510(k) Data Aggregation

    K Number
    K173709

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-01-26

    (53 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K161262

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Zenith Support Catheter is indicated for the introduction of interventional devices into the peripheral and neuro vasculature.

    Device Description

    The InNeuroCo Zenith Support product consists of a catheter, a hemostasis valve, and a split sheath introducer. The Zenith Support catheter is a single lumen, variable stiffness catheter that has an outer polymer shaft and is reinforced with a Stainless Steel cross coil. The Zenith Support catheter has a radiopaque Platinum/Iridium marker band on the distal end. The Zenith Support is available with an internal diameter of 0.074 inches and a nominal outer diameter of 0.086 inches. There are three working lengths available: 95 cm, 105 cm, and 115 cm in either diameter. The Zenith Support catheter has a PTFE-lined lumen throughout the catheter shaft. The catheter is flexible and has a hydrophilic coating. The Zenith Support catheter is inserted through a guide catheter or vascular sheath, provides access to the target site, and once in place, provides a reinforcing conduit for other intravascular devices. Accessories included with the device are a Tuohy-Borst hemostasis valve with an extension luer and a split sheath introducer. The Zenith product line is supplied sterile, non-pyrogenic, and intended for single use only.

    AI/ML Overview

    This request concerns the InNeuroCo Zenith Support Catheter, a percutaneous catheter used for introducing interventional devices into the peripheral and neuro vasculature. The provided document is a 510(k) premarket notification summary, which focuses on demonstrating substantial equivalence to a predicate device rather than a comprehensive study for novel device acceptance criteria.

    The information primarily revolves around non-clinical data and justifications for not re-testing certain parameters due to the device's similarity to its predicate. Therefore, detailed information on ground truth establishment, expert qualifications, adjudication methods, or MRMC studies is not available as this type of submission typically relies on a comparison to an already approved device.

    Here's an analysis based on the provided document:


    1. Table of Acceptance Criteria and Reported Device Performance

    The document provides a "SUMMARY OF NON-CLINICAL DATA" (page 8, "Table 1: Non-Clinical Testing Analysis") which serves this purpose. Many tests were not re-tested for the Zenith Support Catheter because no material changes were made from the predicate device (Zenith 074) and manufacturing processes were identical or within the predicate's tolerance.

    TestAcceptance CriteriaReported Device Performance (Analysis/Outcome)
    BiocompatibilityNo re-testing for Zenith Support Catheter: "There are no material changes from the predicate device. Manufacturing process is identical to the predicate. Only the segment lengths were changed and pitch adjustments were within the predicate device's tolerance." This applies to all biocompatibility tests below.
    - Material Mediated PyrogenThe test article extracts must not cause a febrile reaction greater than 0.5°C in any individual subject.Not re-tested.
    - Cytotoxicity MEM ElutionThe cultures treated with the test article must not have a reactivity grade greater than 2.Not re-tested.
    - Hemolysis (ASTM Method, extract human blood)The hemolytic index above the negative control article must be less than 5%.Not re-tested.
    - Hemolysis (ASTM method, direct contact human blood)The hemolytic index above the negative control article must be less than 5%.Not re-tested.
    - Unactivated Partial Thromboplastin TimeThere must be no statistical decrease between the UPTT of plasma exposed to the test article and to the negative or untreated control.Not re-tested.
    - Complement ActivationThere must be no statistical increase between either the C3a or SC5b-9 concentrations in plasma exposed to the test article as compared to the negative and untreated controls.Not re-tested.
    - Dog ThromboresistanceThe test articles must receive a thrombus formation score less than or equal to that of the control.Not re-tested.
    - Maximization SensitizationThe test article must elicit a positive response in less than 10% of the test animals.Not re-tested.
    - Intracutaneous Toxicity/ReactivityThe test article extracts must not induce a significantly greater biological reaction than the control.Not re-tested.
    - Acute Systemic Toxicity TestThe test article extracts must not induce a significantly greater biological reaction than the control.Not re-tested.
    Animal Testing - Angiographic EvaluationNo units tested can cause vessel injury.Not re-tested. "The tips, materials, and transitions are identical to the predicate, and the initial evaluation for the predicate included the use of the device without the Scout Introducer."
    Chemical CompatibilityChemicals have no negative effect on the catheter or accessories.Not re-tested. "There are no material changes from the predicate device, and pitch adjustments are within the tolerance of the predicate device."
    Radiographic DetectabilityProduct shall be visible under fluoro imaging.Not re-tested. "The marker bands are identical to the predicate."
    Visual InspectionCatheter shall appear free from damage, including a rounded tip and smooth transition points.Not re-tested. "There are no material changes from the predicate device, and pitch adjustments are within the tolerance of the predicate device."
    Working LengthTest samples should be within existing working length specification.Not re-tested. "The segment lengths do not impact the cut to length step in addition the working length test compares the output to the target length."
    Hub compatibilityHub shall meet existing Luer specifications.Not re-tested. "The hub is identical to the predicate device."
    Outside DiameterTest samples should be within existing outside diameter specifications.Not re-tested. "OD is unaffected by the working length change."
    Leak – AirTest samples should be within existing Air - Leak specifications.Not re-tested. "There are no material changes from the predicate device, and pitch adjustments are within the tolerance of the predicate device."
    Leak – LiquidTest samples should be within existing Leak - Liquid specifications.Not re-tested. "There are no material changes from the predicate device, and pitch adjustments are within the tolerance of the predicate device."
    ParticulatesTest samples should be within existing Particulate specifications.Not re-tested. "There are no material changes from the predicate device, and pitch adjustments are within the tolerance of the predicate device."
    Simulated Use - BenchTest samples must meet predetermined user needs.Re-tested Yes. "Zenith Support Catheter test samples met the acceptance criteria for Simulated Use-Bench to demonstrate that it is substantially equivalent to the predicate device."
    Catheter BurstTest sample burst pressures must meet or exceed existing minimum burst pressure specification.Not re-tested. "There are no material changes from the predicate device, and pitch adjustments are within the tolerance of the predicate device."
    TensileTest sample ultimate tensile strength must meet or exceed existing tensile strength specifications.Not re-tested. "The bonds and joints are the same as the predicate Zenith 074 device. Any pitch adjustments are within the tolerance of the predicate device."
    CorrosionTest samples shall exhibit no evidence of corrosion.Not re-tested. "There are no material changes from the predicate device, and pitch adjustments are within the tolerance of the predicate device."
    Packaging - Dye LeakTest sample shall not exhibit any visual leaks or channels.Not re-tested. "The packaging is identical to the cleared reference SDA device." (Note: The reference here is to an "SDA device," not the Zenith 074 predicate, indicating identical packaging to another cleared device.)
    Packaging - PeelTest sample tensile strength must meet or exceed existing tensile strength specifications.Not re-tested. "The packaging is identical to the cleared reference SDA device." (Again, referring to SDA device for packaging.)
    Kink ResistanceTest sample kink resistance must meet or exceed existing Kink Resistance specifications.Not re-tested. "All transitions are the same, only the location of the transition changes, in addition, the pitch adjustments are within the Zenith 074 tolerance."
    TorqueTest sample torque results must meet or exceed existing torque specifications.Not re-tested. "There are no material changes from the predicate device, and pitch adjustments are within the tolerance of the predicate device."
    PTFE Liner InspectionTest sample liner adhesion must meet or exceed existing PTFE Liner inspection specifications.Not re-tested. "There are no material changes from the predicate device, and pitch adjustments are within the tolerance of the predicate device."
    Hydrophilic Coating IntegrityTest sample results must meet or exceed existing Hydrophilic Coating Integrity specifications.Not re-tested. "There are no material changes from the predicate device, and pitch adjustments are within the tolerance of the predicate device."
    Labeling LegibilityTest samples shall demonstrate text legibility.Not re-tested. "The label is almost identical to the predicate device, except that the Scout Introducer is removed from the labeling."
    BarcodeTest samples shall demonstrate readily readable barcodes.Not re-tested. "Barcodes are similar to the predicate."
    SterilizationSterilization load shall pose an equal or lesser challenge to sterilize than existing sterile product loads.Not re-tested. "The Zenith Support Catheter was adopted into the Zenith Family Sterilization Cycle."
    Shelf LifeAged test samples must meet or exceed existing specifications.Not re-tested. "There are no material changes from the predicate device, therefore shelf life testing conducted for the Zenith 074 still applies."

    2. Sample Size Used for the Test Set and the Data Provenance

    Due to the nature of a 510(k) summary relying on substantial equivalence, specific sample sizes for re-tested parameters (like Simulated Use - Bench) are not explicitly stated. For most tests, the data provenance is essentially inherited from the predicate device (Zenith 074) and another reference device (InNeuroCo SDA) because the new device's materials and manufacturing are identical or within tolerance. The document indicates these are non-clinical studies (benchtop and animal models for predicate devices). The geographical origin of the data is not specified beyond the manufacturer being based in Florida, USA. All studies are retrospective in relation to the current device's testing, as they refer to data from predicate devices.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This information is not provided. For non-clinical tests, "ground truth" is typically established by laboratory procedures and measurements against defined standards (e.g., ISO, ASTM). For the "Simulated Use - Bench" test, it states it was conducted "by a physician," implying at least one expert, but qualifications and number are not detailed.

    4. Adjudication Method for the Test Set

    Not applicable/not provided. This type of detail (e.g., 2+1, 3+1 consensus) is typical for clinical studies, especially those involving expert interpretation of medical images, which is not the focus of this non-clinical submission.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a medical device (catheter), not an AI/Software as a Medical Device (SaMD) product. Therefore, MRMC studies and AI-assisted performance metrics are not relevant or performed for this submission.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. The device is a physical catheter, not an algorithm.

    7. The Type of Ground Truth Used

    For the majority of the non-clinical tests not re-performed, the "ground truth" (or accepted performance) is based on:

    • Established performance of the predicate device (Zenith 074): This implies that the predicate device previously met these standards.
    • International Standards (ISO 10993, ISO 10555-1, ISO 594-1, ISO 594-2, ISO 11135) and ASTM standards (ASTM F1929-12, ASTM F88-09, ASTM F1980): These standards define methodologies and acceptable ranges for material properties, biocompatibility, and device functionality.
    • USP 788: For particulates.

    For the re-tested "Simulated Use - Bench" test, the ground truth was "predetermined user needs," likely established through design validation or clinical input.

    8. The Sample Size for the Training Set

    Not applicable. The Zenith Support Catheter is a physical medical device. The concept of "training set" is relevant for AI/machine learning models, which this device is not.

    9. How the Ground Truth for the Training Set was Established

    Not applicable as there is no "training set" for this physical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K172468

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-12-06

    (113 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K161262, K171672, K152202

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The 091 Long Sheath is indicated for the introduction of interventional devices into the peripheral, coronary, and neuro vasulature.

    Device Description

    The InNeuroCo, Inc. 091 Long Sheath is a variable stiffness catheter that has a catheter shaft reinforced with a stainless steel double coil. It has a radiopaque Platinum/lridium marker band on the distal end. The distal 10 cm of the 091 Long Sheath Catheter has a hydrophilic coating. The catheter has a nominal outer diameter of 0.109 inches and a nominal inner diameter of 0.091 inches. It is available in three working lengths: 70 cm, and 90 cm. The 091 Long Sheath has a PTFE-lined lumen. The 091 Long Sheath is inserted at a vascular access point to provide access to the target site and once in place, provides a reinforcing conduit for other intravascular devices. Accessories included with the device are a Tuohy-Borst Hemostasis Valve and a Vessel Dilator. The 091 Long Sheath is supplied sterile, non-pyrogenic, and intended for single use only.

    AI/ML Overview

    The provided text is a 510(k) summary for the InNeuroCo 091 Long Sheath peripheral, coronary, and neuro vasculature catheter. It outlines the device's characteristics and its comparison to a predicate device, along with the results of performance testing to demonstrate substantial equivalence.

    Based on the provided document, here's a description of the acceptance criteria and the study proving the device meets these criteria:

    This document describes the validation of a medical device, the InNeuroCo 091 Long Sheath, for regulatory approval. It is not an AI/ML device, and therefore several of the requested sections (e.g., number of experts, adjudication methods, MRMC study, standalone performance, training set details) are not applicable as they relate to the validation of AI/ML algorithms, not traditional medical devices like catheters.


    Acceptance Criteria and Device Performance Study for the InNeuroCo 091 Long Sheath

    The InNeuroCo 091 Long Sheath is a percutaneous catheter intended for the introduction of interventional devices into the peripheral, coronary, and neuro vasculature. The study conducted was a series of non-clinical (bench and animal) tests to demonstrate substantial equivalence to a legally marketed predicate device (Stryker AXS Infinity LS, K152876).

    1. Table of Acceptance Criteria and Reported Device Performance

    TestTest Method SummaryAcceptance CriteriaReported Device Performance (Conclusions)
    Biocompatibility-Material Mediated Pyrogen TestingCompleted per ISO 10993-11The test article extracts must not cause a febrile reaction greater than 0.5°C in any individual subject.Test results for the InNeuroCo Super Distal Access (K161262) and the Zenith Catheter (K171672) were leveraged as materials and manufacturing processes are equivalent. Met acceptance criteria.
    Biocompatibility-Cytotoxicity MEM ElutionCompleted per ISO 10993-5The cultures treated with the test article must not have a reactivity grade greater than 2.Leveraged results from InNeuroCo Super Distal Access (K161262) and Zenith Catheter (K171672). Met acceptance criteria.
    Biocompatibility-Hemolysis ASTM Method, extract human bloodCompleted per ISO 10993-4The hemolytic index above the negative control article must be less than 5%.Leveraged results from InNeuroCo Super Distal Access (K161262) and Zenith Catheter (K171672). Met acceptance criteria.
    Biocompatibility-Hemolysis, ASTM method, direct contact (human blood)Completed per ISO 10993-4The hemolytic index above the negative control article must be less than 5%.Leveraged results from InNeuroCo Super Distal Access (K161262) and Zenith Catheter (K171672). Met acceptance criteria.
    Biocompatibility-Unactivated Partial Thromboplastin TimeCompleted per ISO 10993-4There must be no statistical decrease between the UPTT of plasma exposed to the test article and to the negative or untreated control.Leveraged results from InNeuroCo Super Distal Access (K161262) and Zenith Catheter (K171672). Met acceptance criteria.
    Biocompatibility-Complement ActivationCompleted per ISO 10993-4There must be no statistical increase between either the C3a or SC5b-9 concentrations in plasma exposed to the test article as compared to the negative and untreated controls.Leveraged results from InNeuroCo Super Distal Access (K161262) and Zenith Catheter (K171672). Met acceptance criteria.
    Biocompatibility-Dog ThrombogenicityCompleted per ISO 10993-4The test articles must receive a thrombus formation score less than or equal to that of the control.Leveraged results from InNeuroCo Super Distal Access (K161262) and Zenith Catheter (K171672). Met acceptance criteria.
    Biocompatibility-Maximization SensitizationCompleted per ISO 10993-10The test article must elicit a positive response in less than 10% of the test animals.Leveraged results from InNeuroCo Super Distal Access (K161262) and Zenith Catheter (K171672). Met acceptance criteria.
    Biocompatibility-Intracutaneous Toxicity/ReactivityCompleted per ISO 10993-10The test article extracts must not induce a significantly greater biological reaction than the control.Leveraged results from InNeuroCo Super Distal Access (K161262) and Zenith Catheter (K171672). Met acceptance criteria.
    Biocompatibility-Acute Systemic Toxicity TestCompleted per ISO 10993-11The test article extracts must not induce a significantly greater biological reaction than the control.Leveraged results from InNeuroCo Super Distal Access (K161262) and Zenith Catheter (K171672). Met acceptance criteria.
    Animal Testing-Angiographic EvaluationProduct evaluated within a porcine model for clinically significant injury.No units tested can cause vessel injury.Test article met the acceptance criteria for Animal Testing.
    Bench Testing - Chemical CompatibilityCatheter exposed to chemicals readily available in a clinical setting.Chemicals have no negative effect on the catheter or accessories.Leveraged results from InNeuroCo Super Distal Access (K161262). Met acceptance criteria.
    Bench Testing - Visual InspectionCompleted per ISO 10555-1Catheter shall appear free from damage, including a rounded tip and smooth transition points.091 Long Sheath test samples met the acceptance criteria.
    Bench Testing - DimensionsCompleted per ISO 10555-1Test samples should be within existing dimensional specifications.091 Long Sheath test samples met the acceptance criteria. Leveraged results for Hemostasis Valve from InNeuroCo Super Distal Access (K161262), which also met the criteria.
    Bench Testing - Hub compatibilityCompleted per ISO 594-1 and ISO 594-2Hub shall meet existing Luer specifications.Leveraged results for the 091 Long Sheath and Dilator from InNeuroCo Intermediate Catheter (K152202) and for the Hemostasis Valve from Super Distal Access (K161262). All tested samples met the acceptance criteria.
    Bench Testing - Leak - AirCompleted per ISO 10555-1Test samples should be within existing Air – Leak specifications.091 Long Sheath test samples met the acceptance criteria.
    Bench Testing - Leak - LiquidCompleted per ISO 10555-1Test samples should be within existing Leak - Liquid specifications.091 Long Sheath test samples met the acceptance criteria.
    Bench Testing - ParticulatesCompleted per USP 788Test samples should be within existing Particulate specifications.091 Long Sheath test samples met the acceptance criteria.
    Bench Testing - Simulated UseLS underwent simulated use testing by a physician in a benchtop model.Test samples must meet predetermined user needs.091 Long Sheath test samples met the acceptance criteria.
    Bench Testing - Catheter BurstCompleted per ISO 10555-1Test sample burst pressures must meet or exceed existing minimum burst pressure specification.091 Long Sheath test samples met the acceptance criteria. ----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
    Bench Testing - TensileCompleted per ISO 10555-1Test sample ultimate tensile strength must meet or exceed existing tensile strength specifications.091 Long Sheath test samples met the acceptance criteria. ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
    Bench Testing - CorrosionCompleted per ISO 10555-1Test samples shall exhibit no evidence of corrosion.Leveraged results from InNeuroCo Intermediate Catheter (K152202). All samples met the acceptance criteria. --------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
    Bench Testing - Packaging - Dye LeakCompleted per ASTM F1929-12Test sample shall not exhibit any visual leaks or channels.091 Long Sheath test samples met the acceptance criteria. --------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
    Bench Testing - Packaging - PeelCompleted per ASTM F88-09Test sample tensile strength must meet or exceed existing tensile strength specifications.091 Long Sheath test samples met the acceptance criteria. ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
    Bench Testing - Kink ResistanceSamples are subjected to different diameters until kink is observed.Test sample kink resistance must meet or exceed existing Kink Resistance specifications.091 Long Sheath test samples met the acceptance criteria. ----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
    Bench Testing - TorqueConditioned samples are torqued to failure.Test sample torque results must meet or exceed existing torque specifications.091 Long Sheath test samples met the acceptance criteria. ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
    Bench Testing - PTFE Liner inspectionLS was challenged to demonstrate liner adherence.Test sample liner adhesion must meet or exceed existing PTFE Liner inspection specifications.091 Long Sheath test samples met the acceptance criteria. ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
    Bench Testing - Hydrophilic Coating IntegrityConditioned samples were repeatedly exposed to friction to demonstrate that the hydrophilic coating is not affected.Test sample results must meet or exceed existing Hydrophilic Coating Integrity specifications.091 Long Sheath test samples met the acceptance criteria. ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
    Bench Testing - Labeling LegibilityLabel is legible after printing.Test samples shall demonstrate text legibility.091 Long Sheath test samples met the acceptance criteria. ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
    Bench Testing - BarcodeBarcode is readable with a standard barcode reader.Test samples shall demonstrate readily readable barcodes.091 Long Sheath test samples met the acceptance criteria. ----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
    SterilizationISO 11135 and AAMI TIR 28Sterilization load shall pose an equal or lesser challenge to sterilize than existing sterile product loads.091 Long Sheath sterilization load met the acceptance criteria. ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
    Bacterial EndotoxinAAMIST72Test samples must meet the bacterial endotoxin acceptance criteria.The 091 Long Sheath samples met the acceptance criteria. ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
    Shelf LifeASTM F1980Aged test samples must meet or exceed existing specifications.091 Long Sheath test samples met the acceptance criteria. -----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The exact sample sizes for each specific test are not explicitly enumerated in the provided summary table. The statements typically indicate "Test samples" or "All samples," implying that a sufficient number were tested according to the referenced ISO or ASTM standards.
    • Data Provenance: The data is a mix of:
      • Direct testing: For many bench tests (e.g., Visual Inspection, Leaks, Particulates, Simulated Use, Catheter Burst, Tensile, Kink Resistance, Torque, PTFE Liner inspection, Hydrophilic Coating Integrity, Labeling Legibility, Barcode, Sterilization, Bacterial Endotoxin, Shelf Life). These tests were conducted on the 091 Long Sheath.
      • Leveraged data: For all biocompatibility tests, Chemical Compatibility, Hub compatibility, and Corrosion, data was leveraged from previously approved InNeuroCo devices (Super Distal Access K161262, Zenith Catheter K171672, and Intermediate Catheter K152202). This leveraging is justified by stating that the materials and manufacturing processes are equivalent.
      • Animal Testing: Product evaluated within a porcine model for angiographic evaluation.
    • Retrospective/Prospective: All described testing appears to be prospective (new tests conducted for this device and leveraged results from other devices), not retrospective analysis of existing clinical data.
    • Country of Origin: Not specified in the document, but given the FDA 510(k) submission, it would be expected to follow US or internationally recognized standards. The company address is in Florida, USA.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts

    • Not Applicable. This is a traditional medical device (catheter) validation, not an AI/ML algorithm. Ground truth for device performance is established through standardized physical, chemical, biological, and animal testing, not human expert interpretation of data.
    • For "Simulated Use" testing, it mentions "by a physician," indicating one or more physicians were involved in that specific assessment, but their number and specific qualifications are not detailed beyond "physician."

    4. Adjudication Method for the Test Set

    • Not Applicable. As this is not an AI/ML study, there is no need for adjudication of conflicting expert opinions on a dataset. Test results are objective measurements against defined acceptance criteria.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    • No. This type of study is specifically designed for evaluating the impact of AI on human reader performance, which is not relevant for the validation of a physical medical device like a catheter.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • Not Applicable. There is no AI algorithm being validated in this submission. The "device" is a physical catheter.

    7. The Type of Ground Truth Used

    • The "ground truth" for this device's performance is established by:
      • Standardized Test Methods: Adherence to established international (ISO, ASTM, USP) and industry standards for material properties, mechanical performance, biocompatibility, and sterility.
      • Physical Measurements: Objective measurements of dimensions, burst pressure, tensile strength, etc.
      • Chemical/Biological Assays: Results from biocompatibility tests (e.g., cytotoxicity, hemolysis, pyrogenicity, sensitization) and bacterial endotoxin tests.
      • Functional Demonstrations: Successful performance in simulated use environments and in an animal model (porcine model for angiographic evaluation, showing no vessel injury).
      • Predetermined Specifications: Comparison against established product specifications and the performance of the predicate device.

    8. The Sample Size for the Training Set

    • Not Applicable. This is not an AI/ML device, so there is no "training set."

    9. How the Ground Truth for the Training Set was Established

    • Not Applicable. As there is no AI/ML training set, this question is not relevant.