Search Filters

Search Results

Found 4 results

510(k) Data Aggregation

    K Number
    K233499

    Validate with FDA (Live)

    Date Cleared
    2024-08-16

    (290 days)

    Product Code
    Regulation Number
    870.5100
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K162209, K192344, K163114, K173680

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Sapphire NC ULTRA Coronary Dilatation Catheter is indicated for:

    · Balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis in □ patients evidencing coronary ischemia for the purpose of improving myocardial perfusion.

    · Balloon dilatation of a coronary artery occlusion for the treatment of acute myocardial □ infarction.

    · In-stent restenosis.

    · Post-delivery expansion of balloon expandable coronary stents.

    Device Description

    The Sapphire NC ULTRA Coronary Dilatation Catheter is designed to allow easy exchange of the catheter using a standard length 0.014 inch guidewire. Balloon diameters range from 1.75mm to 5.0mm. The balloon material is made of a minimally compliant material, 1.75mm to 4.0mm balloons have a rated burst pressure of 20 atmospheres, and 4.5mm to 5.0mm balloons have a rated burst pressure of 18 atmospheres. The minimally compliant balloon material allows high pressure dilatation while maintaining precise control of the balloon diameter and length. The proximal shaft of the catheter is composed of a female luer connector connected to a PTFE coated stainless steel tube. The proximal shaft allows superior proximal pushability and trackability with a smooth transition to a distal shaft. Two radiopaque platinum/iridium marker bands are located within the balloon segment. The guidewire enters the catheter tip and advances coaxially out the distal Rx port, thereby allowing both coaxial guidance and rapid exchange of catheter with a single standard-length guidewire. Two marked sections of 5mm length each located on the proximal shaft indicate catheter position relative to the tip of either a brachial or femoral guiding catheter.

    AI/ML Overview

    The provided text is a 510(k) summary for the Sapphire NC ULTRA Coronary Dilatation Catheter. It outlines various performance tests conducted to establish substantial equivalence to a predicate device. However, it does not contain the specific level of detail required to fully answer all aspects of your request, particularly regarding clinical study design, expert qualifications, or detailed performance metrics against acceptance criteria.

    Here's an attempt to answer your questions based on the available information:

    1. A table of acceptance criteria and the reported device performance

    The document states: "The test results met all acceptance criteria, which are the same or similar to the predicate device and ensure that the Sapphire NC ULTRA Coronary Dilatation Catheter design and construction are suitable for their intended use."

    While it lists the types of tests performed, it does not provide a table with specific acceptance criteria or quantitative performance results for each test. For example, it lists "Balloon Rated Burst Pressure (in-stent)" as a test, but doesn't state what the accepted pressure was or what the device achieved.

    Here's a generalized table based on the types of tests mentioned, but without specific numerical criteria or performance data:

    Acceptance Criteria CategoryReported Device Performance
    In vitro Performance:
    Visual InspectionMet acceptance criteria
    Particulate EvaluationMet acceptance criteria
    Dimensional VerificationMet acceptance criteria
    Balloon ComplianceMet acceptance criteria
    Hub Leakage TestMet acceptance criteria
    Balloon Prep, Deployment, RetractionMet acceptance criteria
    Balloon Inflation/Deflation TimeMet acceptance criteria
    Coating IntegrityMet acceptance criteria
    Balloon Fatigue (in-stent)Met acceptance criteria
    Balloon FatigueMet acceptance criteria
    Balloon Rated Burst Pressure (in-stent)Met acceptance criteria
    Balloon Rated Burst PressureMet acceptance criteria
    Shaft BurstMet acceptance criteria
    Catheter Bond StrengthMet acceptance criteria
    Tip Pull StrengthMet acceptance criteria
    Flexibility and KinkMet acceptance criteria
    Torque StrengthMet acceptance criteria
    Marker Band RadiopacityMet acceptance criteria
    Biocompatibility:
    CytotoxicityMet acceptance criteria
    SensitizationMet acceptance criteria
    Intracutaneous ReactivityMet acceptance criteria
    Acute Systemic ToxicityMet acceptance criteria
    PyrogenicityMet acceptance criteria
    HemolysisMet acceptance criteria
    Partial Thromboplastin TimeMet acceptance criteria
    Platelet and Leukocyte CountsMet acceptance criteria
    Complement ActivationMet acceptance criteria
    Toxicological Risk Assessment (TRA) of Extractable ChemicalsMet acceptance criteria
    Other:
    Packaging and Sterilization ValidationMet acceptance criteria
    Shelf LifeMet acceptance criteria

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document describes in vitro performance tests and biocompatibility tests. This means the "test set" refers to material samples (e.g., catheter units, material extracts) used for these laboratory tests, not a clinical patient dataset.

    • Sample Size: The document does not specify the sample size for the in vitro or biocompatibility tests.
    • Data Provenance: Not applicable in the context of in vitro and biocompatibility testing. These are laboratory tests conducted on device components or finished products.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This question is geared towards clinical studies involving human interpretation or pathology. Since the provided text only details in vitro and biocompatibility testing, there is no information about experts or their qualifications for establishing ground truth as would be required in a clinical setting. These tests typically follow standardized protocols and are evaluated by lab personnel, not medical experts establishing clinical ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Again, this question is relevant for clinical studies with human assessors. For the in vitro and biocompatibility testing described, adjudication methods like N+1 are not applicable. The results are typically quantitative measurements or observations against predefined pass/fail criteria according to established standards (e.g., ISO, FDA guidance).

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    The document does not mention a multi-reader multi-case (MRMC) comparative effectiveness study. The device is an angioplasty catheter, not an AI software or a device that assists human readers in interpreting images. Therefore, this question is not applicable to the information provided.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This question is also not applicable. The Sapphire NC ULTRA Coronary Dilatation Catheter is a physical medical device, not an algorithm or software requiring standalone performance testing in that context.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the in vitro performance tests, the "ground truth" is established by adherence to predefined engineering specifications, material properties, and performance benchmarks derived from industry standards (e.g., those in the "Class II Special Controls Guidance Document for Certain Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheters").

    For biocompatibility testing, the "ground truth" is established by compliance with international standards like ISO 10993-1, which define acceptable biological responses and safety profiles for medical device materials.

    8. The sample size for the training set

    This question is applicable to machine learning or AI models. Since the device is a physical medical catheter and the testing described is primarily in vitro and biocompatibility, there is no concept of a "training set" in this context.

    9. How the ground truth for the training set was established

    As there is no training set for a physical medical device, this question is not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K233505

    Validate with FDA (Live)

    Date Cleared
    2024-08-15

    (289 days)

    Product Code
    Regulation Number
    870.5100
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K163114, K211807, K162209, K192344

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Sapphire ULTRA Coronary Dilatation Catheter (1.0-1.25mm configurations) is indicated for: balloon pre-dilatation of a stenotic portion of a coronary artery or bypass graft stenosis (≥70% stenosis) for the purpose of improving myocardial perfusion.

    The Sapphire ULTRA Coronary Dilatation Catheter (1.5-4.0mm configurations) is indicated for: balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis in . patients evidencing coronary ischemia for the purpose of improving myocardial perfusion. balloon dilatation of a coronary artery occlusion for the treatment of acute myocardial infarction.

    Device Description

    The Sapphire ULTRA Coronary Dilatation Catheter is designed to allow easy exchange of the catheter using a standard length 0.014 inch guidewire. Balloon diameters range from 1.0mm to 4.0mm. The balloon material is made of the semi-compliant material with a rated burst pressure of 14 atmospheres. The proximal shaft of the catheter is composed of a female luer connector connected to a PTFE coated stainless steel tube. The proximal shaft allows superior proximal pushability with a smooth transition to a distal shaft composed of an outer tube and an inner tube with a balloon laser welded to both tubes at the distal tip. Two radiopaque platinum marker bands are located within the balloon segment (there is only one centrally located marker band for Ø1.0-1.5mm configurations). The guidewire enters the catheter tip and advances coaxially out the distal Rx port, thereby allowing both coaxial guidance and rapid exchange of catheter with a single standard length guidewire. Two marked sections, 5mm length each, located on the proximal shaft indicate catheter position relative to the tip of either a brachial or femoral guiding catheter.

    AI/ML Overview

    The provided text is a 510(k) summary for the Sapphire ULTRA Coronary Dilatation Catheter. It outlines the device's description, indications for use, technological characteristics, and performance tests conducted to demonstrate substantial equivalence to predicate devices.

    However, the provided document DOES NOT contain information regarding:

    • Acceptance criteria for an AI/ML device. This document describes a medical device (a catheter), not an AI/ML diagnostic or prognostic tool. Therefore, there are no AI/ML specific acceptance criteria, study designs (like MRMC), or details about ground truth establishment as would be typically required for such devices.
    • Study that proves the device meets AI/ML acceptance criteria. The performance tests listed are for the physical properties and biocompatibility of a medical catheter, not for the performance of an AI/ML algorithm.

    Therefore, I cannot provide the requested information regarding AI/ML acceptance criteria and proof of performance. The document focuses on the physical and biological safety and efficacy of a coronary dilatation catheter, which is a hardware medical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K182360

    Validate with FDA (Live)

    Date Cleared
    2018-11-09

    (71 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K082519, K162209

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Teleport microcatheters are indicated for:
    supporting and facilitating the placement of guidewires in the coronary and peripheral vasculature.
    exchanging guidewires in the coronary and peripheral vasculature.
    the delivery of contrast media into the coronary, peripheral, and abdominal vasculature.

    Device Description

    The Teleport family of microcatheters are single lumen catheters, offered in two shaft sizes (2.0F and 2.1F) with working lengths of 135cm or 150cm, designed for use in the coronary, peripheral, and abdominal vasculature. The shaft profiles gradually decrease from 2.6F (0.034") to 2.0F (0.0265") for the 2.0F configuration (Teleport), and 2.7F (0.036") to 2.1F (0.0275") for the 2.1F configuration (Teleport Control) respectively. The catheter consists of four sections: body shaft, proximal tip shaft, distal tip shaft, and a radiopaque tip. The distal most 60cm of the outer surface is coated with a hydrophilic polymer to increase lubricity and the lumen of the catheter is lined with a fluoropolymer to facilitate movement of the guidewire. The catheter is compatible with a standard 0.014 inch (0.36mm) guidewire.

    AI/ML Overview

    The provided text describes a 510(k) summary for the Teleport Microcatheter, focusing on its substantial equivalence to a predicate device. However, it does not contain a detailed study proving the device meets specific acceptance criteria in the format requested. Instead, it lists the types of performance tests conducted and a general statement that "The Teleport microcatheter test results met all acceptance criteria and were similar to the predicate and reference devices."

    Therefore, I cannot provide a table of acceptance criteria and reported device performance, nor can I provide information on sample sizes for test sets, data provenance, number or qualifications of experts, adjudication methods, MRMC studies, standalone performance, type of ground truth used, or training set details as these are not present in the provided document.

    The document confirms that standalone performance testing was done for the device, as indicated by the list of "Performance Testing" conducted, which did not involve human readers (e.g., flow rate, burst pressure, torque strength).

    Ground truth for these engineering and material performance tests would be established through predefined engineering specifications and standards (e.g., a specific flow rate must be achieved, a certain pressure must be withstood before burst).

    Ask a Question

    Ask a specific question about this device

    K Number
    K173894

    Validate with FDA (Live)

    Date Cleared
    2018-02-09

    (50 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K162209, K141118

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Jade PTA Balloon Dilatation Catheter is indicated for Percutaneous Transluminal Angioplasty in the peripheral vasculature, including iliac, femoral, popliteal, infra-popliteal, and renal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. This device is also indicated for post-dilation of balloon expandable and self-expanding stents in the peripheral vasculature.

    Device Description

    The JADE Percutaneous Transluminal Angioplasty (PTA) Balloon Dilatation Catheter is a rapid exchange balloon catheter for peripheral indications with a working length of 150cm. The minimally compliant balloons, available in diameters from 1.5-6.0mm and lengths from 15-120mm, can be inflated by injecting dilute contrast media solution through the trailing hub of the catheter. The balloon material is made of a minimally compliant material with a rated burst pressure of 20 atm (Ø1.5-4.0mm) or 18atm (Ø4.5-6.0mm). Hydrophilic lubricious coatings are applied to the distal section of the catheter. The proximal shaft of the catheter is composed of a female luer connector bonded to a nylon tube which is internally supported by a stainless steel hypotube. Two radiopaque platinum/iridium marker bands are located within the balloon shoulders. The internal lumen of the catheter accepts a maximum 0.014 inch (0.36mm) guidewire. The guidewire enters the catheter tip and advances coaxially out the RX port, thereby allowing both coaxial guidance and rapid exchange of catheter with a single standard length guidewire. Two marked sections, 3mm in length located on the proximal shaft specifically designed to be highly visible, indicate catheter position relative to the guiding catheter or guiding sheath. The design of this dilatation catheter does not incorporate a lumen for distal dye injections or distal pressure measurements.

    AI/ML Overview

    The provided text is a 510(k) Summary for the Jade PTA Balloon Dilatation Catheter. It describes the device, its intended use, and the performance data submitted to support its substantial equivalence to predicate devices. However, it does not contain information about a study proving the device meets acceptance criteria related to an AI/ML algorithm or its performance with human readers. The information provided is for a physical medical device (a balloon catheter), not an AI/ML diagnostic or assistive tool.

    Therefore, I cannot fulfill your request as the necessary information regarding AI/ML acceptance criteria, human reader studies, ground truth establishment for AI/ML, sample sizes for training/test sets for AI/ML, or expert qualifications for AI/ML ground truth is not present in the provided document.

    The "Performance Data" section in the document refers to in vitro performance tests and biocompatibility tests for the physical catheter, not an AI/ML system.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1