Search Results
Found 3 results
510(k) Data Aggregation
(207 days)
The AViTA Pulse Oximeter is intended for spot check monitoring of oxygen saturation of arterial hemoglobin (SpO2) and pulse rate for adults, adolescents and child in home and professional healthcare facility settings. It is designed for fingers between 0.8cm and 2.3cm (0.3 inches ~0.9 inches) and intended for use during no-motion condition. The device is prescription only
The AViTA Pulse Oximeter is a non-invasive, prescription-use medical device intended for spot-check monitoring measurement of functional oxygen saturation of arterial hemoglobin (SpO₂) and pulse rate in adult, adolescent, and child users in home and professional healthcare environment settings.
The device operates using spectrophotometry by emitting red and infrared light through a pulsatile vascular bed, such as a fingertip, and detecting the light transmitted through the tissue to determine the relative concentrations of oxygenated and deoxygenated hemoglobin.
The device is designed for use on fingers with a thickness range of approximately 0.8 cm to 2.3 cm (0.3 inches to 0.9 inches) and is intended for use under no-motion conditions. The device is intended for spot-check measurements only and is not intended for continuous monitoring.
N/A
Ask a specific question about this device
(161 days)
Vital Signs Monitor MD2000C is a portable device indicated for measuring physiological parameters, such as NIBP, SpO2, PR, and Pulse waveform of adult and three years old and older pediatric patients in hospitals, community hospitals and medical facilities. Vital Signs Monitor is intended for spot-checking and/or continuous monitoring of patients.
The proposed device Vital Signs Monitor MD2000C is a device powered by internal electrical power source and external electrical power source. The MD2000C is used in hospital,community hospitals and medical facilities. It can measure and record functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate (PR), Non-invasive measurement of blood pressure (NIBP) of adult and pediatric patients. Its accessories include SpO2 probe, NIBP cuff for adult, NIBP cuff for pediatric, and power adapter. It has data storage,display,alarm.
Here's a breakdown of the acceptance criteria and study information for the Vital Signs Monitor MD2000C based on the provided text:
Acceptance Criteria and Device Performance
The direct acceptance criteria for the Vital Signs Monitor MD2000C are not explicitly stated as a separate table within the document in terms of specific performance targets for NIBP, SpO2, and PR beyond accuracy ranges. Instead, the document focuses on demonstrating substantial equivalence to a predicate device (MD2000B and MD300M for SpO2, and Multi-parameters health Examination System for NIBP).
The performance data available relates to the MD300M (for SpO2) and Multi-parameters health Examination System (for NIBP), which the new device MD2000C claims to be identical to in those modules.
Table of Acceptance Criteria (Implied by Predicate Performance) and Reported Device Performance:
| Parameter | Implied Acceptance Criteria (from predicate) | Reported Device Performance (from predicate studies) |
|---|---|---|
| SpO2 Accuracy | ARMS of ≤ 2-3% (common for pulse oximeters) | MD300M: ARMS of 1.75% (over 70-100% range) |
| NIBP Accuracy | Mean error within ±5mmHg, Standard deviation within 8mmHg (per ISO 81060-2:2018 or similar) | Multi-parameters health Examination System: Mean error within ±5mmHg, Standard deviation within 8mmHg |
| SpO2 Measuring Range | 70%-100% | 70%-100% |
| SpO2 Resolution | 1% | 1% |
| PR Measuring Range | 30-235 bpm (predicate) | 30-250 bpm (proposed) |
| PR Resolution | 1 bpm | 1 bpm |
| PR Accuracy | 30-100 bpm: ±2bpm; 101-235 bpm: ±2% (predicate) | ±2bpm or ±2% (choose larger) (proposed) |
| NIBP Measurement Method | Oscillometric method | Oscillometric method |
| Static pressure measurement range | 0~270mmHg (predicate) | 20~280mmHg (proposed) |
| Maximum static pressure error | ±3mmHg | ±3mmHg |
Note: The "Implied Acceptance Criteria" are based on the performance of the predicate devices which the new device is seeking equivalence to. The direct acceptance criteria for the specific device are not explicitly listed in a regulatory numerical format but are demonstrated through equivalence to these predicate performances.
Study Information:
2. Sample sizes used for the test set and the data provenance:
- SpO2 Accuracy Study (MD300M):
- Sample Size: 12 healthy adult volunteer subjects (6 females, 6 males)
- Demographics: Ages 21-43 years, 47-82 kg, 155-185 cm, with a representative range of pigmentation.
- Data Provenance: Conducted in Yue Bei People's Hospital. Prospective study. Country of origin not explicitly stated but implied to be China (as the manufacturer is Beijing Choice Electronic Technology Co., Ltd.)
- NIBP Accuracy Study (Multi-parameters health Examination System):
- Sample Size: 85 subjects (35 child subjects aged 3-12 years, 50 adult subjects)
- Data Provenance: Conducted in PLA Rocket Force Characteristic Medical Center. Prospective study. Country of origin not explicitly stated but implied to be China.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- The document does not specify the number or qualifications of experts used to establish ground truth for either the SpO2 or NIBP studies.
- For SpO2, ground truth is typically established using a co-oximeter in controlled desaturation studies (e.g., fractional oxyhemoglobin measurement compared to pulse oximeter readings).
- For NIBP, ground truth is typically established simultaneously with a reference auscultatory or invasive blood pressure measurement taken by qualified personnel.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- The document does not provide details on any adjudication methods used for the test sets. The studies appear to be direct performance comparisons against a reference standard rather than assessments requiring multiple human interpretations.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No MRMC study was performed. This device is a vital signs monitor, not an AI-assisted diagnostic imaging or interpretation tool for human readers. It provides physiological parameters directly.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- Yes, standalone performance was evaluated. The studies described for SpO2 and NIBP accuracy directly measure the device's performance against a reference standard, without human interpretation of the device's output influencing the accuracy metrics. The device itself is designed to provide direct physiological measurements.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- SpO2: The ground truth for SpO2 accuracy studies is typically established by fractional oxyhemoglobin measurement using a co-oximeter during controlled hypoxia conditions. This is considered the reference standard for pulse oximetry.
- NIBP: The ground truth for NIBP accuracy studies is established by simultaneous measurements using a reference auscultatory method (e.g., mercury sphygmomanometer with stethoscope by trained observer) or an invasive arterial line, compared to the device's oscillometric readings.
8. The sample size for the training set:
- The document does not specify sample sizes for training sets. It describes the device's performance through validation studies (test sets) for specific modules that were adopted from previously cleared devices. It's implied that the algorithms within these modules were developed and potentially trained using proprietary datasets not detailed in this 510(k) summary.
9. How the ground truth for the training set was established:
- The document does not specify how ground truth was established for any potential training sets. This information would typically be part of the internal development process for the SpO2 and NIBP algorithms, which predates this 510(k) application (as the modules are reused from previously cleared devices).
Ask a specific question about this device
(209 days)
The Cascade IOMAX™ Intraoperative Monitor with Surgical Studio software (IOMAX) is an electroneurodiagnostic device that acquires, displays and stores physiologic data from peripheral sensory and motor nerves, muscles and the central nervous system, generated either spontaneously or elicited stimuli. The acquired data are necessary to perform somatosensory. auditory and visual evoked potentials (EPs), electroencephalography (EEG), electromyography (EMG), transcranial motor evoked potentials (TcMEPs), direct cortical stimulation, nerve conduction studies and Train of Four (TOF) analysis. SpO2 measures and displays oxygen saturation and heart rate information. The system also delivers direct nerve stimulation required for specific surgical procedures.
Evoked Potentials (EPs): IOMAX provides electrical, auditory or visual stimulation and measures, and stores the electrical activity of the nervous system in response to the stimulation.
EEG: IOMAX measures, displays, records, and stores electrical activity of the brain from two or more electrodes on the head.
Free Run EMG: IOMAX acquires, displays, records, and stores spontaneous EMG activity of motor nerves by continually displaying a live stream of mechanically induced myotome contractions.
TcMEP: IOMAX delivers transcranial stimulation via dedicated outputs for intraoperative assessment. Cortical Stimulation: IOMAX delivers Low Current Stimulation (LCS) during surgical procedures to map various areas of the cortex.
Triggered EMG (TEMG): IOMAX electrically stimulates the motor nerves, and displays, records, and stores the resulting compound muscle action potentials in the innervated muscle.
Nerve Conduction Study (NCS): IOMAX measures, displays, records, and stores sensory and motor nerve conduction time (latency) by applying a stimulus to peripheral nerves, the spinal cord, and the central nervous system.
Train of Four (TOF) or Twitch Test: IOMAX delivers a train of four pulses and measures, and stores the compound muscle action potential amplitude fade for analysis.
SpO2: IOMAX measures and displays oxygen saturation and heart rate information. Remote Reader: IOMAX provides passive, real time remote review of intraoperative monitoring for a physician outside of the operating room.
IOMAX is used by or under the direction of a licensed physician, surgeon, or neurologist in a professional healthcare facility environment for pre-operative, intraoperative and post-operative testing.
The IOMAX is a multimodality intraoperative neuromonitoring (IONM) system. It consists of Cadwell custom hardware, a standard laptop or desktop personal computer (PC) running a standard off-the-shelf (OTS) operating system (OS), and Cadwell custom software.
The modalities recorded, measured and displayed by the IOMAX are:
- . Evoked potential (EP) in the form of:
- Brainstem auditory (BAEP): O
- Visual (VEP); and O
- Somatosensory (SSEP). O
- Transcranial electrical motor evoked potential (TcMEP).
- Electromyography (EMG). ●
- Triggered EMG.
- Electroencephalogram (EEG). ●
- Nerve conduction studies.
- Train of four (TOF). ●
- SpO2 and heart rate values. ●
- Threshold mode. ●
- Cortical stimulation.
Here's an analysis of the provided text regarding the acceptance criteria and supporting studies for the Cadwell IOMAX Intraoperative Monitor:
Acceptance Criteria and Reported Device Performance
The provided document describes the Cadwell IOMAX Intraoperative Monitor, an electroneurodiagnostic device. The acceptance criteria are broadly focused on the device's technical performance, safety, and compliance with various medical device standards. The document does not present specific quantitative acceptance criteria or reported performance values in a distinct table format. Instead, it states that "Test results indicate that the IOMAX complies with its predetermined specifications and the applicable standards." and "Clinical results indicate that the IOMAX complies with the applicable requirements of the standard."
However, we can infer the categories of acceptance criteria based on the performance testing summary. The reported device performance is generally a statement of compliance.
| Acceptance Criteria Category | Reported Device Performance |
|---|---|
| Biocompatibility | Accessories (SpO2 sensor) verified to be biocompatible in accordance with ISO 10993-1:2009. The core IOMAX device has no patient contact materials. |
| Software | Software designed and developed according to a robust software development process, rigorously verified and validated. Complies with predetermined specifications and applicable FDA guidance documents and IEC 62304:2006. |
| Electrical Safety | Complies with ANSI/AAMI ES60601-1:2005 and IEC 60529:1989 (and amendments/corrigenda). |
| Electromagnetic Compatibility | Complies with IEC 60601-1-2:2014. |
| Performance Testing – Bench | Complies with internal requirements and standards including IEC 60068-2-27, IEC 60068-2-64, IEC 60601-2-26, IEC 60601-2-40, IEC 60601-1-6, IEC 62366, and ISO 80601-2-61. |
| Performance Testing – Clinical | Complies with the applicable requirements of ISO 80601-2-61:2011 for pulse oximeter equipment. |
Study Details:
The provided document summarizes various tests rather than detailing a single "study" as one might expect for a clinical trial. However, it does outline the testing performed to demonstrate compliance.
-
Sample size used for the test set and the data provenance:
- Biocompatibility: Not explicitly stated as a "sample size" for a test set in the clinical sense. The verification was on "component materials" of accessories. Data provenance is implied to be laboratory testing of materials.
- Software, Electrical Safety, EMC, Bench Performance: Not applicable in terms of patient sample size. These involve engineering and laboratory testing of the device itself.
- Clinical Performance (SpO2): The document states "Clinical results indicate that the IOMAX complies with the applicable requirements of the standard [ISO 80601-2-61:2011, Medical electrical equipment – Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment]". This standard defines requirements for pulse oximeters, which would necessitate testing on human subjects. However, the specific sample size for this clinical testing is not provided in the given text. The data provenance is clinical testing against a standard. The country of origin is not specified.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- This information is not provided in the document. The testing described is primarily technical compliance with standards, and clinical performance for SpO2 which refers to a standard directly. There is no mention of independent expert review or ground truth establishment in the traditional sense for diagnostic accuracy.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- This information is not provided and is unlikely to be applicable for the types of compliance testing described here. Adjudication methods are typically used for medical image interpretation or diagnostic accuracy studies involving human readers.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No MRMC comparative effectiveness study was done or mentioned. This device is an intraoperative monitor and stimulator; it is not an AI-powered diagnostic imaging tool that would typically involve human readers interpreting cases with or without AI assistance. Therefore, no effect size of human improvement with AI assistance is applicable or provided.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- The IOMAX is an electroneurodiagnostic device that acquires and displays physiological data. Its "performance" is based on the accuracy and reliability of its measurements and stimulation. The testing described (electrical safety, EMC, bench, and clinical for SpO2) assesses the device's inherent function, which is essentially "standalone performance" in the sense that the device directly performs the measurement or stimulation. It doesn't involve an "algorithm" in the AI sense, but rather the hardware and software's ability to accurately capture and present physiological signals or deliver specific stimuli.
- The clinical performance for SpO2 (compliance to ISO 80601-2-61) is a test of the device's standalone accuracy in measuring oxygen saturation and heart rate.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- For the technical compliance tests (biocompatibility, software, electrical safety, EMC, bench performance), the "ground truth" is defined by the requirements of the specified industry standards (e.g., ISO 10993-1, ANSI/AAMI ES60601-1, IEC 60601-1-2, IEC 60068, IEC 60601-2-26, IEC 60601-2-40, IEC 60601-1-6, IEC 62366). The device's output is compared against the expected performance defined by these standards.
- For the clinical performance testing for SpO2, the ground truth would be established by a reference method or device as required by ISO 80601-2-61, which typically involves comparing the pulse oximeter's readings against arterial blood gas analysis results.
-
The sample size for the training set:
- This information is not applicable as the device is not described as an AI/machine learning device that requires a training set in the conventional sense. The "training set" for software development would be the requirements and specifications used during its creation, rather than a dataset for statistical model training.
-
How the ground truth for the training set was established:
- As the concept of a "training set" for an AI model is not applicable here, the establishment of ground truth for such a set is also not applicable. The software's "ground truth" during development would be its specified functional requirements.
Ask a specific question about this device
Page 1 of 1