K Number
K160641

Validate with FDA (Live)

Date Cleared
2017-03-08

(366 days)

Product Code
Regulation Number
870.1250
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Solitaire™ Platinum Revascularization Device is intended to restore blood flow by removing thrombus from a large intracranial vessel in patients experiencing ischemic stroke within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.

Device Description

The subject 6-40-10 Solitaire™ Platinum Revascularization Device is designed to restore blood flow in patients experiencing ischemic stroke due to large intracranial vessel occlusion. The Solitaire™ Platinum Revascularization Device is designed for use in the neurovasculature such as the Internal Carotid Artery (ICA), M1 and M2 segments of the middle cerebral artery, basilar, and the vertebral arteries. The distal nitinol portion of the Solitaire™ Platinum Revascularization Device facilitates clot retrieval. The Solitaire™ Platinum Revascularization Device has Platinum Iridium radiopaque markers on the working cell length, proximal and distal ends. The subject Solitaire™ Platinum Revascularization Device is a portfolio expansion to the predicate Solitaire™ Platinum Revascularization Device (K153071).

AI/ML Overview

Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

Device: Solitaire™ Platinum Revascularization Device (6x40 mm)


1. Table of Acceptance Criteria and Reported Device Performance

The provided document describes both biocompatibility testing and non-clinical bench testing. There is also clinical data leveraged from a registry.

Biocompatibility Acceptance Criteria & Performance:

Test CategoryTest DescriptionMethodAcceptance CriteriaReported Device Performance (Conclusion)
CytotoxicityL929 MTT CytotoxicityISO 10993-5Viability is $\geq$ 70%.Acceptance criteria met
SensitizationGuinea Pig Maximization SensitizationISO 10993-10Test article does not elicit a sensitization response.Acceptance criteria met
IrritationIntracutaneous Irritation TestISO 10993-10Differences in the mean test and control scores of the extract dermal observations are $<$ 1.0.Acceptance criteria met
Acute Systemic ToxicityAcute Systemic Injection TestISO 10993-11No abnormal clinical signs and weight loss in excess of 10%. Temperature rise $\geq$ 0.5°CAcceptance criteria met
Materials Mediated Rabbit Pyrogen(Not specified)(Not specified)Acceptance criteria met
Hemo-compatibilityHemolysisISO 10993-4No significant differences between the test article extract and negative control article results. The test article is considered non-hemolyticAcceptance criteria met
Partial Thromboplastin Time(Not specified)Clotting times are similar to the negative control and the reference material (HDPE) indicating the device materials are not an activator of the intrinsic coagulation pathway.Acceptance criteria met
Platelet and Leukocyte Count(Not specified)Test article does not adversely affect the platelet and leukocyte components of the blood compared to the reference material.Acceptance criteria met
Complement Activation C3a and SC5b-9 Assay(Not specified)Levels of C3a and SC5b-9 are comparable and less than the positive control.Acceptance criteria met
Thrombosis(Not specified)Thrombo-resistance properties are acceptable in clinical use.Acceptance criteria met
GenotoxicityBacterial Mutagenicity TestISO 10993-3Test article is considered non-mutagenicAcceptance criteria met
In-vitro Mouse Lymphoma Assay(Not specified)Test article is considered non-mutagenicAcceptance criteria met
In-vivo Mouse Micronucleus Assay(Not specified)Test article is considered non-mutagenicAcceptance criteria met

Non-Clinical Bench Testing Acceptance Criteria & Performance:

Test DescriptionMethodAcceptance CriteriaReported Device Performance (Conclusion)
Total System LengthTotal system length was measured from the distal tip of the distal marker to the proximal tip of the delivery system.The stent length should be adequate for its intended use for delivery via standard microcatheters.Acceptance criteria met
Marker-to-Marker LengthMarker-to-marker length is representative of the total stent length. Total stent length was measured 100% in process.The stent length should be adequate for its intended use for delivery via standard microcatheters.Acceptance criteria met
Delivery ForceDelivery force was measured through a representative tortuous anatomical model.The stent must be below delivery force specification.Acceptance criteria met
Resheathing ForceResheathing force was measured through a representative tortuous anatomical model.The stent must be below re-sheathing force specification.Acceptance criteria met
Multiple Resheathing DurabilityMultiple resheathing durability was evaluated on the ability to withstand delivery and withdrawal forces in a representative tortuous model beyond the recommended number of passes and re-sheathings allowed per the Instructions for Use.The stent must be able to be deployed and resheathed up to four times.Acceptance criteria met
Body Finger Marker Coil TensileBody finger marker coil tensile was performed to verify the strength of the laser weld of the Platinum/Iridium marker coil to the Nitinol body finger.The body finger marker coil tensile should be greater than or equal to existing tensile strength specification.Acceptance criteria met
Radial ForceRadial force was measured 100% in process.The stent must be within existing radial force specification.Acceptance criteria met
RadiopacityRadiopacity was verified by both a material and dimensional analyses.The radiopaque body markers must be visible using standard catheter laboratory equipment.Acceptance criteria met
Ease of Use (Fluorosafe Marker Location)Ease of use (fluorosafe marker location) was measured from the distal tip of the distal marker to the distal end of the fluorosafe marker.Less required fluoroscopy is better than more required fluoroscopy.Acceptance criteria met
ParticulateParticulate was evaluated for generation under simulated use in a representative tortuous anatomical model.The stent must not generate clinically unacceptable particulate.Acceptance criteria met

Clinical Performance (Leveraged from STRATIS Registry 4x40 Subgroup):

Confirmatory Clinical DataAcceptance Criteria (Implied)Reported Device Performance
Safety Data:
Symptomatic ICHAs low as reasonably achievable, comparable to predicate2.8% (7/252)
Mortality at 90 daysAs low as reasonably achievable, comparable to predicate16.2% (48/296)
Study Device-Related SAEsAs low as reasonably achievable, comparable to predicate0.3% (1/296)
Performance Data:
Successful revascularization (TICI 2b-3)High success rate, comparable to predicate90.4% (227/251)
Functional Independence (mRS 0-2) at 90 daysHigh rate, comparable to predicate60.3% (167/277)

2. Sample Size Used for the Test Set and Data Provenance

  • Non-Clinical Animal Testing: The study used an "in-vivo Yorkshire cross swine model." The number of animals is not explicitly stated, but it evaluated 4 out of 8 proposed target sites (Right Vertebral, Left Vertebral, Right Internal Thoracic, Left Internal Thoracic, Right Costocervical, Left Costocervical, Renal, Ascending Pharyngeal).
    • Provenance: Prospective, animal model.
  • Confirmatory Clinical Data (Leveraged from STRATIS Registry):
    • Sample Size: A total of 984 patients were enrolled in the STRATIS Registry. From this, the analysis for the 4x40 Solitaire devices (predicate and reference) included 296 patients who were treated only with these devices.
    • Provenance: Prospective, multi-center, observational registry conducted in the United States (55 sites).

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

  • Animal Testing: The document does not specify the number of experts or their qualifications for evaluating the animal testing results (e.g., Luminal Thrombus, Endothelial Cell Coverage, etc.).
  • Clinical Data (STRATIS Registry):
    • Clinical Events Committee: An independent Clinical Events Committee was used for adverse event adjudication. The number of experts or their specific qualifications (e.g., type of physician, years of experience) are not detailed.
    • Core Lab: An independent Core Lab was used for procedural image assessment. The number of experts or their specific qualifications are not detailed.

4. Adjudication Method for the Test Set

  • Clinical Data (STRATIS Registry): The document states that an "independent Clinical Events Committee for adverse event adjudication and an independent Core Lab for procedural image assessment" were used. This indicates an independent adjudication process, but the specific method (e.g., 2+1, 3+1 consensus) is not detailed.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

  • No, an MRMC comparative effectiveness study was not done as described for AI assistance. The clinical data leveraged was from an observational registry evaluating clinical outcomes of the device in humans. This study was not comparing human readers with and without AI assistance; rather, it was assessing the device's performance in a real-world clinical setting.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

  • Not applicable. This document describes a medical device (revascularization device), not an algorithm or AI software. Therefore, there is no "standalone" algorithm-only performance to report. The performance discussed for the clinical data is the device's performance in situ within a human patient, with clinicians performing the procedure.

7. The Type of Ground Truth Used

  • Biocompatibility: In vitro and in vivo laboratory test results (e.g., cell viability, immune responses, clotting times, mutagenicity).
  • Bench Testing: Engineering measurements and evaluations against design specifications.
  • Animal Testing: Histopathological and physiological observations in an animal model (e.g., luminal thrombus, endothelial cell coverage, inflammation).
  • Clinical Data (STRATIS Registry):
    • Adverse Events: Adjudicated by an independent Clinical Events Committee, likely based on medical records, clinical observations, and definitions of adverse events.
    • Procedural Image Assessment: Adjudicated by an independent Core Lab, likely measuring revascularization success (TICI scores) from imaging data.
    • Functional Independence (mRS 0-2): Patient outcomes measured by modified Rankin Scale (mRS) at 90 days.

8. The Sample Size for the Training Set

  • Not applicable. For this type of medical device (physical revascularization device), there is no "training set" in the context of an AI/ML algorithm. The "training set" concept is relevant for data-driven models that learn from data.

9. How the Ground Truth for the Training Set Was Established

  • Not applicable. As there is no training set for an AI/ML algorithm, this question is not relevant to the provided document.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized image of an eagle with three heads, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the eagle. The eagle is depicted in black and white, and the text is in a simple, sans-serif font.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 8, 2017

Micro Therapeutics, Inc. d/b/a ev3 Neurovascular Jennifer Correa Senior Regulatory Affairs Specialist 9775 Toledo Way Irvine, California 92618

Re: K160641

Trade/Device Name: Solitaire™ Platinum Revascularization Device (6x40 mm) Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: NRY Dated: February 16, 2017 Received: February 21, 2017

Dear Ms. Correa:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

William J.
Heetderks -S
Digitally signed by William J. Heetderks -
DN: c=US, o=U.S. Government, ou=HHS,
ou=FDA, ou=People,
0.9.2342.19200300.100.1.1=0010149848
cn=William J. Heetderks -S
Date: 2017.03.08 16:48:39-05'00'

for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K160641

Device Name

Solitaire™ Platinum Revascularization Device (6x40 mm)

Indications for Use (Describe)

The Solitaire™ Platinum Revascularization Device is intended to restore blood flow by removing thrombus from a large intracranial vessel in patients experiencing ischemic stroke within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.

Type of Use (Select one or both, as applicable)
× Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SummaryK160641
510(k) Owner:Micro Therapeutics, Inc. d/b/a ev3 Neurovascular9775 Toledo WayIrvine, CA 92618Establishment Registration No. 2029214
Contact Person:Jennifer CorreaSenior Specialist, Regulatory AffairsTelephone: (949) 297-9563E-mail: Jennifer.L.Correa@medtronic.com
Date SummaryPrepared:March 1, 2017
Trade Name ofDevice:SolitaireTM Platinum Revascularization Device (6x40 mm)
Common Name ofDevice:Catheter, Thrombus Retriever
Classification ofDevice:21 CFR 870.1250 - Class II
Product Code:NRY
Predicate Device:SolitaireTM Platinum Revascularization Device510(k)#: K153071
Performance Data:The following non-clinical bench testing and confirmatoryclinical data supports the subject 6-40-10 SolitaireTM PlatinumRevascularization Device:• Total System Length• Marker-to-Marker Length• Delivery Force• Resheathing Force• Multiple Resheathing Durability• Body Finger Marker Coil Tensile• Radial Force• Radiopacity• Ease of Use (Fluorosafe Marker Location)• Particulate• Leveraged confirmatory clinical data using the cleared4x40 Solitaire devices
Conclusion:The materials of construction, design, and manufacturing/packaging processes for the subject 6-40-10 SolitaireTM PlatinumRevascularization Device are identical to the predicate SolitaireTMPlatinum Revascularization Device (K153071) therefore;biocompatibility, sterilization, packaging, and shelf lifevalidations have been leveraged or adopted. Non-clinical animaltesting was performed to demonstrate safety, effectiveness, and

{4}------------------------------------------------

usability of the subject 6-40-10 Solitaire™ Platinum Revascularization Device. Non-clinical bench and animal testing confirmed that the subject 6-40-10 Solitaire™ Platinum Revascularization Device does not introduce new risks or increase existing risks. Confirmatory clinical data using the cleared 4x40 Solitaire devices (predicate 4-40-10 Solitaire™ Platinum and reference 4-40 Solitaire™ 2) was leveraged to support substantial equivalence of the subject 6-40-10 Solitaire™ Platinum Revascularization Device. The subject Solitaire™ Platinum Revascularization Device should be determined to be substantially equivalent to the predicate Solitaire™ Platinum Revascularization Device (K153071) based on non-clinical bench and animal testing, confirmatory clinical data, similarities in design, principles of operation, and indications for use.

Device Description:

The subject 6-40-10 Solitaire™ Platinum Revascularization Device is designed to restore blood flow in patients experiencing ischemic stroke due to large intracranial vessel occlusion. The Solitaire™ Platinum Revascularization Device is designed for use in the neurovasculature such as the Internal Carotid Artery (ICA), M1 and M2 segments of the middle cerebral artery, basilar, and the vertebral arteries. The distal nitinol portion of the Solitaire™ Platinum Revascularization Device facilitates clot retrieval. The Solitaire™ Platinum Revascularization Device has Platinum Iridium radiopaque markers on the working cell length, proximal and distal ends. The subject Solitaire™ Platinum Revascularization Device is a portfolio expansion to the predicate Solitaire™ Platinum Revascularization Device (K153071).

Indications for Use:

The Solitaire™ Platinum Revascularization Device is intended to restore blood flow by removing thrombus from a large intracranial vessel in patients experiencing ischemic stroke within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.

Device Comparison:

The table below provides a comparison of the technological characteristics associated with the portfolio expansion of the subject 6-40-10 Solitaire™ Platinum Revascularization Device and the predicate Solitaire™ Platinum Revascularization Device (K153071).

{5}------------------------------------------------

Predicate DeviceSolitaire™ PlatinumRevascularizationDevice (K153071)Subject Device6-40-10 Solitaire™PlatinumRevascularizationDeviceRationale for Difference(if applicable)
Indication forUseThe Solitaire™PlatinumRevascularizationDevice is intended torestore blood flow byremoving thrombusfrom a largeintracranial vessel inpatients experiencingischemic strokewithin 8 hours ofsymptom onset.Patients who areineligible forintravenous tissueplasminogenactivator (IV t-PA) orwho fail IV t-PAtherapy arecandidates fortreatment.SameN/A
Principles ofOperationThe device is used inthe neurovasculatureto restore blood flowby removingthrombus.SameN/A
Method ofSupplyStored withindispenser coil, Tyvekpouch, and shippingcarton.SameN/A
SterilizationMethodEthylene OxideSameN/A
Predicate DeviceSolitaire™ PlatinumRevascularizationDevice (K153071)Subject Device6-40-10 Solitaire™PlatinumRevascularizationDeviceRationale for Difference(if applicable)
Device Size(s)4-20-104-40-106-20-106-40-10The subject 6-40-10Solitaire™ PlatinumRevascularization Deviceshares similar technologicalcharacteristics to thepredicate 6-20-10 (6 mmstent diameter) and the 4-40-10 (40 mm stent length) ofthe predicate Solitaire™Platinum RevascularizationDevices. Bench testing hasdemonstrated that the subject6-40-10 Solitaire™ PlatinumRevascularization Device iswithin the existing ranges ofthe 4x40 Solitaire devices fordevice length, radial outwardforce and delivery /resheathing forces.
StentNitinol.SameN/A
Push-wireNitinol.SameN/A
Distal MarkerCoil90% Platinum/10% IridiumSameN/A
Body FingerMarker Coil90% Platinum/10% IridiumSameN/A
Push-wireShrink TubingPTFESameN/A
IntroducerSheathPTFE/GrilamidSameN/A

{6}------------------------------------------------

Sterilization and Shelf Life:

The subject 6-40-10 Solitaire™ Platinum Revascularization Device is sterilized using the identical validated, Ethylene Oxide (EO) sterilization cycle as the predicate Solitaire™ Platinum Revascularization Device (K153071). The materials of construction, design, manufacturing process, packaging, and sterile load configurations are identical to the predicate Solitaire™ Platinum Revascularization Device (K153071). Therefore, no additional sterilization validation testing is required.

The subject 6-40-10 Solitaire™ Platinum Revascularization Device and subsequent packaging remain functional and maintain sterility for up to two (2) years. The materials

{7}------------------------------------------------

of construction, design, manufacturing process, packaging, and sterile load configurations are identical to the predicate Solitaire™ Platinum Revascularization Device (K153071). Therefore, no additional shelf life validation testing is required.

Biocompatibility:

The subject 6-40-10 Solitaire™ Platinum Revascularization Device does not introduce any new materials into the finished device nor the manufacturing processes. The material of the subject 6-40-10 Solitaire™ Platinum Revascularization Device is identical to the material of the predicate Solitaire™ Platinum Revascularization Device (K153071). Biocompatibility data for the Solitaire device family was tested for the Solitaire™ FR Revascularization Device. The biocompatibility data for the Solitaire™ FR was adopted for the subject 6-40-10 Solitaire™ Platinum Revascularization Device. A summary of the biocompatibility testing completed for the Solitaire™ FR Revascularization Device includes:

TestCategoryTestDescriptionMethodAcceptance CriteriaConclusion
CytotoxicityL929 MTTCytotoxicityISO 10993-5Viability is $\geq$ 70%.Acceptance criteriamet
SensitizationGuinea PigMaximizationSensitizationISO 10993-10Test article does notelicit a sensitizationresponse.Acceptance criteriamet
IrritationIntracutaneousIrritation TestISO 10993-10Differences in themean test and controlscores of the extractdermal observationsare $<$ 1.0.Acceptance criteriamet
AcuteSystemicToxicityAcute SystemicInjection TestISO 10993-11No abnormal clinicalsigns and weight lossin excess of 10%.Temperature rise$\geq$ 0.5°CAcceptance criteriamet
MaterialsMediatedRabbit Pyrogen
TestCategoryTestDescriptionMethodAcceptance CriteriaConclusion
Hemo-compatibilityHemolysisISO 10993-4No significant differences between the test article extract and negative control article results. The test article is considered non-hemolyticAcceptance criteria met
Partial Thromboplastin TimeClotting times are similar to the negative control and the reference material (HDPE) indicating the device materials are not an activator of the intrinsic coagulation pathway.
Platelet and Leukocyte CountTest article does not adversely affect the platelet and leukocyte components of the blood compared to the reference material.
Complement Activation C3a and SC5b-9 AssayLevels of C3a and SC5b-9 are comparable and less than the positive control.
ThrombosisThrombo-resistance properties are acceptable in clinical use.
GenotoxicityBacterial Mutagenicity TestISO 10993-3Test article is considered non-mutagenicAcceptance criteria met
In-vitro Mouse Lymphoma AssayTest article is considered non-mutagenic
In-vivo Mouse Micronucleus AssayTest article is considered non-mutagenic

{8}------------------------------------------------

{9}------------------------------------------------

Performance Data – Bench:

A summary of the non-clinical bench testing performed for the subject 6-40-10 Solitaire™ Platinum Revascularization Device is summarized in the table below.

Test DescriptionMethodAcceptance CriteriaConclusion
Total System LengthTotal system length was measured from the distal tip of the distal marker to the proximal tip of the delivery system.The stent length should be adequate for its intended use for delivery via standard microcatheters.Acceptance criteria met
Marker-to-Marker LengthMarker-to-marker length is representative of the total stent length. Total stent length was measured 100% in process.The stent length should be adequate for its intended use for delivery via standard microcatheters.Acceptance criteria met
Delivery ForceDelivery force was measured through a representative tortuous anatomical model.The stent must be below delivery force specification.Acceptance criteria met
Resheathing ForceResheathing force was measured through a representative tortuous anatomical model.The stent must be below re-sheathing force specification.Acceptance criteria met
Multiple Resheathing DurabilityMultiple resheathing durability was evaluated on the ability to withstand delivery and withdrawal forces in a representative tortuous model beyond the recommended number of passes and re-sheathings allowed per the Instructions for Use.The stent must be able to be deployed and resheathed up to four times.Acceptance criteria met
Body Finger Marker Coil TensileBody finger marker coil tensile was performed to verify the strength of the laser weld of the Platinum/Iridium marker coil to the Nitinol body finger.The body finger marker coil tensile should be greater than or equal to existing tensile strength specification.Acceptance criteria met
Test DescriptionMethodAcceptance CriteriaConclusion
Radial ForceRadial force was measured 100% in process.The stent must be within existing radial force specification.Acceptance criteria met
RadiopacityRadiopacity was verified by both a material and dimensional analyses.The radiopaque body markers must be visible using standard catheter laboratory equipment.Acceptance criteria met
Ease of Use(Fluorosafe Marker Location)Ease of use (fluorosafe marker location) was measured from the distal tip of the distal marker to the distal end of the fluorosafe marker.Less required fluoroscopy is better than more required fluoroscopy.Acceptance criteria met
ParticulateParticulate was evaluated for generation under simulated use in a representative tortuous anatomical model.The stent must not generate clinically unacceptable particulate.Acceptance criteria met

{10}------------------------------------------------

Performance Data - Animal:

Non-clinical animal testing was conducted to verify the performance of the subject 6-40-10 Solitaire™ Platinum Revascularization Device. Non-clinical animal testing was conducted to verify the subject 6-40-10 Solitaire™ Platinum Revascularization Device is safe and effective in an in-vivo Yorkshire cross swine model. Four (4) of the following eight (8) proposed target sites were evaluated:

  • Right Vertebral ●
  • Left Vertebral ●
  • Right Internal Thoracic ●
  • Left Internal Thoracic ●
  • Right Costocervical ●
  • Left Costocervical ●
  • Renal ●
  • Ascending Pharyngeal

Please note the renal and ascending pharyngeal are considered alternative back-up sites depending on the size of the swine, distal target zone diameter, length, consistency of vessel diameter, and the ability to harvest the vessel post procedure.

{11}------------------------------------------------

Non-clinical animal testing was evaluated at 0-day (acute) and 30-day (chronic) time points. The non-clinical animal testing was designed to show no-clinical inferiority between the control 6-30 Solitaire™ 2 Revascularization Device and the test 6-40-10 Solitaire™ Platinum Revascularization Device. Non-clinical animal testing evaluated the following biological responses of vascular tissue specific to safety:

  • . Luminal Thrombus
  • Endothelial Cell Coverage
  • Inflammation
  • Fibrin
  • Medial Smooth Muscle Cell (SMC) Loss ●
  • Disruption of the Internal Elastic Lamina (IEL) and External Elastic Lamina (EEL)
  • . Adventitia
  • Neointimal Hyperplasia
  • Elastic Fiber Separation ●
  • . Lumen Area Reduction

Performance Testing - Clinical:

Type of data

A clinical study was not performed for the subject 6-40-10 Solitaire™ Platinum Revascularization Device. Confirmatory clinical data using the cleared 4x40 Solitaire devices (predicate 4-40-10 Solitaire™ Platinum and reference 4-40 Solitaire™ 2) was leveraged to support substantial equivalence of the subject 6-40-10 Solitaire™ Platinum Revascularization Device. Leveraging of the 4x40 Solitaire device confirmatory clinical data is appropriate as the subject 6-40-10 Solitaire™ Platinum Revascularization Device is within the existing ranges of the 4x40 Solitaire devices for device length, radial outward force and delivery / resheathing forces. The confirmatory 4x40 Solitaire clinical data was collected as part of the Systematic Evaluation of Patients TReated with Neurothrombectomy Devices for AcuTe IschemicStroke (STRATIS) Registry.

STRATIS Registry Design

A prospective, multi-center, non-randomized, observational registry evaluating the use of Covidien market-released neurothrombectomy devices in patients diagnosed with an acute ischemic stroke due to large intracranial vessel occlusion. The study used an independent Clinical Events Committee for adverse event adjudication and an independent Core Lab for procedural image assessment.

Sample Size

Under the STRATIS Registry, 984 patients were enrolled that were treated with a Medtronic neurothrombectomy device at the discretion of treating physicians and as part of hospital standard practice at 55 sites within the United States. Of the 984 patients treated, a total of 296 were treated using only the cleared 4x40 device. Major patient eligible criteria for consent and enrollment into the Registry were to be treated with the Solitaire device within 8 hours of symptom onset.

{12}------------------------------------------------

Confirmatory Clinical Data

The confirmatory clinical data leveraged from the STRATIS Registry 4x40 group included baseline demographics, safety data and performance data and showed that the cleared 4x40 Solitaire device has a similar clinical performance and safety profile to other cleared Solitaire device dimensions.

STRATIS Registry 4x40 Subgroup
Confirmatory Clinical Data
Baseline Demographics
Age (years)66.4 ± 15.2 (296)67.0 (58.0, 77.0)
NIHSS at baseline16.8 ± 5.5 (296)16.0 (12.0, 21.0)
Male Gender51.4% (152/296)
Safety Data
Symptomatic ICH2.8% (7/252)
Mortality at 90 days16.2% (48/296)
Performance Data
Successfulrevascularization(TICI 2b-3)90.4% (227/251)
FunctionalIndependence (mRS 0-2)at 90 days60.3% (167/277)

Device Related Serious Adverse Events (SAEs)

STRATIS Registry 4x40 SubgroupDevice-Related SAEs
Study Device-RelatedSAEs0.3% (1/296) [1]**

** Subarachnoid hemorrhage

Conclusion

Non-clinical bench, animal testing and the confirmatory clinical data demonstrated that the subject 6-40-10 Solitaire™ Platinum Revascularization Device is substantially equivalent to the predicate Solitaire™ Platinum Revascularization Device (K153071) and does not raise new questions of the safety and effectiveness of the device. Additionally, there are no changes to the indications for use or the fundamental scientific technology of the Solitaire™ Platinum Revascularization Device.

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).