510(k) Data Aggregation

    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    HALYARD® SMART-FOLD Sterilization Wraps are intended to be used to enclose another medical device that is to be sterilized by a healthcare provider using:

    • Pre-vacuum steam at 270°F/132°C for 4 minutes. The wrap was validated for dry times of 30 minutes.
    • 100% ethylene oxide (EO) with a concentration of 725-735 mg/L at 131°F/55°C and 40%-80% relative humidity for 60 minutes. The wrap was validated for aeration times for EO sterilization of 8 hours at 110° F/43.3ºC.

    STERIS V-PRO® Low Temperature Sterilization Systems.

    • · STERIS® V-PRO® 60 (Lumen, Non-Lumen and Flexible Cycles)

    • · STERIS® V-PRO® s2 (Lumen, Non-Lumen and Flexible Cycles)

    • · STERIS® V-PRO® 1 (Lumen Cycle)

    • · STERIS® V-PRO® 1 Plus (Lumen and Non-Lumen Cycle)

    • · STERIS® V-PRO® maX (Lumen, Non-Lumen and Flexible Cycle)

    • · STERIS® V-PRO® maX 2 (Lumen, Non-Lumen and Flexible Cycle)

    • Advanced Sterilization Products STERRAD® Sterilization System

    • STERRAD® 100S

    • · STERRAD® NX®, (Standard Cycle, Advanced Cycle)

    • · STERRAD® NX® with ALLClear® Technology, (Standard Cycle, Advanced Cycle)

    • · STERRAD® 100NX® (Standard Cycle, Flex Cycle, EXPRESS Cycle, DUO Cycle, ULTRA GL Cycle)

    • STERRAD® 100NX® with ALLClear® Technology, (Standard Cycle, EXPRESS Cycle, DUO Cycle, ULTRA GI Cycle)

    HALYARD* ONE-STEP, QUICK CHECK and SEQUENTIAL Sterilization Wraps are intended to be used to enclose another medical device that is to be sterilized by a healthcare provider using:

    · Pre-vacuum steam at 270°F/132°C for 4 minutes. The wrap was validated for dry times of 20 minutes for Models 100 and 200, and for 30 minutes for Models 300, 400, 500, and 600.

    • · 100% ethylene oxide (EO) with a concentration of 725-735 mg/L at 131°F/55°C and 40%-80% relative humidity for 60 minutes. The wrap was validated for aeration times for EO sterilization of 8 hours at 43.3℃.
      · STERIS V-PRO® Low Temperature Sterilization Systems. The wrap was validated to be effectively aerated during the pre-programmed cycles.

    • · STERIS V-PRO® 60 (Lumen, Non-Lumen and Flexible Cycles)

    • · STERIS V-PRO® s2 (Lumen, Non-Lumen and Flexible Cycles)

    • · STERIS V-PRO® 1 (Lumen Cycle)

    • · STERIS V-PRO® 1 Plus (Lumen and Non-Lumen Cycle)

    • · STERIS V-PRO® maX (Lumen, Non-Lumen and Flexible Cycle)

    • · STERIS V-PRO® maX 2 (Lumen, Non-Lumen and Flexible Cycle)

    · Gravity steam at 250°F/121°C for 30 minutes (25 minute dry time for Models 100, 200 and 30 minute dry time for Models 400, 500 and 600)

    • · Advanced Sterilization Products STERRAD® Sterilization System
    • · STERRAD® 50, 100S, and 200
    • · STERRAD® NX®, (Standard Cycle, Advanced Cycle)
    • · STERRAD® NX® with ALLClear® Technology, (Standard Cycle, Advanced Cycle)
    • · STERRAD® 100NX® (Standard Cycle, Flex Cycle, EXPRESS Cycle, DUO Cycle)
    • · STERRAD® 100NX® with ALLClear® Technology, (Standard Cycle, EXPRESS Cycle, DUO Cycle)
    • · STERILUCENT® HC80TT Sterilization Cycles (Lumen and Flexible Cycles)
    • · Stryker Sterizone® VP4 Sterilizer Cycle 1
    Device Description

    HALYARD SEQUENTIAL Sterilization Wrap is supplied to the customer as bulk packages of single sheets, where in accordance with standard hospital practices, two sheets are then used to wrap a medical device or a collection of medical devices for sterilization. HALYARD ONE-STEP, HALYARD QUICK CHECK and HALYARD SMART-FOLD Sterilization Wraps are comprised of two sheets of HALYARD* SEQUENTIAL Sterilization Wrap. This allows for convenient wrapping with two sheets simultaneously.

    The HALYARD ONE-STEP, QUICK CHECK and HALYARD SMART-FOLD Sterilization Wraps are intended to enclose another medical device that is to be sterilized by a healthcare provider.

    AI/ML Overview

    The document describes the regulatory submission for HALYARD sterilization wraps. It focuses on demonstrating substantial equivalence to a predicate device through performance testing. No AI/ML algorithm is involved in the device. Therefore, a complete response is not possible for all aspects of your request.

    Here's a breakdown of the available information:

    1. A table of acceptance criteria and the reported device performance

    PurposeTestAcceptance CriteriaResults
    Sterilant Penetration/EfficacyANSI/AAMI ST79, ANSI/AAMI/ISO 11138-7Achieving a 10^-6 sterility assurance level following processing in a worst-case half-cyclePassed
    Performance Testing (Non-sterile/Sterile)ANSI/AAMI/ISO 11607-1 Annex B, ISO 13938-2, ASTM D4966-12, CPSC 1610Complies with the selected physical propertiesPassed
    Maintenance of Package IntegrityANSI/AAMI/ISO 11607-1, ANSI/AAMI ST79Maintain sterility for 1 year, except for gravity steam (30 days), STERRAD ULTRA GI (3 months) and Sterilucent (6 months).Passed
    BiocompatibilityISO 10993-1, ISO 10993-5, ISO 10993-10, ISO 10993-7Non-cytotoxic, Non-irritating, Non-sensitizingPassed
    ResidualsSteris V-PRO Low Temperature Sterilization System H2O2 residuals, Sterizone VP4 Cycle 1 residual H2O2, ISO 14937 (ULTRA GI Cycle only), ISO 10993-7 EtO and ECHH2O2 ≤ 0.000218 mg/cm2, H2O2 ≤ 0.56 µg/cm2, H2O2 ≤ 9100 µg/cm2, EO < 4 mg, ECH < 9 mgPassed

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    The document does not explicitly state the sample sizes for the test sets of each specific test. It refers to validation loads and specific lumen configurations used in testing, but not the total number of items tested for each criterion. The data provenance is not specified, but the testing would typically be conducted in a laboratory setting under controlled conditions. It's prospective testing as it's for a new or modified device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    This is not applicable as the device is a sterilization wrap, not an AI/ML diagnostic tool requiring expert interpretation for ground truth. The "ground truth" for this type of device is established through scientific and engineering standards for sterility and material properties.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    This is not applicable since there is no human interpretation or subjective assessment involved that would require an adjudication method. The tests are objective and based on measurable physical and chemical properties.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable. The device is a sterilization wrap, not an AI/ML software or a device that impacts human reader performance.

    6. If a standalone (i.e., algorithm only without human-in-the loop performance) was done

    This is not applicable. The device is a physical product (sterilization wrap), not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The ground truth for this device is based on established scientific and regulatory standards for sterilization assurance, material performance, biocompatibility, and residual levels. This includes:

    • Sterility Assurance Level (SAL): A 10^-6 SAL is a universally accepted standard for sterilization, meaning there is a 1 in 1,000,000 chance of a non-sterile item.
    • Physical Properties: Standards like ISO 11607-1, ISO 13938-2, ASTM D4966-12, and CPSC 1610 define acceptable limits for the physical integrity and strength of packaging materials.
    • Biocompatibility: ISO 10993 series standards define acceptable levels for cytotoxicity, irritation, and sensitization to ensure the materials are safe for patient contact.
    • Residuals: Specific thresholds for residual sterilants (e.g., H2O2, EtO, ECH) are defined by ISO standards and validated for safety.

    8. The sample size for the training set

    This is not applicable as the device is a physical product and does not involve AI/ML requiring a training set.

    9. How the ground truth for the training set was established

    This is not applicable as the device is a physical product and does not involve AI/ML requiring a training set with established ground truth.

    Ask a Question

    Ask a specific question about this device

    K Number
    K143053

    Validate with FDA (Live)

    Date Cleared
    2015-04-03

    (162 days)

    Product Code
    Regulation Number
    880.6850
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K082554,K082177

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    KIMGUARD* ONE-STEP* Sterilization Wrap (KC100, KC200) are intended to be used to enclose another medical device that is to be sterilized by a healthcare provider using:

    · Pre-vacuum steam at 270°F / 132°C for 4 minutes. The wrap was validated for dry times of 20 minutes for models KC100 and KC200.

    · 100% ethylene oxide (EO) with a concentration of 725-735 mg/L at 131°F / 55°C and 40-80% relative humidity for 60 minutes. The wrap was validated for aeration times for EO sterilization of 8 hours at 43.3 °C for models KC100 and KC200.

    KIMGUARD* ONE-STEP* Sterilization Wrap (KC100, KC200) are intended to allow sterilization of the enclosed medical device(s) and also maintain sterility of the enclosed device(s) until used.

    Test results validated that KIMGUARD* ONE-STEP* Sterilization Wraps (KC100, KC200) allowed sterilization of the enclosed devices by the ethylene oxide sterilization and by pre-vacuum cycles.

    These models of the KIMGUARD* ONE-STED* Sterilization Wrap have been validated for use with the ethylene oxide and pre-vacuum cycles in Table 1 (see page 2 of 2)

    Device Description

    KIMGUARD* ONE-STEP* Sterilization Wrap is comprised of two sheets of KIMGUARD* Sterilization Wrap ultrasonically seamed on two edges. This allows for convenient wrapping with two sheets simultaneously.

    The sheets of sterilization wrap are square or rectanqular fabric produced using a three-layer SMS (spunbond-meltblown-spunbond) process. The wrap fabric (white or blue color) is composed of polypropylene with the addition of less than 2% by weight of phthalocvanine blue pigment (blue fabric only), less than 1% by weight titanium dioxide pigment, and less than 0.009% by weight of a potassium phosphate anti-static treatment. The wrap allows a sterilized package to be opened aseptically.

    AI/ML Overview

    This FDA 510(k) summary describes KIMGUARD® ONE-STEP® Sterilization Wrap intended to enclose medical devices for sterilization and maintain sterility.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (Measured Performance)Reported Device Performance (Results)
    Maintain Sterility (for Pre-vacuum steam sterilization at 270°F / 132°C for 4 minutes with 20 minutes dry time)Passed
    Maintain Sterility (for 100% ethylene oxide (EO) at 725-735 mg/L, 131°F / 55°C, 40-80% RH for 60 minutes with 8 hours aeration at 43.3°C or 12 hours at 55°C)Passed
    Maintenance of 365-Day Package Integrity (sterility maintenance)Passed
    Ethylene Oxide Sterilant PenetrationPassed
    Pre-Vacuum Steam Sterilant PenetrationPassed
    Post Sterilization Biocompatibility Testing (Cytotoxicity, ISO 10993-5: 2009; Irritation, ISO 10993-10:2010; Sensitization, ISO 10993-10: 2010)Passed
    Maximum Wrapped Package Content Weights for KC100 (Very Light Weight Package)Up to 3 lbs. (Tested with 16 huck towels (17"x 29"))
    Maximum Wrapped Package Content Weights for KC200 (Light Weight Package)Up to 6 lbs. (Tested with 2 huck towels (17"x 29"), 2 fluid resistant U-drapes (68"x 109"), 1 fluid resistant universal bar drape (70" x 108"))

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not explicitly state the specific sample sizes used for the "Maintenance of 365-Day Package Integrity," "Sterilant Penetration," and "Biocompatibility" tests. For the KIMGUARD® ONE-STEP® Sterilization Wrap Recommendations for Use (Table 1), the "test set" can be considered the specific configurations of loads used to validate the maximum wrapped package content weights:

    • KC100: 16 huck towels (17"x 29") weighing 3 lbs.
    • KC200: 2 huck towels (17"x 29"), 2 fluid resistant U-drapes (68"x 109"), 1 fluid resistant universal bar drape (70" x 108") weighing 6 lbs.

    The data provenance is not explicitly mentioned as country of origin, but the applicant is Halyard Health, located in Alpharetta, GA, USA. The studies appear to be prospective as they were conducted to validate the device's performance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This product is a sterilization wrap, not an AI/imaging device. Therefore, the concept of "experts" to establish ground truth in the context of diagnostic interpretation is not applicable here. The validation involves standardized performance tests for sterilization and material properties. Compliance with relevant standards (e.g., ISO 10993 for biocompatibility) would be performed by qualified personnel in testing laboratories.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable for this type of device and study. Adjudication methods are typically used in clinical studies involving interpretation by multiple readers.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI/imaging device.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This is not an AI software. The device itself (the sterilization wrap) has a standalone performance (its ability to maintain sterility and allow sterilant penetration). The nonclinical tests essentially evaluate this "standalone" performance.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for this device's performance is established by:

    • Quantitative measurements and observations based on standardized test methods for:
      • Sterilant penetration (ensuring the sterilant reaches the enclosed medical device).
      • Microbiological challenge (implied by "maintenance of sterility" tests, indicating no microbial ingress).
      • Physical integrity tests to confirm the wrap remains intact over time, preventing contamination.
      • Biocompatibility tests according to ISO standards, which involve specific physiological responses.

    8. The sample size for the training set

    Not applicable. This is not an AI/machine learning device that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K141071

    Validate with FDA (Live)

    Date Cleared
    2014-07-28

    (94 days)

    Product Code
    Regulation Number
    880.6850
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K082177, K091685

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    KIMGUARD* ONE-STEP* Sterilization Wrap (KC100, KC200, KC300, KC400, KC500 and KC600) are intended to enclose another medical device that is to be sterilized by a healthcare provider using:

    • Gravity Steam at 250°F/121°C for 30 minutes. .
      KIMGUARD* ONE-STEP* Sterilization Wrap (KC100, KC200, KC300, KC400, KC500 and KC600) are intended to allow sterilization of the enclosed medical device(s) and also maintain sterility of the enclosed device(s) until used.
    Device Description

    KIMGUARD* ONE-STEP* Sterilization Wrap is comprised of two sheets of KIMGUARD* Sterilization Wrap that is ultrasonically seamed on two edges. This seamed configuration allows for convenient wrapping of an article using two sheets simultaneously.

    The sheets of sterilization wrap are square or rectangular fabric produced using a threelayer SMS (spunbound-meltblown-spunbound) process. The wrap fabric is composed of polypropylene with the addition of less than 2% by weight of phthalocyanine blue pigment, less than 1% by weight titanium dioxide pigment, and less than 0.008% by weight of antistatic treatment. The wrap allows a sterilized package to be opened aseptically.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the KIMGUARD ONE-STEP Sterilization Wrap:

    Acceptance Criteria and Device Performance for KIMGUARD® ONE-STEP® Sterilization Wrap

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategoryReported Device Performance
    Sterilization Efficacy (Gravity Steam)Passed
    Sterilant PenetrationPassed
    Package Integrity/Physical PropertiesPassed
    Drying and AerationPassed
    Maintenance of Package IntegrityPassed
    Biocompatibility (pre- and post-sterilization)Passed
    Maintenance of Package Sterility (Duration)30 days

    Note: The document states that "All results of testing met the same acceptance criteria as the predicate devices demonstrating substantial equivalence to the predicate devices." While specific, numerical acceptance criteria are not explicitly detailed (e.g., minimum tensile strength), the "Passed" designation for each category indicates successful fulfillment of the established criteria, which are identical to those of the predicate devices (K082177 and K091685). The specific loads and weights for each wrap model also represent acceptance criteria for their respective uses.


    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the specific numerical sample size used for the test sets (e.g., how many wraps were tested for sterilant penetration). However, it lists the types of loads used in the "Sterility Maintenance Validation Study" for each wrap model:

    • KC100: 1 tray liner 20" x 25", 12.5" x 9" x 1" Tray, 1 lb of metal mass
    • KC200: 1 tray liner 20" x 25", 10" x 20" x 3 ½ Tray, 3 lbs of metal mass
    • KC300: 1 tray liner 20" x 25", 10" x 20" x 3 ½Tray, 6 lbs of metal mass
    • KC400: 1 tray liner 20" x 25", 10" x 20" x 3 ½ Tray, 10 lbs of metal mass
    • KC500: 1 tray liner 20" x 25", 11" x 22" x 3 ½Tray, 12 lbs of metal mass
    • KC600: 1 tray liner 20" x 25", 11" x 22" x 3 ½Tray, 20 lbs of metal mass

    The document does not provide information on the country of origin of the data or whether the study was retrospective or prospective. It is implied to be prospective testing conducted for regulatory submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the document. The tests described are nonclinical, laboratory-based performance tests rather than studies requiring human expert interpretation for ground truth establishment.

    4. Adjudication Method (e.g., 2+1, 3+1, none) for the Test Set

    This information is not applicable as the tests are nonclinical performance evaluations rather than interpretive studies requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

    No, an MRMC comparative effectiveness study was not conducted. This type of study is typically for evaluating diagnostic or interpretive AI systems, not packaging materials.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    This is not applicable. The device is a physical sterilization wrap, not an algorithm or AI system. The performance evaluation is inherently "standalone" in the sense that it assesses the physical product's characteristics and functionality.

    7. The Type of Ground Truth Used

    The ground truth for the nonclinical tests appears to be established scientific and regulatory standards/methods for sterilization packaging performance. This would include methods for assessing sterilant penetration, package integrity (e.g., burst strength, seal integrity), drying, aeration, and maintenance of sterility (e.g., microbial barrier testing). The "results of testing met the same acceptance criteria as the predicate devices," implying a comparison against established performance benchmarks for similar products.

    8. The Sample Size for the Training Set

    This is not applicable. The KIMGUARD ONE-STEP Sterilization Wrap is a physical medical device, not an AI or Machine Learning algorithm that requires a "training set" in the computational sense.

    9. How the Ground Truth for the Training Set Was Established

    This is not applicable for the reasons stated above.