Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K170421

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-01-11

    (332 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K132219, K123627

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ET Hybrid Abutment is a customized abutment intended for use with HIOSSEN ET dental implant in the edentulous or partially edentulous maxilla or mandible to provide support for prosthetic restorations such as crowns and bridges. All digitally designed copings for use with the ET Hybrid Abutment for CAD/CAM are intended to be sent to a HIOSSEN Inc. manufacturing facility for manufacture.

    Device Description

    ET Hybrid Abutment is intended to be placed onto Hiossen dental implants to provide support for customized prosthetic restorations such as crowns and bridges. The proposed devices are composed of ET Link Abutment and Coping. The ET Link Abutment is premanufactured abutment by Hiossen Inc. only and the coping would be manufactured and bonded into the ET Link Abutment by Hiossen only with design input using CAD software from both by dental laboratories and by Hiossen Inc. The final device of ET Hybrid Abutment is under controlled by Hiossen's Quality System and provided to the customer. ET Hybrid Abutment is indicated for screw-retained single tooth or cement-retained single tooth and bridge restorations. Titanium alloy is the raw material for the Link Abutment and Zirconia material is for the Coping. The ET Hybrid Abutment gives the flexibility to design customized prosthetics.

    AI/ML Overview

    This document is a 510(k) summary for the ET Hybrid Abutment by HIOSSEN Inc., intended for use with HIOSSEN ET dental implants to support prosthetic restorations. The submission relies on the substantial equivalence to a primary predicate device, ET Hybrid Abutment (K162390), and two reference predicates: Straumann Variobase Abutments (K132219) and ET SmartFit Abutment (K123627). No clinical studies were submitted. Therefore, the information provided below will be based on the non-clinical testing and comparison to predicate devices, rather than a clinical study establishing the device's meeting of acceptance criteria from human data.

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state "acceptance criteria" in a quantitative format for the entire device. Instead, it relies on demonstrating substantial equivalence to predicate devices through material similarity, manufacturing processes, intended use, and non-clinical testing (biocompatibility and fatigue). The performance is demonstrated by conformance to relevant ISO standards for dental implants.

    Acceptance Criteria (Implied / Indirect)Reported Device Performance (Demonstrated vs. Predicate)
    Biocompatibility: Meet requirements for implantable devices.Demonstrated by relying on the predicate K162390. The subject device is manufactured using identical materials (Titanium Alloy Ti-6AL 4V, Zirconia Oxide) and manufacturing processes. Chemical composition, body contact (BONE / TISSUE, permanent > 30 days), and sterilization methods are the same as predicate devices. No additional biocompatibility testing was deemed necessary.
    Fatigue Strength: Withstand static and dynamic forces in oral environment.Successfully demonstrated through additional non-clinical testing: static and dynamic fatigue tests according to ISO 14801 in saline at 2Hz on the final finished sterilized abutment. The details of the performance (e.g., specific load cycles endured) are not provided in this summary but are implicit in the conclusion of substantial equivalence.
    Sterilization Efficacy: Ensure sterility after processing.Demonstrated by relying on the predicate device's sterilization validation. The subject device uses the same materials (Titanium & Zirconia, Cement) and the predicate's validation for steam sterilization was conducted according to ISO 17665-1, ISO/TS 17665-2 for gravity displacement on wrapped conditions. The worst-case sample (longest height, largest diameter) was used.
    Dimensional Specifications: Adhere to design limitations.The device specifications (Angulation, Gingival Diameter, Total Height, Contact Diameter, Wall Thickness, Gingival Height, Post Diameter, Post Height) are explicitly listed (e.g., Angulation 030°, Gingival Diameter 4.215 mm), indicating controlled design parameters. These are compared to the predicate, with the only difference noted as the coping part angle.
    Material Composition: Use specified materials.Uses Titanium Alloy Ti-6AL 4V and Zirconia Oxide, identical to the primary predicate device.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set: Specific sample sizes for the fatigue tests (ISO 14801) are not explicitly stated in this 510(k) summary. ISO standards typically specify minimum sample sizes for such tests.
    • Data Provenance: The origin of the testing data is not explicitly stated in terms of country, but it would have been conducted by or for HIOSSEN Inc., which is based in Fairless Hills, Pennsylvania, USA. The testing is prospective as it was conducted to support the
      510(k) submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. This is a medical device (dental abutment) undergoing non-clinical mechanical and biological testing for substantial equivalence, not an AI or diagnostic device requiring expert consensus for a ground truth in image interpretation or patient diagnosis.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. As this is not a diagnostic device or a study involving human interpretation of data, there is no adjudication method used. The testing relies on objective, standardized physical and material property measurements.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI-assisted diagnostic device, and therefore, no MRMC study or AI-related effectiveness study was conducted.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a physical medical device (dental abutment), not an algorithm or software.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" in this context refers to the established standards and specifications for material properties, mechanical performance, and biocompatibility of dental implant abutments. These are based on:

    • International Standards: Primarily ISO 14801 for fatigue testing, and ISO 17665-1, ISO/TS 17665-2 for sterilization validation.
    • Predicate Device Data: The performance and safety profile of the legally marketed predicate devices (K162390, K132219, K123627) serve as the benchmark for demonstrating substantial equivalence.
    • Material Specifications: Established specifications for Titanium Alloy Ti-6AL 4V and Zirconia Oxide.

    8. The sample size for the training set

    Not applicable. There is no training set as this is not a machine learning or AI device.

    9. How the ground truth for the training set was established

    Not applicable. No training set was used.

    Ask a Question

    Ask a specific question about this device

    K Number
    K162390

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-01-30

    (157 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K100245, K123627

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ET Hybrid Abutment is a customized abutment intended for use with HIOSSEN ET dental implant in the edentulous or partially edentulous maxilla or mandible to provide support for prosthetic restorations such as crowns and bridges. All digitally designed copings for use with the ET Hybrid Abutment for CAD/CAM are intended to be sent to a HIOSSEN Inc. manufacturing facility for manufacture.

    Device Description

    ET Hybrid Abutment is intended to be placed onto Hiossen dental implants to provide support for customized prosthetic restorations such as crowns and bridges. The proposed devices are composed of ET Link Abutment and Coping. The ET Link Abutment is premanufactured abutment by Hiossen Inc. only and the coping would be manufactured by Hiossen only with design input using CAD/CAM software from both by dental laboratories and by Hiossen Inc. The final device of ET Hybrid Abutment is under controlled by Hiossen's Quality System and provided to the customer. ET Hybrid Abutment is indicated for screw-retained single tooth or cement-retained single tooth and bridge restorations. Titanium alloy is the raw material for the Link Abutment and Zirconia material is for the Coping. The ET Hybrid Abutment gives the flexibility to design customized prosthetics. The coping is straight only and the ET Link abutment is not intended to provide an angle and/or divergence correction.

    AI/ML Overview

    The provided text is a 510(k) summary for the "ET Hybrid Abutment" by HIOSSEN Inc. This document focuses on demonstrating substantial equivalence to predicate devices, rather than providing detailed acceptance criteria and a study proving that the device meets those criteria.

    Therefore, most of the requested information regarding acceptance criteria, specific device performance, sample sizes, expert qualifications, adjudication methods, multi-reader multi-case studies, standalone performance, and ground truth establishment is not available in the provided text, as this type of information is typically part of performance studies, which are not included here.

    However, based on the non-clinical testing summary, we can glean some information regarding the sterilization validation.

    Here's what can be extracted and what is explicitly not available:

    1. Table of acceptance criteria and the reported device performance:

    Acceptance CriteriaReported Device Performance
    BiocompatibilityRelied upon predicate devices (K123627, K100245) for biocompatibility. Subject device manufactured using identical materials and manufacturing processes to previously cleared predicates.
    Sterilization Validation (Steam sterilization)All tests conducted in accordance with ISO 17665-1 and ISO/TS 17665-2 for gravity displacement on unwarped condition; no discrepancies.

    2. Sample size used for the test set and the data provenance:

    • Not available. The document states "No additional biocompatibility testing was necessary for this device" as it relies on predicate devices. For sterilization validation, the specific sample size for the test set is not provided. Data provenance is also not specified beyond adherence to ISO standards.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable/Not available. No specific ground truth establishment by experts for performance testing is detailed, as most of the evaluation relies on substantial equivalence to predicates and adherence to international standards for sterilization.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not applicable/Not available. No adjudication method is mentioned as there were no clinical studies or reported performance evaluations that would typically require such a method.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This device is a dental abutment, not an AI-powered diagnostic device. Therefore, MRMC studies and AI-related effectiveness are not relevant.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This device is a dental abutment, not a software algorithm.

    7. The type of ground truth used:

    • For biocompatibility: Implicitly, the "ground truth" relies on the established biocompatibility of the materials (Titanium Alloy Ti-6AL 4V and Zirconia) in the predicate devices and their manufacturing processes, which were deemed safe and effective.
    • For sterilization validation: The "ground truth" is adherence to the validated sterilization parameters and protocols defined by ISO 17665-1 and ISO/TS 17665-2.

    8. The sample size for the training set:

    • Not applicable. As this is a medical device (dental abutment) and not an AI/machine learning system, there is no "training set."

    9. How the ground truth for the training set was established:

    • Not applicable. No training set exists for this type of device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K160436

    Validate with FDA (Live)

    Date Cleared
    2016-09-29

    (225 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K091239, K123627

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CreoDent Solidex® Customized Abutment is intended for use with an endosseous implant to support a prosthetic device in patients who are partially or completely endentulous. The device can be used for single or multiple-unit restorations. The prosthesis can be cemented or screw retained to the abutment. An abutment screw is used to secure the abutment to the endosseous implant.

    The CreoDent Solidex® Customized Abutment is compatible with the following:

    • Astra Tech Osseospeed TX Implants 3.5mm, 4.5mm
    • Hiossen TS Implants 3.5mm, 4.0mm
    Device Description

    The Solidex® Customized Abutment is Ti-6A1-4V Eli titanium alloy meets ASTM F-136 Standard and Screw is CP TI Gr4 meets ASTM F67 and is designed to be screw retained for use with endosseous dental implants to provide support for a prosthetic restoration. These abutments are indicated for cement or screw retained restorations. Solidex® Customized Abutments are compatible with:

    • Astra Tech Osseospeed TX Implants 3.5mm, 4.5mm (K024111)
    • Hiossen TS Implants 3.5mm, 4.0mm (K151858)

    The design of subject device is customized to the requirements of each patient as may be specified by the prescribing dentist. Customization is limited by the minimum dimensions for wall thickness, diameter, height, collar height and angulation.

    AI/ML Overview

    The medical device under consideration is the CreoDent Solidex® Customized Abutment. The provided document is a 510(k) premarket notification summary, which focuses on demonstrating substantial equivalence to legally marketed predicate devices rather than proving a device meets specific acceptance criteria through a standalone clinical study. Therefore, some of the requested information, such as sample sizes for test/training sets, number/qualifications of experts for ground truthing, adjudication methods, and MRMC studies, is not typically part of a 510(k) submission for this type of device and is not present in the provided text.

    Based on the information available:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the CreoDent Solidex® Customized Abutment are implicitly derived from its substantial equivalence to predicate devices in material, indications for use, performance characteristics, and dimensions/angulations, backed by non-clinical testing. The performance is assessed against the established standards and the characteristics of the predicate devices.

    Acceptance Criteria CategorySpecific Acceptance Criteria (Derived from Predicate Equivalence)Reported Device Performance (CreoDent Solidex® Customized Abutment)
    MaterialAbutment: Ti-6A1-4V Eli titanium alloy meets ASTM F-136 Standard. Screw: CP TI Gr4 meets ASTM F67 Standard. (Comparable to predicate materials)Abutment: Ti-6A1-4V Eli titanium alloy meets ASTM F-136 Standard. Screw: CP TI Gr4 meets ASTM F67 Standard. (Reported as a higher grade material with more tensile strength than the predicate for the abutment, and comparable for the screw).
    Performance CharacteristicsAllows prosthesis to be cemented or screw-retained to the abutment, with the abutment screw securing it to the endosseous implant. (Comparable to predicate functionality).Allows the prosthesis to be cemented or screw retained to the abutment. The abutment screw is designed to secure the abutment to the endosseous implant.
    Indications for UseIntended for use with an endosseous implant to support a prosthetic device in patients who are partially or completely edentulous, for single or multiple-unit restorations, cemented or screw-retained. Compatible with Astra Tech Osseospeed TX Implants 3.5mm, 4.5mm and Hiossen TS Implants 3.5mm, 4.0mm. (Comparable to predicate except for specific compatible implant bodies).Intended for use with an endosseous implant to support a prosthetic device in patients who are partially or completely edentulous, for single or multiple-unit restorations, either cemented or screw retained. Compatible with Astra Tech Osseospeed TX Implants 3.5mm, 4.5mm and Hiossen TS Implants 3.5mm, 4.0mm.
    Dimensions & AngulationsPersonalized abutment design with minimum dimensions for wall thickness, diameter, height, collar height, and angulation. Angles not to exceed up to 20 degrees from the implant axis. (Compared to predicate's specific compatible implant sizes and angulation limits).Customizable to patient requirements, limited by minimum dimensions. Abutment sizes for specified Astra Tech Osseospeed TX Implants (3.5mm, 4.5mm) and Hiossen TS Implants (3.5mm, 4.0mm). Angles not to exceed up to 20 degrees from the implant axis (for Astra Tech and Hiossen TS compatibility). This is a difference from one reference predicate which had 30 degrees angulation but is mitigated by testing.
    Mechanical Strength & CompatibilitySufficient mechanical strength for intended clinical application and compatibility with specified implant systems, as demonstrated by non-clinical testing (ISO 14801:2007E, reverse engineering, sterilization validation). This is the key performance "acceptance criteria" evaluated via testing.Static/Fatigue testing in accordance with ISO 14801:2007E (worst-case scenario), reverse engineering dimensional analysis, and sterilization validation (ISO 17665-1) were performed. These results demonstrated that the Solidex® customized Abutment has sufficient mechanical strength and is compatible with the Astra Tech Osseospeed TX and Hiossen TS implant systems. This testing specifically addresses the differences in compatible implant bodies and angulation limitations compared to predicates.

    Study Proving Acceptance Criteria:

    The study proving the device meets its acceptance criteria is a non-clinical testing study focused on mechanical performance and compatibility.

    • Study Type: Non-clinical testing, specifically static/fatigue testing, reverse engineering dimensional analysis, and sterilization validation.
    • Standards Followed: ISO 14801:2007E (Dentistry-Implants-Dynamic fatigue test for endosseous dental implants) and ISO 17665-1 (Sterilization validation).
    • Purpose: To demonstrate that the CreoDent Solidex® Customized Abutment has sufficient mechanical strength for its intended clinical application and is compatible with the specified Astra Tech Osseospeed TX and Hiossen TS implant systems. This testing specifically addressed the differences identified when comparing to predicate devices (compatible implant bodies and angulation).

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not specified in the document. Non-clinical mechanical testing typically involves a sufficient number of samples to ensure statistical significance for the specific test (e.g., number of abutment-implant assemblies tested for fatigue) but this exact number is not provided.
    • Data Provenance: Not explicitly stated (e.g., country of origin, retrospective/prospective). As a non-clinical, in-vitro mechanical testing study, the data is generated in a laboratory setting.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    • This is a mechanical testing study, not one requiring expert human interpretation of medical images or patient data to establish ground truth. Therefore, this information is not applicable and not provided. The "ground truth" here is the physical performance of the device under defined test conditions per international standards.

    4. Adjudication Method for the Test Set

    • Not applicable. As a mechanical testing study, adjudication of human interpretations is not involved.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No, a MRMC comparative effectiveness study was not done. This type of study is for evaluating diagnostic or prognostic algorithms read by human experts, which is not relevant for this dental implant abutment.

    6. Standalone (Algorithm Only Without Human-in-the-loop Performance) Study

    • Not applicable. This device is a physical medical device (an abutment for dental implants), not a software algorithm.

    7. Type of Ground Truth Used

    • The "ground truth" for this device's performance evaluation is established through physical measurements and adherence to international mechanical testing standards (ISO 14801:2007E). The benchmarks are defined by these standards, the performance of the legally marketed predicate devices, and the physical properties of the materials and design.

    8. Sample Size for the Training Set

    • Not applicable. This is a physical device, and the validation process described is for its mechanical integrity and compatibility, not for an algorithm that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable as there is no training set for a physical device like this.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1