Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K201085

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2021-01-15

    (267 days)

    Product Code
    Regulation Number
    870.5150
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K122478, K160641

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The NeVa™ PV Thrombectomy Device is indicated for:

    • The non-surgical removal of emboli and thrombi from peripheral blood vessels.
    • Temporary use in peripheral vessel occlusion.
    • Use with aspiration and with the injection or infusion of contrast media and other fluids.
    Device Description

    The NeVa™ PV Thrombectomy Device is a sterile, single-use thrombectomy device designed to for use in the peripheral vasculature to restore blood flow via non-surgical removal of emboli and thrombi from peripheral vessels in the vasculature. The NeVa™ PV device consists of a self-expanding nitinol basket attached to a core pusher wire that can be delivered through a micro catheter, deployed across a clot, and then removed with the clot to enable revascularization of the occluded vessel. To aid in fluoroscopic visualization, radiopaque markers are positioned at the distal end, proximal end, and along the nitinol basket of the NeVa™ PV device. A flexible atraumatic soft tip attached to the distal end of the nitinol basket is radiopaque for enhanced fluoroscopic visualization.

    The NeVaTM PV device is provided sterile and is intended for single patient use. The device is not intended to be re-sterilized or re-used. No accessories are provided with the NeVaTM PV device. The device is intended to be used in conjunction with the appropriately sized microcatheter as indicated in the proposed Instructions for Use.

    AI/ML Overview

    The provided text describes the regulatory clearance of a medical device, the NeVa™ PV Thrombectomy Device, based on its substantial equivalence to a predicate device. It details the device's characteristics, indications for use, and a summary of nonclinical testing performed to demonstrate its performance. However, the document does not contain information about acceptance criteria for a study, nor does it detail a study that proves the device meets specific acceptance criteria in the context of an algorithm or AI performance.

    The document is for a medical device clearance (510(k)) that relies on substantial equivalence to a predicate device (ReVive™ PV Thrombectomy Device), not on meeting specific quantitative performance metrics as one might see for an AI/algorithm-based medical device.

    Therefore, I cannot provide the requested information for acceptance criteria and a study proving device performance in the context of AI/algorithm evaluation. The document focuses on the physical and functional performance of a mechanical thrombectomy device.

    To directly address your request given the provided text, I must state that the information is not present in the provided regulatory submission.

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    Reference Devices :

    K150616, K143077, K133464, K132641, K122478, K120961

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Trevo Retrievers are indicated for use to restore blood flow in the neurovasculature by removing thrombus for the treatment of acute ischemic stroke to reduce disability in patients with a persistent, proximal anterior circulation, large vessel occlusion, and smaller core infarcts who have first received intravenous tissue plasminogen activator (IV t-PA). Endovascular therapy with the device should start within 6 hours of symptom onset.

    Device Description

    The Trevo ProVue and XP ProVue Retrievers consist of a flexible, tapered core wire with a shaped self-expanding section at the distal end for clot capture and removal. Radiopaque platinum wires in the shaped section and radiopaque markers on the distal end allow fluoroscopic visualization. The Trevo Retrievers have a hydrophilic coating to reduce friction during use. A torque device and an insertion tool are provided with the Retrievers. The Trevo Retrievers are delivered to the site of occlusion in the neurovasculature through a microcatheter. The torque device may be used to lock the core wire of the Trevo Retriever to the microcatheter during the procedure, allowing the Trevo Retriever and microcatheter to be retracted as a system through the guide catheter and removed from the body with captured clot.

    AI/ML Overview

    The acceptance criteria for the Trevo ProVue and XP ProVue Retrievers, a neurovascular mechanical thrombectomy device, are derived from the clinical study, bench testing, animal studies, and various regulatory requirements. The study supporting these criteria is primarily the MR CLEAN (Multicenter Randomized Clinical Trial of Endovascular Treatment for Acute Ischemic Stroke in the Netherlands) trial, with additional non-clinical and animal studies.

    Here's a breakdown of the acceptance criteria and reported device performance:

    1. Table of Acceptance Criteria and Reported Device Performance

    Note: The document does not explicitly state "acceptance criteria" for each performance metric with specific thresholds (e.g., "must achieve X%"). Instead, the assessment is based on demonstrating safety and effectiveness as compared to medical management alone, or meeting benchmarks set by non-clinical testing. Where specific targets are not given, the trial's statistically significant improvement or demonstration of comparable safety/effectiveness serves as the "meeting" of the criteria.

    CategoryAcceptance Criteria (Implied / Explicit)Reported Device Performance
    EffectivenessPrimary Effectiveness Endpoint: Clinically meaningful functional independence (mRS ≤ 2) at 90 days, demonstrating superiority over Medical Management (MM) alone.29.2% (28/96) of patients in the Trevo Retriever group achieved mRS ≤ 2 at 90 days, compared to 19.3% (48/249) in the MM group. Odds Ratio [95% CI]: 1.88 [1.07, 3.29], p-value = 0.014. (Demonstrated statistically significant superiority).
    Secondary Effectiveness Endpoint 1: Percentage of patients with no intra-cranial occlusion after 24 hours.77.5% (62/80) in the Trevo group achieved no intra-cranial occlusion, compared to 33.51% (65/194) in the MM group. (Demonstrated significantly higher rates of recanalization).
    Secondary Effectiveness Endpoint 2: Percentage of Trevo patients with recanalization (TICI ≥ 2a).81.3% (78/96) of Trevo patients achieved TICI ≥ 2a.
    Secondary Effectiveness Endpoint 3: Neurological outcomes assessed by NIHSS at 24 hours and 5-7 days post-randomization, demonstrating improvement compared to MM alone.NIHSS (24 Hours): Trevo Mean ± SD: 14.23 ± 9.23 (N=96) vs. MM Mean ± SD: 16.19 ± 7.77 (N=240). NIHSS (5-7 Days): Trevo Mean ± SD: 13.02 ± 12.16 (N=94) vs. MM Mean ± SD: 15.76 ± 11.49 (N=237). (Directionally favorable for Trevo, though statistical significance for these endpoints isn't explicitly stated).
    SafetyPrimary Safety Endpoint: All-cause mortality rate at 90 days, demonstrating non-inferiority or reduced mortality compared to MM alone.Trevo FDA Cohort: 13.3% (16/120) mortality. MM: 22.89% (57/249) mortality. (Lower mortality rate in Trevo group).
    Secondary Safety Endpoint 1: Proportion of patients with symptomatic intracranial hemorrhage (sICH) within 24 hours.Trevo FDA Cohort: 6.67% (8/120) sICH. MM: 4.42% (11/249) sICH. (Slightly higher incidence in Trevo group, but likely considered acceptable within the benefit-risk profile).
    Secondary Safety Endpoint 2: Proportion of patients with neurological deterioration within 5-7 days or discharge.Trevo FDA Cohort: 9.17% (10/109) . MM: 9.57% (22/230). (Comparable rates).
    Other Adverse Events: Monitor rates of various SAEs and adverse events to ensure an acceptable safety profile.Comprehensive reporting of SAEs (Table 10) and MedDRA coded AEs (Table 11) indicates generally comparable or acceptable rates across various categories, with some expected differences due to the invasive nature of the procedure (e.g., higher haemorrhage intracranial in Trevo, but also higher "other infection" and "other complication"). The overall benefit-risk determination concluded that benefits outweigh risks.
    BiocompatibilityDevice materials demonstrated to be biocompatible (ISO 10993-1:2009/AC:2010 compliance).Leveraged from prior submissions (K120961). Tests included hemocompatibility/coagulation, hemolysis, cytotoxicity, maximization test for delayed hypersensitivity, intracutaneous reactivity, acute systemic toxicity, and material-mediated rabbit pyrogenicity. (Met standard).
    SterilitySterility assurance level (SAL) of 10^-6 (ISO 11135-1:2007). Sterilant residuals per ISO 10993-7:2008. Endotoxin testing <2.15 EU/device (USP <161>).Achieved SAL of 10^-6. Sterilant residuals tested. Endotoxin testing met criterion. (Met standards).
    Shelf-Life24-month shelf-life supported by accelerated and real-time shelf-life testing, including packaging integrity and component functional performance.Supported by accelerated testing (Trevo XP ProVue) and real-time testing (Trevo ProVue), including: dimensional verification, tensile strength, radial force, tip flexibility, torque/tensile durability, retriever platinum wire and joint durability, radiopacity, deliverability, simulated use, coating integrity, and particulate testing. Packaging integrity tested to ASTM D4169, F1980, F1929, F2906, F88/M. (Met standard).
    Performance (Bench)Device performs as intended under anticipated conditions of use (Table 2 tests: Dimensional Verification, Tensile Strength, Radial Force, Tip Flexibility, Torque/Tensile Durability, Platinum Wire Joint Strength/Durability, Radiopacity, Deliverability, Particulate Evaluation, Coating Integrity Evaluation, Simulated Use).All listed bench tests were conducted and passed: Verified dimensions; recorded peak tensile force; constrained and released shaped section to record radial force; flexed distal tip to record compression/flex force; applied rotations and tensile cycles for durability; recorded peak tensile force for platinum wires; performed wrap/unwrapped and deploy/reload cycles; assessed radiopacity visually; measured force to push device through tortuous model; measured particulates; performed visual assessment of coating integrity before/after simulated use; simulated use in silicone neurovascular model with thrombus. (All tests passed, indicating performance as intended).
    Performance (Animal)Safety (vessel response) demonstrated in animal studies, including no arterial dissection, perforation, or thrombosis.Acute and chronic swine studies (up to 6 passes). Acute study: no evidence of vessel dissection, perforation, or thrombosis at Day 0. Chronic study: no evidence of dissection, perforation, or thrombosis at Day 0, and no stenosis, vessel irregularity, intimal flap or pseudoaneurysm at Day 30. Histopathology showed arterial healing consistent with routine catheterization. (Demonstrated safety in animal models).

    2. Sample Sizes Used for the Test Set and Data Provenance

    The primary clinical study supporting the device's expanded indication for the Trevo Retrievers was the MR CLEAN trial.

    • Test Set Sample Size:

      • TREVO FDA Cohort: 120 patients (from the IAT arm, met specific criteria) (N=120 for primary safety; N=96 for primary effectiveness after exclusions).
      • Medical Management (MM) Control Group: 249 patients (after exclusions for IV t-PA timing).
      • Total Patients in MR CLEAN (originally randomized): 500 patients (233 IAT, 267 control).
    • Data Provenance:

      • Country of Origin: The Netherlands (MR CLEAN was conducted in every endovascular hospital center in the Netherlands).
      • Retrospective or Prospective: Prospective, randomized, open-label, controlled, multicenter trial.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The document mentions expert review for several endpoints:

    • Recanalization and Occlusion Assessment (Table 4 & 5):
      • Number of Experts: "Consensus review by up to three readers" for occlusion assessment (Table 4). "Consensus review by at least three readers" for TICI score recanalization (Table 5).
      • Qualifications: The document does not explicitly state the specific qualifications (e.g., years of experience, subspecialty) of these readers/experts. However, given the nature of the study (neurovascular stroke treatment), it is highly probable they were experienced neurologists, neuroradiologists, or stroke specialists involved in the study's clinical centers.

    4. Adjudication Method for the Test Set

    • Adjudication Method: "Consensus review by up to three readers" or "at least three readers" was used for evaluating arterial occlusion and recanalization (TICI scores). This implies a method where multiple experts independently (or collectively) assess the imaging and arrive at a joint decision. The specific process (e.g., if there were disagreements, how they were resolved beyond just "consensus") is not detailed beyond "consensus review."

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of Human Reader Improvement

    • No, a typical MRMC comparative effectiveness study, as often seen for diagnostic AI, was not performed in this context.
      • The MR CLEAN trial directly compared patient outcomes with and without the device (with medical management), rather than evaluating how human readers' diagnostic accuracy or treatment decisions improved with AI assistance.
      • The Trevo Retrievers are a mechanical device for thrombectomy, not a diagnostic AI algorithm intended to assist human readers in interpreting images. Therefore, the concept of "how much human readers improve with AI vs. without AI assistance" does not directly apply to this device's clinical evaluation.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Study Was Done

    • Yes, in essence, the "device" (Trevo Retrievers) was evaluated in a standalone manner as a treatment in the context of the clinical trial.
      • The MR CLEAN trial evaluated the performance of the neurovascular mechanical thrombectomy device (Trevo Retrievers) in conjunction with best medical management against best medical management alone. This is effectively a "standalone" evaluation of the device's therapeutic effect, rather than an "algorithm only" evaluation, as it's a physical medical device.
      • The device's performance is measured by its direct impact on patient outcomes when used by clinicians, not as an interpretive AI tool.

    7. The Type of Ground Truth Used

    • Clinical Outcomes/Expert Assessment (Blinded):
      • Primary Effectiveness (mRS): The 90-day mRS assessment was performed by an assessor who was blinded to the subject's treatment allocation. This is a high-quality method for establishing ground truth for functional independence.
      • Imaging-based Endpoints (Occlusion/Recanalization): Assessed by a consensus review by up to three/at least three readers (presumably domain experts) after 24 hours using CTA or MRA and the AOL/TICI scales.
      • Safety Endpoints: Based on collected adverse events, neurological assessments (NIHSS), and mortality data, which are objective clinical measures or physician-reported events.
      • Animal Studies: Safety ground truth established by intra-procedural angiography and histopathology assessments performed by experts.

    8. The Sample Size for the Training Set

    • Not Applicable in the traditional sense for an AI/algorithm: The information provided describes the regulatory approval of a physical medical device (mechanical thrombectomy retriever), not an AI algorithm. Therefore, there is no "training set" for an AI model.
    • The clinical data from the MR CLEAN trial serves as the pivotal clinical evidence supporting the device's effectiveness and safety, analogous to a validation set for an AI, but it's not a "training set" for an algorithm.

    9. How the Ground Truth for the Training Set Was Established

    • Not Applicable: As explained above, this is a physical medical device, not an AI algorithm, so there is no "training set" or ground truth establishment process for an AI model. The "ground truth" for the device's performance is established through rigorous clinical trials and non-clinical testing as described in the previous sections.
    Ask a Question

    Ask a specific question about this device

    K Number
    K132641

    Validate with FDA (Live)

    Date Cleared
    2014-01-13

    (140 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K122478

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Trevo Retriever is intended to restore blood flow in the neurovasculature by removing thrombus in patients experiencing ischemic stroke within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.

    Device Description

    The Modified Trevo ProVue Retriever consists of a flexible, tapered core with a shaped section at the distal end. It is designed to restore blood flow in the neurovasculature by removing thrombus in patients experiencing ischemic stroke. Radiopaque platinum wires in the shaped section and new radiopaque markers on the new truncated distal end allow fluoroscopic visualization. Retriever dimensions are indicated on the product label. The Retriever has a hydrophilic coating to reduce friction during use. A torque device and an insertion tool are provided with the Retriever. The proximal end of the device is compatible with the Abbott guide wire extension to facilitate removal or exchange of a catheter while maintaining the Retriever position in the vessel. The Modified Trevo ProVue Retriever is offered in two sizes.

    AI/ML Overview

    The provided text describes a 510(k) submission for a Modified Trevo ProVue Retriever, a medical device intended to remove thrombus in patients experiencing ischemic stroke. The submission focuses on demonstrating substantial equivalence to a predicate device (Concentric Trevo ProVue Retriever K122478) rather than an AI-powered diagnostic device. Therefore, many of the requested categories related to AI performance, ground truth, expert adjudication, and MRMC studies are not applicable or cannot be extracted from this document.

    Here's the information that can be extracted or inferred from the provided text, with an explanation of why other requested information is not present:


    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance CriteriaReported Device Performance
    Dimensional VerificationMet"Dimensional verification meets acceptance criteria."
    Retriever Proximal Joint Tensile StrengthMet"Retriever Proximal Joint Tensile Strength meets acceptance criteria."
    Retriever Mid Joint Tensile StrengthMet"Retriever Mid Joint Tensile Strength meets acceptance criteria."
    Retriever Tip Tensile StrengthMet"Retriever Tip Tensile Strength meets acceptance criteria."
    Retriever Shaped Section Radial ForceMet"Retriever Shaped Section Radial Force meets acceptance criteria."
    Retriever/Vessel Interaction (Tip Flexibility)Met"Retriever/Vessel Interaction (Tip Flexibility) meets acceptance criteria."
    Retriever Torque Tensile DurabilityMet"Retriever Torque Tensile Durability meets acceptance criteria."
    Retriever Platinum Wire Joint StrengthMet"Retriever Platinum Wire Joint Strength meets acceptance criteria."
    Retriever Platinum Wire and Joint DurabilityMet"Retriever Platinum Wire and Joint Durability meets acceptance criteria."
    RadiopacityAssessed based on visual assessment of the device being used under fluoroscopy."Radiopacity meets acceptance criteria."
    Retriever/Microcatheter DeliverabilityMet"Retriever/Microcatheter Deliverability meets acceptance criteria."
    Proximal Curl ResistanceMet"Proximal Curl Resistance meets acceptance criteria."
    Simulated UseDemonstrated successful performance in a silicone neurovascular model with simulated thrombus."Simulated Use meets acceptance criteria."
    Biocompatibility (Cytotoxicity - ISO Elution Method)Non-cytotoxic (Dilutions: Undiluted (3 - Moderate reactivity - noted as consistent with historical data), 50%, 25%, 12.5%, 6.25%, 3.13% - 0 reactivity)Pass for all dilutions, with specific note on historical consistency for undiluted moderate reactivity.
    Biocompatibility (ASTM Hemolysis)Non-HemolyticPass. Hemolytic index: 0.0% (direct contact) and 0.4% (extract)
    Biocompatibility (Physicochemical Tests Plastics)No heavy metals or leachables (Non-volatile residue: 2 mg, Residue on ignition: ≤ 2 mg, Heavy metal: < 1 ppm, Buffering capacity: < 1.0 mL)Pass. Stated conclusion: "No heavy metals or leachables."

    Explanation: The acceptance criteria are implicitly stated by the "Conclusion" column in the "Testing Summary" and "Biocompatibility" tables, which consistently reports that the device "meets acceptance criteria." The specific quantitative acceptance criteria are not explicitly detailed (e.g., what is the exact tensile strength required), but the summary confirms they were met.


    2. Sample Size Used for the Test Set and the Data Provenance

    • Sample Size: The document does not specify the exact sample sizes (number of devices, number of tests, or number of simulated emboli) used for the various physical and simulated use tests.
    • Data Provenance:
      • Simulated Use: The test used a "silicone neurovascular model cast from actual human neurovascular arteries," replicating tortuosity, diameter, and location of arteries. This indicates a benchtop, in vitro model.
      • Other Tests: The other tests (dimensional, tensile strength, radial force, durability, radiopacity, deliverability, curl resistance) are laboratory benchtop tests performed on physical device samples.
      • All testing was conducted by Concentric Medical, Inc. (the submitter), implying an internal, controlled testing environment.
      • The tests are prospective in nature, as they were conducted specifically for this submission.
      • The country of origin of the data is not explicitly stated but is implicitly the country where Concentric Medical, Inc. conducts its R&D and testing, likely the USA given the submission to the FDA.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This information is Not Applicable to this submission. The Modified Trevo ProVue Retriever is a physical medical device, not an AI diagnostic algorithm. Ground truth for the performance characteristics of the device is established through physical and simulated benchtop testing, not through expert review of clinical data or images. The "Radiopacity" test mentions "visual assessment," which likely implies assessment by trained personnel (e.g., engineers, quality control specialists) using fluoroscopy, but no details on their qualifications or number are provided, as it's a standard engineering/device performance check.


    4. Adjudication Method for the Test Set

    This information is Not Applicable to this submission, as it pertains to expert consensus for clinical ground truth in AI studies. The tests conducted are objective physical and performance measurements.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is Not Applicable. This is a submission for a physical medical device. There is no AI component or human-in-the-loop performance being evaluated in this K132641 summary.


    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is Not Applicable. This is a physical medical device, not a standalone algorithm.


    7. The Type of Ground Truth Used

    The "ground truth" for this device's performance is established through objective, quantifiable physical and mechanical tests, and simulated use in a validated in vitro model. This implicitly serves as the "ground truth" for its functional characteristics.

    • Physical/Mechanical Properties: Measurements directly obtained during testing (e.g., force, dimensions, durability cycles) compared against predetermined specifications.
    • Simulated Use: The ability of the device to successfully retrieve simulated thrombus in a anatomically correct (cast from human arteries) silicone neurovascular model.
    • Biocompatibility: Laboratory test results (cytotoxicity, hemolysis, physicochemical properties) compared against ISO 10993-1 standards and internal specifications.

    8. The Sample Size for the Training Set

    This information is Not Applicable. The device is a physical medical device; there is no AI training set involved.


    9. How the Ground Truth for the Training Set was Established

    This information is Not Applicable. There is no "training set" for this physical medical device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1