K Number
K173680

Validate with FDA (Live)

Device Name
Sapphire II PRO
Date Cleared
2018-03-01

(90 days)

Product Code
Regulation Number
870.5100
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Sapphire II PRO coronary dilatation catheter (1.0-1.25mm configurations) is indicated for: * balloon pre-dilatation of a stenotic portion of a coronary artery or bypass graft stenosis (>70% stenosis) for the purposes of improving myocardial perfusion.

Device Description

The Ø1.0-1.25mm Sapphire II PRO coronary dilatation catheter is a percutaneous transluminal coronary angioplasty (PTCA) balloon catheter with a working length of 140cm. The proximal shaft is a PTFE coated stainless steel hypotube. Hydrophilic lubricious coatings are applied to the distal section. The semi-compliant balloons, available in diameters from 1.0-1.25mm and lengths from 5-15mm, can be inflated by injecting dilute contrast media solution through the trailing hub of the catheter. One radiopaque platinum marker band is centrally located within the balloon segment. The catheter is compatible with 5F or larger guiding catheters. The internal lumen of the catheter accepts a standard 0.014 inch PTCA guidewire. The proximal part of the guidewire enters the catheter tip and advances coaxially out the catheter proximal port, thereby allowing both coaxial guidance and rapid exchange of catheters with a single standard length guidewire. Two marked sections are located on the hypotube shaft to indicate catheter position relative to the tip of either a brachial or femoral guiding catheter. The design of this dilatation catheter does not incorporate a lumen for distal dye injections or distal pressure measurements.

AI/ML Overview

The information provided is about the Sapphire II PRO Coronary Dilatation Catheter (1.0-1.25mm configurations). This is a medical device, and the document describes a clinical study to assess its safety and effectiveness.

Here's the breakdown of the requested information based on the provided text:

1. A table of acceptance criteria and the reported device performance

The document does not explicitly present a "table of acceptance criteria" with corresponding "reported device performance" in a pass/fail format for each criterion. Instead, it describes a primary endpoint for Device Procedural Success and secondary endpoints for effectiveness and safety, along with their observed results. I will construct a table reflecting these key outcomes.

Acceptance Criteria (Study Endpoint Description)Reported Device Performance (Sapphire II PRO)
Primary Endpoint: Device Procedural Success (Composite of):59/61 (96.7%) of patients (site-reported)
1. Successful delivery, inflation, deflation, and withdrawal of the study balloon.60/61 (98.4%) of patients (site-reported)
2. No evidence of vessel perforation, flow-limiting dissection (grade C or higher), or reduction in TIMI flow from baseline related to the study balloon.61/61 (100%) of patients (no procedural complications observed)
3. Final TIMI flow grade of 3 at the conclusion of the PCI procedure.60/61 (98.4%) of patients (site-reported)
Secondary Peri-Procedural Endpoints (Effectiveness):
Successful balloon delivery to the target lesion.61 (100.0%)
Successful inflation at the target lesion.60 (98.4%)
Successful deflation and withdrawal of the study balloon.61 (100.0%)
Absence of vessel perforation.61 (100.0%)
Absence of flow limiting dissection (Grade C or higher).61 (100.0%)
No reduction in TIMI flow from baseline related to the study balloon. (analyzable in 51 patients)51/51 (100.0%)
Final TIMI flow grade of 3 at the conclusion of the PCI procedure.60 (98.4%)
Absence of balloon rupture of the study balloon.60 (98.4%) (1 subject had balloons pin-holed and ruptured)
Improvement in Minimum Lumen Diameter (measured by QCA). (analyzable in 51 patients)49/51 (96.1%)
Lesion success (patient based).60 (98.4%)
Clinically significant arrhythmia (peri-procedural).0 (0.0%)
Secondary In-hospital Clinical Safety and Efficacy Endpoints:
In-hospital MACE (composite of all death, target vessel MI or clinically indicated TLR).1 (1.6%)
Death (Cardiac).0 (0.0%)
Death (Non-cardiac).0 (0.0%)
MI (Target vessel MI).1 (1.6%) (peri-procedural, non-Q-wave MI, elevated CK-MB, did not require treatment)
MI (Non-target vessel MI).0 (0.0%)
Clinically indicated revascularization (TLR).0 (0.0%)
In-hospital Target Lesion Failure (TLF) (composite of cardiac death, target vessel MI or clinically indicated TLR).1 (1.6%)
In-hospital stent thrombosis (ST) within the target vessel.0 (0.0%)
Clinically significant arrhythmias requiring intervention.0 (0.0%)

2. Sample size used for the test set and the data provenance

  • Sample Size: 61 patients (67 lesions)
  • Provenance: Prospective, open-label, multicenter, single-arm, observational study conducted at four (4) US sites.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

The document mentions "Angiographic Core Lab" and "Clinical Event Committee (CEC) adjudication" for various endpoints.

  • Angiographic Core Lab: Responsible for reporting several secondary peri-procedural endpoints (e.g., vessel perforation, dissection, TIMI flow, MLD). The number and qualifications of experts in this core lab are not specified in the provided text.
  • Clinical Event Committee (CEC): Responsible for adjudicating in-hospital clinical safety and efficacy endpoints (e.g., MACE, MI, TLR, TLF, ST). The number and qualifications of experts in the CEC are not specified in the provided text.
  • Site Reported and Adjudicated: Some endpoints were site-reported and then adjudicated, implying review by professionals at the study sites, but specific numbers and qualifications are not provided.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

The document states:

  • "Site Reported and Adjudicated" for some peri-procedural endpoints.
  • "Core Lab Reported" for other peri-procedural endpoints.
  • "Clinical Event Committee (CEC) adjudication" for in-hospital clinical safety and efficacy endpoints.

However, the specific adjudication method (e.g., 2+1, 3+1 consensus) used by the Core Lab or CEC is not detailed in the provided text.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

There is no mention of an MRMC study or AI assistance in this document. This study is a clinical trial for a coronary dilatation catheter, not an AI-powered diagnostic device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This question is not applicable. The device is a physical medical instrument (a catheter) and not an algorithm or AI system.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The ground truth was established through a combination of:

  • Clinical observation and assessment: By the treating physicians at the study sites, reported as "Site Reported."
  • Angiographic Core Lab analysis: Objective assessment of angiograms.
  • Clinical Event Committee (CEC) adjudication: Expert review and classification of clinical events.
  • Outcomes data: Such as the occurrence of MACE, MI, and TLR.

8. The sample size for the training set

This question is not applicable as the study describes a clinical trial for a physical medical device, not a machine learning model. There is no concept of a "training set" for the device itself in this context.

9. How the ground truth for the training set was established

This question is not applicable as there is no training set for the device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.

March 1, 2018

OrbusNeich Medical Trading Inc. John Pazienza Senior Director, Engineering 5363 NW 35th Avenue Fort Lauderdale, Florida 33309

Re: K173680

Trade/Device Name: Sapphire II PRO Regulation Number: 21 CFR 870.5100 Regulation Name: Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheter Regulatory Class: Class II Product Code: LOX Dated: November 28, 2017 Received: December 1, 2017

Dear John Pazienza:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Kenneth J. Cavanaugh -S

for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICESForm Approved: OMB No. 0910-0120
Food and Drug AdministrationExpiration Date: 06/30/2020
See PRA Statement below.

Indications for Use

510(k) Number (if known)K173680
Device NameSapphire II PRO Coronary Dilatation Catheter
Indications for Use (Describe)The Sapphire II PRO coronary dilatation catheter (1.0-1.25mm configurations) is indicated for: * balloon pre-dilatation of a stenotic portion of a coronary artery or bypass graft stenosis (>70% stenosis) for the purposes of improving myocardial perfusion.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer

Paperwork Reduction Act (PRA) Staff

PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."

FORM FDA 3881 (7/17)Page 1 of 1
-----------------------------------

AOTA mmoss.tr-0.tr OA9 II arinqques

A0-10-(x)015-7102

976 to 88

{3}------------------------------------------------

K173680

510(k) Summary

This 510(k) Summary is submitted in accordance with 21 CFR 807.92(c).
Submitter:OrbusNeich Medical Trading, Inc.5363 NW 35th AvenueFort Lauderdale, FL 33309Phone: 954.730.0711Fax: 954.730.7601
Contact Person:John D. Pazienza
Date Prepared:February 2, 2018
Trade Name:Sapphire II PRO Coronary Dilatation Catheter
Common Name:Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheter
Classification Name:Catheters, transluminal coronary angioplasty, percutaneous (21 CFR870.5100(a), Product Code LOX)
Predicate Device:Sapphire II PRO (K163114; cleared January 5, 2017)Mini Trek (K110617; cleared June 2, 2011)Emerge (K130391; cleared July 10, 2013)Sprinter Legend (K103095; cleared October 22, 2010)
Device Description:The Ø1.0-1.25mm Sapphire II PRO coronary dilatation catheter is apercutaneous transluminal coronary angioplasty (PTCA) ballooncatheter with a working length of 140cm. The proximal shaft is aPTFE coated stainless steel hypotube. Hydrophilic lubricious coatingsare applied to the distal section. The semi-compliant balloons,available in diameters from 1.0-1.25mm and lengths from 5-15mm,can be inflated by injecting dilute contrast media solution through thetrailing hub of the catheter. One radiopaque platinum marker band iscentrally located within the balloon segment. The catheter iscompatible with 5F or larger guiding catheters. The internal lumen ofthe catheter accepts a standard 0.014 inch PTCA guidewire. Theproximal part of the guidewire enters the catheter tip and advancescoaxially out the catheter proximal port, thereby allowing both coaxialguidance and rapid exchange of catheters with a single standard lengthguidewire. Two marked sections are located on the hypotube shaft toindicate catheter position relative to the tip of either a brachial orfemoral guiding catheter. The design of this dilatation catheter doesnot incorporate a lumen for distal dye injections or distal pressuremeasurements.
Intended Use:The Sapphire II PRO Coronary Dilatation Catheter (Ø1.0-1.25mmconfigurations) is indicated for:• balloon pre-dilatation of a stenotic portion of a coronaryartery or bypass graft stenosis (≥70% stenosis) for thepurposes of improving myocardial perfusion.

{4}------------------------------------------------

  • Comparisons of the new and predicate devices show that the Technological Characteristics: technological characteristics such as product performance, design, and intended use are substantially equivalent to the currently marketed predicate devices.
  • Performance Data: Both in vitro performance tests, such as dimensional verification, balloon preparation, deployment, and retraction, balloon rated burst pressure, balloon fatigue, balloon compliance, balloon inflation and deflation time, catheter bond strength, tip pull strength, flexibility and kinking, torque strength, radiopacity, coating integrity, and particulate evaluation, and also biocompatibility tests, such as cytotoxicity, sensitization, intracutaneous reactivity, acute systemic toxicity, hemocompatibility (hemolysis, partial thromboplastin time, platelet and leukocyte counts, complement activation, and in vivo thromboresistance), pyrogenicity, and genotoxicity (bacterial mutagenicity and in vitro mouse lymphoma) were conducted on the Ø1.0-1.25mm Sapphire II PRO coronary dilatation catheter. The test results met all acceptance criteria, were similar to predicate devices, and ensure that the Ø1.0-1.25mm Sapphire II PRO coronary dilatation catheter design and construction are suitable for its intended use as recommended by the Class II Special Controls Guidance Document for Certain Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheters (FDA; September 8, 2010).

Clinical Data: Purpose

The purpose of this study was to assess the acute safety and device procedural success of the Sapphire II PRO Ø1.0 and 1.25 mm coronary dilatation catheter in its intended use for the initial dilatation of coronary artery or by-pass graft stenosis.

Study Design

The Sapphire II PRO Study was a prospective, open label, multicenter, single arm, observational study designed to evaluate the acute safety and device procedural success of the Sapphire II PRO Ø1.0 and 1.25 mm coronary dilatation catheters in patients with stenotic coronary arteries or bypass grafts (≥70% diameter stenosis) during percutaneous coronary intervention (PCI).

Sixty-one (61) patients were treated at four (4) US sites with the Sapphire II PRO coronary dilatation catheter to pre-dilate coronary arteries or bypass grafts during their index procedure. All patients were screened according to the protocol inclusion and exclusion criteria and were followed through hospital discharge.

The primary endpoint was Device Procedural Success consisting of a composite of:

  • Successful delivery, inflation, deflation and withdrawal of the . study balloon;
  • No evidence of vessel perforation, flow limiting dissection . (grade C or higher) or reduction in TIMI flow from baseline related to the study balloon;
  • . Final TIMI flow grade of 3 at the conclusion of the PCI procedure.

{5}------------------------------------------------

The secondary peri-procedure related endpoints of study device effectiveness included the individual components of the primary endpoint. along with the angiographic core lab determined improvement in the Minimum Lumen Diameter (MLD) following predilation of the lesion with the study balloon, and Lesion Success defined as successful PCI in the absence of vessel perforation, flow limiting dissection (grade C or higher), or reduction in TIMI flow from baseline related to the study balloon, or clinically significant arrhythmias.

The secondary in-hospital clinical safety and efficacy endpoints reported through hospital discharge included: in-hospital Major Adverse Cardiac Events (MACE), defined as a composite of; all death (cardiac and non-cardiac), Myocardial Infarction (MI) and clinically indicated Target Lesion Revascularization (TLR). Also reported are the individual components of the MACE composite end-point, inhospital Target Lesion Failure (TLF), defined as a composite of cardiac death, target vessel MI, or clinically indicated TLR, along with in-hospital Stent Thrombosis (ST) within the target vessel, or clinically significant arrhythmias (requiring intervention). Both MI and ST were determined by the Academic Research Consortium (ARC) classification criteria.

Demographics and Baseline Lesion Characteristics

A total of 61 patients including 67 lesions were treated in accordance to the protocol at 4 U.S. sites. All patients were screened according to the protocol inclusion and exclusion criteria and were followed through hospital discharge. The analyses were performed on the intent-to-treat (ITT) population and were descriptive.

The mean age of the study population was 67.26 ± 9.35 years and 11/61 (18.0%) were female. The prevalence of traditional cardiovascular risk factors in enrolled patients included 19/61 (31.1%) diabetes mellitus, 10/61 (16.4%) current smoking, and 51/61 (83.6%) hypertension. Angina status for the enrolled patients was reported as 21/61 (34.4%) unstable angina. 31/61 (50.8%) stable angina. 1/61 (1.6%) NSTEMI, and 8/61 (13.1%) were asymptomatic.

A total of 67 lesions were treated in the study, of which 63/67 (94%) were located in native coronary arteries and 4/67 (6.0%) were located in bypass grafts. Treated vessels included: diagonal branch of the LAD 10/67 (14.9%), LAD 22/67 (32.8%), LCx 12/67 (17.9%), marginal branch of the LCx 1/67 (1.5%), posterior descending branch of RCA 2/67 (3.0%), and RCA 16/67 (23.9%). Total occlusion was observed in 9/67 (13.4%) lesions and severe calcification was observed in 24/67 (35.8%) of the lesions. Multi-vessel disease was present in 33/61 (54.1%) of the patients and the average lesion length was 18.42 ± 12.2 mm.

Study Results

As shown in Table 1, Device Procedural Success was site reported as achieved in 59/61 (96.7%) of the patients, including successful delivery, inflation, deflation and withdrawal of all the study balloons. No procedural complications were observed in the intent-to-treat patient population, this includes no vessel perforation, flow limiting dissection (grade C or higher), or reduction in TIMI flow from baseline

{6}------------------------------------------------

related to the study balloon. Device procedural failure was reported in two subjects 2/61 (3.3%): i) one subject presented with a chronic total occlusion (CTO) and the lesion was not able to be successfully dilated with any device such that the final TIMI flow did not reach grade 3 at the conclusion of the PCI procedure, and ii) one subject where two study balloons pin-holed and ruptured at >12 atm inflation pressure however the lesions were able to be fully opened for a successful PCI.

Table 1, Primary Endpoint Outcomes

Primary EndpointSapphire II PRO
Patients (n=61)Lesions (n=67)
Device Procedural Success59/61 (96.7%)64/67 (95.5%)
Success of delivery, inflation/deflation and withdrawal60/61 (98.4%)65/67 (97.0%)
Absence of vessel perforation, flow limiting dissection (grade C orhigher), or reduction in TIMI flow from baseline related to the studyballoon.61/61 (100%)67/67 (100%)
Final TIMI flow grade of 3 at the conclusion of the PCI procedure.60/61 (98.4%)66/67 (98.5%)

As shown in Tables 2a and 2b, the secondary peri-procedure related endpoints of study device effectiveness and in-hospital clinical safety and efficacy measured through hospital discharge and included site reported successful delivery, inflation, deflation and withdrawal of the study balloon, with the absence of vessel perforation, flow limiting dissection (Grade C or Higher), and no reduction in TIMI flow from baseline related to the study balloon. In-hospital Major Adverse Cardiac Events (MACE) was observed in 1/61 (1.6%) of the patients. The peri-procedural, non-Q-wave MI event was characterized by elevation of post-procedure creatine kinase-myoglobin band (CK-MB) levels and did not require treatment. The all-cause death and TLR rates were 0.0% (0/61). There were no reports of TLR 0.0% (0/61), clinically indicated or in-hospital stent thrombosis (ST) 0.0% (0/61) within the target vessel. There were no clinically significant arrhythmias requiring intervention 0.0% (0/61) and no additional device observations reported in the conduct of the study.

Table 2a, Study Device Effectiveness Outcomes
-----------------------------------------------------
Secondary EndpointSite Reported orAngiographic Central CoreLab Reported (QCA)OverallPopulation(n=61 patients)
Peri-procedural end-points of study device effectiveness
Successful balloon delivery to the target lesionSite Reported and Adjudicated61 (100.0%)
Successful inflation at the target lesionSite Reported and Adjudicated60 (98.4%)
Successful deflation and withdrawal of the study balloonSite Reported and Adjudicated61 (100.0%)
Absence of vessel perforationCore Lab Reported61 (100.0%)
Absence of flow limiting dissection (Grade C or higher)Core Lab Reported61 (100.0%)
No reduction in TIMI flow from baseline related to the studyballoon*Core Lab Reported51/51 (100.0%)
Final TIMI flow grade of 3 at the conclusion of the PCI procedureCore Lab Reported60 (98.4%)
Absence of balloon rupture of the study balloonSite Reported and Adjudicated60 (98.4%)
Improvement in Minimum Lumen Diameter (measured byQCA)**Core Lab Reported49/51 (96.1%)
Lesion success (patient based)Core Lab Reported60 (98.4%)
Clinically significant arrhythmiaSite Reported and Adjudicated0 (0.0%)
  • The TIMI flow after study balloon was not analyzable in 10 patients by the Angiographic Core Lab. Theref

reduction in TIMI flow from baseline related to the study balloon could only be analyzed in 51 patients *The Improvement in MLD after use of the study balloon as measured by OCA was not analyzable in 10 patients by the Angiographic Core Lab. Therefore, this value could only be analyzed in 51 patients.

{7}------------------------------------------------

Secondary EndpointSite Reported or ClinicalEvent Committee (CEC)adjudicationOverallPopulation(n=61 patients)
In-hospital clinical safety and efficacy
In-hospital MACE - (composite of all death, target vessel MI orclinically indicated TLR)CEC adjudication1 (1.6%)
Death0 (0.0%)
• CardiacCEC adjudication0 (0.0%)
• Non-cardiacCEC adjudication0 (0.0%)
MI1 (1.6%)
• Target vessel MICEC adjudication1 (1.6%)
• Non-target vessel MICEC adjudication0 (0.0%)
Clinically indicated revascularization
• TLRCEC adjudication0 (0.0%)
In-hospital Target Lesion Failure (TLF)(composite of cardiac death, target vessel MI or clinically indicatedTLR)CEC adjudication1 (1.6%)
In-hospital stent thrombosis (ST) within the target vesselCEC adjudication0 (0.0%)
Clinically significant arrhythmias requiring interventionCEC adjudication0 (0.0%)

Conclusion

The results of the Sapphire II PRO Ø1.0 and 1.25 mm diameter clinical study support the acute safety and device success of the Sapphire II PRO Ø1.0 and 1.25 mm Coronary Dilatation Catheter and its intended use as a pre-dilatation catheter in the stenotic portion of a coronary artery or bypass graft stenosis (≥70% diameter stenosis).

Conclusion: This information supports a determination of substantial equivalence between the Ø1.0-1.25mm Sapphire II PRO coronary dilatation catheter and the predicate devices described above.

§ 870.5100 Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheter.

(a)
Standard PTCA Catheter —(1)Identification. A PTCA catheter is a device that operates on the principle of hydraulic pressurization applied through an inflatable balloon attached to the distal end. A PTCA balloon catheter has a single or double lumen shaft. The catheter features a balloon of appropriate compliance for the clinical application, constructed from a polymer. The balloon is designed to uniformly expand to a specified diameter and length at a specific pressure as labeled, with well characterized rates of inflation and deflation and a defined burst pressure. The device generally features a type of radiographic marker to facilitate fluoroscopic visualization of the balloon during use. A PTCA catheter is intended for balloon dilatation of a hemodynamically significant coronary artery or bypass graft stenosis in patients evidencing coronary ischemia for the purpose of improving myocardial perfusion. A PTCA catheter may also be intended for the treatment of acute myocardial infarction; treatment of in-stent restenosis (ISR) and/or post-deployment stent expansion.(2)
Classification. Class II (special controls). The special control for this device is “Class II Special Controls Guidance Document for Certain Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheters.” See § 870.1(e) for the availability of this guidance document.(b)
Cutting/scoring PTCA Catheter —(1)Identification. A cutting/scoring PTCA catheter is a balloon-tipped catheter with cutting/scoring elements attached, which is used in those circumstances where a high pressure balloon resistant lesion is encountered. A cutting/scoring PTCA catheter is intended for the treatment of hemodynamically significant coronary artery stenosis for the purpose of improving myocardial perfusion. A cutting/scoring PTCA catheter may also be indicated for use in complex type C lesions or for the treatment of in-stent restenosis.(2)
Classification. Class III (premarket approval). As of May 28, 1976, an approval under section 515 of the act is required before this device may be commercially distributed. See § 870.3.