band which is used for visualization under fluoroscopy. The React™ 71 Catheter is introduced into the vasculature through the Split-Y Introducer Sheath. The proximal end of the React™ 71 Catheter is designed with a green thermoplastic elastomer strain relief and a clear hub. The distal end of the React™ 71 Catheter is coated with a Surmodics Serene™ Coating.

    AI/ML Overview

    The provided text describes a 510(k) submission for the React™ 71 Catheter, focusing on demonstrating substantial equivalence to a predicate device after a change in device length. The information primarily covers non-clinical bench testing.

    Here's an analysis of the provided information against your requested categories:

    1. A table of acceptance criteria and the reported device performance

    Acceptance CriteriaReported Device Performance
    The length of the subject device shall be compatible with minimum length 136 cm guide catheter.The subject device met the acceptance criteria.
    The subject device should be able to navigate to the distal M1 segment of the MCA over a microcatheter.The subject device met the acceptance criteria.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size: Not explicitly stated. The text mentions "non-clinical bench testing was conducted," but does not provide details on the number of devices tested for each criterion.
    • Data Provenance: Not specified. "Non-clinical bench testing" typically implies laboratory-based tests. The country of origin of the data is not mentioned. The data is prospective in the sense that new tests were conducted for this submission, but it's not clinical human data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This section is not applicable as the study described is non-clinical bench testing, not a study involving human or animal data requiring expert ground truth for interpretation.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This section is not applicable as the study described is non-clinical bench testing against predefined performance criteria, not a study involving subjective interpretations requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, an MRMC comparative effectiveness study was not done. This submission is for a percutaneous catheter, not an AI-powered diagnostic device.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    No, a standalone algorithm performance study was not done. This submission is for a physical medical device (catheter), not an algorithm or software.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the non-clinical bench tests, the "ground truth" is defined by the acceptance criteria established in the test methods, based on FDA Guidance 1757 and FDA Recognized Consensus Standards 5-125 & 5-129. These are objective engineering and performance standards, not clinical ground truths like pathology or expert consensus.

    8. The sample size for the training set

    This section is not applicable. This is a 510(k) submission for a physical medical device, not a machine learning or AI algorithm that requires a training set.

    9. How the ground truth for the training set was established

    This section is not applicable for the reasons stated in point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K210230

    Validate with FDA (Live)

    Device Name
    Phenom Catheters
    Date Cleared
    2021-02-25

    (28 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K180715

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Phenom™ Catheters are intended for the introduction of interventional devices or diagnostic agents into the neuro, peripheral, and coronary vasculatures.

    Device Description

    The Phenom™ Catheters are variable stiffness, single lumen catheters designed to access small, tortuous vasculature. They are available in a variety of lengths, stiffness and inner and outer diameters. The outer surface of the catheter is coated to enhance navigation in the vessel. The Phenom™ 17, 21, and 27 Catheters have a hydrophilic coating that spans the distal 100cm. The Phenom™ Plus Catheters have a hydrophilic coating that spans the distal 90cm. The catheter also incorporates a liner to facilitate movement of introduction devices passing through its lumen. The distal tip has radiopaque marker(s) to aid visualization and positioning under fluoroscopy. The Phenom™ Catheter is packaged with a shaping mandrel and may be accompanied with a split introducer sheath.

    AI/ML Overview

    The provided text describes the regulatory clearance for the "Phenom Catheters" and includes performance data from bench testing. However, it does not describe any study involving AI, human readers, or the establishment of ground truth for such a study. Therefore, I cannot fulfill your request for information related to AI or human-in-the-loop performance.

    The document focuses on demonstrating substantial equivalence to predicate devices through non-clinical bench testing. There is no mention of acceptance criteria or performance metrics in the context of an AI-based device, a multi-reader multi-case (MRMC) study, or standalone algorithm performance.

    Here's a breakdown of the information that is present in the document, framed against your request:

    1. A table of acceptance criteria and the reported device performance

    The document provides a table of "Tests" with "Test Method Summary" and "Results." The "Results" column implicitly indicates that the acceptance criteria were met for each test, as it states "The Phenom™ Catheters met the acceptance criteria for..." or "acceptable friction force," "compatible for use," etc.

    TestAcceptance Criteria (Implied by Result)Reported Device Performance
    Compatibility – Guide CatheterCompatible with guide catheters in a simulated tortuous path model.The Phenom™ Catheters are compatible for use with a guide catheter.
    Compatibility – Guide WireCompatible with guide wires in a simulated tortuous path model.The Phenom™ Catheters are compatible for use with a guide wire.
    Compatibility – RHVCompatible with a standard RHV.The Phenom™ Catheters are compatible with the standard used RHV.
    Coating Lubricity (Friction Force)Exhibit acceptable friction force under simulated clinical conditions.The Phenom™ Catheters exhibit acceptable friction force under simulated clinical conditions.
    Hub Functional & DimensionalMet the acceptance criteria for hub functional & dimensional requirements (per ISO 80369-7:2016 and ISO 80369-20: 2015).The Phenom™ Catheters met the acceptance criteria for hub functional & dimensional requirements.
    Torque StrengthExhibit acceptable torsional strength integrity.The Phenom™ Catheters exhibit acceptable torsional strength integrity.
    TensileMet the acceptance criteria for tensile strength (per ISO 10555-1:2013, Annex B).The Phenom™ Catheters met the acceptance criteria for tensile strength.
    Air AspirationMet the acceptance criteria for air aspiration (per ISO 10555-1:2013).The Phenom™ Catheters met the acceptance criteria for air aspiration.
    Liquid LeakMet the acceptance criteria for liquid leak (per ISO 10555-1:2013).The Phenom™ Catheters met the acceptance criteria for liquid leak.
    ParticulateMet the acceptance criteria for particulate (per USP <788>).The Phenom™ Catheters met the acceptance criteria for particulate.
    Design ValidationMet user needs and intended use(s) (per ANSI/AAMI HE75:2009/(R) 2018).The Phenom™ Catheters met the user needs and intended use(s) for which it was designed and tested.
    Biocompatibility Tests
    Cytotoxicity – MEM ElutionExhibited slight reactivity (Grade 1).Non-cytotoxic
    Sensitization – Guinea Pig MaximizationExhibited no dermal reactions (Grade 0).Non-sensitizer
    Irritation – Intracutaneous Study in RabbitsExhibited no dermal reactions (Score 0.0).Non-irritant
    Acute Systemic Toxicity – Systemic Toxicity Study in MiceExhibited no mortality or evidence of systemic toxicity.Non-toxic
    Material Mediated Pyrogenicity – USP Rabbit Pyrogen StudyExhibited no temperature rise above 0.5°C.Non-pyrogenic
    Hemocompatibility – Platelet and Leukocyte CountsPlatelet and leukocyte counts within average normalized and control values.Non-activator
    Hemocompatibility – Partial Thromboplastin TimeClotting time within control values.Non-activator
    Hemocompatibility – Hemolysis Direct Contact and Extract MethodsNo hemolysis and within control values.Non-hemolytic
    Hemocompatibility – Complement ActivationComplement activation within control values.Non-activator

    Regarding the other points of your request, the document does not provide this information because it's for a medical device (catheter) and not an AI-based system:

    • 2. Sample size used for the test set and the data provenance: Not applicable. The testing described is bench testing of physical devices, not a test set of data.
    • 3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for data interpretation is not relevant for this type of device submission.
    • 4. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable.
    • 5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is not an AI system.
    • 6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This device is not an algorithm.
    • 7. The type of ground truth used (expert consensus, pathology, outcomes data, etc): Not applicable. The "ground truth" for the catheter is its ability to perform according to engineering specifications and biocompatibility standards.
    • 8. The sample size for the training set: Not applicable. This is not an AI model requiring a training set.
    • 9. How the ground truth for the training set was established: Not applicable.

    In summary, the provided document details the regulatory clearance of a physical medical device (catheter) based on non-clinical performance and biocompatibility testing, not an AI software/device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K201690

    Validate with FDA (Live)

    Date Cleared
    2020-07-21

    (29 days)

    Product Code
    Regulation Number
    870.1330
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K180715

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Hydrophilic Guidewire is indicated for general intravascular use to aid in the selective placement of catheters in the peripheral, visceral and cerebral vasculature during diagnostic and/or therapeutic procedures.

    Device Description

    The Hydrophilic Guidewire is a stainless-steel guidewire with a radiopaque, platinum distal coil. The guidewire has a hydrophilic coating that spans the distal 170 cm. Included within the sterile pouch is a pin vise to assist in guidewire manipulation and an introducer needle to ease the introduction of the guidewire into the catheter hub and/or hemostasis valve.

    AI/ML Overview

    The provided text describes the 510(k) premarket notification for the Mirage™ and X-pedion™ Hydrophilic Guidewires, asserting their substantial equivalence to a predicate device. The information primarily focuses on bench testing rather than AI/ML algorithm performance. Therefore, many of the requested details, such as those related to AI model training, expert ground truth, and human reader studies, are not applicable or cannot be extracted from this document.

    Here's the relevant information based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    TestAcceptance Criteria (Implicit: "met the acceptance criteria")Reported Device Performance
    Performance
    Visual Inspection and Dimensional VerificationMet the acceptance criteria for visual inspection and dimensional verification.The Mirage™ and X-pedion™ Hydrophilic Guidewires met the acceptance criteria for visual inspection and dimensional verification.
    Friction ForceMet the acceptance criteria for friction force.The Mirage™ and X-pedion™ Hydrophilic Guidewires met the acceptance criteria for friction force.
    FlexingMet the acceptance criteria for flexing.The Mirage™ and X-pedion™ Hydrophilic Guidewires met the acceptance criteria for flexing.
    ParticulateMet the acceptance criteria for particulate.The Mirage™ and X-pedion™ Hydrophilic Guidewires met the acceptance criteria for particulate.
    Tensile StrengthMet the acceptance criteria for tensile strength.The Mirage™ and X-pedion™ Hydrophilic Guidewires met the acceptance criteria for tensile strength.
    Corrosion ResistanceMet the acceptance criteria for corrosion resistance.The Mirage™ and X-pedion™ Hydrophilic Guidewires met the acceptance criteria for corrosion resistance.

    2. Sample size used for the test set and the data provenance

    The document does not specify the exact sample sizes (number of guidewires) used for each bench test. The data provenance is from non-clinical bench testing conducted by Micro Therapeutics, Inc. d/b/a ev3 Neurovascular (the applicant).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This is a medical device clearance based on substantial equivalence demonstrated through non-clinical bench testing, not an AI/ML algorithm requiring expert-established ground truth. The "ground truth" for these tests would be the physical properties and performance measurements of the guidewires against engineering specifications.

    4. Adjudication method for the test set

    Not applicable, as this is bench testing of physical device properties, not a study involving human interpretation of data where adjudication would be necessary.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This document is for a physical medical guidewire, not an AI-powered diagnostic or assistive tool. No MRMC study was conducted or mentioned.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is for a physical medical guidewire.

    7. The type of ground truth used

    The "ground truth" for the bench tests would be the physical and mechanical specifications for guidewires, as referenced by standards such as ISO 11070 and USP <788>, as well as internal quality procedures (QP50324, TM0047). The tests evaluated observed performance against these predefined criteria.

    8. The sample size for the training set

    Not applicable. There is no training set for a physical medical device.

    9. How the ground truth for the training set was established

    Not applicable. There is no training set for a physical medical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K193548

    Validate with FDA (Live)

    Date Cleared
    2020-01-18

    (29 days)

    Product Code
    Regulation Number
    870.1330
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K180715, K093681

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Hydrophilic Guidewire is indicated for general intravascular use to aid in the selective placement of catheters in the peripheral, visceral and cerebral vasculature during diagnostic and/or therapeutic procedures.

    Device Description

    The Hydrophilic Guidewire is a stainless-steel guidewire with a radiopaque, platinum distal coil. The guidewire has a hydrophilic coating that spans the distal 170 cm. Included within the sterile pouch is a pin vise to assist in guidewire manipulation and an introducer needle to ease the introduction of the guidewire into the catheter hub and/or hemostasis valve.

    AI/ML Overview

    The provided text describes the 510(k) premarket notification for the Mirage™ Hydrophilic Guidewire and X-pedion™ Hydrophilic Guidewire. This document primarily focuses on demonstrating substantial equivalence to a predicate device through non-clinical (bench and biocompatibility) testing, rather than a study on an AI/ML-based medical device.

    Therefore, the information required to answer most of your questions about acceptance criteria for an AI/ML device, sample sizes for AI test sets, expert ground truthing, MRMC studies, standalone performance, training sets, and ground truth establishment for AI/ML models is not present in this document.

    However, I can extract information related to the acceptance criteria and performance data for the physical medical device as described:


    Acceptance Criteria and Device Performance (Based on Non-Clinical Bench Testing):

    The document details numerous non-clinical bench tests performed to demonstrate the safety and performance of the guidewires. The results consistently state that "The Mirage™ and X-pedion™ Hydrophilic Guidewires met the acceptance criteria for [Test Name]."

    Table of Acceptance Criteria and Reported Device Performance (Non-Clinical):

    Test CategorySpecific Test / ParameterAcceptance Criteria (Implied by positive outcome)Reported Device Performance
    Biocompatibility
    Chemical Characterization (Extractables/Leachables)No risk to the patient from Chemicals of Potential Concern (COPC).The extractables/leachables found are acceptable.
    CytotoxicityNo evidence of cell lysis or toxicity; grade < 2 (Mild).Considered non-cytotoxic.
    SensitizationNo delayed dermal contact sensitization.Does not elicit a sensitization response.
    IrritationDifference between test and control mean score = 0.0.Considered a non-irritant.
    Acute Systemic ToxicityNo mortality or systemic toxicity from extracts.Do not indicate signs of toxicity.
    Hemocompatibility (Hemolytic Index)Hemolytic index for direct contact < threshold (e.g., 0.5% / 0.6%).Hemolytic index was 0.5% (direct contact) and 0.6% (extract).
    Hemocompatibility (Complement System)Not considered to be potential activators of the complement system.Control and test articles are not considered to be potential activators.
    Hemocompatibility (Thrombogenic Potential)Acceptable thrombogenic potential.Demonstrates moderate thrombus formation (less than control article).
    PyrogenicityNo temperature rise of 0.5°C or more above baseline.No animal showed a temperature rise of 0.5°C or more.
    Microbial
    Ethylene Oxide (EO) ResidualMet criteria per ISO 10993-7.Met the acceptance criteria for ethylene oxide residual.
    Ethylene Chlorohydrin (ECH) ResidualMet criteria per ISO 10993-7.Met the acceptance criteria for ethylene chlorohydrin residual.
    Bioburden RecoveryMet criteria per ISO 11737-1.Met the acceptance criteria for bioburden recovery.
    BioburdenMet criteria per ISO 11737-1.Met the acceptance criteria for bioburden.
    Bacterial EndotoxinMet criteria per ANSI/AAMI ST72, USP <161>, and USP <85>.Met the acceptance criteria for bacterial endotoxin.
    Packaging
    Terminally Sterilized Medical DevicesMet criteria per ISO 11607.Met the acceptance criteria for packaging terminally sterilized medical devices.
    Performance (Bench)
    Device Compatibility/Distal AccessCompatible with ancillary devices, navigates tortuous vessels.Met the acceptance criteria for device compatibility/distal access.
    Distal FlexibilityNavigates through tortuous vessels.Met the acceptance criteria for distal flexibility.
    Visual FractureMet criteria per ISO 10555-1 and ISO 11070.Met the acceptance criteria for visual fracture.
    RadiopacityClearly visible during use.Met the acceptance criteria for radiopacity.
    Tip BucklingWithstands forces typical of clinical use.Met the acceptance criteria for tip buckling.
    Tip RetentionSatisfactory tip retention.Met the acceptance criteria for tip retention.
    Tip ShapeabilitySatisfactory tip shapeability.Met the acceptance criteria for tip shapeability.
    Torque ResponseDistal tip responds to proximal manipulations.Met the acceptance criteria for torque response.
    Turns to FailureWithstands torsional forces typical of clinical use.Met the acceptance criteria for turns to failure.
    Visual Inspection and Dimensional VerificationMeets specified dimensions.Met the acceptance criteria for visual inspection and dimensional verification.
    ParticulateMeets criteria per USP <788> for particulate counts and sizes.Met the acceptance criteria for particulate counts and sizes.
    Friction ForceSatisfactory friction force.Met the acceptance criteria for friction force.
    FlexingMet criteria per ISO 11070.Met the acceptance criteria for flexing.
    Corrosion ResistanceMet criteria per ISO 11070.Met the acceptance criteria for corrosion resistance.
    Tensile StrengthMet criteria per ISO 11070.Met the acceptance criteria for tensile strength.

    Since this is a 510(k) submission for a non-AI/ML medical device, the following information is not applicable or not provided in the document:

    1. Sample sizes used for the test set and the data provenance: Not an AI device, so no "test set" in the AI sense. Bench tests follow specific standards, implying sample sizes per those standards, but not explicitly stated here.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable; ground truth for a physical device is established via standardized physical and chemical tests, not expert human interpretation of data for AI.
    3. Adjudication method for the test set: Not applicable for non-AI bench testing.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done: Not applicable.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
    6. The type of ground truth used: For a physical guidewire, "ground truth" is typically defined by adherence to engineering specifications, material properties, and performance under simulated clinical conditions, verified through empirical testing following established ISO and USP standards.
    7. The sample size for the training set: Not applicable; this is not an AI/ML device.
    8. How the ground truth for the training set was established: Not applicable; this is not an AI/ML device.

    In summary, this document is a regulatory submission for a physical medical device demonstrating substantial equivalence, not an AI/ML device requiring clinical validation for diagnostic or prognostic purposes.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1