Found 4 results

510(k) Data Aggregation

    K Number
    K252266

    Validate with FDA (Live)

    Date Cleared
    2025-10-17

    (88 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    All
    Predicate For
    N/A
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    S.I.N. Dental Implant System is intended for placement in the maxillary or mandibular arch to provide support for single-unit or multi-unit restorations. When a one-stage surgical approach is applied, the S.I.N. Dental Implant System may be immediately loaded when good primary stability is achieved and the functional load is appropriate.

    All digitally-designed custom abutments for use with Interface CAD-CAM abutments are to be sent to a S.I.N.-validated milling center for manufacture.

    Device Description

    The purpose of this submission is to add components to the S.I.N. Dental Implant System, which includes components previously cleared K200992, K193096, K170398 and K170392.

    This submission adds Interface Abutments and Healing Abutments to the S.I.N. Dental Implant System. The subject device Interface abutments are provided with MT 16º and CM11.5º connection compatible with the S.I.N. Dental implants.

    Subject Device Interface Abutments

    Subject device Interface Abutments are two-piece abutments for which the second part (or top half) is the ceramic superstructure. The second part is designed and manufactured using CAD-CAM techniques. The titanium component of the Interface Abutments has a cementable platform diameter of 3.5mm, 4.0mm, 4.5mm, or 5.5mm, cementable post-height of 4.0mm or 6.0mm, and a gingival height built into the titanium base component from 0.5mm to 4.0 mm. The two pieces of the Ti-Base are bonded using Panavia Universal Dual-Cure Cement (K150704), as referenced in K190936.

    All subject devices are provided only in straight designs, with no pre-angled conventional abutments or pre-angled CAD-CAM abutments. No subject device abutment is to be customized to create an angled abutment or to correct for angulation, and no superstructure for use with any subject device Interface abutment (CAD-CAM) is to be fabricated to create a final abutment with angulation or to correct for angulation. Therefore, this model has a slight design difference to improve screw insertion and removal. It was purposely designed to optimize fixation and removal of the screw connection. This new presentation allows screw fixation even in cases of anterior aesthetic rehabilitation, avoiding the need to create screw access holes in the sublingual phase. However, it is confirmed that all implant-abutment interfaces are intended exclusively for straight configurations and are not indicated for any angled model. These minor differences in the exact dimensions of the subject device interface abutment as compared to the predicate device and reference devices do not impact safety, effectiveness, or substantial equivalence.

    The subject device interface abutments with the CM11.5º interface are compatible with implants cleared in K200992, K221453, K211921 and K170392 in terms of abutment-implant connection. The subject device with the MT16º interface is compatible with implants cleared in K222231, K193096 and K170392 in terms of abutment-implant connection.

    The design parameters for the CAD-CAM fabrication of patient-specific abutments from CADCAM Abutments are the same for the 11.5° CM implant connection and the MT 16° implant connection. The design limit parameters are:

    • Minimum wall thickness – 0.5 mm
    • Minimum abutment post height for single-unit restoration – 4.0 mm
    • Maximum angle – 0°. Straight only
    • Maximum angle for screw insertion/removal – 20º
    • Maximum gingival height – 5.0mm
    • Minimum Gingival Height – 0.5 mm
    • Maximum allowable abutment Post Height – 6 mm
    • Total Abutment Height – 10mm

    By definition, the abutment post height is considered by FDA to be the "stump" portion above the gingival collar, to which the restorations attach.

    All subject device Interface Abutments are manufactured from titanium alloy (Ti-6Al-4V) conforming to ASTM F136 and the same zirconia that is specified for superstructures in K190396, and are to be used with subject abutment screws. All subject devices are provided sterile. All Interface abutments are provided sterile by gamma irradiation; the same sterilization method used in predicate device K190396 and in the additional devices K200992 and K170392.

    All Interface abutments manufactured form titanium alloy are anodized using a standard anodization process is identical to the anodization process used on abutments cleared in K190396.

    The subject device abutments screws are to be used with Interface Abutments present in this submission and they are substantially equivalent to material (titanium alloy) and designs to abutments screws cleared in K170392.

    Subject Device Healing Caps

    The subject device Healing Cap is substantially equivalent to the corresponding design Healing Cap cleared in K170392. The subject device healing cap is designed for the healing period between implant placement and final abutment placement and is not intended to support a temporary prosthesis.

    All Healing Caps are provided sterile by gamma irradiation; the same sterilization method used in predicate device K170392 and additional predicate device K200992.

    All subject device components are provided sterile by gamma irradiation, the same sterilization method used in K190936, K170392 and K200992.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K251497

    Validate with FDA (Live)

    Date Cleared
    2025-10-16

    (154 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K151455, K011394, K150704, K201860

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Elos Accurate® Hybrid Base™ is intended for attaching to dental implants in order to provide basis for single or multiple tooth prosthetic restorations. The Hybrid Base™ is used as an interface between a dental implant and a zirconia superstructure and will be attached to the implant using a prosthetic screw and attached to the zirconia superstructure by cementing.

    The Elos Accurate® Hybrid Base™ is compatible with the implant systems listed in table 1:

    Table 1:

    Implant Platform compatibilityPlatform diameter [mm]Implant Body diameter [mm]
    Nuventus NV.C NPØ3.5Ø3.5 / Ø4.3
    Nuventus NV.C RPØ5.0Ø5.0

    The zirconia superstructures for use with the Elos Accurate® Hybrid Base™ are either intended to be sent and manufactured at a FDA registered Elos Medtech approved milling facility or to be designed and manufactured according to digital dentistry workflow. The workflow system integrates multiple components of the digital dentistry workflow: scan files from Intra-Oral Scanners, CAD software, CAM software, ceramic material milling machine and associated tooling and accessories.

    Device Description

    The Elos Accurate® Hybrid Base™ is intended for attaching to dental implants in order to provide basis for single or multiple tooth prosthetic restorations. The Hybrid Base™ functions as an interface between a dental implant and a zirconia superstructure. It attaches to the implant using the included prosthetic screw and connects to the zirconia superstructure through cementing. The Elos Accurate® Hybrid Base™ is a two-piece abutment composed of the Hybrid Base as the bottom-half and the zirconia superstructure as the top-half, which when assembled comprises the final finished medical device.

    The Elos Accurate® Hybrid Base™ is available for two types: Engaging and Non-Engaging

    • The Hybrid Base™ Engaging which is intended for single tooth dental restorations and having an indexing feature, which avoid the Hybrid Base from rotating in the implant.
    • The Hybrid Base™ Non-Engaging is intended for multiple tooth dental restorations and has no indexing feature, which allows the Hybrid Base to rotate in the implant.

    The Elos Accurate® Hybrid Base™ is available with 4 different collar heights and is provided with a 7.5mm chimney which can be cut down to several heights by the user, to minimum 3.5 mm.

    The Elos Accurate® Hybrid Base™ consists of a pre-manufactured prosthetic component in Titanium alloy per ASTM F136, as well as supporting digital library file for 510(k) cleared design software (3Shape Abutment Designer™ Software, K151455) which facilitates the design of a patient specific zirconia superstructure by the laboratory/clinician. The Elos Accurate® Hybrid Base™ fits directly to an endosseous dental implant. The laboratory designed superstructure is manufactured from 510(k) cleared Zirconia (Lava Plus, K011394) according to digital dentistry workflow. For all Elos Accurate® Hybrid Base™ models the zirconia superstructure must be designed according to following limits:

    Hybrid Base abutments (zirconia part):
    Min. wall thickness 0.5 mm
    Gingival height min. 0.5mm or max. 5 mm
    Max. angulation 20°
    Min. post height* 4 mm

    *The post height is defined as the cementable height of the abutment.

    The laboratory designed superstructure is attached to the Elos Accurate® Hybrid Base by use of 510(k) cleared cement (Panavia V5, K150704) and the final prosthetic restoration is attached to the implant using a Prosthetic screw.

    The Prosthetic Screw, also made of Titanium Alloy (ASTM F136), provided for the Elos Accurate® Hybrid Base™ is used to secure the final prosthetic restoration to the implant in the patient's mouth. The Prosthetic Screws have a hexalobular driver connection interface. The driver intended to be used with the Elos Accurate® Hexalobular Prosthetic Screws is the Elos Prosthetic Screwdriver 18mm, 26mm or 34mm (Ref. No. 042.0018, 042.0026 or 042.0034). These screwdrivers are Class I Exempt devices per FDA product code NDP.

    The Elos Accurate® Hybrid Base has a gold anodized surface to increase the esthetics of the dental restoration - the same surface as in predicate device K242025.

    The subject prosthetic screws are provided are either non-coated or surface coated with DLC (identical to the DLC coating in reference device K201860).

    The Elos Accurate® Hybrid Base™ and the Prosthetic screw is delivered non-sterile, and the final restoration and corresponding screw is intended to be sterilized at the dental clinic before it is placed in the patient. The recommended sterilization procedure is full cycle pre-vacuum steam sterilization at a temperature of 132 °C (270°F) for 4 minutes. Dry time: 20 minutes.

    The Elos Accurate® Hybrid Base™ is packed in 1 pc. packaging with the Elos Accurate® Hexalobular Prosthetic screw. The Screw is also available packed alone in 1 pcs packaging.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K250967

    Validate with FDA (Live)

    Device Name
    Sherlock
    Manufacturer
    Date Cleared
    2025-06-26

    (87 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K212664, K193335, K163194, K180536, K201225, K150704

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Sherlock abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for single-unit or multi-unit prosthetic restorations.

    All digitally designed CAD/CAM customizations for Sherlock abutments are to be sent to an Open Implants-validated milling center for manufacture.

    Sherlock abutments are compatible with the implant systems listed in the Compatibility Table:

    Compatibility Table

    Compatible Implant SystemsImplant Body Diameter (mm)Implant Platform Diameter (mm)
    Neodent Helix GM®, Drive GM®, Titamax GM®3.5, 3.75, 4.0, 4.3, 5.0, 6.0, 7.03.0
    Device Description

    Sherlock is a dental implant abutment system that is being expanded to include a new compatible implant system, Neodent GM®. The Subject device implant platform diameter is 3.0 mm, and the corresponding compatible implant body diameters range from 3.5 mm to 7.0 mm.

    The abutment designs are Titanium Base, Titanium Blank, Straight Multi-Unit, Multi-Unit Angled 17°, and Multi-Unit Angled 30° Abutments. These abutment designs were previously cleared in the sponsor's K220482 Predicate device and K212664 Reference device submissions. All abutment designs are provided with corresponding abutment screws.

    The Subject device Titanium Base abutments are intended to be used as a two-piece abutment composed of the base bottom-portion (prefabricated titanium base component) with a cemented/bonded CAD-CAM fabricated zirconia top-portion (superstructure) where the final two-piece abutment (base component and cemented superstructure) is the finished device used for the prosthetic restoration. Each patient-specific zirconia superstructure is individually prescribed by the clinician and manufactured by an authorized milling center.

    All Subject device prefabricated titanium base components are provided in a straight design with no angulation in the titanium base post. They are provided with either an indexed/engaging implant connection for crowns or a non-engaging/nonindexed implant connections for bridges. The standard prefabricated titanium base components are provided in gingival heights ranging from 0.8 mm to 3.0 mm and abutment post lengths of 8 mm or 10 mm. The ASC prefabricated titanium base components are provided in gingival heights ranging 0.8 mm to 2.5 mm and abutment post length of 8 mm. Additional gingival height may be provided for both abutment designs in the zirconia superstructure. ASC prefabricated titanium base components are provided with a cutout in the prosthetic post to accommodate a restoration with an angled screw channel when clinically necessary. Standard prefabricated titanium base components and ASC prefabricated titanium base components posts may be reduced to 4 mm to accommodate individual patient occlusion. The zirconia mesostructure may contain an angled post within the established design parameters.

    The overall design parameters for the two-part Standard and ASC CAD/CAM prefabricated titanium base components with zirconia mesostructure are:

    • Minimum Zirconia Wall Thickness – 0.5 mm
    • Minimum Post Height for single-unit abutment* – 4.0 mm
    • Minimum Overall Gingival Height – 0.8 mm (titanium base plus zirconia)
    • Maximum Overall Gingival Height – 5 mm
    • Maximum Correction Angle – 30°

    The required cement for bonding the zirconia superstructure to the Subject device Titanium Bases to create the final two-piece abutment is Kuraray Noritake Dental PANAVIA™ V5 cleared in K150704.

    Titanium Blank abutments, sometimes referred to as "Pre-mill" or "Ti-Blank" abutments are one-part abutments intended for use in a CAD/CAM workflow. Each Subject device Titanium Blank implant abutment has a pre-manufactured indexed implant connection interface with a cylindrical customization section and a milling retention geometry section. The retention geometry holds the component in a milling machine fixture while the patient-specific portion above the implant interface is milled in a dental milling machine. All patient-specific Titanium Blank abutment fabrication is by prescription on the order of the clinician.

    The overall design parameters for the Titanium Blank customized abutments are:

    • Minimum Wall Thickness – 0.75 mm
    • Minimum Post Height for single-unit abutment* – 4.0 mm
    • Minimum Overall Gingival Height – 0.8 mm
    • Maximum Overall Gingival Height – 5 mm
    • Maximum Correction Angle – 30°

    All digitally designed zirconia mesostructures for use with the Subject device titanium base abutments and digitally designed Subject device titanium blank abutments will be fabricated at an Open Implants validated milling center under FDA quality system regulations.

    Multi-Unit Abutments (MUAs) are intended for use with multi-unit restorations. They are considered two-part abutments. The base portion of the MUA is connected directly to the implant either with an integral screw (straight MUA) or with a separate multi-unit abutment screw (angulated abutments). Straight MUAs have a non-indexed connection with the dental implant. The angulated MUAs have an indexed connection with the dental implant. The second part of the MUA is a mating coping which is retained with a prosthetic screw. Multi-Unit Abutments are available in Straight, 17° Angulated and 30° Angulated configurations. The coping and prosthetic screw is compatible with each MUA design/configuration.

    All Subject device abutments and corresponding abutment screws are pre-manufactured from Ti-6Al-4V ELI (Grade 23) titanium conforming to ASTM F136, Standard Specification for Wrought Titanium-6 Aluminum-4 Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications (UNS R56401) and are provided non-sterile to the user. The mesostructure/copings for Titanium base abutments are fabricated from zirconia conforming to ISO 13356, Implants for surgery — Ceramic materials based on yttria-stabilized tetragonal zirconia (Y-TZP).

    AI/ML Overview

    The provided FDA 510(k) Clearance Letter for Sherlock Dental Implant Abutment does not contain the information requested regarding acceptance criteria and a study proving the device meets those criteria.

    This document describes the device's indications for use, its substantial equivalence to predicate devices, and the non-clinical data reviewed by the FDA for clearance. However, it does not detail specific acceptance criteria or an associated study demonstrating compliance with those criteria.

    The "Performance Data" section lists several types of non-clinical data submitted, such as static and fatigue testing according to ISO 14801, a reverse engineering study, biocompatibility testing, cleaning and sterilization validation, and an MRI review. These are the studies performed to support the substantial equivalence claim, but the FDA letter does not specify quantitative acceptance criteria for these tests nor does it report specific performance outcomes that directly address such criteria.

    Therefore, I cannot populate the requested table or provide details for points 1-9 based solely on the provided text. The document focuses on demonstrating substantial equivalence to previously cleared devices rather than presenting a performance study against predefined acceptance criteria.

    Ask a Question

    Ask a specific question about this device

    K Number
    K242025

    Validate with FDA (Live)

    Date Cleared
    2024-12-10

    (152 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K151455, K011394, K150704

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Elos Accurate® Hybrid Base™ is intended for attaching to dental implants in order to provide basis for single or multiple tooth prosthetic restorations. The Hybrid Base™ is used as an interface between a dental implant and a zirconia superstructure and will be attached to the implant using a prosthetic screw and attached to the zirconia superstructure by cementing.

    The Elos Accurate® Hybrid Base™ is compatible with the implant systems listed in table 1: Table 1.

    Implant Platform compatibilityPlatform diameter [mm]Implant Body diameter [mm]
    Astra Tech EV 3.0Ø3Ø3
    Astra Tech EV 3.6Ø3.6Ø3.6
    Astra Tech EV 4.2Ø4.2Ø3.6 & Ø4.2
    Astra Tech EV 4.8Ø4.8Ø4.2 & Ø4.8
    Astra Tech EV 5.4Ø5.4Ø5.4

    The zirconia superstructures for use with the Elos Accurate® Hybrid Base™ are either intended to be sent and manufactured at a FDA registered Elos Medtech approved milling facility or to be designed and manufactured according to digital dentistry workflow system integrates multiple components of the digital dentistry workflow: scan files from Intra-Oral Scanners, CAD software. CAM software, ceramic material, milling machine and associated tooling and accessories,

    Device Description

    The Elos Accurate® Hybrid Base™ is intended for attaching to dental implants in order to provide basis for single or multiple tooth prosthetic restorations. The Hybrid Base™ functions as an interface between a dental implant and a zirconia superstructure. It attaches to the implant using the included prosthetic screw and connects to the zirconia superstructure through cementing. The Elos Accurate® Hybrid Base™ is a two-piece abutment composed of the Hybrid Base as the bottomhalf and the zirconia superstructure as the top-half, which when assembled comprises the final finished medical device.

    The Elos Accurate® Hybrid Base™ is available for two types: Engaging and Non-Engaging.

    • The Hybrid Base™ Engaging which is intended for single tooth dental restorations and having an indexing feature. which avoid the Hybrid Base from rotating in the implant.
    • -The Hybrid Base™ Non-Engaging is intended for multiple tooth dental restorations and has no indexing feature, which allows the Hybrid Base to rotate in the implant.

    The Elos Accurate® Hybrid Base™ is available with 4 different collar heights and is provided with a 7,5mm chimney which can be cut down to several heights by the user, to minimum 3,5 mm.

    The Elos Accurate® Hybrid Base™ consists of a pre-manufactured prosthetic component in Titanium alloy per ASTM F136, as well as supporting digital library file for 510(k) cleared design software (3Shape Abutment Designer™ Software, K151455) which facilitates the design of a patient specific zirconia superstructure by the laboratory/clinician. The Elos Accurate® Hybrid Base™ fits directly to an endosseous dental implant. The laboratory designed superstructure is manufactured from 510(k) cleared Zirconia (Lava Plus, K011394) according to digital dentistry workflow. For all Elos Accurate® Hybrid Base™ models the zirconia superstructure must be designed according to following limits:

    Hybrid Base abutments (zirconia part):
    Min. wall thickness 0.5 mm
    Gingival height min. 0.5mm or max. 5 mm
    Max. angulation 20°.
    Min. post height* 4 mm

    *The post height is defined as the cementable height of the abutment.

    The laboratory designed superstructure is attached to the Elos Accurate® Hybrid Base by use of 510(k) cleared cement (Panavia V5, K150704) and the final prosthetic restoration is attached to the implant using a Prosthetic screw.

    The Prosthetic Screw, also made of Titanium Alloy (ASTM F136), provided for the Elos Accurate® Hybrid Base™ is used to secure the final prosthetic restoration to the implant in the patient's mouth. The Prosthetic Screws have a hexalobular driver connection interface. The intended driver to be used with the Elos Accurate® Hexalobular Prosthetic Screws is the Elos Accurate® Prosthetic Screwdriver 18mm, 26mm or 34mm (Ref. No. PS-AH18-1, PS-AH26-1 or PS-AH34-1). These screwdrivers are Class I Exempt devices per FDA product code NDP.

    The Elos Accurate® Hybrid Base has a gold anodized surface to increase the esthetics of the dental restoration - the same surface as in predicate device K230317.

    The subject prosthetic screws are provided anodized identical to reference device K120414.

    The Elos Accurate® Hybrid Base™ and the Prosthetic screw is delivered non-sterile and the final restoration and corresponding screw is intended to be sterilized at the dental clinic before it is placed in the patient. The recommended sterilization procedure is full cycle pre-vacuum steam sterilization at a temperature of 132 °C (270°F) for 4 minutes. Dry time: 20 minutes.

    The Elos Accurate® Hybrid Base™ is sold in 1 pc. packaging and the Elos Accurate® Prosthetic screw is sold in 1 pc. packaging.

    AI/ML Overview

    The provided text does not describe an AI medical device. It is a 510(k) summary for a dental implant component, the Elos Accurate® Hybrid Base™. Therefore, it is impossible to extract the requested information about acceptance criteria and a study proving an AI device meets those criteria.

    The document focuses on demonstrating the substantial equivalence of the Elos Accurate® Hybrid Base™ to a predicate device, primarily through:

    • Similar Indications for Use and Intended Use: The device serves the same purpose as existing products.
    • Comparable Design and Materials: It uses similar components (titanium alloy, zirconia) and design principles.
    • Non-Clinical Testing: Fatigue testing (meeting ISO 14801), engineering and dimensional analysis, sterilization validation, digital dentistry workflow validation, and biocompatibility testing were performed. The fatigue testing is the closest parallel to performance testing, but it's for mechanical durability rather than diagnostic or analytical accuracy.

    There is no mention of an algorithm, AI, machine learning, or any form of software that provides diagnostic or analytical outputs. Consequently, none of the specific questions about AI device performance metrics, training sets, ground truth establishment, or multi-reader studies can be answered from this document.