    K Number
    K251262
    Date Cleared
    2025-09-26

    (156 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K231127, K193096, K213672

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    S.I.N. Dental Implant System is intended for placement in the maxillary or mandibular arch to provide support for single-unit or multi-unit restorations. When a one-stage surgical approach is applied, the S.I.N. Dental Implant System is intended for immediate loading when good primary stability is achieved and with appropriate occlusal loading.

    All digitally-designed custom abutments for use with Interface CAD-CAM abutments are to be sent to a S.I.N.- validated milling center for manufacture.

    The Abutment Mini Angled Morse 450 is intended for use with the S.I.N. Dental Implant System Zygomatic implants. S.I.N. Dental Implant System Zygomatic implants are intended for placement in the maxillary arch to provide support for fixed or removable dental prostheses in patients with partially or fully edentulous maxillae. When a one-stage surgical approach is applied, the S.I.N. Dental Implant System Zygomatic implants are intended for immediate loading when good primary stability is achieved and with appropriate occlusal loading.

    Device Description

    The purpose of this submission is to add components to the S.I.N. Dental Implant System, which includes components cleared previously in K222231, K231127, and K170392.

    This submission includes dental implants Epikut S with a Morse taper (MT) abutment interface and an acid-etched endosseous surface, and Epikut S Plus implants with an endosseous surface produced by acid- etching followed by application of a hydroxyapatite coating (HAnano) and the same described connection. The implant design and endosseous surfaces are nearly identical to the Epikut S and Epikut S Plus implants cleared in K222231, with the exception of the additional body/platform diameter (6.0 /5.2 mm) with lengths of 8.5, 10, 11.5, 13.

    The subject device dental implants are summarized in the following table.

    Implant LinesBody Ø, mmPlatform Ø, mmLengths, mm
    Epikut S6.05.28.5, 10, 11.5, 13
    Epikut S Plus6.05.28.5, 10, 11.5, 13

    This submission includes Abutment Mini Angled Morse 45° in sizes from 1.5, 2.0, 2.5 with a 16° Morse taper connection mating abutments for screw retained, multi-unit prostheses and a prosthetic platform diameter of 4.8 mm, that is for exclusively use with cleared Zygomatic Plus implant. The corresponding abutment screw is PFMAAM 16A.

    This submission includes Interface abutment MT 16° prosthetic components for fabrication of patient-specific abutments that are compatible with implants from S.I.N. Dental Implant System. Subject device Interface Abutments are two-piece abutments for which the second part (or top half) is the ceramic superstructure. The second part is designed and manufactured using CAD-CAM techniques. The titanium component of the Interface Abutments Morse Taper has a cementable platform diameter of 3.5mm, 4.5mm, or 5.5mm, cementable post-height of 4.0mm or 6.0mm, and a gingival height built into the titanium base component from 0.5mm to 4.0mm. The two pieces of the Ti-Base are bonded using Panavia Universal Dual-Cure Cement (K150704), as referenced in K190936. These abutments are used with a fixation screw PFMT 02A.

    After scanning the intraoral area, the S.I.N.-validated milling center will receive the STL file and design the coping/superstructure using computer software (CAD) according to the design parameters below and the space available.

    Design parameters for all Interface CAD-CAM Abutments superstructure are:

    • Minimum wall thickness – 0.5 mm
    • Minimum abutment post height for single-unit restoration – 4.0 mm
    • Maximum angle – 0°. Straight only
    • Maximum gingival height – 5.0 mm
    • Minimum Gingival height – 0 mm
    • Maximum allowable abutment Post Height – 6 mm
    • Total abutment height – 10mm
    • Abutment diameter – depends on available space (for the patient)

    By definition, the abutment post height is considered by FDA to be the "stump" portion above the gingival collar, to which the restorations attach.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K232821
    Date Cleared
    2024-05-31

    (261 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K193096, K200992, K211921, K221453, K222231

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    S.I.N. Dental Implant System is intended for placement in the maxillary or mandibular arch to provide support for single-unit or multi-unit restorations. When a one-stage surgical approach is applied, the S.I.N. Dental Implant System is intended for immediate loading when good primary stability is achieved and with appropriate occlusal loading.

    Unitite Slim, Unitite, Unitite Compact, Strong SW CM S.I.N Dental Implant System implants with lengths less than 7 mm are intended for delayed loading only.

    Epikut CM, Epikut Plus CM, Epikut S, Epikut S Plus S.I.N. Dental Implant System implants with lengths of 18, 20, 22, or 24 mm may be tilted up to 30°. When used in the mandible or maxilla with implants with lengths of 18, 20, 22, or 24 mm at an angulation of 30°, a minimum of four implants must be used and must be splinted. When placed in the maxilla with lengths of 18, 20, 22, or 24 mm at angulations between 0° and less than 30°, the S.I.N. Dental Implants are only indicated for multiple unit restorations in splinted applications that utilize at least two implants.

    Epikut CM, Epikut Plus CM, Strong SW CM Plus, Strong SW HE Plus, Strong SW HI Plus All digitally-designed custom abutments for use with Interface CAD-CAM abutments are to be sent to a S.I.N.-validated milling center for manufacture.

    Device Description

    The purpose of this submission is to add labeling claims for a surface treatment designated HA™®, which is applied to implants of the S.I.N. Dental Implant System, as previously cleared in K170392. K193096. K200992, K211921, K221453, and K222231. The new labeling claims and the information supporting the claims are shown below in Performance Data.

    The following information is a summary of the designs of the subject device implants, as they have been previously cleared with the HA"" surface treatment. Note that there are no changes to any aspect of the design, nor to the HA" surface treatment, as described in the respective 510(k) Premarket Notifications referenced above.

    The previously cleared implants that are the subject of this submission includes eleven (11) product lines from six (6) 510(k)s. Implant body diameters, platform diameters and lengths for each product line are shown below in Table 10.1 Summary of Subject Device Designs and Sizes.

    K170392: Unitite Slim, Unitite, Unitite Compact K193096: Strong SW CM Plus, Strong SW HE Plus, Strong SW HI Plus K200992: Strong SW Plus K211921: Epikut Plus CM, Epikut Plus HE K221453: Epikut Plus CM (additional lengths) K22231: Epikut S Plus

    All subject device dental implants are manufactured from unalloyed titanium conforming to ASTM F67. The HA1000 surface treatment is identical to that cleared in K211921.

    AI/ML Overview

    The provided text is a 510(k) Premarket Notification from the U.S. FDA for the S.I.N. Dental Implant System. It primarily focuses on demonstrating substantial equivalence to previously cleared predicate devices by comparing their indications for use, designs, materials, and manufacturing processes, with an emphasis on adding new labeling claims related to a surface treatment.

    Crucially, this document states: "No clinical data were included in this submission." This means that the device's performance was not proven through a study involving human subjects or artificial intelligence (AI). The document relies on non-clinical data such as in vitro testing, animal testing referenced from published literature, sterilization, endotoxin, shelf-life, biological evaluation, MR compatibility, and fatigue testing to support the claims and substantial equivalence.

    Therefore, most of the requested information regarding acceptance criteria, study details, sample sizes, expert involvement, and specific performance metrics for an AI/device study cannot be extracted from this document.

    However, I can extract the information related to the non-clinical performance data presented to meet the criteria for substantial equivalence to predicates.

    Here's a breakdown of what can be extracted and an explanation of why other requested information is not present:

    Information that CANNOT be extracted from this document (and why):

    • A table of acceptance criteria and the reported device performance (for clinical/AI studies): Not applicable as no clinical or AI study was performed. The acceptance criteria relate to demonstrating substantial equivalence through non-clinical data, not specific performance metrics against a ground truth for a diagnostic claim.
    • Sample sizes used for the test set and the data provenance: Not applicable.
    • Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable.
    • Adjudication method for the test set: Not applicable.
    • If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable.
    • If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable (not an AI device).
    • The type of ground truth used (expert consensus, pathology, outcomes data, etc.) for test set: Not applicable as no clinical study or diagnostic AI was evaluated.
    • The sample size for the training set: Not applicable (not an AI device).
    • How the ground truth for the training set was established: Not applicable (not an AI device).

    What CAN be extracted/inferred regarding "acceptance criteria" and "study that proves the device meets the acceptance criteria" in the context of this 510(k) submission:

    The "acceptance criteria" for a 510(k) are typically met by demonstrating substantial equivalence to existing legally marketed devices (predicates). The "study that proves the device meets the acceptance criteria" in this case refers to the non-clinical tests and comparisons performed to show this equivalence and support the new labeling claims.

    1. A table of acceptance criteria and the reported "device performance" (in terms of non-clinical criteria):

    The acceptance criteria for this 510(k) submission are implicitly met by demonstrating that the subject device is as safe and effective as the predicate devices and that the new labeling claims are supported by scientific evidence. The "performance" here refers to meeting specific engineering, biological, and material standards, not diagnostic accuracy.

    Acceptance Criterion (Implicitly required for 510(k) & specific claims)Reported Device "Performance" (Evidence provided)
    Equivalence in Intended UseSubject device's Indications for Use (IFUS) are identical or substantially similar to those of predicate and reference devices (K170392, K193096, K200992, K211921, K221453, K222231). This includes general placement, loading conditions, specific length limitations for delayed loading, and angulation requirements for longer implants.
    Equivalence in Design/Physical CharacteristicsThe design (body/platform diameters, lengths, implant/abutment interface) of the subject device implants (including the new HA™® surface) is identical to those previously cleared in predicate/reference 510(k)s (K170392, K193096, K200992, K211921, K221453, K222231).
    Equivalence in MaterialsSubject device is manufactured from unalloyed titanium conforming to ASTM F67, identical to predicate devices. The HA™® surface treatment is identical to that cleared in K211921.
    Sterilization EfficacyDemonstrated by gamma irradiation sterilization to a sterility assurance level of 10^-6 by VDmax25 method, according to ISO 11137-1 and ISO 11137-2 (referenced from K211921).
    Absence of Pyrogens/EndotoxinsBacterial endotoxin testing (LAL test) according to ANSI/AAMI ST72 on manufacturing water (weekly) and sterilized product (quarterly) demonstrated all sterile product meets a limit of < 20 EU/device (referenced from K211921).
    Shelf-Life StabilityShelf life testing, including testing of samples after 4 years of real-time aging according to ASTM F1929 and F88/F88M (packaging sterile barrier) and sterility testing of product (referenced from K211921).
    BiocompatibilityBiological evaluation performed according to ISO 10993-1, with test results leveraged from K170392 for ISO 10993-3 (genotoxicity), ISO 10993-5 (cytotoxicity), ISO 10993-6 (implantation), ISO 10993-10 (sensitization, irritation), and ISO 10993-11 (systemic toxicity) to support biocompatibility.
    MRI Safety and CompatibilityNon-clinical analysis and testing according to ASTM F2052 (magnetically induced displacement force), ASTM F2213 (magnetically induced torque), ASTM F2182 (RF induced heating), and ASTM F2119 (image artifact), and the FDA guidance document "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment" (issued May 2021). No specific results are provided in the summary.
    Mechanical Strength/FatigueFatigue testing according to ISO 14801 referenced from K170392, K193096, K200992, K211921, K221453, and K222231 is applicable to the subject device (the specific results or "performance" against a benchmark are not provided in the summary, just that such testing was leveraged).
    Support for HAnano Surface ClaimsSurface Wettability (Hydrophilicity): Referenced Bezerra et al., 2017. Page 6: "Water contact angle was significantly less on HAnano treated titanium surface than on dual acid-etched titanium surface."Osteoblast Behavior/Differentiation: Referenced Bezerra et al., 2017 (Pages 4, 10) and da Silva et al., 2020 (Page 7). Claims include greater transition to differentiation-related signaling, changes in osteogenic gene marker expression, and spreading morphology different from machined/hydrophilic dual acid-etched surfaces. Results of in vitro testing are not necessarily predictive of human clinical performance.

    2. Sample sized used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):

    • Sample Size: Not applicable. The "test set" here would refer to the samples used in the non-clinical tests (e.g., in vitro cell cultures, physical test samples, animal models where applicable for referenced studies). Exact sample sizes for these referenced non-clinical studies are not provided in this summary.
    • Data Provenance:
      • Country of Origin: Not specified for the referenced studies (Bezerra et al., da Silva et al.). The manufacturing country for the device is Brazil (S.I.N. - Sistema de Implante Nacional S.A., São Paulo, Brazil).
      • Retrospective or Prospective: Not applicable as these were non-clinical, in vitro, or animal studies, not human clinical trials.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable as this is a non-clinical submission for a dental implant, not an AI/diagnostic device requiring expert ground truth for imaging interpretation. The "ground truth" for the non-clinical properties is established by the methods, standards, and scientific literature cited (e.g., ASTM, ISO standards, and published research papers).

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not applicable. There was no expert reader study.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, an MRMC study was not done. This is not an AI-enabled device.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • No, a standalone algorithm performance study was not done. This is not an AI-enabled device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The "ground truth" for the claims made are based on non-clinical in vitro and animal studies, and adherence to recognized standards for material properties, sterilization, biocompatibility, and mechanical performance. Specifically for the HAnano claims, the ground truth is derived from the experimental results reported in the referenced scientific publications (e.g., measurements of water contact angle, observation of cell morphology, assessment of gene marker expression).

    8. The sample size for the training set:

    • Not applicable. This is not an AI device trained on a dataset.

    9. How the ground truth for the training set was established:

    • Not applicable. This is not an AI device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K221453
    Date Cleared
    2022-12-15

    (210 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K211921, K193096, K170392, K200992

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    S.I.N. Dental Implant System is intended for placement in the maxillary or mandibular arch to provide support for single-unit or multi-unit restorations. When a one-stage surgical approach is applied, the S.I.N. Dental Implant System is intended for immediate loading when good primary stability is achieved and with appropriate occlusal loading.

    S.I.N. Dental Implant System implants with lengths of 18, 20, 22, or 24 mm may be tilted up to 30°. When used in the mandible or maxilla with implants with lengths of 18, 20, 22, or 24 mm at an angulation of 30°, a minimum of four implants must be used and must be splinted. When placed in the maxilla with lengths of 18, 20, 22, or 24 mm at angulations between 0° and less than 30°, the S.I.N. Dental Implant System implants are only indicated for multiple unit restorations in splinted applications that utilize at least two implants.

    All digitally-designed custom abutments for use with Interface CAD-CAM abutments are to be sent to a S.I.N.-validated milling center for manufacture.

    Device Description

    The purpose of this submission is to add components to the S.I.N. Dental Implant System, which includes components cleared previously in K211921, K203725, K200992, K193096, K170398, and K051859.

    This submission includes dental implants Epikut CM with a Morse taper (CM) abutment interface and an acid-etched endosseous surface, and Epikut Plus CM implants with an endosseous surface produced by acid-etching followed by application of a hydroxyapatite coating (HA ""0"). The implant design and endosseous surfaces are identical to those cleared in K211921, with the exception of the additional body/platform diameter (4.0 mm) and the longer lengths (18, 20, 22, and 24 mm).

    This submission includes Multifunctional Abutments with Morse taper connections (16°, 11.5°, and 4°), and protectors and temporary cylinders for these abutments; the subject device Multifunctional Abutments are compatible with subject device implants (CM 11.5°) and previously-cleared implants (CM 16° and 4°). This submission also includes Multifunctional Abutment components (cylinders) manufactured from Co-Cr-Mo alloy compatible with the subject Multifunctional Abutments and with abutments cleared in K170392; CAD-CAM abutment components (called "Interface") manufactured from Ti-6A1-4V alloy and Co-Cr-Mo alloy, compatible with the subject Multifunctional Abutments and with abutments cleared in K170392; and screws compatible with subject device components and previously cleared components with the 11.5° Morse taper connection. All subject device abutments are straight, with no angulation allowed.

    All subject device dental implants are manufactured from unalloyed titanium conforming to ASTM F67. The acid etching procedure is applied to all subject device dental implants. The acid etching process in this submission is identical to the process used to manufacture the dental implants cleared in K211921, and the HAM00 surface treatment is identical to that cleared in K211921.

    The subject device implants are compatible with abutments and prosthetic components in this submission and components cleared previously in K200992, K193096, K170392, and K051859.

    The subject device abutments and prosthetic components are compatible with implants and components in this submission and components cleared previously in K211921. K200992. K193096, and K170392.

    AI/ML Overview

    This document is a 510(k) Premarket Notification for the S.I.N. Dental Implant System. It does not describe an acceptance criteria or a study proving device performance against an acceptance criteria in the way typically found for AI/ML-driven medical devices. Instead, it demonstrates substantial equivalence to previously cleared predicate devices through non-clinical performance data and detailed comparisons of technological characteristics.

    Therefore, the requested information elements related to acceptance criteria, specific study designs (test sets, training sets, expert review, MRMC studies, standalone performance), and ground truth establishment are not applicable to this type of submission.

    The document focuses on showing that the new components (implants and abutments) of the S.I.N. Dental Implant System are as safe and effective as existing legally marketed predicate devices. This involves demonstrating equivalency in:

    • Intended Use: The indications for use are substantially equivalent to the predicate devices.
    • Technological Characteristics: Materials, manufacturing processes, design principles, sterilization methods, and performance data (non-clinical) are comparable or identical.

    Here's a breakdown of the requested information based on the provided document, highlighting why most are not directly present:

    1. Table of Acceptance Criteria and Reported Device Performance

    This type of table, with specific numerical acceptance criteria and a corresponding performance metric for an AI device, is not present because this is a 510(k) for a physical medical device (dental implants and components) and not an AI/ML software. The "performance data" provided are non-clinical tests demonstrating material properties, sterilization efficacy, biocompatibility, and mechanical integrity, which are described qualitatively or by reference to established standards (e.g., ISO, ASTM).

    Here's an illustrative table based on the provided "PERFORMANCE DATA" section, but it does not represent AI/ML acceptance criteria:

    Acceptance Criterion (Implicitly based on standards)Reported Device Performance (as summarized)
    Sterility Assurance Level (SAL) of 10⁻⁶ (ISO 11137-1 & 11137-2 for gamma; ANSI/AAMI/ISO 17665-1 & TIR 17665-2 for moist heat)Demonstrated by selecting/substantiating a 25 kGy dose (gamma) and by overkill method (moist heat), referenced from predicate devices.
    Bacterial Endotoxin Limit < 20 EU/device (ANSI/AAMI ST72)Demonstrated by LAL test on water samples (weekly) and sterilized product samples (quarterly), referenced from predicate devices.
    Shelf Life (packaging sterile barrier integrity and product sterility) after 4 years (ASTM F1929, F88/F88M)Demonstrated by testing samples after 4 years of real-time aging, referenced from predicate devices.
    Biocompatibility (ISO 10993-5, -3, -6, -10, -11)Demonstrated for implant (ASTM F67), abutment (ASTM F136, F1537), and zirconia coping (ISO 13356) materials, and for HAnano coating, referenced from predicate devices.
    HAnano Coating Characterization (SEM, XPS, TEM, XRD, adherence)Characterization performed and leveraged from predicate devices.
    Implant Acid-Etched Surface Characterization (SEM)Characterization performed and leveraged from predicate devices.
    MR Environment Safety (ASTM F2052, F2213, F2182, F2119 and FDA guidance)Non-clinical analysis and testing performed to evaluate metallic subject devices in MR environment.
    Mechanical Integrity (Implants with Abutments) (ISO 14801)Engineering analysis demonstrated subject devices do not create a new worst-case construct; previous mechanical testing applicable.

    2. Sample size used for the test set and the data provenance: Not applicable (no AI/ML test set mentioned). The document refers to non-clinical testing of physical device components.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable (no AI/ML ground truth mentioned).

    4. Adjudication method for the test set: Not applicable (no AI/ML test set mentioned).

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable (no AI/ML or human reader study mentioned).

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not applicable (no AI/ML algorithm mentioned).

    7. The type of ground truth used: Not applicable (no AI/ML ground truth mentioned). The "ground truth" for this submission are the established performance characteristics and safety profiles of the predicate devices and general standards for dental implants.

    8. The sample size for the training set: Not applicable (no AI/ML training set mentioned).

    9. How the ground truth for the training set was established: Not applicable (no AI/ML training set mentioned).