Found 4 results

510(k) Data Aggregation

    K Number
    K140467

    Validate with FDA (Live)

    Device Name
    FLEX-MT +
    Manufacturer
    Date Cleared
    2014-10-07

    (225 days)

    Product Code
    Regulation Number
    890.5850
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K083030, K021100

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    As prescribed by a physician for the following:

    TENS- Transcutaneous Nerve Stimulation

    • Symptomatic relief of chronic intractable pain .
    • . Post traumatic and post surgical pain relief

    EMS- Electrical Muscle Stimulation

    • . Relaxation of muscle spasm
    • . Increasing local blood circulation
    • . Muscle re-education
    • . Prevention or retardation of disuse atrophy
    • Prevention of venous thrombosis of the calf muscles immediately after surgery
    • . Maintaining or increase range of motion
    Device Description

    The Flex-MT + is a combination TENS and EMS device which delivers nerve or muscle stimulation by applying an electrical current to electrodes, which are attached on the patient's skin. The output and waveform is adjustable according to the intended treatment of patient. The stimulator has 2 output channels, accessed through jacks at the top of the housing, so that it may stimulate either 2 or 4 patient electrodes. The device is powered by 700 mAh 4.8V Ni MH rechargeable battery pack. A patient compliance timer can memorize 60 sets of operation records; the total recordable time is 999 hours.

    AI/ML Overview

    The provided text is a 510(k) summary for the Flex-MT+ device. It outlines the device's technical specifications and compares them to predicate devices to establish substantial equivalence.

    Here's an analysis of the acceptance criteria and the "study" (in this case, performance testing) that proves the device meets them, based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state "acceptance criteria" for performance, but rather demonstrates substantial equivalence by comparing the Flex-MT+ device's technical characteristics and performance to two predicate devices (K083030 EMSI TENS-EMS-14 (Flex MT) and K021100 EMPI 300 PV). The "Comment" column in the provided tables effectively acts as the "acceptance criteria" by stating whether the subject device's performance is comparable to or within acceptable limits of the predicate devices and relevant standards.

    Acceptance Criteria (Derived) and Reported Device Performance (Flex-MT+)

    CharacteristicAcceptance Criteria (Derived from Predicates & Standards)Reported Device Performance (Flex-MT+)Fulfilled?
    Indications for UseSame as predicate TENS and NMES modes.Same as K083030 EMSI TENS-EMS-14 (Flex MT) and K021100 EMPI 300 PVYes
    Power SourceBattery operated; battery charger available (consistent with EMPI predicate).700 mAh 4.8V Ni-MH, rechargeable battery pack and chargerYes
    Patient Leakage Current< 1 microamp (consistent with EMSI predicate).< 1 microampYes
    Number of Output Modes2 (consistent with EMSI predicate).2Yes
    Number of Output Channels2 (consistent with both predicates).2Yes
    Method of Channel IsolationTransformer (consistent with both predicates).TransformerYes
    Regulated Current or Voltage?Regulated Voltage (consistent with EMSI predicate).Regulated VoltageYes
    Software/Firmware/MicroprocessorYes (consistent with both predicates).YesYes
    Automatic Overload Trip?Provides additional safety (improved over EMSI predicate).YesYes
    Automatic Shut Off?Yes (consistent with both predicates).YesYes
    Patient Override Control?Yes (consistent with both predicates).YesYes
    Indicator Display (On/Off, Low Bat, Voltage/Current)Yes (consistent with both predicates).Yes (1-10 bars displayed for Voltage/Current Level)Yes
    Timer Range (minutes)Minor difference from EMPI predicate is acceptable and does not affect safety/effectiveness.5-90 minutes, or continuousYes
    Compliance with Voluntary StandardsANSI/AAMI ES1-1993, IEC 60601-1-2 as applicable (consistent with EMSI predicate).ANSI/AAMI ES1-1993, IEC 60601-1-2 as applicableYes
    Compliance with 21 CFR 898?Yes (consistent with both predicates).YesYes
    Weight & DimensionsMinor differences do not change device performance.156 g (with battery), 12cm x 5.4cm x 3.3cmYes
    Housing MaterialsPlastic (consistent with both predicates).PlasticYes
    Waveform (TENS/EMS)Asymmetrical biphasic / Biphasic symmetrical (consistent with both predicates).Asymmetrical biphasic / Biphasic symmetricalYes
    Waveform ShapeRectangular (consistent with both predicates).RectangularYes
    Maximum Output VoltageWithin the range of predicates and acceptable limits of IEC 60601-1.43.2V @ 500ΩYes
    Maximum Output CurrentWithin the range of predicates and acceptable limits of IEC 60601-1.86.4mA @ 500ΩYes
    Pulse Width per phaseSame or increased selection as predicates.50-400µsec (consistent with EMPI predicate)Yes
    Pulse FrequencySame available frequencies as predicates.2-150 HzYes
    Net Charge ($\mu$C per pulse)Within acceptable limits of IEC 60601-1.14.6 @ 500Ω (TENS), 0 (EMS)Yes
    Maximum Current Density (mA/cm²)Within acceptable limits of IEC 60601-1.1.83 @500Ω (TENS), 2.82 @500Ω (EMS)Yes
    Maximum Power Density (W/cm²)Within acceptable limits to avoid thermal burn.0.000858W/cm2 @500ΩYes
    Additional FeaturesPatient compliance timer, battery charger (consistent with predicates).Patient compliance timer, battery chargerYes

    2. Sample size used for the test set and the data provenance

    The document explicitly states: "No clinical study data is provided in support of this submission."
    Therefore, there is no "test set" in the sense of clinical data from patients. The evaluation is based on technical, electrical, and mechanical performance testing, likely conducted on a single or a few units of the device to verify its specifications. The document doesn't specify the sample size for these technical performance tests. The data provenance would be laboratory testing by the manufacturer (EMSI).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. Since no clinical study or test set involving human interpretation of medical data was conducted, no experts were used to establish ground truth in that context. The "ground truth" for the technical performance is derived from recognized industry standards (IEC, ANSI/AAMI) and the specifications of the predicate devices.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable, as there was no test set requiring human adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a powered muscle stimulator/transcutaneous nerve stimulator, not an AI-powered diagnostic or assistive tool for human readers. No MRMC study was performed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a stimulator, not an algorithm, and its performance is assessed based on its physical and electrical outputs, not an algorithmic standalone performance.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for evaluating the Flex-MT+ device's technical characteristics and performance is based on:

    • Industry Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10, IEC 60601-1-11, EN 60601-1-4, EN 62304, and ANSI/AAMI ES1-1993. The device's compliance with these standards serves as a form of "ground truth" for safety and basic performance.
    • Predicate Device Specifications: The technical specifications (e.g., maximum output voltage, current, pulse width, frequency) of the legally marketed predicate devices (K083030 EMSI TENS-EMS-14 (Flex MT) and K021100 EMPI 300 PV) serve as the benchmark for substantial equivalence. The argument is that if the new device's specifications are within acceptable ranges of previously cleared devices, it is substantially equivalent.

    8. The sample size for the training set

    Not applicable. This device is not an AI/machine learning model, so there is no training set.

    9. How the ground truth for the training set was established

    Not applicable. As there is no training set for an AI/ML model, there is no ground truth established for a training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K131917

    Validate with FDA (Live)

    Date Cleared
    2014-04-29

    (307 days)

    Product Code
    Regulation Number
    890.5850
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K121369, K021100

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Indications for FES waveform:

    • Stimulation of the muscles in the leg and ankle of partially paralyzed patients to provide flexion of the foot and thus improve the patient's gait.

    Indications for TENS waveform:

    • Symptomatic relief of chronic (long term) intractable pain
    • Symptomatic relief of post-traumatic acute pain and post surgical pain

    Indications for Russian waveform:

    • Temporary relaxation of muscle spasms
    • Prevention or retardation of disuse atrophy in post-injury type conditions
    • Increase local blood circulation
    • Muscle re-education
    • Maintaining or increasing range of motion
    Device Description

    The device NEURODYN II is a four output channel stimulator, operated in power supply 100 to 240 V, 50-60 Hz, with independent controls, graphic display backlit blue, mechanical contact keyboard and ABS cabinet with acrylic cover.
    Used in the following electrical current therapies: TENS (Transcutaneous Electrical Nerve Stimulation) FES (Functional Electrical Stimulation) RUSSIAN CURRENT (Burst Modulated Medium Frequency)
    Characteristics:
    TENS: four 120mA peak to peak channels FES: four 120mA peak to peak channels RUSSIAN: four 120mA peak to peak channels Electrical Feed: 100 a 240V (50/60Hz) Input Power: 85VA Temperature Range During Transport and Storage: 41 to 122°F Environment Operating Temperature Range: 41 to 113°F

    The device NEURODYN COMPACT is a two output channel stimulator. operated in power supply 100 to 240 V, 50-60 Hz, with independent controls, graphic display backlit blue, mechanical contact keyboard and ABS cabinet with acrylic cover.
    Used in the following electrical current therapies: TENS (Transcutaneous Electrical Nerve Stimulation) FES (Functional Electrical Stimulation) RUSSIAN CURRENT (Burst Modulated Medium Frequency)
    Characteristics:
    TENS: four 120mA peak to peak channels FES: four 120mA peak to peak channels RUSSIAN: four 120mA peak to peak channels Electrical Feed: 100 a 240V (50/60Hz) Input Power: 85VA Temperature Range During Transport and Storage: 41 to 122°F Environment Operating Temperature Range: 41 to 113°F

    AI/ML Overview

    The provided submission for the NEURODYN Compact and NEURODYN II devices does not include a detailed study proving the devices meet specific quantitative acceptance criteria. Instead, it relies on demonstrating substantial equivalence to previously cleared predicate devices (K121369 Neurodyn/Aussie Powered Muscle Stimulator by Ibramed and K021100 300 PV Complete Electrotherapy System by EMPI) based on similar indications for use, operating principles, hardware configuration, and compliance with recognized electrical safety and performance standards.

    The acceptance criteria are implicitly defined by compliance with these standards and the established performance characteristics of the predicate devices.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaNEURODYN Compact / NEURODYN II Performance
    Safety StandardsIEC 60601-1 (General requirements for basic safety and essential performance)
    IEC 60601-1-2 (Electromagnetic compatibility)
    IEC 60601-1-4 (Programmable electrical medical systems)
    IEC 60601-2-10 (Nerve and muscle stimulators)
    MaterialBiocompatibility (FDA cleared electrodes)
    Electrical Output Parameters (Demonstrated to be "similar" or "identical" to predicate devices)TENS: 0-120mA peak to peak FES: 0-120mA peak to peak Russian: 0-120mA peak to peak
    Frequency0.5-250 Hz
    Phase Duration50-500μs
    Current Output (at various loads)124 mA @ 500Ω, 110mA @ 2KΩ, 39.2mA @ 10KΩ (for Compact/II)
    Charge per Phase58μC
    Max. avg. power density0.038W/cm²
    Patient Leakage ControlNormal Condition: 0.0347 mA Single Fault Condition: 0.0162 mA

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The submission does not describe a test set involving patient data or clinical trials. The "testing results" mentioned are non-clinical and relate to engineering and electrical safety compliance. Therefore, there is no sample size for a test set or information on data provenance in the context of clinical performance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. The submission focuses on non-clinical engineering and safety standard compliance, not clinical interpretation or ground truth established by medical experts for a diagnostic or treatment outcome.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable, as there is no clinical test set requiring expert adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a powered muscle stimulator, not an AI-assisted diagnostic or imaging device. No MRMC study was performed or mentioned.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This device is a physical medical device, not an algorithm, and its performance inherently involves human-in-the-loop (e.g., a therapist or patient operating the device).

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" in this context is adherence to recognized international electrical safety and performance standards (IEC 60601 series). These standards define acceptable electrical characteristics, safety mechanisms, and electromagnetic compatibility. The device's performance is measured against the specifications of these standards and comparison to predicate devices, rather than clinical outcomes or expert consensus on patient data.

    8. The sample size for the training set

    Not applicable. There is no training set mentioned, as this is not an AI/machine learning device.

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K131629

    Validate with FDA (Live)

    Date Cleared
    2013-11-26

    (175 days)

    Product Code
    Regulation Number
    890.5850
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K031077, K121369, K021100

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Neurodyn Multiwave - Indications for Use:
    As a FES device:

    • Stimulation of the muscles in the leg and ankle of partially paralyzed patients to provide flexion of the foot and thus improve the patient's gait.
      As a TENS device:
    • Symptomatic relief of chronic (long term) intractable pain
    • Symptomatic relief of post-traumatic acute pain and post-surgical pain
      As an Interferential and Premodulated device:
    • Symptomatic relief of chronic intractable pain, acute post traumatic pain, or acute post traumatic surgical pain
      As a Russian device:
    • Temporary relaxation of muscle spasms
    • Prevention or retardation of disuse atrophy in post-injury type conditions
    • Increase local blood circulation
    • Muscle re-education
    • Maintaining or increasing range of motion
      As a Burst Modulated Alternating Current (Aussie) device:
    • Temporary relaxation of muscle spasms
    • Prevention or retardation of disuse atrophy in post-injury type conditions
    • Increase local blood circulation
    • Muscle re-education
    • Maintaining or increasing range of motion
    • Symptomatic relief of chronic intractable pain, acute post traumatic pain, or acute post traumatic surgical pain
      As a Microcurrent device:
    • Symptomatic relief of chronic intractable pain
    • Symptomatic relief of post-traumatic acute pain and post surgical pain
      As a DC/Polarized device:
    • Relaxation of Muscle Spasm

    Aussie Sport - Indications for Use:
    As an Burst Modulated Alternating Current (Aussie) device:

    • Temporary relaxation of muscle spasms
    • Prevention or retardation of disuse atrophy in post-injury type conditions
    • Increase local blood circulation
    • Muscle re-education
    • Maintaining or increasing range of motion
    • Symptomatic relief of chronic intractable pain, acute post traumatic pain, or acute post traumatic surgical pain
    Device Description

    Neurodyn Multiwave and Aussie Sport Neuromuscular Stimulators are intended for the treatment of, relief of chronic (long term) intractable pain as adjunctive treatment of post-surgical and post-traumatic acute pain. Both devices have the same intended uses and incorporate the same technologies as the following predicate devices: Vectra Genisys K031077, Neurodyn/Neurpdun Aussie K121369 and 400PV Complete K021100.

    The Neurodyn Multiwave Muscle Stimulator is a programmable device. It comes equipped with 5 preset clinical programs along with 10 user programs are adjustable and can be changed according to the patient's needs, doctor's recommendations and prescription settings.

    The Aussie Sport Muscle Stimulator has four output channels with independent intensity controls. Thus, four different areas can be stimulated separately or together during a therapy session. It is adjustable and can be changed according to the patient's needs, doctor's recommendations and prescription settings. It generates the medium frequency alternate current(MFAC), burst modulated alternating current (Aussie)- type of sinusoidal current with a frequency carrying 1,000 Hz or 4,000 Hz and a burst duration of 4 ms or 2 ms, modulated in pulse trains (bursts) with a variable frequency from 1 to 120Hz.

    AI/ML Overview

    The provided document is a 510(k) Summary for the Neurodyn Multiwave and Aussie Sport powered muscle stimulators. It focuses on demonstrating substantial equivalence to previously cleared predicate devices rather than presenting a study with specific acceptance criteria and performance metrics for the device itself.

    Therefore, the following information cannot be extracted from the document:

    1. A table of acceptance criteria and the reported device performance: The document does not define specific performance acceptance criteria for the Neurodyn Multiwave or Aussie Sport devices, nor does it report their performance against such criteria. Instead, it relies on demonstrating that the device characteristics are "identical" to predicate devices.
    2. Sample size used for the test set and the data provenance: No test set or associated data provenance is mentioned.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: No experts or ground truth establishment are discussed.
    4. Adjudication method for the test set: No adjudication method is mentioned as there is no test set described.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, an effect size of how much human readers improve with AI vs without AI assistance: No MRMC study or AI assistance is mentioned.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: No standalone algorithm performance is discussed.
    7. The type of ground truth used: No ground truth is specified.
    8. The sample size for the training set: No training set is mentioned.
    9. How the ground truth for the training set was established: No training set or ground truth establishment is discussed.

    Summary of Device Comparison (as provided in the document - not an acceptance criteria table):

    The document provides a "Device Comparison Table" to show substantial equivalence, not to present acceptance criteria and performance. The key points from this table are:

    Feature/CriterionNeurodyn Multiwave (K131629, New Device)Predicate Devices (Neurodyn K121369, 300 PV K021100, Vectra Genisys K031077) & Aussie Sport (K131629, New Device)
    Technological characteristicsIdenticalIdentical (Medium-frequency alternating current - MFAC)
    Device MaterialABS plastic panel, LCD displayABS plastic panel, LCD display
    Width (in)6.8Varied (6.8, 9.75, 1.26)
    Height (in)4.9Varied (4.9, 3.3, 8.75)
    Depth (in)12.4Varied (12.4, 12.75, 4.5)
    Number of Channels44
    Temperature range during transport and storage45°f-110°fVaried (45°f-110°f, -40 to 158°F, -40 to 150 F)
    Environment operating temperature range45°F-110°FVaried (45°F-110°F, 50° to 104°F, 45 to 105° F)
    PerformanceIdenticalIdentical
    BiocompatibilityFDA cleared electrodesFDA cleared electrodes
    Mechanical SafetyIdenticalIdentical
    Anatomical SitesIdenticalIdentical
    Russian CurrentYesYes (Neurodyn, Vectra Genisys); No (300 PV, Aussie Sport)
    Aussie CurrentYesYes (Neurodyn, Vectra Genisys, Aussie Sport)
    Interferential CurrentYesYes (Neurodyn, 300 PV, Vectra Genisys); No (Aussie Sport)
    MicrocurrentYesYes (Neurodyn, Vectra Genisys); No (300 PV, Aussie Sport)
    TENSYesYes (Neurodyn, 300 PV, Vectra Genisys); No (Aussie Sport)
    PremodulatedYesYes (Neurodyn, 300 PV, Vectra Genisys); No (Aussie Sport)
    FESYesYes (300 PV, Vectra Genisys); No (Neurodyn, Aussie Sport)
    DC/PolarizedYesYes (Vectra Genisys); No (Neurodyn, 300 PV, Aussie Sport)
    Voltage Input100/240V 50/60Hz BivoltVaried (100/240V 50/60Hz Bivolt, 3.0V DC, 100/240V 50/60Hz 1.0A)
    Output5A+17VVaried (5A+17V, 1.0A+3.0V DC, 7.3A+24V)
    Method of line current isolationDouble IsolationDouble Isolation
    Patient leakage control (normal condition)0.0508mAVaried (0.0508mA, 0.0502mA, 69μA)
    Patient leakage current (single fault condition)0.0252mAVaried (0.0252mA, 0.0248mA, 31μA)
    Software microprocessorYesYes
    Automatic overload tripNoNo
    Automatic shutoffNoNo
    Locking featureKeyboard lock safety featureKeyboard lock safety feature
    Treatment timer1 to 60 minutesVaried (1 to 60 minutes, 5 to 60 minutes)
    Auto test and repeatTreatment timer with auto shut offTreatment timer with auto shut off / Treatment timer
    Frequency Range50/60HzVaried (50/60Hz, ---)
    Maximum Current Density2.0 mA2.0 mA

    Conclusion: The 510(k) summary concludes that the Neurodyn Muscle Stimulators are substantially equivalent to the predicate devices based on similarities in indications for use, technological design, materials, and various performance characteristics. The argument made is that because the new devices are "identical" or sufficiently similar to cleared predicate devices, they do not raise new issues of safety or effectiveness. No specific study demonstrating the new device's performance against defined acceptance criteria is presented.

    Ask a Question

    Ask a specific question about this device

    K Number
    K110350

    Validate with FDA (Live)

    Device Name
    KNEEHAB XP
    Date Cleared
    2011-08-03

    (177 days)

    Product Code
    Regulation Number
    882.5890
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K083105, K082011, K061516, K021100, K971437

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Kneehab XP, Type 412/421, Indications for use are as follows:

    In NMES mode (Programs 1-6) the Kneehab XP is intended to:

    • Maintain or increase the range of motion.
    • Prevention or retardation of disuse atrophy
    • Re-educate muscles
    • Early post-surgical quadriceps strengthening and improved post surgical knee stability secondary to quadriceps strengthening
    • Relax muscle spasms
    • Increase blood circulation

    In TENS Mode (Programs 7 - 9) the Kneehab XP is intended to:

    • Provide symptomatic relief and management of chronic, intractable pain
    • Provide an adjunctive treatment in the management of acute, post-surgical or post-traumatic pain
    • Provide symptomatic relief and management of intractable pain and relief of pain associated with arthritis
    • Provide an adjunctive therapy in reducing the level of pain and symptoms associated with osteoarthritis of the knee
    Device Description

    The Kneehab XP, Type 412/421 is a portable, battery operated, combination device which can provide both neuromuscular electronic stimulation (NMES) and transcutaneous electrical nerve stimulation (TENS). The device incorporates multipath®, a patented technology developed by neurotech® which enables the Kneehab XP Conductive Garment to deliver highly focused and accurate NMES muscle contractions. This device also provides a method of pain management and relief through the use of TENS technology.

    The Kneehab XP pack consists of a rechargeable control unit, a left or right universally sized garment, a pack of custom adhesive electrodes, a battery charger and instructions for use. The garment is fastened around the thigh and above the kneecap and contains a connector socket into which the control unit is plugged. Power is derived from a 3.6V NiMH rechargeable battery pack that is pre-installed in the unit. A battery charger is included with the device and the device cannot be used while being charged. The adhesive electrodes have an estimated usage capability of 20 sessions when used under the recommended conditions of use.

    All internal connections of the unit are over molded to prevent moisture ingress. The user has no access to the wiring or connectors within the garment and is unable to alter the current path. There are nine treatment programs in total (six NMES and three TENS) with a duration of 20 minutes each. Program details are included in the instructions for use. For purposes of hygiene, the garment may be cleaned and instructions for device care are included in the user manual.

    AI/ML Overview

    The provided document is a 510(k) Summary for the Kneehab XP, Type 412/421. This type of submission is for demonstrating substantial equivalence to a legally marketed predicate device, rather than providing clinical study data to meet specific performance acceptance criteria. Therefore, the document does not describe a study that proves the device meets specific acceptance criteria.

    Instead, it focuses on comparing the technological characteristics and intended uses of the Kneehab XP to existing predicate devices to argue for substantial equivalence. The document primarily highlights technical and functional comparisons rather than performance metrics from a clinical trial.

    Based on the provided document, I cannot fulfill all sections of your request as it does not contain the information regarding an acceptance criteria study.

    However, I can extract the following relevant information regarding the device and its intended use, which would typically form the basis for performance evaluation in a clinical study:

    1. A table of acceptance criteria and the reported device performance
    The document does not provide a table of acceptance criteria or reported device performance from a study. It only provides electrical output specifications.

    Device Electrical Output Parameters (Technical Characteristics, not performance data):

    ParameterKneehab XP (Proposed Device)
    Frequency/Phase Duration of program with highest output power (Program 7)99Hz / 300 µS
    Baseline to Peak Current80mA @ 500Ω, 28mA @ 2kΩ, 3.9mA @ 10kΩ
    Baseline to Peak Output Voltage40.0V @ 500Ω, 55.6V @ 2kΩ, 39.3V @ 10kΩ
    Maximum RMS Output Voltage (+/-10%) Vrms9.3 V @ 500 Ω, 17.1V @ 2kΩ, 14.1V @ 10kΩ
    Maximum RMS Output Current (+/-10%) Irms18.6 mA@ 500 Ω, 8.6 mA@ 2kΩ, 1420 µA@ 10kΩ
    Pulse Width640 µs sum of both phases (300µs + 40µs interphase interval)
    Net Charge (µC per pulse)0µC @ 500Ω
    Maximum Phase Charge @500 (+/-20%)24 µC @ 500Ω
    Maximum Current Density @500 Ω0.22 mA/cm² (using smallest electrode conductive surface area, 83 cm²)
    Maximum Power Density @500 Ω (Program 7)2.1mW/cm² @ 500Ω

    Intended Use and Indications:

    NMES Mode (Programs 1-6):

    • Maintain or increase the range of motion.
    • Prevention or retardation of disuse atrophy.
    • Re-educate muscles.
    • Early post-surgical quadriceps strengthening and improved post-surgical knee stability secondary to quadriceps strengthening.
    • Relax muscle spasms.
    • Increase blood circulation.
      TENS Mode (Programs 7-9):
    • Provide symptomatic relief and management of chronic, intractable pain.
    • Provide an adjunctive treatment in the management of acute, post-surgical or post-traumatic pain.
    • Provide symptomatic relief and management of intractable pain and relief of pain associated with arthritis.
    • Provide an adjunctive therapy in reducing the level of pain and symptoms associated with osteoarthritis of the knee.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
    Not applicable, as no performance study data is presented in this 510(k) summary.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
    Not applicable, as no performance study data is presented in this 510(k) summary.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
    Not applicable, as no performance study data is presented in this 510(k) summary.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
    Not applicable. This is a medical device application for a physical device (electrical stimulator), not an AI/imaging diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
    Not applicable. This is a medical device application for a physical device (electrical stimulator), not an algorithm-only device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
    Not applicable, as no performance study data is presented in this 510(k) summary.

    8. The sample size for the training set
    Not applicable, as no performance study data is presented in this 510(k) summary.

    9. How the ground truth for the training set was established
    Not applicable, as no performance study data is presented in this 510(k) summary.

    Summary of Device and 510(k) Submission:

    The Kneehab XP is a portable, battery-operated combination device that provides both Neuromuscular Electrical Stimulation (NMES) and Transcutaneous Electrical Nerve Stimulation (TENS). Its primary purpose within this 510(k) submission is to demonstrate substantial equivalence to previously cleared predicate devices (Kneehab XP Conductive Garment, MediStim XP, MediTens XP, Staodyn Max Preset, 300 PV Complete Electrotherapy System, Bioncicare® Stimulator System). The submission details the technological characteristics, intended use, and indications for use of the Kneehab XP and compares them to these predicate devices, ensuring that there are "no new technological characteristics that could affect safety or effectiveness." The focus is on meeting international safety standards (IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10) and demonstrating similarity in electrical parameters and functionality to the predicate devices.