    Intended Use

    The Mpact 3D Metal implants are designed for cementless use in total hip arthroplasty in primary or revision surgery.

    The patient should be skeletally mature.

    The patient's condition should be due to one or more of:

    • Severely painful and/or disabled joint: as a result of osteoarthritis, post-traumatic arthritis, rheumatoid arthritis or psoriatic arthritis, congenital hip dysplasia, ankylosing spondylitis.
    • Avascular necrosis of the femoral head.
    • Acute traumatic fracture of the femoral head or neck.
    • Failure of previous hip surgery: joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement where sufficient bone stock is present.
    Device Description

    The Mpact 3D Metal Implants Extension - DMLS Technology are a line extension to the Mpact 3D Metal Acetabular Shells and 3D Metal Augments (K171966, K202568) and to the Mpact Acetabular Systems Shells (K103721, K122641, K132879 and K230011) and are designed to be used with the Medacta Total Hip Prosthesis System. Specifically, the devices subject of this submission are:

    • Acetabular shell size Ø44 Two-holes;
    • Acetabular shell size Ø44 Multi-holes thin;
    • Acetabular shells sizes from Ø42T to Ø58T of both Two-holes and Multi-holes Thin designs, allowing to be coupled with an increase size of liner.

    The subject implants are intended to be used during Total Hip Arthroplasty to replace the acetabulum and they are provided individually packed, sterile and single-use. Similarly to the predicate devices, the subject acetabular shells are manufactured using a Direct Metal Laser Sintering (DMLS) process with titanium alloy powder according to ASTM F2924-14.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K220405

    Validate with FDA (Live)

    Device Name
    Amis K Long
    Date Cleared
    2022-04-12

    (57 days)

    Product Code
    Regulation Number
    888.3353
    Age Range
    All
    Predicate For
    N/A
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The hip prosthesis AMIS-K Long is designed for cemented use in total or partial hip arthroplasty in primary or revision surgery.

    Total hip replacement is indicated in the following cases:
    · Severely painful and/or disabled joint as a result of arthritis, rheumatoid polyarthritis or congenital hip dysplasia

    • Avascular necrosis of the femoral head
    • · Acute traumatic fracture of the femoral head or neck
      · Failure of previous hip surgery: joint reconstruction, arthrodesis, partial hip arthroplasty, hip resurfacing replacement, or total hip arthroplasty.

    Partial hip arthroplasty is indicated in the following cases:
    · Acute traumatic fracture of the femoral head or neck

    • · Non-union of femoral neck fracture
    • · Avascular necrosis of the femoral head.
    • · Primary pathology involving the femoral head but with a non-deformed acetabulum.
    Device Description

    The AMIS-K Long cemented stem is designed for cemented use in total or partial hip arthroplasty in primary or revision surgery.

    The AMIS-K Long is a straight, double-tapered cemented stem whose primary stability is ensured by bone cement.

    The AMIS-K Long implants in this submission are comprised of the following size: from #2 to #5 with 2 different stem body size for #2, #3 and #4, 1 different stem body size for #5. Stem length: 200 - 250 - 300 mm (for each size).

    The AMIS-K Long implants are part of the Medacta Total Hip Prosthesis system.

    The Medacta Total Hip Prosthesis system consists of femoral stemoral heads, and acetabular components.

    The AMIS-K Long are cemented stems manufactured from high nitrogen stainless steel (ISO 5832-9) with a mirror polished surface.

    The acetabular components consist of metal cups and liners made of ultra-high molecular weight polyethylene (UHMWPE), or Highcross highly crosslinked ultra-high polyethylene (HXUHMWE). Acetabular components include the Mpact DM (K143453), VersafitCup (K083116 and K092265), VersafitCup CC Trio (K103352, K120531 and K122911), Mpact (K103721 and K132879), Mpact 3D Metal (K171966), and Medacta Bipolar Head (K091967).

    The AMIS-K Long implants can be combined with the CoCr Ball Heads (K072857 and K080885). Endo Head (K111145), or MectaCer BIOLOX® Forte (K073337) or MectaCer BIOLOX® Delta Femoral Heads (K112115).

    AI/ML Overview

    The provided text is a 510(k) summary for a medical device (Amis K Long hip prosthesis), which primarily focuses on demonstrating substantial equivalence to a predicate device rather than presenting a detailed study proving the device meets specific acceptance criteria in the context of AI/ML or diagnostic performance.

    Therefore, the document does not contain the information requested regarding acceptance criteria and a study proving a device meets these criteria in the context of AI/ML performance. Specifically:

    • No acceptance criteria for AI/ML performance are listed. The acceptance criteria mentioned refer to non-clinical performance tests for the physical hip prosthesis (e.g., pull-off strength, fatigue testing).
    • No study proving AI/ML device performance is described. The document explicitly states "No clinical studies were conducted." and describes non-clinical mechanical tests.
    • The requested elements (sample size, data provenance, expert qualifications, etc.) are not applicable in this context as the document is about a physical implant, not an AI/ML diagnostic or assistive device.

    The study described in the document primarily focuses on non-clinical mechanical tests to validate the physical properties of the hip prosthesis, such as pull-off strength, range of motion, and fatigue. These tests adhere to established international standards (ASTM, EN ISO, ISO).

    In summary, the provided text does not contain the information necessary to answer the questions about acceptance criteria and a study proving device performance as they relate to AI/ML or diagnostic accuracy.

    Ask a Question

    Ask a specific question about this device

    K Number
    K201673

    Validate with FDA (Live)

    Date Cleared
    2020-09-18

    (91 days)

    Product Code
    Regulation Number
    888.3353
    Age Range
    All
    Predicate For
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The hip prosthesis SMS is designed for cementless use in total or partial hip arthroplasty, for primary or revision surgery. Hip replacement is indicated in the following cases:

    · Severely painful and/or disabled joint as a result of arthritis, rheumatoid polyarthritis or congenital hip dysplasia

    · Avascular necrosis of the femoral head

    · Acute traumatic fracture of the femoral head or neck

    · Failure of previous hip surgery: joint reconstruction, internal fixation, arthrodesis, partial hip arthroplasty, hip resurfacing replacement or total hip arthroplasty.

    Device Description

    The SMS femoral stem is a cementless bone preserving short stem designed for proximal fixation in total or partial hip arthroplasty for primary or revision surgery.

    The SMS implants are comprised of the following products:

    • SMS Cementless Solid Standard Stem (available in 11 sizes from size 3 to 13); and ●
    • . SMS Cementless Solid Lateralized Stem (available in 11 sizes - from size 3 to 13).

    Both are available on the US market via the clearance - K181693.

    Concerning the new sizes of the solid version: 1V2, 2V2, size 14, and size 15; both STD and LAT versions have been introduced as a product range extension to the stems currently marketed as the SMS product line (K181693).

    The SMS implants are line extensions to Medacta's Total Hip Prosthesis - AMIStem-H, Quadra-S and Quadra-H Femoral Stems (K093944), AMIStem and Quadra - Line Extension (K121011), AMIStem-P, AMIStem-P Collared and AMIStem-H Proximal Coating Femoral Stems (K173794), Quadra-H and Quadra-R Femoral Stems (K082792), AMIStem-H Proximal Coating (K161635), MiniMAX (K170845), and SMS (K181693).

    The SMS implants are part of the Medacta Total Hip Prosthesis system. The Medacta Total Hip Prosthesis system consists of femoral stems, modular femoral heads and acetabular components. The acetabular components consist of metal cups and liners made of ultra-high molecular weight polyethylene (UHMWPE) or Highcross highly crosslinked ultra-high molecular weight polyethylene (HXUHMWPE). Acetabular components include the Mpact DM (K143453), VersafitCup (K083116 and K092265), VersafitCup CC Trio (K103352), Mpact (K103721 and K132879), Mpact 3D Metal (K171966) and Medacta Bipolar Head (K091967).

    The SMS stems can be combined with the CoCr ball heads (K072857, K080885 and K103721), Endo Head (K111145) or with the MectaCer BIOLOX® Forte (K073337), MectaCer BIOLOX® Delta Femoral Heads (K112115) or MectaCer BIOLOX® Option Heads (K131518).

    AI/ML Overview

    This is a 510(k) premarket notification for the "SMS Cementless Stem" (K201673), a hip prosthesis. The document details the device, its intended use, and a comparison to predicate devices, along with performance data.

    Here's an analysis of the provided text in relation to your request:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document outlines various non-clinical performance tests conducted for the SMS Cementless Stem. However, it does not provide specific acceptance criteria values or the quantitative results of these tests in a table format. It only states that the testing was conducted "according to written protocols with acceptance criteria that were based on standards."

    Here's what can be inferred for a table, but actual criteria values and results are missing in the document:

    Test (Standard)Acceptance Criteria (Not provided explicitly, generally "meet standard requirements")Reported Device Performance (Not provided explicitly, implied "met criteria")
    Range of Motion (EN ISO 21535:2009)Based on standard EN ISO 21535:2009Conducted
    Fatigue Testing - Stemmed Femoral Components (ISO 7206-4:2010 + AME 1:2016)Based on standard ISO 7206-4Conducted
    Fatigue Testing - Head and Neck Region of Stemmed Femoral Components (ISO 7206-6:2013)Based on standard ISO 7206-6Conducted
    Pull-off Force Testing - Taper Connections (ASTM F2009-00 (Reapproved 2011))Based on standard ASTM F2009-00Conducted
    Coating Tests - Hydroxyapatite (ISO 13779-1:2008)Based on standard ISO 13779-1Conducted
    Coating Tests - Tension Testing of Calcium Phosphate and Metal Coatings (ASTM F1147-99)Based on standard ASTM F1147-99Conducted
    Pyrogenicity - Bacterial Endotoxin Test (European Pharmacopoeia §2.6.14 / USP chapter <85>)Based on European Pharmacopoeia §2.6.14 / USP chapter <85>Conducted
    Pyrogenicity - Pyrogen Test (USP chapter <151>)Based on USP chapter <151>Conducted

    2. Sample Size Used for the Test Set and Data Provenance

    The document explicitly states: "No clinical studies were conducted." Therefore, there is no test set in the context of human clinical data or patient outcomes. The "performance data" refers to non-clinical mechanical and material tests. The sample sizes for these mechanical tests are not provided in this summary. Data provenance is not applicable here as no human data was used.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    Not applicable as no clinical studies were conducted and no ground truth in the clinical sense was established. The "ground truth" for the non-clinical tests would be the specifications and requirements of the referenced ISO and ASTM standards.

    4. Adjudication Method for the Test Set

    Not applicable as no clinical studies were conducted which would require adjudication of expert opinions or outcomes.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and its effect size

    No MRMC study was mentioned or performed, as "No clinical studies were conducted." This device is a mechanical implant, not an AI or diagnostic tool that would typically undergo MRMC studies.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a hip implant, not an algorithm or AI system.

    7. The Type of Ground Truth Used

    For the non-clinical performance and coating tests, the "ground truth" implicitly refers to the
    requirements and specifications outlined in the referenced international standards (ISO and ASTM) and the European Pharmacopoeia/USP chapters. The device is expected to meet these pre-defined engineering and material standards. This is a form of engineering/material specification compliance as ground truth.

    8. The Sample Size for the Training Set

    Not applicable. This device is a mechanical implant, not an AI or machine learning model that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this type of medical device's approval.

    Ask a Question

    Ask a specific question about this device

    K Number
    K181693

    Validate with FDA (Live)

    Device Name
    SMS Femoral Stem
    Date Cleared
    2018-11-28

    (154 days)

    Product Code
    Regulation Number
    888.3353
    Age Range
    All
    Predicate For
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The hip prosthesis SMS is designed for cementless use in total or partial hip arthroplasty, for primary or revision surgery. Hip replacement is indicated in the following cases:

    · Severely painful and/or disabled joint as a result of arthritis, rheumatoid polyarthritis or congenital hip dysplasia

    · Avascular necrosis of the femoral head

    · Acute traumatic fracture of the femoral head or neck

    · Failure of previous hip surgery: joint reconstruction, internal fixation, arthroplasty, hip resurfacing replacement or total hip arthroplasty.

    Device Description

    The SMS femoral stem is a cementless bone preserving short stem designed for proximal fixation in total or partial hip arthroplasty for primary or revision surgery. The SMS implants subject of this submission are comprised of the following products:

    • SMS Cementless Solid Standard Stem (available in 11 sizes); and ●
    • SMS Cementless Solid Lateralized Stem (available in 11 sizes).

    The SMS implants are line extensions to Medacta's Total Hip Prosthesis - AMIStem-H, Quadra-S and Quadra-H Femoral Stems (K093944), AMIStem and Quadra - Line Extension (K121011), AMIStem-P, AMIStem-P Collared and AMIStem-H Proximal Coating Femoral Stems (K173794), Quadra-H and Quadra-R Femoral Stems (K082792), AMIStem-H Proximal Coating (K161635), and MiniMAX (K170845).

    The SMS implants are part of the Medacta Total Hip Prosthesis system. The Medacta Total Hip Prosthesis system consists of femoral stems, modular femoral heads and acetabular components. The acetabular components consist of metal cups and liners made of ultra-high molecular weight polyethylene (UHMWPE) or Highcross highly crosslinked ultra-high molecular weight polvethylene (HXUHMWPE).

    Acetabular components include the Mpact DM (K143453), VersafitCup (K083116 and K092265), VersafitCup CC Trio (K103352), Mpact (K103721 and K132879), Mpact 3D Metal (K171966) and Medacta Bipolar Head (K091967).

    The SMS stems can be combined with the CoCr ball heads (K072857, K080885 and K103721), Endo Head (K111145) or with the MectaCer BIOLOX® Forte (K073337), MectaCer BIOLOX® Delta Femoral Heads (K112115) or MectaCer BIOLOX® Option Heads (K131518).

    AI/ML Overview

    This document describes the premarket notification (510(k)) for the SMS Femoral Stem, a cementless bone-preserving short stem designed for proximal fixation in total or partial hip arthroplasty.

    Here's an analysis of the provided text in relation to acceptance criteria and supporting studies, though it's important to note that this is a medical device submission, not an AI/ML device submission, so the questions regarding human readers, training sets, etc., are not directly applicable here. The document focuses on mechanical performance and biocompatibility for a physical implant.

    1. A table of acceptance criteria and the reported device performance

    The document lists performance tests and the standards they adhere to, implying that meeting these standards constitutes the acceptance criteria. However, it does not explicitly state numerical acceptance criteria and does not present specific performance data values in a table format. It merely states that "Testing was conducted according to written protocols with acceptance criteria that were based on standards." and that the studies "demonstrated substantial equivalence."

    Here's a generalized representation based on the provided information:

    Acceptance Criterion (Implicit)Reported Device Performance
    Conformance to EN ISO 21535:2009 (Range of Motion)Testing conducted, supporting substantial equivalence.
    Conformance to ISO 7206-4:2010 (Fatigue Testing - Stem)Testing conducted, supporting substantial equivalence.
    Conformance to ISO 7206-6:2013 (Fatigue Testing - Head/Neck)Testing conducted, supporting substantial equivalence.
    Conformance to ISO 7206-10:2003 (Static Fatigue - Modular Heads)Testing conducted, supporting substantial equivalence.
    Conformance to ASTM F2009-00 (Axial Disassembly Force)Testing conducted, supporting substantial equivalence.
    Conformance to ISO 13779-1:2008 (Coating - Hydroxyapatite)Testing conducted, supporting substantial equivalence.
    Conformance to ASTM F1147-99 (Tension Testing - Coatings)Testing conducted, supporting substantial equivalence.
    Conformance to European Pharmacopoeia §2.6.14/USP <85> (BET)Testing conducted (LAL test), supporting substantial equivalence.
    Conformance to USP <151> (Pyrogen Test)Testing conducted, supporting substantial equivalence.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document lists various non-clinical mechanical and coating tests, which typically involve physical samples of the device components. It does not specify the sample sizes (number of stems tested for fatigue, etc.) for each test. The provenance of the data is implicit: it's generated from laboratory testing of the device components. There is no clinical data or patient-derived data mentioned.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This question is not applicable to this type of medical device submission. Ground truth and expert adjudication are relevant for AI/ML diagnostic devices, not for physical implants that undergo mechanical and material testing. The "ground truth" for these tests are the established scientific principles and measurement techniques outlined in the specified ISO and ASTM standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This question is not applicable for the same reasons as #3. Mechanical testing results are typically objectively measured and evaluated against a standard, not subject to expert adjudication in the same way as diagnostic interpretations.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This question is not applicable as this is not an AI/ML device. No MRMC study was conducted.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This question is not applicable as this is not an AI/ML device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the non-clinical tests, the "ground truth" is defined by the specifications and acceptable ranges outlined in the referenced international (ISO) and American (ASTM) standards. For instance, a fatigue test determines if the device can withstand a certain number of cycles at a specific load without failure, as per the standard's requirements. Pyrogenicity is determined by established biological testing methods (LAL test, USP pyrogen test).

    8. The sample size for the training set

    This question is not applicable as this is not an AI/ML device. There is no concept of a "training set" for a physical implant undergoing mechanical and material characterization.

    9. How the ground truth for the training set was established

    This question is not applicable as this is not an AI/ML device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K181254

    Validate with FDA (Live)

    Device Name
    Quadra P
    Date Cleared
    2018-10-22

    (164 days)

    Product Code
    Regulation Number
    888.3353
    Age Range
    All
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The hip prosthesis QUADRA-P is designed for cementless use in total or partial hip arthroplasty; for primary or revision surgery. Hip replacement is indicated in the following cases:

    • Severely painful and/or disabled joint as a result of arthritis, rheumatoid polyarthritis or congenital hip dysplasia

    · Avascular necrosis of the femoral head

    · Acute traumatic fracture of the femoral head or neck

    · Failure of previous hip surgery: joint reconstruction, internal fixation, arthroplasty, hip resurfacing replacement or total hip arthroplasty.

    Device Description

    The Quadra-P implants are line extensions to Medacta's Total Hip Prosthesis – AMIStem-H, Quadra-S, and Quadra-H Femoral Stems (K093944); AMIStem and Quadra - Line Extension (K121011); AMIStem-P, AMIStem-P Collared, and AMIStem-H Proximal Coating Femoral Stems (K173794); Quadra-H and Quadra-R Femoral Stems (K082792); and AMIStem-H Proximal Coating (K161635).

    The Quadra-P implants subject of this submission are comprised of the following products:

    • . Quadra-P STD. Stem sizes 0 - 10;
    • Quadra-P LAT, Stem sizes 0 - 10.
      The Quadra-P implants are part of the Medacta Total Hip Prosthesis system. The Medacta Total Hip Prosthesis system consists of femoral stems, modular femoral heads, and acetabular components. The acetabular components consist of metal cups and liners made of ultra-high molecular weight polyethylene (UHMWPE) or Highcross highly crosslinked ultra-high molecular weight polyethylene (HXUHMWPE). Acetabular components include the Mpact DM (K143453), VersafitCup (K083116 and K092265), VersafitCup CC Trio (K103352), Mpact (K103721 and K132879), Mpact 3D Metal (K171966), and Medacta Bipolar Head (K091967).

    The Quadra-P implants can be combined with the CoCr Ball Heads (K072857 and K080885), Endo Head (K111145), MectaCer Biolox Option Heads (K131518), or with the MectaCer BIOLOX® Forte (K073337) or MectaCer BIOLOX® Delta Femoral Heads (K112115).

    AI/ML Overview

    This document is a 510(k) premarket notification for the "Quadra-P" hip prosthesis, specifically describing the physical and mechanical testing performed to demonstrate substantial equivalence to previously cleared predicate devices. It does not contain information about the performance or acceptance criteria of an AI/Software device, nor does it describe a study involving humans or the establishment of ground truth for such a device.

    Therefore, I cannot extract the information required to answer your questions about acceptance criteria and studies proving the device meets those criteria, as the provided document pertains to a physical medical device (hip prosthesis) and its mechanical engineering performance tests, not an AI or software-based medical device.

    The document discusses:

    • Device Name: Quadra-P (hip prosthesis)
    • Regulation Number: 21 CFR 888.3353 (Hip Joint Metal/Ceramic/Polymer Semi-Constrained Cemented Or Nonporous Uncemented Prosthesis)
    • Performance Data (Non-Clinical Studies):
      • Range of Motion (ROM) testing (EN ISO 21535:2009)
      • Fatigue testing (ISO 7206-4 and ISO 7206-6)
      • Pull off force testing (ASTM F2009-00)
      • Pyrogenicity testing (European Pharmacopoeia §2.6.14/USP <85> and USP <151>)
    • Clinical Studies: "No clinical studies were conducted."

    Since the request is specifically about a device proving it meets acceptance criteria through a study involving a test set, expert ground truth, adjudication methods, and potentially human readers (which implies an AI/Software device or diagnostic device), the provided text is entirely irrelevant to those questions.